Thin laser light sheet microscope for microbial oceanography

Open Access Open Access

Abstract

Despite a growing need, oceanographers are limited by existing technological constrains and are unable to observe aquatic microbes in their natural setting. In order to provide a simple and easy to implement solution for such studies, a new Thin Light Sheet Microscope (TLSM) has been developed. The TLSM utilizes a well-defined sheet of laser light, which has a narrow (23 micron) axial dimension over a 1 mm × 1 mm field of view. This light sheet is positioned precisely within the depth of field of the microscope’s objective lens. The technique thus utilizes conventional microscope optics but replaces the illumination system. The advantages of the TLSM are two-fold: First, it concentrates light only where excitation is needed, thus maximizing the efficiency of the illumination source. Secondly, the TLSM maximizes image sharpness while at the same time minimizing the level of background noise. Particles that are not located within the objective’s depth of field are not illuminated and therefore do not contribute to an out-of-focus image. Images from a prototype system that used SYBR Green I fluorescence stain in order to localize single bacteria are reported. The bacteria were in a relatively large and undisturbed volume of 4ml, which contained natural seawater. The TLSM can be used for fresh water studies of bacteria with no modification. The microscope permits the observation of interactions at the microscale and has potential to yield insights into how microbes structure pelagic ecosystems.

©2002 Optical Society of America

Full Article  |  PDF Article
More Like This
Structured light sheet fluorescence microscopy based on four beam interference

Ming Lei and Andreas Zumbusch
Opt. Express 18(18) 19232-19241 (2010)

Adaptive illumination based on direct wavefront sensing in a light-sheet fluorescence microscope

Dean Wilding, Paolo Pozzi, Oleg Soloviev, Gleb Vdovin, and Michel Verhaegen
Opt. Express 24(22) 24896-24906 (2016)

Spatio-temporal performance in an incoherent holography lattice light-sheet microscope (IHLLS)

Mariana Potcoava, Christopher Mann, Jonathan Art, and Simon Alford
Opt. Express 29(15) 23888-23901 (2021)

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (4)

Figure 1:
Figure 1: Thin light sheet microscopy (TLSM), schematic diagram.
Figure 2:
Figure 2: Thin light sheet - beam characteristics.
Figure 3
Figure 3 (a): Sample illuminated with the thin light sheet illumination. (b) Same sample with a broad beam illumination. Marine bacterial isolates in 0.22 μm filtered seawater, stained. Particle size cannot be scaled due to sensor blooming. Particles appear about 1.7 times larger than their actual size.
Figure 4:
Figure 4: Different size organisms imaged with the TLSM of total seawater. Particles appear about 1.7 times larger than their actual size.

Equations (2)

Equations on this page are rendered with MathJax. Learn more.

δ = n λ 0 N A 2 + n e NA M