A frequency domain near-infrared spectroscopy oximeter using high-speed, direct analog to digital conversion

Bernhard Zimmermann, Juliette Selb, Stefan Carp, Qianqian Fang, Joe Stadtmiller, Robert Dewsnap, Ron Altman, and David Boas  »View Author Affiliations


http://dx.doi.org/10.1364/BIOMED.2012.BSu3A.78


View Full Text Article

Acrobat PDF (41 KB) Note that full-text PDFs from conferences typically contain 1-3 pages of content, some or all of which might be an abstract, summary, or miscellaneous items.





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations
  • Export Citation/Save Click for help

Abstract

A frequency-encoded frequency domain near-infrared spectroscopy (FD-NIRS) oximeter with two wavelengths has been constructed. The system digitizes the 67.5MHz and 75MHz waveforms using a high speed (180MSPS) 16-bit analog to digital converter.

© 2012 OSA

OCIS Codes
(170.0110) Medical optics and biotechnology : Imaging systems
(170.3660) Medical optics and biotechnology : Light propagation in tissues
(170.3890) Medical optics and biotechnology : Medical optics instrumentation

Citation
B. Zimmermann, J. Selb, S. Carp, Q. Fang, J. Stadtmiller, R. Dewsnap, R. Altman, and D. Boas, "A frequency domain near-infrared spectroscopy oximeter using high-speed, direct analog to digital conversion," in Biomedical Optics and 3-D Imaging, OSA Technical Digest (Optical Society of America, 2012), paper BSu3A.78.