A digital phasemeter for precision length measurements

Malcolm Gray, Terry McRae, Magnus Hsu, Jan Herrmann, and Daniel Shaddock  »View Author Affiliations


http://dx.doi.org/10.1364/CLEO_AT.2012.JW1C.4


View Full Text Article

Acrobat PDF (1508 KB) Note that full-text PDFs from conferences typically contain 1-3 pages of content, some or all of which might be an abstract, summary, or miscellaneous items.





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations
  • Export Citation/Save Click for help

Abstract

We report on a digital phasemeter for application in precision length measurement using laser interferometry. Our interferometer demonstrates a length sensitivity of 0.5pm/√Hz above 5Hz, and a cyclic error of better than 50pm.

© 2012 OSA

OCIS Codes
(120.3180) Instrumentation, measurement, and metrology : Interferometry
(120.5050) Instrumentation, measurement, and metrology : Phase measurement