Properties of indium antimonide detectors for use as transfer standards for detector calibrations

Louis-Philippe Boivin  »View Author Affiliations


Applied Optics, Vol. 37, Issue 10, pp. 1924-1929 (1998)
http://dx.doi.org/10.1364/AO.37.001924


View Full Text Article

Enhanced HTML    Acrobat PDF (190 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

The properties of 4-mm-diameter and 7-mm-diameter liquid-nitrogen-cooled indium antimonide detectors were investigated to determine their suitability as transfer standards (i.e., reference detectors) for the detector calibration facility at the National Research Council of Canada. The spectral responsivity of the InSb detectors in the spectral range 1000–3000 nm was determined by a two-step procedure involving the use of germanium transfer standards and suitably characterized thermopiles. It is shown that the long-term reproducibility of the InSb detector calibrations is approximately ±1%. The uniformity and linearity as a function of wavelength, as well as the background current noise and drift of the two detectors, are compared. It is shown that the 7-mm-diameter detector is clearly superior to the 4-mm one for use as a transfer standard. It is estimated that the overall accuracy of the calibrations of the two InSb detectors is ±2% in the range 1200–2800 nm.

© 1998 Optical Society of America

OCIS Codes
(040.0040) Detectors : Detectors
(040.5160) Detectors : Photodetectors
(120.5630) Instrumentation, measurement, and metrology : Radiometry

History
Original Manuscript: July 11, 1997
Published: April 1, 1998

Citation
Louis-Philippe Boivin, "Properties of indium antimonide detectors for use as transfer standards for detector calibrations," Appl. Opt. 37, 1924-1929 (1998)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-37-10-1924


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. L. P. Boivin, K. Gibb, “Monochromator-based cryogenic radiometry at the NRC,” Metrologia. 32, 565–570 (1995/1996). [CrossRef]
  2. R. Köhler, R. Goebel, R. Pello, “Results of an international comparison of spectral responsivity of silicon photodetectors,” Metrologia. 32, 463–468 (1995/1996). [CrossRef]
  3. L. P. Boivin, A. A. Gaertner, D. S. Gignac, “Realization of the New Candela (1979) at NRC,” Metrologia. 24, 139–152 (1987). [CrossRef]
  4. J. Campos, A. Corrons, A. Pons, R. Corredera, “Realization of the candela from a partial filtering V(λ) detector traceable to a cryogenic radiometer,” Metrologia. 32, 675–679 (1995/1996). [CrossRef]
  5. C. L. Cromer, G. Eppeldauer, J. E. Hardis, T. C. Larason, Y. Ohno, A. C. Parr, “The NIST detector-based luminous intensity scale,” J. Res. Natl. Inst. Stand. Technol. 101, 109–132 (1996). [CrossRef]
  6. W. S. Cleveland, “Robust locally weighted regression and smoothing scatterplots,” J. Am. Statist. Assoc. 74, 829–836 (1979). [CrossRef]
  7. L. P. Boivin, “Automated absolute and relative spectral linearity measurements on photovoltaic detectors,” Metrologia. 30, 355–360 (1993). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited