OSA's Digital Library

Applied Optics

Applied Optics

APPLICATIONS-CENTERED RESEARCH IN OPTICS

  • Vol. 41, Iss. 10 — Apr. 1, 2002
  • pp: 1929–1935

Study of bandwidth effects in monochromator-based spectral responsivity measurements

Louis-Philippe Boivin  »View Author Affiliations


Applied Optics, Vol. 41, Issue 10, pp. 1929-1935 (2002)
http://dx.doi.org/10.1364/AO.41.001929


View Full Text Article

Enhanced HTML    Acrobat PDF (144 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

The errors caused by monochromator bandwidth in spectral responsivity measurements with a monochromator-based apparatus are discussed. Bandwidth effects are not negligible in high-accuracy cryogenic radiometer-based calibrations. A simple numerical method is used to calculate bandwidth effects for different types of detectors, monochromator slit scattering functions, and monochromator output spectral distributions. The method uses low-order Lagrange polynomials fitted segmentwise to measured spectral responsivity and monochromator spectral distribution data in order to make the calculations. It is shown that the shape of the slit function has only a small influence on the bandwidth errors, whereas the output spectral distribution of the monochromator can strongly affect bandwidth errors. It is also shown that in most cases the magnitude of bandwidth effects will vary as the square of the bandwidth. Bandwidth error calculations are presented for various types of detectors (silicon, silicon trap, germanium, InGaAs), for a V(λ) detector, and for a typical filter radiometer. A comparison is made between calculated and measured bandwidth effects to validate the method used. In general, calculations of bandwidth effects will be mostly useful for determining uncertainties associated with monochromator bandwidth in spectral responsivity measurements; however, in certain cases the calculations can be used to apply corrections for such effects.

© 2002 Optical Society of America

OCIS Codes
(040.6040) Detectors : Silicon
(120.5630) Instrumentation, measurement, and metrology : Radiometry

History
Original Manuscript: August 6, 2001
Revised Manuscript: October 12, 2001
Published: April 1, 2002

Citation
Louis-Philippe Boivin, "Study of bandwidth effects in monochromator-based spectral responsivity measurements," Appl. Opt. 41, 1929-1935 (2002)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-41-10-1929


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. E. F. Zalewski, C. R. Duda, “Silicon photodiode device with 100% external quantum efficiency,” Appl. Opt. 22, 2867–2873 (1983). [CrossRef] [PubMed]
  2. J. L. Gardner, “Transmission trap detectors,” Appl. Opt. 33, 5914–5918 (1994). [CrossRef] [PubMed]
  3. T. R. Gentile, J. M. Houston, J. E. Hardis, C. L. Cromer, A. C. Parr, “National Institute of Standards and Technology high-accuracy cryogenic radiometer,” Appl. Opt. 35, 1056–1068 (1996). [CrossRef] [PubMed]
  4. K. D. Stock, H. Hofer, “PTB primary standard for optical radiant power: transfer-optimized facility in the clean-room centre,” Metrologia 32, 545–549 (1996). [CrossRef]
  5. L. P. Boivin, K. Gibb, “Monochromator-based cryogenic radiometry at the NRC,” Metrologia 32, 565–570 (1995/1996). [CrossRef]
  6. C. A. Schrama, R. Bosma, K. Gibb, H. Reijn, P. Bloembergen, “Comparison of monochromator-based and laser-based cryogenic radiometry,” Metrologia 35, 431–435 (1998). [CrossRef]
  7. R. Nemecek, P. Nemecek, “Spectral bandwidth correction by responsivity measurement,” in 11th Slovak-Czech-Polish Optical Conference on Wave and Quantum Aspects of Contemporary Optics, M. Hrabovský, A. Štrba, W. Urbanczyk, eds., Proc. SPIE3820, 322–328 (1999). [CrossRef]
  8. A. Corrons, E. F. Zalewski, “Detector spectral response from 350 to 1200 nm using a monochromator based spectral comparator,” Natl. Bur. Stand. Tech. Note 988 (National Bureau of Standards, Gaithersburg, Md., 1978).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited