Characterization of an absolute cryogenic radiometer as a standard detector for radiant-power measurements

R. U. Datla, K. Stock, A. C. Parr, C. C. Hoyt, P. J. Miller, and P. V. Foukal  »View Author Affiliations


Applied Optics, Vol. 31, Issue 34, pp. 7219-7225 (1992)
http://dx.doi.org/10.1364/AO.31.007219


View Full Text Article

Enhanced HTML    Acrobat PDF (871 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

An active cavity radiometer of the electrical substitution type with a cone receiver that operates at 2–4 K has been developed for measuring radiant fluxes in the dynamic range of 20 nW to 100 μW within an uncertainty of ±1% (2σ level). It is a broadband absolute detector with a flat overall absorption efficiency that is > 99% for radiation from the visible to long-wavelength IR. The system is designed based on thermal modeling and experimental measurements of concepts. It has been installed in the cryogenic chamber for low-background infrared radiation calibrations at the National Institute of Standards and Technology (NIST) for testing cryogenic blackbody sources, detectors, and optical components. Its time constant, responsivity, and nonequivalence error have been measured. They are in agreement with design predictions. Radiant power measurements of an amplitude-stabilized He–Ne laser beam with the radiometer and an industry standard photodiode detector, QED-200, have been intercompared and found to be in agreement. The intercomparison ratio of the measurements with the absolute cryogenic radiometer and QED-200 was 1.004 in the 75–100-μW range with an uncertainty of 0.5% (the 3σ level).

© 1992 Optical Society of America

History
Original Manuscript: January 27, 1992
Published: December 1, 1992

Citation
R. U. Datla, K. Stock, A. C. Parr, C. C. Hoyt, P. J. Miller, and P. V. Foukal, "Characterization of an absolute cryogenic radiometer as a standard detector for radiant-power measurements," Appl. Opt. 31, 7219-7225 (1992)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-31-34-7219


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. F. Hengstberger, “The absolute measurement of radiant power,” in Absolute Radiometry: Electrically Calibrated Thermal Detectors of Optical Radiation, F. Hengstberger, ed. (Academic, San Diego, Calif., 1989), pp. 1–117.
  2. P. V. Foukal, C. Hoyt, H. Kochling, P. Miller, “Cryogenic absolute radiometers as laboratory irradiance standards, remote sensing detectors, and pyroheliometers,” Appl. Opt. 29, 988–993 (1990). [CrossRef] [PubMed]
  3. C. R. Yokely, “Long wave infrared testing at NBS,” in Applications of Optical Metrology: Techniques and Measurements II, J. J. Lee, ed., Proc. Soc. Photo-Opt. Instrum. Eng.416, 2–7 (1983).
  4. A. C. Parr, J. Fowler, S. Ebner, “Low background infrared calibration facility at the National Bureau of Standards,” in Infrared Scene Simulation: Systems, Requirements, Calibration, Devices, and Modeling, R. B. Johnson, M. J. Triplett, eds., Proc. Soc. Photo-Opt. Instrum. Eng.940, 26–33 (1988).
  5. S. C. Ebner, A. C. Parr, C. C. Hoyt, “Update on the low background IR calibration facility at the National Institute of Standards and Technology,” in Imaging Infrared: Scene Simulations, Modeling, and Real Image Tracking, A. J. Huber, M. J. Triplett, J. R. Wolverton, eds., Proc. Soc. Photo-Opt. Instrum. Eng.1110, 49–60 (1989).
  6. E. F. Zalewski, C. R. Duda, “Silicon photodiode device with 100% external quantum efficiency,” Appl. Opt. 22, 2867–2873 (1983). [CrossRef] [PubMed]
  7. A. F. Clark, “Thermal expansion,” in Materials at Low Temperatures, P. R. Reed, A. F. Clark, eds. (American Society for Metals, Metals Park, Ohio, 1983), pp. 75–128.
  8. Product of Lord Corporation, Industrial Coatings Division, 2000 West Grandview Boulevard, Erie, Pa. 16514–0038.
  9. Laser Power Controller system, Models LPC-VIS and RD-40, manufactured by Cambridge Research and Instrumentation, Inc., 21 Erie Street, Cambridge, Mass. 02139.

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited