Properties of Indium Antimonide Detectors for use as Transfer Standards for Detector Calibrations
Applied Optics, Vol. 37, Issue 10, pp. 1924-1929 (1998)
http://dx.doi.org/10.1364/AO.37.001924
Acrobat PDF (190 KB)
Abstract
The properties of 4-mm-diameter and 7-mm-diameter liquid-nitrogen-cooled indium antimonide detectors were investigated to determine their suitability as transfer standards (i.e., reference detectors) for the detector calibration facility at the National Research Council of Canada. The spectral responsivity of the InSb detectors in the spectral range 1000–3000 nm was determined by atwo-step procedure involving the use of germanium transfer standardsand suitably characterized thermopiles. It is shown that thelong-term reproducibility of the InSb detector calibrations is approximately ∓1%. The uniformity and linearity as a function of wavelength, as well as the background current noise and drift of thetwo detectors, are compared. It is shown that the 7-mm-diameter detector is clearly superior to the 4-mm one for use as a transfer standard. It is estimated that the overall accuracy of the calibrations of the two InSb detectors is ∓2% in the range 1200–2800 nm.
© 1998 Optical Society of America
[Optical Society of America ]
OCIS Codes
(040.0040) Detectors : Detectors
(040.5160) Detectors : Photodetectors
(120.5630) Instrumentation, measurement, and metrology : Radiometry
Citation
Louis-Philippe Boivin, "Properties of Indium Antimonide Detectors for use as Transfer Standards for Detector Calibrations," Appl. Opt. 37, 1924-1929 (1998)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-37-10-1924
You do not have subscription access to this journal. Citation lists with outbound citation links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription
You do not have subscription access to this journal. Article level metrics are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription





OSA is a member of 