Improvements in the Accuracy and the Repeatability of Long Trace Profiler Measurements
Applied Optics, Vol. 38, Issue 25, pp. 5468-5479 (1999)
http://dx.doi.org/10.1364/AO.38.005468
Acrobat PDF (216 KB)
Abstract
Modifications of the long trace profiler at the Advanced Photon Source at Argonne National Laboratory have significantly improved its accuracy and repeatability for measuring the figure of large flat and long-radius mirrors. Use of a Dove prism in the reference beam path corrects phasing problems between mechanical errors and thermally induced system errors. A single reference correction now completely removes both of these error signals from the measured surface profile. The addition of a precision air conditioner keeps the temperature in the metrology enclosure constant to within ∓0.1 °C over a 24-h period and has significantly improved the stability and the repeatability of the measurements. Long-radius surface curvatures can now be measured absolutely with a high degree of confidence. These improved capabilities are illustrated with a series of measurements of a 500-mm-long mirror with a 5-km radius of curvature. The standard deviation in the average of ten slope profile scans is 0.3 μrad, and the corresponding standard deviation in the height error is 4.6 nm.
© 1999 Optical Society of America
[Optical Society of America ]
OCIS Codes
(120.3940) Instrumentation, measurement, and metrology : Metrology
(120.6650) Instrumentation, measurement, and metrology : Surface measurements, figure
(340.6720) X-ray optics : Synchrotron radiation
(340.7470) X-ray optics : X-ray mirrors
Citation
Peter Z. Takacs, Eugene L. Church, Cynthia J. Bresloff, and Lahsen Assoufid, "Improvements in the Accuracy and the Repeatability of Long Trace Profiler Measurements," Appl. Opt. 38, 5468-5479 (1999)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-38-25-5468
You do not have subscription access to this journal. Citation lists with outbound citation links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription
You do not have subscription access to this journal. Article level metrics are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription





OSA is a member of 