OSA's Digital Library

Applied Optics

Applied Optics

APPLICATIONS-CENTERED RESEARCH IN OPTICS

  • Vol. 39, Iss. 34 — Dec. 1, 2000
  • pp: 6390–6395

Optimizing precision of fixed-polarizer, rotating-polarizer, sample, and fixed-analyzer spectroscopic ellipsometry

Zhiming Huang and Junhao Chu  »View Author Affiliations


Applied Optics, Vol. 39, Issue 34, pp. 6390-6395 (2000)
http://dx.doi.org/10.1364/AO.39.006390


View Full Text Article

Enhanced HTML    Acrobat PDF (119 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

General equations describing the theoretical uncertainty in measurements of the complex reflectance ratio, ρ, are given for fixed-polarizer, rotating-polarizer, sample, and fixed-analyzer (PPrSA) ellipsometers operating under shot-noise-limited (SNL) and detector-noise-limited (DNL) conditions. Optimum values of the fixed-polarizer and fixed-analyzer azimuths are calculated as a function of ellipsometric parameters (ψ and Δ) for SNL and DNL systems. The uncertainty of complex reflectance ratio, calculated from the optimum values of the fixed polarizer and analyzer azimuths, decreases linearly in ψ as ψ → 0 under the SNL condition but approaches arbitrary large under the DNL condition. This suggests that a compensator is not needed under the SNL condition to achieve high precision on dielectric surfaces but that measurements at small ψ values should be avoided under the DNL condition. A discrepancy of the fixed polarizer or analyzer azimuth from its optimum value for various Δ is discussed with ψ = 5° and 45° under the SNL condition.

© 2000 Optical Society of America

OCIS Codes
(120.0120) Instrumentation, measurement, and metrology : Instrumentation, measurement, and metrology
(120.2130) Instrumentation, measurement, and metrology : Ellipsometry and polarimetry

History
Original Manuscript: May 15, 2000
Revised Manuscript: August 14, 2000
Published: December 1, 2000

Citation
Zhiming Huang and Junhao Chu, "Optimizing precision of fixed-polarizer, rotating-polarizer, sample, and fixed-analyzer spectroscopic ellipsometry," Appl. Opt. 39, 6390-6395 (2000)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-39-34-6390


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. B. D. Cahan, R. F. Spanier, “A high-speed automatic ellipsometer,” Surf. Sci. 16, 166–176 (1969). [CrossRef]
  2. P. S. Hauge, F. H. Dill, “Design and operation of ETA, an automated ellipsometer,” IBM J. Res. Dev. 17, 472–489 (1973). [CrossRef]
  3. D. E. Aspnes, “Optimizing precision of rotating-analyzer ellipsometers,” J. Opt. Soc. Am. 64, 639–646 (1974). [CrossRef]
  4. L. Y. Chen, D. W. Lynch, “Scanning ellipsometer by rotating polarizer and analyzer,” Appl. Opt. 26, 5221–5228 (1987). [CrossRef] [PubMed]
  5. L. Y. Chen, X. W. Feng, Y. Su, H. Z. Ma, Y. H. Qian, “Design of a scanning ellipsometer by synchronous rotation of the polarizer and analyzer,” Appl. Opt. 33, 1299–1305 (1994). [CrossRef] [PubMed]
  6. R. M. A. Azzam, “A simple Fourier photopolarimeter with rotating polarizer and analyzer for measuring Jones and Mueller matrices,” Opt. Commun. 25, 137–140 (1978). [CrossRef]
  7. P. S. Hauge, “Recent developments in instrumentation in ellipsometry,” Surf. Sci. 96, 108–140 (1980). [CrossRef]
  8. S. H. Russev, “Correction for nonlinearity and polarization-dependent sensitivity in the detection system of analyzer ellipsometers,” Appl. Opt. 28, 1504–1507 (1989). [CrossRef] [PubMed]
  9. R. W. Collins, “Automatic rotating element ellipsometers: calibration, operation, and real-time application,” Rev. Sci. Instrum. 61, 2029–2062 (1990). [CrossRef]
  10. R. W. Stobie, B. Rao, M. J. Dignam, “Automatic ellipsometer with high sensitivity and special advanges for infrared spectroscopy of adsorbed species,” Appl. Opt. 14, 999–1003 (1975); R. W. Stobie, B. Rao, M. J. Dignam, “Analysis of a novel ellipsometric technique with special advantages for infrared spectroscopy,” J. Opt. Soc. Am. 65, 25–28 (1975). [CrossRef] [PubMed]
  11. S. Bertucci, A. Pawlowski, N. Nicolas, L. Johann, A. El Ghemmaz, N. Stein, R. Kleim, “Systemic errors in fixed polarizer, rotating polarizer, sample, fixed analyzer spectroscopic ellipsometry,” Thin Solid Films 313–314, 73–78 (1998).
  12. R. J. Archer, Manual on Ellipsometry (Gaertner, Chicago, 1968).
  13. E. Schmidt, “Precision of ellipsometric measurement,” J. Opt. Soc. Am. 60, 490–494 (1970). [CrossRef]
  14. R. M. A. Azzam, N. M. Bashara, “Choice of compensator azimuth and position in ellipsometry,” J. Opt. Soc. Am. 62, 700–701 (1972). [CrossRef]
  15. R. M. A. Azzam, N. M. Bashara, Ellipsometry and Polarized Light (North-Holland, Amsterdam, 1977).
  16. O. S. Heavens, Optical Properties of Thin Solid Films (Butterworths, London, 1955).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited