Absolute linearity characterization of lock-in amplifiers

E. Theocharous  »View Author Affiliations


Applied Optics, Vol. 47, Issue 8, pp. 1090-1096 (2008)
http://dx.doi.org/10.1364/AO.47.001090


View Full Text Article

Enhanced HTML    Acrobat PDF (815 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

In applications where low signal-to-noise ratios are encountered, the use of phase-sensitive detection (PSD) is highly advantageous and is applied widely. However, the characteristics of lock-in amplifiers that utilize the PSD technique, such as the linearity of the output and the gain of these instruments, have not been extensively evaluated. The author proposes a method for measuring the linearity characteristics of lock-in amplifiers and describes how this method was used to measure the linearity factor of three nominally identical commercially available digital lock-in amplifiers. The results of this study show that the linearity factor of these lock-in amplifiers can deviate from unity by more than 0.1% over a factor of 2 change in output. Moreover, the linearity characteristics of these lock-in amplifiers vary from one instrument to the next. The linearity characteristics were shown to be independent of the reference frequency, the time constant, and the temporal profile of the signal being analyzed. However, they were found to be dependent on the sensitivity settings of these instruments. The linearity characteristics of these instruments were observed not to change with time. This implies that, where the uncertainty contribution due to the nonlinearity of a lock-in amplifier is significant, the nonlinearity can be evaluated by using the method described in this article and can be used to apply corrections.

© 2008 Optical Society of America

OCIS Codes
(000.2170) General : Equipment and techniques
(120.1880) Instrumentation, measurement, and metrology : Detection
(120.3930) Instrumentation, measurement, and metrology : Metrological instrumentation

ToC Category:
Instrumentation, Measurement, and Metrology

History
Original Manuscript: December 4, 2007
Revised Manuscript: January 23, 2008
Manuscript Accepted: January 24, 2008
Published: March 7, 2008

Citation
E. Theocharous, "Absolute linearity characterization of lock-in amplifiers," Appl. Opt. 47, 1090-1096 (2008)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-47-8-1090


Sort:  Year  |  Journal  |  Reset  

References

  1. E. Theocharous, “The establishment of the NPL infrared relative spectral responsivity scale using cavity pyroelectric detectors,” Metrologia 43, S115-S119, (2006). [CrossRef]
  2. D. P. Blair and P. H. Sydenham, “Phase sensitive detection as a means to recover signals buried in noise,” J. Phys. E 8, 621-627, (1975). [CrossRef]
  3. E. Theocharous, J. Ishii, and N. P. Fox, “Absolute linearity measurements on HgCdTe detectors in the infrared,” Appl. Opt. 43, 4182-4188 (2004). [CrossRef] [PubMed]
  4. C. L. Sanders, “Accurate measurements of and corrections for nonlinearities in radiometers,” J. Res. Nat. Bur. Stand. Sect. A 76, 437-453 (1972).
  5. The output of digital lock-in amplifiers represents the root mean square value of the fundamental (reference) frequency component of the electrical signal being analyzed; so the output of a lock-in amplifier depends on the temporal profile of the electrical signal being analyzed. The sensitivity settings are calibrated for input signals of a square-wave temporal profile. The effective gain defines the lock-in amplifier sensitivity for that particular temporal profile of the signal being analyzed.
  6. L. P. Boivin, “Automated absolute and relative spectral linearity measurements on photovoltaic detectors,” Metrologia 30, 355-360 (1993). [CrossRef]
  7. E. Theocharous, “Reply to comments on 'Absolute linearity measurements on a PbS detector in the infrared',” Appl. Opt. 46, 6495-6497 (2007). [CrossRef] [PubMed]
  8. E. Theocharous, “Absolute linearity measurements on PbS detectors in the infrared,” Appl. Opt. 45, 2381-2386(2006). [CrossRef] [PubMed]
  9. E. Theocharous, “Absolute linearity measurements on a PbSe detector in the infrared,” Infrared Phys. Technol. 50, 63-69 (2007). [CrossRef]
  10. W. Budde, “Physical detectors of radiation,” in Optical Radiation Measurements (Academic Press, 1983), Vol. 4, pp. 70-82.
  11. I. Vayshenkev, S. Yang, and R. Swafford, “Nonlinearity of high-power optical power meters at 1480 nm,” Appl. Opt. 45, 1098-1101 (2006). [CrossRef]
  12. E. Theocharous, “How linear is the response of your detector?,” Europhoton. Mag. , February 2007, pp. 33-35.
  13. E. Theocharous, N. P. Fox, V. I. Sapritsky, S. N. Mekhontsev, and S. P. Morozova, “Absolute measurements of black-body emitted radiance,” Metrologia 35, 549-554 (1998). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited