Abstract
A confocal pore size measurement based on super-resolution image restoration is proposed to obtain a fast and accurate measurement for submicrometer pore size of nuclear track-etched membranes (NTEMs). This method facilitates the online inspection of the pore size evolution during etching. Combining confocal microscopy with super-resolution image restoration significantly improves the lateral resolution of the NTEM image, yields a reasonable circle edge-setting criterion of 0.2408, and achieves precise pore edge detection. Theoretical analysis shows that the minimum measuring diameter can reach 0.19 μm, and the root mean square of the residuals is only 1.4 nm. Edge response simulation and experiment reveal that the edge response of the proposed method is better than 80 nm. The NTEM pore size measurement results obtained by the proposed method agree well with that obtained by scanning electron microscopy.
© 2014 Optical Society of America
Full Article | PDF ArticleMore Like This
Han Cui, Weiqian Zhao, Yun Wang, Ying Fan, Lirong Qiu, and Ke Zhu
Opt. Express 24(10) 10767-10776 (2016)
Fang Tang, Yun Wang, Lirong Qiu, Weiqian Zhao, and Yingbin Sun
Appl. Opt. 53(31) 7407-7414 (2014)
Sheng-Hui Ding, Qi Li, Rui Yao, and Qi Wang
Appl. Opt. 49(36) 6834-6839 (2010)