A Constant Output Monochromator
Applied Optics, Vol. 7, Issue 4, pp. 599-602 (1968)
http://dx.doi.org/10.1364/AO.7.000599
Acrobat PDF (533 KB)
Abstract
A constant output monochromator has been developed that produces a power of 100 ± 2 µW over a wavelength range from 375 nm to 1400 nm with a spectral bandpass of 20 nm. The output is regulated by controlling the source intensity and the monochromator transmission. A thermopile is incorporated into the instrument for calibrating the output of the open loop system.
Citation
Gary B. Maxwell and Douglas D. Doonan, "A Constant Output Monochromator," Appl. Opt. 7, 599-602 (1968)
http://www.opticsinfobase.org/ao/abstract.cfm?URI=ao-7-4-599