1. Introduction
Digital imaging devices based on charge-coupled device (CCD) or
complementary metal–oxide–semiconductor (CMOS) are widely used in a
variety of scientific applications [
1G. C. Holst and T. S. Lomheim, CMOS/CCD Sensors and
Camera Systems
(SPIE, 2007).
]. In
radiometry, imaging devices are of great importance for carrying out certain types
of measurements that should otherwise be performed by means of scanning devices.
They are also the essential elements for multispectral imaging, including color
imaging, that becomes a powerful tool in many application fields like biomedical
science, remote sensing, arts, and color science.
For scientific applications of digital imaging
devices, the technical specifications,
such as spatial resolution, quantum efficiency, spectral responsivity, linearity,
geometric fidelity, and response time are
required to be carefully characterized.
In the radiometric and colorimetric applications, measurement of spectral
responsivity is of particular importance, which is defined as the ratio of output
digital reading to input radiometric quantity as a function of wavelength [
2“Spectral
responsivity measurements of detectors, radiometers, and
photometers,” International Commission on
Illumination (CIE) Publication No. 202
(2011).
]. Since the gain of digital sensors can be
easily adjusted and manipulated, the
normalized values of spectral
responsivity provide the essential information for further characterization and
calibration of the radiometric and colorimetric imaging
devices [
3H.-C. Lee, Introduction to Color Imaging Science
(Cambridge University,
2005).
]. Once the normalized spectral responsivity is measured, absolute
calibration of the instrument, i.e., adjustment of the electronic gain and data
acquisition algorithm, can be performed at a spectrally integrated calibration
source suitable for the target application.
A straightforward way to measure the spectral
responsivity of an imaging device is to
irradiate it uniformly so that each individual detector receives the same and known
amount of optical radiation from a wavelength-tunable monochromatic source.
A uniform source can be effectively
realized by using an integrating sphere
[
4W. A. Hovis and J. S. Knoll, “Characteristics of an internally
illuminated calibration sphere,” Appl.
Opt.
22, 4004–4007
(1983). [CrossRef]
]. However, the choice of a
wavelength-tunable monochromatic source requires a careful consideration. The
conventional choice for this is the combination of a white-spectrum (incandescent or
discharge) lamp and a monochromator. Such a monochromator-based tunable source can
scan a wide wavelength range with adjustable interval and bandwidth, but the
spectral power and spectral purity are
limited by the performance of the used
monochromator [
2“Spectral
responsivity measurements of detectors, radiometers, and
photometers,” International Commission on
Illumination (CIE) Publication No. 202
(2011).
]. In order to achieve a high
spectral purity, a double-grating configuration for a low stray level is required
which, on the other hand, reduces the spectral output power passing through the
monochromator. Recently, tunable laser-based
integrating sphere sources are
demonstrated at several national
metrology institutes as the
high-accuracy solution for spectral
responsivity measurement of imaging devices [
5A. Ferrero, J. Campos, and A. Pons, “Low-uncertainty absolute radiometric
calibration of a CCD,” Metrologia
43, S17–S21
(2006). [CrossRef]
V. E. Anderson, N. P. Fox, and D. H. Nettleton, “Highly stable, monochromatic and tunable
optical radiation source and its application to high accuracy
spectrophotometry,” Appl. Opt.
31, 536–545
(1992). [CrossRef]
S. W. Brown, G. P. Eppeldauer, and K. R. Lykke, “Facility for spectral irradiance and
radiance responsivity calibrations using uniform
sources,” Appl. Opt.
45, 8218–8237
(2006). [CrossRef]
–
8M. Schuster, S. Nevas, A. Sperling, and S. Voelker, “Spectral calibration of radiometric
detectors using tunable laser sources,” Appl.
Opt.
51, 1950–1961
(2012). [CrossRef]
]. However, the coverage of a
large wavelength range requires a large and
expensive laser facility, which is far
from a practical instrumentation for industrial or field applications.
In this paper, we present a new instrumentation scheme for
measuring the normalized spectral responsivity of digital imaging devices, which
consists of light-emitting diodes (LEDs), a single-grating monochromator, and an
integrating sphere [
9K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, “LED-based tunable monochromatic uniform
source for calibration of
imaging sensors and
cameras,” in Proceedings of the 11th
International Conference on New
Developments and Applications in Optical Radiometry
(NEWRAD, 2011),
pp. 118–119.
,
10K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, “Measurement of normalized spectral
response of digital imaging
sensors using a LED-based
tunable uniform source,” in Proceedings of
the XX IMEKO World Congress
(IMEKO, 2012), paper TC7-O-14.
]. LEDs are
compact, cost-effective, efficient, and
commercially available at different wavelengths covering from UV
through IR. Various tunable
quasi-monochromatic sources based on LEDs for radiometric applications are already
reported [
11I. Fryc, S. W. Brown, G. P. Eppeldauer, and Y. Ohno, “LED-based spectrally tunable source for
radiometric, photometric, and colorimetric
applications,” Opt. Eng.
44, 111309 (2005). [CrossRef]
G. Zaid, S.-N. Park, S. Park, and D.-H. Lee, “Differential
spectra responsivity
measurement of photovoltaic detectors with a LED-based integrating sphere
source,” Appl. Opt.
49,
6772–6783
(2010). [CrossRef]
–
13B. H. Hamadani, J. Roller, B. Dougherty, and H. W. Yoon,
“Versatile
light-emitting-diode-based spectral response measurement system for
photovoltaic device characterization,” Appl.
Opt.
51, 4469–4476
(2012). [CrossRef]
]. However, the relatively
large spectral bandwidth ranging from 20 to 50 nm limits the use of LEDs as a
spectral source. To overcome this shortcoming, we use a compact
single-grating monochromator as a
tunable spectral filter between LEDs
and an integrating sphere to develop a uniform source with a spectral bandwidth
smaller than 5 nm. By selecting a proper LED in accordance to the wavelength
setting of the monochromator, the spectral power could be optimized without reducing
the spectral purity.
We show measurement results for two digital imaging device samples, one monochrome
(B/W) CCD and one trichromic CMOS, by using the LED-based tunable uniform source.
The measurements are performed both in irradiance mode for bare sensors
and in radiance mode for cameras.
Measurement accuracy is verified by
evaluating the measurement uncertainty
and systematic error sources for the case of normalized spectral responsivity. The
achieved performance shows a high potential of the instrument as a practical
solution for industrial applications with a compact and cost-effective setup.
2. Instrumentation
A. Experimental Setup
Figure
1 schematically shows the
experimental setup for measuring the normalized spectral response of
digital imaging sensors and cameras. The LED-based tunable
uniform source consists of LEDs mounted
on a rotational disk, a compact
single-grating monochromator, and a barium sulfate (
) coated integrating sphere with an inner
diameter of 50 mm. The LED disk operates the selected one from multiple
mounted LEDs at a predefined position [
12G. Zaid, S.-N. Park, S. Park, and D.-H. Lee, “Differential
spectra responsivity
measurement of photovoltaic detectors with a LED-based integrating sphere
source,” Appl. Opt.
49,
6772–6783
(2010). [CrossRef]
]. We used 25 LEDs of different wavelengths to cover the wavelength
range from 380 to 900 nm.
Fig. 1. Schematic diagram of the normalized spectral responsivity measurement
setup (top view).
The radiation emitted from the selected LED is coupled into the monochromator
(Spectral Products, model CM110, focal length 110 mm,
, slit width 0.6 mm) by using two
planoconvex lenses with a diameter of 25.4 mm and a focal length of
20 mm. The spectrally filtered radiation from the monochromator is then
coupled into the integrating sphere by using a focusing lens through a sphere
port with a diameter of 12.7 mm. A focusing lens is used to form an image
of the monochromator output slit with a size of
on the sphere inner wall opposite to
the input port so that the light is
distributed inside the sphere only after the first diffuse reflection at the
sphere wall. No baffle is used inside the sphere as
the direct irradiation of the ports
by the incident light is effectively avoided though the beam spot
control. The choice of the sphere
size was the result of a compromise between the irradiance level and
the spatial uniformity of the
irradiance distribution inside the sphere. The irradiance level inside the
integrating sphere is monitored by
a Si photodiode attached at another sphere port with a diameter of 12.7 mm.
As we calibrate the signal of this photodiode with respect to spectral
irradiance at the position of the device under test (DUT), we refer it as the
reference detector (REF). The third port of the integrating sphere has a
diameter of 25.4 mm and is used as the output port of the tunable uniform
source. All three ports are placed so that the normal axes of all the ports are
mutually perpendicular (see Fig.
1).
Two different types of DUT can be measured with the LED-based tunable uniform
source. A bare imaging sensor,
which needs to be calibrated with
respect to spectral irradiance at a
specific position, is placed in front of the output port at a distance of
approximately 10 mm (irradiance mode measurement). Alternatively, an
imaging camera including a lens system for calibration with respect to spectral
radiance is placed at a distance corresponding to the
focal length of the lens system
(radiance mode measurement). For the irradiance mode measurement, the DUT sensor
is mounted on a computer-controlled three-axis translation stage for a precise
and reproducible positioning. The complete setup is mounted on a ground plate
with a size of less than
and controlled with computer
software for an automated measurement.
B. Source Characteristics and Measurement Procedure
The basic characteristics of the LED-based tunable uniform source, which can
influence the accuracy of the
normalized spectral responsivity measurement, are experimentally tested. The
measurement procedure is then optimized to avoid or minimize the
related error or uncertainty
sources.
1. Output Radiant Power
Figure
2(a) shows spectral power distribution of 25 LEDs used
for the tunable uniform source, which is measured at the entrance slit of
the monochromator by using a spectroradiometer (Instrument System, model
CAS140) with a resolution bandwidth of
approximately 3 nm.
Depending on materials and constructions of LEDs, each LED provides
different power level and different spectral bandwidth. Note that we used
only the LEDs, which are commercially available (supplied by Roithner Laser
Technik, Austria; super bright lamp type, 5 mm diameter round package
with clear dome lens, 8° viewing angle).
Fig. 2. (a) Spectral power distribution of the individual LEDs used for
the LED-based tunable uniform source. Different colors of the curves
correspond to different 25 LEDs, which had been lit one by one
during the wavelength scan. (b) Plot of the relative irradiance
available at the output port of the integrating sphere source as a
function of wavelength measured with an interval of 5 nm (dot
symbols). The error bars correspond to repeatability uncertainties
of each measurement points.
Selection of LED is correlated with the wavelength setting of the
monochromator: as the measurement wavelength is set by an operator, the
computer software automatically controls the monochromator, the rotational
disk position, and the LED driving circuit to turn on one LED that can
provide the maximum radiant power at that wavelength. The output
radiant power from the
monochromator, therefore, depends not only on the input LED power but also
on the coupling and
transmission through the grating monochromator at each wavelength.
Figure
2(b) shows the
relative change of irradiance as the wavelength scans from 380 to
900 nm with an interval of 5 nm. The irradiance at each wavelength
is measured by using a calibrated photodiode with an active area of
at a distance of 10 mm in front of
the sphere output port. The
uncertainties shown in Fig.
2(b)
as error bars are calculated from repeated scans of the normalized
irradiance measurement. We can
confirm that the repeatability uncertainty is large at the wavelengths with
low irradiance values, which will propagate to the uncertainty of spectral
responsivity measurement. The absolute values of
irradiance are not accurately
known because we could not accurately determine the size of the active area
as well as the uniformity of its responsivity. Nevertheless, based on the
specification of the photodiode, we could estimate that all the values of
the irradiance in
Fig.
2(b) lie within a range
from 10 to
. Considering the geometric condition, this
range of irradiance corresponds to a range of
radiance from 5 to
. We note that we have intentionally reduced
the coupling efficiency of LED
at several set wavelengths, at which the intrinsic output LED power is
extremely high compared to
other LEDs, in order to minimize error due to nonlinearity of signal
readings.
2. Spectral Purity
The spectral characteristics of output radiant power at each wavelength are
checked by using the
scanning-type spectroradiometer
(Instrument System, model Spectro 320) with a resolution bandwidth of
0.25 nm. The spectral bandwidth is measured to
be less than 5 nm as a
full width at half-maximum for the whole tuning range from 380 to
900 nm. The measured spectral bandwidth was in accordance with the
specification of the used single-grating monochromator with a
-number of 3.9 and a slit width of
0.6 mm.
In addition, the spectral stray light is measured
by scanning the
spectroradiometer with a large
resolution bandwidth of
5 nm in the range far from the set wavelength of the monochromator. Up
to the dynamic range limit of
, no out-of-band stray light could be
identified. This high spectral purity with low stray light is one of the
important advantage of the colored LED-monochromator system in
comparison with the
conventional white
lamp-mochromator system [
2“Spectral
responsivity measurements of detectors, radiometers, and
photometers,” International Commission on
Illumination (CIE) Publication No. 202
(2011).
]. Since the LEDs are
intrinsically
quasi-monochromatic with emission bandwidth of less than 50 nm, no
output of higher-order diffraction is expected. Therefore, no order-sorting
filter is required, which is indispensable for the lamp system to cover a
wide wavelength tuning range. Moreover, the ratio between the in-band power
and the out-of-band power for the LED system is much less than the case for
the lamp system, which results in a reduced stray light level. As a
consequence, the LED-monochromator system could be realized based on a
simple high-throughput, single-grating monochromator without concern about
the spectral purity.
3. Temporal Power Stability
The measurement of normalized spectral responsivity is
based on a comparison between the REF signal and the DUT signal at the same
radiation provided by the tunable uniform source. Temporal instability of
radiant power during the both signal readings, therefore, causes an error in
their ratio measurement. In order to achieve a stable radiant power, each
LED selected for a specific wavelength is operated with a constant forward
current by using a precision source meter. However, we noticed that each LED
shows a characteristic power drift within a duration of several minutes
after turning on, which is probably due to a thermalization in the LED
[
9K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, “LED-based tunable monochromatic uniform
source for calibration of
imaging sensors and
cameras,” in Proceedings of the 11th
International Conference on New
Developments and Applications in Optical Radiometry
(NEWRAD, 2011),
pp. 118–119.
,
10K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, “Measurement of normalized spectral
response of digital imaging
sensors using a LED-based
tunable uniform source,” in Proceedings of
the XX IMEKO World Congress
(IMEKO, 2012), paper TC7-O-14.
]. Since the emission characteristics of LEDs is sensitive to
temperature at the p–n junction, such a power drift is not to be
avoided for the setup where the junction temperature of the LEDs is not
actively stabilized [
14S. Park, Y.-W. Kim, D.-H. Lee, and S.-N. Park, “Preparation
of a standard light-emitting
diode (LED) for photometric measurements by functional
seasoning,” Metrologia
43, 299–305
(2006). [CrossRef]
].
In order to minimize the error due to the temporal power instability, we
built a drift check algorithm in the measurement software. At each set value
of wavelength, the readings of
the REF and DUT signals is repeated twice in the sequence
, where and denote the recordings of the REF and DUT
signals at a time of , respectively. The recording of the REF
corresponds to a reading of the averaged REF photocurrent by a calibrated
ammeter, while the recording of the DUT corresponds to capturing of its
digital image data as a single shot. The software compares then the relative
difference of the REF readings at and , which corresponds to the relative power
drift during the signal recordings for the ratio measurement. The software
repeats the measurement sequence at this wavelength until the relative power
drift between and is less than a predefined criterion of
0.1%. This has a consequence that the measurement software waits until the
LED lit for that wavelength reaches a sufficient power stability. The
measurement error due to a
power drift is limited to be smaller than 0.1% for the whole wavelength
range. We note that the short-term power instability is additionally
considered as a repeatability uncertainty of the readings of the averaged
REF photocurrent both at and with a number of sampling of more than 10
for each reading.
4. Spatial Uniformity
For an integrating sphere with a finite port size,
the spectral irradiance at the
output port is not perfectly uniform, but shows a distribution which varies
smoothly across the port [
4W. A. Hovis and J. S. Knoll, “Characteristics of an internally
illuminated calibration sphere,” Appl.
Opt.
22, 4004–4007
(1983). [CrossRef]
]. If two
detectors are to be compared at
such a nonuniform irradiance field, the difference in detector size or
position causes a systematic error [
12G. Zaid, S.-N. Park, S. Park, and D.-H. Lee, “Differential
spectra responsivity
measurement of photovoltaic detectors with a LED-based integrating sphere
source,” Appl. Opt.
49,
6772–6783
(2010). [CrossRef]
]. Therefore, characterization of
spatial uniformity of the
irradiance distribution at the position of DUT provides useful information
for error and uncertainty analysis.
Figure
3(a)
shows the spatial irradiance distribution of the LED-based tunable uniform
source, which is measured at a distance of 10 mm from the output port
through a horizontal scan at different wavelengths from 400 to 800 nm.
For this measurement, a monochrome CCD sensor is used as a DUT mounted on
the computer-controlled translation stage. While the CCD sensor is moved on
the horizontal axis crossing the output port through the center
(
-axis in Fig.
1), the digital images are captured at different
positions. The raw data at each position was the averaged value of the
pixels in a fixed area of interest (AOI) with a size of
pixels corresponding to approximately
. The measured raw data are then normalized
to the value at the center position of the output port. Such a measurement
of the normalized irradiance distribution is repeated at different
wavelengths. We verify in Fig.
3(a) that the spatial uniformity of irradiance is limited only
in a small area at the center of the output port. Such a nonuniform
distribution can cause a significant error when the absolute irradiance
responsivity should be measured by comparing two detectors with different
sizes [
12G. Zaid, S.-N. Park, S. Park, and D.-H. Lee, “Differential
spectra responsivity
measurement of photovoltaic detectors with a LED-based integrating sphere
source,” Appl. Opt.
49,
6772–6783
(2010). [CrossRef]
].
Fig. 3. (a) Spatial irradiance distribution of the LED-based tunable
uniform source measured at a distance of 10 mm from the output
port through a horizontal scan in 1 mm steps at different
wavelengths (dot symbols). The connecting curves are only for better
visibility. (b) Plot of the relative difference of the spatial
irradiance distribution with respect to the average values of the
different wavelengths within a range of at the center of the output
port.
However, concerning only measurement of the
normalized values of the
spectral irradiance responsivity, the nonuniformity of irradiance
distribution itself does not
affect the accuracy, as long as the
positioning of the DUT is
reproducible. Instead, the major concern is the spectral dependence of the
spatial irradiance
distribution. From Fig.
3(a), we
confirm that the spatial irradiance distribution measured at different
wavelengths show a good overlap. In order to verify it more precisely, we
plotted in Fig.
3(b) the relative difference of the
spatial irradiance distribution with respect to the average values of the
different wavelengths, which are calculated from the data of Fig.
3(a) within a range of
at the center of the output
port. We confirm that the wavelength dependence of irradiance distribution
is smaller than
in the range of
near the center, which is comparable to the
uncertainty of the signal
readings (see Table
2 and
Fig.
5). Outside the central
range, the random scattering of the
data increases, as expected,
from the decreased level of the DUT signal. No significant wavelength
dependence of spatial irradiance distribution could be
identified within the limit of
the measurement uncertainty. From the spatial uniformity measurements in
Fig.
3, we conclude that the
error due to spatial nonuniformity of irradiance can be minimized for the
normalized spectral responsivity measurement by positioning a calibration
detector or a test detector with high precision and reproducibility. In our
setup of Fig.
1, we used the
translation stage which can control the absolute position within
(Sigma Koki, SGSP model series).
5. Stray Light
The REF detector attached to the integrating sphere monitors the radiant
power level delivered from the LED-based tunable source into the sphere, and
will be calibrated against the spectral irradiance at a fixed point in front
of the output port (see Fig.
1).
Any ambient stray light which enters through the output port will change the
REF reading and cause a measurement error. In general, the ambient stray can
be easily corrected by a dark signal reading with a shutter inside the
monochromator. We performed the dark signal readings of both REF and DUT
prior to the main measurement procedure and subtracted them from the final
signal readings. The complete setup is placed inside a light-tight box to
keep the ambient stray as low as possible.
The specific problem related to our instrumentation is the stray due to
reflection from the DUT when it is placed close to the output port for the
irradiance mode measurement. In this condition, the light from the output
port is reflected from the detector surface and chassis of the test sensor
and comes back into the integrating sphere as an additional stray. Since
this stray exists only in the presence of the radiant power input, it cannot
be corrected by the dark readings. We have experimentally tested the amount
of the reflection stray by using different chassis materials at the position
of DUT. We compared a chassis of bare black-anodized Aluminum and a chassis
coated with a low-reflectance paint (3M/Mankiewicz, NEXTEL Velvet 811-21) at
a distance of below 20 mm from the output port. As a result, we
confirmed a significant change of the REF readings depending on the chassis
materials. Especially for wavelengths above 700 nm, where the
reflectance of black-anodized
aluminum increases, the
difference between two chassis materials were as large as 10% in the REF
signal.
In order to eliminate such an error due to the stray by
reflection, we designed the measurement procedure so that the DUT in the
irradiance mode is moved to a predefined “zero” position
outside the output port, whenever the REF reading is performed. For the DUT
reading, the DUT is moved back to the test position in front of the output
port. This procedure is easily realized with the computer-controlled
translation stage mounting the DUT in the irradiance measurement mode. We
note that such a DUT movement is not required for the radiance mode
measurement with a DUT at a large distance () from the output port, because the change
of the REF reading due to reflection from the DUT was negligible.
C. Instrument Calibration
Calibration of the instrument is performed in two steps.
First, the wavelength scale of the
monochromator is
calibrated. Two pencil-type calibration lamps of Hg(Ar) and Ar are placed at the
input slit of the monochromator and
the transmitted light is recorded at the output slit by using a photodiode while
scanning the wavelength with a small step of 0.2 nm. The instrument values
of the peak wavelengths of selected
isolated emission lines are compared with the literature values. The maximum
difference of the instrument
wavelength from the literature value was 0.73 nm in a range from 380 to
900 nm. From this measurement, we declared that the wavelength scale of the
tunable source is calibrated with a standard uncertainty of wavelength
, which is estimated from
by assuming a rectangular probability
distribution [
15Evaluation of measurement data—guide
to the expression of uncertainty in measurement
(JCGM, 2008).
].
Second, the responsivity of the REF photocurrent reading (in A) is calibrated
against the irradiance (in ) at the position of the DUT close to the output
port as a function of wavelength. However, as our aim is to measure the
normalized spectral responsivity of the DUT, the absolute values of the REF
responsivity (in ) is not of interest. Instead, we use only the
normalized values of the REF responsivity.
For calibration of the REF responsivity, a single-element Si photodiode
(Hamamatsu S1337-1010BQ, active
area
) with the known spectral responsivity is
mounted as a standard detector (STD) at the place of DUT in the irradiance mode.
The spectral power responsivity of the STD, which we denote as
, is calibrated against the KRISS spectral
responsivity scale from 300 to 1000 nm with relative standard uncertainties
from 0.2% to 0.4% (
) depending on wavelength [
16R. Goebel and M. Stock, “Report on the comparison CCPR-K2.b of
spectral responsivity measurements in the range 300 nm to
1000 nm,” Metrologia
41, 02004 (2004). [CrossRef]
]. The calibration equation for the REF responsivity
is then given by
where
and
denote the readings of the REF and of the STD,
respectively. Note that the reading of detectors used in this paper always
corresponds to the dark-subtracted signal. For the measurement of the normalized
spectral responsivity of an imaging sensor or camera, we use only the relative
values of the REF spectral responsivity
which are normalized, e.g., to its maximum
value:
Figure
4 shows the experimental
calibration result of the instrument which is a plot of the normalized REF
responsivity
as a function of wavelength measured from 380
to 900 nm with a step of 5 nm. The calibration measurements are
repeated and averaged at different distances of the STD to the
output port from 10 to 25 mm
with a step of 5 mm. The
distance dependence of the normalized REF
responsivity is included as a
random uncertainty component into the uncertainty budget. The error bars in
Fig.
4 indicate the combined
expanded uncertainty with a coverage factor
for a confidence level of approximately 95%
[
15Evaluation of measurement data—guide
to the expression of uncertainty in measurement
(JCGM, 2008).
]. The uncertainty
depends on wavelength mainly
because the radiant power of the LED-based tunable source strongly varies with
wavelength as shown in Fig.
2(b). In
addition, the uncertainty of the STD responsivity is also wavelength
dependent.
Fig. 4. Spectral responsivity of the REF as measured against the standard
photodiode mounted at a close distance to the output port. The error
bars represent the associated uncertainties of this calibration.
The uncertainty budget for the REF spectral responsivity
according to Eq. (
1) is shown in Table
1, as an example, at 435 nm, where
the combined uncertainty was the lowest. Expressed as relative uncertainties,
the budget is valid also for the normalized values in Eq. (
2). The uncertainty component of
type A and B correspond to the component evaluated with a statistical method and
with a nonstatistical method, respectively [
15Evaluation of measurement data—guide
to the expression of uncertainty in measurement
(JCGM, 2008).
]. For the type A components, the number of repetition is also
indicated in the parentheses.
Table 1. Uncertainty Budget of the REF Responsivity Calibration at 435 nm
| Related Quantity in
Eq. (1) | Uncertainty Component
Description | Evaluation
Type | Probability
Distribution | Relative Standard
Uncertainty |
|---|
| Repeatability of the REF reading | A (50) | t | 0.03% |
| Nonlinearity of current measurement device | B | Rectangular | 0.05% |
| Repeatability of the STD reading | A (50) | t | 0.02% |
| Nonlinearity of current measurement device | B | Rectangular | 0.3% |
| Wavelength calibration of monochromator | B | Rectangular | 0.01% |
| Distance dependence of calibration | A (5) | t | 0.22% |
| Radiant power drift | B | Rectangular | |
| Calibration uncertainty of the STD responsivity | B | Normal | 0.25% |
| Combined standard uncertainty of the REF responsivity | | Normal | 0.46% |
All the uncertainty components in Table
1 can be evaluated directly from the data of each quantity in
Eq. (
1) except the
uncertainty of the signal ratio with respect to the wavelength calibration
uncertainty. The propagation of the wavelength uncertainty
to the uncertainty of a ratio measurement is
described in [
17S. Park, D.-H. Lee, Y.-W. Kim, and S.-N. Park, “Uncertainty evaluation for the
spectroradiometric measurement of the averaged light-emitting diode
intensity,” Appl. Opt.
46, 2851–2858
(2007). [CrossRef]
]. From Eq. (
1), we derive the relation between the relative uncertainty of the
ratio
and the absolute uncertainty of wavelength
as
Note that the uncertainty of a ratio of two quantities due to an inaccurate
setting of wavelength depends on
the difference of the spectral slopes of both the quantities.
D. Data Acquisition of Imaging Device
Once the instrument is calibrated, the standard photodiode is removed and an
imaging device as a DUT is mounted for measurement. If the DUT is a bare digital
image sensor without a lens system, it is measured in the irradiance mode at a
distance of approximately 10 mm to the output port. If the DUT is a digital
camera with an imaging optics, the measurement is performed in the radiance mode
at a distance of the focal length of the imaging lens system (see
Fig.
1). The imaging device is
connected with the computer via an USB interface for control and data
communication.
In the measurement procedure of normalized spectral responsivity of the DUT, one
DUT reading at each wavelength corresponds to a capture of its digital image
data as a single shot. All the DUT readings during the wavelength scan of the
whole measurement range should be performed at a fixed gain and exposure of the
DUT, which are to be set for an optimal signal-to-noise ratio prior to the scan
measurement. The collected batch of digital images for one wavelength scan is
then treated by a separate data acquisition program, which calculates the
average and the standard deviation of pixel values within a predefined AOI for
each image. Normally, we set
pixels at the center of the image as AOI. The
average value of the pixel data in the AOI is
denoted as the reading of the DUT
at a wavelength of
,
, and the measurement equation of the normalized
spectral responsivity of the DUT,
, is now given by
If necessary, the final result
can be renormalized to its maximum
value or to any value at a specific normalization wavelength
:
3. Measurement Results
A. Monochrome CCD Sensor and Camera
The first DUT we have measured was a monochrome (B/W) progressive scan CCD sensor
(Sony, model ICX205AL,
pixels, pixel size ), which is integrated into a digital camera
device with an 8 bit analog-to-digital (A/D) converter (ThorLabs, model DCU
224M).
Figure
5 shows the measured
normalized spectral responsivity of the CCD sensor with expanded uncertainty as
error bars (
). We see the increased
uncertainties at several
wavelengths due to weak radiant power of the LED-based tunable source. As the
average digital counts at these wavelengths are low, the uncertainty of
resolution becomes significant, which also causes abrupt changes of the spectral
data. Nevertheless, most of the measurement data in the range from 380 to
900 nm have the relative expanded uncertainties of less than 1%
(
). The lowest uncertainty is achieved again at
the points, where the average count of the DUT was the highest.
Fig. 5. Normalized spectral responsivity of a monochrome CCD sensor measured in
the irradiance mode. The error bars represent the associated
uncertainties of this measurement.
The uncertainty budget of the measurement result of the monochrome CCD sensor in
Fig.
5 is representatively
presented in Table
2 for the
wavelength of 435 nm. The most significant component beside the calibration
uncertainty is the resolution uncertainty of the DUT reading. As the resolution
of the digital data is fixed to be unity, the resolution uncertainty depends on
the average count of the pixel data
in the AOI. For
at 435 nm, we calculate the relative
uncertainty due to resolution by
by assuming a rectangular probability
distribution. Compared to Table
1,
several systematic components are absent in Table
2
because these components are once
considered in the uncertainty of the REF responsivity. For Table
2, the equation for the uncertainty due
to wavelength inaccuracy in Eq. (
3) should be modified in accordance with Eq. (
4) to
Table 2. Uncertainty Budget of the Normalized Spectral Responsivity Measurement
for a Monochrome CCD Sensor at 435 nm
| Related Quantity in
Eq. (4) | Uncertainty Component
Description | Evaluation
Type | Probability
Distribution | Relative Standard
Uncertainty |
|---|
| Repeatability of the DUT reading | A (400) | t | 0.10% |
| Resolution | B | Rectangular | 0.12% |
| Repeatability of the REF reading | A (50) | t | 0.03% |
| Nonlinearity of current measurement device | B | Rectangular | 0.05% |
| Wavelength calibration of monochromator | B | Rectangular | 0.01% |
| Radiant power drift | B | Rectangular | |
| Calibration uncertainty of the REF responsivity (see
Table 1) | B | Normal | 0.46% |
| Combined standard uncertainty of the REF responsivity | | Normal | 0.49% |
The radiance mode measurement is demonstrated for an imaging camera, in which the
same monochrome CCD sensor is used. A camera lens system (AVENIR zoom lens, back
focal length 50 mm, ) is attached to the CCD sensor, and the camera
is placed at a distance of approximately 500 mm. The focal length of the
lens system is adjusted so that we see a well-focused image of the output port.
The camera is aligned so that the AOI is placed at the center of the output port
while the size of the AOI is kept much smaller than that of the output port
image.
Figure
6 shows the measured
normalized spectral responsivity of the CCD camera as dot symbols. The error
bars indicate the expanded uncertainty of the measurement
(
). The dashed curve corresponds to the CCD
sensor result of Fig.
5, which is
plotted together for comparison. The measurement uncertainty of the camera is
not very different from that of the sensor. However, we can clearly recognize
the effect of the lens system to the spectral responsivity from comparison of
the two results. Especially in the short wavelength range, the transmission of
the used lens system seems to be low so that the responsivity of the camera
tends to nearly zero. These measurement results demonstrate the practicability
of our instrument, which can measure both imaging sensors and cameras without
extra calibration.
Fig. 6. Normalized spectral responsivity of a monochrome CCD camera measured in
the radiance mode (dot symbols). The error bars represent the associated
uncertainties of this
measurement. The dashed
curve is the result of the CCD sensor in the camera, which is measured
in the irradiance mode and shown in Fig.
5.
B. Trichromatic CMOS Sensor and Camera
As the second DUT, we measured a trichromatic CMOS-based device with an
8 bit A/D converter (HVS, model HVR-2300CA, pixels, pixel size ). The measurement procedure was the same as the
case with the CCD device. However, the data acquisition is modified because
three different values of red (R), green (G), and blue (B) channels are assigned
to each pixel. We treated three channels as independent devices. However, the
normalization of three channels is
done to the same wavelength of 550 nm in order to compare their
responsivities to each other.
Figure
7 shows the measured
normalized spectral responsivity of the trichromatic CMOS sensor in the
irradiance mode with expanded uncertainty as error bars
(
). The spectral shapes and overlaps of
the three channels can be clearly
recognized. The responsivity above 680 nm is close to zero because
the CMOS sensor contains an
infrared-cutting filter in front of the sensor.
Fig. 7. Normalized spectral responsivity of a trichromatic CMOS sensor measured
in the irradiance mode. The error bars represent the associated
uncertainties of this measurement.
Figure
8 shows the result of the CMOS
camera. For this measurement, the camera lens system used for the monochromatic
CCD sensor is attached to the CMOS sensor. For comparison, the results of the
CMOS sensor of Fig.
7 are plotted
together as dashed curves. We see a significant and complex change of the
spectral responsivity for the case of the CMOS camera through attaching the lens
system. We presume that a multireflection interference between the
lens system and the
infrared-cutting filter can cause such a change. Figure
8 shows that such influences from lens
systems or other optical components to the spectral responsivity of the device
can be effectively identified and investigated by using our instrument.
Fig. 8. Normalized spectral responsivity of a tri-chromatic CMOS camera measured
in the radiance mode (dot symbols). The error bars represent the
associated uncertainties of this measurement. The dashed curve is the
result of the CMOS sensor in the camera, which is measured in the
irradiance mode and shown in Fig.
7.
4. Summary
We realized a new instrumentation scheme for measurement of normalized spectral
responsivity of imaging sensors and
cameras, which is based on a LED-based tunable spectral source with an integrating
sphere. A combination of 25 LEDs with different colors and a single-grating
monochromator provided tunable radiation with a bandwidth of less than 5 nm in
a range from 380 to 900 nm. A high spectral purity could be achieved without
using order-sorting filters and double-grating configuration.
We experimentally tested the characteristics of the LED-based tunable uniform source,
which can affect the accuracy of the measurement. The measurement procedure is then
designed and modified to minimize the error and uncertainty sources. The calibration
of the instrument is performed by using spectral lamps and a standard photodiode for
spectral responsivity.
The instrument is applied to a monochrome CCD device and a
trichrome CMOS device. The normalized spectral responsivity is measured both in the
irradiance mode for the bare sensors and in the radiance mode for the cameras. The
measurement uncertainty is evaluated to be less than 1% except several wavelengths
where the radiant power was too low.
In conclusion, we demonstrated a high potential
of the LED-based
tunable uniform source for practical metrology applications. Replacement of
conventional light source by LEDs provides a possibility to realize a cost-effective
and compact setup of a monochromatic source. For instance, the intrinsically low
spectral stray of the LED-based source makes the double-monochromator configuration
and the order-sorting filters unnecessary. Although LEDs cannot provide sufficient
radiant power at several wavelengths, careful uncertainty evaluation of the spectral
measurement provides sufficient information to identify and bridge these points.
Further improvements of the LED-based source are expected by using, e.g., its
excellent modulation property.
Acknowledgments
This work was supported by the Korea Research Institute of
Standards and Science (KRISS) under the project “Establishment of National
Physical Measurement Standards and Improvements of Calibration/Measurement
Capability,” grant 12011002.
References
1. | G. C. Holst and T. S. Lomheim, CMOS/CCD Sensors and
Camera Systems
(SPIE, 2007). |
2. | “Spectral
responsivity measurements of detectors, radiometers, and
photometers,” International Commission on
Illumination (CIE) Publication No. 202
(2011). |
3. | H.-C. Lee, Introduction to Color Imaging Science
(Cambridge University,
2005). |
4. | W. A. Hovis and J. S. Knoll, “Characteristics of an internally
illuminated calibration sphere,” Appl.
Opt.
22, 4004–4007
(1983). [CrossRef] |
5. | A. Ferrero, J. Campos, and A. Pons, “Low-uncertainty absolute radiometric
calibration of a CCD,” Metrologia
43, S17–S21
(2006). [CrossRef] |
6. | V. E. Anderson, N. P. Fox, and D. H. Nettleton, “Highly stable, monochromatic and tunable
optical radiation source and its application to high accuracy
spectrophotometry,” Appl. Opt.
31, 536–545
(1992). [CrossRef] |
7. | S. W. Brown, G. P. Eppeldauer, and K. R. Lykke, “Facility for spectral irradiance and
radiance responsivity calibrations using uniform
sources,” Appl. Opt.
45, 8218–8237
(2006). [CrossRef] |
8. | M. Schuster, S. Nevas, A. Sperling, and S. Voelker, “Spectral calibration of radiometric
detectors using tunable laser sources,” Appl.
Opt.
51, 1950–1961
(2012). [CrossRef] |
9. | K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, “LED-based tunable monochromatic uniform
source for calibration of
imaging sensors and
cameras,” in Proceedings of the 11th
International Conference on New
Developments and Applications in Optical Radiometry
(NEWRAD, 2011),
pp. 118–119. |
10. | K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, “Measurement of normalized spectral
response of digital imaging
sensors using a LED-based
tunable uniform source,” in Proceedings of
the XX IMEKO World Congress
(IMEKO, 2012), paper TC7-O-14. |
11. | I. Fryc, S. W. Brown, G. P. Eppeldauer, and Y. Ohno, “LED-based spectrally tunable source for
radiometric, photometric, and colorimetric
applications,” Opt. Eng.
44, 111309 (2005). [CrossRef] |
12. | G. Zaid, S.-N. Park, S. Park, and D.-H. Lee, “Differential
spectra responsivity
measurement of photovoltaic detectors with a LED-based integrating sphere
source,” Appl. Opt.
49,
6772–6783
(2010). [CrossRef] |
13. | B. H. Hamadani, J. Roller, B. Dougherty, and H. W. Yoon,
“Versatile
light-emitting-diode-based spectral response measurement system for
photovoltaic device characterization,” Appl.
Opt.
51, 4469–4476
(2012). [CrossRef] |
14. | S. Park, Y.-W. Kim, D.-H. Lee, and S.-N. Park, “Preparation
of a standard light-emitting
diode (LED) for photometric measurements by functional
seasoning,” Metrologia
43, 299–305
(2006). [CrossRef] |
15. | Evaluation of measurement data—guide
to the expression of uncertainty in measurement
(JCGM, 2008). |
16. | R. Goebel and M. Stock, “Report on the comparison CCPR-K2.b of
spectral responsivity measurements in the range 300 nm to
1000 nm,” Metrologia
41, 02004 (2004). [CrossRef] |
17. | S. Park, D.-H. Lee, Y.-W. Kim, and S.-N. Park, “Uncertainty evaluation for the
spectroradiometric measurement of the averaged light-emitting diode
intensity,” Appl. Opt.
46, 2851–2858
(2007). [CrossRef] |