Abstract
A commercial quadrupole mass spectrometer (residual gas analyzer, RGA) system has been modified to operate at large separations of its electronics console and sensing head. The methods implemented have made practical applications, as well as operation, of such a system possible for the first time. This advance was stimulated by a need for placing sensors at remote (45 or 60 m) chamber locations for space simulation testing of orbital flight spacecraft. Emphasis is placed on instrument functional requirements and describing the hardware changes and adjustment techniques necessary to assure operation at the extended cable lengths.
PDF Article
More Like This
Portable Remote Imaging Spectrometer coastal ocean sensor: design, characteristics, and first flight results
Pantazis Mouroulis, Byron Van Gorp, Robert O. Green, Heidi Dierssen, Daniel W. Wilson, Michael Eastwood, Joseph Boardman, Bo-Cai Gao, David Cohen, Brian Franklin, Frank Loya, Sarah Lundeen, Alan Mazer, Ian McCubbin, David Randall, Brandon Richardson, Jose I. Rodriguez, Charles Sarture, Eugenio Urquiza, Rudolph Vargas, Victor White, and Karl Yee