Abstract
A simple Flow Injection/Inductively Coupled Plasma Atomic Emission Spectrometry system is described. The analytical performance, in terms of linearity, precision, and detection limits, was investigated for both aqueous and organic solvents. Acceptable detection limits were obtained with the use of approximately one-hundreth of the minimum sample volume required for continuous aspiration. Some deterioration of detection limits, however, was observed. The suitability of this system to biological samples was demonstrated by the analysis of iron- and cobalt-rubredoxins and National Bureau of Standards' Bovine Liver standard reference material.
PDF Article
More Like This
Practical and highly sensitive elemental analysis for aqueous samples containing metal impurities employing electrodeposition on indium-tin oxide film samples and laser-induced shock wave plasma in low-pressure helium gas
Koo Hendrik Kurniawan, Marincan Pardede, Rinda Hedwig, Syahrun Nur Abdulmadjid, Kurnia Lahna, Nasrullah Idris, Eric Jobiliong, Hery Suyanto, Maria Margaretha Suliyanti, May On Tjia, Tjung Jie Lie, Zener Sukra Lie, Davy Putra Kurniawan, and Kiichiro Kagawa
Appl. Opt. 54(25) 7592-7597 (2015)