Surface-Enhanced Raman Spectroscopy at a Silver Electrode as a Real-Time Detector in Flowing Streams
Applied Spectroscopy, Vol. 46, Issue 1, pp. 147-151 (1992)
Acrobat PDF (534 KB)
Abstract
An analytical application for Surface-Enhanced Raman Spectroscopy at a silver electrode is described. Real-time SER spectra of adenine and cytosine have been recorded in a 10-μL spectroelectrochemical flow cell under flowing conditions. Charge-coupled-device detection allowed high-quality spectra spanning a ~1200 cm−1 region to be recorded with integration times of 4 seconds. A low-power He.Ne laser was used as a source. SERS at the silver electrode offers rapid time response to adsorption/desorption by appropriate potential modulation. The technique is extremely reproducible and insensitive to temperature and flow rate. The effects of incident photon energy and applied potential on the intensity of the Raman signal are discussed.
Citation
Neil J. Pothier and R. Ken Forcé, "Surface-Enhanced Raman Spectroscopy at a Silver Electrode as a Real-Time Detector in Flowing Streams," Appl. Spectrosc. 46, 147-151 (1992)
http://www.opticsinfobase.org/as/abstract.cfm?URI=as-46-1-147
You do not have subscription access to this journal. Citation lists with outbound citation links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Log in to access OSA Member Subscription





OSA is a member of 