MEMS segmented-based adaptive optics scanning laser ophthalmoscope

Silvestre Manzanera, Michael A. Helmbrecht, Carl J. Kempf, and Austin Roorda  »View Author Affiliations


Biomedical Optics Express, Vol. 2, Issue 5, pp. 1204-1217 (2011)
http://dx.doi.org/10.1364/BOE.2.001204


View Full Text Article

Acrobat PDF (1622 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

The performance of a MEMS (micro-electro-mechanical-system) segmented deformable mirror was evaluated in an adaptive optics (AO) scanning laser ophthalmoscope. The tested AO mirror (Iris AO, Inc, Berkeley, CA) is composed of 37 hexagonal segments that allow piston/tip/tilt motion up to 5 μm stroke and ±5 mrad angle over a 3.5 mm optical aperture. The control system that implements the closed-loop operation employs a 1:1 matched 37-lenslet Shack-Hartmann wavefront sensor whose measurements are used to apply modal corrections to the deformable mirror. After a preliminary evaluation of the AO mirror optical performance, retinal images from 4 normal subjects over a 0.9°x0.9° field size were acquired through a 6.4 mm ocular pupil, showing resolved retinal features at the cellular level. Cone photoreceptors were observed as close as 0.25 degrees from the foveal center. In general, the quality of these images is comparable to that obtained using deformable mirrors based on different technologies.

© 2011 OSA

1. Introduction

In the early 1980s, Webb et al. [1

1. R. H. Webb, G. W. Hughes, and O. Pomerantzeff, “Flying spot TV ophthalmoscope,” Appl. Opt. 19(17), 2991–2997 (1980). [CrossRef] [PubMed]

,2

2. R. H. Webb and G. W. Hughes, “Scanning laser ophthalmoscope,” IEEE Trans. Biomed. Eng. BME-28(7), 488–492 (1981). [CrossRef] [PubMed]

] devised a new ophthalmoscope to image the living human retina. This instrument, called the confocal scanning laser ophthalmoscope (SLO), was based on a confocal optical imaging system and a scanning laser light beam illuminating the retina. It was shown to provide higher image contrast and light efficiency than conventional imaging systems [3

3. R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]

].

But, as in any other ophthalmoscope, the objective lens of the imaging system is the eye’s optical system comprised of the cornea and crystalline lens, which are affected by optical aberrations that degrade the quality of the retinal images. In order to solve this problem, adaptive optics (AO) was incorporated in the SLO. The new instrument, the adaptive optics scanning laser ophthalmoscope (AOSLO) [4

4. A. Roorda, F. Romero-Borja, W. Donnelly Iii, H. Queener, T. J. Hebert, and M. C. W. Campbell, “Adaptive optics scanning laser ophthalmoscopy,” Opt. Express 10(9), 405–412 (2002). [PubMed]

] allowed visualization of the living human retina at cellular level with improved lateral and axial resolution [5

5. F. Romero-Borja, K. Venkateswaran, A. Roorda, and T. Hebert, “Optical slicing of human retinal tissue in vivo with the adaptive optics scanning laser ophthalmoscope,” Appl. Opt. 44(19), 4032–4040 (2005). [CrossRef] [PubMed]

,6

6. K. Venkateswaran, A. Roorda, and F. Romero-Borja, “Theoretical modeling and evaluation of the axial resolution of the adaptive optics scanning laser ophthalmoscope,” J. Biomed. Opt. 9(1), 132–138 (2004). [CrossRef] [PubMed]

].

AO is a technique that has proven to be effective in improving the resolution of ground-based telescope images [7

7. H. W. Babcock, “The possibility of compensating astronomical seeing,” Publ. Astron. Soc. Pac. 65(386), 229–236 (1953). [CrossRef]

9

9. R. Tyson, Principles of Adaptive Optics (CRC Press, 2010).

] and correcting the eye’s aberrations to obtain enhanced in vivo retinal images [10

10. H. Hofer, L. Chen, G. Y. Yoon, B. Singer, Y. Yamauchi, and D. R. Williams, “Improvement in retinal image quality with dynamic correction of the eye’s aberrations,” Opt. Express 8(11), 631–643 (2001). [CrossRef] [PubMed]

]. Briefly, an AO system consists of two elements, a wavefront sensor (WFS) that measures the aberrations and a corrector that removes the wavefront error. Significant efforts have been made to build and test these correcting devices. Some of them are based on the optical properties of liquid crystals [11

11. E. J. Fernández, P. M. Prieto, and P. Artal, “Wave-aberration control with a liquid crystal on silicon (LCOS) spatial phase modulator,” Opt. Express 17(13), 11013–11025 (2009). [CrossRef] [PubMed]

13

13. F. Vargas-Martín, P. M. Prieto, and P. Artal, “Correction of the aberrations in the human eye with a liquid-crystal spatial light modulator: limits to performance,” J. Opt. Soc. Am. A 15(9), 2552–2562 (1998). [CrossRef] [PubMed]

] and others, the most widely used, are deformable mirrors (DM) that make use of different types of technologies to modify the shape of a mirror surface [14

14. J. Z. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef] [PubMed]

17

17. E. Dalimier and C. Dainty, “Comparative analysis of deformable mirrors for ocular adaptive optics,” Opt. Express 13(11), 4275–4285 (2005). [CrossRef] [PubMed]

]. Among these, AO mirrors based on MEMS (micro-electro-mechanical-systems) technology have proved to be especially well suited for applications in the eye [18

18. N. Doble, G. Yoon, L. Chen, P. Bierden, B. Singer, S. Olivier, and D. R. Williams, “Use of a microelectromechanical mirror for adaptive optics in the human eye,” Opt. Lett. 27(17), 1537–1539 (2002). [CrossRef] [PubMed]

], due to their reduced size, and capability of fine spatial control of the wavefront. Making use of these advantages, Zhang et al. [19

19. Y. H. Zhang, S. Poonja, and A. Roorda, “MEMS-based adaptive optics scanning laser ophthalmoscopy,” Opt. Lett. 31(9), 1268–1270 (2006). [CrossRef] [PubMed]

] built a compact AOSLO suitable for a clinical environment.

In all of the aforementioned studies, the DMs employed were of the continuous face sheet type. That is, the mirror surface employs a single reflecting deformable layer. But the problem of shaping a reflecting surface can also be tackled using an array of tiled flat mirrors, so-called segmented mirrors, whose facets can be driven independently to shape the wavefront. Initially, bulky segmented mirrors were built for astronomical applications. In 1990 Hulburd and colleagues [20

20. B. Hulburd and D. Sandler, “Segmented mirrors for atmospheric compensation,” Opt. Eng. 29(10), 1186–1190 (1990). [CrossRef]

] described and characterized a 512-segment, 22-cm diameter segmented mirror. And in 1992 [21

21. D. S. Acton and R. C. Smithson, “Solar imaging with a segmented adaptive mirror,” Appl. Opt. 31(16), 3161–3169 (1992). [CrossRef] [PubMed]

] a 19-segment hexagonally packed mirror, 2.8-cm across each segment, was used to image the Sun.

MEMS technology allowed the manufacture of segmented mirrors with much reduced sizes. In 1997, the first experimental demonstration of aberration correction using a MEMS hexagonal piston micromirror array was reported [22

22. M. C. Roggeman, V. M. Bright, B. M. Welsh, S. R. Hick, P. C. Roberts, W. D. Cowan, and J. H. Comtois, “Use of micro-electro-mechanical deformable mirrors to control aberrations in optical systems: theoretical and experimental results,” Opt. Eng. 36(5), 1326–1338 (1997). [CrossRef]

]. Since then, a number of other studies have characterized other piston-only [23

23. W. D. Cowan, M. K. Lee, B. M. Welsh, V. M. Bright, and M. C. Roggemann, “Surface micromachined segmented mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 5(1), 90–101 (1999). [CrossRef]

,24

24. A. Tuantranont and V. M. Bright, “Segmented silicon-micromachined microelectromechanical deformable mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 8(1), 33–45 (2002). [CrossRef]

] and piston-tip-tilt [25

25. D. J. Dagel, W. D. Cowan, O. B. Spahn, G. D. Grossetete, A. J. Grine, M. J. Shaw, P. J. Resnick, and B. Jokiel, “Large-stroke MEMS deformable mirrors for adaptive optics,” J. Microelectromech. Syst. 15(3), 572–583 (2006). [CrossRef]

] segmented mirrors. In general, although these devices showed to be good aberration correctors they also showed poor optical efficiency. The reasons for the latter were the big spaces between segments or reduced fill factor and the poor flatness of the segments, both which caused light losses and diffraction effects. Although some effort was made to overcome this problem by using a lenslet array in front of the mirror to focus light on the segments, other technical issues arose [22

22. M. C. Roggeman, V. M. Bright, B. M. Welsh, S. R. Hick, P. C. Roberts, W. D. Cowan, and J. H. Comtois, “Use of micro-electro-mechanical deformable mirrors to control aberrations in optical systems: theoretical and experimental results,” Opt. Eng. 36(5), 1326–1338 (1997). [CrossRef]

,24

24. A. Tuantranont and V. M. Bright, “Segmented silicon-micromachined microelectromechanical deformable mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 8(1), 33–45 (2002). [CrossRef]

].

Why would one choose to use segmented mirrors for vision applications? Segmented mirrors do not offer immediate optical benefits over more conventional DM technology [26

26. N. Doble, D. T. Miller, G. Yoon, and D. R. Williams, “Requirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes,” Appl. Opt. 46(20), 4501–4514 (2007). [CrossRef] [PubMed]

]. Nor do they represent a path toward a less expensive device compared to other MEMS technologies. In some respects, the 37-segment mirror used in this report represents a step backward in terms of optical control. By comparison, the Boston Micromachines deformable mirror, which is currently used in our laboratory, has similar stroke with 140 actuators behind a continuous face-sheet reflector. Nevertheless, segmented mirrors offer some unique advantages for future AO applications for vision science. For example, segmented mirrors are more scalable than continuous face-sheet designs. Each actuator is independent, so there are no new optical constraints imposed by adding more elements. Second, there is more flexibility for coatings on the segmented reflectors. Advanced coatings may prove useful for specific imaging applications like 2-photon [27

27. J. J. Hunter, B. Masella, A. Dubra, R. Sharma, L. Yin, W. H. Merigan, G. Palczewska, K. Palczewski, and D. R. Williams, “Images of photoreceptors in living primate eyes using adaptive optics two-photon ophthalmoscopy,” Biomed. Opt. Express 2(1), 139–148 (2011). [CrossRef] [PubMed]

] or autofluorescence [28

28. J. I. W. Morgan, J. J. Hunter, W. H. Merigan, and D. R. Williams, “The reduction of retinal autofluorescence caused by light exposure,” Invest. Ophthalmol. Vis. Sci. 50(12), 6015–6022 (2009). [CrossRef] [PubMed]

] where light efficiency is paramount. Finally, the ability to tip and tilt actuators can be used to deflect regions of the pupil out of the optical path, offering an efficient and effective way to control the pupil aperture. Pupil control can be used to limit pupil size and shape, or can be used to bypass scattering caused by small cataracts or highly aberrated regions arising from keratoconus or other disorders. None of these applications are implemented in this paper, because our first interest was in assessing the basic performance of a segmented mirror in a modern AOSLO design. In our opinion, the performance of a system depends as much on the optical design of the instrument as on the quality of wavefront sensor and corrector. Given the evolution and proven benefits of careful optical design for ophthalmic AO systems [29

29. E. A. Rossi, M. Chung, A. Dubra, J. J. Hunter, W. H. Merigan, and D. R. Williams, “Imaging retinal mosaics in the living eye,” Eye (Lond.) 25(3), 301–308 (2011). [CrossRef] [PubMed]

], it’s worth reassessing alternate DM technology in such a modern design. If a 37-element segmented mirror design works reasonably well, then the prospects are good for better performance with a MEMS segmented DM that has higher actuator counts.

In order to make these devices useful for vision applications, as pointed out by some authors [26

26. N. Doble, D. T. Miller, G. Yoon, and D. R. Williams, “Requirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes,” Appl. Opt. 46(20), 4501–4514 (2007). [CrossRef] [PubMed]

,30

30. D. Miller, L. Thibos, and X. Hong, “Requirements for segmented correctors for diffraction-limited performance in the human eye,” Opt. Express 13(1), 275–289 (2005). [CrossRef] [PubMed]

], the fill factor should be well above 90%. In the last years, a new generation of MEMS segmented mirrors [31

31. M. A. Helmbrecht, M. He, T. Juneau, M. Hart, and N. Doble, “Segmented MEMS deformable-mirror for wavefront correction,” Proc. SPIE 6376, 63760D, 63760D-9 (2006). [CrossRef]

] that have a 98% fill factor, flatten below 7-nm rms (for some models) and have very flat segments (3-20 nm rms, depending on model) have become available (Iris AO, Inc., Berkeley, CA).

Recently, an Iris AO mirror was used in an AO retinal imaging system to explore retinal lesions in primates after infrared radiation overexposure [32

32. J. W. Oliver, G. Pocock, R. Vincelette, S. Kumru, G. D. Noojin, K. J. Schuster, D. Stolarski, A. Shingledecker, and B. A. Rockwell, “In vivo investigation of near infrared retinal lesions utilizing two adaptive optics enhanced imaging modalities,” Proc. SPIE 7175, 71750H, 71750H-9 (2009). [CrossRef]

]. In the human eye, after confirming its capabilities as a wavefront corrector [33

33. N. Doble, C. Kempf, M. Helmbrecht, and A. Roorda, “Closed loop adaptive optics in the human eye using a segmented MEMS deformable mirror,” Invest. Ophthalmol. Vis. Sci. 49, ARVO E-Abstract 4195 (2008).

] it was used in a flood illuminated fundus camera [34

34. N. Doble and S. Choi, “Widefield imaging of the human retina using adaptive optics,” Invest. Ophthalmol. Vis. Sci. 50, ARVO E-Abstract 1062 (2009).

]. But, to date, the performance of MEMS segmented mirrors in an AOSLO for high resolution retinal imaging in the living human eye has not been tested yet. In the present work, an Iris AO 37-segment mirror was used as the corrector element in an AOSLO and its performance was evaluated.

2. Methods

2.1. The AOSLO

The DM is integrated into the AOSLO as depicted in Fig. 1
Fig. 1 Schematic diagram of the AOSLO setup. LD: light delivery; PMT: photomultiplier tube; WFS: wavefront sensor; DM: deformable mirror; HS: horizontal scanner; VS: vertical scanner. The optical setup is built along two perpendicular planes to remove astigmatism.
. The light delivery (LD) module, through a beam splitter, introduces an 840 nm (50-nm bandwidth) collimated beam in the optical system, reaching the eye after reflecting off the DM, the horizontal scanner (HS) and the vertical scanner (VS). The infrared light comes from a superluminescent diode (Broadlighter S840, Superlum, Russia). The HS and the VS are a resonant scanner and a galvanometric scanner, respectively, manufactured by Electro-Optics Products Corp, Flushing Meadows, NY. The resonant scanner works at 16 kHz producing a sinusoidal scan that is coupled to the galvanometric scanner which generates a sawtooth pattern at 1/525th of the fast scan frequency. The diffusely reflected light from each point of the drawn raster scan on the retina transmits inversely along the ingoing path. Part of this light is then reflected on a beam splitter, focused on a pinhole by means of a collecting lens (L1) and finally detected by a GaAs photo multiplier tube (PMT) (Hamamatsu, Japan). The ideal pinhole diameter is around 20 μm [35

35. T. Wilson, “The role of the pinhole in confocal imaging systems,” in The Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum, New York, 1990), pp. 99–113.

] although a 80-μm pinhole was finally chosen to increase throughput. Dedicated software, an electronic module and a frame grabbing board (GenesisLC, Matrox, Canada) convert the detected signal by the PMT into an image of the retina. The rest of light reflected off the retina is used to measure the wavefront error and feed the AO control system that corrects for the aberrations of the eye.

Pairs of spherical mirrors forming 4f systems assure that the eye’s pupil, the VS, the HS and the DM planes are optically conjugated. Finally, to relay the wavefront onto the lenslet array in the WFS two achromatic doublets (L2, L3) are used, yielding a 1:1 magnification between the DM and the WFS, as required. The mirror focal lengths are chosen in such a way that the whole aperture of the DM, 3.5 mm, is used to correct a 6.4 mm pupil in the eye. Three flat mirrors fold the path to meet the space and system design requirements.

One of the most important requirements was to remove, in the eye’s pupil plane, the residual astigmatism induced by off-axis reflections from the spherical mirrors. To achieve this goal, equal amounts of astigmatism was generated along two perpendicular planes, parallel and perpendicular to the optical table [3

3. R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]

,36

36. S. A. Burns, R. Tumbar, A. E. Elsner, D. Ferguson, and D. X. Hammer, “Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope,” J. Opt. Soc. Am. A 24(5), 1313–1326 (2007). [CrossRef] [PubMed]

]. The combination left only defocus at the eye’s pupil, which could be removed by simple axial displacement of the mirrors. An optimization procedure was carried out using optical design software (Zemax©, Bellevue, WA) to determine the angles of incidence of the light on each mirror necessary to yield a diffraction limited system over a 1.6°x1.6° field of view with minimum raster scan distortion.

2.2. The AO mirror

The AO mirror is a MEMS segmented mirror manufactured by Iris AO, Inc. (Berkeley, CA), model PTT111-5 with a 7.0-nm RMS flatness. Each of the 37 hexagonal segments that constitute the array is driven by three actuators that allow piston/tip/tilt motion up to 5-μm stroke and ± 5 mrad angle. The segments are tightly packed to achieve a 98% fill factor over a 3.5-mm optical aperture. Figure 2
Fig. 2 Image of the Iris AO mirror and a detailed view of the hexagonal segment arrangement. The optical aperture is 3.5 mm and each facet is 0.7 mm from vertex to vertex. Courtesy of Iris AO, Inc.
shows a general view of the AO mirror and a schematic diagram of the segment arrangement.

The reflecting surface is attached to an actuator platform that is elevated above the substrate by three flexures. The flexures provide mechanical restoring forces that counter the attractive electrostatic forces of the three underlying diamond-shaped electrodes. Equal voltages on each of the electrodes cause the segment to be pulled downward in a piston motion whereas different voltage distributions produce tip and tilt motions. An illustrative diagram of the latter can be seen in Fig. 3
Fig. 3 Schematic diagram of the structure underlying each mirror segment that allows its different motions. The mirror segment is attached to an actuator platform that can be pushed and tilted by applying the necessary voltage distribution to a set of three electrodes. The flexures provide mechanical restoring forces that counter the attractive electrostatic forces. Courtesy of Iris AO, Inc.
. These movements are not independent and an excessive use of one of them compromises the ability to actuate the others. In order to allow push-pull motions an initial voltage is applied to every segment to put them in a bias position.

2.3. The AO system

3. Results

3.1. Preliminary test on a model eye

Prior to using the AOSLO to obtain retinal images from human eyes, the quality of the AO correction was evaluated using a model eye, consisting of a lens that served as the refractive elements in the eye (cornea and crystalline lens) and a diffuse surface acting as the retina.

The first parameter selected to evaluate the quality of the AO correction is, usually, the residual RMS of the wavefront measured by the WFS. For the model eye with different amounts of induced defocus, the residual RMS was always below 0.025 μm through the 6.4 mm pupil. However, for this particular AO system, the RMS must not be considered alone in order to evaluate its performance. First, because the wavefront error is sampled with a limited number of lenslets, there could potentially be undetected high spatial frequency features in the wavefront. Secondly, the discontinuous nature of the AO mirror surface could induce diffractive effects, which affect the AOSLO image but are not detected by the WFS.

Hence, to overcome the aforementioned issues, the quality of the AO correction was evaluated through two measurements, the PSF provided by the AO mirror -which greatly determines the image resolution- and the double pass image (DPI) that is formed on the confocal pinhole in the AOSLO, which is the ultimate determination of the optical image quality.

3.1.1. PSF evaluation

The PSF was obtained by illuminating the flat AO mirror with a 633-nm collimated laser beam and focusing the reflected light on a CCD camera. To achieve this while keeping the AO mirror in its place in the AOSLO as shown in Fig. 1, an additional optical setup was built. This is constituted by a He-Ne laser, a spatial filter, a collimating lens to generate an aberration-free beam and a set of flat mirrors to direct the light onto the DM and finally to the CCD, where the PSF is registered after being focused by a 100-mm focal length lens. The angle of incidence on the AO mirror was less than 2 degrees, the minimum achievable considering the space requirements. Through the maximum aperture of the AO mirror −3.5 mm- a sequence of three snapshots were taken and averaged to yield the final PSF.

The so obtained image is shown in Fig. 4
Fig. 4 a) PSF produced by the flat AO mirror. Pupil size is 3.5 mm, wavelength is 633 nm and the focal length of the focusing lens is 100 mm. The energy distribution is normalized to unity and the scale on x-y axis represents microns on the CCD. b) Measured energy distribution along the x-axis. For comparison, the theoretically expected PSF from a diffraction limited system, labeled as “ideal”, is also shown. c) Wide-field PSFs obtained saturating the CCD with increasing intensity from left to right to show the dimmest details.
. For comparison, together with the experimentally obtained PSF, the ideal PSF expected under the same conditions was calculated and is also shown on the same graph. The measured PSF (FWHM = 31.2 μm) is very similar to the Airy disk (FWHM = 18.7 μm), revealing the flatness of the AO mirror and the absence of relevant diffraction effects that could affect the AOSLO image quality. Only when the CCD is saturated is possible to reveal -see panel c)- the dimmest diffraction structure caused by the hexagonal segment arrangement.

3.1.2. Double pass image evaluation

Another valuable element to assess the quality of the AO correction is the actual light intensity distribution on the confocal pinhole in the AOSLO which is the ultimate determination of the optical image quality. To the aim of registering this light distribution, the model eye was used together with a removable flat mirror placed between lenses L2 and L3 that directed the focusing light from L2 onto a CCD camera located at its focal plane. The light source used was the AOSLO’s infrared laser (840 nm). The so obtained intensity distribution is the DPI through the AOSLO and model eye optics. This DPI can be easily rescaled to the actual light distribution on the pinhole simply considering the ratio between the focal length of lenses L1 and L2. Although there is a non-common path between lenses L1 and L2 both are doublets carefully aligned and the obtained DPIs will serve to the purpose of determining whether the optical performance is affected by diffraction or imperfections in the DM.

The DPIs obtained after correcting different amounts of defocus (−0.50 D, −0.25 D, + 0.25 D, + 0.50 D) induced by placing the corresponding trial lenses in front of the model eye were recorded. The procedure consisted of making the AO system correct for the aberrations present in the optical system, then holding the AO mirror and placing the removable flat mirror in its location to record the DPI. To reduce noise, 10 images were averaged to yield the final DPI for each correction. Figure 5a
Fig. 5 a) DPI obtained after correcting + 0.25 D of defocus. The energy distribution is normalized to unity and the scale on the x-y axis represents microns on the pinhole plane. b) Energy distributions along the x-axis after correcting the different amounts of defocus. For comparison, the theoretically expected DPI from a diffraction limited system (autoconvolution of the Airy disk) is also shown. The dashed line rectangle indicates the relative size of the 80 μm diameter pinhole employed as the confocal aperture. Wavelength is 840 nm, data are rescaled to a 100 mm focal length focusing lens and the beam diameter over this lens is 3.5 mm.
shows the resulting DPI after correcting + 0.25 D of defocus. For the rest of the defocus corrections the registered energy distribution along the x-axis was extracted from each final DPI for all the corrections (Fig. 5b). For comparison, the theoretically expected DPI from a diffraction limited system, computed as the autoconvolution of the Airy disk, is also shown on the same graph.

As expected from the PSF evaluation, the recorded DPIs (FWHM = 53.8 μm) are close to the ideal energy distribution (FWHM = 31.7 μm) under the same conditions and no significant diffraction effects are noticed. The DPI analysis also allows us to evaluate the optical performance not only for a flat AO mirror, as in the PSF evaluation case, but when the AO mirror is correcting for different amount of aberrations, which has a major practical interest. Figure 5 shows that all the DPIs are very similar through the range of tested defocus and there is only a slight decrease in the maximum intensity for the + 0.50 D case. From these results a uniform retinal image quality is expected when the AO system corrects for different amounts and types of aberrations.

3.1.3. Checking the flatness of the segments

For this particular AO mirror, another aspect to be considered in the evaluation of the AO correction is the performance in correcting defocus with opposite signs. If there is a high enough bow in the AO mirror segments [22

22. M. C. Roggeman, V. M. Bright, B. M. Welsh, S. R. Hick, P. C. Roberts, W. D. Cowan, and J. H. Comtois, “Use of micro-electro-mechanical deformable mirrors to control aberrations in optical systems: theoretical and experimental results,” Opt. Eng. 36(5), 1326–1338 (1997). [CrossRef]

] then this could lead to a better fit to defocused wavefronts with a certain sign. For example, concave segments would fit better to concave wavefronts, giving a better correction and better images. To determine if this particular AO mirror has similar effects, the model eye was used to obtain AOSLO images after correcting induced defocus from −0.50 D to + 0.50 D in 0.25 D steps, using spherical trial lenses and a parallel lens for the 0 D case. The mean gray level in the image was chosen as a simple parameter to compare the AO correction quality through the different cases. The results are graphed in Fig. 6
Fig. 6 Mean gray level (right y-axis) in the images recorded by the AOSLO after correcting different amounts of defocus induced by the corresponding trial lenses. On the left y-axis are represented the initial RMS before the AO correction and the final RMS achieved after the said correction.
, showing that the mean gray level remains nearly constant and hence there is no different performance when correcting defocus with different signs. Although there is a slight decrease in the mean gray level at + 0.50 D, any bow in the segments is not likely to be responsible for it, since a difference would have been observed between −0.25 D and + 0.25 D. Together with the mean gray level, the initial RMS and the final RMS achieved after each correction are also plotted. The symmetry of both graphs around 0 D shows that the equal image quality performance for positive and negative defocus is not biased neither by different initial requirements (Initial RMS) or different performance on the wavefront corrections (Final RMS).

3.2. AO performance on real eyes

To obtain the retinal images shown in both figures, we made use of the tools -described in the methods section- available in the AO control, in particular the possibility of selecting the pupil size that is used to estimate the aberrations on the whole pupil (6.4 mm). Starting with the largest possible pupil we switched to a smaller one when the AO control was not able to provide high quality images. For subject A the AO correction was performed making use of the wavefront sampling provided by the inner 19 lenslets (4.6 mm diameter sensing region). For subjects C and D the inner 31 lenslets (5.9 mm diam.) were used. And for subject B the best images were acquired when all the 37 lenslets (6.4 mm diam.) were employed. Independently of the sensing region used to estimate the complete wavefront, the retinal images were always obtained through the 6.4 mm pupil, except for subject D whose pupil was not big enough to fill the maximum aperture.

Figure 9
Fig. 9 AO performance for subjects A, B, C and D. The initial RMS wavefront error is measured before the AO correction but after correcting defocus and astigmatism with trial lenses. The final RMS is measured after the AO correction. RMS is computed using up to 5th order Zernike terms. Six measurements were used to obtain the RMS values and their error bars, which represent ± 1SD. (See text for details). Pupil size is 6.4 mm.
plots the performance of the AO system, showing the obtained RMS considering up to 5th order Zernike terms, before and after the AO correction on each of the 4 subjects. Six different AO corrections were carried out on each subject, and the last 151 measurements (2.5 s) after the closed-loop was stable, were averaged. These 6 mean values were, in turn, averaged and their standard deviation calculated to yield the final value and error bars respectively. Note that in this case error bars do not represent the stability through the closed-loop operation, but the stability through different AO corrections. The same applies for the RMS measured before initiating the closed-loop. As can be observed in Fig. 9, in all subjects the final RMS was always below 0.11 μm. It should be remarked that these RMS data were obtained for a 6.4 mm pupil, independently of the sensing region that was actually used to estimate the wavefront correction that was applied to the DM.

The initial RMS was measured after correcting defocus and astigmatism by means of trial lenses placed in front of the eye to mitigate against the stroke limits of the AO mirror.

In order to understand the AO correction performance in more detail, Fig. 10
Fig. 10 Wavefront error in terms of Zernike polynomials up to 5th order measured before and after AO correction for subjects A, B, C and D. Values are the average through 6 different measurements and error bars represent ± 1SD (see text for further details). Pupil size is 6.4 mm.
shows plots of the Zernike coefficient values that fit the measured wavefront errors for the six AO corrections performed on each subject and that were used to compute the RMS values graphed on Fig. 9. The procedure followed to obtain the final values and the significance of their corresponding error bars is the same as that of the RMS values. Figure 10 shows that the AO control system provided an excellent correction over those Zernike terms it was asked to correct. The corresponding bars on the graph are so small that they are not distinguishable and those terms averaged almost to zero (some noise may have remained but is not revealed in this plot due to the averaging method employed). In addition, the equally negligible error bars indicate that this final correction state was very stable through the different closed-loop operations performed.

4. Discussion

We do not claim that the images presented in this paper represent the best possible performance of an AO ophthalmoscope (see Rossi et al. [29

29. E. A. Rossi, M. Chung, A. Dubra, J. J. Hunter, W. H. Merigan, and D. R. Williams, “Imaging retinal mosaics in the living eye,” Eye (Lond.) 25(3), 301–308 (2011). [CrossRef] [PubMed]

], for the best examples). But it is important to note that these images were obtained with a modest number of actuators and using a Shack-Hartmann WFS with a limited number of lenslets (37) to sample the wavefront. Although 37 lenslets should be high enough to correct the most important aberrations in the general population [40

40. J. Porter, A. Guirao, I. G. Cox, and D. R. Williams, “Monochromatic aberrations of the human eye in a large population,” J. Opt. Soc. Am. A 18(8), 1793–1803 (2001). [CrossRef] [PubMed]

], the numbers shown in Fig. 9 and Fig. 10 do not imply that perfectly error-free wavefronts were achieved. Moreover, the AO correction seems to perform equally well in both, higher and lower order terms, which is unexpected considering that the AO corrector utilizes only 37 segments. Therefore, what actually can be claimed from Fig. 10 is that the AO control system performs quite well correcting those aberrations that is able to detect.

One of the most important concerns when employing segmented mirrors is the presence of undesired diffraction effects. This is particularly important in retinal imaging where the size and quality of the PSF determines the achievable image resolution. But with a 98% fill factor and carefully cophased segments, the AO mirror tested in this work has shown, under the performed test, to be relatively free of these effects.

5. Conclusions

A MEMS segmented AO mirror (37 segments) has been used as the aberration corrector in an AOSLO to evaluate the capabilities and potential of this type of technology in vision systems.

Some tests using a model eye were initially performed. They showed: i) that the flatness of the segments in the mirror was high enough to not affect the AO performance in the sense that the same image quality is obtained correcting positive or negative defocus and ii), that neither the PSF recorded from a flat AO mirror nor the DPIs registered after correcting different amounts of defocus showed unwanted diffraction effects from the segments.

Retinal images located at the fovea and 1° temporal were acquired from 4 normal subjects. The AO correction yielded a wavefront error (sampled by 37 lenslets) always below 0.11 μm and, despite the moderate number of segments in the AO mirror, cone photoreceptors were imaged as close as 0.25° from the foveal center in one of the subjects. It is possible, considering the normal inter-subject variability, to acquire retinal images with similar quality to that obtained with AOSLOs that make use of AO mirrors based on different technologies.

Acknowledgments

This work has been funded by Ministerio de Educación y Ciencia, Human Resources Mobility National Program I-D + I 2008-2011 (Spain), National Science Foundation grant NSF AST 9876783 and National Institutes of Health grant NIH EY14735.

References and links

1.

R. H. Webb, G. W. Hughes, and O. Pomerantzeff, “Flying spot TV ophthalmoscope,” Appl. Opt. 19(17), 2991–2997 (1980). [CrossRef] [PubMed]

2.

R. H. Webb and G. W. Hughes, “Scanning laser ophthalmoscope,” IEEE Trans. Biomed. Eng. BME-28(7), 488–492 (1981). [CrossRef] [PubMed]

3.

R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]

4.

A. Roorda, F. Romero-Borja, W. Donnelly Iii, H. Queener, T. J. Hebert, and M. C. W. Campbell, “Adaptive optics scanning laser ophthalmoscopy,” Opt. Express 10(9), 405–412 (2002). [PubMed]

5.

F. Romero-Borja, K. Venkateswaran, A. Roorda, and T. Hebert, “Optical slicing of human retinal tissue in vivo with the adaptive optics scanning laser ophthalmoscope,” Appl. Opt. 44(19), 4032–4040 (2005). [CrossRef] [PubMed]

6.

K. Venkateswaran, A. Roorda, and F. Romero-Borja, “Theoretical modeling and evaluation of the axial resolution of the adaptive optics scanning laser ophthalmoscope,” J. Biomed. Opt. 9(1), 132–138 (2004). [CrossRef] [PubMed]

7.

H. W. Babcock, “The possibility of compensating astronomical seeing,” Publ. Astron. Soc. Pac. 65(386), 229–236 (1953). [CrossRef]

8.

F. Roddier, Adaptive Optics in Astronomy (Cambridge University Press, 1999).

9.

R. Tyson, Principles of Adaptive Optics (CRC Press, 2010).

10.

H. Hofer, L. Chen, G. Y. Yoon, B. Singer, Y. Yamauchi, and D. R. Williams, “Improvement in retinal image quality with dynamic correction of the eye’s aberrations,” Opt. Express 8(11), 631–643 (2001). [CrossRef] [PubMed]

11.

E. J. Fernández, P. M. Prieto, and P. Artal, “Wave-aberration control with a liquid crystal on silicon (LCOS) spatial phase modulator,” Opt. Express 17(13), 11013–11025 (2009). [CrossRef] [PubMed]

12.

P. Prieto, E. Fernández, S. Manzanera, and P. Artal, “Adaptive optics with a programmable phase modulator: applications in the human eye,” Opt. Express 12(17), 4059–4071 (2004). [CrossRef] [PubMed]

13.

F. Vargas-Martín, P. M. Prieto, and P. Artal, “Correction of the aberrations in the human eye with a liquid-crystal spatial light modulator: limits to performance,” J. Opt. Soc. Am. A 15(9), 2552–2562 (1998). [CrossRef] [PubMed]

14.

J. Z. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef] [PubMed]

15.

E. J. Fernández, I. Iglesias, and P. Artal, “Closed-loop adaptive optics in the human eye,” Opt. Lett. 26(10), 746–748 (2001). [CrossRef] [PubMed]

16.

M. Glanc, E. Gendron, F. Lacombe, D. Lafaille, J. F. Le Gargasson, and P. Lena, “Towards wide-field retinal imaging with adaptive optics,” Opt. Commun. 230(4-6), 225–238 (2004). [CrossRef]

17.

E. Dalimier and C. Dainty, “Comparative analysis of deformable mirrors for ocular adaptive optics,” Opt. Express 13(11), 4275–4285 (2005). [CrossRef] [PubMed]

18.

N. Doble, G. Yoon, L. Chen, P. Bierden, B. Singer, S. Olivier, and D. R. Williams, “Use of a microelectromechanical mirror for adaptive optics in the human eye,” Opt. Lett. 27(17), 1537–1539 (2002). [CrossRef] [PubMed]

19.

Y. H. Zhang, S. Poonja, and A. Roorda, “MEMS-based adaptive optics scanning laser ophthalmoscopy,” Opt. Lett. 31(9), 1268–1270 (2006). [CrossRef] [PubMed]

20.

B. Hulburd and D. Sandler, “Segmented mirrors for atmospheric compensation,” Opt. Eng. 29(10), 1186–1190 (1990). [CrossRef]

21.

D. S. Acton and R. C. Smithson, “Solar imaging with a segmented adaptive mirror,” Appl. Opt. 31(16), 3161–3169 (1992). [CrossRef] [PubMed]

22.

M. C. Roggeman, V. M. Bright, B. M. Welsh, S. R. Hick, P. C. Roberts, W. D. Cowan, and J. H. Comtois, “Use of micro-electro-mechanical deformable mirrors to control aberrations in optical systems: theoretical and experimental results,” Opt. Eng. 36(5), 1326–1338 (1997). [CrossRef]

23.

W. D. Cowan, M. K. Lee, B. M. Welsh, V. M. Bright, and M. C. Roggemann, “Surface micromachined segmented mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 5(1), 90–101 (1999). [CrossRef]

24.

A. Tuantranont and V. M. Bright, “Segmented silicon-micromachined microelectromechanical deformable mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 8(1), 33–45 (2002). [CrossRef]

25.

D. J. Dagel, W. D. Cowan, O. B. Spahn, G. D. Grossetete, A. J. Grine, M. J. Shaw, P. J. Resnick, and B. Jokiel, “Large-stroke MEMS deformable mirrors for adaptive optics,” J. Microelectromech. Syst. 15(3), 572–583 (2006). [CrossRef]

26.

N. Doble, D. T. Miller, G. Yoon, and D. R. Williams, “Requirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes,” Appl. Opt. 46(20), 4501–4514 (2007). [CrossRef] [PubMed]

27.

J. J. Hunter, B. Masella, A. Dubra, R. Sharma, L. Yin, W. H. Merigan, G. Palczewska, K. Palczewski, and D. R. Williams, “Images of photoreceptors in living primate eyes using adaptive optics two-photon ophthalmoscopy,” Biomed. Opt. Express 2(1), 139–148 (2011). [CrossRef] [PubMed]

28.

J. I. W. Morgan, J. J. Hunter, W. H. Merigan, and D. R. Williams, “The reduction of retinal autofluorescence caused by light exposure,” Invest. Ophthalmol. Vis. Sci. 50(12), 6015–6022 (2009). [CrossRef] [PubMed]

29.

E. A. Rossi, M. Chung, A. Dubra, J. J. Hunter, W. H. Merigan, and D. R. Williams, “Imaging retinal mosaics in the living eye,” Eye (Lond.) 25(3), 301–308 (2011). [CrossRef] [PubMed]

30.

D. Miller, L. Thibos, and X. Hong, “Requirements for segmented correctors for diffraction-limited performance in the human eye,” Opt. Express 13(1), 275–289 (2005). [CrossRef] [PubMed]

31.

M. A. Helmbrecht, M. He, T. Juneau, M. Hart, and N. Doble, “Segmented MEMS deformable-mirror for wavefront correction,” Proc. SPIE 6376, 63760D, 63760D-9 (2006). [CrossRef]

32.

J. W. Oliver, G. Pocock, R. Vincelette, S. Kumru, G. D. Noojin, K. J. Schuster, D. Stolarski, A. Shingledecker, and B. A. Rockwell, “In vivo investigation of near infrared retinal lesions utilizing two adaptive optics enhanced imaging modalities,” Proc. SPIE 7175, 71750H, 71750H-9 (2009). [CrossRef]

33.

N. Doble, C. Kempf, M. Helmbrecht, and A. Roorda, “Closed loop adaptive optics in the human eye using a segmented MEMS deformable mirror,” Invest. Ophthalmol. Vis. Sci. 49, ARVO E-Abstract 4195 (2008).

34.

N. Doble and S. Choi, “Widefield imaging of the human retina using adaptive optics,” Invest. Ophthalmol. Vis. Sci. 50, ARVO E-Abstract 1062 (2009).

35.

T. Wilson, “The role of the pinhole in confocal imaging systems,” in The Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum, New York, 1990), pp. 99–113.

36.

S. A. Burns, R. Tumbar, A. E. Elsner, D. Ferguson, and D. X. Hammer, “Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope,” J. Opt. Soc. Am. A 24(5), 1313–1326 (2007). [CrossRef] [PubMed]

37.

K. Grieve, P. Tiruveedhula, Y. H. Zhang, and A. Roorda, “Multi-wavelength imaging with the adaptive optics scanning laser Ophthalmoscope,” Opt. Express 14(25), 12230–12242 (2006). [CrossRef] [PubMed]

38.

M. A. Helmbrecht and T. Juneau, “Piston-tip-tilt positioning of a segmented MEMS deformable mirror,” Proc. SPIE 6467, 64670M. (2007)

39.

C. R. Vogel, D. W. Arathorn, A. Roorda, and A. Parker, “Retinal motion estimation in adaptive optics scanning laser ophthalmoscopy,” Opt. Express 14(2), 487–497 (2006). [CrossRef] [PubMed]

40.

J. Porter, A. Guirao, I. G. Cox, and D. R. Williams, “Monochromatic aberrations of the human eye in a large population,” J. Opt. Soc. Am. A 18(8), 1793–1803 (2001). [CrossRef] [PubMed]

OCIS Codes
(330.4460) Vision, color, and visual optics : Ophthalmic optics and devices
(110.1080) Imaging systems : Active or adaptive optics

ToC Category:
Ophthalmology Applications

History
Original Manuscript: March 10, 2011
Revised Manuscript: April 12, 2011
Manuscript Accepted: April 13, 2011
Published: April 13, 2011

Virtual Issues
Cellular Imaging of the Retina (2011) Biomedical Optics Express

Citation
Silvestre Manzanera, Michael A. Helmbrecht, Carl J. Kempf, and Austin Roorda, "MEMS segmented-based adaptive optics scanning laser ophthalmoscope," Biomed. Opt. Express 2, 1204-1217 (2011)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-2-5-1204


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. R. H. Webb, G. W. Hughes, and O. Pomerantzeff, “Flying spot TV ophthalmoscope,” Appl. Opt. 19(17), 2991–2997 (1980). [CrossRef] [PubMed]
  2. R. H. Webb and G. W. Hughes, “Scanning laser ophthalmoscope,” IEEE Trans. Biomed. Eng. BME-28(7), 488–492 (1981). [CrossRef] [PubMed]
  3. R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]
  4. A. Roorda, F. Romero-Borja, W. Donnelly Iii, H. Queener, T. J. Hebert, and M. C. W. Campbell, “Adaptive optics scanning laser ophthalmoscopy,” Opt. Express 10(9), 405–412 (2002). [PubMed]
  5. F. Romero-Borja, K. Venkateswaran, A. Roorda, and T. Hebert, “Optical slicing of human retinal tissue in vivo with the adaptive optics scanning laser ophthalmoscope,” Appl. Opt. 44(19), 4032–4040 (2005). [CrossRef] [PubMed]
  6. K. Venkateswaran, A. Roorda, and F. Romero-Borja, “Theoretical modeling and evaluation of the axial resolution of the adaptive optics scanning laser ophthalmoscope,” J. Biomed. Opt. 9(1), 132–138 (2004). [CrossRef] [PubMed]
  7. H. W. Babcock, “The possibility of compensating astronomical seeing,” Publ. Astron. Soc. Pac. 65(386), 229–236 (1953). [CrossRef]
  8. F. Roddier, Adaptive Optics in Astronomy (Cambridge University Press, 1999).
  9. R. Tyson, Principles of Adaptive Optics (CRC Press, 2010).
  10. H. Hofer, L. Chen, G. Y. Yoon, B. Singer, Y. Yamauchi, and D. R. Williams, “Improvement in retinal image quality with dynamic correction of the eye’s aberrations,” Opt. Express 8(11), 631–643 (2001). [CrossRef] [PubMed]
  11. E. J. Fernández, P. M. Prieto, and P. Artal, “Wave-aberration control with a liquid crystal on silicon (LCOS) spatial phase modulator,” Opt. Express 17(13), 11013–11025 (2009). [CrossRef] [PubMed]
  12. P. Prieto, E. Fernández, S. Manzanera, and P. Artal, “Adaptive optics with a programmable phase modulator: applications in the human eye,” Opt. Express 12(17), 4059–4071 (2004). [CrossRef] [PubMed]
  13. F. Vargas-Martín, P. M. Prieto, and P. Artal, “Correction of the aberrations in the human eye with a liquid-crystal spatial light modulator: limits to performance,” J. Opt. Soc. Am. A 15(9), 2552–2562 (1998). [CrossRef] [PubMed]
  14. J. Z. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef] [PubMed]
  15. E. J. Fernández, I. Iglesias, and P. Artal, “Closed-loop adaptive optics in the human eye,” Opt. Lett. 26(10), 746–748 (2001). [CrossRef] [PubMed]
  16. M. Glanc, E. Gendron, F. Lacombe, D. Lafaille, J. F. Le Gargasson, and P. Lena, “Towards wide-field retinal imaging with adaptive optics,” Opt. Commun. 230(4-6), 225–238 (2004). [CrossRef]
  17. E. Dalimier and C. Dainty, “Comparative analysis of deformable mirrors for ocular adaptive optics,” Opt. Express 13(11), 4275–4285 (2005). [CrossRef] [PubMed]
  18. N. Doble, G. Yoon, L. Chen, P. Bierden, B. Singer, S. Olivier, and D. R. Williams, “Use of a microelectromechanical mirror for adaptive optics in the human eye,” Opt. Lett. 27(17), 1537–1539 (2002). [CrossRef] [PubMed]
  19. Y. H. Zhang, S. Poonja, and A. Roorda, “MEMS-based adaptive optics scanning laser ophthalmoscopy,” Opt. Lett. 31(9), 1268–1270 (2006). [CrossRef] [PubMed]
  20. B. Hulburd and D. Sandler, “Segmented mirrors for atmospheric compensation,” Opt. Eng. 29(10), 1186–1190 (1990). [CrossRef]
  21. D. S. Acton and R. C. Smithson, “Solar imaging with a segmented adaptive mirror,” Appl. Opt. 31(16), 3161–3169 (1992). [CrossRef] [PubMed]
  22. M. C. Roggeman, V. M. Bright, B. M. Welsh, S. R. Hick, P. C. Roberts, W. D. Cowan, and J. H. Comtois, “Use of micro-electro-mechanical deformable mirrors to control aberrations in optical systems: theoretical and experimental results,” Opt. Eng. 36(5), 1326–1338 (1997). [CrossRef]
  23. W. D. Cowan, M. K. Lee, B. M. Welsh, V. M. Bright, and M. C. Roggemann, “Surface micromachined segmented mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 5(1), 90–101 (1999). [CrossRef]
  24. A. Tuantranont and V. M. Bright, “Segmented silicon-micromachined microelectromechanical deformable mirrors for adaptive optics,” IEEE J. Sel. Top. Quantum Electron. 8(1), 33–45 (2002). [CrossRef]
  25. D. J. Dagel, W. D. Cowan, O. B. Spahn, G. D. Grossetete, A. J. Grine, M. J. Shaw, P. J. Resnick, and B. Jokiel, “Large-stroke MEMS deformable mirrors for adaptive optics,” J. Microelectromech. Syst. 15(3), 572–583 (2006). [CrossRef]
  26. N. Doble, D. T. Miller, G. Yoon, and D. R. Williams, “Requirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes,” Appl. Opt. 46(20), 4501–4514 (2007). [CrossRef] [PubMed]
  27. J. J. Hunter, B. Masella, A. Dubra, R. Sharma, L. Yin, W. H. Merigan, G. Palczewska, K. Palczewski, and D. R. Williams, “Images of photoreceptors in living primate eyes using adaptive optics two-photon ophthalmoscopy,” Biomed. Opt. Express 2(1), 139–148 (2011). [CrossRef] [PubMed]
  28. J. I. W. Morgan, J. J. Hunter, W. H. Merigan, and D. R. Williams, “The reduction of retinal autofluorescence caused by light exposure,” Invest. Ophthalmol. Vis. Sci. 50(12), 6015–6022 (2009). [CrossRef] [PubMed]
  29. E. A. Rossi, M. Chung, A. Dubra, J. J. Hunter, W. H. Merigan, and D. R. Williams, “Imaging retinal mosaics in the living eye,” Eye (Lond.) 25(3), 301–308 (2011). [CrossRef] [PubMed]
  30. D. Miller, L. Thibos, and X. Hong, “Requirements for segmented correctors for diffraction-limited performance in the human eye,” Opt. Express 13(1), 275–289 (2005). [CrossRef] [PubMed]
  31. M. A. Helmbrecht, M. He, T. Juneau, M. Hart, and N. Doble, “Segmented MEMS deformable-mirror for wavefront correction,” Proc. SPIE 6376, 63760D, 63760D-9 (2006). [CrossRef]
  32. J. W. Oliver, G. Pocock, R. Vincelette, S. Kumru, G. D. Noojin, K. J. Schuster, D. Stolarski, A. Shingledecker, and B. A. Rockwell, “In vivo investigation of near infrared retinal lesions utilizing two adaptive optics enhanced imaging modalities,” Proc. SPIE 7175, 71750H, 71750H-9 (2009). [CrossRef]
  33. N. Doble, C. Kempf, M. Helmbrecht, and A. Roorda, “Closed loop adaptive optics in the human eye using a segmented MEMS deformable mirror,” Invest. Ophthalmol. Vis. Sci. 49, ARVO E-Abstract 4195 (2008).
  34. N. Doble and S. Choi, “Widefield imaging of the human retina using adaptive optics,” Invest. Ophthalmol. Vis. Sci. 50, ARVO E-Abstract 1062 (2009).
  35. T. Wilson, “The role of the pinhole in confocal imaging systems,” in The Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum, New York, 1990), pp. 99–113.
  36. S. A. Burns, R. Tumbar, A. E. Elsner, D. Ferguson, and D. X. Hammer, “Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope,” J. Opt. Soc. Am. A 24(5), 1313–1326 (2007). [CrossRef] [PubMed]
  37. K. Grieve, P. Tiruveedhula, Y. H. Zhang, and A. Roorda, “Multi-wavelength imaging with the adaptive optics scanning laser Ophthalmoscope,” Opt. Express 14(25), 12230–12242 (2006). [CrossRef] [PubMed]
  38. M. A. Helmbrecht and T. Juneau, “Piston-tip-tilt positioning of a segmented MEMS deformable mirror,” Proc. SPIE 6467, 64670M. (2007)
  39. C. R. Vogel, D. W. Arathorn, A. Roorda, and A. Parker, “Retinal motion estimation in adaptive optics scanning laser ophthalmoscopy,” Opt. Express 14(2), 487–497 (2006). [CrossRef] [PubMed]
  40. J. Porter, A. Guirao, I. G. Cox, and D. R. Williams, “Monochromatic aberrations of the human eye in a large population,” J. Opt. Soc. Am. A 18(8), 1793–1803 (2001). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited