Self-starting, self-regulating Fourier domain mode locked fiber laser for OCT imaging

Kartikeya Murari, Jessica Mavadia, Jiefeng Xi, and Xingde Li  »View Author Affiliations


Biomedical Optics Express, Vol. 2, Issue 7, pp. 2005-2011 (2011)
http://dx.doi.org/10.1364/BOE.2.002005


View Full Text Article

Acrobat PDF (1300 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We present a Fourier domain mode locking (FDML) fiber laser with a feedback loop allowing automatic startup without a priori knowledge of the fundamental drive frequency. The feedback can also regulate the drive frequency making the source robust against environmental variations. A control system samples the energy of the light traversing the FDML cavity and uses a voltage controlled oscillator (VCO) to drive the tunable fiber Fabry-Perot filter in order to maximize that energy. We demonstrate a prototype self-starting, self-regulating FDML operating at 40 kHz with a full width tuning range of 140 nm around 1305 nm and a power output of ~40 mW. The laser starts up with no operator intervention in less than 5 seconds and exhibits improved spectral stability over a conventional FDML source. In OCT applications the source achieved over 120 dB detection sensitivity and an ~8.9-µm axial resolution.

© 2011 OSA

1. Introduction

Swept lasers have become an attractive light source for optical coherence tomography (OCT), offering high imaging speed, narrow instantaneous linewidth and improved detection sensitivity [1

1. J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28(21), 2067–2069 (2003). [CrossRef] [PubMed]

3

3. M. Choma, M. Sarunic, C. Yang, and J. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]

]. Several implementations of swept lasers have been described in literature such as mechanical scanning [4

4. S. R. Chinn, E. A. Swanson, and J. G. Fujimoto, “Optical coherence tomography using a frequency-tunable optical source,” Opt. Lett. 22(5), 340–342 (1997). [CrossRef] [PubMed]

7

7. S. H. Yun, C. Boudoux, G. J. Tearney, and B. E. Bouma, “High-speed wavelength-swept semiconductor laser with a polygon-scanner-based wavelength filter,” Opt. Lett. 28(20), 1981–1983 (2003). [CrossRef] [PubMed]

], tunable filtering of fiber ring lasers [3

3. M. Choma, M. Sarunic, C. Yang, and J. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]

] and Fourier domain mode locking (FDML) lasers [8

8. R. Huber, M. Wojtkowski, and J. G. Fujimoto, “Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography,” Opt. Express 14(8), 3225–3237 (2006). [CrossRef] [PubMed]

]. In particular, an FDML fiber laser incorporates a delay line in a circular loop wherein a range of wavelengths are continuously propagating and being amplified. Particular wavelengths are made available as a function of time using a tunable filter synchronized to the round-trip propagation delay.

One of the many advantages of FDML lasing is that within the loop cavity, lasing occurs at all wavelengths simultaneously avoiding the buildup time for laser operation leading to sweep rates limited only by the tunable filter. The recursive amplification also leads to high power output. These advantages have led to considerable interest in FDML sources, with recent reports on ultra-high speed OCT imaging [9

9. M. Gora, K. Karnowski, M. Szkulmowski, B. J. Kaluzny, R. Huber, A. Kowalczyk, and M. Wojtkowski, “Ultra high-speed swept source OCT imaging of the anterior segment of human eye at 200 kHz with adjustable imaging range,” Opt. Express 17(17), 14880–14894 (2009). [CrossRef] [PubMed]

12

12. T. Klein, W. Wieser, C. M. Eigenwillig, B. R. Biedermann, and R. Huber, “Megahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laser,” Opt. Express 19(4), 3044–3062 (2011). [CrossRef] [PubMed]

], OCT data sampling algorithms [13

13. J. F. Xi, L. Huo, J. S. Li, and X. D. Li, “Generic real-time uniform K-space sampling method for high-speed swept-source optical coherence tomography,” Opt. Express 18(9), 9511–9517 (2010). [CrossRef] [PubMed]

], high output power sources [14

14. S. Marschall, T. Klein, W. Wieser, B. R. Biedermann, K. Hsu, K. P. Hansen, B. Sumpf, K. H. Hasler, G. Erbert, O. B. Jensen, C. Pedersen, R. Huber, and P. E. Andersen, “Fourier domain mode-locked swept source at 1050 nm based on a tapered amplifier,” Opt. Express 18(15), 15820–15831 (2010). [CrossRef] [PubMed]

] among others.

The operation and stability of an FDML source critically depends on the synchronization of the tunable filter drive frequency with the light circulation frequency in the delay loop. This poses two challenges. First, conventional FDML sources use a high resolution function generator to manually tune the drive frequency while the extent of synchronization is assessed by the operator by observing the time-dependent spectrum of the source. This is an inherently subjective procedure requiring some expertise. Secondly, due to changes in environmental conditions like the temperature and humidity, the optical path length of the delay loop may change leading to a loss of synchronization between the present drive frequency and the circulation frequency. This manifests as spectral fluctuations in the source output which can be detrimental to the system operation.

We present an FDML source with a control system incorporating a voltage controlled oscillator (VCO) in a feedback configuration that seeks to adjust the drive frequency of the tunable filter such that the energy of the light traversing the loop is maximized. This approach addresses both of the issues described earlier. The source can be started up without a priori knowledge of the fundamental drive frequency. The control system scans a broad range of frequencies, iteratively narrowing the range and arriving at the optimal frequency with no manual intervention. After startup, the control system monitors the power output and can retune the drive frequency if a deviation exceeds a predetermined threshold. We analyze the hypothesis underlying the feedback scheme and describe the electrical and optical implementation thereof. We prototyped a self-starting, self-regulating FDML (SS-FDML) source operating with a 140 nm full width tuning range around 1305 nm sweeping at 40 kHz with a power output of ~40 mW. Characterization results indicating optimal and rapid automatic startup and improved spectral stability over a conventional FDML source are presented. OCT imaging at 40,000 axial scans per second using the source is also demonstrated.

2. System architecture

The hypothesis behind our proposed scheme is that an FDML source works optimally when the tunable filter drive frequency is perfectly matched with the light circulation frequency. Intuitively, this is easy to envision considering a single wavelength. When perfect synchronization is achieved, the wavelength passes through the filter with minimal attenuation and is recursively amplified till some equilibrium value of output power is reached. In case of a mismatch, the wavelength will undergo larger than minimal attenuation through the filter and the equilibrium value will be smaller. The argument can be extended to a continuum of wavelengths and thus the average energy will peak at perfect synchronization.

2.1 Optical subsystem

Figure 1a
Fig. 1 Schematic diagrams of (a) the self-starting, self-regulating FDML laser and (b) swept source OCT system. PC: polarization controller, ISO: isolator, SOA: semiconductor optical amplifier, OC: optical coupler, PD: photodiode, M: mirror, CIR: circulator, GV: galvanometer mirror, CL: collimating lens, FL: focusing lens, PC: polarization controller, BD: balanced detector, DAQ: data acquisition card
shows a schematic of the SS-FDML source. The FDML loop consisted of a 5-km-long SMF-28e fiber loop, a semiconductor optical amplifier (BOA1132, Covega, MD) with a full width half maximum (FWHM) bandwidth of 90 nm centered at 1305 nm, a tunable filter (Lambda Quest, CA) with a finesse of 700 and a free spectral range (FSR) of 160 nm, a polarization controller, two isolators and a 70/30 coupler OC1 which feeds 30% of the light back into the loop. Based on the optical path length the nominal light transit time is 25 µs leading to a drive frequency around 40 kHz. The timing gate [8

8. R. Huber, M. Wojtkowski, and J. G. Fujimoto, “Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography,” Opt. Express 14(8), 3225–3237 (2006). [CrossRef] [PubMed]

] of the system is ~15 ns. A subsequent off-loop 95/5 coupler OC2 diverted 5% of the light to a photodiode and the rest of the light was sent to the imaging system. The instantaneous output of the photodiode represents the light intensity traversing the FDML loop.

2.2 Electronic subsystem

The photodiode output was sent to a custom-designed energy detector based on a 10 MHz four-quadrant analog multiplier integrated circuit (MPY634, Texas Instruments, TX). The multiplier output was digitized to 16 bits at 800 kS/s using a data acquisition (DAQ) card. The DAQ card also implemented a 16 bit digital-to-analog converter (DAC) which was used to drive a signal generator (33220A, Agilent, CA) configured as a voltage controlled oscillator (VCO) by using the external modulation function. The VCO has an RMS jitter of 3.5 ns which is within the timing gate. The VCO output frequency is given by fc + Δf∙vin where fc and Δf are programmable VCO parameters and the vin is the drive voltage. The VCO outputs a square wave while the tunable filter typically requires a sinusoidal drive waveform. Taking advantage of the fact that a square wave consists only of the fundamental frequency (f) and odd harmonics (3f, 5f ...), we designed a low pass filter to convert it to a sine wave. For stable operation, the filter drive waveform must have very low distortion, requiring a very steep filter roll-off over one half of a decade (f to 3f). Thus, we used a three-stage Sallen-Key filter and implemented a 6th order low pass Butterworth filter with a cutoff frequency of 50 kHz and a 120 dB/decade roll-off. This filter can convert square waves between 35 to 45 kHz to sine waves with a distortion between −48 to −62 dB and a maximum signal attenuation of 1.1 dB.

2.3 Feedback control system

2.4 OCT subsystem

The OCT setup is shown in Fig. 1b and consisted of a 95/5 coupler OC3 which sends 5% of the light to a Mach-Zehnder Interferometer (MZI), a 70/30 coupler OC4 which sends 70% of the light to the sample arm and the rest to the reference arm via a circulator in each branch. The sample arm light was scanned using a galvanometer mirror GV. The light coming back from both arms was recombined using a 50/50 coupler OC5 and sent to a balanced detector BD. The MZI and the balanced detector outputs were sent to a high-speed data acquisition module for digitization, display and storage.

3. Results

3.1 System characterization

Figure 2
Fig. 2 Spectra of filter output for 35, 40 and 45 kHz square wave inputs indicating low distortion square-sine conversion.
illustrates the measured characteristics of the low-pass filter. Square wave inputs and the filter output were observed on an oscilloscope and saved for offline Fourier analysis. The figure shows the power spectra of the filter output at 35, 40 and 45 kHz. The data indicates square-to-sine conversion over the frequency range of interest with a worst case third harmonic distortion of −47.9 dB at 35-kHz square wave input.

In order to evaluate the stability of the system, we monitored the output spectrum of the source every 10 minutes for a period of 2 hours using an optical spectrum analyzer. Figure 5a
Fig. 5 (a) Self-starting, self-regulating FDML (SS-FDML) spectra measured every ten minutes over two hours; (b) Spectral fluctuations with and without feedback over two hours; (c)Effect of temperature variation on conventional and SS-FDML.
shows all the thirteen spectra. For each wavelength, we defined the spectral fluctuation as the ratio of the standard deviation to the mean of the intensity measurements. Figure 5b shows that this data for both the conventional and the SS-FDML. This measurement was done in a controlled laboratory environment leading to a relatively low instability in the conventional FDML However, the self-starting, self-regulating FDML exhibits increased stability due to the feedback control. To further demonstrate the regulating ability of the SS-FDML we used a thermal blanket to wrap the fiber spool and heat the air around it. For a conventional FDML, a 4 centigrade increase caused the RMS energy to drop to ~40% of the initial value and the MZI

signal deteriorated as shown in Fig. 5c. The figure also shows the output for the SS-FDML where the power was maintained at 95% of the initial level. The feedback loop achieves this by sensing the power loss and re-optimizing the drive frequency.

3.2 Optical coherence tomography

We tested the performance of the SS-FDML source for OCT imaging in the setup shown in Fig. 1b and described in Sec. 2.4. Figure 6a
Fig. 6 Point spread function (PSF) measured for a range of imaging depths indicating a 7.5 dB roll-off over 2.3 mm (b) Detail of the PSF showing an 8.9 µm axial resolution.
shows the point spread function (PSF) measured across a 2.3 mm imaging depth indicating a roll-off of 7.5 dB. The spurious peaks seen in the data are due to back reflections in the system. Figure 6b illustrates the detail of one of the PSFs showing an axial resolution of 8.9 µm in air. Measurements also indicated a detection sensitivity of over 120 dB with a power of 10 mW in the sample arm.

OCT imaging of a human fingertip was performed at 40,000 axial scans per second using a handheld scanning probe. Figure 7
Fig. 7 One frame from a 40 fps video sequence acquired using a swept-source OCT system with the self-starting, self-regulating FDML source. Skin layers comprising the epidermis (stratum corneum - SC, stratum lucidum - SL, stratum granulosum - SG, stratum spinosum - SS, stratum germinativum - SG), the dermis and sweat ducts (SD) can be observed.
shows a representative image from a video sequence obtained at 40 frames per second. The imaging field of view is about 3 mm (2048 pixels) lateral × 1.8 mm (1024 pixels) deep. The keratinized dead cells forming the top layer of the epidermis, the stratum corneum, followed by the living cell epidermal layers - stratum lucidum, stratum granulosum, stratum spinosum and stratum germinativum, the superficial dermis and sweat ducts be clearly identified.

4. Summary

Acknowledgments

This work was supported in part by the National Institutes of Health under grants R01CA120480 and R01EB007636.

References and links

1.

J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28(21), 2067–2069 (2003). [CrossRef] [PubMed]

2.

R. Leitgeb, C. Hitzenberger, and A. Fercher, “Performance of Fourier domain vs. time domain optical coherence tomography,” Opt. Express 11(8), 889–894 (2003). [CrossRef] [PubMed]

3.

M. Choma, M. Sarunic, C. Yang, and J. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]

4.

S. R. Chinn, E. A. Swanson, and J. G. Fujimoto, “Optical coherence tomography using a frequency-tunable optical source,” Opt. Lett. 22(5), 340–342 (1997). [CrossRef] [PubMed]

5.

B. Golubovic, B. E. Bouma, G. J. Tearney, and J. G. Fujimoto, “Optical frequency-domain reflectometry using rapid wavelength tuning of a Cr4+:forsterite laser,” Opt. Lett. 22(22), 1704–1706 (1997). [CrossRef] [PubMed]

6.

S. Yun, G. Tearney, J. de Boer, N. Iftimia, and B. Bouma, “High-speed optical frequency-domain imaging,” Opt. Express 11(22), 2953–2963 (2003). [CrossRef] [PubMed]

7.

S. H. Yun, C. Boudoux, G. J. Tearney, and B. E. Bouma, “High-speed wavelength-swept semiconductor laser with a polygon-scanner-based wavelength filter,” Opt. Lett. 28(20), 1981–1983 (2003). [CrossRef] [PubMed]

8.

R. Huber, M. Wojtkowski, and J. G. Fujimoto, “Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography,” Opt. Express 14(8), 3225–3237 (2006). [CrossRef] [PubMed]

9.

M. Gora, K. Karnowski, M. Szkulmowski, B. J. Kaluzny, R. Huber, A. Kowalczyk, and M. Wojtkowski, “Ultra high-speed swept source OCT imaging of the anterior segment of human eye at 200 kHz with adjustable imaging range,” Opt. Express 17(17), 14880–14894 (2009). [CrossRef] [PubMed]

10.

B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010). [CrossRef] [PubMed]

11.

W. Wieser, B. R. Biedermann, T. Klein, C. M. Eigenwillig, and R. Huber, “Multi-megahertz OCT: High quality 3D imaging at 20 million A-scans and 4.5 GVoxels per second,” Opt. Express 18(14), 14685–14704 (2010). [CrossRef] [PubMed]

12.

T. Klein, W. Wieser, C. M. Eigenwillig, B. R. Biedermann, and R. Huber, “Megahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laser,” Opt. Express 19(4), 3044–3062 (2011). [CrossRef] [PubMed]

13.

J. F. Xi, L. Huo, J. S. Li, and X. D. Li, “Generic real-time uniform K-space sampling method for high-speed swept-source optical coherence tomography,” Opt. Express 18(9), 9511–9517 (2010). [CrossRef] [PubMed]

14.

S. Marschall, T. Klein, W. Wieser, B. R. Biedermann, K. Hsu, K. P. Hansen, B. Sumpf, K. H. Hasler, G. Erbert, O. B. Jensen, C. Pedersen, R. Huber, and P. E. Andersen, “Fourier domain mode-locked swept source at 1050 nm based on a tapered amplifier,” Opt. Express 18(15), 15820–15831 (2010). [CrossRef] [PubMed]

OCIS Codes
(110.4500) Imaging systems : Optical coherence tomography
(140.3600) Lasers and laser optics : Lasers, tunable

ToC Category:
Optical Coherence Tomography

History
Original Manuscript: May 12, 2011
Revised Manuscript: June 16, 2011
Manuscript Accepted: June 16, 2011
Published: June 22, 2011

Citation
Kartikeya Murari, Jessica Mavadia, Jiefeng Xi, and Xingde Li, "Self-starting, self-regulating Fourier domain mode locked fiber laser for OCT imaging," Biomed. Opt. Express 2, 2005-2011 (2011)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-2-7-2005


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28(21), 2067–2069 (2003). [CrossRef] [PubMed]
  2. R. Leitgeb, C. Hitzenberger, and A. Fercher, “Performance of Fourier domain vs. time domain optical coherence tomography,” Opt. Express 11(8), 889–894 (2003). [CrossRef] [PubMed]
  3. M. Choma, M. Sarunic, C. Yang, and J. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]
  4. S. R. Chinn, E. A. Swanson, and J. G. Fujimoto, “Optical coherence tomography using a frequency-tunable optical source,” Opt. Lett. 22(5), 340–342 (1997). [CrossRef] [PubMed]
  5. B. Golubovic, B. E. Bouma, G. J. Tearney, and J. G. Fujimoto, “Optical frequency-domain reflectometry using rapid wavelength tuning of a Cr4+:forsterite laser,” Opt. Lett. 22(22), 1704–1706 (1997). [CrossRef] [PubMed]
  6. S. Yun, G. Tearney, J. de Boer, N. Iftimia, and B. Bouma, “High-speed optical frequency-domain imaging,” Opt. Express 11(22), 2953–2963 (2003). [CrossRef] [PubMed]
  7. S. H. Yun, C. Boudoux, G. J. Tearney, and B. E. Bouma, “High-speed wavelength-swept semiconductor laser with a polygon-scanner-based wavelength filter,” Opt. Lett. 28(20), 1981–1983 (2003). [CrossRef] [PubMed]
  8. R. Huber, M. Wojtkowski, and J. G. Fujimoto, “Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography,” Opt. Express 14(8), 3225–3237 (2006). [CrossRef] [PubMed]
  9. M. Gora, K. Karnowski, M. Szkulmowski, B. J. Kaluzny, R. Huber, A. Kowalczyk, and M. Wojtkowski, “Ultra high-speed swept source OCT imaging of the anterior segment of human eye at 200 kHz with adjustable imaging range,” Opt. Express 17(17), 14880–14894 (2009). [CrossRef] [PubMed]
  10. B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010). [CrossRef] [PubMed]
  11. W. Wieser, B. R. Biedermann, T. Klein, C. M. Eigenwillig, and R. Huber, “Multi-megahertz OCT: High quality 3D imaging at 20 million A-scans and 4.5 GVoxels per second,” Opt. Express 18(14), 14685–14704 (2010). [CrossRef] [PubMed]
  12. T. Klein, W. Wieser, C. M. Eigenwillig, B. R. Biedermann, and R. Huber, “Megahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laser,” Opt. Express 19(4), 3044–3062 (2011). [CrossRef] [PubMed]
  13. J. F. Xi, L. Huo, J. S. Li, and X. D. Li, “Generic real-time uniform K-space sampling method for high-speed swept-source optical coherence tomography,” Opt. Express 18(9), 9511–9517 (2010). [CrossRef] [PubMed]
  14. S. Marschall, T. Klein, W. Wieser, B. R. Biedermann, K. Hsu, K. P. Hansen, B. Sumpf, K. H. Hasler, G. Erbert, O. B. Jensen, C. Pedersen, R. Huber, and P. E. Andersen, “Fourier domain mode-locked swept source at 1050 nm based on a tapered amplifier,” Opt. Express 18(15), 15820–15831 (2010). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Supplementary Material


» Media 1: MOV (2183 KB)     

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited