Optoacoustic monitoring of cerebral venous blood oxygenation through extracerebral blood

I. Y. Petrov, Y. Petrov, D. S. Prough, D. J. Deyo, I. Cicenaite, and R. O. Esenaliev  »View Author Affiliations


Biomedical Optics Express, Vol. 3, Issue 1, pp. 125-136 (2012)
http://dx.doi.org/10.1364/BOE.3.000125


View Full Text Article

Acrobat PDF (922 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

There is strong clinical evidence that controlling cerebral venous oxygenation (oxyhemoglobin saturation) is critically important for patients with severe traumatic brain injury as well as for patients undergoing cardiac surgery. However, the only available method for cerebral venous blood oxygenation monitoring is invasive and requires catheterization of the internal jugular vein. We designed and built a novel optoacoustic monitor of cerebral venous oxygenation as measured in the superior sagittal sinus (SSS), the large midline cerebral vein. To the best of our knowledge, optical monitoring of cerebral venous blood oxygenation through overlying extracerebral blood is reported for the first time in this paper. The system was capable of detecting SSS signals in vivo at 700, 800, and 1064 nm through the thick (5–6 mm) sheep skull containing the circulating blood. The high (submillimeter) in-depth resolution of the system provided identification of the SSS peaks in the optoacoustic signals. The SSS peak amplitude closely followed the actual SSS blood oxygenation measured invasively using catheterization, blood sampling, and “gold standard” CO-Oximetry. Our data indicate the system may provide accurate measurement of the SSS blood oxygenation in patients with extracerebral blood over the SSS.

© 2011 OSA

1. Introduction

There is strong clinical evidence that monitoring cerebral venous oxygenation (oxyhemoglobin saturation) can be useful in patients with severe traumatic brain injury and in patients undergoing cardiac surgery because cerebral venous oxygenation reflects cerebral ischemia early and can be used to guide therapeutic interventions [1

1. M. Soehle, M. Jaeger, and J. Meixensberger, “Online assessment of brain tissue oxygen autoregulation in traumatic brain injury and subarachnoid hemorrhage,” Neurol. Res. 25(4), 411–417 (2003). [CrossRef] [PubMed]

,2

2. J. M. Murkin, “Perioperative detection of brain oxygenation and clinical outcomes in cardiac surgery,” Semin. Cardiothorac. Vasc. Anesth. 8(1), 13–14 (2004). [CrossRef] [PubMed]

]. Cerebral venous oxygenation below 50% is associated with death or severe neurologic complications (normal range 55–75%) [3

3. C. Metz, M. Holzschuh, T. Bein, C. Woertgen, R. Rothoerl, B. Kallenbach, K. Taeger, and A. Brawanski, “Monitoring of cerebral oxygen metabolism in the jugular bulb: reliability of unilateral measurements in severe head injury,” J. Cereb. Blood Flow Metab. 18(3), 332–343 (1998). [CrossRef] [PubMed]

]. However, the only clinically useful technique for cerebral venous blood oxygenation monitoring is invasive, requiring catheterization of the internal jugular vein and measurement of blood oxygenation in the jugular bulb [4

4. W. J. Stevens, “Multimodal monitoring: head injury management using SjvO2 and LICOX,” J. Neurosci. Nurs. 36(6), 332–339 (2004). [CrossRef] [PubMed]

]. Another invasive method measures regional brain tissue pO2 (partial oxygen tension) with an intracranial probe inserted directly into the brain parenchyma. Both methods can cause complications due to the invasiveness of the procedures and have technical limitations associated with frequent recalibration. The intraparenchymal probe can measure only local brain tissue pO2, while the catheter in the jugular bulb often provides false values associated with contact with the vessel wall. Near-infrared (NIR) spectroscopy, a non-invasive optical technique, can monitor tissue oxygenation based on detection of light diffusively scattered from a target tissue [5

5. F. F. Jöbsis, “Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters,” Science 198(4323), 1264–1267 (1977). [CrossRef] [PubMed]

,6

6. J. H. Choi, M. Wolf, V. Toronov, U. Wolf, C. Polzonetti, D. Hueber, L. P. Safonova, R. Gupta, A. Michalos, W. Mantulin, and E. Gratton, “Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach,” J. Biomed. Opt. 9(1), 221–229 (2004). [CrossRef] [PubMed]

]; however, NIR spectroscopy measures only volume-averaged oxygenation of brain tissue and cannot distinguish between venous, capillary, and arterial blood.

We proposed to use an optoacoustic technique for monitoring of blood oxygenation including cerebral venous blood oxygenation [7

7. R. O. Esenaliev, K. V. Larin, I. V. Larina, M. Motamedi, and D. S. Prough, “Optoacoustic technique for non-invasive continuous monitoring of blood oxygenation,” in Biomedical Topical Meetings (Optical Society of America, Washington DC, 2000), pp. 272–274.

,8

8. R. O. Esenaliev, I. V. Larina, K. V. Larin, D. J. Deyo, M. Motamedi, and D. S. Prough, “Optoacoustic technique for noninvasive monitoring of blood oxygenation: a feasibility study,” Appl. Opt. 41(22), 4722–4731 (2002). [CrossRef] [PubMed]

]. This technique detects optoacoustic (ultrasound) waves generated in tissue due to absorption of NIR pulses followed by thermo-elastic expansion of the irradiated volume. In vivo experiments conducted by our group [9

9. Y. Y. Petrov, D. S. Prough, D. J. Deyo, M. Klasing, M. Motamedi, and R. O. Esenaliev, “Optoacoustic, noninvasive, real-time, continuous monitoring of cerebral blood oxygenation: an in vivo study in sheep,” Anesthesiology 102(1), 69–75 (2005). [CrossRef] [PubMed]

11

11. H. P. Brecht, D. S. Prough, Y. Y. Petrov, I. Patrikeev, I. Y. Petrova, D. J. Deyo, I. Cicenaite, and R. O. Esenaliev, “In vivo monitoring of blood oxygenation in large veins with a triple-wavelength optoacoustic system,” Opt. Express 15(24), 16261–16269 (2007). [CrossRef] [PubMed]

] confirmed the clinically useful potential of this technique. Recently, we demonstrated high accuracy of optoacoustically measured oxygenation in the superior sagittal sinus (SSS), a large central cerebral vein [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

]. Although photoacoustic tomography was studied for quantitative imaging of oxyhemoglobin saturation in the cerebral vasculature of small animals [13

13. X. Wang, X. Xie, G. Ku, L. V. Wang, and G. Stoica, “Noninvasive imaging of hemoglobin concentration and oxygenation in the rat brain using high-resolution photoacoustic tomography,” J. Biomed. Opt. 11(2), 024015 (2006). [CrossRef] [PubMed]

], the imaging depth in these experiments was not sufficient for human applications. Moreover, the small vessel size in these animals did not allow for validation of cerebral blood vessel oxygenation with “gold standard” invasive methods requiring blood samples.

To the best of our knowledge, accurate monitoring of cerebral venous blood oxygenation through overlying extracerebral blood is reported for the first time in this paper. The in vivo measurements were performed through the thick (5–6 mm) sheep skull using our novel multi-wavelength, optoacoustic system by probing the SSS which is located on top of the brain and collects blood from both hemispheres. In this study we tested the capability of the optoacoustic technique to probe the SSS through extracerebral blood and evaluated accuracy of the SSS blood oxygenation measurements through extracerebral blood. We validated the optoacoustic monitoring of the SSS blood oxygenation through extracerebral blood using the “gold standard” CO-Oximetry. This situation simulated much more challenging and clinically relevant conditions than those studied without diploic and other veins in [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

] because sometimes diploic and other veins can overlie cerebral blood vessels and complicate cerebral blood oxygenation monitoring.

2. Materials and methods

A recently developed multi-wavelength optoacoustic system was used in the study (Fig. 1
Fig. 1 Schematic drawing of the experimental setup used in this study.
). For a source of pulsed tunable NIR radiation, the system uses a compact optical parametric oscillator (OPO; Opolette 532 II, Opotek Inc., Carlsbad, CA) with the following parameters: range of available wavelengths 680–2440 nm; pulse duration, 10 ns; repetition rate, 20 Hz. We developed a highly sensitive optoacoustic probe, which incorporated a broadband (3 MHz) piezoelectric transducer and a 4-fiber light-delivery system. The fibers had a core diameter of 1 mm and were mounted surrounding the transducer. The signals from the probe were amplified and digitized with a 100-MHz digitizer (National Instruments Corp., Austin, TX). Using a specially developed software package and a laptop, the OPO was controlled and the digitized signals were acquired and processed in real time.

The system was calibrated with as a “gold standard” CO-Oximeter (IL 682, Instrumentation Laboratories, Lexington, MA) using a procedure described in detail in [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

]. The calibration was performed using model blood vessels in a tissue phantom with optical properties close to that of tissue in the NIR spectral range at three different wavelengths (700, 800/805, and 1064 nm) in a wide range of blood oxygenation (17–94%). Simultaneously, actual blood oxygenation was measured using blood sampling and the CO-Oximeter. The calibration curve obtained in [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

] was used in this study to predict the SSS blood oxygenation.

The three wavelengths were used because they provide accurate measurement of blood oxygenation due to the strong hemoglobin absorption coefficient dependence on oxygenation at 1064 and 700 nm [14

14. W.-F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron. 26(12), 2166–2185 (1990). [CrossRef]

16

16. Y. Y. Petrov, D. S. Prough, D. J. Deyo, I. Y. Petrova, M. Motamedi, and R. O. Esenaliev, “In vivo noninvasive monitoring of cerebral blood with optoacoustic technique,” in 26th Annual International Conference of IEEE Engineering in Medicine and Biology Society (IEEE, NY, 2004), pp. 2052–2054.

]. At 1064 nm the hemoglobin absorption increases with oxygenation, while at 700 nm it decreases. Moreover, at 1064 nm, both melanin and water have low absorption and scattering in tissues [17

17. S. Jacques, “Optical absorption of melanin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/melanin/mua.html.

20

20. V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis (SPIE Press, Bellingham, WA, 2000).

]. In contrast, 800 nm and 805 nm were used for reference measurements and normalization of signals because they are in the middle of the typical range of 795 nm to 810 nm reported in the literature for the isosbestic point [15

15. B. L. Horecker, “The absorption spectra of hemoglobin and its derivatives in the visible and near infra-red regions,” J. Biol. Chem. 148, 173–183 (1943).

,16

16. Y. Y. Petrov, D. S. Prough, D. J. Deyo, I. Y. Petrova, M. Motamedi, and R. O. Esenaliev, “In vivo noninvasive monitoring of cerebral blood with optoacoustic technique,” in 26th Annual International Conference of IEEE Engineering in Medicine and Biology Society (IEEE, NY, 2004), pp. 2052–2054.

,21

21. S. Prahl, “Optical absorption of hemoglobin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/hemoglobin/index.html.

], where optical absorption of blood does not depend on oxygenation. The difference in hemoglobin absorption at these two wavelengths is very small that allows for using them interchangeably in oxygenation studies.

After performing these calibration studies, we conducted in vivo experiments in 6 adult merino sheep. Optoacoustic signals were detected from the SSS through the intact skull (thickness: 5-6 mm). Although the adult sheep skull is thinner than the adult human skull (8-10 mm), the bone over the SSS is thick enough to demonstrate the capability of optoacoustic technique to measure cerebral blood oxygenation. In addition, the structure of the sheep skull is anatomically similar to that of the human skull and both have areas with increased blood content consisting of diploic veins and/or increased capillary network.

The Institutional Animal Care and Use Committee at the University of Texas Medical Branch (UTMB) approved the protocol for the study. The animals were housed at the UTMB’s Animal Resources Center under daily supervision of full-time veterinarians. During the experiments, the sheep were anesthetized with a 1.5% to 2.0% isoflurane and kept in a prone position. Oral-tracheal intubation was used for the delivery of both isoflurane and a gas mixture of oxygen and nitrogen to the animal.

We were able to change venous blood oxygenation throughout a wide range of 10%–100% by varying the fraction of oxygen (FiO2) in the inhaled gas mixture in the range from 10% to 100%. To generate the gas mixtures varying from 10% to 100% inspired oxygen, we combined medical grade oxygen and nitrogen using precision flow meters to produce a total gas flow rate of 2L per minute that provides an adequate flow for anesthesia. We performed at least 2-3 complete cycles of decreasing SSS blood oxygenation and increasing it in each animal. Some sheep were subjected to up to 5 complete cycles.

The scalp was cut along the midline, reflected to the sides, and the skin and other soft tissues over the skull at the site of measurement were removed before attaching the optoacoustic probe. A small craniotomy was made close to the site of optoacoustic measurements to insert a blood-sampling catheter into the SSS to draw samples immediately after each optoacoustic measurement. A small (5 x 10 mm) burr hole was placed in the skull over the SSS and the SSS was cannulated by direct visualization using PE-50 tubing. The cannula was secured with bone wax and sutures in the scalp. Samples were drawn and blood oxygenation was measured with the CO-Oximeter. Although the diploic and other extracerebral veins are small, sometimes they are visible as darker spots in the skull. When the probe was placed on these areas, the optoacoustic signal had an additional (second) peak after the first peak, which was generated on the skull surface.

During the experiment we continuously monitored the animal’s vital signs. Blood pressure was measured using a catheter inserted into femoral artery, while a pulse oximeter attached to lip, tongue, or ear continuously monitored arterial blood oxygenation. Heart rate and cardiac rhythm were monitored by electrocardiography. At the end of the experiment the sheep was given saturated KCl solution intravenously (about 1 cc/kg) under deep (>4%) isoflurane anesthesia.

A thin layer of ultrasound gel ensured acoustic coupling of the optoacoustic probe with tissue. To obtain SSS signals with the highest amplitude, we scanned the probe using a 3D translation stage. Laser pulse energy was attenuated to provide incident laser fluence of about 4 mJ/cm2, which is well below the maximum permissible exposure for skin (20–100 mJ/cm2) in this spectral range [22

22. “ANSI Z136.1—2000,” in American National Standard for Safe Use of Lasers (The Laser Institute of America, Orlando, FL, 2000).

].

At each oxygenation level we consecutively acquired optoacoustic signals at 700 nm, 800 nm, and 1064 nm. Every recorded signal was an average of 400 acquired signals to increase signal-to-noise ratio. A blood sample was drawn from the SSS after each set of the three-wavelength measurements that took approximately 1.5–2 min.

3. Results and discussion

All the signals in Fig. 2 were normalized to the amplitude of the surface peak to minimize influence on the SSS oxygenation measurements of instability in acoustic contact and in the OPO pulse energy. Although, the optoacoustic signal amplitude depends on the light fluence on the skull surface and absorption coefficient at the employed wavelengths, the blood content of the upper dense layer of the skull bone (the outer table) is extremely low. It implies that the changes in the amplitude of the surface peak were not produced by the variations of blood oxygenation, but rather by the instability of acoustic contact and incident light fluence. Hence, the normalization of the signals by this amplitude removes this dependence and increases the accuracy of the SSS oxygenation measurements. The optical absorption of the skull is wavelength-dependent, but it does not influence substantially the accuracy of the SSS blood oxygenation measurements, because the bone absorption has a weak dependence at these wavelengths [23

23. P. Taroni, D. Comelli, A. Farina, A. Pifferi, and A. Kienle, “Time-resolved diffuse optical spectroscopy of small tissue samples,” Opt. Express 15(6), 3301–3311 (2007). [CrossRef] [PubMed]

]. The peak-to-peak SSS signal amplitude measured at 700 nm decreased with the SSS oxygenation. At 1064 nm, the opposite dependence was valid: the amplitude increased with the SSS oxygenation. Since 805 nm is in the isosbestic range in the near-infrared spectrum of whole blood absorption [15

15. B. L. Horecker, “The absorption spectra of hemoglobin and its derivatives in the visible and near infra-red regions,” J. Biol. Chem. 148, 173–183 (1943).

,16

16. Y. Y. Petrov, D. S. Prough, D. J. Deyo, I. Y. Petrova, M. Motamedi, and R. O. Esenaliev, “In vivo noninvasive monitoring of cerebral blood with optoacoustic technique,” in 26th Annual International Conference of IEEE Engineering in Medicine and Biology Society (IEEE, NY, 2004), pp. 2052–2054.

,21

21. S. Prahl, “Optical absorption of hemoglobin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/hemoglobin/index.html.

], the optical absorption of blood does not depend on oxygenation and the amplitude of the SSS peak at 805 nm was changing due to hemodynamic phenomena described below.

Despite the presence of the diploic vein that absorbed some amount of light, the signal from the SSS was strong and easily identifiable. The fact that the SSS signal was detectable through the blood above the SSS is very encouraging from clinical point of view. In patients with head trauma, the brain may be inadequately perfused, while extracerebral blood oxygenation can be normal. Therefore it is important to detect these two blood vessels separately. Figure 3 demonstrates that the optoacoustic technique is capable of separating the signals generated in extracerebral and cerebral blood.

At the same time, hypoxia triggers splenic contraction. This process is much slower than vasodilation and results in an additional red blood cell (RBC) release into the circulation [26

26. I. Kuwahira, U. Kamiya, T. Iwamoto, Y. Moue, T. Urano, Y. Ohta, and N. C. Gonzalez, “Splenic contraction-induced reversible increase in hemoglobin concentration in intermittent hypoxia,” J. Appl. Physiol. 86(1), 181–187 (1999). [PubMed]

28

28. T. Oka, T. Itoi, and K. Hamaoka, “Impaired transient elevation of blood hemoglobin in response to acute hypoxia in neonates with asplenia,” Pediatr. Int. 49(6), 898–902 (2007). [CrossRef] [PubMed]

]. After splenic contraction, the oxygen delivery to organs is maintained by the increase of the blood hemoglobin concentration, and the regional blood flow re-distribution returns to normal. As the absorption coefficient of blood is linearly dependent on the concentration of hemoglobin, this transient elevation of blood hemoglobin concentration manifests itself in the higher SSS signal amplitude. When the hypoxic state ends, the spleen again sequesters excessive RBCs, and the SSS signal decreases.

Figure 7(a)
Fig. 7 Total hemoglobin concentration (red line with dots, left axis) and the peak-to-peak SSS signal amplitude at 800 nm (green line with dots, right axis). The data are for two different sheep: (a) and (b), respectively.
shows the same [THb] trend as in Fig. 6(a), superimposed now with the peak-to-peak SSS signal amplitude measured at 800 nm. The correlation of these two parameters is easily seen, although the [THb] trend was somewhat distorted due to the reasons of infused Ringer’s lactate as discussed above. However, the correlation was sometimes almost ideal as shown in Fig. 7(b) (it shows the data obtained during one cycle of FiO2 change in another sheep).

Another reason for changes of the SSS peak amplitude at 800 nm can be gradual distortion of the alignment of the optoacoustic probe with the SSS. Although the sheep’s head was fixed in a custom-made head frame, some motion due to cranial muscle contraction could lead to a small displacement of the probe from the best-aligned position and distort the signal.

The optoacoustic signals were acquired at the three wavelengths as rapidly as possible within one set of measurements to keep them close in time and thus, to provide similar measurement conditions. In this case, the three confounding factors discussed above had similar influence on the signals measured at all the three wavelengths. This allowed a significant increase in measurement accuracy by normalizing the SSS signal amplitudes at 700 nm and 1064 nm using the amplitude at 800 nm. The normalized peak-to-peak SSS signal amplitudes were calculated as the ratios of the amplitudes measured at 700 nm or 1064 nm to those measured at 800 nm (Fig. 8
Fig. 8 Invasively measured SSS blood oxygenation (black line with dots) and the ratios of the SSS signal amplitudes (colored line with triangles) measured at 700 nm (a) and 1064 nm (b) to those measured at 800 nm.
). The correlation between the amplitude ratio and actual SSS blood oxygenation measured invasively is substantially higher compared to that for the original amplitudes (Fig. 5 ((a) and (c)), colored lines). Figure 9
Fig. 9 Correlation of the ratios of the SSS signal amplitudes at 700 nm (blue triangles) and 1064 nm (red dots) to those measured at 800 nm with actual SSS blood oxygenation. The lines of the corresponding color are linear fit to the data sets (R2 = 0.71 and 0.91 for 700 nm and 1064 nm, respectively).
demonstrates that the amplitude ratio is linearly dependent on blood oxygenation (data for the fourth cycle from Fig. 8 are presented). The lines of the corresponding color fit the data point sets linearly. The correlation coefficients are much higher, R2 = 0.71 and 0.91 at 700 nm and 1064 nm, respectively, while before the normalization they were 0.569 and 0.776, respectively (data not shown).

We predicted the SSS blood oxygenation values and assessed accuracy of the optoacoustic measurements. Figure 10(a)
Fig. 10 (a) Correlation between optoacoustically predicted and actual SSS blood oxygenation. (b) Standard deviation and bias of the difference between predicted and actual oxygenation.
shows high correlation between the optoacoustically predicted and actual SSS blood oxygenation measured with the CO-Oximeter (R2 = 0.91). To predict the SSS blood oxygenation, we calculated the ratio of the SSS peak amplitude measured at 1064 nm to that measured at 700 nm and then used the calibration curve reported in [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

]. The calibration curve is the dependence of the ratio of the signal amplitude measured from sheep blood in the phantom at 1064 nm to that measured at 700 nm vs. blood oxygenation.

To assess accuracy of the optoacoustic measurements, we calculated the bias and standard deviation using the difference between the optoacoustically predicted and actual oxygenation measured with the CO-Oximeter (Fig. 10(b)). These data demonstrate that, despite the presence of the extracerebral blood, the accuracy of the optoacoustically predicted SSS blood oxygenation approaches that of measured without extracerebral blood that was reported in [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

] (R2 = 0.91 vs. 0.965, bias = 2.0% vs. −9.3%, and SD = 8.5% vs. 4.2%, respectively).

Although, no tissue phantom can provide perfect calibration for in vivo measurements, the calibration obtained from blood in vitro in the tissue phantom in [12

12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

] allowed for accurate prediction of SSS blood oxygenation. This is because: 1) the phantom had optical properties (the effective attenuation coefficient and the spectral dependence of the attenuation coefficient) close to that of tissue; 2) the blood signal amplitude was measured in vivo at 1064 nm and 700 nm within a short time to minimize changes in the tissue optical properties during the measurement; and 3) although the diploic and other extracerebral veins produce a noticeable signal, they are small and their effect on the fluence attenuation is not strong. The veins produce a noticeable signal because they are close to the irradiated surface and fluence at these depths is not strongly attenuated. The data presented in Figs. 10(a) and 10(b) on accuracy of the optoacoustic measurement of SSS blood oxygenation through the extracerebral blood confirm that the diploic and other extracerebral veins are not optically thick and do not substantially reduce accuracy of the SSS blood oxygenation measurements.

Since we use the ratio measurements for predicting the SSS blood oxygenation, the acoustic wave aberration and attenuation effects on the accuracy of the SSS blood oxygenation measurements are minimal. Acoustic reverberations do not influence the accuracy of the SSS oxygenation measurements because they arrive at the optoacoustic probe later than the SSS signal due to longer acoustic path. Moreover, the reverberation signal amplitude is low due to stronger acoustic attenuation in the bone over the longer path.

To avoid the lateral scanning of the optoacoustic probe, one can use optoacoustic arrays for fast detection of the SSS signals. Recently developed optoacoustic arrays that provide optoacoustic images deeply in highly scattering media and tissues in vivo [29

29. K. Passler, R. Nuster, S. Gratt, P. Burgholzer, and G. Paltauf, “Piezoelectric annular array for large depth of field photoacoustic imaging,” Biomed. Opt. Express 2(9), 2655–2664 (2011). [CrossRef] [PubMed]

32

32. J. Jose, R. G. H. Willemink, S. Resink, D. Piras, J. C. G. van Hespen, C. H. Slump, W. Steenbergen, T. G. van Leeuwen, and S. Manohar, “Passive element enriched photoacoustic computed tomography (PER PACT) for simultaneous imaging of acoustic propagation properties and light absorption,” Opt. Express 19(3), 2093–2104 (2011). [CrossRef] [PubMed]

] can be adapted for the SSS signal detection. This will further improve system performance and minimize influence of motion artifacts.

4. Conclusions

This study demonstrated that the multi-wavelength optoacoustic system provides accurate monitoring of cerebral venous blood oxygenation through overlying extracerebral blood. The system is capable of detecting SSS signals in vivo through the thick sheep skull containing circulating blood. The high in-depth resolution of the system provided easy identification of the SSS peaks in the optoacoustic signals. The SSS peak amplitude closely followed the actual SSS blood oxygenation measured invasively using catheterization, blood sampling, and “gold standard” CO-Oximetry. Our data indicate that this system may provide accurate measurement of the SSS blood oxygenation in patients.

Acknowledgments

This work is supported in part by the National Institutes of Health (Grants R01 EB00763 and U54EB007954 from the National Institute of Biomedical Imaging and Bioengineering, Grant R01 NS044345 from the National Institute of Neurological Disorders and Stroke, Grant R41HL10309501 from the National Heart, Lung and Blood Institute), the Moody Center for Brain and Spinal Cord Injury Research/Mission Connect of UTMB, Texas Emerging Technology Fund, and UTMB Seed Grant Program. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIBIB or NIH. Drs. Prough and Esenaliev are co-owners of Noninvasix, Inc., a UTMB-based startup that has licensed the rights to optoacoustic monitoring technology.

References and links

1.

M. Soehle, M. Jaeger, and J. Meixensberger, “Online assessment of brain tissue oxygen autoregulation in traumatic brain injury and subarachnoid hemorrhage,” Neurol. Res. 25(4), 411–417 (2003). [CrossRef] [PubMed]

2.

J. M. Murkin, “Perioperative detection of brain oxygenation and clinical outcomes in cardiac surgery,” Semin. Cardiothorac. Vasc. Anesth. 8(1), 13–14 (2004). [CrossRef] [PubMed]

3.

C. Metz, M. Holzschuh, T. Bein, C. Woertgen, R. Rothoerl, B. Kallenbach, K. Taeger, and A. Brawanski, “Monitoring of cerebral oxygen metabolism in the jugular bulb: reliability of unilateral measurements in severe head injury,” J. Cereb. Blood Flow Metab. 18(3), 332–343 (1998). [CrossRef] [PubMed]

4.

W. J. Stevens, “Multimodal monitoring: head injury management using SjvO2 and LICOX,” J. Neurosci. Nurs. 36(6), 332–339 (2004). [CrossRef] [PubMed]

5.

F. F. Jöbsis, “Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters,” Science 198(4323), 1264–1267 (1977). [CrossRef] [PubMed]

6.

J. H. Choi, M. Wolf, V. Toronov, U. Wolf, C. Polzonetti, D. Hueber, L. P. Safonova, R. Gupta, A. Michalos, W. Mantulin, and E. Gratton, “Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach,” J. Biomed. Opt. 9(1), 221–229 (2004). [CrossRef] [PubMed]

7.

R. O. Esenaliev, K. V. Larin, I. V. Larina, M. Motamedi, and D. S. Prough, “Optoacoustic technique for non-invasive continuous monitoring of blood oxygenation,” in Biomedical Topical Meetings (Optical Society of America, Washington DC, 2000), pp. 272–274.

8.

R. O. Esenaliev, I. V. Larina, K. V. Larin, D. J. Deyo, M. Motamedi, and D. S. Prough, “Optoacoustic technique for noninvasive monitoring of blood oxygenation: a feasibility study,” Appl. Opt. 41(22), 4722–4731 (2002). [CrossRef] [PubMed]

9.

Y. Y. Petrov, D. S. Prough, D. J. Deyo, M. Klasing, M. Motamedi, and R. O. Esenaliev, “Optoacoustic, noninvasive, real-time, continuous monitoring of cerebral blood oxygenation: an in vivo study in sheep,” Anesthesiology 102(1), 69–75 (2005). [CrossRef] [PubMed]

10.

Y. Y. Petrov, I. Y. Petrova, I. A. Patrikeev, R. O. Esenaliev, and D. S. Prough, “Multiwavelength optoacoustic system for noninvasive monitoring of cerebral venous oxygenation: a pilot clinical test in the internal jugular vein,” Opt. Lett. 31(12), 1827–1829 (2006). [CrossRef] [PubMed]

11.

H. P. Brecht, D. S. Prough, Y. Y. Petrov, I. Patrikeev, I. Y. Petrova, D. J. Deyo, I. Cicenaite, and R. O. Esenaliev, “In vivo monitoring of blood oxygenation in large veins with a triple-wavelength optoacoustic system,” Opt. Express 15(24), 16261–16269 (2007). [CrossRef] [PubMed]

12.

I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express 17(9), 7285–7294 (2009). [CrossRef] [PubMed]

13.

X. Wang, X. Xie, G. Ku, L. V. Wang, and G. Stoica, “Noninvasive imaging of hemoglobin concentration and oxygenation in the rat brain using high-resolution photoacoustic tomography,” J. Biomed. Opt. 11(2), 024015 (2006). [CrossRef] [PubMed]

14.

W.-F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron. 26(12), 2166–2185 (1990). [CrossRef]

15.

B. L. Horecker, “The absorption spectra of hemoglobin and its derivatives in the visible and near infra-red regions,” J. Biol. Chem. 148, 173–183 (1943).

16.

Y. Y. Petrov, D. S. Prough, D. J. Deyo, I. Y. Petrova, M. Motamedi, and R. O. Esenaliev, “In vivo noninvasive monitoring of cerebral blood with optoacoustic technique,” in 26th Annual International Conference of IEEE Engineering in Medicine and Biology Society (IEEE, NY, 2004), pp. 2052–2054.

17.

S. Jacques, “Optical absorption of melanin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/melanin/mua.html.

18.

G. M. Hale and M. R. Querry, “Optical constants of water in the 200-nm to 200-µm wavelength region,” Appl. Opt. 12(3), 555–563 (1973). [CrossRef] [PubMed]

19.

C. R. Simpson, M. Kohl, M. Essenpreis, and M. Cope, “Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the Monte Carlo inversion technique,” Phys. Med. Biol. 43(9), 2465–2478 (1998). [CrossRef] [PubMed]

20.

V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis (SPIE Press, Bellingham, WA, 2000).

21.

S. Prahl, “Optical absorption of hemoglobin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/hemoglobin/index.html.

22.

“ANSI Z136.1—2000,” in American National Standard for Safe Use of Lasers (The Laser Institute of America, Orlando, FL, 2000).

23.

P. Taroni, D. Comelli, A. Farina, A. Pifferi, and A. Kienle, “Time-resolved diffuse optical spectroscopy of small tissue samples,” Opt. Express 15(6), 3301–3311 (2007). [CrossRef] [PubMed]

24.

F. J. Fry and J. E. Barger, “Acoustical properties of the human skull,” J. Acoust. Soc. Am. 63(5), 1576–1590 (1978). [CrossRef] [PubMed]

25.

J. Enderle, S. Blanchard, and J. Bronzino, Introduction to Biomedical Engineering (Academic, San Diego, CA, 2000), Chap. 15.

26.

I. Kuwahira, U. Kamiya, T. Iwamoto, Y. Moue, T. Urano, Y. Ohta, and N. C. Gonzalez, “Splenic contraction-induced reversible increase in hemoglobin concentration in intermittent hypoxia,” J. Appl. Physiol. 86(1), 181–187 (1999). [PubMed]

27.

M. X. Richardson, R. de Bruijn, and E. Schagatay, “Hypoxia augments apnea-induced increase in hemoglobin concentration and hematocrit,” Eur. J. Appl. Physiol. 105(1), 63–68 (2009). [CrossRef] [PubMed]

28.

T. Oka, T. Itoi, and K. Hamaoka, “Impaired transient elevation of blood hemoglobin in response to acute hypoxia in neonates with asplenia,” Pediatr. Int. 49(6), 898–902 (2007). [CrossRef] [PubMed]

29.

K. Passler, R. Nuster, S. Gratt, P. Burgholzer, and G. Paltauf, “Piezoelectric annular array for large depth of field photoacoustic imaging,” Biomed. Opt. Express 2(9), 2655–2664 (2011). [CrossRef] [PubMed]

30.

V. G. Andreev, Y. Y. Petrov, D. S. Prough, I. Y. Petrova, and R. O. Esenaliev, “Novel optoacoustic array for noninvasive monitoring of blood parameters,” Proc. SPIE 7177, 71770O, 71770O-6 (2009). [CrossRef]

31.

A. Taruttis, E. Herzog, D. Razansky, and V. Ntziachristos, “Real-time imaging of cardiovascular dynamics and circulating gold nanorods with multispectral optoacoustic tomography,” Opt. Express 18(19), 19592–19602 (2010). [CrossRef] [PubMed]

32.

J. Jose, R. G. H. Willemink, S. Resink, D. Piras, J. C. G. van Hespen, C. H. Slump, W. Steenbergen, T. G. van Leeuwen, and S. Manohar, “Passive element enriched photoacoustic computed tomography (PER PACT) for simultaneous imaging of acoustic propagation properties and light absorption,” Opt. Express 19(3), 2093–2104 (2011). [CrossRef] [PubMed]

OCIS Codes
(120.3890) Instrumentation, measurement, and metrology : Medical optics instrumentation
(170.1460) Medical optics and biotechnology : Blood gas monitoring
(170.1610) Medical optics and biotechnology : Clinical applications
(170.4580) Medical optics and biotechnology : Optical diagnostics for medicine
(170.6510) Medical optics and biotechnology : Spectroscopy, tissue diagnostics

ToC Category:
Cardiovascular Applications

History
Original Manuscript: November 14, 2011
Revised Manuscript: December 8, 2011
Manuscript Accepted: December 12, 2011
Published: December 14, 2011

Citation
I. Y. Petrov, Y. Petrov, D. S. Prough, D. J. Deyo, I. Cicenaite, and R. O. Esenaliev, "Optoacoustic monitoring of cerebral venous blood oxygenation through extracerebral blood," Biomed. Opt. Express 3, 125-136 (2012)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-3-1-125


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. M. Soehle, M. Jaeger, and J. Meixensberger, “Online assessment of brain tissue oxygen autoregulation in traumatic brain injury and subarachnoid hemorrhage,” Neurol. Res.25(4), 411–417 (2003). [CrossRef] [PubMed]
  2. J. M. Murkin, “Perioperative detection of brain oxygenation and clinical outcomes in cardiac surgery,” Semin. Cardiothorac. Vasc. Anesth.8(1), 13–14 (2004). [CrossRef] [PubMed]
  3. C. Metz, M. Holzschuh, T. Bein, C. Woertgen, R. Rothoerl, B. Kallenbach, K. Taeger, and A. Brawanski, “Monitoring of cerebral oxygen metabolism in the jugular bulb: reliability of unilateral measurements in severe head injury,” J. Cereb. Blood Flow Metab.18(3), 332–343 (1998). [CrossRef] [PubMed]
  4. W. J. Stevens, “Multimodal monitoring: head injury management using SjvO2 and LICOX,” J. Neurosci. Nurs.36(6), 332–339 (2004). [CrossRef] [PubMed]
  5. F. F. Jöbsis, “Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters,” Science198(4323), 1264–1267 (1977). [CrossRef] [PubMed]
  6. J. H. Choi, M. Wolf, V. Toronov, U. Wolf, C. Polzonetti, D. Hueber, L. P. Safonova, R. Gupta, A. Michalos, W. Mantulin, and E. Gratton, “Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach,” J. Biomed. Opt.9(1), 221–229 (2004). [CrossRef] [PubMed]
  7. R. O. Esenaliev, K. V. Larin, I. V. Larina, M. Motamedi, and D. S. Prough, “Optoacoustic technique for non-invasive continuous monitoring of blood oxygenation,” in Biomedical Topical Meetings (Optical Society of America, Washington DC, 2000), pp. 272–274.
  8. R. O. Esenaliev, I. V. Larina, K. V. Larin, D. J. Deyo, M. Motamedi, and D. S. Prough, “Optoacoustic technique for noninvasive monitoring of blood oxygenation: a feasibility study,” Appl. Opt.41(22), 4722–4731 (2002). [CrossRef] [PubMed]
  9. Y. Y. Petrov, D. S. Prough, D. J. Deyo, M. Klasing, M. Motamedi, and R. O. Esenaliev, “Optoacoustic, noninvasive, real-time, continuous monitoring of cerebral blood oxygenation: an in vivo study in sheep,” Anesthesiology102(1), 69–75 (2005). [CrossRef] [PubMed]
  10. Y. Y. Petrov, I. Y. Petrova, I. A. Patrikeev, R. O. Esenaliev, and D. S. Prough, “Multiwavelength optoacoustic system for noninvasive monitoring of cerebral venous oxygenation: a pilot clinical test in the internal jugular vein,” Opt. Lett.31(12), 1827–1829 (2006). [CrossRef] [PubMed]
  11. H. P. Brecht, D. S. Prough, Y. Y. Petrov, I. Patrikeev, I. Y. Petrova, D. J. Deyo, I. Cicenaite, and R. O. Esenaliev, “In vivo monitoring of blood oxygenation in large veins with a triple-wavelength optoacoustic system,” Opt. Express15(24), 16261–16269 (2007). [CrossRef] [PubMed]
  12. I. Y. Petrova, Y. Y. Petrov, R. O. Esenaliev, D. J. Deyo, I. Cicenaite, and D. S. Prough, “Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system,” Opt. Express17(9), 7285–7294 (2009). [CrossRef] [PubMed]
  13. X. Wang, X. Xie, G. Ku, L. V. Wang, and G. Stoica, “Noninvasive imaging of hemoglobin concentration and oxygenation in the rat brain using high-resolution photoacoustic tomography,” J. Biomed. Opt.11(2), 024015 (2006). [CrossRef] [PubMed]
  14. W.-F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron.26(12), 2166–2185 (1990). [CrossRef]
  15. B. L. Horecker, “The absorption spectra of hemoglobin and its derivatives in the visible and near infra-red regions,” J. Biol. Chem.148, 173–183 (1943).
  16. Y. Y. Petrov, D. S. Prough, D. J. Deyo, I. Y. Petrova, M. Motamedi, and R. O. Esenaliev, “In vivo noninvasive monitoring of cerebral blood with optoacoustic technique,” in 26th Annual International Conference of IEEE Engineering in Medicine and Biology Society (IEEE, NY, 2004), pp. 2052–2054.
  17. S. Jacques, “Optical absorption of melanin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/melanin/mua.html .
  18. G. M. Hale and M. R. Querry, “Optical constants of water in the 200-nm to 200-µm wavelength region,” Appl. Opt.12(3), 555–563 (1973). [CrossRef] [PubMed]
  19. C. R. Simpson, M. Kohl, M. Essenpreis, and M. Cope, “Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the Monte Carlo inversion technique,” Phys. Med. Biol.43(9), 2465–2478 (1998). [CrossRef] [PubMed]
  20. V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis (SPIE Press, Bellingham, WA, 2000).
  21. S. Prahl, “Optical absorption of hemoglobin,” Oregon Medical Laser Center, http://omlc.ogi.edu/spectra/hemoglobin/index.html .
  22. “ANSI Z136.1—2000,” in American National Standard for Safe Use of Lasers (The Laser Institute of America, Orlando, FL, 2000).
  23. P. Taroni, D. Comelli, A. Farina, A. Pifferi, and A. Kienle, “Time-resolved diffuse optical spectroscopy of small tissue samples,” Opt. Express15(6), 3301–3311 (2007). [CrossRef] [PubMed]
  24. F. J. Fry and J. E. Barger, “Acoustical properties of the human skull,” J. Acoust. Soc. Am.63(5), 1576–1590 (1978). [CrossRef] [PubMed]
  25. J. Enderle, S. Blanchard, and J. Bronzino, Introduction to Biomedical Engineering (Academic, San Diego, CA, 2000), Chap. 15.
  26. I. Kuwahira, U. Kamiya, T. Iwamoto, Y. Moue, T. Urano, Y. Ohta, and N. C. Gonzalez, “Splenic contraction-induced reversible increase in hemoglobin concentration in intermittent hypoxia,” J. Appl. Physiol.86(1), 181–187 (1999). [PubMed]
  27. M. X. Richardson, R. de Bruijn, and E. Schagatay, “Hypoxia augments apnea-induced increase in hemoglobin concentration and hematocrit,” Eur. J. Appl. Physiol.105(1), 63–68 (2009). [CrossRef] [PubMed]
  28. T. Oka, T. Itoi, and K. Hamaoka, “Impaired transient elevation of blood hemoglobin in response to acute hypoxia in neonates with asplenia,” Pediatr. Int.49(6), 898–902 (2007). [CrossRef] [PubMed]
  29. K. Passler, R. Nuster, S. Gratt, P. Burgholzer, and G. Paltauf, “Piezoelectric annular array for large depth of field photoacoustic imaging,” Biomed. Opt. Express2(9), 2655–2664 (2011). [CrossRef] [PubMed]
  30. V. G. Andreev, Y. Y. Petrov, D. S. Prough, I. Y. Petrova, and R. O. Esenaliev, “Novel optoacoustic array for noninvasive monitoring of blood parameters,” Proc. SPIE7177, 71770O, 71770O-6 (2009). [CrossRef]
  31. A. Taruttis, E. Herzog, D. Razansky, and V. Ntziachristos, “Real-time imaging of cardiovascular dynamics and circulating gold nanorods with multispectral optoacoustic tomography,” Opt. Express18(19), 19592–19602 (2010). [CrossRef] [PubMed]
  32. J. Jose, R. G. H. Willemink, S. Resink, D. Piras, J. C. G. van Hespen, C. H. Slump, W. Steenbergen, T. G. van Leeuwen, and S. Manohar, “Passive element enriched photoacoustic computed tomography (PER PACT) for simultaneous imaging of acoustic propagation properties and light absorption,” Opt. Express19(3), 2093–2104 (2011). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited