Multispectral image alignment using a three channel endoscope in vivo during minimally invasive surgery

Neil T. Clancy, Danail Stoyanov, David R. C. James, Aimee Di Marco, Vincent Sauvage, James Clark, Guang-Zhong Yang, and Daniel S. Elson  »View Author Affiliations


Biomedical Optics Express, Vol. 3, Issue 10, pp. 2567-2578 (2012)
http://dx.doi.org/10.1364/BOE.3.002567


View Full Text Article

Acrobat PDF (2035 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Sequential multispectral imaging is an acquisition technique that involves collecting images of a target at different wavelengths, to compile a spectrum for each pixel. In surgical applications it suffers from low illumination levels and motion artefacts. A three-channel rigid endoscope system has been developed that allows simultaneous recording of stereoscopic and multispectral images. Salient features on the tissue surface may be tracked during the acquisition in the stereo cameras and, using multiple camera triangulation techniques, this information used to align the multispectral images automatically even though the tissue or camera is moving. This paper describes a detailed validation of the set-up in a controlled experiment before presenting the first in vivo use of the device in a porcine minimally invasive surgical procedure. Multispectral images of the large bowel were acquired and used to extract the relative concentration of haemoglobin in the tissue despite motion due to breathing during the acquisition. Using the stereoscopic information it was also possible to overlay the multispectral information on the reconstructed 3D surface. This experiment demonstrates the ability of this system for measuring blood perfusion changes in the tissue during surgery and its potential use as a platform for other sequential imaging modalities.

© 2012 OSA

1. Introduction

The acquisition of a stack of multispectral images may take several hundred milliseconds or longer, depending on the exposure time of the camera, the number of wavelengths required and the switching time of the filter. During endoscopic investigation or surgery the camera and tissue may also change position, orientation and shape during the acquisition, since the surgeon holds the endoscope, the patient breathes and peristalsis occurs. This poses a problem for this technique, as the multispectral image stack must be aligned in order to extract information on the spatial variation in chromophore concentration. To deal with the motion problem, a number of solutions have been proposed including gating of image acquisition to the cardiac cycle using ECG [15

15. S. P. Nighswander-Rempel, R. A. Shaw, V. V. Kupriyanov, J. Rendell, B. Xiang, and H. H. Mantsch, “Mapping tissue oxygenation in the beating heart with near-infrared spectroscopic imaging,” Vib. Spectrosc. 32(1), 85–94 (2003). [CrossRef]

]. However, this approach is not suitable for tissues where the dominant source of motion may be a less predictable source such as those mentioned earlier. Alternatively, ‘snapshot’ systems exist which aim to capture all the wavelength data simultaneously using a Wollaston prism or image slicer, and a large format CCD [16

16. L. Gao, R. T. Kester, N. Hagen, and T. S. Tkaczyk, “Snapshot Image Mapping Spectrometer (IMS) with high sampling density for hyperspectral microscopy,” Opt. Express 18(14), 14330–14344 (2010). [CrossRef] [PubMed]

,17

17. A. Gorman, D. W. Fletcher-Holmes, and A. R. Harvey, “Generalization of the Lyot filter and its application to snapshot spectral imaging,” Opt. Express 18(6), 5602–5608 (2010). [CrossRef] [PubMed]

]. While these systems overcome the motion problem, they do so at the expense of spatial resolution and are incompatible with some sequential imaging modalities such as polarisation imaging.

2. Materials and methods

The imaging system is built on a customised trinocular (three channel) rigid endoscope (Intuitive Surgical, Inc., Sunnyvale, USA) with computer vision techniques developed for image alignment.

2.1. Three channel endoscope

The optical design of the endoscope is shown in Fig. 1 (a)
Fig. 1 (a) Optical set-up of the system [22]. (b) Photograph of the experimental arrangement.
, consisting of a pair of outputs for stereo viewing and a wide-angle central channel. A pair of 12.7 mm diameter, 50 mm focal length lenses were used to image onto the colour stereo cameras (IDS Imaging, uEye 2230-C). A 45° plane mirror was used to reflect the central channel to the multispectral optics. A 25.4 mm diameter, 75 mm focal length lens was used to provide higher magnification in the wide-angle central channel in order to set its field of view to be approximately the same as the stereo channels. The wider diameter lens also improves light collection efficiency from what is already a low light transmission channel due to its comparatively small diameter. The light source was a xenon lamp (xenon 300; Karl Storz GmbH) and was connected to the endoscope with a fibre-optic light cable.

The LCTF (Varispec, CRI, Inc) was mounted after the 75 mm focal length lens along with a monochrome camera (Thorlabs DCU 223M). The LCTF has a spectral range running from 400 to 720 nm and a resolution (FWHM) that varies from 7 nm at 400 nm to 20 nm at 720 nm. A custom-written LabVIEW (National Instruments Corporation, USA) program was used to synchronise the multispectral camera with the LCTF so that an image was saved each time the filter tuned over a user-specified wavelength range and resolution, with an image saved by the camera at each step. Camera properties such as integration time, frame rate and gain could also be set from within the program. Data transfer between PC and devices was conducted using the USB 2.0 interface. Colour images were recorded at a resolution of 1024 × 768 pixels while 2 × 2 binning was used to improve the signal to noise ratio in the multispectral camera (512 × 384 pixels). At a working distance of 5 cm, the spatial resolution of the multispectral camera was approximately 0.7 mm.

Since the principle of the alignment procedure relies on tracking features that are moving in time, the images from each camera must be acquired simultaneously. Due to the low transmission of the LCTF, a long integration time is required for the multispectral camera, making this the speed-limiting element. The cameras were synchronised using a programmable digital output (NI USB-9472, National Instruments Corporation, USA), providing a 9 V square wave (T = 1 s) to their digital inputs. The software for controlling the three cameras and LCTF was combined in one program, while the trigger signal was generated from a separate LabVIEW program running in parallel. For the in vivo experiment, the integration time and gain were set to 700 ms and 25 respectively for the multispectral camera, and 90 ms and 20 for the stereo cameras. For the colour-checker card experiment, the increased reflectivity of the surface meant that a lower gain setting could be afforded.

2.2. Stereo reconstruction and multispectral projection

In this paper, a number of computer vision techniques developed for 3D reconstruction using stereoscopic cameras [23

23. D. Stoyanov, A. Darzi, and G.-Z. Yang, “Dense 3D depth recovery for soft tissue deformation during robotically assisted laparoscopic surgery,” in MICCAI, C. Barillot, D. R. Haynor, and P. Hellier, eds. (Springer-Verlag, Rennes/Saint-Malo, France, 2004), pp. 41–48.

] and tracking the motion of tissue in vivo [24

24. D. Stoyanov, G. P. Mylonas, F. Deligianni, A. Darzi, and G.-Z. Yang, “Soft-tissue motion tracking and structure estimation for robotic assisted MIS procedures,” in MICCAI, J. Duncan, and G. Gerig, eds. (Springer-Verlag, Palm Springs, USA, 2005), pp. 139–146.

,25

25. D. Stoyanov and G.-Z. Yang, “Stabilization of image motion for robotic assisted beating heart surgery,” in MICCAI'07 Proceedings of the 10th international conference on Medical image computing and computer-assisted intervention, N. Ayache, S. Ourselin, and A. Maeder, eds. (Springer-Verlag, Brisbane, Australia, 2007), Vol. 1, Part I, pp. 417–424.

] are combined to aid alignment of the multispectral images and visualisation of the processed data [22

22. N. T. Clancy, D. Stoyanov, V. Sauvage, D. R. C. James, G.-Z. Yang, and D. S. Elson, “A Triple Endoscope System for Alignment of Multispectral Images of Moving Tissue,” in Biomedical Optics, Technical Digest (CD) (Optical Society of America, 2010), paper BTuD27.

].

Each camera was geometrically modelled using the pinhole projection model [26

26. R. Hartley and A. Zisserman, Multiple View Geometry in Computer Vision (Cambridge University Press, 2000).

], mapping points in 3D space, M = [X Y Z] onto each image plane in pixel coordinates m = [x y] as a matrix multiplication:

mtL=KL[I|0]Mt,mtR=KR[RR|tR]Mt,mtM=KM[RM|tM]Mt.
(1)

The intrinsic camera model parameters, K, for each camera were determined with a photogrammetry-based technique using multiple views of a planar calibration object of known dimensions [27

27. Z. Zhang, “A flexible new technique for camera calibration,” IEEE Trans. Pattern Anal. 22(11), 1330–1334 (2000). [CrossRef]

]. The pose of each camera, described by a rotation matrix R and translation vector t, with respect to a reference coordinate system (taken as the left camera) could also be computed during the calibration process. Thus, with a fully calibrated system, the mathematical relationship between common features in each camera is characterised and the projection of 3D tissue surface points onto the 2D image planes of the cameras can be readily performed using the derived matrix multiplication factors. An illustration of this geometric relationship is shown in Fig. 2
Fig. 2 Illustration of trinocular endoscope imaging geometry. Geometric calibration of the system means that image points in the left and right white light images can be used to triangulate the 3D position of points on the tissue surface. These can then be reprojected into multispectral image coordinates. We track the motion of points in the white light images as these have consistent light appearance and we use the reprojection capability of the calibrated system to maintain a track of the respective region in the multispectral image.
.

By matching image primitives across the multiple views acquired with the system, it was possible to quantitatively recover the 3D structure of the subject by triangulating line-of-sight rays, q, from the two white light channels as shown in Fig. 2 [28

28. D. Scharstein and R. Szeliski, “A taxonomy and evaluation of dense two-frame stereo correspondence algorithms,” Int. J. Comput. Vis. 47(1/3), 7–42 (2002). [CrossRef]

]. The triangulation process involves a least squares solution by rearranging the equations for the left and right projection from Eq. (1). For soft-tissue undergoing deformation, the 3D motion of the surface could be determined by tracking surface points [23

23. D. Stoyanov, A. Darzi, and G.-Z. Yang, “Dense 3D depth recovery for soft tissue deformation during robotically assisted laparoscopic surgery,” in MICCAI, C. Barillot, D. R. Haynor, and P. Hellier, eds. (Springer-Verlag, Rennes/Saint-Malo, France, 2004), pp. 41–48.

] or a geometric surface parameterisation [24

24. D. Stoyanov, G. P. Mylonas, F. Deligianni, A. Darzi, and G.-Z. Yang, “Soft-tissue motion tracking and structure estimation for robotic assisted MIS procedures,” in MICCAI, J. Duncan, and G. Gerig, eds. (Springer-Verlag, Palm Springs, USA, 2005), pp. 139–146.

].

2.2.1 Feature tracking and image alignment

The main processing steps for each image triplet are summarised in Fig. 3
Fig. 3 Image processing algorithm schematic. The raw image stacks are acquired simultaneously using the synchronised endoscope cameras. Features are identified and tracked in 3D throughout the image stack. The transformations needed to align these features are derived and implemented. Each feature is then projected through 3D space onto the multispectral camera using the calibration relations, so that they are automatically aligned in the multispectral 2D images. These can then be processed to extract the reflectance spectrum at each pixel location over the stack and compute relative concentrations of chromophores of interest. The processed image may then be reprojected onto 3D space to aid visualisation.
. After the cameras have been calibrated, the surface shape of the object (phantom, in vivo tissue) under investigation is reconstructed by triangulating corresponding features in each of the stereo views using the geometry shown in Fig. 2. Once this is done, the position of these features, which are selected based on intensity gradient information, can be tracked in time over the full set of images by employing stereo-temporal constraints using a stereoscopic variant of the Lucas-Kanade algorithm [25

25. D. Stoyanov and G.-Z. Yang, “Stabilization of image motion for robotic assisted beating heart surgery,” in MICCAI'07 Proceedings of the 10th international conference on Medical image computing and computer-assisted intervention, N. Ayache, S. Ourselin, and A. Maeder, eds. (Springer-Verlag, Brisbane, Australia, 2007), Vol. 1, Part I, pp. 417–424.

].

Once the location of corresponding 3D features are identified over the full set of images (across time and the stereo pair), a function may be derived that maps each point onto a reference image (the first image acquired in the series) by using the calibration information with Eq. (1) and the temporal correspondence of the points. Using the calibration matrix multiplications, these features can then be back-projected from 3D space to the multispectral image plane, aligning each of these images automatically. The chief advantage of this procedure is that the algorithm only requires the feature information from the sharp, full colour stereo images for the alignment. The multispectral images, which can suffer from low light transmission, wavelength-dependent features and motion blur due to the longer integration times required, are automatically aligned using the data from the controlled calibration. Of course, there are limitations and feature tracking can fail in the presence of occlusions due to specular highlights, surgical instruments or dynamic effects such as bleeding. These can potentially be overcome by more complex matching strategies and motion modelling predicting physiological tissue motions, however, this was beyond the scope of our current study.

Furthermore, for subjects with a known geometric structure, such as planes, the multispectral images can also be warped using projective transformations such as homographies [26

26. R. Hartley and A. Zisserman, Multiple View Geometry in Computer Vision (Cambridge University Press, 2000).

] in order to fully align a stack of images. This technique is first used in a proof-of-concept experiment, where a planar object is moving with respect to the endoscope. Subsequent to this, the first in vivo results obtained with the system are presented, in which an arbitrary tissue surface was imaged.

2.3. Multispectral processing

Once the images from the multispectral channel are aligned, it is possible to process them and extract the reflectance spectra of the tissue of interest. The absorbance spectrum at each pixel is calculated by taking the logarithm of the ratio of reflected intensity from the sample (IS) to that from a reflectance standard (Specralon; Labsphere, Inc., USA) (IR):

A=log(ISIDIRID),
(2)

where ID is the ‘dark intensity,’ recorded at each pixel when the illumination source is switched off and accounts for any background signal from ambient light. For the in vivo experiment, this term was neglected as the procedure was carried out laparoscopically. The term inside the brackets in Eq. (2) is the reflectance. Given that the main absorber of light in tissue is haemoglobin, and assuming that other chromophores are negligible and that the effects of scattering are flat across the wavelength range of interest (visible range), the absorbance is given by

A(λ)=[HbO2]εHbO2λ+[Hb]εHbλ+D,
(3)

3. Results

3.1. Phantom validation

To demonstrate the capabilities of the system, a standard Macbeth colour chart was chosen as the object. It was placed in front of the endoscope and images recorded as the position of the chart was changed and the LCTF scanned through the 420-700 nm wavelength range so that the final data set consisted of synchronised images from the multispectral and colour cameras of the object in different orientations. The 3D reconstruction method described above was then used to warp and align each of the images, allowing recovery of the multispectral data.

To test the accuracy of the reproduction of the images, a reflectance spectrum for each of the colour panels was calculated and compared to spectra obtained in a control experiment while the target remained static. The red, green and blue filter response of the colour stereo cameras was calculated by acquiring images of the reflectance standard through the LCTF (decoupled from the endoscope) and dividing by the reference spectrum to correct for the CCD sensitivity. This data was used to reconstruct the colour of each panel in the aligned multispectral images by integrating them over the measured filter response. The results seen in Fig. 4 (d)
Fig. 4 (a) Reconstructed colour image of checker card using misaligned multispectral images. (b) Colour checker reconstructed using multispectral images of stationary target. (c) Image of colour checker card acquired using colour CCD camera. (d) Colour image of moving chart reconstructed using the aligned multispectral camera images along with reflectance spectra (normalised intensity vs. wavelength) for each panel. The reflectance spectrum for each colour panel calculated using the aligned images (blue dots) is compared with that calculated from images of a stationary target (red dots).
show that the shape of the normalised reflectance spectrum in each colour panel can be reproduced, along with the colour of each panel, in spite of the fact that it is moving significantly (approximately 3 cm in directions perpendicular to camera axis, 5 cm in directions parallel to camera axis, rotation ≈10°) in comparison to typical movements associated with breathing. The movement was equivalent to an average shift of 71 pixels (36% of the diagonal length of one of the colour panels).

This agreement between the spectra was quantified using Bland-Altman analysis [36

36. J. M. Bland and D. G. Altman, “Statistical methods for assessing agreement between two methods of clinical measurement,” Lancet 327 (8476), 307–310 (1986). [CrossRef] [PubMed]

], where it was found that there was negligible bias in the spectra measured by the trinocular system (average difference in reflectance = 0.03), and 95% of the differences were found, on average, in the range ±0.20 (normalised reflected intensity). The full results for each of the colour panels (starting from the top left as seen in Fig. 4) are given in Table 1

Table 1. Bland-Altman analysis of the agreement between spectra measured using multispectral images of a static colour checker card, and images of a moving card aligned using the 3D reconstruction and tracking algorithm

table-icon
View This Table
.

The blur on the edges of the colour panels is due to slight misalignment of the ‘dewarped’ images. However, this error is small (≈0.5 mm) in comparison with the gross displacement of the target, and does not prevent recovery of the spectral data.

3.2. In vivo testing

In vivo testing was performed on a 75 kg female pig under general anaesthesia during a laparoscopic procedure (UK Home Office approved licence PPL 80/2297). The trinocular system was inserted into the abdomen through an 11 mm laparoscopic port and held stationary using two endoscope clamps fixed to the operating table. The LCTF was set to cycle through the wavelength range 440 nm to 720 nm in 10 nm steps, resulting in a stack of 29 multispectral images. Acquisition of the full stack of image triplets took approximately 30 seconds, during which time movement due to breathing was evident. It should be noted that subsequently only the 500-630 nm range was used in the spectral analysis as the signal outside this range was close to the level of system noise. The mathematical model applied to the reflectance spectra was of the form shown in Eq. (3), which assumed that haemoglobin absorption was the only significant attenuator of light. For processing the images and testing the alignment algorithm, small sections or ‘patches’ of an image were assessed individually, each centred on a corresponding feature.

Breathing was the main source of motion during the experiment, but other irregular movements to be compensated for included peristalsis and small variations in the endoscope’s position in its mechanical holder or minor motion of the surgeon’s hand, which was also supporting the device. The movement of a sample patch tracked using the system is shown in Fig. 5
Fig. 5 Motion tracking in vivo. The location of the patch is tracked in the colour stereo cameras (right camera; top row). The feature is back-projected onto the multispectral camera (middle row). Multispectral images are aligned using the back-projected feature (bottom row). See Media 1, Media 2, and Media 3.
. In this case, the factors mentioned above caused displacements of 15-20 mm in the x-y plane. This is illustrated in the attached video files that show tracking of a feature in one of the stereo cameras, for three different patches, alongside multispectral data acquisition (intensity equalised for clarity) before and after alignment (Media 1, Media 2, Media 3). The back-projected locations of the same feature in the raw multispectral images show that the algorithm can track the tissue without depending on salient features. Once this tracking has been accomplished, a simple transformation aligns the image patches so that the tracked feature is matched in the full image series.

Once each patch was aligned across all the acquisitions in a particular experiment, the multispectral images were processed by fitting the absorbance at each pixel to the model described by Eq. (3). This resulted in a concentration value for the haemoglobins and the offset value D. Colour images of the tissue in each patch were also reconstructed using the same technique as that used in the phantom validation (Section 3.1). These colour images are shown in Fig. 6
Fig. 6 Colour images of selected patches of intestinal tissue reconstructed from the multispectral stack of images, and corresponding maps of total haemoglobin. Arrows indicate smaller blood vessels that are washed out in the raw images but become visible after alignment. The smear effect on Patch 4 and 5 is due to pixel padding after the region of interest around those features moved outside the boundary of the raw multispectral image.
along with maps of Hbt, before and after alignment.

In the raw images (before alignment), the largest blood vessel in the field of view is still visible but is blurred due to motion artefacts. However, after alignment these vessels are more defined and have higher contrast with the surrounding tissue. Smaller blood vessels branching off the side of the large vessel are clearly visible in the aligned images of several patches but are completely averaged out in the raw images. This is further supported by the Hbt maps, which show that the areas of highest blood content correspond to the locations of the vessels after alignment. In the raw images, erroneously high Hbt values are seen in the areas around the large blood vessel and low values in the regions corresponding to the locations of smaller vessels.

The implications that misalignment has for the recorded spectra are particularly noticeable in regions near blood vessels. In these areas, pre-alignment, the signal over the full wavelength range could variously come from an area with a high blood content (vessel) or low blood content (parenchyma) depending on the movement of the tissue during acquisition. This can result in fitting errors due to the spatial averaging of the spectrum, causing overestimation of Hbt and SaO2 in the parenchyma and underestimation in the vessels.

The visualisation capabilities of the system are demonstrated in Fig. 7
Fig. 7 Stereo pair (left and right) used for feature tracking and 3D reconstruction shown alongside the finished surface. The processed Hbt data from the patches analysed in Fig. 6 are overlaid onto the surface.
. A dense reconstruction of the surface is provided by the stereo cameras while the locations of the patches analysed in Fig. 6 are indicated. After processing it was possible to reproject pixels from the multispectral camera back onto the surface, providing functional information overlaid on the white light image.

4. Discussion and conclusions

The results of the in vivo experiment showed that when applied in this complex surgical environment, the system was capable of following the arbitrary shape of intestinal tissue and compensating for the motion associated with breathing and peristalsis. Using the aligned images, it was possible to extract the absorbance spectrum at each point in the field of view and calculate the relative blood content and oxygen saturation. After processing, it was possible to project this information back onto the colour view. The main limitation of the current system is the acquisition time for a single stack of multispectral images which is close to 30 s for 30 different wavelengths. This is due to the LCTF: its switching time was of the order of 50 ms, but its poor optical throughput (5-50%) meant that long integration times also had to be used for each wavelength. With a faster filter such as an acousto-optic tuneable filter (AOTF), a more sensitive camera such as an EMCCD [18

18. R. Leitner, M. De Biasio, T. Arnold, C. Viet Dinh, M. Loog, and R. P. W. Duin, “Multi-spectral video endoscopy system for the detection of cancerous tissue,” Pattern Recognit. Lett. (to be published).

] and real-time 3D reconstruction and tracking algorithms [25

25. D. Stoyanov and G.-Z. Yang, “Stabilization of image motion for robotic assisted beating heart surgery,” in MICCAI'07 Proceedings of the 10th international conference on Medical image computing and computer-assisted intervention, N. Ayache, S. Ourselin, and A. Maeder, eds. (Springer-Verlag, Brisbane, Australia, 2007), Vol. 1, Part I, pp. 417–424.

], this system could achieve real-time operation suitable for use in the clinic. Furthermore, the light propagation model used will be updated to improve the fit to the experimental results by accounting for differential pathlength variations across the spectrum [37

37. M. B. Bouchard, B. R. Chen, S. A. Burgess, and E. M. C. Hillman, “Ultra-fast multispectral optical imaging of cortical oxygenation, blood flow, and intracellular calcium dynamics,” Opt. Express 17(18), 15670–15678 (2009). [CrossRef] [PubMed]

]. Other multivariate analysis techniques such as support vector machines [38

38. H. Akbari, L. V. Halig, D. M. Schuster, A. Osunkoya, V. Master, P. T. Nieh, G. Z. Chen, and B. Fei, “Hyperspectral imaging and quantitative analysis for prostate cancer detection,” J. Biomed. Opt. 17(7), 076005 (2012). [CrossRef] [PubMed]

] will be implemented to classify tissue according to oxygenation or disease-status.

Although applied to multispectral imaging here, this technique could be applied to any of the sequential image acquisition modalities previously mentioned (multispectral fluorescence, narrowband or polarisation-resolved imaging) as well as in dynamic small animal imaging and microscopy. The system may be integrated into robotic-assisted minimally invasive surgical systems such as the da Vinci, to provide morphological and functional feedback to the surgeon, since the stereo view of the tissue is a standard feature of this console.

Acknowledgements

We gratefully acknowledge the loan of the laparoscope by Intuitive Surgical, Inc and the assistance of Northwick Park Institute for Medical Research (NPIMR) for trial arrangements. Funding for this project was provided by ERC grant 242991, and UK EPSRC and Technology Strategy Board grants EP/E06342X/1 and DT/E011101/1. Danail Stoyanov would like to acknowledge the financial support of a Royal Academy of Engineering/EPSRC Fellowship.

References and links

1.

M. A. Ilias, E. Häggblad, C. Anderson, and E. G. Salerud, “Visible, hyperspectral imaging evaluating the cutaneous response to ultraviolet radiation,” Proc. SPIE 6441, 644103 (2007). [CrossRef]

2.

B. S. Sorg, B. J. Moeller, O. Donovan, Y. Cao, and M. W. Dewhirst, “Hyperspectral imaging of hemoglobin saturation in tumor microvasculature and tumor hypoxia development,” J. Biomed. Opt. 10(4), 044004 (2005). [CrossRef] [PubMed]

3.

S. P. Nighswander-Rempel, R. Anthony Shaw, J. R. Mansfield, M. Hewko, V. V. Kupriyanov, and H. H. Mantsch, “Regional variations in myocardial tissue oxygenation mapped by near-infrared spectroscopic imaging,” J. Mol. Cell. Cardiol. 34(9), 1195–1203 (2002). [CrossRef] [PubMed]

4.

M. G. Sowa, J. R. Payette, M. D. Hewko, and H. H. Mantsch, “Visible-near infrared multispectral imaging of the rat dorsal skin flap,” J. Biomed. Opt. 4(4), 474–481 (1999). [CrossRef]

5.

R. T. Bryan, L. J. Billingham, and D. M. A. Wallace, “Narrow-band imaging flexible cystoscopy in the detection of recurrent urothelial cancer of the bladder,” BJU Int. 101(6), 702–706 (2008). [CrossRef] [PubMed]

6.

J. H. Takano, T. Yakushiji, I. Kamiyama, T. Nomura, A. Katakura, N. Takano, and T. Shibahara, “Detecting early oral cancer: narrowband imaging system observation of the oral mucosa microvasculature,” Int. J. Oral Maxillofac. Surg. 39(3), 208–213 (2010). [CrossRef] [PubMed]

7.

H. C. Wolfsen, J. E. Crook, M. Krishna, S. R. Achem, K. R. Devault, E. P. Bouras, D. S. Loeb, M. E. Stark, T. A. Woodward, L. L. Hemminger, F. K. Cayer, and M. B. Wallace, “Prospective, controlled tandem endoscopy study of narrow band imaging for dysplasia detection in Barrett’s Esophagus,” Gastroenterology 135(1), 24–31 (2008). [CrossRef] [PubMed]

8.

V. X. D. Yang, P. J. Muller, P. Herman, and B. C. Wilson, “A multispectral fluorescence imaging system: design and initial clinical tests in intra-operative Photofrin-photodynamic therapy of brain tumors,” Lasers Surg. Med. 32(3), 224–232 (2003). [CrossRef] [PubMed]

9.

L. Zhou and W. S. El-Deiry, “Multispectral fluorescence imaging,” J. Nucl. Med. 50(10), 1563–1566 (2009). [CrossRef] [PubMed]

10.

J. Qi, C. Barrière, T. C. Wood, and D. S. Elson, “Polarized multispectral imaging in a rigid endoscope based on elastic light scattering spectroscopy,” Biomed. Opt. Express 3(9), 2087–2099 (2012). [CrossRef]

11.

P. K. Frykman, E. H. Lindsley, M. Gaon, and D. L. Farkas, “Spectral imaging for precise surgical intervention in Hirschsprung’s disease,” J Biophotonics 1(2), 97–103 (2008). [CrossRef] [PubMed]

12.

N. T. Clancy, V. Sauvage, S. Saso, D. Stoyanov, D. J. Corless, M. Boyd, D. E. Noakes, G.-Z. Yang, S. Ghaem-Maghami, J. R. Smith, and D. S. Elson, “Registration and analysis of multispectral images acquired during uterine transplantation surgery,” in Biomedical Optics and 3D Imaging, Technical Digest (CD) (Optical Society of America, 2012), paper BSu3A.73.

13.

S. L. Best, A. Thapa, M. J. Holzer, N. Jackson, S. A. Mir, J. A. Cadeddu, and K. J. Zuzak, “Minimal arterial in-flow protects renal oxygenation and function during porcine partial nephrectomy: confirmation by hyperspectral imaging,” Urology 78(4), 961–966 (2011). [CrossRef] [PubMed]

14.

K. J. Zuzak, S. C. Naik, G. Alexandrakis, D. Hawkins, K. Behbehani, and E. H. Livingston, “Intraoperative bile duct visualization using near-infrared hyperspectral video imaging,” Am. J. Surg. 195(4), 491–497 (2008). [CrossRef] [PubMed]

15.

S. P. Nighswander-Rempel, R. A. Shaw, V. V. Kupriyanov, J. Rendell, B. Xiang, and H. H. Mantsch, “Mapping tissue oxygenation in the beating heart with near-infrared spectroscopic imaging,” Vib. Spectrosc. 32(1), 85–94 (2003). [CrossRef]

16.

L. Gao, R. T. Kester, N. Hagen, and T. S. Tkaczyk, “Snapshot Image Mapping Spectrometer (IMS) with high sampling density for hyperspectral microscopy,” Opt. Express 18(14), 14330–14344 (2010). [CrossRef] [PubMed]

17.

A. Gorman, D. W. Fletcher-Holmes, and A. R. Harvey, “Generalization of the Lyot filter and its application to snapshot spectral imaging,” Opt. Express 18(6), 5602–5608 (2010). [CrossRef] [PubMed]

18.

R. Leitner, M. De Biasio, T. Arnold, C. Viet Dinh, M. Loog, and R. P. W. Duin, “Multi-spectral video endoscopy system for the detection of cancerous tissue,” Pattern Recognit. Lett. (to be published).

19.

G. N. Stamatas, M. Southall, and N. Kollias, “In vivo monitoring of cutaneous edema using spectral imaging in the visible and near infrared,” J. Invest. Dermatol. 126(8), 1753–1760 (2006). [CrossRef] [PubMed]

20.

D. Stoyanov, A. Rayshubskiy, and E. Hillman, “Robust registration of multispectral images of the cortical surface in neurosurgery,” in Proceedings of 9th IEEE International Symposium on Biomedical Imaging (Institute of Electrical and Electronic Engineers, Barcelona, 2012), pp. 1643–1646.

21.

K. J. Zuzak, S. C. Naik, G. Alexandrakis, D. Hawkins, K. Behbehani, and E. H. Livingston, “Characterization of a near-infrared laparoscopic hyperspectral imaging system for minimally invasive surgery,” Anal. Chem. 79(12), 4709–4715 (2007). [CrossRef] [PubMed]

22.

N. T. Clancy, D. Stoyanov, V. Sauvage, D. R. C. James, G.-Z. Yang, and D. S. Elson, “A Triple Endoscope System for Alignment of Multispectral Images of Moving Tissue,” in Biomedical Optics, Technical Digest (CD) (Optical Society of America, 2010), paper BTuD27.

23.

D. Stoyanov, A. Darzi, and G.-Z. Yang, “Dense 3D depth recovery for soft tissue deformation during robotically assisted laparoscopic surgery,” in MICCAI, C. Barillot, D. R. Haynor, and P. Hellier, eds. (Springer-Verlag, Rennes/Saint-Malo, France, 2004), pp. 41–48.

24.

D. Stoyanov, G. P. Mylonas, F. Deligianni, A. Darzi, and G.-Z. Yang, “Soft-tissue motion tracking and structure estimation for robotic assisted MIS procedures,” in MICCAI, J. Duncan, and G. Gerig, eds. (Springer-Verlag, Palm Springs, USA, 2005), pp. 139–146.

25.

D. Stoyanov and G.-Z. Yang, “Stabilization of image motion for robotic assisted beating heart surgery,” in MICCAI'07 Proceedings of the 10th international conference on Medical image computing and computer-assisted intervention, N. Ayache, S. Ourselin, and A. Maeder, eds. (Springer-Verlag, Brisbane, Australia, 2007), Vol. 1, Part I, pp. 417–424.

26.

R. Hartley and A. Zisserman, Multiple View Geometry in Computer Vision (Cambridge University Press, 2000).

27.

Z. Zhang, “A flexible new technique for camera calibration,” IEEE Trans. Pattern Anal. 22(11), 1330–1334 (2000). [CrossRef]

28.

D. Scharstein and R. Szeliski, “A taxonomy and evaluation of dense two-frame stereo correspondence algorithms,” Int. J. Comput. Vis. 47(1/3), 7–42 (2002). [CrossRef]

29.

S. A. Prahl, “Tabulated molar extinction coefficient for hemoglobin in water” (Oregon Medical Laser Center, 2009), http://omlc.ogi.edu/spectra/hemoglobin/summary.html

30.

K. J. Zuzak, M. D. Schaeberle, M. T. Gladwin, R. O. I. Cannon 3rd, and I. W. Levin, “Noninvasive determination of spatially resolved and time-resolved tissue perfusion in humans during nitric oxide inhibition and inhalation by use of a visible-reflectance hyperspectral imaging technique,” Circulation 104(24), 2905–2910 (2001). [CrossRef] [PubMed]

31.

D. R. C. James, V. Sauvage, N. T. Clancy, J. Clark, A. W. Darzi, G.-Z. Yang, and D. S. Elson, “Towards intra operative diagnosis of mesenteric oxygenation with hyperspectral imaging,” presented at the Society of American Gastrointestinal and Endoscopic Surgeons (SAGES) 12th World Congress of Endoscopic Surgery, National Harbour, Maryland, USA. 2010.

32.

V. Sauvage, D. James, K. Koh, T. Wood, and D. S. Elson, “Development of a hyperspectral laparoscope system for intraoperative diagnosis of intestinal ischemia,” Proc. SPIE 7555, 7555-8 (2010).

33.

D. Yudovsky, A. Nouvong, and L. Pilon, “Hyperspectral imaging in diabetic foot wound care,” J. Diabetes Sci. Tech. 4(5), 1099–1113 (2010). [PubMed]

34.

J. R. Payette, E. Kohlenberg, L. Leonardi, A. Pabbies, P. Kerr, K.-Z. Liu, and M. G. Sowa, “Assessment of skin flaps using optically based methods for measuring blood flow and oxygenation,” Plast. Reconstr. Surg. 115(2), 539–546 (2005). [CrossRef] [PubMed]

35.

C. R. Tracy, J. D. Terrell, R. P. Francis, E. F. Wehner, J. Smith, M. Litorja, D. L. Hawkins, M. S. Pearle, J. A. Cadeddu, and K. J. Zuzak, “Characterization of renal ischemia using DLP hyperspectral imaging: a pilot study comparing artery-only occlusion versus artery and vein occlusion,” J. Endourol. 24(3), 321–325 (2010). [CrossRef] [PubMed]

36.

J. M. Bland and D. G. Altman, “Statistical methods for assessing agreement between two methods of clinical measurement,” Lancet 327 (8476), 307–310 (1986). [CrossRef] [PubMed]

37.

M. B. Bouchard, B. R. Chen, S. A. Burgess, and E. M. C. Hillman, “Ultra-fast multispectral optical imaging of cortical oxygenation, blood flow, and intracellular calcium dynamics,” Opt. Express 17(18), 15670–15678 (2009). [CrossRef] [PubMed]

38.

H. Akbari, L. V. Halig, D. M. Schuster, A. Osunkoya, V. Master, P. T. Nieh, G. Z. Chen, and B. Fei, “Hyperspectral imaging and quantitative analysis for prostate cancer detection,” J. Biomed. Opt. 17(7), 076005 (2012). [CrossRef] [PubMed]

OCIS Codes
(170.2150) Medical optics and biotechnology : Endoscopic imaging
(170.3010) Medical optics and biotechnology : Image reconstruction techniques
(170.6510) Medical optics and biotechnology : Spectroscopy, tissue diagnostics

ToC Category:
Endoscopes, Catheters and Micro-Optics

History
Original Manuscript: June 18, 2012
Revised Manuscript: September 7, 2012
Manuscript Accepted: September 11, 2012
Published: September 14, 2012

Virtual Issues
BIOMED 2012 (2012) Biomedical Optics Express

Citation
Neil T. Clancy, Danail Stoyanov, David R. C. James, Aimee Di Marco, Vincent Sauvage, James Clark, Guang-Zhong Yang, and Daniel S. Elson, "Multispectral image alignment using a three channel endoscope in vivo during minimally invasive surgery," Biomed. Opt. Express 3, 2567-2578 (2012)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-3-10-2567


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. M. A. Ilias, E. Häggblad, C. Anderson, and E. G. Salerud, “Visible, hyperspectral imaging evaluating the cutaneous response to ultraviolet radiation,” Proc. SPIE6441, 644103 (2007). [CrossRef]
  2. B. S. Sorg, B. J. Moeller, O. Donovan, Y. Cao, and M. W. Dewhirst, “Hyperspectral imaging of hemoglobin saturation in tumor microvasculature and tumor hypoxia development,” J. Biomed. Opt.10(4), 044004 (2005). [CrossRef] [PubMed]
  3. S. P. Nighswander-Rempel, R. Anthony Shaw, J. R. Mansfield, M. Hewko, V. V. Kupriyanov, and H. H. Mantsch, “Regional variations in myocardial tissue oxygenation mapped by near-infrared spectroscopic imaging,” J. Mol. Cell. Cardiol.34(9), 1195–1203 (2002). [CrossRef] [PubMed]
  4. M. G. Sowa, J. R. Payette, M. D. Hewko, and H. H. Mantsch, “Visible-near infrared multispectral imaging of the rat dorsal skin flap,” J. Biomed. Opt.4(4), 474–481 (1999). [CrossRef]
  5. R. T. Bryan, L. J. Billingham, and D. M. A. Wallace, “Narrow-band imaging flexible cystoscopy in the detection of recurrent urothelial cancer of the bladder,” BJU Int.101(6), 702–706 (2008). [CrossRef] [PubMed]
  6. J. H. Takano, T. Yakushiji, I. Kamiyama, T. Nomura, A. Katakura, N. Takano, and T. Shibahara, “Detecting early oral cancer: narrowband imaging system observation of the oral mucosa microvasculature,” Int. J. Oral Maxillofac. Surg.39(3), 208–213 (2010). [CrossRef] [PubMed]
  7. H. C. Wolfsen, J. E. Crook, M. Krishna, S. R. Achem, K. R. Devault, E. P. Bouras, D. S. Loeb, M. E. Stark, T. A. Woodward, L. L. Hemminger, F. K. Cayer, and M. B. Wallace, “Prospective, controlled tandem endoscopy study of narrow band imaging for dysplasia detection in Barrett’s Esophagus,” Gastroenterology135(1), 24–31 (2008). [CrossRef] [PubMed]
  8. V. X. D. Yang, P. J. Muller, P. Herman, and B. C. Wilson, “A multispectral fluorescence imaging system: design and initial clinical tests in intra-operative Photofrin-photodynamic therapy of brain tumors,” Lasers Surg. Med.32(3), 224–232 (2003). [CrossRef] [PubMed]
  9. L. Zhou and W. S. El-Deiry, “Multispectral fluorescence imaging,” J. Nucl. Med.50(10), 1563–1566 (2009). [CrossRef] [PubMed]
  10. J. Qi, C. Barrière, T. C. Wood, and D. S. Elson, “Polarized multispectral imaging in a rigid endoscope based on elastic light scattering spectroscopy,” Biomed. Opt. Express3(9), 2087–2099 (2012). [CrossRef]
  11. P. K. Frykman, E. H. Lindsley, M. Gaon, and D. L. Farkas, “Spectral imaging for precise surgical intervention in Hirschsprung’s disease,” J Biophotonics1(2), 97–103 (2008). [CrossRef] [PubMed]
  12. N. T. Clancy, V. Sauvage, S. Saso, D. Stoyanov, D. J. Corless, M. Boyd, D. E. Noakes, G.-Z. Yang, S. Ghaem-Maghami, J. R. Smith, and D. S. Elson, “Registration and analysis of multispectral images acquired during uterine transplantation surgery,” in Biomedical Optics and 3D Imaging, Technical Digest (CD) (Optical Society of America, 2012), paper BSu3A.73.
  13. S. L. Best, A. Thapa, M. J. Holzer, N. Jackson, S. A. Mir, J. A. Cadeddu, and K. J. Zuzak, “Minimal arterial in-flow protects renal oxygenation and function during porcine partial nephrectomy: confirmation by hyperspectral imaging,” Urology78(4), 961–966 (2011). [CrossRef] [PubMed]
  14. K. J. Zuzak, S. C. Naik, G. Alexandrakis, D. Hawkins, K. Behbehani, and E. H. Livingston, “Intraoperative bile duct visualization using near-infrared hyperspectral video imaging,” Am. J. Surg.195(4), 491–497 (2008). [CrossRef] [PubMed]
  15. S. P. Nighswander-Rempel, R. A. Shaw, V. V. Kupriyanov, J. Rendell, B. Xiang, and H. H. Mantsch, “Mapping tissue oxygenation in the beating heart with near-infrared spectroscopic imaging,” Vib. Spectrosc.32(1), 85–94 (2003). [CrossRef]
  16. L. Gao, R. T. Kester, N. Hagen, and T. S. Tkaczyk, “Snapshot Image Mapping Spectrometer (IMS) with high sampling density for hyperspectral microscopy,” Opt. Express18(14), 14330–14344 (2010). [CrossRef] [PubMed]
  17. A. Gorman, D. W. Fletcher-Holmes, and A. R. Harvey, “Generalization of the Lyot filter and its application to snapshot spectral imaging,” Opt. Express18(6), 5602–5608 (2010). [CrossRef] [PubMed]
  18. R. Leitner, M. De Biasio, T. Arnold, C. Viet Dinh, M. Loog, and R. P. W. Duin, “Multi-spectral video endoscopy system for the detection of cancerous tissue,” Pattern Recognit. Lett. (to be published).
  19. G. N. Stamatas, M. Southall, and N. Kollias, “In vivo monitoring of cutaneous edema using spectral imaging in the visible and near infrared,” J. Invest. Dermatol.126(8), 1753–1760 (2006). [CrossRef] [PubMed]
  20. D. Stoyanov, A. Rayshubskiy, and E. Hillman, “Robust registration of multispectral images of the cortical surface in neurosurgery,” in Proceedings of 9th IEEE International Symposium on Biomedical Imaging (Institute of Electrical and Electronic Engineers, Barcelona, 2012), pp. 1643–1646.
  21. K. J. Zuzak, S. C. Naik, G. Alexandrakis, D. Hawkins, K. Behbehani, and E. H. Livingston, “Characterization of a near-infrared laparoscopic hyperspectral imaging system for minimally invasive surgery,” Anal. Chem.79(12), 4709–4715 (2007). [CrossRef] [PubMed]
  22. N. T. Clancy, D. Stoyanov, V. Sauvage, D. R. C. James, G.-Z. Yang, and D. S. Elson, “A Triple Endoscope System for Alignment of Multispectral Images of Moving Tissue,” in Biomedical Optics, Technical Digest (CD) (Optical Society of America, 2010), paper BTuD27.
  23. D. Stoyanov, A. Darzi, and G.-Z. Yang, “Dense 3D depth recovery for soft tissue deformation during robotically assisted laparoscopic surgery,” in MICCAI, C. Barillot, D. R. Haynor, and P. Hellier, eds. (Springer-Verlag, Rennes/Saint-Malo, France, 2004), pp. 41–48.
  24. D. Stoyanov, G. P. Mylonas, F. Deligianni, A. Darzi, and G.-Z. Yang, “Soft-tissue motion tracking and structure estimation for robotic assisted MIS procedures,” in MICCAI, J. Duncan, and G. Gerig, eds. (Springer-Verlag, Palm Springs, USA, 2005), pp. 139–146.
  25. D. Stoyanov and G.-Z. Yang, “Stabilization of image motion for robotic assisted beating heart surgery,” in MICCAI'07 Proceedings of the 10th international conference on Medical image computing and computer-assisted intervention, N. Ayache, S. Ourselin, and A. Maeder, eds. (Springer-Verlag, Brisbane, Australia, 2007), Vol. 1, Part I, pp. 417–424.
  26. R. Hartley and A. Zisserman, Multiple View Geometry in Computer Vision (Cambridge University Press, 2000).
  27. Z. Zhang, “A flexible new technique for camera calibration,” IEEE Trans. Pattern Anal.22(11), 1330–1334 (2000). [CrossRef]
  28. D. Scharstein and R. Szeliski, “A taxonomy and evaluation of dense two-frame stereo correspondence algorithms,” Int. J. Comput. Vis.47(1/3), 7–42 (2002). [CrossRef]
  29. S. A. Prahl, “Tabulated molar extinction coefficient for hemoglobin in water” (Oregon Medical Laser Center, 2009), http://omlc.ogi.edu/spectra/hemoglobin/summary.html
  30. K. J. Zuzak, M. D. Schaeberle, M. T. Gladwin, R. O. I. Cannon, and I. W. Levin, “Noninvasive determination of spatially resolved and time-resolved tissue perfusion in humans during nitric oxide inhibition and inhalation by use of a visible-reflectance hyperspectral imaging technique,” Circulation104(24), 2905–2910 (2001). [CrossRef] [PubMed]
  31. D. R. C. James, V. Sauvage, N. T. Clancy, J. Clark, A. W. Darzi, G.-Z. Yang, and D. S. Elson, “Towards intra operative diagnosis of mesenteric oxygenation with hyperspectral imaging,” presented at the Society of American Gastrointestinal and Endoscopic Surgeons (SAGES) 12th World Congress of Endoscopic Surgery, National Harbour, Maryland, USA. 2010.
  32. V. Sauvage, D. James, K. Koh, T. Wood, and D. S. Elson, “Development of a hyperspectral laparoscope system for intraoperative diagnosis of intestinal ischemia,” Proc. SPIE7555, 7555-8 (2010).
  33. D. Yudovsky, A. Nouvong, and L. Pilon, “Hyperspectral imaging in diabetic foot wound care,” J. Diabetes Sci. Tech.4(5), 1099–1113 (2010). [PubMed]
  34. J. R. Payette, E. Kohlenberg, L. Leonardi, A. Pabbies, P. Kerr, K.-Z. Liu, and M. G. Sowa, “Assessment of skin flaps using optically based methods for measuring blood flow and oxygenation,” Plast. Reconstr. Surg.115(2), 539–546 (2005). [CrossRef] [PubMed]
  35. C. R. Tracy, J. D. Terrell, R. P. Francis, E. F. Wehner, J. Smith, M. Litorja, D. L. Hawkins, M. S. Pearle, J. A. Cadeddu, and K. J. Zuzak, “Characterization of renal ischemia using DLP hyperspectral imaging: a pilot study comparing artery-only occlusion versus artery and vein occlusion,” J. Endourol.24(3), 321–325 (2010). [CrossRef] [PubMed]
  36. J. M. Bland and D. G. Altman, “Statistical methods for assessing agreement between two methods of clinical measurement,” Lancet327 (8476), 307–310 (1986). [CrossRef] [PubMed]
  37. M. B. Bouchard, B. R. Chen, S. A. Burgess, and E. M. C. Hillman, “Ultra-fast multispectral optical imaging of cortical oxygenation, blood flow, and intracellular calcium dynamics,” Opt. Express17(18), 15670–15678 (2009). [CrossRef] [PubMed]
  38. H. Akbari, L. V. Halig, D. M. Schuster, A. Osunkoya, V. Master, P. T. Nieh, G. Z. Chen, and B. Fei, “Hyperspectral imaging and quantitative analysis for prostate cancer detection,” J. Biomed. Opt.17(7), 076005 (2012). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Supplementary Material


» Media 1: AVI (685 KB)     
» Media 2: AVI (684 KB)     
» Media 3: AVI (685 KB)     

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited