Maximum-likelihood estimation in Optical Coherence Tomography in the context of the tear film dynamics

Jinxin Huang, Eric Clarkson, Matthew Kupinski, Kye-sung Lee, Kara L. Maki, David S. Ross, James V. Aquavella, and Jannick P. Rolland  »View Author Affiliations


Biomedical Optics Express, Vol. 4, Issue 10, pp. 1806-1816 (2013)
http://dx.doi.org/10.1364/BOE.4.001806


View Full Text Article

Acrobat PDF (1266 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Understanding tear film dynamics is a prerequisite for advancing the management of Dry Eye Disease (DED). In this paper, we discuss the use of optical coherence tomography (OCT) and statistical decision theory to analyze the tear film dynamics of a digital phantom. We implement a maximum-likelihood (ML) estimator to interpret OCT data based on mathematical models of Fourier-Domain OCT and the tear film. With the methodology of task-based assessment, we quantify the tradeoffs among key imaging system parameters. We find, on the assumption that the broadband light source is characterized by circular Gaussian statistics, ML estimates of 40 nm +/− 4 nm for an axial resolution of 1 μm and an integration time of 5 μs. Finally, the estimator is validated with a digital phantom of tear film dynamics, which reveals estimates of nanometer precision.

© 2013 OSA

1. Introduction

Dry Eye Disease (DED) is a multifactorial disease of the tears and the ocular surface. Symptoms include discomfort, visual disturbance, and tear film instability, and in some cases damage to the ocular surface [1

1. M. A. Lemp, C. Baudouin, and J. Baum., “The definition and classification of dry eye disease: report of the definition and classification subcommittee of the International Dry Eye WorkShop (2007),” Ocul. Surf. 5(2), 75–92 (2007). [CrossRef] [PubMed]

]. DED, a serious public health problem that affects more than 100 million people worldwide, degrades the quality of life seriously [2

2. B. Miljanović, R. Dana, D. A. Sullivan, and D. A. Schaumberg, “Impact of dry eye syndrome on vision-related quality of life,” Am. J. Ophthalmol. 143(3), 409–415, 415.e2 (2007). [CrossRef] [PubMed]

,3

3. R. Montés-Micó, “Role of the tear film in the optical quality of the human eye,” J. Cataract Refract. Surg. 33(9), 1631–1635 (2007). [CrossRef] [PubMed]

]. However, therapies for DED remain elusive for two reasons: there is no quantitative, highly repeatable, objective diagnosis method, and there is little or no correlation between signs and symptoms [4

4. D. A. Sullivan, K. M. Hammitt, D. A. Schaumberg, B. D. Sullivan, C. G. Begley, P. Gjorstrup, J. S. Garrigue, M. Nakamura, Y. Quentric, S. Barabino, M. Dalton, and G. D. Novack, “Report of the TFOS/ARVO symposium on global treatments for dry eye disease: an unmet need,” Ocul. Surf. 10(2), 108–116 (2012). [CrossRef] [PubMed]

]. In this work, we hypothesize that tear film instability is a key indicator of DED and can be quantified by tear film dynamics, specifically the thinning rate of the tear film across the cornea.

Optical coherence tomography (OCT) is a biomedical imaging method that has emerged in the last two decades [5

5. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]

]. In OCT, one collects the back-reflected, or scattered, light from tissue. The method is similar to ultrasound, but it operates in the near infrared region of the spectrum. OCT has been used extensively in ophthalmology, especially for imaging the retina. In this work, we propose to use OCT to monitor tear film dynamics. Considering the speed of changes in tear film thicknesses, Fourier-Domain OCT (FD-OCT) is tailored to this challenge.

Statistical decision theory has been widely used in various biomedical fields in which random processes associated with light emission and detection must be accounted for. Statistical decision theory was introduced into time domain OCT for classification tasks [6

6. J. Rolland, J. O’Daniel, C. Akcay, T. DeLemos, K. S. Lee, K. I. Cheong, E. Clarkson, R. Chakrabarti, and R. Ferris, “Task-based optimization and performance assessment in optical coherence imaging,” J. Opt. Soc. Am. A 22(6), 1132–1142 (2005). [CrossRef] [PubMed]

,7

7. A. C. Akcay, E. Clarkson, and J. P. Rolland, “Effect of source spectral shape on task-based assessment of detection and resolution in optical coherence tomography,” Appl. Opt. 44(35), 7573–7580 (2005). [CrossRef] [PubMed]

]. We recently reported on key results of using statistical decision theory to estimate tear film thickness [8

8. J. Huang, K. S. Lee, E. Clarkson, M. Kupinski, K. L. Maki, D. S. Ross, J. V. Aquavella, and J. P. Rolland, “Phantom study of tear film dynamics with optical coherence tomography and maximum-likelihood estimation,” Opt. Lett. 38(10), 1721–1723 (2013). [CrossRef] [PubMed]

]. In this paper, we present the detailed mathematical framework that underlies these results, and we further report on the limit of the estimation. To our knowledge, this work represents the first time a maximum-likelihood (ML) estimator has been used to interpret OCT data.

In section 2 of this paper, we present a mathematical model of an FD-OCT system in the context of the ML estimation. We present a model of the tear film in section 3. Next, in section 4, we detail the generation of simulated spectra that serve as inputs to the ML estimator. The principle of the ML estimator is then detailed in section 5, while its performance is investigated in section 6. In the last section of the paper, the ML estimator is validated using a digital phantom.

2. Mathematical model of FD-OCT in the context of the ML estimation

In this work, we consider a free-space setup of an FD-OCT system that is shown schematically in Fig. 1
Fig. 1 FD-OCT setup.
. It is based on a Michelson-like interferometer geometry, in which the detector is a spectrometer.

The broadband light source emits an electric field that can be regarded as a superposition of plane waves and expressed as
Es(t)=E0(ω)exp(iωt)dω.
(1)
The electric field for a plane wave with an angular frequency ω can be written as
Es(ω,t)=E0(ω)exp(iωt).
(2)
These plane waves are split at the beam splitter. One of the two resulting waves travels into a reference arm, the other travels into a sample arm. The electric field propagating in the reference arm will accumulate a phase delay ϕ1(ω), which accounts for the optical path length, the dispersion, and any other shifts of phase from the reference arm. We let α(ω) represent the amplitude loss caused by reflection, absorption, and the loss at the beam splitter. The electric field propagating in the sample arm will, similarly, experience a phase delay ϕ2(ω). We Let β(ω) represent the amplitude loss and phase shifts caused by reflection, beam splitting, and interaction with the specimen. These two modulated electric fields will recombine at the beam splitter and propagate to the detector. The detector in FD-OCT is a complementary metal–oxide–semiconductor (CMOS) or charge-coupled device (CCD) camera-based spectrometer, whose output is a number array. Each element of the array is proportional to the number of electrons accumulated at an associated pixel during one integration time.

The electric field on the xth pixel of the detector can be written as
Ed(x,t)=ωxΔωxωx[α(ω)exp(iϕ1(ω))+β(ω)exp(iϕ2(ω))]E0(ω)exp(iωt)dω.
(3)
For simplicity, we define
m(ω)=α(ω)exp(iϕ1(ω))+β(ω)exp(iϕ2(ω)).
(4)
Therefore Eq. (3) may be rewritten as
Ed(x,t)=ωxΔωxωxm(ω)E0(ω)exp(iωt)dω;
(5)
where the angular frequencies of the incidence waves on the xth pixel span from ωxωx to ωx.

KNp(x,x;Δt)=R(x)eδxxΔtωxΔωxωx|m(ω)|2S(ω)dω+δxx0Δt0ΔtR(x)R(x)e2ωxΔωxωxωxΔωxωx|m(ω)|2|m(ω)|2S(ω)S(ω)exp(i(ωω)(tt))dωdωdtdt=R(x)eδxxΔtωxΔωxωx|m(ω)|2S(ω)dω+δxxΔtR(x)R(x)e2ωxΔωxωx|m(ω)|4S2(ω)dω.
(24)

3. Tear film model

The normal tear film consists of three parts: the lipid layer, the aqueous layer, and the mucin layer. Because the aqueous layer has the largest volume in the tear film, here we only consider the aqueous layer [11

11. P. E. King-Smith, B. A. Fink, R. M. Hill, K. W. Koelling, and J. M. Tiffany, “The thickness of the tear film,” Curr. Eye Res. 29(4-5), 357–368 (2004). [CrossRef] [PubMed]

]; we take this as a first approximation, to be refined in future investigations. We model the tear film as a one layer structure with two interfaces, the interface of the air and the tear film and the interface of the tear film and the cornea; these are shown schematically in Fig. 2
Fig. 2 Tear film model.
, where n1, n2, and n3 are the refractive indexes of the air, tear film, and cornea, respectively. In the figure, l is the thickness of the tear film, and r1 and r2 are the reflection coefficients of the two interfaces.

The interface of the tear film and the cornea is considered rough (and is illustrated as such), as it is known that the corneal surface has microplicae and glycocalyx [11

11. P. E. King-Smith, B. A. Fink, R. M. Hill, K. W. Koelling, and J. M. Tiffany, “The thickness of the tear film,” Curr. Eye Res. 29(4-5), 357–368 (2004). [CrossRef] [PubMed]

]. In our model, we use the root mean squared deviation in the height of the surface, σ, to characterize the surface roughness. We assume that deviations of the corneal surface from a smooth plane can be described by a stochastic process with a Gaussian distribution. We assume that the incident light strikes the corneal surface at a right angle. Using the perturbation theory developed in [12

12. T. A. Leskova, A. A. Maradudin, and I. V. Novikov, “Scattering of light from the random interface between two dielectric media with low contrast,” J. Opt. Soc. Am. A 17(7), 1288–1300 (2000). [CrossRef] [PubMed]

], the following equation has been established [13

13. Y. Ashtamker, V. Freilikher, and J. C. Dainty, “Ambiguity of optical coherence tomography measurements due to rough surface scattering,” Opt. Express 19(22), 21658–21664 (2011). [CrossRef] [PubMed]

]
r2=n2n3n2+n3exp(2ε2σ2k2),
(25)
where ε2 is the dielectric constant of the tear film. At the interface of the air and the tear film, which we consider to be smooth, the reflection coefficient, r1, is given by Fresnel’s law of reflection as
r1=n1n2n1+n2.
(26)
The total response m(ω) from the OCT system can be written as m(ω)=rrefexp(i2kreflref)+(n1n2n1+n2+n2n3n2+n3exp(2ε2σ2ktf2)exp(i2ktfl))exp(i2ksamlsam), (27)where rrefis the reflectivity of the reference mirror, lrefand lsam are the lengths of the reference arm and the sample arm, respectively, and kref, ksam and ktfare the wave numbers in the corresponding medium.

4. Generate simulated OCT imaging data sets

To generate one simulated image for a tear film with thickness l, we assume that one instance of the output at the xth pixel follows a normal distribution:
Nlg(x,Δt)~Normal(Nlg(x,Δt)D|P|GP|GG,KNlg(x,x;Δt)).
(28)
Thus we use a normal random number generator to simulate a spectrum Nlg, which is a vector representing the output along the camera’s pixel array. It is important to note that we simulate spectra to benchmark the framework; in a real measurement, Nlg will be a measured spectrum, which will be the input to the ML estimator detailed in section 5.

5. Principle of the ML estimator

On the assumption that all pixels are independent of each other, the probability that a particular spectrum Nlg is generated by a tear film of thickness d can be expressed as
P(Nlg|d)=1(2π)M2Det[KNdg]12exp[12x(Nlg(x,Δt)Ndg(x,Δt)D|P|GP|GG)2KNdg(x,x;Δt)],
(29)
where M is the number of pixels, and Det is the determinant.

The ML estimator estimates parameters by maximizing P(Nlg|d) with respect to d, which is equivalent to minimizing the following expression
qln(P(Nlg|d)).
(30)
Therefore, the estimate from one data set Nlg will be given by

d^=argmind[x(Nlg(x,Δt)Ndg(x,Δt))22KNdg(x,x;Δt)+12xln(KNdg(x,x;Δt))+M2ln(2π)].
(31)

Figure 3
Fig. 3 Principle of the ML estimator illustrated with different axial resolutions of 1 μm, 2 μm, and 4 μm.
illustrates the principle of our ML estimation. We consider spectra that are Gaussian-shaped, with 1 μm, 2 μm, and 4 μm axial resolutions, respectively. Axial resolution, in this paper, is conventionally defined as the half width half maximum of the axial point spread function [7

7. A. C. Akcay, E. Clarkson, and J. P. Rolland, “Effect of source spectral shape on task-based assessment of detection and resolution in optical coherence tomography,” Appl. Opt. 44(35), 7573–7580 (2005). [CrossRef] [PubMed]

,14

14. C. Akcay, P. Parrein, and J. P. Rolland, “Estimation of longitudinal resolution in optical coherence imaging,” Appl. Opt. 41(25), 5256–5262 (2002). [CrossRef] [PubMed]

]. The source power is set at 1 mW to be in compliance with the ANSI standard for eye safety in this spectral region [15

15. American National Standards Institute, “American national standard for safe use of lasers,” ANSI Z136.1–2007 (2007).

]. The integration time is 5 μs. Ground truth of the tear film thickness is set to be 0.5 μm, and we assume a root mean squared height of the corneal surface of 150 nm [16

16. H. B. Chen, S. Yamabayashi, B. Ou, Y. Tanaka, S. Ohno, and S. Tsukahara, “Structure and composition of rat precorneal tear film. A study by an in vivo cryofixation,” Invest. Ophthalmol. Vis. Sci. 38(2), 381–387 (1997). [PubMed]

] to characterize the corneal roughness.

The plots in red frames of Fig. 3 show the negative log of the conditional probability, which characterizes the probability that one observed spectrum Nlg is generated by a tear film thickness d. Simulations for three different optical axial resolutions are shown. Results show that the conditional probability oscillates when the possible thickness changes. The distance between two adjacent peaks is half of the center wavelength in the sample, or 300 nm (in the tear film). For a tear film whose thickness is 0.5 μm the three resolutions of 1, 2, and 4 μm yield estimates of 505 nm, 485 nm, and 545 nm, respectively, where each estimate is determined by the position of the minimum point in the corresponding conditional probability plot shown in Fig. 3. We will characterize precision of this measurement in section 6.1. Note that with this method we can obtain estimates with precision greater than the optical axial resolution of the system. It is clear from Fig. 3 that for low resolution (i.e., 4 μm) the estimator misses the minimum value because nearby thicknesses yield values comparable to the minimum.

6. Investigation of the performance of the ML estimator

6.1 Impact of detector integration time

Imaging speed is a key factor in the performance of the system. The two parameters that may affect the imaging speed, the repetition rate of the pulsed source and the detector integration time, must be considered. In OCT, a first requirement is that the source speed is at least as great as the detector’s response rate. Therefore we will study the impact of the detector integration time. To quantify how the system performance changes with integration time, we set up a simulation in which we consider a fixed tear film thickness of 0.5 μm and we repeat the estimation 5000 times. We then compute the root mean squared error (RMSE) to evaluate the performance for different optical axial resolutions as a function of integration time. We plot the resulting RMSE as a function of integration time in Fig. 4
Fig. 4 Impact of integration time on the RMSE of the estimation.
.

These results show that for a fixed axial resolution, the RMSE decreases exponentially with increasing integration time, eventually reaching a plateau value in the nanometer range. For the same integration time, higher axial resolution results in a higher precision. There are two costs to consider in pushing the system to highest axial resolution: one is the cost of the light source, and the other is the possible reduction in imaging depth. Indeed, as the axial resolution increases, the spectral bandwidth increases as well, but may still be collected on an array of the same size. Therefore, the spectral resolution typically decreases, unless a more costly detector is used; if one uses the same detector, the depth of imaging is typically reduced.

6.2 Limit of the estimation

In estimating thickness, it is important to identify the limit of precision of the ML estimation. To study this limit, we vary the thickness of the ground truth in the simulation and calculate the relative error, which is defined as the ratio of the RMSE to the tear film thickness. In this simulation, the integration time is fixed to be 5 μs and the RMSE is calculated based on 5000 simulated data sets. Results are shown in Fig. 5
Fig. 5 Estimation limit in relative error of the ML estimator as a function of the tear-film thickness.
. For axial resolutions of 1 μm and 4 μm, the relative error diminishes as the tear film thickness increases. For a maximum 10% relative error (shown as a red dashed line in Fig. 5), the ML estimator can estimate thicknesses as low as 0.04 μm and 1.3 μm for axial resolutions of 1 μm and 4 μm, respectively, where each estimate corresponds to the intersection of the red dashed line and the curve of relative error.

7. Validation of the ML estimator with a digital phantom

8. Conclusion

We have developed and presented a mathematical model of FD-OCT with a single layer (aqueous) tear film model laid on a rough corneal surface. To benchmark the framework of ML estimation in FD-OCT, we implemented an ML estimator to interpret simulated OCT data sets that were generated from the mathematical models. We quantified the tradeoff between the detector integration time and the precision of the estimation. Results show that a combined axial resolution of 1 μm (defined conventionally as the half width at half maximum of the axial point spread function) and an integration time of 5 μs allow estimates in the nanometer range (e.g. 40 nm). Finally, the estimator was validated with a digital phantom of tear film dynamics. Results reveal estimates of nanometer precision.

Acknowledgments

This research was supported by the NYSTAR Foundation C050070, NIH grants RC1-EB010974 and R37-EB000803, and by Research to Prevent Blindness.

References and links

1.

M. A. Lemp, C. Baudouin, and J. Baum., “The definition and classification of dry eye disease: report of the definition and classification subcommittee of the International Dry Eye WorkShop (2007),” Ocul. Surf. 5(2), 75–92 (2007). [CrossRef] [PubMed]

2.

B. Miljanović, R. Dana, D. A. Sullivan, and D. A. Schaumberg, “Impact of dry eye syndrome on vision-related quality of life,” Am. J. Ophthalmol. 143(3), 409–415, 415.e2 (2007). [CrossRef] [PubMed]

3.

R. Montés-Micó, “Role of the tear film in the optical quality of the human eye,” J. Cataract Refract. Surg. 33(9), 1631–1635 (2007). [CrossRef] [PubMed]

4.

D. A. Sullivan, K. M. Hammitt, D. A. Schaumberg, B. D. Sullivan, C. G. Begley, P. Gjorstrup, J. S. Garrigue, M. Nakamura, Y. Quentric, S. Barabino, M. Dalton, and G. D. Novack, “Report of the TFOS/ARVO symposium on global treatments for dry eye disease: an unmet need,” Ocul. Surf. 10(2), 108–116 (2012). [CrossRef] [PubMed]

5.

D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]

6.

J. Rolland, J. O’Daniel, C. Akcay, T. DeLemos, K. S. Lee, K. I. Cheong, E. Clarkson, R. Chakrabarti, and R. Ferris, “Task-based optimization and performance assessment in optical coherence imaging,” J. Opt. Soc. Am. A 22(6), 1132–1142 (2005). [CrossRef] [PubMed]

7.

A. C. Akcay, E. Clarkson, and J. P. Rolland, “Effect of source spectral shape on task-based assessment of detection and resolution in optical coherence tomography,” Appl. Opt. 44(35), 7573–7580 (2005). [CrossRef] [PubMed]

8.

J. Huang, K. S. Lee, E. Clarkson, M. Kupinski, K. L. Maki, D. S. Ross, J. V. Aquavella, and J. P. Rolland, “Phantom study of tear film dynamics with optical coherence tomography and maximum-likelihood estimation,” Opt. Lett. 38(10), 1721–1723 (2013). [CrossRef] [PubMed]

9.

H. H. Barrett and K. J. Myers, Foundations of Image Science (Wiley, Hoboken, 2004), Chap. 14.

10.

H. H. Barrett and K. J. Myers, Foundations of Image Science (Wiley, Hoboken, 2004), Chap. 8.

11.

P. E. King-Smith, B. A. Fink, R. M. Hill, K. W. Koelling, and J. M. Tiffany, “The thickness of the tear film,” Curr. Eye Res. 29(4-5), 357–368 (2004). [CrossRef] [PubMed]

12.

T. A. Leskova, A. A. Maradudin, and I. V. Novikov, “Scattering of light from the random interface between two dielectric media with low contrast,” J. Opt. Soc. Am. A 17(7), 1288–1300 (2000). [CrossRef] [PubMed]

13.

Y. Ashtamker, V. Freilikher, and J. C. Dainty, “Ambiguity of optical coherence tomography measurements due to rough surface scattering,” Opt. Express 19(22), 21658–21664 (2011). [CrossRef] [PubMed]

14.

C. Akcay, P. Parrein, and J. P. Rolland, “Estimation of longitudinal resolution in optical coherence imaging,” Appl. Opt. 41(25), 5256–5262 (2002). [CrossRef] [PubMed]

15.

American National Standards Institute, “American national standard for safe use of lasers,” ANSI Z136.1–2007 (2007).

16.

H. B. Chen, S. Yamabayashi, B. Ou, Y. Tanaka, S. Ohno, and S. Tsukahara, “Structure and composition of rat precorneal tear film. A study by an in vivo cryofixation,” Invest. Ophthalmol. Vis. Sci. 38(2), 381–387 (1997). [PubMed]

17.

K. L. Maki, R. J. Braun, P. Ucciferro, W. D. Henshaw, and P. E. King-Smith, “Tear film dynamics on an eye-shaped domain II: flux boundary conditions,” J. Fluid Mech. 647, 361–389 (2010). [CrossRef]

OCIS Codes
(030.0030) Coherence and statistical optics : Coherence and statistical optics
(110.3000) Imaging systems : Image quality assessment

ToC Category:
Ophthalmology Applications

History
Original Manuscript: June 10, 2013
Revised Manuscript: August 17, 2013
Manuscript Accepted: August 17, 2013
Published: August 29, 2013

Citation
Jinxin Huang, Eric Clarkson, Matthew Kupinski, Kye-sung Lee, Kara L. Maki, David S. Ross, James V. Aquavella, and Jannick P. Rolland, "Maximum-likelihood estimation in Optical Coherence Tomography in the context of the tear film dynamics," Biomed. Opt. Express 4, 1806-1816 (2013)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-4-10-1806


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. M. A. Lemp, C. Baudouin, J. Baum, and ., “The definition and classification of dry eye disease: report of the definition and classification subcommittee of the International Dry Eye WorkShop (2007),” Ocul. Surf.5(2), 75–92 (2007). [CrossRef] [PubMed]
  2. B. Miljanović, R. Dana, D. A. Sullivan, and D. A. Schaumberg, “Impact of dry eye syndrome on vision-related quality of life,” Am. J. Ophthalmol.143(3), 409–415, 415.e2 (2007). [CrossRef] [PubMed]
  3. R. Montés-Micó, “Role of the tear film in the optical quality of the human eye,” J. Cataract Refract. Surg.33(9), 1631–1635 (2007). [CrossRef] [PubMed]
  4. D. A. Sullivan, K. M. Hammitt, D. A. Schaumberg, B. D. Sullivan, C. G. Begley, P. Gjorstrup, J. S. Garrigue, M. Nakamura, Y. Quentric, S. Barabino, M. Dalton, and G. D. Novack, “Report of the TFOS/ARVO symposium on global treatments for dry eye disease: an unmet need,” Ocul. Surf.10(2), 108–116 (2012). [CrossRef] [PubMed]
  5. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. Fujimoto, “Optical coherence tomography,” Science254(5035), 1178–1181 (1991). [CrossRef] [PubMed]
  6. J. Rolland, J. O’Daniel, C. Akcay, T. DeLemos, K. S. Lee, K. I. Cheong, E. Clarkson, R. Chakrabarti, and R. Ferris, “Task-based optimization and performance assessment in optical coherence imaging,” J. Opt. Soc. Am. A22(6), 1132–1142 (2005). [CrossRef] [PubMed]
  7. A. C. Akcay, E. Clarkson, and J. P. Rolland, “Effect of source spectral shape on task-based assessment of detection and resolution in optical coherence tomography,” Appl. Opt.44(35), 7573–7580 (2005). [CrossRef] [PubMed]
  8. J. Huang, K. S. Lee, E. Clarkson, M. Kupinski, K. L. Maki, D. S. Ross, J. V. Aquavella, and J. P. Rolland, “Phantom study of tear film dynamics with optical coherence tomography and maximum-likelihood estimation,” Opt. Lett.38(10), 1721–1723 (2013). [CrossRef] [PubMed]
  9. H. H. Barrett and K. J. Myers, Foundations of Image Science (Wiley, Hoboken, 2004), Chap. 14.
  10. H. H. Barrett and K. J. Myers, Foundations of Image Science (Wiley, Hoboken, 2004), Chap. 8.
  11. P. E. King-Smith, B. A. Fink, R. M. Hill, K. W. Koelling, and J. M. Tiffany, “The thickness of the tear film,” Curr. Eye Res.29(4-5), 357–368 (2004). [CrossRef] [PubMed]
  12. T. A. Leskova, A. A. Maradudin, and I. V. Novikov, “Scattering of light from the random interface between two dielectric media with low contrast,” J. Opt. Soc. Am. A17(7), 1288–1300 (2000). [CrossRef] [PubMed]
  13. Y. Ashtamker, V. Freilikher, and J. C. Dainty, “Ambiguity of optical coherence tomography measurements due to rough surface scattering,” Opt. Express19(22), 21658–21664 (2011). [CrossRef] [PubMed]
  14. C. Akcay, P. Parrein, and J. P. Rolland, “Estimation of longitudinal resolution in optical coherence imaging,” Appl. Opt.41(25), 5256–5262 (2002). [CrossRef] [PubMed]
  15. American National Standards Institute, “American national standard for safe use of lasers,” ANSI Z136.1–2007 (2007).
  16. H. B. Chen, S. Yamabayashi, B. Ou, Y. Tanaka, S. Ohno, and S. Tsukahara, “Structure and composition of rat precorneal tear film. A study by an in vivo cryofixation,” Invest. Ophthalmol. Vis. Sci.38(2), 381–387 (1997). [PubMed]
  17. K. L. Maki, R. J. Braun, P. Ucciferro, W. D. Henshaw, and P. E. King-Smith, “Tear film dynamics on an eye-shaped domain II: flux boundary conditions,” J. Fluid Mech.647, 361–389 (2010). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.