Photoacoustic microscopy: a potential new tool for evaluation of angiogenesis inhibitor

Sung-Liang Chen, Joseph Burnett, Duxin Sun, Xunbin Wei, Zhixing Xie, and Xueding Wang  »View Author Affiliations


Biomedical Optics Express, Vol. 4, Issue 11, pp. 2657-2666 (2013)
http://dx.doi.org/10.1364/BOE.4.002657


View Full Text Article

Acrobat PDF (3910 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

The feasibility of photoacoustic microscopy (PAM) for evaluation of angiogenesis inhibitor was investigated on a chick embryo model in vivo. Different concentrations of the angiogenesis inhibitor, Sunitinib, were applied to the chorioallantoic membrane (CAM) of the chick embryos. Imaging of microvasculature in embryo CAMs was acquired using a laser-scanning PAM system; while the optical microscopy (OM) capturing the microvascular images of the same set of CAMs for comparison served as a gold standard for validating the results from PAM. The microvascular density as a function of applied Sunitinib concentration has been quantified in both PAM and OM images. The results from these two modalities have a good agreement, suggesting that PAM could provide an unbiased quantification of microvascular density for objective evaluation of anti-angiogenesis medication. In comparison with conventional OM which enables only two-dimensional enface imaging, PAM is capable of three-dimensional analysis of microvessels, including not only morphology but also functions, as demonstrated in part by the imaging result on a canine bladder model. The emerging PAM technique shows promise to be used in clinical and preclinical settings for comprehensive and objective evaluation of anti-angiogenesis medications.

© 2013 Optical Society of America

1. Introduction

Several methodologies have been demonstrated to study 3D angiogenesis. The layer-by-layer biopsies of bone marrows were used for computer-aided 3D reconstruction images to reveal a complex arborizing and branching of microvessels in bone marrow with leukemia [3

3. A. R. Perez-Atayde, S. E. Sallan, U. Tedrow, S. Connors, E. Allred, and J. Folkman, “Spectrum of tumor angiogenesis in the bone marrow of children with acute lymphoblastic leukemia,” Am. J. Pathol. 150(3), 815–821 (1997). [PubMed]

]. The serial sectioning of tissue samples mechanically restricts the biopsy method from applications in vivo. Laser scanning confocal microscopy can capture stacks of confocal optical slices to digitally reconstitute a 3D projection of the vascular endothelial cells with high resolution. This technology, however, usually requires staining of vascular endothelial cells for angiogenesis imaging [4

4. I. Rajantie, M. Ilmonen, A. Alminaite, U. Ozerdem, K. Alitalo, and P. Salven, “Adult bone marrow-derived cells recruited during angiogenesis comprise precursors for periendothelial vascular mural cells,” Blood 104(7), 2084–2086 (2004). [CrossRef] [PubMed]

]. 3D power Doppler ultrasound enables 3D study of cancer associated hemodynamic changes, and can render both anatomic (e.g. vascularization) and functional images (e.g. blood flow) information [5

5. H. Pairleitner, H. Steiner, G. Hasenoehrl, and A. Staudach, “Three-dimensional power Doppler sonography: imaging and quantifying blood flow and vascularization,” Ultrasound Obstet. Gynecol. 14(2), 139–143 (1999). [CrossRef] [PubMed]

,6

6. J. W. Xuan, M. Bygrave, H. Jiang, F. Valiyeva, J. Dunmore-Buyze, D. W. Holdsworth, J. I. Izawa, G. Bauman, M. Moussa, S. F. Winter, N. M. Greenberg, J. L. Chin, M. Drangova, A. Fenster, and J. C. Lacefield, “Functional neoangiogenesis imaging of genetically engineered mouse prostate cancer using three-dimensional power Doppler ultrasound,” Cancer Res. 67(6), 2830–2839 (2007). [CrossRef] [PubMed]

]. However, improving the spatial resolution of high-frequency power Doppler ultrasound beyond 100 μm is challenging, which makes it difficult to study the morphologies and functions of small capillaries from angiogenesis [6

6. J. W. Xuan, M. Bygrave, H. Jiang, F. Valiyeva, J. Dunmore-Buyze, D. W. Holdsworth, J. I. Izawa, G. Bauman, M. Moussa, S. F. Winter, N. M. Greenberg, J. L. Chin, M. Drangova, A. Fenster, and J. C. Lacefield, “Functional neoangiogenesis imaging of genetically engineered mouse prostate cancer using three-dimensional power Doppler ultrasound,” Cancer Res. 67(6), 2830–2839 (2007). [CrossRef] [PubMed]

]. Optical coherence tomography (OCT), with the ability to quantitatively assess hemodynamic parameter such as blood vessel diameter and flow using Doppler algorithms, has been adapted to noninvasive microangiography of the human retina and other organs such as brain [7

7. S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14(17), 7821–7840 (2006). [CrossRef] [PubMed]

9

9. J. A. Izatt, M. D. Kulkarni, S. Yazdanfar, J. K. Barton, and A. J. Welch, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett. 22(18), 1439–1441 (1997). [CrossRef] [PubMed]

]. Neither OCT nor Doppler ultrasound, however, has the direct access to the blood oxygenation, one of the key functional parameters of tissue micro-environment.

Photoacoustic microscopy (PAM) is an emerging hybrid imaging modality based on the detection of laser generated ultrasound signals [10

10. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

14

14. L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nat. Photonics 3(9), 503–509 (2009). [CrossRef] [PubMed]

]. Presenting the intrinsic optical absorption contrast among soft tissues with excellent sensitivity and spatial resolution, PAM holds the promise to provide high-quality 3D angiogenesis images. Based on how the focus is achieved for generating microscopic images, PAM has now been divided into two categories including acoustic-resolution PAM (AR-PAM) and optical-resolution PAM (OR-PAM) [10

10. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

,11

11. H. F. Zhang, K. Maslov, and L. V. Wang, “In vivo imaging of subcutaneous structures using functional photoacoustic microscopy,” Nat. Protoc. 2(4), 797–804 (2007). [CrossRef] [PubMed]

,13

13. K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

]. The former has achieved lateral resolution of 45 μm and axial resolution of 15 μm with imaging depth beyond 3 mm. In comparison with AR-PAM, OR-PAM, although with limited imaging depth (> 0.7 mm [13

13. K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

]), enables further elevated spatial resolution which is crucial for quantitative evaluation of 3D distributed small capillaries. The OR-PAM system with axial resolution of 15 µm has been shown as a promising tool for mapping neovascular network architecture during angiogenesis inhibitor responsiveness [15

15. S. S. Oladipupo, S. Hu, A. C. Santeford, J. Yao, J. R. Kovalski, R. V. Shohet, K. Maslov, L. V. Wang, and J. M. Arbeit, “Conditional HIF-1 induction produces multistage neovascularization with stage-specific sensitivity to VEGFR inhibitors and myeloid cell independence,” Blood 117(15), 4142–4153 (2011). [CrossRef] [PubMed]

]. Potentially, the OR-PAM system with axial resolution down to the size of capillaries (5-10 µm) should furnish better quantification of the 3D change of the capillary network in response to angiogenesis inhibitor. For both AR-PAM and OR-PAM, the axial resolution is determined by the bandwidth of detected photoacoustic signals instead of the confocal parameter. Thus, low NA lenses may be used without sacrificing the depth resolution. Similar to the situation in OCT, working with relatively low NA objectives for PAM allows imaging with large fields of view, enabling the synthesis of microscopic and macroscopic information [16

16. V. J. Srinivasan, A. C. Chan, and E. Y. Lam, “Doppler OCT and OCT angiography for in vivo imaging of vascular physiology,” Chapter 2 in Selected Topics in Optical Coherence Tomography, book edited by Gangjun Liu (2012).

]. Using ultrasonic detectors with extremely high receiving bandwidth, axial resolution less than 10 μm has been realized [17

17. Z. Xie, S.-L. Chen, T. Ling, L. J. Guo, P. L. Carson, and X. Wang, “Pure optical photoacoustic microscopy,” Opt. Express 19(10), 9027–9034 (2011). [CrossRef] [PubMed]

,18

18. C. Zhang, K. Maslov, J. Yao, and L. V. Wang, “In vivo photoacoustic microscopy with 7.6-µm axial resolution using a commercial 125-MHz ultrasonic transducer,” J. Biomed. Opt. 17(11), 116016 (2012). [CrossRef] [PubMed]

]. This excellent axial resolution enables the sectioning along the depth direction of tissues in vivo, facilitating the visualization of 3D features in an intact biological object from tissue level to cellular level.

In this work, we explored the feasibility of OR-PAM for objective assessment of an angiogenesis inhibitor through quantitative imaging of microvasculature. The capability of our imaging system in 3D rendering and analysis of microvasculature was first demonstrated through the experiment on a canine bladder model, paving the road toward in vivo noninvasive biopsy of bladder cancer. Then by using a chick embryo model, we explored the performance of this system in quantifying the microvascular density and its change in response to angiogenesis inhibitor at different doses. The chick embryo model not only provides a reliable and easy-to-operate assay for evaluating agents with antiangiogenic activity but also enables conventional optical microscopy (OM) and digital image analysis (DIA) to be used as a gold standard to verify PAM results [19

19. C. Peifer and G. Dannhardt, “A novel quantitative chick embryo assay as an angiogenesis model using digital image analysis,” Anticancer Res. 24(3a), 1545–1551 (2004). [PubMed]

,20

20. C. N. Doukas, I. Maglogiannis, and A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]

].

2. Methods

The PAM images were obtained over a region of 1.22 mm × 1.22 mm with 256 × 256 pixels, and were displayed in 256 grey levels, i.e., 8 bits. A 4X objective was used in the OM to capture optical images with a size comparable to PAM images. The 8-bits color OM images were stored in BMP format. In general, we took OM images and then PAM images of the same embryo at the place marked by the filter paper. The quantification of both PAM and OM images from the chick embryos was realized following the DIA method described in the literature [20

20. C. N. Doukas, I. Maglogiannis, and A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]

]. Before analysis, the OM color images were converted into grayscale. Then, an adaptive thresholding was applied to convert the grayscale PAM and OM images to black/white images, i.e. ‘1’ and ‘0’ representing vessels and surrounding tissues, respectively. During adaptive thresholding, we partitioned the original image into many sub-images, and normalized every pixel value to between 0 and 1 in each sub-image. The mean value M in that sub-image was calculated. Then, for every pixel value p, if p > M + C, where C is a user-defined constant, p was set to ‘1’ as the blood vessels; otherwise ‘0’ as the surrounding tissues was assigned. Finally, the vessel density was determined by calculating the ratio of the number of pixels with ‘1’ to that of total pixels (i.e., the area of white pixels over the total imaging area). In the adaptive thresholding, different values of the number of sub-images and the constant C were selected to optimize the similarity of the vascular pattern between the input grayscale images and the output black/white images, and were dependent on the characteristics of images including lighting or signal uniformity, vessel density, and vessel texture. The DIA method for calculating the vessel density following the description in Ref [20

20. C. N. Doukas, I. Maglogiannis, and A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]

]. can be directly applied to 2D images of chick embryos. For the calculation of vessel density in the 3D image of the canine bladder, same idea was followed and the 3D vessel density was calculated by counting the volume of white pixels over the total imaging volume.

3. Results

Label-free PAM images presenting the 3D microvasculature in the inner wall of a canine bladder and the image analysis are shown in Fig. 2
Fig. 2 3D PAM of canine bladder, demonstrating the excellent capability of PAM in mapping the spatially distributed microvasculatures in biological samples. (a) The MAP images on the XY plane. (b) and (c) 2D cuts of 3D PAM image along the lines indicated in (a), where the left and right columns are for the images before (Grayscale) and after processing (Black/white). (d) The shortest A-line photoacoustic signal and its Hilbert transform. (e) The depth-encoded MAP image. Images (a)-(c) and (e) share the same scale bar shown in the bottom.
. The quantitative evaluation of vessel density using the DIA method described above was conducted. In Figs. 2(a)-2(c), the left column shows the imaging result in grayscale which, after image processing, was turned into black/white images, as shown in the right column. The constant value C of 0.2 and the number of sub-images of 36 (partitioned along X and Y, i.e., a 2D partition) were chosen in image conversion. Figures 2(b) and 2(c) show the 2D cuts from the 3D PAM image along the two lines indicated in Fig. 2(a). In these 2D B-scan images, the microvasculature with 2-3 layers in the sample can be clearly observed. A vessel density in 3D of 2.4% over a volume of 1.04 mm × 1.04 mm × 0.52 mm was extracted from the converted black/white images. Besides, facilitated by the excellent spatial resolution of PAM, the vasculature from large to small vessels can all be recognized clearly with satisfactory contrast-to-noise ratio, as can be seen from the maximum amplitude projection (MAP) image in Fig. 2(a). The highest axial resolution of this system was quantified as 8 μm [17

17. Z. Xie, S.-L. Chen, T. Ling, L. J. Guo, P. L. Carson, and X. Wang, “Pure optical photoacoustic microscopy,” Opt. Express 19(10), 9027–9034 (2011). [CrossRef] [PubMed]

]. The shortest length scale of imaged vessels was ~20 μm, as shown in the A-line photoacoustic signal in Fig. 2(d). Figure 2(e) shows the depth-encoded MAP image on the XY plane, demonstrating the sectioning capability of PAM.

Figure 3(a)
Fig. 3 Co-registered PAM and OM images taken from a chick embryo CAM. (a) The images taken on EDD 5. (b) The images taken on EDD 6 which was one day after the embryo was treated with 100-μM Sunitinib. The vasculature patterns in PAM are in good agreement with the counterpart OM image.
shows an example of co-registered PAM and OM images of the vasculature on a chick embryo CAM on EDD 5. As can be seen, the two images from PAM and OM have good match for either large vessels of ~160 µm or networks of micro-vessels down to ~10 µm, as well as the branching points. Figure 3(b) depicts the characteristic microvasculature of the embryo on EDD 6 after 100-µM Sunitinib was applied on EDD 5. The significantly decreased vessel density on the CAM as a result of the applied angiogenesis inhibitor can be clearly recognized in both PAM and OM images, which verifies that the anti-angiogenesis medication works well on the chick embryo CAM model. The signal-to-noise ratios (SNRs) of Figs. 3(a) and 3(b) are 22 dB and 24 dB, respectively.

For quantitative study, chick embryos were divided into four groups and treated on EDD 5 with 0.9% NaCl (control), or 10 µl of Sunitinib at 30, 100, and 300 µM concentrations respectively. For each group, the number of embryos was ≥ 3. PAM and OM images were taken 24 hrs after the treatment (EDD 6). On each embryo, images were taken from at least two different locations both within the treated area. Therefore, for each condition, we have at least 2 × 3 = 6 samples which facilitate the calculation of the average and the standard deviation (STD). Figure 4
Fig. 4 Representative PAM images of CAM on EDD 6 when the chick embryos were treated with 10 µl of Sunitinib at different concentrations for a period of 24 hours. The SNRs of PAM images from left to right are 24, 23, 23, and 25 dB, respectively.
shows the representative PAM images from two individual chick embryos as examples, demonstrating clearly the gradual changes in microvascular density from plentiful to sparse as a result of the increased concentration of the Sunitinib applied on the CAM.

The DIA method described above for the extraction of angiogenic density was pretested through the observation by the eye to determine the most suitable value of C and the number of sub-images [20

20. C. N. Doukas, I. Maglogiannis, and A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]

]. More specifically, we visually compared the images before (grayscale) and after the thresholding process (black/white) and saw if the processed images still rendered the characteristic vascular features. This visual assessment was conducted over all the PAM and OM images from all the four sample groups in order to decide the optimal parameters which include the constant value C of 0.2 and the number of sub-images of 36. Figure 5
Fig. 5 Conversion of PAM images from grayscale to black/white by using 36 sub-images and C = 0.2. (a) The example result from specimens with high vessel density (treated with Sunitinib at 30 µM). (b) The example results from specimens with low vessel density (treated with Sunitinib at 300 µM). The SNRs of PAM images before image processing are 24 dB and 26 dB in (a) and (b), respectively.
illustrates the examples of PAM image conversion from grayscale to black/white using the selected parameters, where Figs. 5(a) and 5(b) show the results from embryos with high and low vascular densities respectively. For both of these two cases, the majority of vascular features were kept well after the conversion. The converted images in black/white were then used for quantification of the vessel densities in each specimen.

Figure 6
Fig. 6 Statistical results of vessel density in chick embryo CAM vs. applied drug concentration quantified by PAM (solid line) and OM (dashed line), respectively. The two curves are shifted a little along the x-axis to avoid the overlap of the error bars. The inset shows the normalized curves for better comparison.
shows the statistical results of the vessel density in each group extracted from PAM and OM images. By comparing the two curves from PAM and OM respectively, a similar dependence (i.e. slope) of the vessel density over applied drug concentration from 30 to 300 µM can be seen. A lower-than-expected vessel density and a relatively large STD are observed in the control condition of the PAM curve, which is believed to be a result of the large sample-to-sample difference. Moreover, the inherent large variation in vessel density exists among different regions on the same CAM surface. The discrepancy in absolute values of vessel density between PAM and OM will be discussed later. For better comparison, the two curves from PAM and OM are normalized respectively, as shown in the inset of Fig. 6. For normalization, the average value of vessel densities over three conditions (i.e. 30, 100 and 300 µM) is subtracted from the vessel density at each condition. The data of the control condition is excluded due to the issue mentioned above. The two curves after normalization showing the relative vessel densities are in good agreement except at the control condition.

4. Discussion and conclusions

In this work, we demonstrated ex vivo imaging of canine bladder (Fig. 2) using the microring resonator, to prove the depth sectioning capability, and in vivo imaging of chick embryos (Figs. 3-5) using the PZT needle hydrophone, to validate the capability in quantitative evaluation of vascular density in response to angiogenesis inhibitor. Microring, although providing a better axial resolution [17

17. Z. Xie, S.-L. Chen, T. Ling, L. J. Guo, P. L. Carson, and X. Wang, “Pure optical photoacoustic microscopy,” Opt. Express 19(10), 9027–9034 (2011). [CrossRef] [PubMed]

], was not used for chick embryos mainly because it comes with a large footprint (~1 cm wide) and, therefore, is not convenient for reflection-mode imaging. PZT needle hydrophone, with a diameter of 1 mm, facilitates a convenient imaging setup on chick embryos; while the limited axial resolution of hydrophone was not a problem for imaging 2D vasculature in CAM of chick embryos. Further development on the microring resonator to facilitate reflection-mode operation is part of the future work [29

29. S.-L. Chen, Z. Xie, L. J. Guo, and X. Wang, “A fiber-optic system for dual-modality photoacoustic microscopy and confocal fluorescence microscopy using miniature components,” Photoacoustics 1(2), 30–35 (2013). [CrossRef]

].

Due to the spatial sensitivity of the PZT hydrophone, PA images in Figs. 3-5 show brighter in the center region and dimmer on the edge. The adopted adaptive thresholding helps alleviate the issue of nonuniform spatial sensitivity of PA images by converting the original grayscale PA images to black/white images. In the work described in Ref [20

20. C. N. Doukas, I. Maglogiannis, and A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]

], the issue of nonuniform lighting effect (i.e., nonuniform spatial sensitivity) was also effectively addressed by using the same adaptive thresholding method.

As we can see from Fig. 3 as an example, there is no obvious difference between PAM and OM in describing large vessels. However, there is noticeable discontinuity in small vessels and capillaries in the PAM images, which does not exist in the OM images. The main reason is that OM images are captured by a complementary metal–oxide–semiconductor sensor which has a much longer temporal integration; while in PAM, when no signal averaging is involved, the signal at each image pixel is an integration over a very short period of 2 ns as determined by laser pulse duration. Therefore, unlike OM, PAM shows the transient image of the red blood cells sparsely distributed in the capillaries, which is the main reason for the discontinuity in the PAM images. This also, at least in part, explains why the blood vessel densities quantified by PAM are lower than those by OM, as observed in Fig. 6.

Acknowledgments

Supports from the National Institutes of Health grant R01AR060350, University of Michigan-Shanghai Jiao Tong University (UM-SJTU) Joint Grant, and Samsung GRO Program are gratefully acknowledged.

References and links

1.

A. W. Griffioen and G. Molema, “Angiogenesis: potentials for pharmacologic intervention in the treatment of cancer, cardiovascular diseases, and chronic inflammation,” Pharmacol. Rev. 52(2), 237–268 (2000). [PubMed]

2.

J. Folkman, “Anti-angiogenesis: new concept for therapy of solid tumors,” Ann. Surg. 175(3), 409–416 (1972). [CrossRef] [PubMed]

3.

A. R. Perez-Atayde, S. E. Sallan, U. Tedrow, S. Connors, E. Allred, and J. Folkman, “Spectrum of tumor angiogenesis in the bone marrow of children with acute lymphoblastic leukemia,” Am. J. Pathol. 150(3), 815–821 (1997). [PubMed]

4.

I. Rajantie, M. Ilmonen, A. Alminaite, U. Ozerdem, K. Alitalo, and P. Salven, “Adult bone marrow-derived cells recruited during angiogenesis comprise precursors for periendothelial vascular mural cells,” Blood 104(7), 2084–2086 (2004). [CrossRef] [PubMed]

5.

H. Pairleitner, H. Steiner, G. Hasenoehrl, and A. Staudach, “Three-dimensional power Doppler sonography: imaging and quantifying blood flow and vascularization,” Ultrasound Obstet. Gynecol. 14(2), 139–143 (1999). [CrossRef] [PubMed]

6.

J. W. Xuan, M. Bygrave, H. Jiang, F. Valiyeva, J. Dunmore-Buyze, D. W. Holdsworth, J. I. Izawa, G. Bauman, M. Moussa, S. F. Winter, N. M. Greenberg, J. L. Chin, M. Drangova, A. Fenster, and J. C. Lacefield, “Functional neoangiogenesis imaging of genetically engineered mouse prostate cancer using three-dimensional power Doppler ultrasound,” Cancer Res. 67(6), 2830–2839 (2007). [CrossRef] [PubMed]

7.

S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14(17), 7821–7840 (2006). [CrossRef] [PubMed]

8.

R. K. Wang, S. L. Jacques, Z. Ma, S. Hurst, S. R. Hanson, and A. Gruber, “Three dimensional optical angiography,” Opt. Express 15(7), 4083–4097 (2007). [CrossRef] [PubMed]

9.

J. A. Izatt, M. D. Kulkarni, S. Yazdanfar, J. K. Barton, and A. J. Welch, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett. 22(18), 1439–1441 (1997). [CrossRef] [PubMed]

10.

H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

11.

H. F. Zhang, K. Maslov, and L. V. Wang, “In vivo imaging of subcutaneous structures using functional photoacoustic microscopy,” Nat. Protoc. 2(4), 797–804 (2007). [CrossRef] [PubMed]

12.

L. V. Wang, “Tutorial on photoacoustic microscopy and computed tomography,” IEEE J. Sel. Top. Quantum Electron. 14(1), 171–179 (2008). [CrossRef]

13.

K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

14.

L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nat. Photonics 3(9), 503–509 (2009). [CrossRef] [PubMed]

15.

S. S. Oladipupo, S. Hu, A. C. Santeford, J. Yao, J. R. Kovalski, R. V. Shohet, K. Maslov, L. V. Wang, and J. M. Arbeit, “Conditional HIF-1 induction produces multistage neovascularization with stage-specific sensitivity to VEGFR inhibitors and myeloid cell independence,” Blood 117(15), 4142–4153 (2011). [CrossRef] [PubMed]

16.

V. J. Srinivasan, A. C. Chan, and E. Y. Lam, “Doppler OCT and OCT angiography for in vivo imaging of vascular physiology,” Chapter 2 in Selected Topics in Optical Coherence Tomography, book edited by Gangjun Liu (2012).

17.

Z. Xie, S.-L. Chen, T. Ling, L. J. Guo, P. L. Carson, and X. Wang, “Pure optical photoacoustic microscopy,” Opt. Express 19(10), 9027–9034 (2011). [CrossRef] [PubMed]

18.

C. Zhang, K. Maslov, J. Yao, and L. V. Wang, “In vivo photoacoustic microscopy with 7.6-µm axial resolution using a commercial 125-MHz ultrasonic transducer,” J. Biomed. Opt. 17(11), 116016 (2012). [CrossRef] [PubMed]

19.

C. Peifer and G. Dannhardt, “A novel quantitative chick embryo assay as an angiogenesis model using digital image analysis,” Anticancer Res. 24(3a), 1545–1551 (2004). [PubMed]

20.

C. N. Doukas, I. Maglogiannis, and A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]

21.

S.-L. Chen, Z. Xie, P. L. Carson, X. Wang, and L. J. Guo, “In vivo flow speed measurement of capillaries by photoacoustic correlation spectroscopy,” Opt. Lett. 36(20), 4017–4019 (2011). [CrossRef] [PubMed]

22.

S.-L. Chen, Z. Xie, T. Ling, L. J. Guo, X. Wei, and X. Wang, “Miniaturized all-optical photoacoustic microscopy based on microelectromechanical systems mirror scanning,” Opt. Lett. 37(20), 4263–4265 (2012). [CrossRef] [PubMed]

23.

D. Ribatti, “Chick embryo chorioallantoic membrane as a useful tool to study angiogenesis,” Int Rev Cell Mol Biol 270, 181–224 (2008). [CrossRef] [PubMed]

24.

A. M. Cimpean, D. Ribatti, and M. Raica, “The chick embryo chorioallantoic membrane as a model to study tumor metastasis,” Angiogenesis 11(4), 311–319 (2008). [CrossRef] [PubMed]

25.

A. Mazzocca, E. Fransvea, G. Lavezzari, S. Antonaci, and G. Giannelli, “Inhibition of transforming growth factor beta receptor I kinase blocks hepatocellular carcinoma growth through neo-angiogenesis regulation,” Hepatology 50(4), 1140–1151 (2009). [CrossRef] [PubMed]

26.

I. Goze, A. Cetin, and A. Goze, “Investigation of effects of essential oils of Origanum minutiflorum O Schwarz PH Davis and Cyclotrichium niveum (Labiatae) plants on angiogenesis in shell-less chick embryo culture,” Afr. J. Biotechnol. 9, 2156–2160 (2010).

27.

P. Nowak-Sliwinska, A. Weiss, J. R. van Beijnum, T. J. Wong, J.-P. Ballini, B. Lovisa, H. van den Bergh, and A. W. Griffioen, “Angiostatic kinase inhibitors to sustain photodynamic angio-occlusion,” J. Cell. Mol. Med. 16(7), 1553–1562 (2012). [CrossRef] [PubMed]

28.

H. C. Yalcin, A. Shekhar, A. A. Rane, and J. T. Butcher, “An ex-ovo chicken embryo culture system suitable for imaging and microsurgery applications,” J. Vis. Exp. 44(44), 2154 (2010). [PubMed]

29.

S.-L. Chen, Z. Xie, L. J. Guo, and X. Wang, “A fiber-optic system for dual-modality photoacoustic microscopy and confocal fluorescence microscopy using miniature components,” Photoacoustics 1(2), 30–35 (2013). [CrossRef]

30.

E. M. Strohm, E. S. L. Berndl, and M. C. Kolios, “Probing red blood cell morphology using high-frequency photoacoustics,” Biophys. J. 105(1), 59–67 (2013). [CrossRef] [PubMed]

OCIS Codes
(170.0180) Medical optics and biotechnology : Microscopy
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(170.5120) Medical optics and biotechnology : Photoacoustic imaging

ToC Category:
Microscopy

History
Original Manuscript: August 28, 2013
Revised Manuscript: October 7, 2013
Manuscript Accepted: October 12, 2013
Published: October 25, 2013

Citation
Sung-Liang Chen, Joseph Burnett, Duxin Sun, Xunbin Wei, Zhixing Xie, and Xueding Wang, "Photoacoustic microscopy: a potential new tool for evaluation of angiogenesis inhibitor," Biomed. Opt. Express 4, 2657-2666 (2013)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-4-11-2657


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. A. W. Griffioen, G. Molema, “Angiogenesis: potentials for pharmacologic intervention in the treatment of cancer, cardiovascular diseases, and chronic inflammation,” Pharmacol. Rev. 52(2), 237–268 (2000). [PubMed]
  2. J. Folkman, “Anti-angiogenesis: new concept for therapy of solid tumors,” Ann. Surg. 175(3), 409–416 (1972). [CrossRef] [PubMed]
  3. A. R. Perez-Atayde, S. E. Sallan, U. Tedrow, S. Connors, E. Allred, J. Folkman, “Spectrum of tumor angiogenesis in the bone marrow of children with acute lymphoblastic leukemia,” Am. J. Pathol. 150(3), 815–821 (1997). [PubMed]
  4. I. Rajantie, M. Ilmonen, A. Alminaite, U. Ozerdem, K. Alitalo, P. Salven, “Adult bone marrow-derived cells recruited during angiogenesis comprise precursors for periendothelial vascular mural cells,” Blood 104(7), 2084–2086 (2004). [CrossRef] [PubMed]
  5. H. Pairleitner, H. Steiner, G. Hasenoehrl, A. Staudach, “Three-dimensional power Doppler sonography: imaging and quantifying blood flow and vascularization,” Ultrasound Obstet. Gynecol. 14(2), 139–143 (1999). [CrossRef] [PubMed]
  6. J. W. Xuan, M. Bygrave, H. Jiang, F. Valiyeva, J. Dunmore-Buyze, D. W. Holdsworth, J. I. Izawa, G. Bauman, M. Moussa, S. F. Winter, N. M. Greenberg, J. L. Chin, M. Drangova, A. Fenster, J. C. Lacefield, “Functional neoangiogenesis imaging of genetically engineered mouse prostate cancer using three-dimensional power Doppler ultrasound,” Cancer Res. 67(6), 2830–2839 (2007). [CrossRef] [PubMed]
  7. S. Makita, Y. Hong, M. Yamanari, T. Yatagai, Y. Yasuno, “Optical coherence angiography,” Opt. Express 14(17), 7821–7840 (2006). [CrossRef] [PubMed]
  8. R. K. Wang, S. L. Jacques, Z. Ma, S. Hurst, S. R. Hanson, A. Gruber, “Three dimensional optical angiography,” Opt. Express 15(7), 4083–4097 (2007). [CrossRef] [PubMed]
  9. J. A. Izatt, M. D. Kulkarni, S. Yazdanfar, J. K. Barton, A. J. Welch, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett. 22(18), 1439–1441 (1997). [CrossRef] [PubMed]
  10. H. F. Zhang, K. Maslov, G. Stoica, L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]
  11. H. F. Zhang, K. Maslov, L. V. Wang, “In vivo imaging of subcutaneous structures using functional photoacoustic microscopy,” Nat. Protoc. 2(4), 797–804 (2007). [CrossRef] [PubMed]
  12. L. V. Wang, “Tutorial on photoacoustic microscopy and computed tomography,” IEEE J. Sel. Top. Quantum Electron. 14(1), 171–179 (2008). [CrossRef]
  13. K. Maslov, H. F. Zhang, S. Hu, L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]
  14. L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nat. Photonics 3(9), 503–509 (2009). [CrossRef] [PubMed]
  15. S. S. Oladipupo, S. Hu, A. C. Santeford, J. Yao, J. R. Kovalski, R. V. Shohet, K. Maslov, L. V. Wang, J. M. Arbeit, “Conditional HIF-1 induction produces multistage neovascularization with stage-specific sensitivity to VEGFR inhibitors and myeloid cell independence,” Blood 117(15), 4142–4153 (2011). [CrossRef] [PubMed]
  16. V. J. Srinivasan, A. C. Chan, and E. Y. Lam, “Doppler OCT and OCT angiography for in vivo imaging of vascular physiology,” Chapter 2 in Selected Topics in Optical Coherence Tomography, book edited by Gangjun Liu (2012).
  17. Z. Xie, S.-L. Chen, T. Ling, L. J. Guo, P. L. Carson, X. Wang, “Pure optical photoacoustic microscopy,” Opt. Express 19(10), 9027–9034 (2011). [CrossRef] [PubMed]
  18. C. Zhang, K. Maslov, J. Yao, L. V. Wang, “In vivo photoacoustic microscopy with 7.6-µm axial resolution using a commercial 125-MHz ultrasonic transducer,” J. Biomed. Opt. 17(11), 116016 (2012). [CrossRef] [PubMed]
  19. C. Peifer, G. Dannhardt, “A novel quantitative chick embryo assay as an angiogenesis model using digital image analysis,” Anticancer Res. 24(3a), 1545–1551 (2004). [PubMed]
  20. C. N. Doukas, I. Maglogiannis, A. A. Chatziioannou, “Computer-supported angiogenesis quantification using image analysis and statistical averaging,” IEEE Trans. Inf. Technol. Biomed. 12(5), 650–657 (2008). [CrossRef] [PubMed]
  21. S.-L. Chen, Z. Xie, P. L. Carson, X. Wang, L. J. Guo, “In vivo flow speed measurement of capillaries by photoacoustic correlation spectroscopy,” Opt. Lett. 36(20), 4017–4019 (2011). [CrossRef] [PubMed]
  22. S.-L. Chen, Z. Xie, T. Ling, L. J. Guo, X. Wei, X. Wang, “Miniaturized all-optical photoacoustic microscopy based on microelectromechanical systems mirror scanning,” Opt. Lett. 37(20), 4263–4265 (2012). [CrossRef] [PubMed]
  23. D. Ribatti, “Chick embryo chorioallantoic membrane as a useful tool to study angiogenesis,” Int Rev Cell Mol Biol 270, 181–224 (2008). [CrossRef] [PubMed]
  24. A. M. Cimpean, D. Ribatti, M. Raica, “The chick embryo chorioallantoic membrane as a model to study tumor metastasis,” Angiogenesis 11(4), 311–319 (2008). [CrossRef] [PubMed]
  25. A. Mazzocca, E. Fransvea, G. Lavezzari, S. Antonaci, G. Giannelli, “Inhibition of transforming growth factor beta receptor I kinase blocks hepatocellular carcinoma growth through neo-angiogenesis regulation,” Hepatology 50(4), 1140–1151 (2009). [CrossRef] [PubMed]
  26. I. Goze, A. Cetin, A. Goze, “Investigation of effects of essential oils of Origanum minutiflorum O Schwarz PH Davis and Cyclotrichium niveum (Labiatae) plants on angiogenesis in shell-less chick embryo culture,” Afr. J. Biotechnol. 9, 2156–2160 (2010).
  27. P. Nowak-Sliwinska, A. Weiss, J. R. van Beijnum, T. J. Wong, J.-P. Ballini, B. Lovisa, H. van den Bergh, A. W. Griffioen, “Angiostatic kinase inhibitors to sustain photodynamic angio-occlusion,” J. Cell. Mol. Med. 16(7), 1553–1562 (2012). [CrossRef] [PubMed]
  28. H. C. Yalcin, A. Shekhar, A. A. Rane, J. T. Butcher, “An ex-ovo chicken embryo culture system suitable for imaging and microsurgery applications,” J. Vis. Exp. 44(44), 2154 (2010). [PubMed]
  29. S.-L. Chen, Z. Xie, L. J. Guo, X. Wang, “A fiber-optic system for dual-modality photoacoustic microscopy and confocal fluorescence microscopy using miniature components,” Photoacoustics 1(2), 30–35 (2013). [CrossRef]
  30. E. M. Strohm, E. S. L. Berndl, M. C. Kolios, “Probing red blood cell morphology using high-frequency photoacoustics,” Biophys. J. 105(1), 59–67 (2013). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited