Study on chromatic aberration in a population of Chinese myopic eyes by means of optical design

Yuanqing He, Yan Wang, Zhaoqi Wang, Chao Fang, Yongji Liu, Lin Zhang, Shaolin Zheng, Lu Wang, and Shengjiang Chang  »View Author Affiliations


Biomedical Optics Express, Vol. 4, Issue 5, pp. 667-679 (2013)
http://dx.doi.org/10.1364/BOE.4.000667


View Full Text Article

Acrobat PDF (1027 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Two kinds of individual eye models, involving and without involving the angle between visual axis and optical axis, are established by means of optical design. We use them to study the properties of the transverse chromatic aberration (TCA) and longitudinal chromatic aberration (LCA) over the visible spectrum. Then the effects of the LCA and TCA on the visual quality of human eyes are evaluated. The statistical averages of TCA and LCA over the visible spectrum for Chinese myopic eyes are obtained. Results show that both TCA and LCA restrict the visual performance, and LCA is more detrimental than TCA.

© 2013 OSA

1. Introduction

In recent years, there has been an increased interest in the study of the visual quality, especially the aberrations of human eyes. The human eye's optical media exhibit significant chromatic dispersion [1

1. J. G. Sivak and T. Mandelman, “Chromatic dispersion of the ocular media,” Vision Res. 22(8), 997–1003 (1982). [CrossRef] [PubMed]

], which is detrimental to visual performance. It is common to separate the chromatic aberration in longitudinal chromatic aberration (LCA), and transverse chromatic aberration (TCA). LCA manifests as a change in the focal point as a function of the wavelength and is also referred to as chromatic focal shift, while TCA occurs when chief rays of the spectrum passing through the center of the pupil strike on the retinal surface at different positions.

The LCA has been studied for a long time. In 1947, Wald and Griffin [2

2. G. Wald and D. R. Griffin, “The change in refractive power of the human eye in dim and bright light,” J. Opt. Soc. Am. 37(5), 321–336 (1947). [CrossRef] [PubMed]

] measured the LCA of 14 eyes over the spectrum from 405nm to 750nm and found that the average magnitude is about 2.32D. This study proceeded over a broad spectrum mostly covering the visible spectrum, but it is inevitable to introduce subjective error in the process of subjective participation. With the development of wavefront technology, it is possible to measure the LCA objectively. In 1999, Susana Marcos et al. [3

3. S. Marcos, S. A. Burns, E. Moreno-Barriusop, and R. Navarro, “A new approach to the study of ocular chromatic aberrations,” Vision Res. 39(26), 4309–4323 (1999). [CrossRef] [PubMed]

] obtained the magnitude of LCA through measuring the wavefront aberration and it was about 1.26D over the spectrum from 450nm to 650nm.

For the measurement of TCA, two-color, vernier alignment technique is the most common. In 1995, Rynders [4

4. M. Rynders, B. Lidkea, W. Chisholm, and L. N. Thibos, “Statistical distribution of foveal transverse chromatic aberration, pupil centration, and angle psi in a population of young adult eyes,” J. Opt. Soc. Am. A 12(10), 2348–2357 (1995). [CrossRef] [PubMed]

] obtained the statistical distribution of the subjective TCA for 170 eyes over the spectrum from 497nm to 605nm by means of a two-dimensional, two-color, vernier alignment technique. Through rearranging to make the horizontally oriented lines and vertically oriented lines in the same target, it is allowed to measure horizontal TCA and vertical TCA respectively. Results show that the absolute magnitude of TCA is 0.82arcmin in average. However, this experiment needs the participation of subjects, and only the TCA over the fixed spectrum band can be obtained while this spectrum band does not cover the visible spectrum properly.

Furthermore, several studies attempted to evaluate the effect of the LCA and TCA on the visual quality. Susana Marcos et al. [3

3. S. Marcos, S. A. Burns, E. Moreno-Barriusop, and R. Navarro, “A new approach to the study of ocular chromatic aberrations,” Vision Res. 39(26), 4309–4323 (1999). [CrossRef] [PubMed]

] theoretically computed the effect of LCA and TCA on the visual quality and evaluated which one was more detrimental to the visual performance of human eye. They found that TCA could degrade the visual quality as much as LCA. Correcting TCA and leaving LCA uncorrected could improve the visual quality more than correcting the LCA and leaving the TCA uncorrected.

In this paper, two kinds of individual eye models, with and without involving the angle between visual axis and optical axis are established for 80 myopic eyes with the optical design software ZEMAX, combined with the measurement data of subjects including wavefront aberrations, axial lengths and corneal topography. With the eye models, we investigate the properties of the TCA and LCA over the visible spectrum from 400nm to 760nm and evaluate the effects on visual quality. This method is objective, which avoids possible errors of subjective measurements.

It is reported clinically that the patients who have received refractive surgery may possess a varied color perception. Therefore it is meaningful to statistically investigate the chromatic aberrations of myopic eyes before and after the surgery. This paper shows our preliminary results in this aspect.

2. Methods

A group of 80 eyes (41 OD and 39 OS) of 46 Chinese subjects is selected. Among them 68 eyes are from 34 subjects. All of these eyes are myopic with spherical refractions in the range of −0.75D to −8.5D. The subjects have neither ocular disease nor corneal surgery. The wavefront aberrations and the corneal topographies are respectively measured by Hartmann-Shack wavefront aberrometer [5

5. J. Liang and D. R. Williams, “Aberrations and retinal image quality of the normal human eye,” J. Opt. Soc. Am. A 14(11), 2873–2883 (1997). [CrossRef] [PubMed]

] – Wavescan (VISX, Santa Clara, CA, USA) and Pentacam (Oculus GmbH, Wetzlar, Germany) in the case of darkroom with natural pupils of the diameter range from 6mm to 7.5mm. During the measurements, subjects are asked to stare at the target in the apparatus to make the visual axis of the eye coincident with the optical axis of the apparatus. The axial lengths are measured by the Ultrasonic A/B Scanner (MEDA Co., Tianjin, China). We construct two kinds of individual eye models, involving and without involving the angle between visual axis and optical axis, with the measurement data to investigate the properties of the TCA and the LCA respectively.

2.1 Construction of the fundamental eye model

2.2 Study on LCA by the individual eye model of visual axis coincident with optical axis

Corneal surface is the most powerful refracting surface in the eye, providing 2/3 of the eye’s focusing power. Especially, the angle between visual axis and optical axis is determined by the shape of the corneal surface [9

9. R. Navarro, L. González, and J. L. Hernández, “Optics of the average normal cornea from general and canonical representations of its surface topography,” J. Opt. Soc. Am. A 23(2), 219–232 (2006). [CrossRef]

]. In this research, curvature radii and elevation maps of corneal surfaces are obtained with Pentacam. The Pentacam is a rotating Scheimpflug camera and the rotational measuring procedure generates Scheimpflug images. It calculates 3-dimensional model of the anterior eye segment from 25000 true elevation points. The elevation of corneal surface can be expressed as the Zernike Fringe Sag type [10

10. Zemax Optical Design Program User’s Guide (Zemax Development Corporation, 2005), Chap. 11, pp. 229–230.

], expressed as:
Z=z0+c(x2+y2)1+[1-c2(x2+y2)]12+i=1NAiZi(x,y)
(1)
Where c is the corneal surface curvature provided by Pentacam, Ziis theith Zernike Fringe polynomial, Ai is the coefficient on the ith Zernike Fringe polynomial, N is the number of Zernike coefficients in the series and z0 is a constant. The fitting procedure is implemented with MATLAB. To build up the individual eye model, we substitute the curvature radii and elevation data of the anterior and posterior corneal surfaces for the correspondent surface data in the fundamental eye model.

The thicknesses of cornea, anterior chamber and vitreous in the fundamental eye model are then substituted with the correspondent measurement values of the axial lengths provided by the ultrasonic thickness gauge.

In ZEMAX, we set entrance pupil diameter to be 6mm and wavelength to be 555nm for the eye model. One of the procedures of building up individual eye model is to transfer the wavefront aberrations to the fundamental eye model [11

11. H. Guo, Z. Wang, Q. Zhao, W. Quan, and Y. Wang, “Individual eye model based on wavefront aberration,” Optik (Stuttg.) 116(2), 80–85 (2005). [CrossRef]

]. In the measurement of wavefront aberrations, the pupil diameters of subjects are different. To ensure the consistency, a conversion of aberrations from the actual pupil diameter to the fixed pupil diameter of 6mm is carried out by programming in MATLAB. Although it is more correct to take the wavefront aberrations over the actual pupil diameter in the first place, the error introduced by the conversion is quite small [12

12. G. M. Dai, “Scaling Zernike expansion coefficients to smaller pupil sizes: a simpler formula,” J. Opt. Soc. Am. A 23(3), 539–543 (2006). [CrossRef] [PubMed]

]. In order to make the wavefront aberrations of the eye model equal to that of the actual eye, we take the aberrations as the target of the merit function and optimize the anterior surface of the crystalline lens of the eye model. In this way the individual eye model of visual axis coincident with optical axis is established. Figure 1
Fig. 1 Individual eye model of visual axis coincident with optical axis.
shows the layout of the model. It can be seen that there is an obvious defocus due to the myopia.

To investigate the properties of LCA over the visible spectrum, the working spectrum of eye model is set from 400nm to 760nm with the primary wavelength of 555nm. Figure 2
Fig. 2 LCA of the individual eye model.
shows the LCA of the individual eye model over the visible spectrum by means of ray trace. The ray of primary wavelength (green) does not focus to the image surface (retina) exactly because of myopia, and the ray of 400nm (blue) focuses in front of the focal point of the ray of primary wavelength while the ray of 760nm (red) focuses in the behind. The distance between focal points is the magnitude of LCA over the visible spectrum.

In ZEMAX, the chromatic focal shift plot shows the distance between the focal point of the ray at certain wavelength and that at the primary wavelength with unit of μm, which represents the LCA. Figure 3
Fig. 3 Chromatic focal shift plot of individual eye model.
is the plot of an individual eye model where abscissa is the wavelength and the ordinate is the chromatic focal shift. It can be seen that the trend of LCA is flatter in long wavelengths band compared to that in short wavelengths band.

In term of the Newton formula, we can also express the LCA with refraction in diopters (D):
ΔP=Δxff
(2)
Where Δxis the magnitude of the chromatic focal shift over the visible spectrum, fand fare the focal length in object space and in image space respectively.

The effect of LCA on the image quality of human eyes is the expansion of the image spot on retina. Ideally, the image spot of the ray of primary wavelength (555nm) ought to be airy disk in the case of diffraction limitation. The existence of LCA enlarges the scale of image spot. The relational expression of the diameter of the spot as the chromatic focal shift is [13

13. R. Li, Z. Wang, Y. Liu, and G. Mu, “A method to design aspheric spectacles for correction of high-order aberrations of human eye,” Sci. China Technol. Sci. 55(5), 1391–1401 (2012). [CrossRef]

]:
d=LΔx1+fΔx
(3)
Where d is the diameter of spot, L is the entrance pupil diameter,f is the focal length in object space, and Δx is the magnitude of chromatic focal shift over the spectrum from 400nm to 555nm or from 555nm to 760nm. We choose the one yielding bigger image spot on retina as reference. The calculation is under the photopic condition (L = 3mm).

2.3 Study on TCA by the individual eye model involving the angle between visual and optical axis

The visual axis of human eye does not coincide with the optical axis, so the TCA is not zero at the fovea. The angle between visual axis and optical axis is determined by the elevation data of the anterior corneal surface [9

9. R. Navarro, L. González, and J. L. Hernández, “Optics of the average normal cornea from general and canonical representations of its surface topography,” J. Opt. Soc. Am. A 23(2), 219–232 (2006). [CrossRef]

]. The raw elevation data measured by Pentacam can be fit to a parametric model, expressed as the sum of two terms:
Z(x,y)=S(x,y)+R(x,y)
(4)
Where S(x,y) is a regular basis representing the overall shape of the corneal surface, and R(x,y)is a residual, accounting for irregularities and departures from the overall shape of the basis. It is reasonable to choose the best fitting shape of the corneal surface as the basis and make the residual R(x,y)as low as possible. In this way, the analysis of the basis S(x,y) will provide meaningful overall properties. From earlier studies, the overall shape of the corneal surface is close to ellipsoid [14

14. H. Burek and W. A. Douthwaite, “Mathematical models of the general corneal surface,” Ophthalmic Physiol. Opt. 13(1), 68–72 (1993). [CrossRef] [PubMed]

]. To fit the basis properly, we use the Cartesian coordinate system and set the z-axis of the coordinate system along the visual axis. In this coordinate system, the basis surface can be expressed as a general quadric format:

a11x2+a22y2+a33z2+a12xy+a13xz+a23yz+a1x+a2y+a3z+a0=0
(5)

Then we set a new coordinate system(x0,y0,z0) in which the quadric takes its canonical form expressed as:

x02a2+y02b2+z02c2=1
(6)

The center of the ellipsoid is the origin of this coordinate, with three axes along three orthogonal axes of the ellipsoid and the z0-axis along the optical axis. We transform the general quadric system to the canonical system by means of orthogonal transformation, including one rotation and one translation. The rotation angle is just equivalent to the angle between visual axis and optical axis. This fitting procedure is implemented with MATLAB.

In ZEMAX, the lateral color plot provides the magnitude of TCA. Figure 5
Fig. 5 Lateral color plot of individual eye model.
is the lateral color plot of the individual eye model where the abscissa indicates the magnitude of TCA (μm) and the ordinate represents the field of view (FOV) with 1.0 standing for the whole FOV. In this case, the angle between visual axis and optical axis is obtained by means of orthogonal transformation, which is 4.64° in horizontal direction and 1.5° in vertical direction. Therefore, we set 4.64° FOV in X-direction and 1.5° FOV in Y-direction for the eye model. The black lines show the scale of airy disk under the photopic condition.

The magnitude of horizontal or vertical TCA can be obtained by setting FOV in the correspondent direction only. Figures 6(a)
Fig. 6 Lateral color plot of individual eye model, (a) 4.64°FOV in X-direction, (b) 1.5°FOV in Y-direction.
and 6(b) are the lateral color plots with the set of 4.64° FOV in X-direction and the set of 1.5° FOV in Y-direction respectively.

3. Results

3.1 Statistical results of TCA

TCA over the visible spectrum from 400nm to 760nm of 80 eyes with myopia are investigated based on the individual eye model involving the angle, and the results of horizontal, vertical and the absolute TCA are listed in columns 4 to 7 in Table 2

Table 2. Statistics of spherical refraction (Ps), cylinder (Pc), horizontal TCA in arcmin, vertical TCA in arcmin, absolute TCA in arcmin, absolute TCA in μm, diameters of airy disk, chromatic focal shift, LCA in D, and diameter of image spot caused by LCA over the visible spectrum under the photopic condition

table-icon
View This Table
| View All Tables
. The magnitudes are read from “lateral color”, and the signs are obtained by ray tracing. Besides, the spherical refractions (Ps) and the cylinder refractions (Pc) are listed in columns 2 and 3, respectively. To evaluate the effect of the TCA on the visual quality of human eyes, the diameters of airy disk under the photopic condition are also listed in column 8. The last two rows give the average and the maximum values respectively.

In Table 2, NO.1 to NO.41 eyes are the right eyes with the signs of horizontal TCA being negative, and the rest 39 eyes are the left eyes with the signs of horizontal TCA being positive. The absolute values are taken for the calculation of the average of TCA. The average of horizontal TCA is 1.38 ± 0.61 arcmin while the average of vertical TCA is 0.82 ± 0.56 arcmin. The statistical average of the absolute TCA over the visible spectrum is 8.86 ± 3.56μm, corresponding to 1.67 ± 0.68 arcmin in angular expression, and the maximum is 16.69μm or 3.17 arcmin. TCA varies widely across subjects.

The distributions of the horizontal and vertical TCA are also plotted in Fig. 7
Fig. 7 Distribution of the horizontal TCA and vertical TCA over the visible spectrum from 400nm to 760nm for 80 eyes.
according to the data in Table 2. The abscissa indicates the horizontal TCA while the ordinate indicates vertical TCA. The black square represents the right eye and the red star represents the left eye. It can be seen in horizontal direction that the distribution of TCA is at the negative half of the x-axis for all right eyes, and it is at the positive half for all left eyes. This indicates that the orientations of the visual axis in horizontal direction trend to the nasal side for all eyes. In vertical direction, the signs of TCA are irregular, indicating that the orientations of visual axis in vertical direction are indeterminate.

As shown in Table 2, the diameter of airy disk is about 7.79μm in average. There are 31 eyes with the magnitudes of TCA less than the diameter of airy disk accounting for 38.75% of all eyes, 35 eyes with the magnitudes ranging from 1 to 1.5 times as large as the diameter of airy disk accounting for 43.75%, and 14 eyes with the magnitudes larger than 1.5 times of the diameter of airy disk accounting for 17.5%. Statistically, the magnitude of TCA is slightly larger than the diameter of airy disk, but the maximum is about 2.17 times as large as the diameter of airy disk. We conclude that the effect of TCA on the visual quality of human eyes cannot be ignored.

3.2 Statistical results of LCA

LCA over the visible spectrum from 400nm to 760nm of 80 eyes are investigated based on the individual eye model of visual axis coincident with optical axis. The chromatic focal shift, the magnitudes of LCA and the diameters of image spot caused by LCA under the photopic condition are listed in columns 9, 10 and 11 in Table 2, respectively.

The statistical average of LCA over the visible spectrum is 2.13 ± 0.05D, corresponding to the chromatic focal shift of 782 ± 21μm in average. It can be seen that the LCA varies little across subjects. Under the photopic condition, the diameter of image spot caused by the LCA is 70.42μm in average, which is about 8 times larger than the average of TCA (8.86μm) and 4 times larger than the maximum of TCA (16.96μm). This indicates that both TCA and LCA restrict the visual performance and LCA is more detrimental than TCA.

To further illustrate the effects of LCA and TCA on the visual quality we present the MTF curves of the eye model in both single wavelength (555nm) and visible light (from 473nm to 601nm as [3

3. S. Marcos, S. A. Burns, E. Moreno-Barriusop, and R. Navarro, “A new approach to the study of ocular chromatic aberrations,” Vision Res. 39(26), 4309–4323 (1999). [CrossRef] [PubMed]

]). The eye model used is the fundamental eye model to simulate the monochromatic-aberration-free eyes. The variation of the MTF at 0° FOV represents the influence of LCA-only on the visual quality, and that at 3° FOV represents the influence of the sum of TCA and LCA. This is because the influence of the angle between visual axis and optical axis is just as the influence of the FOV at the same angle, and we cannot separate the TCA from the total chromatic aberrations in the eye model. The angle of 3° is the average angle formed by visual axis and optical axis in this research. Figure 8
Fig. 8 MTF for the fundamental eye model, (a) with single wavelength.(b) with polychromatic light.
shows the MTF curves with (a) by single wavelength setting and (b) by visible light setting. Symbols of 1 and 2 represent the FOV at 0° and 3° receptively, and the black curve represents the diffraction limit. It can be seen that curve 1 in (b) drops seriously compared to that in (a), indicating the serious influence of LCA on the visual quality. Curve 2 is slightly lower than curve 1 in (b), indicating the minor influence of TCA on the visual quality in the presence of LCA. Sowmya Ravikumar et al. [15

15. S. Ravikumar, L. N. Thibos, and A. Bradley, “Calculation of retinal image quality for polychromatic light,” J. Opt. Soc. Am. A 25(10), 2395–2407 (2008). [CrossRef] [PubMed]

] demonstrated that in the presence of typical levels of monochromatic aberrations, the effect of TCA was greatly reduced. Our result is somehow in an agreement with their conclusion.

The spot diagrams of this eye model are presented in Fig. 9
Fig. 9 Spot diagram for the fundamental eye model, (a) and (b) with single wavelength, (c) and (d) with polychromatic light.
. Where (a) and (b) are the spot diagrams in single wavelength of 555nm at 0° FOV and 3° FOV respectively, while (c) and (d) are the spot diagrams in visible light at 0° FOV and 3° FOV respectively. The black circles show the scale of airy disk. It can be seen in a comparison of (c) and (d) that the influence of TCA is negligible in the presence of LCA.

Figure 10
Fig. 10 Spot diagrams of the eye with typical magnitude of monochromatic aberrations. (a) Without involving the angle, (b) involving the angle.
is the spot diagrams of the eye with typical magnitude of monochromatic aberrations, the astigmatism and spherical aberration of one of the subjects. Where (a) is for the eye model without involving the angle between visual axis and optical axis, and (b) is for the eye model involving the angle. It can be seen as compared with Fig. 9 that the influence of the monochromatic aberrations on visual quality is less than the influence of LCA.

Figure 11
Fig. 11 The relevance between refraction and chromatic aberration.
shows the relevance between TCA and refraction, and between LCA and refraction. The abscissa is the refraction of subjects in D, and the ordinate is the magnitude of TCA in arcmin and LCA in D. The black dots represent the TCA while the red stars represent the LCA. It can be seen that both of TCA and LCA have no obvious relevance to the refraction.

4. Conclusions

The properties of TCA and LCA over visible spectrum from 400nm to 760nm of 80 myopic eyes are studied by constructing two kinds of individual eye models, involving and without involving the angle formed by visual axis and optical axis.

The statistical average of LCA is 2.13 ± 0.05D which is similar to the results in [2

2. G. Wald and D. R. Griffin, “The change in refractive power of the human eye in dim and bright light,” J. Opt. Soc. Am. 37(5), 321–336 (1947). [CrossRef] [PubMed]

,3

3. S. Marcos, S. A. Burns, E. Moreno-Barriusop, and R. Navarro, “A new approach to the study of ocular chromatic aberrations,” Vision Res. 39(26), 4309–4323 (1999). [CrossRef] [PubMed]

]. The trend of LCA is flatter in long wavelength band as compared to that in short wavelength band. LCA varies little across subjects.

The effects of both LCA and TCA on the visual quality are evaluated with the scale of image spot and the MTF curves. The magnitude of TCA is slightly larger than the diameter of airy disk, but the maximum is about 2.17 times as large as the diameter of airy disk. The effect of TCA on the image quality cannot be ignored. Under the photopic condition, the diameter of image spot caused by LCA is 70.42μm in average, which is about 8 times larger than the average magnitude of TCA (8.86μm) and 4 times larger than the maximum (16.96μm). Therefore, both TCA and LCA restrict the visual performance, and LCA is more detrimental than TCA.

It is shown with the MTF analysis that the chromatic aberrations degrade the visual performance to a certain extent, with the effect of LCA more serious. In the presence of LCA the influence of TCA is greatly reduced.

Acknowledgment

This work is supported by the National Nature Science Foundation of China (No. 81170873), and the National Nature Science Foundation of China (No. 11104149).

References and links

1.

J. G. Sivak and T. Mandelman, “Chromatic dispersion of the ocular media,” Vision Res. 22(8), 997–1003 (1982). [CrossRef] [PubMed]

2.

G. Wald and D. R. Griffin, “The change in refractive power of the human eye in dim and bright light,” J. Opt. Soc. Am. 37(5), 321–336 (1947). [CrossRef] [PubMed]

3.

S. Marcos, S. A. Burns, E. Moreno-Barriusop, and R. Navarro, “A new approach to the study of ocular chromatic aberrations,” Vision Res. 39(26), 4309–4323 (1999). [CrossRef] [PubMed]

4.

M. Rynders, B. Lidkea, W. Chisholm, and L. N. Thibos, “Statistical distribution of foveal transverse chromatic aberration, pupil centration, and angle psi in a population of young adult eyes,” J. Opt. Soc. Am. A 12(10), 2348–2357 (1995). [CrossRef] [PubMed]

5.

J. Liang and D. R. Williams, “Aberrations and retinal image quality of the normal human eye,” J. Opt. Soc. Am. A 14(11), 2873–2883 (1997). [CrossRef] [PubMed]

6.

D. A. Atchison and G. Smith, “Chromatic dispersions of the ocular media of human eyes,” J. Opt. Soc. Am. A 22(1), 29–37 (2005). [CrossRef] [PubMed]

7.

H.-L. Liou and N. A. Brennan, “Anatomically accurate, finite model eye for optical modeling,” J. Opt. Soc. Am. A 14(8), 1684–1695 (1997). [CrossRef] [PubMed]

8.

Y. Zhang, Applied Optics (Publishing House of Electronics Industry, 2008), Chap. 14.

9.

R. Navarro, L. González, and J. L. Hernández, “Optics of the average normal cornea from general and canonical representations of its surface topography,” J. Opt. Soc. Am. A 23(2), 219–232 (2006). [CrossRef]

10.

Zemax Optical Design Program User’s Guide (Zemax Development Corporation, 2005), Chap. 11, pp. 229–230.

11.

H. Guo, Z. Wang, Q. Zhao, W. Quan, and Y. Wang, “Individual eye model based on wavefront aberration,” Optik (Stuttg.) 116(2), 80–85 (2005). [CrossRef]

12.

G. M. Dai, “Scaling Zernike expansion coefficients to smaller pupil sizes: a simpler formula,” J. Opt. Soc. Am. A 23(3), 539–543 (2006). [CrossRef] [PubMed]

13.

R. Li, Z. Wang, Y. Liu, and G. Mu, “A method to design aspheric spectacles for correction of high-order aberrations of human eye,” Sci. China Technol. Sci. 55(5), 1391–1401 (2012). [CrossRef]

14.

H. Burek and W. A. Douthwaite, “Mathematical models of the general corneal surface,” Ophthalmic Physiol. Opt. 13(1), 68–72 (1993). [CrossRef] [PubMed]

15.

S. Ravikumar, L. N. Thibos, and A. Bradley, “Calculation of retinal image quality for polychromatic light,” J. Opt. Soc. Am. A 25(10), 2395–2407 (2008). [CrossRef] [PubMed]

OCIS Codes
(330.4595) Vision, color, and visual optics : Optical effects on vision
(330.7325) Vision, color, and visual optics : Visual optics, metrology
(330.7326) Vision, color, and visual optics : Visual optics, modeling

ToC Category:
Vision and Visual Optics

History
Original Manuscript: January 9, 2013
Revised Manuscript: March 15, 2013
Manuscript Accepted: March 30, 2013
Published: April 3, 2013

Citation
Yuanqing He, Yan Wang, Zhaoqi Wang, Chao Fang, Yongji Liu, Lin Zhang, Shaolin Zheng, Lu Wang, and Shengjiang Chang, "Study on chromatic aberration in a population of Chinese myopic eyes by means of optical design," Biomed. Opt. Express 4, 667-679 (2013)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-4-5-667


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. J. G. Sivak and T. Mandelman, “Chromatic dispersion of the ocular media,” Vision Res.22(8), 997–1003 (1982). [CrossRef] [PubMed]
  2. G. Wald and D. R. Griffin, “The change in refractive power of the human eye in dim and bright light,” J. Opt. Soc. Am.37(5), 321–336 (1947). [CrossRef] [PubMed]
  3. S. Marcos, S. A. Burns, E. Moreno-Barriusop, and R. Navarro, “A new approach to the study of ocular chromatic aberrations,” Vision Res.39(26), 4309–4323 (1999). [CrossRef] [PubMed]
  4. M. Rynders, B. Lidkea, W. Chisholm, and L. N. Thibos, “Statistical distribution of foveal transverse chromatic aberration, pupil centration, and angle psi in a population of young adult eyes,” J. Opt. Soc. Am. A12(10), 2348–2357 (1995). [CrossRef] [PubMed]
  5. J. Liang and D. R. Williams, “Aberrations and retinal image quality of the normal human eye,” J. Opt. Soc. Am. A14(11), 2873–2883 (1997). [CrossRef] [PubMed]
  6. D. A. Atchison and G. Smith, “Chromatic dispersions of the ocular media of human eyes,” J. Opt. Soc. Am. A22(1), 29–37 (2005). [CrossRef] [PubMed]
  7. H.-L. Liou and N. A. Brennan, “Anatomically accurate, finite model eye for optical modeling,” J. Opt. Soc. Am. A14(8), 1684–1695 (1997). [CrossRef] [PubMed]
  8. Y. Zhang, Applied Optics (Publishing House of Electronics Industry, 2008), Chap. 14.
  9. R. Navarro, L. González, and J. L. Hernández, “Optics of the average normal cornea from general and canonical representations of its surface topography,” J. Opt. Soc. Am. A23(2), 219–232 (2006). [CrossRef]
  10. Zemax Optical Design Program User’s Guide (Zemax Development Corporation, 2005), Chap. 11, pp. 229–230.
  11. H. Guo, Z. Wang, Q. Zhao, W. Quan, and Y. Wang, “Individual eye model based on wavefront aberration,” Optik (Stuttg.)116(2), 80–85 (2005). [CrossRef]
  12. G. M. Dai, “Scaling Zernike expansion coefficients to smaller pupil sizes: a simpler formula,” J. Opt. Soc. Am. A23(3), 539–543 (2006). [CrossRef] [PubMed]
  13. R. Li, Z. Wang, Y. Liu, and G. Mu, “A method to design aspheric spectacles for correction of high-order aberrations of human eye,” Sci. China Technol. Sci.55(5), 1391–1401 (2012). [CrossRef]
  14. H. Burek and W. A. Douthwaite, “Mathematical models of the general corneal surface,” Ophthalmic Physiol. Opt.13(1), 68–72 (1993). [CrossRef] [PubMed]
  15. S. Ravikumar, L. N. Thibos, and A. Bradley, “Calculation of retinal image quality for polychromatic light,” J. Opt. Soc. Am. A25(10), 2395–2407 (2008). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited