In vivo quantification of the scattering properties of tissue using multi-diameter single fiber reflectance spectroscopy

F. van Leeuwen-van Zaane, U. A. Gamm, P. B. A. A. van Driel, T. J. A. Snoeks, H. S. de Bruijn, A. van der Ploeg-van den Heuvel, I. M. Mol, C. W. G. M. Löwik, H. J. C. M. Sterenborg, A. Amelink, and D. J. Robinson  »View Author Affiliations


Biomedical Optics Express, Vol. 4, Issue 5, pp. 696-708 (2013)
http://dx.doi.org/10.1364/BOE.4.000696


View Full Text Article

Acrobat PDF (1390 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Multi diameter single fiber reflectance (MDSFR) spectroscopy is a non-invasive optical technique based on using multiple fibers of different diameters to determine both the reduced scattering coefficient (μs′) and a parameter γ that is related to the angular distribution of scattering, where γ = (1-g2)/(1-g1) and g1 and g2 the first and second moment of the phase function, respectively. Here we present the first in vivo MDSFR measurements of μs′(λ) and γ(λ) and their wavelength dependence. MDSFR is performed on nineteen mice in four tissue types including skin, liver, normal tongue and in an orthotopic oral squamous cell carcinoma. The wavelength-dependent slope of μs′(λ) (scattering power) is significantly higher for tongue and skin than for oral cancer and liver. The reduced scattering coefficient at 800 nm of oral cancer is significantly higher than of normal tongue and liver. Gamma generally increases with increasing wavelength; for tumor it increases monotonically with wavelength, while for skin, liver and tongue γ(λ) reaches a plateau or even decreases for longer wavelengths. The mean γ(λ) in the wavelength range 400-850 nm is highest for liver (1.87 ± 0.07) and lowest for skin (1.37 ± 0.14). Gamma of tumor and normal tongue falls in between these values where tumor exhibits a higher average γ(λ) (1.72 ± 0.09) than normal tongue (1.58 ± 0.07). This study shows the potential of using light scattering spectroscopy to optically characterize tissue in vivo.

© 2013 OSA

1. Introduction

At large source-detector separations, light transport can be considered diffuse and is therefore only dependent on the absorption coefficient (μa) and the reduced scattering coefficient (μs′), given by μs=μs(1g1) where g1=cos(θ) is the first moment of the scattering phase function (PF), also called the scattering anisotropy. At these large distances, light transport is insensitive to the exact shape of the PF. However, in device configurations with small source-detector separations, an additional, PF dependent variable γ is necessary to describe the effect of large angle scattering events on the reflectance signal [4

4. H. Tian, Y. Liu, and L. Wang, “Influence of the third-order parameter on diffuse reflectance at small source-detector separations,” Opt. Lett. 31(7), 933–935 (2006). [CrossRef] [PubMed]

7

7. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]

]. This parameter γ is defined as
γ=1g21g1
(1)
where g2 is the second Legendre moment of the PF [5

5. F. Bevilacqua and C. Depeursinge, “Monte Carlo study of diffuse reflectance at source−detector separations close to one transport mean free path,” J. Opt. Soc. Am. A 16(12), 2935–2945 (1999). [CrossRef]

].

Our group has previously developed single fiber reflectance spectroscopy (SFR), where one fiber serves as both source and detector. One advantage of this geometry is that SFR is sensitive to superficial tissues and their microvasculature, which are susceptible to early changes in morphology caused by (pre-)malignant disease. Furthermore, the diameter of a single fiber probe is such that it can be easily guided through the working channel of and endoscope to sample internal organs. The analysis of SFR spectra is performed using a semi-empirical model (cf. Eq. (2)) that describes the PF-dependent relationship between SFR signal and the dimensionless reduced scattering (which is defined as the product of the reduced scattering coefficient and the fiber diameter μs′ df) [6

6. S. C. Kanick, U. A. Gamm, M. Schouten, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Measurement of the reduced scattering coefficient of turbid media using single fiber reflectance spectroscopy: fiber diameter and phase function dependence,” Biomed. Opt. Express 2(6), 1687–1702 (2011). [CrossRef] [PubMed]

]. This semi-empirical model was based on Monte-Carlo simulations and validated in tissue mimicking phantoms [8

8. S. C. Kanick, D. J. Robinson, H. J. Sterenborg, and A. Amelink, “Monte Carlo analysis of single fiber reflectance spectroscopy: photon path length and sampling depth,” Phys. Med. Biol. 54(22), 6991–7008 (2009). [CrossRef] [PubMed]

,9

9. S. C. Kanick, H. J. Sterenborg, and A. Amelink, “Empirical model of the photon path length for a single fiber reflectance spectroscopy device,” Opt. Express 17(2), 860–871 (2009). [CrossRef] [PubMed]

]. SFR spectroscopy allows us to determine tissue absorption without prior knowledge of the scattering coefficient [10

10. S. C. Kanick, D. J. Robinson, H. J. C. M. Sterenborg, and A. Amelink, “Method to quantitate absorption coefficients from single fiber reflectance spectra without knowledge of the scattering properties,” Opt. Lett. 36(15), 2791–2793 (2011). [CrossRef] [PubMed]

]. Recently, SFR has been extended to multi-diameter single fiber reflectance spectroscopy (MDSFR), in which two fiber diameters can be used to determine the μs′ and γ [7

7. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]

,11

11. S. C. Kanick, U. A. Gamm, H. J. C. M. Sterenborg, D. J. Robinson, and A. Amelink, “Method to quantitatively estimate wavelength-dependent scattering properties from multidiameter single fiber reflectance spectra measured in a turbid medium,” Opt. Lett. 36(15), 2997–2999 (2011). [CrossRef] [PubMed]

,12

12. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Measurement of tissue scattering properties using multi-diameter single fiber reflectance spectroscopy: in silico sensitivity analysis,” Biomed. Opt. Express 2(11), 3150–3166 (2011). [CrossRef] [PubMed]

]. An MDSFR measurement consists of several co-localized SFR measurements using different fiber diameters [11

11. S. C. Kanick, U. A. Gamm, H. J. C. M. Sterenborg, D. J. Robinson, and A. Amelink, “Method to quantitatively estimate wavelength-dependent scattering properties from multidiameter single fiber reflectance spectra measured in a turbid medium,” Opt. Lett. 36(15), 2997–2999 (2011). [CrossRef] [PubMed]

], which are each individually corrected for absorption [10

10. S. C. Kanick, D. J. Robinson, H. J. C. M. Sterenborg, and A. Amelink, “Method to quantitate absorption coefficients from single fiber reflectance spectra without knowledge of the scattering properties,” Opt. Lett. 36(15), 2791–2793 (2011). [CrossRef] [PubMed]

], resulting in the reflectance in the absence of absorption,RSF0, for every SFR measurement. The co-localized, multi-fiber diameter RSF0values are then simultaneously fitted to the model described in Eq. (2),
RSF0(λ,df)=ηlim(1+p1ep3μs(λ)df)[(μs(λ)df)p2p3+(μs(λ)df)p2]
(2)
where ηlim is the diffuse limit of the single fiber collection efficiency and [p1 p2 p3] are γ(λ)-dependent coefficients, the values of which are derived from Monte Carlo simulations. The spectral shape of μs′(λ) is constrained according to a biologically realistic power law decay. Solving the equation with two unknown variables for two different fiber diameters allows wavelength dependent quantification of both μs′ and γ. The MDSFR approach has been validated both in silico and in optical phantoms containing polystyrene microspheres over a wide range of biologically relevant values for μs′(λ) and γ(λ), with and without the presence of absorbers [7

7. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]

,12

12. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Measurement of tissue scattering properties using multi-diameter single fiber reflectance spectroscopy: in silico sensitivity analysis,” Biomed. Opt. Express 2(11), 3150–3166 (2011). [CrossRef] [PubMed]

].

In the current study we present, for the first time, differences in the wavelength dependence of μs′ and γ in vivo, measured using MDSFR. We have acquired MDSFR spectra in vivo from tumor, normal tongue tissue, skin and liver in an orthotopic xenograft model for oral cancer in mice using consecutive SFR measurements of two different fiber diameters (0.4 and 0.8 mm).

2. Methods

2.1 MDSFR reflectance model

The aim of this first step of the analysis is to remove the effects of absorption on the individual SFR spectra. It was shown previously that accurate absorption coefficients from individual SFR spectra can be obtained without prior knowledge of the tissue scattering properties [7

7. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]

,10

10. S. C. Kanick, D. J. Robinson, H. J. C. M. Sterenborg, and A. Amelink, “Method to quantitate absorption coefficients from single fiber reflectance spectra without knowledge of the scattering properties,” Opt. Lett. 36(15), 2791–2793 (2011). [CrossRef] [PubMed]

] if an optimized set [p1 p2 p3 CPF] is used, i.e. p1 = 1.55, p2 = 0.969, p3 = 6.82, and CPF = 0.944. The background scattering model used in this first step of individual analysis of each SFR spectrum is based on a 4th order polynomial, i.e.
μs(λ)=a1(λλ0)1+a2(λλ0)2+a3(λλ0)3+a4(λλ0)4
(10)
This model allows sufficient degrees of freedom to correct for the physically incorrect assumption that [p1 p2 p3 CPF] are independent of γ. This aspect of the model will be further addressed in the discussion section. After removal of the effect of absorption on the individual SFR spectra, the MDSFR analysis is performed by solving Eq. (2) for multiple fiber diameters simultaneously. Here, the reduced scattering coefficient is modeled using the power law function:
μs(λ)=a1(λλ0)a2
(11)
and Eq. (2) and Eq. (4) are adapted to yield a wavelength-dependent γ, by replacing the coefficients [p1 p2 p3] with p1=0.63γ2p2=0.57γp3=2.31γ2 in Eq. (2) [11

11. S. C. Kanick, U. A. Gamm, H. J. C. M. Sterenborg, D. J. Robinson, and A. Amelink, “Method to quantitatively estimate wavelength-dependent scattering properties from multidiameter single fiber reflectance spectra measured in a turbid medium,” Opt. Lett. 36(15), 2997–2999 (2011). [CrossRef] [PubMed]

] and CPF=0.68γ0.6 in Eq. (4). A Levenberg-Marquart non-linear fit procedure is used to estimate the variables μs′(λ) and γ(λ) from the MDSFR spectra of two different fiber diameters simultaneously.

Finally, since the MDSFR analysis yields estimates of μs′(λ) and γ(λ), these values can now be used in the individual SFR fits to re-estimate the absorption coefficient. In the first step, the absorption coefficient was estimated without knowledge of the scattering properties by assuming fixed (wavelength independent) values for [p1 p2 p3 CPF] and using a 4th order polynomial for μs′(λ); the accuracy of that approach is now verified by using the extracted wavelength dependent estimates of γ and using a biologically realistic shape for μs′(λ), i.e. a power law.

2.2 Experimental setup

The illustration in Fig. 1
Fig. 1 Schematic diagram of the measurement setup. Reflectance and fluorescence are measured through a single fiber of either 0.4 or 0.8 mm. Two identical setups are used to accommodate 2 fiber diameters.
represents a single probe-unit of the MDSFR spectroscopic setup.During reflectance measurements, white light emitted by a halogen light source [HL-2000-FHSA; Ocean Optics; Duiven, NL] is directed through the first fiber of a quadfurcation into a solid core fiber-optic probe and guided onto the tissue. A small fraction of the light is reflected from the tissue into the acceptance cone of the probe fiber and directed via a second fiber of the quadfurcation into a spectrometer [SD-2000; Ocean Optics; Duiven, NL] for detection and further analysis. After the reflectance measurement, a low power fluorescence measurement is performed, using a 405 nm laser directed into another fiber of the quadfurcation to illuminate the tissue through the same probe. The emitted fluorescence is then guided through the last fiber of the quadfurcation, which leads through a 450 nm long-pass filter into a second spectrograph [QE-65000; Ocean Optics; Duiven, NL]. The fluorescence measurement is incorporated to ensure the presence of tumor cells, lenti-virally infected with GFP, in the measurement volume. The complete MDSFR system consists of two identical probe-units containing fiber diameters of 0.4 and 0.8 mm. The fiber probes are sequentially placed in contact with the tissue under investigation. All probes are polished under an angle of 15° to minimize internal specular reflections from the probe tip. A calibration procedure was performed as described previously [7

7. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]

], consisting of a measurement in water in a dark container (Iwater) and a measurement in a liquid phantom containing 20% Intralipid diluted to yield μs′(800nm) = 1.2 mm−1, (IILcal), where
RSFmeas=IILsim[ImeasIwaterIILcalIwater]
(12)
and IILsim is the absolute SFR signal obtained through Monte Carlo simulations.

2.3 Animal model

The study protocol was approved by the Animal Welfare Committee of the Leiden University Medical Center. Housing of BALB/cByJ nu/nu female mice (aged 4-6 weeks; Charles River Laboratories), the experiments and euthanazation were performed in accordance with the guidelines of this committee. Chlorophyll free food and sterilized water were provided without restriction. A set of 19 mice were injected with OSC19 cells lenti-virally transfected with luc and GFP (6x104 cells, Biocat, Heidelberg, Germany) [15

15. T. Yokoi, A. Yamaguchi, T. Odajima, and K. Furukawa, “Establishment and characterization of a human cell line derived from a squamous cell carcinoma of the tongue,” Tumor Res. 23, 43–57 (1988).

,16

16. F. Carlotti, M. Bazuine, T. Kekarainen, J. Seppen, P. Pognonec, J. A. Maassen, and R. C. Hoeben, “Lentiviral vectors efficiently transduce quiescent mature 3T3-L1 adipocytes,” Mol. Ther. 9(2), 209–217 (2004). [CrossRef] [PubMed]

]. After an incubation period of 10-13 days, animals were included in the study and anesthetized with 2-3% Isoflurane. Reflectance spectra were acquired from tumor, normal tongue tissue (both 5 measurements), and skin and liver (both 3 measurements). The consecutive measurements were taken by removing and carefully repositioning the fiber between measurements, without imparting undue pressure on each tissue. For tumor, the fiber tip was positioned directly on top of the visible tumor. An additional MDSFR measurement of liver was carried out immediately before animals were sacrificed. After the animals were sacrificed, the tongue was harvested for microscopic tissue analysis.

2.4 Data reduction and statistical analysis

To verify the presence of tumor tissue in the measurement volume with the probe on top of the tumor, co-localized fluorescence spectra are acquired with both the 0.4 and 0.8 mm fiber. Since the quantum yield of CopGFP, incorporated to identify tumor tissue, is high, qualitative analysis of these spectra was sufficient to exclude measurements where tumor tissue was not present in the measurement volume. Two animals did not develop a visual tumor, and lacked GFP fluorescence in all five tumor measurements for both fiber diameters. These animals were excluded from the analysis of optical properties of tumor tissue.

Variation in probe pressure and probe location in consecutive measurements may lead to slight variations in outcome, which can interfere with a proper analysis of the averaged MDSFR data. Therefore, all SFR spectra with R(800 nm) deviating >2σ from the average value of R(800 nm) of the consecutive measurements on each tissue site, were removed from the analysis after the first analysis step. To account for remaining variations in subsequent SFR measurements, the standard deviation of R(800 nm) for each fiber diameter was used as a weight factor in the non-linear MDSFR fit procedure. Therefore, SFR fiber measurements with large variations contribute less to the best-fit outcome. Statistical analysis of optical properties for various tissues is done by using a paired Student T-test with P<0.05 as the level of significance.

3. Results

After an incubation period of 10-13 days, 17 mice had a visible tumor at the tip of the tongue. Figure 2
Fig. 2 Confocal fluorescence microscopy image of mouse tongue, using 488 nm excitation and 520-540 nm detection, showing the distribution of GFP-expressing OSC tumor in green and transmitted 488 nm light in grey. Scale bar 1 mm.
shows a representative example of a combined confocal fluorescence and white light transmission microscopy image of a frozen section cut vertically through the center of the tongue. In this example the fiber tip was positioned in contact with the surface of the tongue in the upper left region of the image where tumor cells are close to the surface of the tongue. The image shows a thin but variable layer of epithelium overlying the GFP-fluorescent tumor at the tip of the tongue.A total of 588 reflectance spectra taken from 19 mice were analyzed in this study. Figure 3
Fig. 3 Typical 0.8 mm SFR data. Plotted are the measured RSF (black dots), individual SFR fits (solid lines) and calculatedRSF0(dashed lines) for normal tongue, tumor, skin and liver tissue.
shows representative reflectance spectra of tumor, tongue, skin and liver tissue, together with their best fits and calculated absorption-corrected reflectance RSF0 for a 0.8 mm fiber (after the 3rd analysis step, based on power-law scattering and wavelength-dependent γ).

In this example, the spectra taken on skin show a significantly higher reflectance over the whole wavelength range compared to normal tongue and tumor tissue, while reflectance for liver is lowest. RSF0 spectra (dashed lines in Fig. 3) of five consecutive co-localized measurements per fiber diameter are averaged, and these averaged spectra and their standard deviations serve as input for the MDSFR analysis. Figure 4
Fig. 4 Averaged SFR spectra of normal tongue tissue; fiber diameters are 0.4 and 0.8 mm.
shows the average RSF0 spectra of normal tongue tissue in a single mouse, fitted with the MDSFR model. As expected, a larger fiber diameter results in a higher reflectance.

Figures 5(a)
Fig. 5 μs′ (a) and γ (b) for different tissues, measured in 1 representative mouse.
and 5(b) show μs′(λ) and γ (λ) resulting from MDSFR fits on different tissues for a representative mouse. For all locations, μs′ decreases with increasing wavelength [17

17. A. E. Cerussi, A. J. Berger, F. Bevilacqua, N. Shah, D. Jakubowski, J. Butler, R. F. Holcombe, and B. J. Tromberg, “Sources of absorption and scattering contrast for near-infrared optical mammography,” Acad. Radiol. 8(3), 211–218 (2001). [CrossRef] [PubMed]

]. It is helpful to describe μs′ in terms of two parameters, the reduced scattering coefficient at 800 nm μs′(800nm), and its wavelength dependent slope, which is often termed the scattering power (resp. a1 and a2 in Eq. (11)).

In Fig. 6
Fig. 6 Gamma per tissue, averaged over n = 19.
, we show the average γ(λ) for the four different tissue types, based on the measurements on all 19 mice.

Figures 7(a)
Fig. 7 Average scattering power (a), μs′ (800nm) (b) and Gamma (c) of 19 mice, for 4 different tissues.
, 7(b), and 7(c) show the median and quartiles of μs′(800), the scattering power, and γ averaged over the 400-850 nm wavelength range, for normal tongue, tumor tissue, skin and liver.Both Fig. 6 and Fig. 7(c) clearly show that γ is lowest for skin and highest for liver. Furthermore, γ for tumor increases monotonically with increasing wavelength, and increases more over the measured wavelength range than for normal tissue, although this difference is not significant. Table 1

Table 1. Scattering Power, μs′(800nm) and Average Gamma for Four Different Tissues

table-icon
View This Table
summarizes the average values for scattering power, μs′(800) and average gamma for the four tissue types.

From Fig. 7(a) and Table 1, it can be seen that the scattering powers of tumor and liver are significantly lower than for normal tongue and skin (Paired t-test, P<0.05). The μs′(800nm) for tumor is significantly higher than for normal tongue and liver. The average value for γ of skin differs significantly from all other tissues.

An interesting observation is that although the reflectance is highest for skin, the μs′(λ) is not. This suggests that the angular distribution of scattering represented by γ(λ) can strongly influence the observed reflectance. The low γ(λ) for skin indicates that scattering is more directed backwards for skin than for other tissues, resulting in a higher total reflectance. For liver the opposite occurs; the low reflectance in this tissue is not a result of a low μs′(λ), but the high γ(λ) which indicates that most of the scattered light is directed forward.

4. Discussion and conclusion

In this paper we have presented the first in vivo measurements of μs′(λ) and γ(λ) obtained using multi-diameter single fiber reflectance spectroscopy. We determined μs′(λ) and γ(λ) for 4 different types of mouse tissue, including skin, liver, normal tongue and oral squamous cell carcinoma. Measurements of μs′(λ) and γ(λ) show distinct differences between tissue types.

4.1 Comparison to previous studies: μs(λ)

4.2 Comparison to previous studies: γ(λ)

4.3 Step-analysis of fit procedure

MDSFR is based on the reflectance in the absence of absorption. In the vast majority of in vivo situations, where absorption is present, one first needs to determine the contribution of absorption to the total reflectance intensity using the model that we previously developed for individual SFR measurements. To achieve this, our analysis of tissue absorption and scattering in an individual SFR measurement was based on a background scattering model consisting of a polynomial (n = 4), together with an optimized set of wavelength-independent coefficients representing γ [10

10. S. C. Kanick, D. J. Robinson, H. J. C. M. Sterenborg, and A. Amelink, “Method to quantitate absorption coefficients from single fiber reflectance spectra without knowledge of the scattering properties,” Opt. Lett. 36(15), 2791–2793 (2011). [CrossRef] [PubMed]

]. The choice of such a non-physical background scattering model is necessary in order to correct for the ill-posed assumption that γ is wavelength independent; the background scattering model needs sufficient freedom in this first step to correct for this assumption. It is important to emphasize that in the first step, we only need to remove the effect of absorption on the individual SFR spectra by implementing the approach reported previously [11

11. S. C. Kanick, U. A. Gamm, H. J. C. M. Sterenborg, D. J. Robinson, and A. Amelink, “Method to quantitatively estimate wavelength-dependent scattering properties from multidiameter single fiber reflectance spectra measured in a turbid medium,” Opt. Lett. 36(15), 2997–2999 (2011). [CrossRef] [PubMed]

]. The MDSFR analysis itself is then performed with the physically realistic power-law scattering model (that follows from the definition of the Whittle-Matern correlation function), which yields values for the reduced scattering coefficient (defined by the scattering power and μs′(800nm)) and γ. To verify that the absorption was correctly estimated in the first step, the extracted values for γ from the MDSFR analysis were used to re-analyse the individual SFR measurements using the power-law scattering model and the extracted wavelength-dependent γ. We found that the absorption spectra extracted in the first step were almost identical to the absorption spectra extracted using the power-law scattering model and the wavelength dependent γ, as indicated by an average Pearson correlation coefficient of R2 = 0.99 between extracted absorption spectra.

Clearly when more information becomes available about the wavelength dependence of γ, that knowledge could be used to spectrally constrain γ and thus complete an MDSFR fit in a single step fitting routine.

4.4 Limitations

4.4.1 Probe replacement

A potential limitation of the present study is caused by the fact that multiple fibers with different diameters are used sequentially. Therefore, slight differences in measurement volume are not only due to different fiber diameters but can also occur from measuring a slightly different spot in each measurement. Sequentially measuring the exact same tissue location with the exact same pressure for various fiber diameters is not only time consuming, but also makes the current measurement procedure cumbersome and sensitive to measurement artifacts. Clearly the use of a single measurement probe that is capable of rapidly changing its effective aperture would be a major step forward. New technological developments should overcome this problem in the near future and provide true co-localized measurement positions for various fiber diameters [29

29. C. L. Hoy, U. A. Gamm, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Use of a coherent fiber bundle for multi-diameter single fiber reflectance spectroscopy,” Biomed. Opt. Express 3(10), 2452–2464 (2012). [CrossRef] [PubMed]

]. Despite the uncertainty in probe repositioning, the standard deviations for an individual mouse on the MDSFR fitted values for μs′ and γ (Figs. 5(a) and 5(b)) are all within 10% of the fitted value (with the exception of μs′ for skin), which is very encouraging.

4.4.2 Layered tissue

It is well known that skin and oral mucosa are comprised of layered tissue. In skin the stratum corneum and epidermis overlay the underlying dermis. In the oral cavity the normal tongue is comprised of a superficial keratinized layer with underlying epithelial layer and deeper lying connective tissue comprised of orthogonal muscle fiber bundles. The two, 0.4 and 0.8 mm, fibers are expected to probe the relatively homogeneous layers of tissue comprising mouse epithelium and dermis in the skin and the tongue epithelium and underlying connective tissue in normal tongue. OSC in the tongue develops beneath the superficial keratinized layer but can potentially be overlain by a layer of epithelial tissue of variable thickness. Microscope images of sections of OSC in the tongue (Fig. 2) show variation in thickness of the tissue layer covering the tumor between animals. Although tumor size and position varied somewhat between animals, placing the fiber probe on a position where the tumor was visible minimized the distance between tumor and fiber tip. Since the measurement volume and depth increase with increasing fiber diameter, the smallest fiber diameter was specifically chosen to be 0.4 mm to incorporate tumor within the measurement volume. However, the contribution of the top layer of the tongue epithelium to the total acquired signal will be slightly different for each fiber diameter. Given these concerns we analyzed spectra acquired using each fiber diameter at a single location. This analysis showed consistent values for scattering power between fiber diameters. While this result suggests that the influence of epithelium overlying the tumor is small, it could be a reason why we observe only a relatively small difference in scattering signature between normal tongue and tumor tissue. The potential confounding effect of overlying epithelium could be removed in the future by studying a chemically induced model of oral cancer in animals or interrogating superficial human OSC.

4.5 Future developments

Quantification of tissue optical properties does not only yield direct diagnostic information, it also facilitates the use of quantitative in vivo fluorescence measurements. To quantify the concentration of an (exogenous) fluorophore in tissue, for example to monitor photodynamic therapy, knowledge of local tissue optical properties is necessary to correct for the effects of scattering and absorption on the collected fluorescence [30

30. B. C. Wilson, M. S. Patterson, and L. Lilge, “Implicit and explicit dosimetry in photodynamic therapy: a New paradigm,” Lasers Med. Sci. 12(3), 182–199 (1997). [CrossRef] [PubMed]

]. Our setup allows for reflectance and fluorescence measurements to be taken with the same probe, providing a method to determine fluorescence corrected for the influence of tissue optical properties, which opens the way to quantitatively compare fluorescence from different tissues and patients.

In conclusion, we have presented the first in vivo MDSFR analysis of four different tissue types, based on measurements of 19 animals. Analysis of scattering power, reduced scattering at 800nm and average γ resulted in significant differences for the four tissue types studied. Future work will address the relation between μs′(λ) and γ(λ) and the properties of the tissue refractive index correlation function D and lc.

References and links

1.

R. Reif, O. A’Amar, and I. J. Bigio, “Analytical model of light reflectance for extraction of the optical properties in small volumes of turbid media,” Appl. Opt. 46(29), 7317–7328 (2007). [CrossRef] [PubMed]

2.

A. Amelink, H. J. C. M. Sterenborg, M. P. L. Bard, and S. A. Burgers, “In vivo measurement of the local optical properties of tissue by use of differential path-length spectroscopy,” Opt. Lett. 29(10), 1087–1089 (2004). [CrossRef] [PubMed]

3.

G. Zonios, L. T. Perelman, V. Backman, R. Manoharan, M. Fitzmaurice, J. Van Dam, and M. S. Feld, “Diffuse reflectance spectroscopy of human adenomatous colon polyps in vivo,” Appl. Opt. 38(31), 6628–6637 (1999). [CrossRef] [PubMed]

4.

H. Tian, Y. Liu, and L. Wang, “Influence of the third-order parameter on diffuse reflectance at small source-detector separations,” Opt. Lett. 31(7), 933–935 (2006). [CrossRef] [PubMed]

5.

F. Bevilacqua and C. Depeursinge, “Monte Carlo study of diffuse reflectance at source−detector separations close to one transport mean free path,” J. Opt. Soc. Am. A 16(12), 2935–2945 (1999). [CrossRef]

6.

S. C. Kanick, U. A. Gamm, M. Schouten, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Measurement of the reduced scattering coefficient of turbid media using single fiber reflectance spectroscopy: fiber diameter and phase function dependence,” Biomed. Opt. Express 2(6), 1687–1702 (2011). [CrossRef] [PubMed]

7.

U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]

8.

S. C. Kanick, D. J. Robinson, H. J. Sterenborg, and A. Amelink, “Monte Carlo analysis of single fiber reflectance spectroscopy: photon path length and sampling depth,” Phys. Med. Biol. 54(22), 6991–7008 (2009). [CrossRef] [PubMed]

9.

S. C. Kanick, H. J. Sterenborg, and A. Amelink, “Empirical model of the photon path length for a single fiber reflectance spectroscopy device,” Opt. Express 17(2), 860–871 (2009). [CrossRef] [PubMed]

10.

S. C. Kanick, D. J. Robinson, H. J. C. M. Sterenborg, and A. Amelink, “Method to quantitate absorption coefficients from single fiber reflectance spectra without knowledge of the scattering properties,” Opt. Lett. 36(15), 2791–2793 (2011). [CrossRef] [PubMed]

11.

S. C. Kanick, U. A. Gamm, H. J. C. M. Sterenborg, D. J. Robinson, and A. Amelink, “Method to quantitatively estimate wavelength-dependent scattering properties from multidiameter single fiber reflectance spectra measured in a turbid medium,” Opt. Lett. 36(15), 2997–2999 (2011). [CrossRef] [PubMed]

12.

U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Measurement of tissue scattering properties using multi-diameter single fiber reflectance spectroscopy: in silico sensitivity analysis,” Biomed. Opt. Express 2(11), 3150–3166 (2011). [CrossRef] [PubMed]

13.

R. L. P. van Veen, W. Verkruysse, and H. J. C. M. Sterenborg, “Diffuse-reflectance spectroscopy from 500 to 1060 nm by correction for inhomogeneously distributed absorbers,” Opt. Lett. 27(4), 246–248 (2002). [CrossRef] [PubMed]

14.

N. Rajaram, A. Gopal, X. Zhang, and J. W. Tunnell, “Experimental validation of the effects of microvasculature pigment packaging on in vivo diffuse reflectance spectroscopy,” Lasers Surg. Med. 42(7), 680–688 (2010). [CrossRef] [PubMed]

15.

T. Yokoi, A. Yamaguchi, T. Odajima, and K. Furukawa, “Establishment and characterization of a human cell line derived from a squamous cell carcinoma of the tongue,” Tumor Res. 23, 43–57 (1988).

16.

F. Carlotti, M. Bazuine, T. Kekarainen, J. Seppen, P. Pognonec, J. A. Maassen, and R. C. Hoeben, “Lentiviral vectors efficiently transduce quiescent mature 3T3-L1 adipocytes,” Mol. Ther. 9(2), 209–217 (2004). [CrossRef] [PubMed]

17.

A. E. Cerussi, A. J. Berger, F. Bevilacqua, N. Shah, D. Jakubowski, J. Butler, R. F. Holcombe, and B. J. Tromberg, “Sources of absorption and scattering contrast for near-infrared optical mammography,” Acad. Radiol. 8(3), 211–218 (2001). [CrossRef] [PubMed]

18.

V. Turzhitsky, N. N. Mutyal, A. J. Radosevich, and V. Backman, “Multiple scattering model for the penetration depth of low-coherence enhanced backscattering,” J. Biomed. Opt. 16(9), 097006 (2011). [CrossRef] [PubMed]

19.

V. Turzhitsky, A. Radosevich, J. D. Rogers, A. Taflove, and V. Backman, “A predictive model of backscattering at subdiffusion length scales,” Biomed. Opt. Express 1(3), 1034–1046 (2010). [CrossRef] [PubMed]

20.

J. D. Rogers, I. R. Capoğlu, and V. Backman, “Nonscalar elastic light scattering from continuous random media in the Born approximation,” Opt. Lett. 34(12), 1891–1893 (2009). [CrossRef] [PubMed]

21.

A. Kim and B. C. Wilson, Optical-Thermal Response of Laser-Irradiated Tissue (Springer, 2011), Chap. 8.

22.

P. Thueler, I. Charvet, F. Bevilacqua, M. St. Ghislain, G. Ory, P. Marquet, P. Meda, B. Vermeulen, and C. Depeursinge, “In vivo endoscopic tissue diagnostics based on spectroscopic absorption, scattering, and phase function properties,” J. Biomed. Opt. 8(3), 495–503 (2003). [CrossRef] [PubMed]

23.

J. Yi and V. Backman, “Imaging a full set of optical scattering properties of biological tissue by inverse spectroscopic optical coherence tomography,” Opt. Lett. 37(21), 4443–4445 (2012). [CrossRef] [PubMed]

24.

G. Hall, S. L. Jacques, K. W. Eliceiri, and P. J. Campagnola, “Goniometric measurements of thick tissue using Monte Carlo simulations to obtain the single scattering anisotropy coefficient,” Biomed. Opt. Express 3(11), 2707–2719 (2012). [CrossRef] [PubMed]

25.

M. Xu and R. R. Alfano, “Fractal mechanisms of light scattering in biological tissue and cells,” Opt. Lett. 30(22), 3051–3053 (2005). [CrossRef] [PubMed]

26.

A. J. Gomes, S. Ruderman, M. DelaCruz, R. K. Wali, H. K. Roy, and V. Backman, “In vivo measurement of the shape of the tissue-refractive-index correlation function and its application to detection of colorectal field carcinogenesis,” J. Biomed. Opt. 17(4), 047005 (2012). [CrossRef] [PubMed]

27.

F. Bevilacqua, D. Piguet, P. Marquet, J. D. Gross, B. J. Tromberg, and C. Depeursinge, “In vivo local determination of tissue optical properties: applications to human brain,” Appl. Opt. 38(22), 4939–4950 (1999). [CrossRef] [PubMed]

28.

S. Chamot, E. Migacheva, O. Seydoux, P. Marquet, and C. Depeursinge, “Physical interpretation of the phase function related parameter γ studied with a fractal distribution of spherical scatterers,” Opt. Express 18(23), 23664–23675 (2010). [CrossRef] [PubMed]

29.

C. L. Hoy, U. A. Gamm, H. J. Sterenborg, D. J. Robinson, and A. Amelink, “Use of a coherent fiber bundle for multi-diameter single fiber reflectance spectroscopy,” Biomed. Opt. Express 3(10), 2452–2464 (2012). [CrossRef] [PubMed]

30.

B. C. Wilson, M. S. Patterson, and L. Lilge, “Implicit and explicit dosimetry in photodynamic therapy: a New paradigm,” Lasers Med. Sci. 12(3), 182–199 (1997). [CrossRef] [PubMed]

OCIS Codes
(060.2310) Fiber optics and optical communications : Fiber optics
(170.6510) Medical optics and biotechnology : Spectroscopy, tissue diagnostics
(300.6550) Spectroscopy : Spectroscopy, visible
(170.6935) Medical optics and biotechnology : Tissue characterization

ToC Category:
Spectroscopic Diagnostics

History
Original Manuscript: December 19, 2012
Revised Manuscript: February 20, 2013
Manuscript Accepted: February 23, 2013
Published: April 9, 2013

Citation
F. van Leeuwen-van Zaane, U. A. Gamm, P. B. A. A. van Driel, T. J. A. Snoeks, H. S. de Bruijn, A. van der Ploeg-van den Heuvel, I. M. Mol, C. W. G. M. Löwik, H. J. C. M. Sterenborg, A. Amelink, and D. J. Robinson, "In vivo quantification of the scattering properties of tissue using multi-diameter single fiber reflectance spectroscopy," Biomed. Opt. Express 4, 696-708 (2013)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-4-5-696


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. R. Reif, O. A’Amar, I. J. Bigio, “Analytical model of light reflectance for extraction of the optical properties in small volumes of turbid media,” Appl. Opt. 46(29), 7317–7328 (2007). [CrossRef] [PubMed]
  2. A. Amelink, H. J. C. M. Sterenborg, M. P. L. Bard, S. A. Burgers, “In vivo measurement of the local optical properties of tissue by use of differential path-length spectroscopy,” Opt. Lett. 29(10), 1087–1089 (2004). [CrossRef] [PubMed]
  3. G. Zonios, L. T. Perelman, V. Backman, R. Manoharan, M. Fitzmaurice, J. Van Dam, M. S. Feld, “Diffuse reflectance spectroscopy of human adenomatous colon polyps in vivo,” Appl. Opt. 38(31), 6628–6637 (1999). [CrossRef] [PubMed]
  4. H. Tian, Y. Liu, L. Wang, “Influence of the third-order parameter on diffuse reflectance at small source-detector separations,” Opt. Lett. 31(7), 933–935 (2006). [CrossRef] [PubMed]
  5. F. Bevilacqua, C. Depeursinge, “Monte Carlo study of diffuse reflectance at source−detector separations close to one transport mean free path,” J. Opt. Soc. Am. A 16(12), 2935–2945 (1999). [CrossRef]
  6. S. C. Kanick, U. A. Gamm, M. Schouten, H. J. Sterenborg, D. J. Robinson, A. Amelink, “Measurement of the reduced scattering coefficient of turbid media using single fiber reflectance spectroscopy: fiber diameter and phase function dependence,” Biomed. Opt. Express 2(6), 1687–1702 (2011). [CrossRef] [PubMed]
  7. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, A. Amelink, “Quantification of the reduced scattering coefficient and phase-function-dependent parameter γ of turbid media using multidiameter single fiber reflectance spectroscopy: experimental validation,” Opt. Lett. 37(11), 1838–1840 (2012). [CrossRef] [PubMed]
  8. S. C. Kanick, D. J. Robinson, H. J. Sterenborg, A. Amelink, “Monte Carlo analysis of single fiber reflectance spectroscopy: photon path length and sampling depth,” Phys. Med. Biol. 54(22), 6991–7008 (2009). [CrossRef] [PubMed]
  9. S. C. Kanick, H. J. Sterenborg, A. Amelink, “Empirical model of the photon path length for a single fiber reflectance spectroscopy device,” Opt. Express 17(2), 860–871 (2009). [CrossRef] [PubMed]
  10. S. C. Kanick, D. J. Robinson, H. J. C. M. Sterenborg, A. Amelink, “Method to quantitate absorption coefficients from single fiber reflectance spectra without knowledge of the scattering properties,” Opt. Lett. 36(15), 2791–2793 (2011). [CrossRef] [PubMed]
  11. S. C. Kanick, U. A. Gamm, H. J. C. M. Sterenborg, D. J. Robinson, A. Amelink, “Method to quantitatively estimate wavelength-dependent scattering properties from multidiameter single fiber reflectance spectra measured in a turbid medium,” Opt. Lett. 36(15), 2997–2999 (2011). [CrossRef] [PubMed]
  12. U. A. Gamm, S. C. Kanick, H. J. Sterenborg, D. J. Robinson, A. Amelink, “Measurement of tissue scattering properties using multi-diameter single fiber reflectance spectroscopy: in silico sensitivity analysis,” Biomed. Opt. Express 2(11), 3150–3166 (2011). [CrossRef] [PubMed]
  13. R. L. P. van Veen, W. Verkruysse, H. J. C. M. Sterenborg, “Diffuse-reflectance spectroscopy from 500 to 1060 nm by correction for inhomogeneously distributed absorbers,” Opt. Lett. 27(4), 246–248 (2002). [CrossRef] [PubMed]
  14. N. Rajaram, A. Gopal, X. Zhang, J. W. Tunnell, “Experimental validation of the effects of microvasculature pigment packaging on in vivo diffuse reflectance spectroscopy,” Lasers Surg. Med. 42(7), 680–688 (2010). [CrossRef] [PubMed]
  15. T. Yokoi, A. Yamaguchi, T. Odajima, K. Furukawa, “Establishment and characterization of a human cell line derived from a squamous cell carcinoma of the tongue,” Tumor Res. 23, 43–57 (1988).
  16. F. Carlotti, M. Bazuine, T. Kekarainen, J. Seppen, P. Pognonec, J. A. Maassen, R. C. Hoeben, “Lentiviral vectors efficiently transduce quiescent mature 3T3-L1 adipocytes,” Mol. Ther. 9(2), 209–217 (2004). [CrossRef] [PubMed]
  17. A. E. Cerussi, A. J. Berger, F. Bevilacqua, N. Shah, D. Jakubowski, J. Butler, R. F. Holcombe, B. J. Tromberg, “Sources of absorption and scattering contrast for near-infrared optical mammography,” Acad. Radiol. 8(3), 211–218 (2001). [CrossRef] [PubMed]
  18. V. Turzhitsky, N. N. Mutyal, A. J. Radosevich, V. Backman, “Multiple scattering model for the penetration depth of low-coherence enhanced backscattering,” J. Biomed. Opt. 16(9), 097006 (2011). [CrossRef] [PubMed]
  19. V. Turzhitsky, A. Radosevich, J. D. Rogers, A. Taflove, V. Backman, “A predictive model of backscattering at subdiffusion length scales,” Biomed. Opt. Express 1(3), 1034–1046 (2010). [CrossRef] [PubMed]
  20. J. D. Rogers, I. R. Capoğlu, V. Backman, “Nonscalar elastic light scattering from continuous random media in the Born approximation,” Opt. Lett. 34(12), 1891–1893 (2009). [CrossRef] [PubMed]
  21. A. Kim and B. C. Wilson, Optical-Thermal Response of Laser-Irradiated Tissue (Springer, 2011), Chap. 8.
  22. P. Thueler, I. Charvet, F. Bevilacqua, M. St. Ghislain, G. Ory, P. Marquet, P. Meda, B. Vermeulen, C. Depeursinge, “In vivo endoscopic tissue diagnostics based on spectroscopic absorption, scattering, and phase function properties,” J. Biomed. Opt. 8(3), 495–503 (2003). [CrossRef] [PubMed]
  23. J. Yi, V. Backman, “Imaging a full set of optical scattering properties of biological tissue by inverse spectroscopic optical coherence tomography,” Opt. Lett. 37(21), 4443–4445 (2012). [CrossRef] [PubMed]
  24. G. Hall, S. L. Jacques, K. W. Eliceiri, P. J. Campagnola, “Goniometric measurements of thick tissue using Monte Carlo simulations to obtain the single scattering anisotropy coefficient,” Biomed. Opt. Express 3(11), 2707–2719 (2012). [CrossRef] [PubMed]
  25. M. Xu, R. R. Alfano, “Fractal mechanisms of light scattering in biological tissue and cells,” Opt. Lett. 30(22), 3051–3053 (2005). [CrossRef] [PubMed]
  26. A. J. Gomes, S. Ruderman, M. DelaCruz, R. K. Wali, H. K. Roy, V. Backman, “In vivo measurement of the shape of the tissue-refractive-index correlation function and its application to detection of colorectal field carcinogenesis,” J. Biomed. Opt. 17(4), 047005 (2012). [CrossRef] [PubMed]
  27. F. Bevilacqua, D. Piguet, P. Marquet, J. D. Gross, B. J. Tromberg, C. Depeursinge, “In vivo local determination of tissue optical properties: applications to human brain,” Appl. Opt. 38(22), 4939–4950 (1999). [CrossRef] [PubMed]
  28. S. Chamot, E. Migacheva, O. Seydoux, P. Marquet, C. Depeursinge, “Physical interpretation of the phase function related parameter γ studied with a fractal distribution of spherical scatterers,” Opt. Express 18(23), 23664–23675 (2010). [CrossRef] [PubMed]
  29. C. L. Hoy, U. A. Gamm, H. J. Sterenborg, D. J. Robinson, A. Amelink, “Use of a coherent fiber bundle for multi-diameter single fiber reflectance spectroscopy,” Biomed. Opt. Express 3(10), 2452–2464 (2012). [CrossRef] [PubMed]
  30. B. C. Wilson, M. S. Patterson, L. Lilge, “Implicit and explicit dosimetry in photodynamic therapy: a New paradigm,” Lasers Med. Sci. 12(3), 182–199 (1997). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.