Fiber-based polarization-sensitive OCT of the human retina with correction of system polarization distortions

Boy Braaf, Koenraad A. Vermeer, Mattijs de Groot, Kari V. Vienola, and Johannes F. de Boer  »View Author Affiliations


Biomedical Optics Express, Vol. 5, Issue 8, pp. 2736-2758 (2014)
http://dx.doi.org/10.1364/BOE.5.002736


View Full Text Article

Acrobat PDF (6346 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

In polarization-sensitive optical coherence tomography (PS-OCT) the use of single-mode fibers causes unpredictable polarization distortions which can result in increased noise levels and erroneous changes in calculated polarization parameters. In the current paper this problem is addressed by a new Jones matrix analysis method that measures and corrects system polarization distortions as a function of wavenumber by spectral analysis of the sample surface polarization state and deeper located birefringent tissue structures. This method was implemented on a passive-component depth-multiplexed swept-source PS-OCT system at 1040 nm which was theoretically modeled using Jones matrix calculus. High-resolution B-scan images are presented of the double-pass phase retardation, diattenuation, and relative optic axis orientation to show the benefits of the new analysis method for in vivo imaging of the human retina. The correction of system polarization distortions yielded reduced phase retardation noise, and better estimates of the diattenuation and the relative optic axis orientation in weakly birefringent tissues. The clinical potential of the system is shown by en face visualization of the phase retardation and optic axis orientation of the retinal nerve fiber layer in a healthy volunteer and a glaucoma patient with nerve fiber loss.

© 2014 Optical Society of America

1. Introduction

In the past two decades optical coherence tomography (OCT) has become a standard clinical tool in ophthalmology for non-invasive cross-sectional imaging of retinal and choroidal structures [1

1. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]

,2

2. W. Drexler and J. G. Fujimoto, “State-of-the-art retinal optical coherence tomography,” Prog. Retin. Eye Res. 27(1), 45–88 (2008). [CrossRef] [PubMed]

]. OCT is based on optical low-coherence interferometry and provides high axial (2-15 µm) and lateral (5-30 µm) resolutions which makes it ideal for in vivo visualization of single tissue layer morphology and the diagnosis of pathology on a microscopic level [2

2. W. Drexler and J. G. Fujimoto, “State-of-the-art retinal optical coherence tomography,” Prog. Retin. Eye Res. 27(1), 45–88 (2008). [CrossRef] [PubMed]

].

In order to facilitate optical alignment and allow the use of fiber-based sample interfaces (e.g. endoscopes) PS-OCT systems are often built with single-mode optical fibers (SMFs) [21

21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

]. In SMFs stress-induced birefringence and imperfections in the fiber core symmetry pose a problem by changing the polarization state of the OCT light to an unknown state, which can become aligned in parallel or orthogonal to the sample optical axis [21

21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

,22

22. B. H. Park, C. Saxer, S. M. Srinivas, J. S. Nelson, and J. F. de Boer, “In vivo burn depth determination by high-speed fiber-based polarization sensitive optical coherence tomography,” J. Biomed. Opt. 6(4), 474–479 (2001). [CrossRef] [PubMed]

]. Polarization modulation and multiplexing methods have therefore been developed for fiber-based PS-OCT that probe the sample with several distinct (orthogonal) polarization states to solve this problem [21

21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

28

28. B. Baumann, W. Choi, B. Potsaid, D. Huang, J. S. Duker, and J. G. Fujimoto, “Swept source/Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unit,” Opt. Express 20(9), 10229–10241 (2012). [CrossRef] [PubMed]

]. It was recognized early on that these methods only could achieve optimal results when the changes in the polarization states are wavelength independent, i.e., if polarization mode dispersion (PMD) in the PS-OCT system is negligible [21

21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

]. In SMFs PMD causes a differential group delay between the two principal polarization states, which results in axial displacements between the detected polarization states in PS-OCT [21

21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

,29

29. B. Elmaanaoui, B. Wang, J. C. Dwelle, A. B. McElroy, S. S. Liu, H. G. Rylander, and T. E. Milner, “Birefringence measurement of the retinal nerve fiber layer by swept source polarization sensitive optical coherence tomography,” Opt. Express 19(11), 10252–10268 (2011). [CrossRef] [PubMed]

,30

30. M. Villiger, E. Z. Zhang, S. Nadkarni, W. Y. Oh, B. E. Bouma, and B. J. Vakoc, “Artifacts in polarization-sensitive optical coherence tomography caused by polarization mode dispersion,” Opt. Lett. 38(6), 923–925 (2013). [CrossRef] [PubMed]

]. It was shown recently by Villiger et al. [30

30. M. Villiger, E. Z. Zhang, S. Nadkarni, W. Y. Oh, B. E. Bouma, and B. J. Vakoc, “Artifacts in polarization-sensitive optical coherence tomography caused by polarization mode dispersion,” Opt. Lett. 38(6), 923–925 (2013). [CrossRef] [PubMed]

] that PMD can add significant amounts of noise and even cause erroneous local retardation changes in PS-OCT measurements. It is therefore evident that PMD and other wavelength-dependent effects should be minimized or corrected. The amount by which the SMF length can be reduced in fiber-based PS-OCT systems is limited and therefore the development of numerical solutions for PMD correction is required. In PS-OCT microscopy on muscle/tendon tissue Zhang et al. [31

31. E. Z. Zhang, W. Y. Oh, M. L. Villiger, L. Chen, B. E. Bouma, and B. J. Vakoc, “Numerical compensation of system polarization mode dispersion in polarization-sensitive optical coherence tomography,” Opt. Express 21(1), 1163–1180 (2013). [CrossRef] [PubMed]

] showed that a set of three sample arm calibration signals can be used to measure and numerically compensate PMD effects. In a separate study, Villiger et al. [32

32. M. Villiger, E. Z. Zhang, S. K. Nadkarni, W. Y. Oh, B. J. Vakoc, and B. E. Bouma, “Spectral binning for mitigation of polarization mode dispersion artifacts in catheter-based optical frequency domain imaging,” Opt. Express 21(14), 16353–16369 (2013). [CrossRef] [PubMed]

] presented a robust Stokes-analysis method that worked directly on endoscopic PS-OCT sample data, and mitigated PMD effects at the expense of a lower axial resolution using a spectral binning method.

In this paper we present a method to correct wavelength-dependent polarization distortions introduced by the setup in ophthalmic PS-OCT measurements of the retina. A passive-component depth-multiplexed swept-source PS-OCT setup at 1-µm was developed and is presented together with its theoretical Jones matrix modeling. The correction method extends the conventional PS-OCT Jones matrix analysis method of Park et al. [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

] with a spectral binning method to measure the polarization states of the retinal surface and deeper located birefringent tissue structures as a function of wavelength. Together with the theoretical model of the setup the analysis of these two tissue structures allowed the determination and correction of the system polarization distortions. This included the simultaneous correction of PMD effects and wavelength-dependent power differences between the polarization states in the interferometer reference and sample arms. High-resolution images are presented of the polarization properties in the macula and optic nerve head areas of a healthy subject and a glaucoma patient.

2. Experimental system description and Jones matrix modeling

The developed PS-OCT setup was based on the passive-component depth-multiplexed PS-OCT methods as published by Lim et al. [27

27. Y. Lim, Y. J. Hong, L. Duan, M. Yamanari, and Y. Yasuno, “Passive component based multifunctional Jones matrix swept source optical coherence tomography for Doppler and polarization imaging,” Opt. Lett. 37(11), 1958–1960 (2012). [CrossRef] [PubMed]

] and Baumann et al. [28

28. B. Baumann, W. Choi, B. Potsaid, D. Huang, J. S. Duker, and J. G. Fujimoto, “Swept source/Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unit,” Opt. Express 20(9), 10229–10241 (2012). [CrossRef] [PubMed]

]. The optical layout of the setup and an example of the acquired OCT images are shown in Fig. 1
Fig. 1 The optical layout of the PS-OCT setup showing the interferometer (left panel), the ophthalmic interface (upper right panel), and an example of the acquired images (lower right panel). In the sample arm the input light was converted in two depth-multiplexed orthogonal polarization states by a polarization delay unit (PDU) which extended the optical path of the horizontal linearly polarized state. Both states were launched simultaneously into the eye via an ophthalmic interface to record two depth-multiplexed OCT images with a dual-channel polarization-sensitive detection unit (H-det. & V-det.). In the reference arm 10% of the light was fed to a fiber-based Mach-Zehnder interferometer (K-clock) to create a stable interference signal that was used to clock the data-acquisition hardware linearly in wavenumbers. Component abbreviations, PBS: polarizing beam splitter, P: polarizer, PC: polarization controller, C: calibration mirror, GS: galvanometer scanners, DM: dichroic mirror, GP: glass plate, FT: fixation target, CAM: pupil camera.
, and are respectively discussed in sections 2.1 and 2.2. This is followed by the Jones matrix modeling of the PS-OCT setup in section 2.3.

2.1 Interferometer and ophthalmic interface hardware

The setup used a 1-µm wavelength swept-source (Axsun Technologies, Inc., MA, USA) with a fixed repetition rate of 100 kHz and an average output power of 23 mW. Light from the swept-source was coupled into a single-mode fiber-based interferometer (left panel Fig. 1) where it was split into a sample and a reference arm in a 70/30 ratio by a fiber coupler.

In the sample arm the light was coupled into a polarization-delay unit (PDU, marked by a dashed box in Fig. 1) where two orthogonal polarization states were created and sent simultaneously to the eye. This was achieved by splitting the incoming light in its two polarization components with a polarizing beam splitter (PBS), after which the two components each travelled a different path length through air before they were recombined by another PBS. This created an optical path delay for the horizontal linear polarized state with respect to the vertical linear polarized state and consequently multiplexed them in depth in the OCT image. Each state was additionally purified by a polarizer (P) to ensure orthogonality of the polarization states, and to suppress inter-state interference as described by Ju et al. [34

34. M. J. Ju, Y. J. Hong, S. Makita, Y. Lim, K. Kurokawa, L. Duan, M. Miura, S. Tang, and Y. Yasuno, “Advanced multi-contrast Jones matrix optical coherence tomography for Doppler and polarization sensitive imaging,” Opt. Express 21(16), 19412–19436 (2013). [CrossRef] [PubMed]

]. The optical power was set equal for both states using a polarization controller (PC) on the input fiber of the PDU. The light launched from the PDU was divided by a 20/80 coupler in two arms going respectively to the ophthalmic interface for imaging and a calibration mirror (C) for the generation of a reference signal. Returning light from both arms was analyzed in a polarization-sensitive detection unit (described below).

In the reference arm 10% of the light was split off by a fiber coupler to feed a fiber-based Mach-Zehnder Interferometer (MZI). The MZI used two 50/50 couplers with unbalanced air-paths and a balanced detector (1837 GHz-Nirvana, New Focus, CA, USA) to generate a stable interference signal that was used for k-clocking of the data acquisition. The remaining 90% of the reference arm light was transmitted through open air to match its path length and chromatic dispersion to the sample arm.

The polarization-sensitive detection unit consisted of a 50/50 fiber coupler where the light from the sample and reference arms was recombined, and PBS cubes to split the light from both output leads into its orthogonal polarization components. The horizontal and vertical polarization components were recorded by separate balanced detectors (PDB430C, Thorlabs Inc., NJ, USA) denoted as H-det. and V-det. in Fig. 1 respectively. Polarization controllers were installed to align the polarization states of the coupler's output leads, and to distribute the reference arm power equally over both detectors. This alignment was optimized to yield equal optical power and spectral shape for all detector photo-diodes as measured by their fast monitor outputs.

The ophthalmic interface (upper right panel Fig. 1) directed the OCT light (red paths) via a set of galvanometer scanners (GS) and a telescope to the human eye to image the posterior segment. The bundle incident on the cornea had an optical power of 1.5 mW and a diameter (1/e2) of 1.1 mm to provide a theoretical diffraction-limited spot-size (1/e2) of 20 µm on the retina. In order to improve the subject's fixation stability and facilitate instrument alignment a fixation target LED grid (FT) and a pupil camera (CAM) (blue paths) were coupled together via a glass plate (GP) and combined with the OCT light using a dichroic mirror (DM). The DM was placed under 45° in between the scan lens and the ophthalmic lens to ensure a constant incidence angle on the DM during scanning of the OCT beam with negligible (angle dependent) polarization effects.

All single-mode fiber components were used at standard lengths (1 - 2.5 m lead length) and together had a physical length of ~14 m per interferometer arm.

2.2 Data-acquisition and signal pre-processing

After detection, the signals from H-det. and V-det. were low-pass filtered (SBLP-200+ , 200 MHz −3 dB cutoff frequency, Mini-circuits, NY, USA) to reduce high-frequency noise and recorded by a 14-bit resolution dual-channel data-acquisition (DAQ) board (X5-400M, Innovative Integration, CA, USA). The k-clock signal from the MZI was high-pass filtered (SHP-20 + , 15.5 MHz −3 dB cutoff frequency, Mini-circuits) to remove low-frequency signal offsets, and provided the DAQ sampling clock. The MZI air-gap mismatch was set to 8.4 mm to create a k-clock that supported an OCT imaging depth range of 4.2 mm. The sweep-trigger of the swept-source initiated the digitization of A-scans linearly in wavenumbers with 1536 samples for an optical bandwidth of 92 nm centered at 1040 nm. A system sensitivity of 92.2 dB was measured from a mirror near the zero-delay depth as the incoherent intensity sum of the four signals. The signal roll-off with depth was −6 dB over 4.5 mm, and the axial resolution was 8.2 µm in air.

In the lower right panel of Fig. 1 an example is given of the acquired B-scan images for both detectors. By extending the optical path of the horizontal linearly polarized state in the PDU with half the OCT image depth range (2.1 mm), two depth-multiplexed images were recorded per detection channel. In what follows, the corresponding OCT signals are denoted as H1 and H2 for H-det., and V1 and V2 for V-det. for the horizontal and vertical states respectively. The calibration mirror created two sets of reference signals (C) that were used to phase-stabilize successive A-scans by cross-correlation. This procedure was implemented as described earlier by our group [35

35. B. Braaf, K. A. Vermeer, V. A. Sicam, E. van Zeeburg, J. C. van Meurs, and J. F. de Boer, “Phase-stabilized optical frequency domain imaging at 1-µm for the measurement of blood flow in the human choroid,” Opt. Express 19(21), 20886–20903 (2011). [CrossRef] [PubMed]

] without the need for numerical k-space resampling in this case. Afterwards fixed-pattern noise was subtracted from the data [36

36. N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12(3), 367–376 (2004). [CrossRef] [PubMed]

] and optical chromatic dispersion was corrected digitally [37

37. N. V. Iftimia, D. X. Hammer, C. E. Bigelow, D. I. Rosen, T. Ustun, A. A. Ferrante, D. Vu, and R. D. Ferguson, “Toward noninvasive measurement of blood hematocrit using spectral domain low coherence interferometry and retinal tracking,” Opt. Express 14(8), 3377–3388 (2006). [CrossRef] [PubMed]

] before polarization-sensitive analysis was performed as described in sections 3 and 4.

2.3 Jones matrix modeling of the passive-component depth-multiplexed PS-OCT setup

3. Jones matrix analysis to correct system polarization distortions

The principles for tissue polarization analysis using Jones matrix modeling in fiber-optic PS-OCT systems were described by Park et al. [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

]. The core element of this analysis is the Jones matrix modeling of the polarization state at the sample surface, which acts as a reference state to assess polarization changes at deeper locations. In this section the analysis of Park et al. [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

] was used as the basis for a more elaborate analysis in which also wavelength- and depth-dependent setup distortions are addressed.

3.1 Signal roll-off correction and depth-delay demodulation

The Jones matrix analysis started with the selection of a single B-scan from an in vivo retinal data set for which the data was pre-processed as described in section 2.2. Subsequently the complex electric field information was obtained via the digital Hilbert transform step of Eq. (2) with the addition of a division by the amplitude roll-off curve in the depth domain to correct for βz) and βz + γ) in Eq. (8). The amplitude roll-off curve was obtained in a separate experiment by measuring the signal amplitude on a mirror that was displaced over the full OCT imaging depth. The intensity roll-off difference between the two depth-multiplexed states was maximally −3.6 dB, which is equivalent to an amplitude factor of 0.66.

In addition, the depth-displacement γ between the two launched polarization states was measured from the calibration mirror signals by determining the first order phase difference (slope) as a function of wavenumber. The depth-delayed state was demodulated (frequency-shifted) on top of the undelayed state by applying an opposing first order phase-ramp. This typically corrected γ up to a quarter of a pixel, since PMD elsewhere in the setup caused slight under- or over-correction.

Rewriting Eq. (8) for these corrections, the sample data is described by Esample(k):
Esample(k)=[RH(k)00RV(k)]Jout(k)Js(k)Jin(k)[α(k)ei(kγ'+ρ(k))001]ein(k),
(9)
where γ' is the residual γ after correction.

3.2 K-dependent sample surface state determination

The PS-OCT data was first Fourier transformed without spectral binning and the location of the retinal surface pixels were determined by image segmentation. The PS-OCT data was then separately processed with spectral binning for 15 overlapping spectral bands of 384 samples (~23 nm) in width that were equally spaced over the OCT spectrum with interval steps of 82 samples (~5 nm). For every spectral bin the data was Fourier transformed and the pixels at the retinal surface were extracted. The surface state was then created by complex averaging over all surface pixels for each of the four electric field components (EH1, EH2, EV1, and EV2). In the averaging process first every pixel was normalized to (divided by) EH1 to account for the amplitude and phase offset of each pixel (see also section 4.4) [46

46. Y. Lim, M. Yamanari, S. Fukuda, Y. Kaji, T. Kiuchi, M. Miura, T. Oshika, and Y. Yasuno, “Birefringence measurement of cornea and anterior segment by office-based polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 2(8), 2392–2402 (2011). [CrossRef] [PubMed]

]. In Fig. 2(B) the surface state components EH1, EH2, EV1, and EV2 are plotted respectively for their real and imaginary parts at the sample number of each spectral bin center. It can be clearly seen that EH2, EV1, and EV2 varied gradually but with significant amplitude across the spectrum while EH1 was constant as a consequence of the normalization step before the complex averaging. The smooth variation of the electric field components enabled the use of spline extra- and interpolation to extend the surface state to all wavenumbers of the OCT spectrum (solid lines in Fig. 2(B)).

Since the sample surface pixels were not (or minimally) affected by the polarization properties of the retinal tissues, the surface state Esurf(k) was modeled according to Eq. (8) with exclusion of Js(k) and only contained terms describing the setup components and the cornea:
Esurf(k)=[RH(k)00RV(k)]Jout(k)Jin(k)[α(k)ei(kγ'+ρ(k))001].
(10)
Equations (9) and (10) were used to calculate the similarity transformation of Js(k):
Esurf1(k)Esample(k)=[α(k)ei(kγ'+ρ(k))001]1Jin1(k)Js(k)Jin(k)[α(k)ei(kγ'+ρ(k))001]ein(k).
(11)
In Eq. (11) Jout(k) and the reference arm diagonal matrix are cancelled out, which removed the optical path sections with the longest SMF lengths (together ~21 m) and therefore eliminated the largest contributors of PMD. The preservation of Jin(k) was further preferred over more common methods that preserve Jout(k) [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

], since Jin(k) is non-diattenuating and can be described as a unitary matrix of the SU(2) group. This was not the case in the current setup for Jout(k) and the reference arm diagonal matrix, since polarization dependent losses within the bulk optics of the polarization-sensitive detection circuit caused deviations from SU(2) [47

47. G. D. VanWiggeren and R. Roy, “Transmission of linearly polarized light through a single-mode fiber with random fluctuations of birefringence,” Appl. Opt. 38(18), 3888–3892 (1999). [CrossRef] [PubMed]

] and complicated their exact modeling.

3.3 Eigenvector analysis of deeper located birefringent tissue structures

In this section the remaining system polarization distortions are determined by simultaneously characterizing the unknown setup parameters Jin(k) and α(k)·ei(kγ'+ρ(k)). This is only possible for Eq. (11) when Js(k) has distinct eigenvalues and eigenvectors (i.e. Js(k) is not defective) which prevents Eq. (11) from cancelling itself to an identity matrix. In practice this condition is satisfied when the measurement is subject to birefringence or diattenuation. The analysis described in this section was therefore performed only on selected tissue pixels with significant phase retardation (>0.1 rad) and signal-to-noise ratio (SNR >10 dB). A large quantity of pixels in a single retinal B-scan is normally suited, since the birefringent RNFL and Henle's fiber layer impose phase retardation on all deeper tissue layers. These pixels were obtained and analyzed for different spectral bands by applying the spectral binning procedure of section 3.2 to Eq. (11).

The eigendecomposition of Eq. (11) gives the left eigenvectors:
ν(k)=[α(k)ei(kγ'+ρ(k))001]1Jin1(k)Rs1[s100s2],
(12)
where Rs represents a unitary change of basis for Js(k), and s1 and s2 are two complex scalars resulting from the eigendecomposition algorithm (LAPACK, eig() function, Matlab 2010b, The Mathworks, Inc.) to normalize the length of the eigenvectors to 1. Rs is modeled as a standard rotation matrix and describes the tissue optic axis as shown later in section 4.1. In order to use ν(k) for the correction of the remaining system polarization distortions, s1 and s2 have to be removed since they are artificial and do not represent any physical parts of the system. This is achieved by decomposing ν(k) in its individual terms and reconstructing the most common realization of ν(k) without s1 and s2.

Jin(k) describes the sample arm optical path from the exit of the PDU through the setup and the cornea, and can be separated in two Jones matrices: Jin(k) = Jc(k)·Jsys(k). In this decomposition Jc(k) describes the local corneal polarization properties and Jsys(k) describes the setup components between the PDU and the eye. The matrices Jc(k), Jsys(k) and Rs are members of the SU(2) group and for simplicity Jin−1(k)·Rs−1 = Jsys−1(k)·Jc−1(k)·Rs−1 can therefore be replaced in Eq. (12) by a wavelength-dependent SU(2) matrix in the form [x(k), -y*(k); y(k), x*(k)] with |x(k)|2 + |y(k)|2 = 1. Rewriting Eq. (12) according to these steps gives:
ν(k)=[ν11ν12ν21ν22]=[α(k)ei(kγ'+ρ(k))001]1Jsys1(k)Jc1(k)Rs1[s100s2]=[α(k)ei(kγ'+ρ(k))001]1[x(k)y(k)*y(k)x(k)*][s100s2]=[s1x(k)/(α(k)ei(kγ'+ρ(k)))s2y(k)*/(α(k)ei(kγ'+ρ(k)))s1y(k)s2x(k)*].
(13)
Subsequently Eq. (13) was used to determine α(k) and the amplitudes of s1, s2, x(k) and y(k):
α(k)=(|ν21||ν22|)/(|ν11||ν12|),
(14)
|s1|=|ν11α(k)|2+|ν21|2,|s2|=|ν12α(k)|2+|ν22|2,
(15)
|x(k)|=|ν11|α(k)/|s1|=|ν22|/|s2|,|y(k)|=|ν12|α(k)/|s2|=|ν21|/|s1|.
(16)
The phase of the individual variables of Eq. (13) could not be obtained separately from s1 and s2. However the phase φν(k) of the product of x(k), y(k)* and e-i(kγ'+ρ(k)) could be determined from each column vector of ν(k):

φν(k)=arg(x(k)y(k)*ei(kγ'+ρ(k)))=arg(ν11ν21*)=arg(ν12ν22*).
(17)

The obtained values for α(k), |x(k)|, |y(k)|, and φν(k) were used to construct a matrix Q(k) that represents an estimate of the most common realization of ν(k), and could be rewritten according to the matrix terms of Eq. (13) without s1 and s2:
Q(k)=[(|x(k)|/α(k))eiφν2(|y(k)|/α(k))eiφν2|y(k)|eiφν2|x(k)|eiφν2]=[x(k)¯/(α(k)ei(kγ'+ρ(k)))y(k)*¯/(α(k)ei(kγ'+ρ(k)))y(k)¯x(k)*¯][eiφxy(k)200eiφxy(k)2]ei(kγ'+ρ(k))2=[α(k)ei(kγ'+ρ(k))001]1Jsys1(k)Jc1(k)¯Rs1¯[eiφxy(k)200eiφxy(k)2]ei(kγ'+ρ(k))2,
(18)
where φxy(k) = arg(x(k)·y(k)) and the overbars denote the most common realization of a variable or matrix. In general the polarization properties of the setup components were constant for all A-scans in a single data set and therefore the product between Q−1(k) and ν(k) cancelled them out:
κ(k)=[κ11κ12κ21κ22]=Q(k)1ν(k)=[eiφxy(k)200eiφxy(k)2]Rs¯Jc(k)¯Jc1(k)Rs1[s100s2]ei(kγ'+ρ(k))2.
(19)
Equation (19) is close to an identity matrix, but the variation of Jc−1 and Rs−1 around their most common realizations caused considerable amplitude in the off-diagonal elements for the majority of the realizations of κ(k), and still allowed the determination of φxy(k) from each column vector of κ(k) via:
φxy(k)=arg(κ11κ21*)=arg(κ22*κ12).
(20)
In Fig. 3(A) the lower right histogram shows the distribution of φxy(k) for the center spectral bin, which had a clear distinct peak although wider and lower in peak-height than was observed for φν(k). The lower graph in Fig. 3(B) shows the variation across wavenumber of φxy(k) in red, which was small since the larger part of the system distortions was already corrected.

3.4 Defining system correction factors

The analysis derived in sections 3.2–3.3 allowed the determination of system correction factors from a single B-scan of a volumetric data set that subsequently could be applied to all other B-scans. Individual correction factors were obtained for setup distortions of both the optical path into the eye, as well as for the optical path out of the eye. The former is obtained from Eqs. (18) and (20) and is described as:
Qin(k)=Q(k)[eiφxy(k)200eiφxy(k)2]=[α(k)ei(kγ'+ρ(k))001]1Jsys1(k)Jc1(k)¯Rs1¯ei(kγ'+ρ(k))2,
(21)
and describes the most common realization of ν(k) multiplied with a phase factor. The latter is obtained out of the surface state description of Eq. (10) and Qin(k) as:
Qout(k)=(Esurf(k)Qin(k))1=Rs¯Jout1(k)[RH(k)00RV(k)]1ei(kγ'+ρ(k))2.
(22)
Using Qin(k) and Qout(k) the original sample data Esample(k) as given by Eq. (9) of every B-scan was corrected for all system distortions:
E^sample(k)=Qout(k)Esample(k)Qin(k)=Rs¯Jc1(k)¯Jc(k)Js(k)Jc(k)Jc1(k)¯Rs1¯ein(k)=Rs¯JrcT(k)Js(k)Jrc(k)Rs1¯ein(k),
(23)
where the residual corneal effect Jc(k)Jc1(k)¯ is replaced with its equivalent retarder Jrc(k).

4. Extraction of tissue polarization parameters

4.1 Depth-resolved tissue Jones matrix modeling

The nondepolarizing properties of tissues are normally modeled by Jones matrices in the form J = R−1·D·P·R [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

,48

48. S. Jiao, W. Yu, G. Stoica, and L. V. Wang, “Optical-fiber-based Mueller optical coherence tomography,” Opt. Lett. 28(14), 1206–1208 (2003). [CrossRef] [PubMed]

]. In this model R is a standard rotation matrix R = [cos(θ), -sin(θ); sin(θ), cos(θ)] that defines a common optic axis with orientation angle θ for the tissue birefringence and diattenuation [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

]. P and D are diagonal matrices that describe respectively the birefringence in the form of phase retardation η with complex phase factors as P = [eiη/2, 0; 0, e-iη/2] and the diattenuation with polarization dependent attenuation factors ε1,2 as D = [ε1, 0; 0, ε2]. J can be modeled as a function of wavenumber similar to first-order PMD models [49

49. D. Penninckx and V. Morénas, “Jones matrix of polarization mode dispersion,” Opt. Lett. 24(13), 875–877 (1999). [CrossRef] [PubMed]

] as J(k) = R−1·D·P(k)·R. Here only P(k) is considered wavenumber-dependent with η(k) as a linear function of k to model a depth-shift between the principle polarization states due to birefringence. In the current study this model applied to Js(k), Jc(k), and Jrc(k).

4.2 Intensity, phase retardation and diattenuation

The intensity of every pixel was calculated from Eq. (24) as the sum of the intensities for the individual field components:
I(z)=|E^H1(z)|2+|E^V1(z)|2+|E^H2(z)|2+|E^V2(z)|2.
(25)
In accordance with Park et al. [33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

] a unitary similarity transformation of Js(z) was calculated by extracting the surface state which was inversely multiplied with Eq. (24):
E^surf=Rs¯JrcTJrcRs1¯,
(26)
E^surf1E^sample(z)=Rs¯Jrc-1Js(z)JrcRs1¯ein(z).
(27)
Êsurf varies slowly across the B-scan as earlier pointed out by Cense et al. [4

4. B. Cense, T. C. Chen, B. H. Park, M. C. Pierce, and J. F. de Boer, “In vivo depth-resolved birefringence measurements of the human retinal nerve fiber layer by polarization-sensitive optical coherence tomography,” Opt. Lett. 27(18), 1610–1612 (2002). [CrossRef] [PubMed]

] and Pircher et al. [50

50. M. Pircher, E. Götzinger, B. Baumann, and C. K. Hitzenberger, “Corneal birefringence compensation for polarization sensitive optical coherence tomography of the human retina,” J. Biomed. Opt. 12(4), 041210 (2007). [CrossRef] [PubMed]

]. Local averaging of Êsurf along the B-scan was therefore applied to reduce noise while preserving the local corneal polarization properties. This was implemented using the spatial electric field averaging method described in section 4.4 with a kernel size of 40 A-scans. It was observed that strong specular reflections from the retinal surface can cause distortions in the extracted Êsurf. In these cases Êsurf was determined below the specular reflection to avoid artifacts in the calculated polarization parameters.

The eigendecomposition of Eq. (27) now yields the eigenvalues Ds(z)·Ps(z) = [λ1, 0; 0, λ2] from which the double-pass phase retardation η(z) (DPPR) was calculated as the phase-difference between the diagonal elements:
η(z)=|arg(λ1λ2*)|,
(28)
and the double-pass diattenuation d(z) as the relative difference in intensity:

d(z)=|ε12ε22ε12+ε22|=||λ1|2|λ2|2|λ1|2+|λ2|2|.
(29)

4.3 Relative optic axis orientation

The optic axis of Eq. (27) is the product of the sample optic axis defined by Rs and the unitary transformation JrcRs1¯. As previously described by Park et al. [51

51. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Optic axis determination accuracy for fiber-based polarization-sensitive optical coherence tomography,” Opt. Lett. 30(19), 2587–2589 (2005). [CrossRef] [PubMed]

], the plane of optic axes given by Rs is restricted to the QU-plane of the Poincaré sphere, but is arbitrarily rotated out of this plane by the unitary transformation. This results in an offset and sign ambiguity on the measured optic axis, and allows only the evaluation of optic axis differences instead of absolute values. In Eq. (27) the local corneal polarization properties within Jrc caused variations of the unitary transformation between A-scans and complicated the comparison of optic axis changes between different locations in the tissue. Consequently, these variations were corrected to confine Rs for all A-scans to the same Poincaré plane and optic axis offset.

The left eigenvectors of Eq. (27) are modeled as νz=Rs¯Jrc1Rs1Sz, where Sz is the eigendecomposition diagonal matrix with normalization factors. Assuming the cornea is free from diattenuation [52

52. F. Fanjul-Vélez, M. Pircher, B. Baumann, E. Götzinger, C. K. Hitzenberger, and J. L. Arce-Diego, “Polarimetric analysis of the human cornea measured by polarization-sensitive optical coherence tomography,” J. Biomed. Opt. 15(5), 056004 (2010). [CrossRef] [PubMed]

], νz consists only of unitary matrices and the normalization factors therefore only introduce a complex phase factor. In this case Jrc is modeled as Jrc=Rrc1PrcRrc and the eigendecomposition of Eq. (26) provides eigenvalues λrc=Prc2 and eigenvectors νrc=Rs¯Rrc1. These were used to construct a corneal correction factor for every A-scan: Qrc=νrcλrcνrc1=Rs¯JrcRs1¯. The corneal correction factor was then applied to νz for which the result is written in electric field notation:
νaxis=Qrcνz=Rs¯Rs1Sz=[Eν1xeiSz1Eν2xeiSz2Eν1yeiSz1Eν2yeiSz2],
(30)
where eiSz1 and eiSz2 are the eigendecomposition normalization phase factors, and Eν1x, Eν1y, Eν2x, and Eν2y are the electric field components of the polarization eigenstates. In Eq. (30) the local corneal effects of Jrc are cancelled out, but Rs¯ remains for every A-scan. Subsequently the Stokes vectors were calculated for both eigenstates which cancelled out the normalization phase factors. The obtained Stokes vectors point to opposite locations on the Poincaré sphere and define an optic axis Asam. For all sample locations Asam lied now on the same plane in the Poincaré sphere. The average Stokes vector over all pixels was calculated to define a reference axis Aref, which was used to calculate the relative optic axis orientation angle θ as the negative angle between Asam and Aref. Since Eq. (27) is corrected for the retinal surface polarization state, the obtained θ measures the change in the optic axis orientation relative to the surface with inclusion of an arbitrary offset induced by Rs¯ and Aref. In practice Aref was calculated from a single B-scan and used as the reference axis for the whole data set.

4.4 Spatial electric field averaging for SNR improvement and speckle reduction

Speckle noise and low SNR can significantly degrade the quality of measured tissue polarization parameters [46

46. Y. Lim, M. Yamanari, S. Fukuda, Y. Kaji, T. Kiuchi, M. Miura, T. Oshika, and Y. Yasuno, “Birefringence measurement of cornea and anterior segment by office-based polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 2(8), 2392–2402 (2011). [CrossRef] [PubMed]

,53

53. E. Götzinger, M. Pircher, B. Baumann, T. Schmoll, H. Sattmann, R. A. Leitgeb, and C. K. Hitzenberger, “Speckle noise reduction in high speed polarization sensitive spectral domain optical coherence tomography,” Opt. Express 19(15), 14568–14585 (2011). [CrossRef] [PubMed]

]. A spatial averaging method for complex electric fields was therefore developed to simultaneously improve the SNR and reduce speckle noise, at the expense of a small reduction in resolution. This method uses similar basic concepts as described for the Jones matrix averaging methods of Lim et al. [46

46. Y. Lim, M. Yamanari, S. Fukuda, Y. Kaji, T. Kiuchi, M. Miura, T. Oshika, and Y. Yasuno, “Birefringence measurement of cornea and anterior segment by office-based polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 2(8), 2392–2402 (2011). [CrossRef] [PubMed]

] and Ju et al. [34

34. M. J. Ju, Y. J. Hong, S. Makita, Y. Lim, K. Kurokawa, L. Duan, M. Miura, S. Tang, and Y. Yasuno, “Advanced multi-contrast Jones matrix optical coherence tomography for Doppler and polarization sensitive imaging,” Opt. Express 21(16), 19412–19436 (2013). [CrossRef] [PubMed]

].

Each pixel is subject to an individual offset in amplitude and phase as given by ein(z) in Eq. (24). The phase offset is seemingly random for every pixel and spatial complex averaging between pixels is therefore only effective when these phase offsets are corrected. The simplest way to do so is to divide the electric field components of each pixel through their individual ÊH1 (or any other) field component, which removes the offsets in amplitude and phase and aligns ÊH1 for all pixels. This simple method is however not optimal since the alignment on a single field component is subject to (speckle) noise. The method is therefore extended by including the other field components and neighboring pixels in the alignment process. A spatial kernel was defined including n pixels. For every pixel the phase angle between the reference electric field component ÊH1 and the other components ÊV1, ÊH2, and ÊV2 were calculated. Since these angles depended on the local polarization properties of the tissue, they had a strong spatial consistency, and were therefore averaged over all n pixels in the kernel with a weighting by their amplitudes to suppress (speckle) noise influences:
ψV1¯=arg(nE^H1E^V1*),ψH2¯=arg(nE^H1E^H2*),ψV2¯=arg(nE^H1E^V2*).
(31)
The phase angles of Eq. (31) were used to correct ÊV1, ÊH2, and ÊV2 to the phase angle of ÊH1, and subsequently the phase angle of the vector sum over all field components was calculated:
ψcor=arg(E^H1+E^V1eiψV1¯+E^H2eiψH2¯+E^V2eiψV2¯),
(32)
which gave the estimated noise-free phase angle of ÊH1. The angle ψcor was used to correct the four original electric field components of the center pixel of the kernel. This procedure was applied to all pixels in a B-scan to obtain a uniform alignment of the phase offset. Afterwards, spatial complex averaging was applied per electric field component using the same spatial kernel size to improve the SNR and reduce the speckle noise for every pixel.

In practice, the current method used one of the diagonal electric field components ÊH1 or ÊV2 as the reference field component, since in the nearly similar matrix form of Eq. (24) these have the largest amplitudes and give therefore the best performance. The method is however easily adapted to use other reference components in case the amplitudes of ÊH1 or ÊV2 are insufficient.

5. System performance and imaging results

5.1 Technical system evaluation

In order to evaluate the system polarization distortions a healthy volunteer was measured and the polarization state at the retinal surface was determined as a function of wavenumber as described by Eq. (10). The sample surface state effectively describes the polarization transfer function of the system including the cornea, and can be used to estimate the depth shift between the input polarization states due to PMD in the system [30

30. M. Villiger, E. Z. Zhang, S. Nadkarni, W. Y. Oh, B. E. Bouma, and B. J. Vakoc, “Artifacts in polarization-sensitive optical coherence tomography caused by polarization mode dispersion,” Opt. Lett. 38(6), 923–925 (2013). [CrossRef] [PubMed]

]. Ten data sets were acquired with different orientations of the sample arm fiber to determine the average PMD depth shift of the system. An average PMD depth shift of 1.6 ± 0.44 µm (0.29 ± 0.08 pixel) was measured which is equivalent to a time delay of 11 ± 2.9 fs. This value is in agreement with earlier measurements of Villiger et al. [30

30. M. Villiger, E. Z. Zhang, S. Nadkarni, W. Y. Oh, B. E. Bouma, and B. J. Vakoc, “Artifacts in polarization-sensitive optical coherence tomography caused by polarization mode dispersion,” Opt. Lett. 38(6), 923–925 (2013). [CrossRef] [PubMed]

] in a swept-source OCT system at 1310 nm without fiber-optic circulators.

The compensation of the system polarization distortions by the correction steps described in sections 3.2-3.4 is illustrated in Fig. 4
Fig. 4 Normalized Stokes vector traces plotted in the Poincaré sphere as a function of wavenumber for the vertical linear input state at the retinal surface state (row (A)) and for birefringent RNFL tissue at a depth of 159 µm (row (B)) without (left) and with (right) correction for system polarization distortions. (A) Traces are plotted for the first (blue), middle (red), and last (green) B-scans of a volumetric data set. The Stokes vector traces described identical paths before correction over nearly a quarter of the globe, and reduced to single locations after correction. (B) The normalized Stokes vector trace of the RNFL tissue described a similar path over the Poincaré sphere as the retinal surface state before correction, and reduced to a small location after correction.
. In row (A) the retinal surface polarization state was plotted as a function of wavenumber in normalized Stokes vectors on the Poincaré sphere. The first (blue), the middle (red), and the last (green) B-scans from a typical volumetric data set of the retina were selected, for which the vertical linearly polarized input state was plotted. The B-scans were acquired 3.3 s apart from each other over a total volume acquisition time of 6.6 s. The left Poincaré sphere of Fig. 4(A) shows the Stokes vector traces for the case without correction, which traveled over an arc of approximately a quarter of the globe from the -U to the +V pole. The nearly identical traces indicate that the system polarization distortions were constant over the acquisition time of the data set. Small shifts in the global location of the traces are attributed to changes in corneal polarization properties. The right Poincaré sphere of Fig. 4(A) shows the same Stokes vector traces after applying the correction for system polarization distortions. The system correction factors were calculated from the middle B-scan and subsequently applied to the other B-scans. It can be clearly seen that after correction the Stokes vector traces reduced to single locations on the Poincaré sphere, which indicates the complete correction of the system polarization distortions. In addition, the locations near the -Q pole were consistent with vertical linearly polarized light, while the offsets from this pole are attributed to different corneal birefringence for these B-scans. The horizontal linear input state showed the same behavior at the exact other side of the Poincaré sphere (not shown). These results confirmed that it is possible to apply the system correction factors obtained from a single B-scan to the whole data set, while maintaining optimal correction.

The same analysis was performed for the polarization state of birefringent RNFL tissue that was obtained slightly inferior from the optic nerve head at a depth of 159 µm. The analysis was performed on a cross-sectional tissue area of 276 µm × 36 µm (width × depth) with an average DPPR of 0.53 rad. In row (B) of Fig. 4 the left Poincaré sphere shows the average normalized Stokes vector trace for the vertical linearly polarized input state as a function of wavenumber without correction for system polarization distortions. In this case the polarization state of the RNFL tissue follows a trace similar as was observed for the uncorrected polarization state at the retinal surface. The right Poincaré sphere shows the same Stokes vector trace after correction for the system polarization distortions. It can be clearly seen that after correction the Stokes vector trace reduced to a small location on the Poincaré sphere, which indicates a successful correction of the system polarization distortions. These results confirm the applicability of the developed correction method to in vivo birefringent tissues.

In addition a quarter-wave plate (10RP34-1064, Newport, CA, USA) was measured at different orientation angles to evaluate the accuracy in determining DPPR and optic axis orientation [25

25. M. Yamanari, S. Makita, and Y. Yasuno, “Polarization-sensitive swept-source optical coherence tomography with continuous source polarization modulation,” Opt. Express 16(8), 5892–5906 (2008). [CrossRef] [PubMed]

,33

33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

]. The measured DPPR with and without correction for system polarization distortions was respectively 3.05 ± 0.07 rad and 3.05 ± 0.06 rad, for a manufacturer specification of 3.08 rad at 1040 nm. The difference between the set and measured axis orientation with and without correction was respectively 1.8 ± 1.6 deg and 1.5 ± 1.2 deg. The equivalent results with and without correction indicated that the system polarization distortions have minimal impact on the measurement of strong birefringent samples, and accurate values were obtained for the phase retardation and optic axis orientation in both cases.

5.2 Cross-sectional PS-OCT imaging in a healthy volunteer

In Fig. 5
Fig. 5 The correction of system polarization distortions in retinal B-scans of a healthy volunteer. The images of the polarization parameters are paired and displayed without and with correction respectively for a size of 1.3 mm × 6.6 mm (depth × width). (A) Intensity. (B&C) DPPR for a π/2 range. (D&E) DPPR for a π/4 range. (F&G) Diattenuation. (H&I) Relative optic axis orientation in Poincaré representation.
the effect of correcting system polarization distortions is shown in cross-sectional PS-OCT images of the macula in a healthy volunteer. The images were obtained slightly inferior to the foveal center, consist of 2000 A-scans, and are displayed over 1.3 mm in depth (in tissue) and 6.6 mm (22°) in width. Spatial electric field averaging was performed over 3 × 7 pixels (depth × width) to reduce (speckle) noise and improve the SNR. Pixels outside the tissue were detected by an intensity threshold (SNR < 8 dB) and are displayed in gray in the images of the polarization parameters. In Fig. 5(A) the intensity information visualizes the structures of the fovea (right side) and a part of the inferior nerve fiber bundle (left side).

Images of the DPPR without and with correction for the system polarization distortions are shown respectively in Figs. 5(B) and 5(C). In both images a clear increase in DPPR was observed for the inferior nerve fiber bundle and the sclera. Especially the DPPR of the sclera increased rapidly with depth, and saturated the image dynamic range. The most dominant difference was the more uniform (less noisy) appearance of the DPPR after correction. Also the increased DPPR caused by Henle's fiber layer was shown more clearly after correction, which was observed in the inner/outer segment junctional layer (IS/OS) flanking the center of the fovea. These differences were even better appreciated when the dynamic range of the images was scaled back from π/2 to π/4 in Figs. 5(D) and 5(E) to visualize small changes in DPPR without and with correction. In these images it is evident that the noise on the DPPR was lower after correction, which was confirmed by the average measured DPPR of 0.20 ± 0.19 rad without and 0.14 ± 0.11 rad with correction for a 10 pixel thick tissue layer starting from the retinal surface. In addition the average DPPR of Henle's fiber layer was measured as 0.22 ± 0.08 rad without and 0.28 ± 0.06 rad with correction. The latter is in close agreement with results reported by Cense et al. [6

6. B. Cense, Q. Wang, S. Lee, L. Zhao, A. E. Elsner, C. K. Hitzenberger, and D. T. Miller, “Henle fiber layer phase retardation measured with polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 4(11), 2296–2306 (2013). [CrossRef] [PubMed]

] who used a PS-OCT system with minimal SMF length and consequently should have been minimally affected by PMD.

Figures 5(H) and 5(I) show the relative optic axis orientation without and with correction for the system polarization distortions, respectively. The relative optic axis orientation was noisy for pixels with a low SNR or a low phase retardation since the eigenvectors are poorly defined in these cases. This occurs especially at the retinal surface where the surface state correction removed all polarizing effects, but was also evident around the fovea where polarizing tissue structures were absent. In the image without correction there is however a false indication of a uniform optic axis orientation around the fovea, which was due to the data deviating from the SU(2) form in the analysis of section 4.3. The optic axis orientation was however well defined in structures with strong DPPR such as the RNFL inferior nerve bundle and the sclera, and shows equivalent results with and without correction. This is in contrast with the axis orientation cast by Henle's fiber layer in the IS/OS layer, which only revealed the expected radial distribution after correction and is seen here in cross-section.

In Fig. 6
Fig. 6 Cross-sectional imaging of the superior optic nerve bundle of a healthy volunteer. Image sizes: 0.9 mm × 5.2 mm (depth × width). (A) Intensity. (B) Diattenuation. (C) DPPR. (D) Optic axis orientation in Poincaré representation.
images are shown for the superior optic nerve bundle with correction for system polarization distortions. The images consist of 2000 A-scans and are displayed over 0.9 mm indepth (in tissue) and 5.2 mm (17.3°) in width. The thick RNFL as shown in Fig. 6(A) clearly demonstrates the negligible diattenuation for this tissue in Fig. 6(B). In addition, the nerve fiber bundles created strong DPPR and are shown in clusters separated by blood vessels in Fig. 6(C). In Fig. 6(D) the relative optic axis orientation visualizes the direction of the fiber bundles in leaving the optic nerve head area. The nearly twofold color cycle in the lateral direction indicates a full rotation of the optic axes in the Poincaré sphere. This is consistent with the nerve fibers leaving the optic nerve head in directions distributed over a 180° angle in real space and is in agreement with earlier PS-OCT findings of Pircher et al. [7

7. M. Pircher, E. Götzinger, R. Leitgeb, H. Sattmann, O. Findl, and C. Hitzenberger, “Imaging of polarization properties of human retina in vivo with phase resolved transversal PS-OCT,” Opt. Express 12(24), 5940–5951 (2004). [CrossRef] [PubMed]

].

5.3 En face retinal PS-OCT imaging in a healthy volunteer and a glaucoma patient

The performance of the developed PS-OCT system for clinical measurements is demonstrated by imaging the macula and optic nerve head areas from a healthy volunteer and a glaucoma patient. For each area a volumetric PS-OCT data set was obtained with 300 B-scans and 2000 A-scans / B-scan in an acquisition time of 6.6 s. Compiled en face PS-OCT maps were created for the intensity, the DPPR, and the relative optic axis orientation. The intensity en face map was calculated by summing the logarithmic scaled intensity over depth including both the retina and choroid. The en face maps of the polarization parameters were obtained from the IS/OS since this layer displays the polarization effects for the complete retina with high SNR [7

7. M. Pircher, E. Götzinger, R. Leitgeb, H. Sattmann, O. Findl, and C. Hitzenberger, “Imaging of polarization properties of human retina in vivo with phase resolved transversal PS-OCT,” Opt. Express 12(24), 5940–5951 (2004). [CrossRef] [PubMed]

,17

17. S. Zotter, M. Pircher, T. Torzicky, B. Baumann, H. Yoshida, F. Hirose, P. Roberts, M. Ritter, C. Schütze, E. Götzinger, W. Trasischker, C. Vass, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Large-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology,” Biomed. Opt. Express 3(11), 2720–2732 (2012). [CrossRef] [PubMed]

]. In addition, the RNFL thickness of both volunteers was measured with Scanning Laser Polarimetry (SLP) using a GDxPRO (Carl Zeiss Meditec Inc., CA, USA). This instrument scans the retina with polarized light at 785 nm and converts the detected DPPR to a measure of RNFL thickness using a fixed factor of 0.67 nm/µm (single pass phase retardation to thickness) [56

56. Q. Zhou, J. Reed, R. Betts, P. Trost, P. Lo, C. Wallace, R. Bienias, G. Li, R. Winnick, W. Papworth, and M. Sinai, “Detection of glaucomatous retinal nerve fiber layer damage by scanning laser polarimetry with variable corneal compensation,” Proc. SPIE 4951, 32–41 (2003). [CrossRef]

]. The SLP thickness maps are plotted in their native clinically used color map, and cropped and positioned according to the PS-OCT field of view.

The results for the right eye of the healthy volunteer are shown in Fig. 7
Fig. 7 En face maps for the right eye of a healthy volunteer. Image sizes: 5.3 mm × 9.6 mm (height × width). (A) Intensity projection. (B) DPPR. (C) Optic axis orientation in Poincaré representation. (D) SLP RNFL thickness map, 200 µm RNFL thickness is equal to 2.15 rad DPPR. Strong retardation and a well defined optic axis orientation were observed for the nerve fiber bundles leaving the optic nerve head and for the Henle fibers in the fovea. This is consistent with the RNFL thickness map obtained by SLP, which showed a similar retardation pattern.
for a field of view of 5.3 × 9.6 mm2 (17.8° × 32.1°). In Fig. 7(A) the en face intensity image displays retinal structures such as several blood vessels leaving the optic nerve head. In Fig. 7(B) the en face map of the retinal DPPR shows the strong retardation of the nerve fiber bundles leaving the optic nerve head in the superior and inferior directions. Around the fovea a (donut) ring of increased retardation can be seen caused by Henle's fiber layer. The optic axis orientation is shown in Fig. 7(C), which clearly visualizes the orientation of the nerve fibers leaving the optic nerve head in all radial directions, while the radial extension of the Henle fibers is clearly observed in the fovea. Note that in PS-OCT the measured optic axis orientation angles in the Stokes convention are subject to a pi-ambiguity in the Poincaré sphere [51

51. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Optic axis determination accuracy for fiber-based polarization-sensitive optical coherence tomography,” Opt. Lett. 30(19), 2587–2589 (2005). [CrossRef] [PubMed]

]. This leads to a wrapping of the optic axis orientation every 90° in real space, and consequently gives a fourfold wrapping for a full 360° rotation as is shown for the fovea and optic nerve head [6

6. B. Cense, Q. Wang, S. Lee, L. Zhao, A. E. Elsner, C. K. Hitzenberger, and D. T. Miller, “Henle fiber layer phase retardation measured with polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 4(11), 2296–2306 (2013). [CrossRef] [PubMed]

]. The RNFL thickness map as measured by SLP is given in Fig. 7(D) and shows an equivalent retardation pattern as observed with PS-OCT including the retardation ring from Henle's fiber layer and the nerve bundles superior and inferior of the optic nerve head.

The results for the left eye of the glaucoma patient are shown in Fig. 8
Fig. 8 En face maps for the left eye of a glaucoma patient. Image sizes: 5.8 mm × 8.9 mm (height × width). (A) Intensity projection. (B) DPPR. (C) Optic axis orientation in Poincaré representation. (D) SLP RNFL thickness map, 200 µm RNFL thickness is equal to 2.15 rad DPPR. Reduced retardation was observed for the superior and inferior nerve fiber bundles in comparison with the healthy volunteer. The loss in the inferior bundle was severe enough to lose information on the optic axis orientation in the lower left image corner. Also the SLP RNFL thickness map showed a reduction in retardation for the inferior nerve fiber bundle.
for a field of view of 5.8 × 8.9 mm2 (19.3° × 29.7°). Similar to the healthy subject, Fig. 8(A) shows the en face intensity image with the retinal structures. Figure 8(B) shows the en face DPPR map with significantly lower retardation in the superior nerve bundle and an almost absent retardation in the inferior bundle in comparison with the healthy volunteer. In contrast the retardation ring of Henle's fiber layer was well preserved. The optic axis orientation in Fig. 8(C) shows the severe loss in the inferior nerve bundle as an absence in orientation which is uniformly colored in cyan in the lower left image corner. The SLP thickness map shows a similar pattern with a significantly decreased thickness for the inferior RNFL bundle as can be seen in Fig. 8(D). This is in agreement with the PS-OCT findings, although seemingly more retardation was observed by SLP than with PS-OCT. This difference might be attributed to an imperfect corneal compensation for the SLP [19

19. E. Götzinger, M. Pircher, B. Baumann, C. Hirn, C. Vass, and C. K. Hitzenberger, “Analysis of the origin of atypical scanning laser polarimetry patterns by polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 49(12), 5366–5372 (2008). [CrossRef] [PubMed]

]. The retardation ring of Henle's fiber layer was however observed similarly to the PS-OCT.

6. Discussion

Recently Zhang et al. [31

31. E. Z. Zhang, W. Y. Oh, M. L. Villiger, L. Chen, B. E. Bouma, and B. J. Vakoc, “Numerical compensation of system polarization mode dispersion in polarization-sensitive optical coherence tomography,” Opt. Express 21(1), 1163–1180 (2013). [CrossRef] [PubMed]

] demonstrated how three calibration signals can be used to solve for all the system polarization distortions in a fiber-based PS-OCT setup if no a priori knowledge on the theoretical setup modeling is considered. In the current study we show that through careful Jones matrix modeling of the system, a similar method was developed in which only two calibration signals were extracted from the sample data (tissue surface state & deeper located birefringent structures) to solve and correct for the system polarization distortions. The advantage of the current method is that no additional calibration signals and associated hardware were needed, which simplified the setup and maximized the OCT imaging depth range for sample measurements. Although the current method used spectral binning for its analysis similar to Villiger et al. [32

32. M. Villiger, E. Z. Zhang, S. K. Nadkarni, W. Y. Oh, B. J. Vakoc, and B. E. Bouma, “Spectral binning for mitigation of polarization mode dispersion artifacts in catheter-based optical frequency domain imaging,” Opt. Express 21(14), 16353–16369 (2013). [CrossRef] [PubMed]

], this did not reduce the axial resolution of the final data, since the corrections were always applied to the original full spectrum PS-OCT data. In the current study a small resolution penalty was however accepted for spatial pixel averaging in order to improve the SNR and reduce the speckle noise.

The current method takes advantage of the sample polarization properties and extracts setup effects from polarization changes caused by birefringent tissue structures. The eye is a suitable sample for this method since birefringent tissues are readily available in the RNFL, Henle's fiber layer and the sclera. However, the birefringent tissue analysis of section 3.3 used histograms to (indirectly) measure the most common optic axis orientation, and was sensitive in its performance to (heavy) variations in the optic axis. This analysis step is therefore best performed on data obtained from tissues with a uniform optic axis orientation. In the eye this data can be easily obtained from B-scans over a line through the center of the fovea, or from horizontal line B-scans through the optic nerve head. The method presented in this paper can therefore be applied with minimal modifications to other OCT applications (e.g. skin imaging) where also birefringent tissue with a locally uniform optic axis orientation is available.

The impact of the system polarization distortions in this study was found to be significant with respect to the weak birefringent tissues in the retina. In contrast, a quarter-wave plate, a strong artificial retarder, was found to be negligibly affected. This implies that the impact of system polarization distortions scales inversely with the strength of the retarder, and that it is important for measurements in low birefringent tissue to correct even the moderate PMD levels that were reported in this study.

Further methods to reduce the system polarization distortions in the current setup can be found in hardware modifications. Obviously, the PMD could be minimized by shortening the significant SMF lengths [21

21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

,29

29. B. Elmaanaoui, B. Wang, J. C. Dwelle, A. B. McElroy, S. S. Liu, H. G. Rylander, and T. E. Milner, “Birefringence measurement of the retinal nerve fiber layer by swept source polarization sensitive optical coherence tomography,” Opt. Express 19(11), 10252–10268 (2011). [CrossRef] [PubMed]

]. In addition, a long coherence length swept-source, such as the VCSEL swept-source [57

57. I. Grulkowski, J. J. Liu, B. Potsaid, V. Jayaraman, C. D. Lu, J. Jiang, A. E. Cable, J. S. Duker, and J. G. Fujimoto, “Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers,” Biomed. Opt. Express 3(11), 2733–2751 (2012). [CrossRef] [PubMed]

], could be used to minimize distortions due to signal roll-off along imaging depth.

In the current study the data processing was not optimized for computational efficiency and had a relatively long computation time with an average duration of 15.4 seconds per B-scan consisting of 2200 A-scans on a computer system equipped with an Intel Xeon W3680 hexa-core processor @ 3.33 GHz and 18 GB of RAM. The analysis and the correction of the system polarization distortions (section 3) contributed to ~5% of the computation time, which was significantly less than the ~80% computation time that was spent on the extraction of the depth-resolved polarization parameters (sections 4.2 and 4.3).

7. Conclusions

In conclusion, we have demonstrated PS-OCT of the in vivo human retina with a novel analysis method to correct for system polarization distortions that works directly on the sample data. The correction method reduced noise and corrected erroneous offsets on the measured double-pass phase retardation, diattenuation, and relative optic axis orientation. This allowed for high-resolution cross-sectional and en face visualization of the polarization parameters in a healthy volunteer and a glaucoma patient. The clinical potential of the technology was demonstrated by the detection of nerve fiber bundle loss in the latter case. It is therefore expected that PS-OCT can also be a valuable tool to study other types of neurodegenerative diseases with retinal nerve fiber loss, such as multiple sclerosis [58

58. A. Petzold, J. F. de Boer, S. Schippling, P. Vermersch, R. Kardon, A. Green, P. A. Calabresi, and C. Polman, “Optical coherence tomography in multiple sclerosis: a systematic review and meta-analysis,” Lancet Neurol. 9(9), 921–932 (2010). [CrossRef] [PubMed]

]. Especially when PS-OCT is combined with eye tracking techniques to reduce motion artifacts [59

59. M. Sugita, S. Zotter, M. Pircher, T. Makihira, K. Saito, N. Tomatsu, M. Sato, P. Roberts, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Motion artifact and speckle noise reduction in polarization sensitive optical coherence tomography by retinal tracking,” Biomed. Opt. Express 5(1), 106–122 (2014). [CrossRef] [PubMed]

,60

60. K. V. Vienola, B. Braaf, C. K. Sheehy, Q. Yang, P. Tiruveedhula, D. W. Arathorn, J. F. de Boer, and A. Roorda, “Real-time eye motion compensation for OCT imaging with tracking SLO,” Biomed. Opt. Express 3(11), 2950–2963 (2012). [CrossRef] [PubMed]

] and complemented with OCT angiography techniques [34

34. M. J. Ju, Y. J. Hong, S. Makita, Y. Lim, K. Kurokawa, L. Duan, M. Miura, S. Tang, and Y. Yasuno, “Advanced multi-contrast Jones matrix optical coherence tomography for Doppler and polarization sensitive imaging,” Opt. Express 21(16), 19412–19436 (2013). [CrossRef] [PubMed]

,61

61. B. Braaf, K. V. Vienola, C. K. Sheehy, Q. Yang, K. A. Vermeer, P. Tiruveedhula, D. W. Arathorn, A. Roorda, and J. F. de Boer, “Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO,” Biomed. Opt. Express 4(1), 51–65 (2013). [CrossRef] [PubMed]

] the capabilities for studying and diagnosing ocular pathology are expected to be promising, and this is therefore the focus for future work.

Acknowledgements

This research was supported by grants from Stichting Wetenschappelijk Onderzoek Oogziekenhuis Prof. Dr. H.J. Flieringa (SWOO), Combined Ophthalmic Research Rotterdam (CORR), the Dutch MS Research Foundation, and a VICI (JFdB) from the Netherlands Organization for Scientific Research (NWO). The authors would like to thank J. Li and F. Feroldi for helpful discussions, J. Weda for hardware assistance, and R. Eijlers and G. Thepass for performing the SLP measurements.

References and links

1.

D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]

2.

W. Drexler and J. G. Fujimoto, “State-of-the-art retinal optical coherence tomography,” Prog. Retin. Eye Res. 27(1), 45–88 (2008). [CrossRef] [PubMed]

3.

E. Götzinger, M. Pircher, M. Sticker, A. F. Fercher, and C. K. Hitzenberger, “Measurement and imaging of birefringent properties of the human cornea with phase-resolved, polarization-sensitive optical coherence tomography,” J. Biomed. Opt. 9(1), 94–102 (2004). [CrossRef] [PubMed]

4.

B. Cense, T. C. Chen, B. H. Park, M. C. Pierce, and J. F. de Boer, “In vivo depth-resolved birefringence measurements of the human retinal nerve fiber layer by polarization-sensitive optical coherence tomography,” Opt. Lett. 27(18), 1610–1612 (2002). [CrossRef] [PubMed]

5.

B. Cense, T. C. Chen, B. H. Park, M. C. Pierce, and J. F. de Boer, “Thickness and birefringence of healthy retinal nerve fiber layer tissue measured with polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 45(8), 2606–2612 (2004). [CrossRef] [PubMed]

6.

B. Cense, Q. Wang, S. Lee, L. Zhao, A. E. Elsner, C. K. Hitzenberger, and D. T. Miller, “Henle fiber layer phase retardation measured with polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 4(11), 2296–2306 (2013). [CrossRef] [PubMed]

7.

M. Pircher, E. Götzinger, R. Leitgeb, H. Sattmann, O. Findl, and C. Hitzenberger, “Imaging of polarization properties of human retina in vivo with phase resolved transversal PS-OCT,” Opt. Express 12(24), 5940–5951 (2004). [CrossRef] [PubMed]

8.

E. Götzinger, M. Pircher, W. Geitzenauer, C. Ahlers, B. Baumann, S. Michels, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Retinal pigment epithelium segmentation by polarization sensitive optical coherence tomography,” Opt. Express 16(21), 16410–16422 (2008). [CrossRef] [PubMed]

9.

M. Pircher, E. Götzinger, O. Findl, S. Michels, W. Geitzenauer, C. Leydolt, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Human macula investigated in vivo with polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 47(12), 5487–5494 (2006). [CrossRef] [PubMed]

10.

M. Yamanari, Y. Lim, S. Makita, and Y. Yasuno, “Visualization of phase retardation of deep posterior eye by polarization-sensitive swept-source optical coherence tomography with 1- µm probe,” Opt. Express 17(15), 12385–12396 (2009). [CrossRef] [PubMed]

11.

B. Baumann, E. Götzinger, M. Pircher, and C. K. Hitzenberger, “Single camera based spectral domain polarization sensitive optical coherence tomography,” Opt. Express 15(3), 1054–1063 (2007). [CrossRef] [PubMed]

12.

M. R. Hee, D. Huang, E. A. Swanson, and J. G. Fujimoto, “Polarization-sensitive low-coherence reflectometer for birefringence characterization and ranging,” J. Opt. Soc. Am. B 9(6), 903–908 (1992). [CrossRef]

13.

J. F. de Boer, T. E. Milner, M. J. van Gemert, and J. S. Nelson, “Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography,” Opt. Lett. 22(12), 934–936 (1997). [CrossRef] [PubMed]

14.

M. Pircher, C. K. Hitzenberger, and U. Schmidt-Erfurth, “Polarization sensitive optical coherence tomography in the human eye,” Prog. Retin. Eye Res. 30(6), 431–451 (2011). [CrossRef] [PubMed]

15.

F. G. Schlanitz, B. Baumann, T. Spalek, C. Schütze, C. Ahlers, M. Pircher, E. Götzinger, C. K. Hitzenberger, and U. Schmidt-Erfurth, “Performance of automated drusen detection by polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 52(7), 4571–4579 (2011). [CrossRef] [PubMed]

16.

E. Götzinger, M. Pircher, B. Baumann, C. Ahlers, W. Geitzenauer, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Three-dimensional polarization sensitive OCT imaging and interactive display of the human retina,” Opt. Express 17(5), 4151–4165 (2009). [CrossRef] [PubMed]

17.

S. Zotter, M. Pircher, T. Torzicky, B. Baumann, H. Yoshida, F. Hirose, P. Roberts, M. Ritter, C. Schütze, E. Götzinger, W. Trasischker, C. Vass, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Large-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology,” Biomed. Opt. Express 3(11), 2720–2732 (2012). [CrossRef] [PubMed]

18.

J. Dwelle, S. Liu, B. Wang, A. McElroy, D. Ho, M. K. Markey, T. Milner, and H. G. Rylander 3rd, “Thickness, phase retardation, birefringence, and reflectance of the retinal nerve fiber layer in normal and glaucomatous non-human primates,” Invest. Ophthalmol. Vis. Sci. 53(8), 4380–4395 (2012). [CrossRef] [PubMed]

19.

E. Götzinger, M. Pircher, B. Baumann, C. Hirn, C. Vass, and C. K. Hitzenberger, “Analysis of the origin of atypical scanning laser polarimetry patterns by polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 49(12), 5366–5372 (2008). [CrossRef] [PubMed]

20.

C. Ahlers, E. Götzinger, M. Pircher, I. Golbaz, F. Prager, C. Schütze, B. Baumann, C. K. Hitzenberger, and U. Schmidt-Erfurth, “Imaging of the retinal pigment epithelium in age-related macular degeneration using polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 51(4), 2149–2157 (2010). [CrossRef] [PubMed]

21.

C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett. 25(18), 1355–1357 (2000). [CrossRef] [PubMed]

22.

B. H. Park, C. Saxer, S. M. Srinivas, J. S. Nelson, and J. F. de Boer, “In vivo burn depth determination by high-speed fiber-based polarization sensitive optical coherence tomography,” J. Biomed. Opt. 6(4), 474–479 (2001). [CrossRef] [PubMed]

23.

W. Y. Oh, B. J. Vakoc, S. H. Yun, G. J. Tearney, and B. E. Bouma, “Single-detector polarization-sensitive optical frequency domain imaging using high-speed intra A-line polarization modulation,” Opt. Lett. 33(12), 1330–1332 (2008). [CrossRef] [PubMed]

24.

W. Y. Oh, S. H. Yun, B. J. Vakoc, M. Shishkov, A. E. Desjardins, B. H. Park, J. F. de Boer, G. J. Tearney, and B. E. Bouma, “High-speed polarization sensitive optical frequency domain imaging with frequency multiplexing,” Opt. Express 16(2), 1096–1103 (2008). [CrossRef] [PubMed]

25.

M. Yamanari, S. Makita, and Y. Yasuno, “Polarization-sensitive swept-source optical coherence tomography with continuous source polarization modulation,” Opt. Express 16(8), 5892–5906 (2008). [CrossRef] [PubMed]

26.

K. H. Kim, B. H. Park, Y. Tu, T. Hasan, B. Lee, J. Li, and J. F. de Boer, “Polarization-sensitive optical frequency domain imaging based on unpolarized light,” Opt. Express 19(2), 552–561 (2011). [CrossRef] [PubMed]

27.

Y. Lim, Y. J. Hong, L. Duan, M. Yamanari, and Y. Yasuno, “Passive component based multifunctional Jones matrix swept source optical coherence tomography for Doppler and polarization imaging,” Opt. Lett. 37(11), 1958–1960 (2012). [CrossRef] [PubMed]

28.

B. Baumann, W. Choi, B. Potsaid, D. Huang, J. S. Duker, and J. G. Fujimoto, “Swept source/Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unit,” Opt. Express 20(9), 10229–10241 (2012). [CrossRef] [PubMed]

29.

B. Elmaanaoui, B. Wang, J. C. Dwelle, A. B. McElroy, S. S. Liu, H. G. Rylander, and T. E. Milner, “Birefringence measurement of the retinal nerve fiber layer by swept source polarization sensitive optical coherence tomography,” Opt. Express 19(11), 10252–10268 (2011). [CrossRef] [PubMed]

30.

M. Villiger, E. Z. Zhang, S. Nadkarni, W. Y. Oh, B. E. Bouma, and B. J. Vakoc, “Artifacts in polarization-sensitive optical coherence tomography caused by polarization mode dispersion,” Opt. Lett. 38(6), 923–925 (2013). [CrossRef] [PubMed]

31.

E. Z. Zhang, W. Y. Oh, M. L. Villiger, L. Chen, B. E. Bouma, and B. J. Vakoc, “Numerical compensation of system polarization mode dispersion in polarization-sensitive optical coherence tomography,” Opt. Express 21(1), 1163–1180 (2013). [CrossRef] [PubMed]

32.

M. Villiger, E. Z. Zhang, S. K. Nadkarni, W. Y. Oh, B. J. Vakoc, and B. E. Bouma, “Spectral binning for mitigation of polarization mode dispersion artifacts in catheter-based optical frequency domain imaging,” Opt. Express 21(14), 16353–16369 (2013). [CrossRef] [PubMed]

33.

B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett. 29(21), 2512–2514 (2004). [CrossRef] [PubMed]

34.

M. J. Ju, Y. J. Hong, S. Makita, Y. Lim, K. Kurokawa, L. Duan, M. Miura, S. Tang, and Y. Yasuno, “Advanced multi-contrast Jones matrix optical coherence tomography for Doppler and polarization sensitive imaging,” Opt. Express 21(16), 19412–19436 (2013). [CrossRef] [PubMed]

35.

B. Braaf, K. A. Vermeer, V. A. Sicam, E. van Zeeburg, J. C. van Meurs, and J. F. de Boer, “Phase-stabilized optical frequency domain imaging at 1-µm for the measurement of blood flow in the human choroid,” Opt. Express 19(21), 20886–20903 (2011). [CrossRef] [PubMed]

36.

N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12(3), 367–376 (2004). [CrossRef] [PubMed]

37.

N. V. Iftimia, D. X. Hammer, C. E. Bigelow, D. I. Rosen, T. Ustun, A. A. Ferrante, D. Vu, and R. D. Ferguson, “Toward noninvasive measurement of blood hematocrit using spectral domain low coherence interferometry and retinal tracking,” Opt. Express 14(8), 3377–3388 (2006). [CrossRef] [PubMed]

38.

B. H. Park and J. F. De Boer, “Polarization-Sensitive Optical Coherence Tomography” in Optical Coherence Tomography: Technology and Applications, W. Drexler, and J. G. Fujimoto, eds. (Springer, 2008), 653–695.

39.

M. E. Brezinski, “Optical Coherence Tomography Theory: Appendix 5-1” in Optical Coherence Tomography: Principles and Applications, M. E. Brezinski, ed. (Elsevier, 2006), 141–142.

40.

S. Makita, M. Yamanari, and Y. Yasuno, “Generalized Jones matrix optical coherence tomography: performance and local birefringence imaging,” Opt. Express 18(2), 854–876 (2010). [CrossRef] [PubMed]

41.

S. Jiao and L. V. Wang, “Jones-matrix imaging of biological tissues with quadruple-channel optical coherence tomography,” J. Biomed. Opt. 7(3), 350–358 (2002). [CrossRef] [PubMed]

42.

R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, “Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography,” Opt. Lett. 25(11), 820–822 (2000). [CrossRef] [PubMed]

43.

G. van Soest, M. Villiger, E. Regar, G. J. Tearney, B. E. Bouma, and A. F. van der Steen, “Frequency domain multiplexing for speckle reduction in optical coherence tomography,” J. Biomed. Opt. 17(7), 076018 (2012). [CrossRef] [PubMed]

44.

D. Hillmann, T. Bonin, C. Lührs, G. Franke, M. Hagen-Eggert, P. Koch, and G. Hüttmann, “Common approach for compensation of axial motion artifacts in swept-source OCT and dispersion in Fourier-domain OCT,” Opt. Express 20(6), 6761–6776 (2012). [CrossRef] [PubMed]

45.

W. Choi, B. Baumann, E. A. Swanson, and J. G. Fujimoto, “Extracting and compensating dispersion mismatch in ultrahigh-resolution Fourier domain OCT imaging of the retina,” Opt. Express 20(23), 25357–25368 (2012). [CrossRef] [PubMed]

46.

Y. Lim, M. Yamanari, S. Fukuda, Y. Kaji, T. Kiuchi, M. Miura, T. Oshika, and Y. Yasuno, “Birefringence measurement of cornea and anterior segment by office-based polarization-sensitive optical coherence tomography,” Biomed. Opt. Express 2(8), 2392–2402 (2011). [CrossRef] [PubMed]

47.

G. D. VanWiggeren and R. Roy, “Transmission of linearly polarized light through a single-mode fiber with random fluctuations of birefringence,” Appl. Opt. 38(18), 3888–3892 (1999). [CrossRef] [PubMed]

48.

S. Jiao, W. Yu, G. Stoica, and L. V. Wang, “Optical-fiber-based Mueller optical coherence tomography,” Opt. Lett. 28(14), 1206–1208 (2003). [CrossRef] [PubMed]

49.

D. Penninckx and V. Morénas, “Jones matrix of polarization mode dispersion,” Opt. Lett. 24(13), 875–877 (1999). [CrossRef] [PubMed]

50.

M. Pircher, E. Götzinger, B. Baumann, and C. K. Hitzenberger, “Corneal birefringence compensation for polarization sensitive optical coherence tomography of the human retina,” J. Biomed. Opt. 12(4), 041210 (2007). [CrossRef] [PubMed]

51.

B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Optic axis determination accuracy for fiber-based polarization-sensitive optical coherence tomography,” Opt. Lett. 30(19), 2587–2589 (2005). [CrossRef] [PubMed]

52.

F. Fanjul-Vélez, M. Pircher, B. Baumann, E. Götzinger, C. K. Hitzenberger, and J. L. Arce-Diego, “Polarimetric analysis of the human cornea measured by polarization-sensitive optical coherence tomography,” J. Biomed. Opt. 15(5), 056004 (2010). [CrossRef] [PubMed]

53.

E. Götzinger, M. Pircher, B. Baumann, T. Schmoll, H. Sattmann, R. A. Leitgeb, and C. K. Hitzenberger, “Speckle noise reduction in high speed polarization sensitive spectral domain optical coherence tomography,” Opt. Express 19(15), 14568–14585 (2011). [CrossRef] [PubMed]

54.

M. Todorović, S. Jiao, L. V. Wang, and G. Stoica, “Determination of local polarization properties of biological samples in the presence of diattenuation by use of Mueller optical coherence tomography,” Opt. Lett. 29(20), 2402–2404 (2004). [CrossRef] [PubMed]

55.

N. Kemp, H. Zaatari, J. Park, H. G. Rylander III, and T. Milner, “Form-biattenuance in fibrous tissues measured with polarization-sensitive optical coherence tomography (PS-OCT),” Opt. Express 13(12), 4611–4628 (2005). [CrossRef] [PubMed]

56.

Q. Zhou, J. Reed, R. Betts, P. Trost, P. Lo, C. Wallace, R. Bienias, G. Li, R. Winnick, W. Papworth, and M. Sinai, “Detection of glaucomatous retinal nerve fiber layer damage by scanning laser polarimetry with variable corneal compensation,” Proc. SPIE 4951, 32–41 (2003). [CrossRef]

57.

I. Grulkowski, J. J. Liu, B. Potsaid, V. Jayaraman, C. D. Lu, J. Jiang, A. E. Cable, J. S. Duker, and J. G. Fujimoto, “Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers,” Biomed. Opt. Express 3(11), 2733–2751 (2012). [CrossRef] [PubMed]

58.

A. Petzold, J. F. de Boer, S. Schippling, P. Vermersch, R. Kardon, A. Green, P. A. Calabresi, and C. Polman, “Optical coherence tomography in multiple sclerosis: a systematic review and meta-analysis,” Lancet Neurol. 9(9), 921–932 (2010). [CrossRef] [PubMed]

59.

M. Sugita, S. Zotter, M. Pircher, T. Makihira, K. Saito, N. Tomatsu, M. Sato, P. Roberts, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Motion artifact and speckle noise reduction in polarization sensitive optical coherence tomography by retinal tracking,” Biomed. Opt. Express 5(1), 106–122 (2014). [CrossRef] [PubMed]

60.

K. V. Vienola, B. Braaf, C. K. Sheehy, Q. Yang, P. Tiruveedhula, D. W. Arathorn, J. F. de Boer, and A. Roorda, “Real-time eye motion compensation for OCT imaging with tracking SLO,” Biomed. Opt. Express 3(11), 2950–2963 (2012). [CrossRef] [PubMed]

61.

B. Braaf, K. V. Vienola, C. K. Sheehy, Q. Yang, K. A. Vermeer, P. Tiruveedhula, D. W. Arathorn, A. Roorda, and J. F. de Boer, “Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO,” Biomed. Opt. Express 4(1), 51–65 (2013). [CrossRef] [PubMed]

OCIS Codes
(110.0110) Imaging systems : Imaging systems
(110.4500) Imaging systems : Optical coherence tomography
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(170.4470) Medical optics and biotechnology : Ophthalmology
(260.5430) Physical optics : Polarization

ToC Category:
Optical Coherence Tomography

History
Original Manuscript: May 28, 2014
Revised Manuscript: July 11, 2014
Manuscript Accepted: July 14, 2014
Published: July 22, 2014

Citation
Boy Braaf, Koenraad A. Vermeer, Mattijs de Groot, Kari V. Vienola, and Johannes F. de Boer, "Fiber-based polarization-sensitive OCT of the human retina with correction of system polarization distortions," Biomed. Opt. Express 5, 2736-2758 (2014)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-5-8-2736


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science254(5035), 1178–1181 (1991). [CrossRef] [PubMed]
  2. W. Drexler and J. G. Fujimoto, “State-of-the-art retinal optical coherence tomography,” Prog. Retin. Eye Res.27(1), 45–88 (2008). [CrossRef] [PubMed]
  3. E. Götzinger, M. Pircher, M. Sticker, A. F. Fercher, and C. K. Hitzenberger, “Measurement and imaging of birefringent properties of the human cornea with phase-resolved, polarization-sensitive optical coherence tomography,” J. Biomed. Opt.9(1), 94–102 (2004). [CrossRef] [PubMed]
  4. B. Cense, T. C. Chen, B. H. Park, M. C. Pierce, and J. F. de Boer, “In vivo depth-resolved birefringence measurements of the human retinal nerve fiber layer by polarization-sensitive optical coherence tomography,” Opt. Lett.27(18), 1610–1612 (2002). [CrossRef] [PubMed]
  5. B. Cense, T. C. Chen, B. H. Park, M. C. Pierce, and J. F. de Boer, “Thickness and birefringence of healthy retinal nerve fiber layer tissue measured with polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.45(8), 2606–2612 (2004). [CrossRef] [PubMed]
  6. B. Cense, Q. Wang, S. Lee, L. Zhao, A. E. Elsner, C. K. Hitzenberger, and D. T. Miller, “Henle fiber layer phase retardation measured with polarization-sensitive optical coherence tomography,” Biomed. Opt. Express4(11), 2296–2306 (2013). [CrossRef] [PubMed]
  7. M. Pircher, E. Götzinger, R. Leitgeb, H. Sattmann, O. Findl, and C. Hitzenberger, “Imaging of polarization properties of human retina in vivo with phase resolved transversal PS-OCT,” Opt. Express12(24), 5940–5951 (2004). [CrossRef] [PubMed]
  8. E. Götzinger, M. Pircher, W. Geitzenauer, C. Ahlers, B. Baumann, S. Michels, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Retinal pigment epithelium segmentation by polarization sensitive optical coherence tomography,” Opt. Express16(21), 16410–16422 (2008). [CrossRef] [PubMed]
  9. M. Pircher, E. Götzinger, O. Findl, S. Michels, W. Geitzenauer, C. Leydolt, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Human macula investigated in vivo with polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.47(12), 5487–5494 (2006). [CrossRef] [PubMed]
  10. M. Yamanari, Y. Lim, S. Makita, and Y. Yasuno, “Visualization of phase retardation of deep posterior eye by polarization-sensitive swept-source optical coherence tomography with 1- µm probe,” Opt. Express17(15), 12385–12396 (2009). [CrossRef] [PubMed]
  11. B. Baumann, E. Götzinger, M. Pircher, and C. K. Hitzenberger, “Single camera based spectral domain polarization sensitive optical coherence tomography,” Opt. Express15(3), 1054–1063 (2007). [CrossRef] [PubMed]
  12. M. R. Hee, D. Huang, E. A. Swanson, and J. G. Fujimoto, “Polarization-sensitive low-coherence reflectometer for birefringence characterization and ranging,” J. Opt. Soc. Am. B9(6), 903–908 (1992). [CrossRef]
  13. J. F. de Boer, T. E. Milner, M. J. van Gemert, and J. S. Nelson, “Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography,” Opt. Lett.22(12), 934–936 (1997). [CrossRef] [PubMed]
  14. M. Pircher, C. K. Hitzenberger, and U. Schmidt-Erfurth, “Polarization sensitive optical coherence tomography in the human eye,” Prog. Retin. Eye Res.30(6), 431–451 (2011). [CrossRef] [PubMed]
  15. F. G. Schlanitz, B. Baumann, T. Spalek, C. Schütze, C. Ahlers, M. Pircher, E. Götzinger, C. K. Hitzenberger, and U. Schmidt-Erfurth, “Performance of automated drusen detection by polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.52(7), 4571–4579 (2011). [CrossRef] [PubMed]
  16. E. Götzinger, M. Pircher, B. Baumann, C. Ahlers, W. Geitzenauer, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Three-dimensional polarization sensitive OCT imaging and interactive display of the human retina,” Opt. Express17(5), 4151–4165 (2009). [CrossRef] [PubMed]
  17. S. Zotter, M. Pircher, T. Torzicky, B. Baumann, H. Yoshida, F. Hirose, P. Roberts, M. Ritter, C. Schütze, E. Götzinger, W. Trasischker, C. Vass, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Large-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology,” Biomed. Opt. Express3(11), 2720–2732 (2012). [CrossRef] [PubMed]
  18. J. Dwelle, S. Liu, B. Wang, A. McElroy, D. Ho, M. K. Markey, T. Milner, and H. G. Rylander, “Thickness, phase retardation, birefringence, and reflectance of the retinal nerve fiber layer in normal and glaucomatous non-human primates,” Invest. Ophthalmol. Vis. Sci.53(8), 4380–4395 (2012). [CrossRef] [PubMed]
  19. E. Götzinger, M. Pircher, B. Baumann, C. Hirn, C. Vass, and C. K. Hitzenberger, “Analysis of the origin of atypical scanning laser polarimetry patterns by polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.49(12), 5366–5372 (2008). [CrossRef] [PubMed]
  20. C. Ahlers, E. Götzinger, M. Pircher, I. Golbaz, F. Prager, C. Schütze, B. Baumann, C. K. Hitzenberger, and U. Schmidt-Erfurth, “Imaging of the retinal pigment epithelium in age-related macular degeneration using polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.51(4), 2149–2157 (2010). [CrossRef] [PubMed]
  21. C. E. Saxer, J. F. de Boer, B. H. Park, Y. Zhao, Z. Chen, and J. S. Nelson, “High-speed fiber based polarization-sensitive optical coherence tomography of in vivo human skin,” Opt. Lett.25(18), 1355–1357 (2000). [CrossRef] [PubMed]
  22. B. H. Park, C. Saxer, S. M. Srinivas, J. S. Nelson, and J. F. de Boer, “In vivo burn depth determination by high-speed fiber-based polarization sensitive optical coherence tomography,” J. Biomed. Opt.6(4), 474–479 (2001). [CrossRef] [PubMed]
  23. W. Y. Oh, B. J. Vakoc, S. H. Yun, G. J. Tearney, and B. E. Bouma, “Single-detector polarization-sensitive optical frequency domain imaging using high-speed intra A-line polarization modulation,” Opt. Lett.33(12), 1330–1332 (2008). [CrossRef] [PubMed]
  24. W. Y. Oh, S. H. Yun, B. J. Vakoc, M. Shishkov, A. E. Desjardins, B. H. Park, J. F. de Boer, G. J. Tearney, and B. E. Bouma, “High-speed polarization sensitive optical frequency domain imaging with frequency multiplexing,” Opt. Express16(2), 1096–1103 (2008). [CrossRef] [PubMed]
  25. M. Yamanari, S. Makita, and Y. Yasuno, “Polarization-sensitive swept-source optical coherence tomography with continuous source polarization modulation,” Opt. Express16(8), 5892–5906 (2008). [CrossRef] [PubMed]
  26. K. H. Kim, B. H. Park, Y. Tu, T. Hasan, B. Lee, J. Li, and J. F. de Boer, “Polarization-sensitive optical frequency domain imaging based on unpolarized light,” Opt. Express19(2), 552–561 (2011). [CrossRef] [PubMed]
  27. Y. Lim, Y. J. Hong, L. Duan, M. Yamanari, and Y. Yasuno, “Passive component based multifunctional Jones matrix swept source optical coherence tomography for Doppler and polarization imaging,” Opt. Lett.37(11), 1958–1960 (2012). [CrossRef] [PubMed]
  28. B. Baumann, W. Choi, B. Potsaid, D. Huang, J. S. Duker, and J. G. Fujimoto, “Swept source/Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unit,” Opt. Express20(9), 10229–10241 (2012). [CrossRef] [PubMed]
  29. B. Elmaanaoui, B. Wang, J. C. Dwelle, A. B. McElroy, S. S. Liu, H. G. Rylander, and T. E. Milner, “Birefringence measurement of the retinal nerve fiber layer by swept source polarization sensitive optical coherence tomography,” Opt. Express19(11), 10252–10268 (2011). [CrossRef] [PubMed]
  30. M. Villiger, E. Z. Zhang, S. Nadkarni, W. Y. Oh, B. E. Bouma, and B. J. Vakoc, “Artifacts in polarization-sensitive optical coherence tomography caused by polarization mode dispersion,” Opt. Lett.38(6), 923–925 (2013). [CrossRef] [PubMed]
  31. E. Z. Zhang, W. Y. Oh, M. L. Villiger, L. Chen, B. E. Bouma, and B. J. Vakoc, “Numerical compensation of system polarization mode dispersion in polarization-sensitive optical coherence tomography,” Opt. Express21(1), 1163–1180 (2013). [CrossRef] [PubMed]
  32. M. Villiger, E. Z. Zhang, S. K. Nadkarni, W. Y. Oh, B. J. Vakoc, and B. E. Bouma, “Spectral binning for mitigation of polarization mode dispersion artifacts in catheter-based optical frequency domain imaging,” Opt. Express21(14), 16353–16369 (2013). [CrossRef] [PubMed]
  33. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Jones matrix analysis for a polarization-sensitive optical coherence tomography system using fiber-optic components,” Opt. Lett.29(21), 2512–2514 (2004). [CrossRef] [PubMed]
  34. M. J. Ju, Y. J. Hong, S. Makita, Y. Lim, K. Kurokawa, L. Duan, M. Miura, S. Tang, and Y. Yasuno, “Advanced multi-contrast Jones matrix optical coherence tomography for Doppler and polarization sensitive imaging,” Opt. Express21(16), 19412–19436 (2013). [CrossRef] [PubMed]
  35. B. Braaf, K. A. Vermeer, V. A. Sicam, E. van Zeeburg, J. C. van Meurs, and J. F. de Boer, “Phase-stabilized optical frequency domain imaging at 1-µm for the measurement of blood flow in the human choroid,” Opt. Express19(21), 20886–20903 (2011). [CrossRef] [PubMed]
  36. N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express12(3), 367–376 (2004). [CrossRef] [PubMed]
  37. N. V. Iftimia, D. X. Hammer, C. E. Bigelow, D. I. Rosen, T. Ustun, A. A. Ferrante, D. Vu, and R. D. Ferguson, “Toward noninvasive measurement of blood hematocrit using spectral domain low coherence interferometry and retinal tracking,” Opt. Express14(8), 3377–3388 (2006). [CrossRef] [PubMed]
  38. B. H. Park and J. F. De Boer, “Polarization-Sensitive Optical Coherence Tomography” in Optical Coherence Tomography: Technology and Applications, W. Drexler, and J. G. Fujimoto, eds. (Springer, 2008), 653–695.
  39. M. E. Brezinski, “Optical Coherence Tomography Theory: Appendix 5-1” in Optical Coherence Tomography: Principles and Applications, M. E. Brezinski, ed. (Elsevier, 2006), 141–142.
  40. S. Makita, M. Yamanari, and Y. Yasuno, “Generalized Jones matrix optical coherence tomography: performance and local birefringence imaging,” Opt. Express18(2), 854–876 (2010). [CrossRef] [PubMed]
  41. S. Jiao and L. V. Wang, “Jones-matrix imaging of biological tissues with quadruple-channel optical coherence tomography,” J. Biomed. Opt.7(3), 350–358 (2002). [CrossRef] [PubMed]
  42. R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, “Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography,” Opt. Lett.25(11), 820–822 (2000). [CrossRef] [PubMed]
  43. G. van Soest, M. Villiger, E. Regar, G. J. Tearney, B. E. Bouma, and A. F. van der Steen, “Frequency domain multiplexing for speckle reduction in optical coherence tomography,” J. Biomed. Opt.17(7), 076018 (2012). [CrossRef] [PubMed]
  44. D. Hillmann, T. Bonin, C. Lührs, G. Franke, M. Hagen-Eggert, P. Koch, and G. Hüttmann, “Common approach for compensation of axial motion artifacts in swept-source OCT and dispersion in Fourier-domain OCT,” Opt. Express20(6), 6761–6776 (2012). [CrossRef] [PubMed]
  45. W. Choi, B. Baumann, E. A. Swanson, and J. G. Fujimoto, “Extracting and compensating dispersion mismatch in ultrahigh-resolution Fourier domain OCT imaging of the retina,” Opt. Express20(23), 25357–25368 (2012). [CrossRef] [PubMed]
  46. Y. Lim, M. Yamanari, S. Fukuda, Y. Kaji, T. Kiuchi, M. Miura, T. Oshika, and Y. Yasuno, “Birefringence measurement of cornea and anterior segment by office-based polarization-sensitive optical coherence tomography,” Biomed. Opt. Express2(8), 2392–2402 (2011). [CrossRef] [PubMed]
  47. G. D. VanWiggeren and R. Roy, “Transmission of linearly polarized light through a single-mode fiber with random fluctuations of birefringence,” Appl. Opt.38(18), 3888–3892 (1999). [CrossRef] [PubMed]
  48. S. Jiao, W. Yu, G. Stoica, and L. V. Wang, “Optical-fiber-based Mueller optical coherence tomography,” Opt. Lett.28(14), 1206–1208 (2003). [CrossRef] [PubMed]
  49. D. Penninckx and V. Morénas, “Jones matrix of polarization mode dispersion,” Opt. Lett.24(13), 875–877 (1999). [CrossRef] [PubMed]
  50. M. Pircher, E. Götzinger, B. Baumann, and C. K. Hitzenberger, “Corneal birefringence compensation for polarization sensitive optical coherence tomography of the human retina,” J. Biomed. Opt.12(4), 041210 (2007). [CrossRef] [PubMed]
  51. B. H. Park, M. C. Pierce, B. Cense, and J. F. de Boer, “Optic axis determination accuracy for fiber-based polarization-sensitive optical coherence tomography,” Opt. Lett.30(19), 2587–2589 (2005). [CrossRef] [PubMed]
  52. F. Fanjul-Vélez, M. Pircher, B. Baumann, E. Götzinger, C. K. Hitzenberger, and J. L. Arce-Diego, “Polarimetric analysis of the human cornea measured by polarization-sensitive optical coherence tomography,” J. Biomed. Opt.15(5), 056004 (2010). [CrossRef] [PubMed]
  53. E. Götzinger, M. Pircher, B. Baumann, T. Schmoll, H. Sattmann, R. A. Leitgeb, and C. K. Hitzenberger, “Speckle noise reduction in high speed polarization sensitive spectral domain optical coherence tomography,” Opt. Express19(15), 14568–14585 (2011). [CrossRef] [PubMed]
  54. M. Todorović, S. Jiao, L. V. Wang, and G. Stoica, “Determination of local polarization properties of biological samples in the presence of diattenuation by use of Mueller optical coherence tomography,” Opt. Lett.29(20), 2402–2404 (2004). [CrossRef] [PubMed]
  55. N. Kemp, H. Zaatari, J. Park, H. G. Rylander, and T. Milner, “Form-biattenuance in fibrous tissues measured with polarization-sensitive optical coherence tomography (PS-OCT),” Opt. Express13(12), 4611–4628 (2005). [CrossRef] [PubMed]
  56. Q. Zhou, J. Reed, R. Betts, P. Trost, P. Lo, C. Wallace, R. Bienias, G. Li, R. Winnick, W. Papworth, and M. Sinai, “Detection of glaucomatous retinal nerve fiber layer damage by scanning laser polarimetry with variable corneal compensation,” Proc. SPIE4951, 32–41 (2003). [CrossRef]
  57. I. Grulkowski, J. J. Liu, B. Potsaid, V. Jayaraman, C. D. Lu, J. Jiang, A. E. Cable, J. S. Duker, and J. G. Fujimoto, “Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers,” Biomed. Opt. Express3(11), 2733–2751 (2012). [CrossRef] [PubMed]
  58. A. Petzold, J. F. de Boer, S. Schippling, P. Vermersch, R. Kardon, A. Green, P. A. Calabresi, and C. Polman, “Optical coherence tomography in multiple sclerosis: a systematic review and meta-analysis,” Lancet Neurol.9(9), 921–932 (2010). [CrossRef] [PubMed]
  59. M. Sugita, S. Zotter, M. Pircher, T. Makihira, K. Saito, N. Tomatsu, M. Sato, P. Roberts, U. Schmidt-Erfurth, and C. K. Hitzenberger, “Motion artifact and speckle noise reduction in polarization sensitive optical coherence tomography by retinal tracking,” Biomed. Opt. Express5(1), 106–122 (2014). [CrossRef] [PubMed]
  60. K. V. Vienola, B. Braaf, C. K. Sheehy, Q. Yang, P. Tiruveedhula, D. W. Arathorn, J. F. de Boer, and A. Roorda, “Real-time eye motion compensation for OCT imaging with tracking SLO,” Biomed. Opt. Express3(11), 2950–2963 (2012). [CrossRef] [PubMed]
  61. B. Braaf, K. V. Vienola, C. K. Sheehy, Q. Yang, K. A. Vermeer, P. Tiruveedhula, D. W. Arathorn, A. Roorda, and J. F. de Boer, “Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO,” Biomed. Opt. Express4(1), 51–65 (2013). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.