Automatic dynamic tear meniscus measurement in optical coherence tomography

Maciej M. Bartuzel, Dorota H. Szczesna-Iskander, and D. Robert Iskander  »View Author Affiliations


Biomedical Optics Express, Vol. 5, Issue 8, pp. 2759-2768 (2014)
http://dx.doi.org/10.1364/BOE.5.002759


View Full Text Article

Acrobat PDF (3604 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

An image processing algorithm is developed for quantitative assessment of tear meniscus dynamics from continuous optical coherence tomography (OCT) measurements. Clinical utility of dynamic OCT tear meniscus measurement is assessed in studies of tear meniscus parameters. The results indicate that any apparent changes in the early post-blink phase meniscus parameters are essentially related to the longitudinal movements of the eye and not to the formation of tear meniscus corresponding to tear film build-up. Dynamic acquisition of tear film meniscus is essential for providing reliable estimates of its parameters such as height, depth, and area.

© 2014 Optical Society of America

1. Introduction

Tear film is important to the health and optics of the human eye. Measurement of tear film properties provides clinicians with investigative capabilities in the evaluation of ocular surface health [1

1. C. S. de Paiva and S. C. Pflugfelder, “Tear clearance implications for ocular surface health,” Exp. Eye Res. 78(3), 395–397 (2004). [CrossRef] [PubMed]

, 2

2. S. C. Tseng and K. Tsubota, “Important concepts for treating ocular surface and tear disorders,” Am. J. Ophthalmol. 124(6), 825–835 (1997). [PubMed]

], dry eye diagnosis [3

3. D. R. Korb, “Survey of preferred tests for diagnosis of the tear film and dry eye,” Cornea 19(4), 483–486 (2000). [CrossRef] [PubMed]

5

5. S. Khanal, A. Tomlinson, A. McFadyen, C. Diaper, and K. Ramaesh, “Dry eye diagnosis,” Invest. Ophthalmol. Vis. Sci. 49(4), 1407–1414 (2008). [CrossRef] [PubMed]

], or the assessment of contact lens wear [6

6. J. J. Nichols and L. T. Sinnott, “Tear film, contact lens, and patient-related factors associated with contact lens-related dry eye,” Invest. Ophthalmol. Vis. Sci. 47(4), 1319–1328 (2006). [CrossRef] [PubMed]

, 7

7. D. H. Szczesna-Iskander, D. R. Iskander, S. A. Read, and D. Alonso-Caneiro, “Noninvasive in vivo assessment of soft contact lens type on tear film surface quality,” Invest. Ophthalmol. Vis. Sci. 53(1), 525–531 (2012). [CrossRef] [PubMed]

].

Depending on the approach taken, tear film measurement techniques can be divided into invasive and noninvasive [8

8. N. Yokoi and A. Komuro, “Non-invasive methods of assessing the tear film,” Exp. Eye Res. 78(3), 399–407 (2004). [CrossRef] [PubMed]

, 9

9. D. H. Szczesna-Iskander and D. R. Iskander, “Future directions in non-invasive measurements of tear film surface kinetics,” Optom. Vis. Sci. 89(5), 749–759 (2012). [CrossRef] [PubMed]

], subjective and objective [10

10. S. C. Pflugfelder, S. C. Tseng, O. Sanabria, H. Kell, C. G. Garcia, C. Felix, W. Feuer, and B. L. Reis, “Evaluation of subjective assessments and objective diagnostic tests for diagnosing tear-film disorders known to cause ocular irritation,” Cornea 17(1), 38–56 (1998). [CrossRef] [PubMed]

], and into static and dynamic [11

11. D. R. Iskander and M. J. Collins, “Applications of high-speed videokeratoscopy,” Clin. Exp. Optom. 88(4), 223–231 (2005). [CrossRef] [PubMed]

]. Measurement of tear meniscus parameters such as its height, depth, area, and calculation of its volume can be used for assisting dry eye diagnosis [12

12. M. Shen, J. Li, J. Wang, H. Ma, C. Cai, A. Tao, Y. Yuan, and F. Lu, “Upper and lower tear menisci in the diagnosis of dry eye,” Invest. Ophthalmol. Vis. Sci. 50(6), 2722–2726 (2009). [CrossRef] [PubMed]

], assessing the efficacy of artificial tear supplements [13

13. J. Wang, P. Simmons, J. Aquavella, J. Vehige, J. Palakuru, S. Chung, and C. Feng, “Dynamic distribution of artificial tears on the ocular surface,” Arch. Ophthalmol. 126(5), 619–625 (2008). [CrossRef] [PubMed]

], evaluating the effects of contact lens wear [14

14. Q. Chen, J. Wang, M. Shen, C. Cai, J. Li, L. Cui, J. Qu, and F. Lu, “Lower volumes of tear menisci in contact lens wearers with dry eye symptoms,” Invest. Ophthalmol. Vis. Sci. 50(7), 3159–3163 (2009). [CrossRef] [PubMed]

, 15

15. Q. Chen, J. Wang, A. Tao, M. Shen, S. Jiao, and F. Lu, “Ultrahigh-resolution measurement by optical coherence tomography of dynamic tear film changes on contact lenses,” Invest. Ophthalmol. Vis. Sci. 51(4), 1988–1993 (2010). [CrossRef] [PubMed]

], and facilitating, using mathematical models, estimation of tear film thickness [16

16. J. L. Creech, L. T. Do, I. Fatt, and C. J. Radke, “In vivo tear-film thickness determination and implications for tear-film stability,” Curr. Eye Res. 17(11), 1058–1066 (1998). [CrossRef] [PubMed]

, 17

17. T. R. Golding, A. S. Bruce, and J. C. Mainstone, “Relationship between tear-meniscus parameters and tear-film breakup,” Cornea 16(6), 649–661 (1997). [CrossRef] [PubMed]

]. Methods of tear meniscus measurement include en-face slit lamp observation, reflective meniscometry [18

18. H. Oguz, N. Yokoi, and S. Kinoshita, “The height and radius of the tear meniscus and methods for examining these parameters,” Cornea 19(4), 497–500 (2000). [CrossRef] [PubMed]

20

20. S. Bandlitz, C. Purslow, P. J. Murphy, H. Pult, and A. J. Bron, “A new portable digital meniscometer,” Optom. Vis. Sci. 91(1), e1–e8 (2014). [CrossRef] [PubMed]

] and cross-sectional measurement with optical coherence tomography (OCT) [21

21. L. W. Jones, S. Rahman, R. Leech, T. Simpson, D. Fonn, J. G. Vehige, and P. A. Simmons, “Determination of tear meniscus height using a novel method based upon optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 44, 2461 (2003). [PubMed]

24

24. J. Wang, J. Aquavella, J. Palakuru, S. Chung, and C. Feng, “Relationships between central tear film thickness and tear menisci of the upper and lower eyelids,” Invest. Ophthalmol. Vis. Sci. 47(10), 4349–4355 (2006). [CrossRef] [PubMed]

]. The latter belongs to the class of noninvasive methods concerning tear meniscus parameters that has the potential to be conducted objectively and in a dynamic fashion.

The majority of methods for extracting tear meniscus parameters from OCT images require manual input from an operator [23

23. G. Savini, P. Barboni, and M. Zanini, “Tear meniscus evaluation by optical coherence tomography,” Ophthalmic Surg. Lasers Imaging 37(2), 112–118 (2006). [PubMed]

, 25

25. R. Fukuda, T. Usui, T. Miyai, S. Yamagami, and S. Amano, “Tear meniscus evaluation by anterior segment swept-source optical coherence tomography,” Am. J. Ophthalmol. 155(4), 620–624 (2013). [CrossRef] [PubMed]

]. To the best of our knowledge the only method that was able to perform the task of extracting tear meniscus parameters in an automatic fashion was proposed by Tsechpenakis and Wang [26

26. G. Tsechpenakis and J. Wang, “CRF-based segmentation of human tear meniscus obtained with optical coherence tomography” in Proceedings of IEEE International Conference on Image Processing (Institute of Electrical and Electronics Engineers, New York, 2007), pp. V509–V512. [CrossRef]

]. The method, based on conditional random field theory, appears to be accurate but the result presented in that work was for a static single measurement and there was no indication how the method performs for a set of dynamic OCT images acquired immediately after a blink. As the authors conclude, an automatic quantitative estimation of the tear menisci is a challenging task.

The repeatability of OCT based tear meniscus measurement has been considered but it was hindered by the necessity of the manual input from an operator. Johnson and Murphy [27

27. M. E. Johnson and P. J. Murphy, “The agreement and repeatability of tear meniscus height measurement methods,” Optom. Vis. Sci. 82(12), 1030–1037 (2005). [CrossRef] [PubMed]

] studied the tear meniscus height using two static OCT measurements on two different days. Wang and colleagues [28

28. J. Wang, J. Aquavella, J. Palakuru, and S. Chung, “Repeated measurements of dynamic tear distribution on the ocular surface after instillation of artificial tears,” Invest. Ophthalmol. Vis. Sci. 47(8), 3325–3329 (2006). [CrossRef] [PubMed]

] used dynamic acquisition of tear meniscus at eight frames per second. However, to evaluate the repeatability of tear meniscus parameters, only static measurements (the first good image in each acquisition) were taken into consideration. Bitton and colleagues [29

29. E. Bitton, A. Keech, T. Simpson, and L. Jones, “Variability of the analysis of the tear meniscus height by optical coherence tomography,” Optom. Vis. Sci. 84(9), 903–908 (2007). [CrossRef] [PubMed]

] assessed the variability of tear meniscus height using 10 consecutive scans using two graders but they had to reevaluate their results after a training session among the observers. Hence, manual estimation of the tear meniscus parameters is subjective and very laborious.

The dynamics of OCT based tear meniscus measurement has not been fully investigated particularly in the sense of tear meniscus formation as well as the reliability of a single post-blink static acquisition. Measurement of tear meniscus dynamics can be conducted with commercially available instrumentation. Many OCT systems have the capability of performing multiple B-scan measurements at the standard video rate of 24 frames per second. At that sampling frequency, an average blink of 250 ms is confounded to about six frames [30

30. M. J. Doughty, “Consideration of three types of spontaneous eyeblink activity in normal humans: during reading and video display terminal use, in primary gaze, and while in conversation,” Optom. Vis. Sci. 78(10), 712–725 (2001). [CrossRef] [PubMed]

] while the tear film build-up phase interlocked with further slower post-blink movement of the upper eyelid may take additional 12 frames [31

31. J. Németh, B. Erdélyi, B. Csákány, P. Gáspár, A. Soumelidis, F. Kahlesz, and Z. Lang, “High-speed videotopographic measurement of tear film build-up time,” Invest. Ophthalmol. Vis. Sci. 43(6), 1783–1790 (2002). [PubMed]

, 32

32. D. H. Szczesna and D. R. Iskander, “Lateral shearing interferometry for analysis of tear film surface kinetics,” Optom. Vis. Sci. 87(7), 513–517 (2010). [PubMed]

].

Until now, there were no automatic techniques that could facilitate quantitative assessment of tear meniscus dynamics with OCT. In this paper, we develop an automatic image processing procedure for dynamic OCT based tear meniscus measurement and assess its clinical utility in studies of tear meniscus parameters.

2. Material and methods

2.1 Measurement protocol

B-scans of the inferior tear meniscus (see Fig. 1
Fig. 1 An example of B-scan of the inferior meniscus taken from a series of post-blink measurements. The position of the B-scan with respect to the eye (left) and the corresponding raw OCT image (right).
) were obtained using commercially available Spectral Optical Coherence Tomography instrument – SOCT Copernicus HR (Optopol, Poland). The instrument offers several scanning protocols for the anterior eye segment, from which the animation mode was chosen. This protocol allows observing the dynamics of tear meniscus. Setting examination parameters include the scanning width, number of A-scans per B-scan, the total number of B-scans and the scanning angle. The scanning angle was set to 90 degrees. As the region of interest is relatively small, the scanning width was set to the minimum available value of 4 mm. This parameter does not affect other parameters. The maximum number of A-scans per B-scan in this mode is 1800 and it determines the highest possible image quality. The maximum possible number of B-scans, up to which the number of A-scans did not decrease, was set to 90.

Each subject was asked to place his chin on the chin rest and lean his forehead on the headrest. To measure the inferior tear meniscus, the subject was asked to look straight ahead (with the help of the fellow eye and additionally placed mirrors on the sides of the instrument) and refrain from any head or eye movements. The orientation of the B-scan plane was central (with respect to the iris outline) to the posterior region of the eye-eyelid junction and normal to the eyelid.

2.2 Image processing procedure

The SOCT Copernicus HR is primarily designed for posterior segment imaging. The ability to image an anterior segment is achieved with an additional corrective lens module that is attached to the head of the instrument. The acquired images of the anterior segment are therefore distorted. Our aim was to design an image processing methodology for extracting information on tear meniscus from the set of dynamically acquired anterior segment images that is general and not restricted to a particular commercially available instrument. Hence, all the parameters of the tear meniscus estimated through the procedure will be given in pixels rather than in their corresponding physical dimensions.

The size of single B-scan is 900x1009 pixels. The region of interest (ROI) is extracted from the original image, in every iteration starting form the last scan, and it has an empirically chosen size of 351 × 201 pixels. The normalized correlation function is used to track the ROI position from a scan to a scan. Since the OCT images contain substantial amount of noise in both background and imaged structures that change from an image to image, applying of-the-shelf image processing techniques did not result in a successful estimation of the tear meniscus parameters.

The proposed tear meniscus parameter estimation algorithm is depicted in Fig. 2
Fig. 2 The schematics of the proposed tear meniscus parameter estimation algorithm.
. Using an image with ROI, the first step of the algorithm is generation of a binary mask that separates unwanted interference and meniscus profile information from the image components containing eye and eyelid structure. This mask is then used in the algorithm that estimates the Eye-Eyelid Edge (EEE) while its complement is used for estimating the Tear Meniscus Profile (TMP). Information on EEE and TMP is then merged to produce estimates of tear meniscus area (TMA), tear meniscus depth (TMD), and tear meniscus height (TMH).

Mask generation is realized through three important steps (see Fig. 3
Fig. 3 The three important steps of mask generation. Filtering (A), multi-threshold segmentation (B), and class merging with post-processing (C).
): filtering, multi-threshold segmentation, and class merging. Filtering is performed by convolving the original image with a circular mask of empirically chosen radius of two. Multi-threshold segmentation is performed with Otsu method for five classes [33

33. N. Otsu, “A threshold selection method from gray-level histograms,” IEEE Trans. Syst. Man Cybern. 9(1), 62–66 (1979). [CrossRef]

]. Class merging transforms the five level output of the Otsu’s algorithm, pre-filtered with a median filter that ensures discontinuities in the meniscus profile, to a binary image, in which the last four classes are combined. Finally, an addition of a U-frame and the flood-fill operation is performed. This particular procedure of the binary mask was dictated by the statistics and the structural content of the ROI image, where the same intensities and similar structures were present in both unwanted and useful signal areas. The smallest number of classes in the multi-thresholding segmentation that would successfully separate the two areas was sought and empirically set to five. Most of the unwanted interference was then confound to class 1, while the useful signal was distributed among classes 2 to 5. Hence, merging of the latter resulted in a binary mask that was further employed for EEE and TMP extraction.

The binary mask was directly used to extract EEE by column-wise searching for the last zero-value pixel corresponding to the estimate of the edge at that location. The edge pixels were divided into two parts representing the eye edge and the eyelid edge. To provide a smooth estimate of the EEE, both parts were separately fitted by the fifth and the second order polynomial expansions, respectively. Finally they were both merged again to create the EEE (see Fig. 4
Fig. 4 An example of extracted and estimated EEE (left) and its position within the ROI image (right).
).

To extract TMP information, the ROI image is first multiplied by an eroded complement of the binary mask. The erosion was performed to rule out high intensity pixels (i.e., those with the intensity comparable to that of TMP pixels) possibly protruding from the eye and eyelid structure to the mask area. Further, similarly to the steps employed for the mask generation itself, average filtering is performed on the image in order to strengthen useful information and suppress the unwanted interference. This is followed by Otsu segmentation to separate TMP pixels from the noise. Since the intensity statistics in this area of ROI substantially change from measurement to measurement, the number of classes in Otsu method needs to be selected conditionally. Given the particular OCT images used in study, the area in which the possible TMP pixels could be found is estimated to contain no more than 200 pixels. Thus if the number of pixels in the highest Otsu class is greater than 200, the number of classes is increased by one and the condition is checked again. Empirically it was observed that the class number properly extracting the TMP information was never less than three. Figure 5
Fig. 5 An example of extracted and estimated tear meniscus profile.
shows a representative example of extracted TMP information (denoted by gray dots).

The TMP information, in terms of pixels is then inspected for outliers. Further, for each horizontal pixel value, a median of the corresponding vertical pixel values was calculated. This ensured a unique x to y characterization of the TMP that could be parametrically modeled, for example with an even order polynomial expansion. Since many measurements do not contain meniscus information at the junctions with eye and eyelid surfaces, and to avoid extrapolation, additional boundary points from the EEE are taken before TMP fitting is performed. Those boundary points were set to fulfill the condition of YEEE < min(YTMP). It is worth noting that a polynomial model is arbitrary. It is likely that the water surface of the TMP at the contact with the tissue would be tangent. Hence, a circle tangent to the EEE would probably be the model of choice in case the images were appropriately corrected for distortions.

The final step of the image processing procedure consists of merging the fitted EEE and TMP curves and estimating the position of meniscus-eyelid junction (point A, see Fig. 2), meniscus-eye junction (point C), and the eye-eyelid junction (point B). Additionally, point D is estimated using a bisector of the eye-eyelid angle. Estimated positions of points A, B, C, and D are then used to estimate the following tear meniscus parameters:

  • TMA – defined as a number of pixels contained within the ABC area enclosed by the EEE and TMP,
  • TMD – defined as the distance between points B and D,
  • TMH – defined as the distance between points A and C.

The image processing procedure is repeated for all post-blink scans and results in time-varying estimates of TMA, TMD, and TMH.

3. Results with discussion

In general, time-varying estimates of TMA, TMD, and TMH revealed little variation in the post-blink meniscus dynamics. An example in which the post-blink increase in tear meniscus parameters is clearly visible is shown in Fig. 6
Fig. 6 An example of automatically extracted parameters of tear film meniscus dynamics. Dashed vertical lines indicate the first resolvable post-blink scan.
. In particular, it was difficult to regularly capture the phase of tear meniscus formation. Nevertheless, the average values of TMA, TMD, and TMH were calculated over about two and half second period in which the first 0.5 second was omitted to account for any tear film build-up and post-blink longitudinal movement of the eye. This corresponded to 50 scans that were taken from each series into consideration.

The subject average (over multiple measurements) of TMA, TMD, and TMH are shown in Fig. 7
Fig. 7 The average over multiple measurements of the TMA (square root), TMD, and TMH for each of the considered subjects (A to J).
. To keep all the results within the same scale and units of pixels, square root of TMA is shown. Error bars denote one standard deviation. The result of the Levene’s variance test (p < 0.01) prohibited the use of standard ANOVA. The Kruskal Wallis revealed statistically significant differences (p<0.001) between the subjects, indicating that the OCT based tear meniscus measurement is able to relatively precisely characterize tear meniscus parameters of an individual.

To ascertain the source of variability (or lack of) in the post-blink phase of tear film meniscus, an experienced subject was asked to exercise regular tiny eye twitching during the measurement. Twitching was evident in slight movement of the upper and lower eyelid and corresponded to subject-induced longitudinal eye movements of amplitude much great then than naturally occurring in the eye during steady viewing conditions [34

34. D. Robert Iskander and H. T. Kasprzak, “Dynamics in longitudinal eye movements and corneal shape,” Ophthalmic Physiol. Opt. 26(6), 572–579 (2006). [CrossRef] [PubMed]

]. An example of the results from this exercise is depicted in Fig. 8
Fig. 8 An example of automatically extracted parameters of tear film meniscus dynamics where subject was asked to exercise regular eye twitching. Dashed vertical lines indicate the first resolvable post-blink scan.
. Note that the changes in magnitude of TMA, TMD, and TMH are substantial. However, similar variability but less regular were observed in the earlier standard measurements of tear meniscus parameters.

The proposed algorithm for automatic extraction of tear meniscus parameters from OCT images has some limitations. Figure 9
Fig. 9 Examples of different OCT tear meniscus measurement scans. Ticks and crosses indicate whether the proposed algorithm could or could not successfully process the type of acquired image. Both signs indicate a probabilistic behavior of the algorithm.
shows a collection of different types of OCT tear meniscus measurement acquisitions encountered in the study. In the top row there are two images with complete TMP information, one with discontinued TMP, and two with partial TMP information. Ticks and crosses in each image indicate whether the proposed algorithm could or could not successfully process the particular type of acquisition. Placing both signs in an image indicate a probabilistic behavior of the algorithm resulting in a noisy time series of TMA, TMD, and TMH. In the bottom row of Fig. 9 there are three images with specular reflex (weak and central, weak with small meniscus, and strong and non-central), an image with tear debris, and an image of a subject with irregular scleral topography. The proposed procedure of automatic meniscus extraction could successfully resolve the majority of images with specular reflexes unless it was very strong.

4. Conclusions

A methodology for automatic quantitative extraction of tear meniscus parameters has been proposed. The method performs well for different types of eye-eyelid junction morphology, a wide range of post blink dynamics, and a varied size of tear meniscus. The method is also robust to moderate amounts of interference resulting from the specular reflex.

The results of this study can be easily generalized to the measurements of upper tear meniscus parameters. The procedure of image processing is equally applicable with the difference of flipping the acquired image of the upper tear meniscus about the central vertical axis.

The results of the study indicate that any observed changes in the early post-blink phase meniscus parameters have to be viewed with caution as they are most likely related to the longitudinal movement of the eye [34

34. D. Robert Iskander and H. T. Kasprzak, “Dynamics in longitudinal eye movements and corneal shape,” Ophthalmic Physiol. Opt. 26(6), 572–579 (2006). [CrossRef] [PubMed]

] rather than to the tear meniscus formation immediately after a blink that would correspond to the phase of tear film build-up [32

32. D. H. Szczesna and D. R. Iskander, “Lateral shearing interferometry for analysis of tear film surface kinetics,” Optom. Vis. Sci. 87(7), 513–517 (2010). [PubMed]

, 39

39. N. Yokoi, H. Yamada, Y. Mizukusa, A. J. Bron, J. M. Tiffany, T. Kato, and S. Kinoshita, “Rheology of tear film lipid layer spread in normal and aqueous tear-deficient dry eyes,” Invest. Ophthalmol. Vis. Sci. 49(12), 5319–5324 (2008). [CrossRef] [PubMed]

].

Taking into account the combination of interfering factors related to head, eye and eyelid movements, it is concluded that a single acquisition of tear film meniscus in OCT does not provide reliable estimates of tear meniscus parameters. The clinical utility of dynamic OCT based tear meniscus measurement lies in the necessity of continuous acquisition of post-blink tear film meniscus to provide reliable static estimates of its parameters such as height, depth, and area calculated as averages or medians over a few second post blink interval.

Acknowledgment

Supported by the European Regional Development Fund within Innovative Economy Operational Program co-financed by the Foundation for Polish Science (POMOST/2012-5/8/0072 to DS-I).

References and links

1.

C. S. de Paiva and S. C. Pflugfelder, “Tear clearance implications for ocular surface health,” Exp. Eye Res. 78(3), 395–397 (2004). [CrossRef] [PubMed]

2.

S. C. Tseng and K. Tsubota, “Important concepts for treating ocular surface and tear disorders,” Am. J. Ophthalmol. 124(6), 825–835 (1997). [PubMed]

3.

D. R. Korb, “Survey of preferred tests for diagnosis of the tear film and dry eye,” Cornea 19(4), 483–486 (2000). [CrossRef] [PubMed]

4.

A. J. Bron, “Diagnosis of dry eye,” Surv. Ophthalmol. 45(Suppl 2), S221–S226 (2001). [CrossRef] [PubMed]

5.

S. Khanal, A. Tomlinson, A. McFadyen, C. Diaper, and K. Ramaesh, “Dry eye diagnosis,” Invest. Ophthalmol. Vis. Sci. 49(4), 1407–1414 (2008). [CrossRef] [PubMed]

6.

J. J. Nichols and L. T. Sinnott, “Tear film, contact lens, and patient-related factors associated with contact lens-related dry eye,” Invest. Ophthalmol. Vis. Sci. 47(4), 1319–1328 (2006). [CrossRef] [PubMed]

7.

D. H. Szczesna-Iskander, D. R. Iskander, S. A. Read, and D. Alonso-Caneiro, “Noninvasive in vivo assessment of soft contact lens type on tear film surface quality,” Invest. Ophthalmol. Vis. Sci. 53(1), 525–531 (2012). [CrossRef] [PubMed]

8.

N. Yokoi and A. Komuro, “Non-invasive methods of assessing the tear film,” Exp. Eye Res. 78(3), 399–407 (2004). [CrossRef] [PubMed]

9.

D. H. Szczesna-Iskander and D. R. Iskander, “Future directions in non-invasive measurements of tear film surface kinetics,” Optom. Vis. Sci. 89(5), 749–759 (2012). [CrossRef] [PubMed]

10.

S. C. Pflugfelder, S. C. Tseng, O. Sanabria, H. Kell, C. G. Garcia, C. Felix, W. Feuer, and B. L. Reis, “Evaluation of subjective assessments and objective diagnostic tests for diagnosing tear-film disorders known to cause ocular irritation,” Cornea 17(1), 38–56 (1998). [CrossRef] [PubMed]

11.

D. R. Iskander and M. J. Collins, “Applications of high-speed videokeratoscopy,” Clin. Exp. Optom. 88(4), 223–231 (2005). [CrossRef] [PubMed]

12.

M. Shen, J. Li, J. Wang, H. Ma, C. Cai, A. Tao, Y. Yuan, and F. Lu, “Upper and lower tear menisci in the diagnosis of dry eye,” Invest. Ophthalmol. Vis. Sci. 50(6), 2722–2726 (2009). [CrossRef] [PubMed]

13.

J. Wang, P. Simmons, J. Aquavella, J. Vehige, J. Palakuru, S. Chung, and C. Feng, “Dynamic distribution of artificial tears on the ocular surface,” Arch. Ophthalmol. 126(5), 619–625 (2008). [CrossRef] [PubMed]

14.

Q. Chen, J. Wang, M. Shen, C. Cai, J. Li, L. Cui, J. Qu, and F. Lu, “Lower volumes of tear menisci in contact lens wearers with dry eye symptoms,” Invest. Ophthalmol. Vis. Sci. 50(7), 3159–3163 (2009). [CrossRef] [PubMed]

15.

Q. Chen, J. Wang, A. Tao, M. Shen, S. Jiao, and F. Lu, “Ultrahigh-resolution measurement by optical coherence tomography of dynamic tear film changes on contact lenses,” Invest. Ophthalmol. Vis. Sci. 51(4), 1988–1993 (2010). [CrossRef] [PubMed]

16.

J. L. Creech, L. T. Do, I. Fatt, and C. J. Radke, “In vivo tear-film thickness determination and implications for tear-film stability,” Curr. Eye Res. 17(11), 1058–1066 (1998). [CrossRef] [PubMed]

17.

T. R. Golding, A. S. Bruce, and J. C. Mainstone, “Relationship between tear-meniscus parameters and tear-film breakup,” Cornea 16(6), 649–661 (1997). [CrossRef] [PubMed]

18.

H. Oguz, N. Yokoi, and S. Kinoshita, “The height and radius of the tear meniscus and methods for examining these parameters,” Cornea 19(4), 497–500 (2000). [CrossRef] [PubMed]

19.

N. Yokoi, A. Bron, J. Tiffany, N. Brown, J. Hsuan, and C. Fowler, “Reflective meniscometry: a non-invasive method to measure tear meniscus curvature,” Br. J. Ophthalmol. 83(1), 92–97 (1999). [CrossRef] [PubMed]

20.

S. Bandlitz, C. Purslow, P. J. Murphy, H. Pult, and A. J. Bron, “A new portable digital meniscometer,” Optom. Vis. Sci. 91(1), e1–e8 (2014). [CrossRef] [PubMed]

21.

L. W. Jones, S. Rahman, R. Leech, T. Simpson, D. Fonn, J. G. Vehige, and P. A. Simmons, “Determination of tear meniscus height using a novel method based upon optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 44, 2461 (2003). [PubMed]

22.

L. W. Jones, S. Rahman, R. Leech, T. Simpson, and D. Fonn, “Determination of inferior tear meniscus height and inferior tear meniscus volume using optical coherence tomography,” Invest. Ophthalmol. Vis. Sci. 45, 144 (2004). [PubMed]

23.

G. Savini, P. Barboni, and M. Zanini, “Tear meniscus evaluation by optical coherence tomography,” Ophthalmic Surg. Lasers Imaging 37(2), 112–118 (2006). [PubMed]

24.

J. Wang, J. Aquavella, J. Palakuru, S. Chung, and C. Feng, “Relationships between central tear film thickness and tear menisci of the upper and lower eyelids,” Invest. Ophthalmol. Vis. Sci. 47(10), 4349–4355 (2006). [CrossRef] [PubMed]

25.

R. Fukuda, T. Usui, T. Miyai, S. Yamagami, and S. Amano, “Tear meniscus evaluation by anterior segment swept-source optical coherence tomography,” Am. J. Ophthalmol. 155(4), 620–624 (2013). [CrossRef] [PubMed]

26.

G. Tsechpenakis and J. Wang, “CRF-based segmentation of human tear meniscus obtained with optical coherence tomography” in Proceedings of IEEE International Conference on Image Processing (Institute of Electrical and Electronics Engineers, New York, 2007), pp. V509–V512. [CrossRef]

27.

M. E. Johnson and P. J. Murphy, “The agreement and repeatability of tear meniscus height measurement methods,” Optom. Vis. Sci. 82(12), 1030–1037 (2005). [CrossRef] [PubMed]

28.

J. Wang, J. Aquavella, J. Palakuru, and S. Chung, “Repeated measurements of dynamic tear distribution on the ocular surface after instillation of artificial tears,” Invest. Ophthalmol. Vis. Sci. 47(8), 3325–3329 (2006). [CrossRef] [PubMed]

29.

E. Bitton, A. Keech, T. Simpson, and L. Jones, “Variability of the analysis of the tear meniscus height by optical coherence tomography,” Optom. Vis. Sci. 84(9), 903–908 (2007). [CrossRef] [PubMed]

30.

M. J. Doughty, “Consideration of three types of spontaneous eyeblink activity in normal humans: during reading and video display terminal use, in primary gaze, and while in conversation,” Optom. Vis. Sci. 78(10), 712–725 (2001). [CrossRef] [PubMed]

31.

J. Németh, B. Erdélyi, B. Csákány, P. Gáspár, A. Soumelidis, F. Kahlesz, and Z. Lang, “High-speed videotopographic measurement of tear film build-up time,” Invest. Ophthalmol. Vis. Sci. 43(6), 1783–1790 (2002). [PubMed]

32.

D. H. Szczesna and D. R. Iskander, “Lateral shearing interferometry for analysis of tear film surface kinetics,” Optom. Vis. Sci. 87(7), 513–517 (2010). [PubMed]

33.

N. Otsu, “A threshold selection method from gray-level histograms,” IEEE Trans. Syst. Man Cybern. 9(1), 62–66 (1979). [CrossRef]

34.

D. Robert Iskander and H. T. Kasprzak, “Dynamics in longitudinal eye movements and corneal shape,” Ophthalmic Physiol. Opt. 26(6), 572–579 (2006). [CrossRef] [PubMed]

35.

D. R. Korb, J. V. Greiner, T. Glonek, R. Esbah, V. M. Finnemore, and A. C. Whalen, “Effect of periocular humidity on the tear film lipid layer,” Cornea 15(2), 129–134 (1996). [CrossRef] [PubMed]

36.

M. J. Collins, D. R. Iskander, A. Saunders, S. Hook, E. Anthony, and R. Gillon, “Blinking patterns and corneal staining,” Eye Contact Lens 32(6), 287–293 (2006). [CrossRef] [PubMed]

37.

M. Shen, J. Li, J. Wang, H. Ma, C. Cai, A. Tao, Y. Yuan, and F. Lu, “Upper and lower tear menisci in the diagnosis of dry eye,” Invest. Ophthalmol. Vis. Sci. 50(6), 2722–2726 (2009). [CrossRef] [PubMed]

38.

N. Yokoi, A. J. Bron, J. M. Tiffany, K. Maruyama, A. Komuro, and S. Kinoshita, “Relationship between tear volume and tear meniscus curvature,” Arch. Ophthalmol. 122(9), 1265–1269 (2004). [CrossRef] [PubMed]

39.

N. Yokoi, H. Yamada, Y. Mizukusa, A. J. Bron, J. M. Tiffany, T. Kato, and S. Kinoshita, “Rheology of tear film lipid layer spread in normal and aqueous tear-deficient dry eyes,” Invest. Ophthalmol. Vis. Sci. 49(12), 5319–5324 (2008). [CrossRef] [PubMed]

OCIS Codes
(100.0100) Image processing : Image processing
(100.2960) Image processing : Image analysis
(170.4470) Medical optics and biotechnology : Ophthalmology
(170.4500) Medical optics and biotechnology : Optical coherence tomography

ToC Category:
Ophthalmology Applications

History
Original Manuscript: April 29, 2014
Revised Manuscript: June 26, 2014
Manuscript Accepted: July 6, 2014
Published: July 23, 2014

Citation
Maciej M. Bartuzel, Dorota H. Szczesna-Iskander, and D. Robert Iskander, "Automatic dynamic tear meniscus measurement in optical coherence tomography," Biomed. Opt. Express 5, 2759-2768 (2014)
http://www.opticsinfobase.org/boe/abstract.cfm?URI=boe-5-8-2759


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. C. S. de Paiva and S. C. Pflugfelder, “Tear clearance implications for ocular surface health,” Exp. Eye Res.78(3), 395–397 (2004). [CrossRef] [PubMed]
  2. S. C. Tseng and K. Tsubota, “Important concepts for treating ocular surface and tear disorders,” Am. J. Ophthalmol.124(6), 825–835 (1997). [PubMed]
  3. D. R. Korb, “Survey of preferred tests for diagnosis of the tear film and dry eye,” Cornea19(4), 483–486 (2000). [CrossRef] [PubMed]
  4. A. J. Bron, “Diagnosis of dry eye,” Surv. Ophthalmol.45(Suppl 2), S221–S226 (2001). [CrossRef] [PubMed]
  5. S. Khanal, A. Tomlinson, A. McFadyen, C. Diaper, and K. Ramaesh, “Dry eye diagnosis,” Invest. Ophthalmol. Vis. Sci.49(4), 1407–1414 (2008). [CrossRef] [PubMed]
  6. J. J. Nichols and L. T. Sinnott, “Tear film, contact lens, and patient-related factors associated with contact lens-related dry eye,” Invest. Ophthalmol. Vis. Sci.47(4), 1319–1328 (2006). [CrossRef] [PubMed]
  7. D. H. Szczesna-Iskander, D. R. Iskander, S. A. Read, and D. Alonso-Caneiro, “Noninvasive in vivo assessment of soft contact lens type on tear film surface quality,” Invest. Ophthalmol. Vis. Sci.53(1), 525–531 (2012). [CrossRef] [PubMed]
  8. N. Yokoi and A. Komuro, “Non-invasive methods of assessing the tear film,” Exp. Eye Res.78(3), 399–407 (2004). [CrossRef] [PubMed]
  9. D. H. Szczesna-Iskander and D. R. Iskander, “Future directions in non-invasive measurements of tear film surface kinetics,” Optom. Vis. Sci.89(5), 749–759 (2012). [CrossRef] [PubMed]
  10. S. C. Pflugfelder, S. C. Tseng, O. Sanabria, H. Kell, C. G. Garcia, C. Felix, W. Feuer, and B. L. Reis, “Evaluation of subjective assessments and objective diagnostic tests for diagnosing tear-film disorders known to cause ocular irritation,” Cornea17(1), 38–56 (1998). [CrossRef] [PubMed]
  11. D. R. Iskander and M. J. Collins, “Applications of high-speed videokeratoscopy,” Clin. Exp. Optom.88(4), 223–231 (2005). [CrossRef] [PubMed]
  12. M. Shen, J. Li, J. Wang, H. Ma, C. Cai, A. Tao, Y. Yuan, and F. Lu, “Upper and lower tear menisci in the diagnosis of dry eye,” Invest. Ophthalmol. Vis. Sci.50(6), 2722–2726 (2009). [CrossRef] [PubMed]
  13. J. Wang, P. Simmons, J. Aquavella, J. Vehige, J. Palakuru, S. Chung, and C. Feng, “Dynamic distribution of artificial tears on the ocular surface,” Arch. Ophthalmol.126(5), 619–625 (2008). [CrossRef] [PubMed]
  14. Q. Chen, J. Wang, M. Shen, C. Cai, J. Li, L. Cui, J. Qu, and F. Lu, “Lower volumes of tear menisci in contact lens wearers with dry eye symptoms,” Invest. Ophthalmol. Vis. Sci.50(7), 3159–3163 (2009). [CrossRef] [PubMed]
  15. Q. Chen, J. Wang, A. Tao, M. Shen, S. Jiao, and F. Lu, “Ultrahigh-resolution measurement by optical coherence tomography of dynamic tear film changes on contact lenses,” Invest. Ophthalmol. Vis. Sci.51(4), 1988–1993 (2010). [CrossRef] [PubMed]
  16. J. L. Creech, L. T. Do, I. Fatt, and C. J. Radke, “In vivo tear-film thickness determination and implications for tear-film stability,” Curr. Eye Res.17(11), 1058–1066 (1998). [CrossRef] [PubMed]
  17. T. R. Golding, A. S. Bruce, and J. C. Mainstone, “Relationship between tear-meniscus parameters and tear-film breakup,” Cornea16(6), 649–661 (1997). [CrossRef] [PubMed]
  18. H. Oguz, N. Yokoi, and S. Kinoshita, “The height and radius of the tear meniscus and methods for examining these parameters,” Cornea19(4), 497–500 (2000). [CrossRef] [PubMed]
  19. N. Yokoi, A. Bron, J. Tiffany, N. Brown, J. Hsuan, and C. Fowler, “Reflective meniscometry: a non-invasive method to measure tear meniscus curvature,” Br. J. Ophthalmol.83(1), 92–97 (1999). [CrossRef] [PubMed]
  20. S. Bandlitz, C. Purslow, P. J. Murphy, H. Pult, and A. J. Bron, “A new portable digital meniscometer,” Optom. Vis. Sci.91(1), e1–e8 (2014). [CrossRef] [PubMed]
  21. L. W. Jones, S. Rahman, R. Leech, T. Simpson, D. Fonn, J. G. Vehige, and P. A. Simmons, “Determination of tear meniscus height using a novel method based upon optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.44, 2461 (2003). [PubMed]
  22. L. W. Jones, S. Rahman, R. Leech, T. Simpson, and D. Fonn, “Determination of inferior tear meniscus height and inferior tear meniscus volume using optical coherence tomography,” Invest. Ophthalmol. Vis. Sci.45, 144 (2004). [PubMed]
  23. G. Savini, P. Barboni, and M. Zanini, “Tear meniscus evaluation by optical coherence tomography,” Ophthalmic Surg. Lasers Imaging37(2), 112–118 (2006). [PubMed]
  24. J. Wang, J. Aquavella, J. Palakuru, S. Chung, and C. Feng, “Relationships between central tear film thickness and tear menisci of the upper and lower eyelids,” Invest. Ophthalmol. Vis. Sci.47(10), 4349–4355 (2006). [CrossRef] [PubMed]
  25. R. Fukuda, T. Usui, T. Miyai, S. Yamagami, and S. Amano, “Tear meniscus evaluation by anterior segment swept-source optical coherence tomography,” Am. J. Ophthalmol.155(4), 620–624 (2013). [CrossRef] [PubMed]
  26. G. Tsechpenakis and J. Wang, “CRF-based segmentation of human tear meniscus obtained with optical coherence tomography” in Proceedings of IEEE International Conference on Image Processing (Institute of Electrical and Electronics Engineers, New York, 2007), pp. V509–V512. [CrossRef]
  27. M. E. Johnson and P. J. Murphy, “The agreement and repeatability of tear meniscus height measurement methods,” Optom. Vis. Sci.82(12), 1030–1037 (2005). [CrossRef] [PubMed]
  28. J. Wang, J. Aquavella, J. Palakuru, and S. Chung, “Repeated measurements of dynamic tear distribution on the ocular surface after instillation of artificial tears,” Invest. Ophthalmol. Vis. Sci.47(8), 3325–3329 (2006). [CrossRef] [PubMed]
  29. E. Bitton, A. Keech, T. Simpson, and L. Jones, “Variability of the analysis of the tear meniscus height by optical coherence tomography,” Optom. Vis. Sci.84(9), 903–908 (2007). [CrossRef] [PubMed]
  30. M. J. Doughty, “Consideration of three types of spontaneous eyeblink activity in normal humans: during reading and video display terminal use, in primary gaze, and while in conversation,” Optom. Vis. Sci.78(10), 712–725 (2001). [CrossRef] [PubMed]
  31. J. Németh, B. Erdélyi, B. Csákány, P. Gáspár, A. Soumelidis, F. Kahlesz, and Z. Lang, “High-speed videotopographic measurement of tear film build-up time,” Invest. Ophthalmol. Vis. Sci.43(6), 1783–1790 (2002). [PubMed]
  32. D. H. Szczesna and D. R. Iskander, “Lateral shearing interferometry for analysis of tear film surface kinetics,” Optom. Vis. Sci.87(7), 513–517 (2010). [PubMed]
  33. N. Otsu, “A threshold selection method from gray-level histograms,” IEEE Trans. Syst. Man Cybern.9(1), 62–66 (1979). [CrossRef]
  34. D. Robert Iskander and H. T. Kasprzak, “Dynamics in longitudinal eye movements and corneal shape,” Ophthalmic Physiol. Opt.26(6), 572–579 (2006). [CrossRef] [PubMed]
  35. D. R. Korb, J. V. Greiner, T. Glonek, R. Esbah, V. M. Finnemore, and A. C. Whalen, “Effect of periocular humidity on the tear film lipid layer,” Cornea15(2), 129–134 (1996). [CrossRef] [PubMed]
  36. M. J. Collins, D. R. Iskander, A. Saunders, S. Hook, E. Anthony, and R. Gillon, “Blinking patterns and corneal staining,” Eye Contact Lens32(6), 287–293 (2006). [CrossRef] [PubMed]
  37. M. Shen, J. Li, J. Wang, H. Ma, C. Cai, A. Tao, Y. Yuan, and F. Lu, “Upper and lower tear menisci in the diagnosis of dry eye,” Invest. Ophthalmol. Vis. Sci.50(6), 2722–2726 (2009). [CrossRef] [PubMed]
  38. N. Yokoi, A. J. Bron, J. M. Tiffany, K. Maruyama, A. Komuro, and S. Kinoshita, “Relationship between tear volume and tear meniscus curvature,” Arch. Ophthalmol.122(9), 1265–1269 (2004). [CrossRef] [PubMed]
  39. N. Yokoi, H. Yamada, Y. Mizukusa, A. J. Bron, J. M. Tiffany, T. Kato, and S. Kinoshita, “Rheology of tear film lipid layer spread in normal and aqueous tear-deficient dry eyes,” Invest. Ophthalmol. Vis. Sci.49(12), 5319–5324 (2008). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited