A blackbody and an electrical method of absolute radiometry employing direct current instruments for the measurement of current and resistance are described. The blackbody, designed for performance near 1000°C, is an electrically heated tubular furnace containing an oxidized steel chamber as the radiating source. The radiometer consists, essentially, of a coil of platinum wire wound spirally and bifilar-wise on a blackened, thin, silver disk from which it is electrically insulated but in good thermal contact. The change in resistance of the coil is a measure of the radiation absorbed by the coated disk. By heating the coil electrically, the energy required to produce this change is readily calculated; thus, the deviation of the radiant energy of the source from the corresponding computed perfect blackbody radiation can be determined. This method indicated that the radiant flux emitted by the above source is within 1% of that from such an ideal blackbody at a temperature between 800°C and 1100°C. Included is a description of the apparatus used in the assemblage, as well as of the procedure followed in obtaining data, with a discussion of the probable sources of error.

Citation
MARION EPPLEY and ALTON R. KAROLI, "Absolute Radiometry Based on a Change in Electrical Resistance," J. Opt. Soc. Am. 47, 748-754 (1957)
http://www.opticsinfobase.org/josa/abstract.cfm?URI=josa-47-8-748


Sort:  Author  |  Journal  |  Reset

References

  1. W. W. Coblentz, Bull. Bur. Standards 12, 553 (1915–1916).
  2. F. Kurlbaum, Ann. Physik (3) 65, 746 (1898).
  3. W. Wien and O. Lummer, Ann. Physik (3) 56, 451 (1895).
  4. O. Lummer and F. Kurlbaum, Verhandl. physik. Ges. (Berlin) 17, 106 (1898).
  5. Snyder, Gier, and Dunkle, Trans. Am. Soc. Mech. Engrs. 77, 1011 (1955).
  6. A. Gouffé, Rev. opt. 24, 1 (1945),
  7. H. F. Stimson, J. Research Natl. Bur. Standards 42, 209 (1949).
  8. W. F. Roeser and H. T. Wensel in Temperature—Its Measurement and Control in Science and Industry (Reinhold Publishing Corporation, New York, 1941).
  9. W. W. Coblentz, Bull. Bur. Standards 10, 28 (1914).
  10. W. W. Coblentz and W. B. Emerson, Bull. Bur. Standards 12, 503 (1915–1916).
  11. E. F. Mueller, Bull. Bur. Standards 13, 547 (1916–1917).
  12. M. Eppley, Rev. Sci. Instr. 3, 687 (1932).
  13. A. J. Jaffey, Rev. Sci. Instr. 25, 349–354 (1954).
  14. G. A. W. Rutgers, Physica 17,131 (1951).
  15. R. T. Birge, Phys. Rev. Suppl. 1, 1 (1929).
  16. L. Harris and K. F. Cuff, J. Opt. Soc. Am. 46, 160 (1956).
  17. W. Gerlach, Ann. Physik (4) 50, 233 (1916).
  18. C. Müller, Z. Physik 82, 1 (1933).
  19. J. Guild, Proc. Roy. Soc. (London) A161, 1 (1937).
  20. E. J. Gillham, Brit. J. Appl. Phys. 4, 151 (1953).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited