Radiometric calibration of the IKFS-2 on-board IR Fourier spectrometer for temperature and humidity probing of the Earth’s atmosphere

D. A. Kozlov  »View Author Affiliations


Journal of Optical Technology, Vol. 80, Issue 2, pp. 110-114 (2013)
http://dx.doi.org/10.1364/JOT.80.000110


View Full Text Article

Acrobat PDF (790 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Calibrating the data of a Fourier spectrometer consists of converting the interferograms measured by the device into the spectra of the input radiation with the required radiometric and spectral accuracy. The main problems of terrestrial radiometric calibration of an on-board Fourier spectrometer are to correct the nonlinearity of the photodetector system, to determine the spectral radiance of the radiation of an on-board blackbody, and also to determine the threshold spectral brightness of the device. This article presents the results of the radiometric calibration of a flying sample of the IKFS-2 apparatus, obtained in thermovacuum tests of the device at the Keldysh Center.

© 2013 Optical Society of America

OCIS Codes
(120.6200) Instrumentation, measurement, and metrology : Spectrometers and spectroscopic instrumentation
(300.6190) Spectroscopy : Spectrometers
(300.6300) Spectroscopy : Spectroscopy, Fourier transforms
(300.6340) Spectroscopy : Spectroscopy, infrared

History
Original Manuscript: May 21, 2012
Published: April 9, 2013

Citation
D. A. Kozlov, "Radiometric calibration of the IKFS-2 on-board IR Fourier spectrometer for temperature and humidity probing of the Earth’s atmosphere," J. Opt. Technol. 80, 110-114 (2013)
http://www.opticsinfobase.org/jot/abstract.cfm?URI=jot-80-2-110


Sort:  Author  |  Year  |  Journal  |  Reset

References

  1. A. V. Desyatov, Yu. M. Golovin, and F. S. Zavelevich, “Calibration of the IKFS-2 on-board IR Fourier spectrometer,” Sovremennye problemy distantsionnogo zondirovaniya Zemli iz kosmosa 7, No. 2, 241 (2010).
  2. F. S. Zavelevich, Yu. M. Golovin, and A. V. Desyatov, “Technological sample of IKFS-2 on-board IR Fourier spectrometer,” Sovremennye problemy distantsionnogo zondirovaniya Zemli iz kosmosa 1, 259 (2009).
  3. L. Fiedler, S. Newman, and S. Bakan, “Correction of detector nonlinearity in Fourier transform spectroscopy with a low-temperature blackbody,” Appl. Opt. 4, 5532 (2005).
  4. “Algorithm theoretical basis document for the Cross Track Infrared Sounder (CrIS): Sensor Data Records (SDR),” Release Date: 20.02.2009, p. 199.
  5. “Dossier de définition des algorithmes IASI,” Release Date: 06.01.2011, p. 502.

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited