Effect of air-flow on the evaluation of refractive surgery ablation patterns

Carlos Dorronsoro, Silvia Schumacher, Pablo Pérez-Merino, Jan Siegel, Michael Mrochen, and Susana Marcos  »View Author Affiliations


Optics Express, Vol. 19, Issue 5, pp. 4653-4666 (2011)
http://dx.doi.org/10.1364/OE.19.004653


View Full Text Article

Enhanced HTML    Acrobat PDF (1303 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

An Allegretto Eye-Q laser platform (Wavelight GmbH, Erlangen, Germany) was used to study the effect of air-flow speed on the ablation of artificial polymer corneas used for testing refractive surgery patterns. Flat samples of two materials (PMMA and Filofocon A) were ablated at four different air flow conditions. The shape and profile of the ablated surfaces were measured with a precise non-contact optical surface profilometer. Significant asymmetries in the measured profiles were found when the ablation was performed with the clinical air aspiration system, and also without air flow. Increasing air-flow produced deeper ablations, improved symmetry, and increased the repeatability of the ablation pattern. Shielding of the laser pulse by the plume of smoke during the ablation of plastic samples reduced the central ablation depth by more than 40% with no-air flow, 30% with clinical air aspiration, and 5% with 1.15 m/s air flow. A simple model based on non-inertial dragging of the particles by air flow predicts no central shielding with 2.3 m/s air flow, and accurately predicts (within 2 μm) the decrease of central ablation depth by shielding. The shielding effects for PMMA and Filofocon A were similar despite the differences in the ablation properties of the materials and the different full-shielding transmission coefficient, which is related to the number of particles ejected and their associated optical behavior. Air flow is a key factor in the evaluation of ablation patterns in refractive surgery using plastic models, as significant shielding effects are found with typical air-flow levels used under clinical conditions. Shielding effects can be avoided by tuning the air flow to the laser repetition rate.

© 2011 OSA

OCIS Codes
(170.1020) Medical optics and biotechnology : Ablation of tissue
(170.3890) Medical optics and biotechnology : Medical optics instrumentation
(220.1000) Optical design and fabrication : Aberration compensation
(330.4460) Vision, color, and visual optics : Ophthalmic optics and devices
(330.5370) Vision, color, and visual optics : Physiological optics

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: December 20, 2010
Revised Manuscript: February 2, 2011
Manuscript Accepted: February 2, 2011
Published: February 24, 2011

Virtual Issues
Vol. 6, Iss. 3 Virtual Journal for Biomedical Optics

Citation
Carlos Dorronsoro, Silvia Schumacher, Pablo Pérez-Merino, Jan Siegel, Michael Mrochen, and Susana Marcos, "Effect of air-flow on the evaluation of refractive surgery ablation patterns," Opt. Express 19, 4653-4666 (2011)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-19-5-4653


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. C. Dorronsoro, D. Cano, J. Merayo-Lloves, and S. Marcos, “Experiments on PMMA models to predict the impact of corneal refractive surgery on corneal shape,” Opt. Express 14(13), 6142–6156 (2006). [CrossRef] [PubMed]
  2. E. Hauge, S. A. Naroo, and W. N. Charman, “Poly(methyl methacrylate) model study of optical surface quality after excimer laser photorefractive keratectomy,” J. Cataract Refract. Surg. 27(12), 2026–2035 (2001). [CrossRef] [PubMed]
  3. W. Bachmann, B. Jean, T. Bende, M. Wohlrab, and H. J. Thiel, “Silicone replica technique and automatic confocal topometry for determination of corneal surface roughness,” Ger. J. Ophthalmol. 2(6), 400–403 (1993). [PubMed]
  4. C. B. Odonnell, J. Kemner, and F. E. Odonnell, “Surface roughness in PMMA is linearly related to the amount of excimer laser ablation,” J. Refract. Surg. 12(1), 171–174 (1996).
  5. S. A. Naroo and W. N. Charman, “Surface roughness after excimer laser ablation using a PMMA model: profilometry and effects on vision,” J. Refract. Surg. 21(3), 260–268 (2005). [PubMed]
  6. J. D. Gottsch, E. V. Rencs, J. L. Cambier, D. Hall, D. T. Azar, and W. J. Stark, “Excimer laser calibration system,” J. Refract. Surg. 12(3), 401–411 (1996). [PubMed]
  7. S. Marcos, C. Dorronsoro, and D. Cano, “Spherical aberration prevention method in e.g. laser refractive surgery system,” (Patent WO 2005/122873 A1, 2005).
  8. C. Dorronsoro, L. Remon, J. Merayo-Lloves, and S. Marcos, “Experimental evaluation of optimized ablation patterns for laser refractive surgery,” Opt. Express 17(17), 15292–15307 (2009). [CrossRef] [PubMed]
  9. C. Roberts, and W. Dupps, “Corneal biomechanics and their role in corneal ablative procedures,” in Customized corneal ablation: The quest for super vision, S. McRae, R. Krueger, and R. Applegate, eds. (Stack publising, 2001).
  10. FDA, “Opthalmic devices panel 110th meeting,” (Food and Drug Administration, 2008).
  11. B. Drum, “Evaluating the Safety and Effectiveness of “Aberration-Free” Ophthalmic Refractive Surgery,” in 9th Annual FDA Science Forum, (Washington, DC, 2003).
  12. B. Drum, “Radial efficiency function in refractive surgery: Ablation losses caused by corneal curvature,” in 11th annual FDA Science Forum, (Washington, DC, 2005).
  13. B. Drum, “The Evolution of the Optical Zone in Corneal Refractive Surgery,” in Wavefront Congress, (Santa Fe, New Mexico, 2007)
  14. C. Dorronsoro, J. Siegel, L. Remon, and S. Marcos, “Suitability of Filofocon A and PMMA for experimental models in excimer laser ablation refractive surgery,” Opt. Express 16(25), 20955–20967 (2008). [CrossRef] [PubMed]
  15. G. H. Pettit and R. Sauerbrey, “Pulsed ultraviolet-laser ablation,” Appl. Phys., A Mater. Sci. Process. 56(1), 51–63 (1993). [CrossRef]
  16. D. W. Hahn, M. N. Ediger, and G. H. Pettit, “Dynamics of ablation plume particles generated during excimer laser corneal ablation,” Lasers Surg. Med. 16(4), 384–389 (1995). [CrossRef] [PubMed]
  17. J. T. Walsh and P. T. Staveteig, “Effect of hydrogen bonding on far-ultraviolet water absorption and potential implications for 193-nm arf excimer laser-tissue interaction,” Proc. SPIE 2391, 176–183 (1995). [CrossRef]
  18. B. T. Fisher and D. W. Hahn, “Development and numerical solution of a mechanistic model for corneal tissue ablation with the 193 nm argon fluoride excimer laser,” J. Opt. Soc. Am. A 24(2), 265–277 (2007). [CrossRef]
  19. B. T. Fisher and D. W. Hahn, “Measurement of small-signal absorption coefficient and absorption cross section of collagen for 193-nm excimer laser light and the role of collagen in tissue ablation,” Appl. Opt. 43(29), 5443–5451 (2004). [CrossRef] [PubMed]
  20. Z. Bor, B. Hopp, B. Rácz, G. Szabó, I. Ratkay, I. Süveges, A. Füst, and J. Mohay, “Plume emission, shock wave and surface wave formation during excimer laser ablation of the cornea,” Refract. Corneal Surg. 9(2Suppl), S111–S115 (1993). [PubMed]
  21. R. R. Krueger, J. S. Krasinski, C. Radzewicz, K. G. Stonecipher, and J. J. Rowsey, “Photography of shock waves during excimer laser ablation of the cornea. Effect of helium gas on propagation velocity,” Cornea 12(4), 330–334 (1993). [CrossRef] [PubMed]
  22. C. A. Puliafito, D. Stern, R. R. Krueger, and E. R. Mandel, “High-speed photography of excimer laser ablation of the cornea,” Arch. Ophthalmol. 105(9), 1255–1259 (1987). [PubMed]
  23. H. Schmidt, J. Ihlemann, B. Wolff-Rottke, K. Luther, and J. Troe, “Ultraviolet laser ablation of polymers: spot size, pulse duration, and plume attenuation effects explained,” J. Appl. Phys. 83(10), 5458–5468 (1998). [CrossRef]
  24. R. Srinivasan and B. Braren, “Ultraviolet-laser ablation of organic polymers,” Chem. Rev. 89(6), 1303–1316 (1989). [CrossRef]
  25. I. Zinovik and A. Povitsky, “Dynamics of multiple plumes in laser ablation: Modeling of the shielding effect,” J. Appl. Phys. 100(2), 024911 (2006). [CrossRef]
  26. R. Sauerbrey and G. H. Pettit, “Theory for the etching of organic materials by ultraviolet-laser pulses,” Appl. Phys. Lett. 55(5), 421–423 (1989). [CrossRef]
  27. H. Lubatschowski, O. Kermani, H. Welling, and W. Ertmer, “A scanning and rotating slit arF excimer laser delivery system for refractive surgery,” J. Refract. Surg. 14(2Suppl), S186–S191 (1998). [PubMed]
  28. J. Noack, R. Tonnies, K. Hohla, R. Birngruber, and A. Vogel, “Influence of ablation plume dynamics on the formation of central islands in excimer laser photorefractive keratectomy,” Ophthalmology 104(5), 823–830 (1997). [PubMed]
  29. C. R. Munnerlyn, M. E. Arnoldussen, A. L. Munnerlyn, and B. A. Logan, “Theory concerning the ablation of corneal tissue with large-area, 193-nm excimer laser beams,” J. Biomed. Opt. 11(6), 064032 (2006). [CrossRef]
  30. C. Dorbecker, H. Lubatschowski, S. Lohmann, C. Ruff, O. Kermani, and W. Ertmer, “Influence of the ablation plume on the removal process during ArF-excimer laser photoablation,” Proc. SPIE 2632, 2–9 (1996). [CrossRef]
  31. L. M. Shanyfelt, P. L. Dickrell, H. F. Edelhauser, and D. W. Hahn, “Effects of laser repetition rate on corneal tissue ablation for 193-nm excimer laser light,” Lasers Surg. Med. 40(7), 483–493 (2008). [CrossRef] [PubMed]
  32. M. Mrochen, M. Kaemmerer, and T. Seiler, “Wavefront-guided laser in situ keratomileusis: early results in three eyes,” J. Refract. Surg. 16(2), 116–121 (2000). [PubMed]
  33. M. Mrochen, R. R. Krueger, M. Bueeler, and T. Seiler, “Aberration-sensing and wavefront-guided laser in situ keratomileusis: management of decentered ablation,” J. Refract. Surg. 18(4), 418–429 (2002). [PubMed]
  34. A. J. Kanellopoulos, J. Conway, and L. H. Pe, “LASIK for hyperopia with the WaveLight excimer laser,” J. Refract. Surg. 22(1), 43–47 (2006). [PubMed]
  35. M. R. Jankov, S. I. Panagopoulou, N. S. Tsiklis, G. C. Hajitanasis, M. Aslanides, and G. Pallikaris, “Topography-guided treatment of irregular astigmatism with the waveLight excimer laser,” J. Refract. Surg. 22(4), 335–344 (2006). [PubMed]
  36. T. Koller, H. P. Iseli, F. Hafezi, M. Mrochen, and T. Seiler, “Q-factor customized ablation profile for the correction of myopic astigmatism,” J. Cataract Refract. Surg. 32(4), 584–589 (2006). [CrossRef] [PubMed]
  37. K. G. Stonecipher and G. M. Kezirian, “Wavefront-optimized versus wavefront-guided LASIK for myopic astigmatism with the ALLEGRETTO WAVE: Three-month results of a prospective FDA trial,” J. Refract. Surg. 24(4), S424–S430 (2008). [PubMed]
  38. D. T. C. Lin, S. R. Holland, K. M. Rocha, and R. R. Krueger, “Method for optimizing topography-guided ablation of highly aberrated eyes with the ALLEGRETTO WAVE Excimer Laser,” J. Refract. Surg. 24(4), S439–S445 (2008). [PubMed]
  39. M. Mrochen, C. Donitzky, C. Wullner, and J. Loffler, “Wavefront-optimized ablation profiles: theoretical background,” J. Cataract Refract. Surg. 30(4), 775–785 (2004). [CrossRef] [PubMed]
  40. M. Mrochen, U. Schelling, C. Wuellner, and C. Donitzky, “Effect of time sequences in scanning algorithms on the surface temperature during corneal laser surgery with high-repetition-rate excimer laser,” J. Cataract Refract. Surg. 35(4), 738–746 (2009). [CrossRef] [PubMed]
  41. M. Mrochen, U. Schelling, C. Wuellner, and C. Donitzky, “Influence of spatial and temporal spot distribution on the ocular surface quality and maximum ablation depth after photoablation with a 1050 Hz excimer laser system,” J. Cataract Refract. Surg. 35(2), 363–373 (2009). [CrossRef] [PubMed]
  42. M. Mrochen, C. Wuellner, K. Rose, and C. Donitzky, “Experimental setup to determine the pulse energies and radiant exposures for excimer lasers with repetition rates ranging from 100 to 1050 Hz,” J. Cataract Refract. Surg. 35(10), 1806–1814 (2009). [CrossRef] [PubMed]
  43. R. Artigas, F. Laguarta, and C. Cadevall, “Dual-technology optical sensor head for 3D surface shape measurements on the micro- and nanoscales,” Proc. SPIE 5457, 166–174 (2004). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Supplementary Material


» Media 1: MOV (2761 KB)     
» Media 2: MOV (2714 KB)     
» Media 3: MOV (2601 KB)     

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited