RECIST versus volume measurement in medical CT using ellipsoids of known size |
Optics Express, Vol. 18, Issue 8, pp. 8151-8159 (2010)
http://dx.doi.org/10.1364/OE.18.008151
Enhanced HTML
Acrobat PDF (162 KB)
Browse Datasets: MIDAS
Full-Text PDF contains links to datasets. See ISP homepage for software requirements and other information.
Abstract
Two hundred eighty three uniaxial ellipsoids with sizes from 4 mm to 11 mm were measured with a coordinate measuring matching (CMM) and also scanned using a medical computed tomography (CT) machine. Their volumes were determined by counting voxels over a threshold, as well as using equivalent volumes from the length given by the RECIST 1.1 criterion (Response Evaluation Criteria in Solid Tumors). The volumetric measurements yield an order of magnitude reduction in residuals compared to the CMM measurements than the residuals of the RECIST measurements also compared to the CMM measurements.
© 2010 OSA
OCIS Codes
(170.3890) Medical optics and biotechnology : Medical optics instrumentation
(170.7440) Medical optics and biotechnology : X-ray imaging
ToC Category:
Medical Optics and Biotechnology
History
Original Manuscript: June 18, 2009
Revised Manuscript: December 22, 2009
Manuscript Accepted: February 8, 2010
Published: April 1, 2010
Virtual Issues
Vol. 5, Iss. 8 Virtual Journal for Biomedical Optics
Imaging in Diagnosis and Treatment of Lung Cancer (2010) Optics Express
Citation
Zachary H. Levine, Bruce R. Borchardt, Nolan J. Brandenburg, Charles W. Clark, Bala Muralikrishnan, Craig M. Shakarji, Joseph J. Chen, and Eliot L. Siegel, "RECIST versus volume measurement in medical CT using ellipsoids of known size," Opt. Express 18, 8151-8159 (2010)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-18-8-8151
Sort: Year | Journal | Reset
References
- E. A. Eisenhauer, P. Therasse, J. Bogaerts, L. H. Schwartz, D. Sargent, R. Ford, J. Dancey, S. Arbuck, S. Gwyther, M. Mooney, L. Rubinstein, L. Shankar, L. Dodd, R. Kaplan, D. Lacombe, J. Verweij, and J. Verweij, “New response evaluation criteria in solid tumours: revised RECIST guideline (version 1.1),” Eur. J. Cancer 45(2), 228–247 (2009). [CrossRef]
- M. A. Gavrielides, L. M. Kinnard, K. J. Myers, and N. Petrick, “Noncalcified lung nodules: volumetric assessment with thoracic CT,” Radiology 251(1), 26–37 (2009). [CrossRef] [PubMed]
- C. I. Henschke, D. I. McCauley, D. F. Yankelevitz, D. P. Naidich, G. McGuinness, O. S. Miettinen, D. M. Libby, M. W. Pasmantier, J. Koizumi, N. K. Altorki, and J. P. Smith, “Early Lung Cancer Action Project: overall design and findings from baseline screening,” Lancet 354(9173), 99–105 (1999). [CrossRef] [PubMed]
- C. I. Henschke and D. F. Yankelevitz, “CT screening for lung cancer: update 2007,” Oncologist 13(1), 65–78 (2008).<jrn[REMOVED IF= FIELD]></jrn> [CrossRef] [PubMed]
- C. I. Henschke and D. F. Yankelevitz, “Erratum for ‘CT screening for lung cancer: Update 2007,’,” Oncologist 13(5), 619 (2008) (</jrn>).<jrn[REMOVED IF= FIELD]> [CrossRef]
- A. P. Reeves, A. B. Chan, D. F. Yankelevitz, C. I. Henschke, B. Kressler, and W. J. Kostis, “On measuring the change in size of pulmonary nodules,” IEEE Trans. Med. Imaging 25(4), 435–450 (2006). [CrossRef] [PubMed]
- K. W. Clark, D. S. Gierada, S. M. Moore, D. R. Maffitt, P. Koppel, S. R. Phillips, and F. W. Prior, “Creation of a CT Image Library for the Lung Screening Study of the National Lung Screening Trial,” J. Digit. Imaging 20(1), 23–31 (2007). [CrossRef]
- D. M. Xu, H. J. van der Zaag-Loonen, M. Oudkerk, Y. Wang, R. Vliegenthart, E. T. Scholten, J. Verschakelen, M. Prokop, H. J. de Koning, and R. J. van Klaveren, “Smooth or attached solid indeterminate nodules detected at baseline CT screening in the NELSON study: cancer risk during 1 year of follow-up,” Radiology 250(1), 264–272 (2009). [CrossRef]
- H. MacMahon, J. H. M. Austin, G. Gamsu, C. J. Herold, J. R. Jett, D. P. Naidich, E. F. Patz, and S. J. Swensen, “Guidelines for management of small pulmonary nodules detected on CT scans: a statement from the Fleischner Society,” Radiology 237(2), 395–400 (2005). [CrossRef] [PubMed]
- G. N. Hounsfield, “Nobel Award address: Computed medical imaging,” Med. Phys. 7(4), 283–290 (1980). [CrossRef] [PubMed]
- N. D. Prionas, S. Ray, and J. M. Boone, “Volume Assessment Accuracy in Computed Tomography: A Phantom Study,” unpublished.
Cited By |
OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.
Multimedia
| Multimedia Files | Recommended Software |
| » Media 1: CSV (8 KB) | QuickTime |
| » Media 2: CSV (9 KB) | QuickTime |
| » Media 3: CSV (9 KB) | QuickTime |
| » Media 4: CSV (9 KB) | QuickTime |
| » Media 5: CSV (9 KB) | QuickTime |
| » Media 6: CSV (9 KB) | QuickTime |
| » Media 7: CSV (9 KB) | QuickTime |
| » Media 8: CSV (9 KB) | QuickTime |
| » Media 9: CSV (9 KB) | QuickTime |
| » Media 10: CSV (9 KB) | QuickTime |
| » Media 11: CSV (9 KB) | QuickTime |





OSA is a member of 