High-speed spectral-domain optical coherence tomography at 1.3 µm wavelength

S. H. Yun, G. J. Tearney, B. E. Bouma, B. H. Park, and J. F. de Boer  »View Author Affiliations


Optics Express, Vol. 11, Issue 26, pp. 3598-3604 (2003)
http://dx.doi.org/10.1364/OE.11.003598


View Full Text Article

Acrobat PDF (295 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We demonstrate a high-speed spectral domain optical coherence tomography (SD-OCT) system capable of acquiring individual axial scans in 24.4 µs at a rate of 19,000 axial scans per second, using an InGaAs line scan camera and broadband light source centered at 1.31 µm. Sensitivity of >105 dB over a 2-mm depth range was obtained with a free-space axial resolution of 12–14 µm, in agreement with our signal-to-noise ratio predictions. Images of human tissue obtained in vivo with SD-OCT show similar penetration depths to those obtained with state-of-the-art time domain OCT despite the ten-fold higher image acquisition speed. These results demonstrate the potential of 1.3 µm SD-OCT for high-speed and high-sensitivity imaging in patients.

© 2003 Optical Society of America

1. Introduction

2. Principle

2.1 SD-OCT System Configuration

Figure 1 shows a schematic of the SD-OCT system. Amplified spontaneous emission from a semiconductor optical amplifier provided a broadband un-polarized Gaussian-like spectrum with full-width-half-maximum (FWHM) of 66 nm centered at 1315 nm. The total output power was 18.7 mW. A single-mode fiber interferometer consisting of a low-loss circulator and wavelength-flattened fiber-optic coupler with a 10/90 splitting ratio, was used for efficient sample arm illumination. At the distal end of the sample arm, a galvanometer-mounted mirror was driven with a saw-tooth waveform to provide transverse beam scanning over 5 mm at the sample. The numerical aperture of the probe was 0.054, resulting in confocal parameter of 1.1 mm and 1/e2 transverse resolution of 30 µm. The light returned from the two arms was directed to a spectrometer via a circulator. In the spectrometer, the optical spectrum was dispersed by a blazed diffraction grating (1200 lines per mm) and imaged by an achromatic doublet lens (focal length: 150 mm) onto an InGaAs CCD array line scan camera (Sensors Unlimited Inc., SU512LX). The total photon-to-electron conversion efficiency of the spectrometer was measured to be 0.8 and 0.3 for s- and p-plane polarization states, respectively. The conversion efficiency includes the diffraction efficiency of the grating and quantum efficiency (~85%) of the InGaAs CCD. The CCD had 512 detector pixels with a 50 µm pitch. The CCD detected a 106-nm bandwidth centered at 1315 nm, where each pixel was separated by 0.208 nm. This sampling interval resulted in a depth range of 2.08 mm. From the measured beam diameter at the diffraction grating, the spectral resolution was calculated to be 0.063 nm, which was narrower than that given by the pixel spacing of the array. The CCD camera was operated at its maximum readout rate of 18.94 kHz. The output of the camera was digitized by using a four channel data acquisition board (DAQ) with 12 bit resolution at a sampling rate of 5 MS/s per channel. The sampled data was transferred continuously to computer memory. A discrete Fourier transform (DFT) was performed on each set of 512 data points acquired by the CCD to produce an axial depth profile of the sample (A-line).

Fig. 1. Schematic of the experimental setup. PC, polarization controller; GM, galvanometer-mounted mirror; DG, diffraction grating; FL, focusing lens; LSC, InGaAs line scan camera; DAQ, data acquisition board.

2.2 SD-OCT SNR Analysis

Fundamental noise sources in heterodyne detection include the electrical noise of the photodetector, shot noise, and relative intensity noise (RIN) from the reference arm light [17

17. W. V. Sorin and D. M. Baney, “A simple intensity noise reduction technique for optical low-coherence reflectometry,” IEEE Photon. Technol. Lett. 4, 1404–1406 (1994). [CrossRef]

]. Maximum SNR is achieved when the RIN from the reference arm power is equal to the electrical noise. Shot noise limited detection is achieved if the shot noise dominates the electrical noise and RIN at this reference arm power. The amount of noise generated per CCD pixel can be conveniently expressed in terms of the number of noise-equivalent electrons accumulated in each CCD scan (A-line). With a CCD gain of 400 nV/electron, the total electrical noise, including the read out and dark noise of the camera and the quantization noise in the analog-to-digital conversion, was found to be Nel=1617. RIN noise is given by NRIN=(fν)1/2 Nref where Nref is the number of electrons per pixel generated by the reference arm light, f is the detection bandwidth or reciprocal of twice the exposure time of the CCD, and Δν=36.4 GHz denotes the FWHM spectral bandwidth of the reference light received by a single pixel (given a square-like spectrum). At its maximum scan rate of 18.94 kHz, our camera operated at an exposure time of 24.4 µs due to its finite readout time (46% duty cycle). This lead to f=20.5 kHz. For the spectrometer efficiency of 0.55 (averaged over two polarization states), a maximum SNR in a single pixel was reached at Nref=2.15×106 where NRIN=Nel. This corresponded to 43% of the full well depth (5×106 electrons) and a reference arm optical power of 24.6 nW. The number of shot noise electrons is given by Nsh=(Nref)1/2. At Nref=2.15×106, Nsh=1466; therefore, the total noise power of our system was 3.4 times larger than the shot-noise limit.

It can be shown [5

5. R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of Fourier domain vs. time domain optical coherence tomography,” Opt. Express 11, 889–894 (2003), http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-8-889 [CrossRef] [PubMed]

8

8. S. H. Yun, G. J. Tearney, J. F. de Boer, N. Iftimia, and B. E. Bouma, “High-speed optical frequency-domain imaging,” Opt. Express 11, 2953–2963 (2003), http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-18-2183 [CrossRef] [PubMed]

] that the sensitivity S, defined by the reciprocal of the noise equivalent reflectivity in the sample, is given by

S[dB]=10×log(Ns1+Nel2Nref+α(fΔν)Nref),
(1)

where ∑Ns denotes the sum of electrons over the entire array generated by sample arm light returning from a 100% reflector, and α=1 for an un-polarized (α=2 for polarized) light source. We note that this expression is a valid approximation for uniform spectral density. In general, the actual signal and noise power for individual CCD pixels should be integrated over the spectral profile to obtain the overall SNR. We measured a total sample arm power of 6.3 mW returning from a gold mirror to the detector arm of the interferometer. This optical power corresponded to ∑Ns=5.56×1011. Therefore, for the noise power described earlier, Eq. (1) predicted a sensitivity of 112.1 dB. A more detailed calculation considering the actual Gaussian-like spectrum of the source and polarization-dependent spectrometer efficiency predicted a maximum theoretical sensitivity of 110.3 dB.

3. Experiment

3.1 Sensitivity

In our experiment, we set the total reference power to be between 7 and 11 µW (6.2×108 - 9.7×108 electrons) so that both the SNR and dynamic range were optimized. Figure 2 shows a typical point spread function (PSF), plotted in a log power scale, measured with a neutral density filter (total attenuation: -55 dB) and a gold mirror in the sample arm (Curve A, black line). The curve was obtained by averaging over 500 PSF’s in power to reduce the fluctuation of the noise floor and present the fine structure in the signal more clearly. The reference arm mirror was positioned to produce a path length difference of 1.3 mm. To correct for the nonlinear k-space sampling interval, 512 sampled points per A-line scan were mapped to uniform frequency spacing by linear interpolation. Interpolation was performed by computing a DFT of the sampled data, zero padding, inverse DFT, and re-sampling of the resultant data at regular intervals [18

18. C. Dorrer, N. Belabas, J-P Likforman, and M. Joffre, “Spectral resolution and sampling issues in Fourier-transform spectral interferometry,” J. Opt. Soc. Am. B 17, 1795–1802 (2000). [CrossRef]

]. This zero-padding process was essential to obtain a sidelobe-free PSF. The small noise peaks at other depth locations may be attributed to multiplicative noise due to fine structure in the source spectrum. Curve B (red line) represents the noise floor, averaged over 500 A-lines, obtained with the sample arm blocked. From the SNR of 51.7 dB, the ratio of the peak value of the signal power (curve A) and the noise floor (curve B), a sensitivity of 106.7 dB was obtained. Curve C (green line) shows the electrical noise level obtained with both the reference and sample arms blocked. The electrical noise was approximately 4.5 dB lower than the total noise level, in agreement with the noise analysis described above.

Fig. 2. Typical point spread function obtained with a partial reflector with -55 dB reflectivity (curve A, black); noise floor measured with the reference light only (curve B, red); camera read out noise (curve C, green). All the curves were obtained by averaging over 500 consecutive measurements to facilitate comparison.
Fig. 3. (a) Sensitivity measured as a function of depth (circles, black dotted line); theoretical fit (curve A’, green); theoretical sensitivity for shot-noise-limited SD-OCT (curve B’, red) and TD-OCT (curve C’, blue). (b) Axial resolution measured as the FWHM

Figure 3(a) shows the sensitivity of the system measured as a function of depth (solid circles, black). The sensitivity is >109 dB from 0 to 0.5 mm and >105 dB up to 1.8 mm. The sensitivity decreases with depth because the finite resolution of the spectrometer reduces fringe visibility more strongly at higher fringe frequencies [12

12. G. Hausler and M. W. Lindner, “Coherence radar and spectral radar - new tools for dermatological diagnosis,” J. Biomed. Opt. 3, 21–31 (1998). [CrossRef]

,18

18. C. Dorrer, N. Belabas, J-P Likforman, and M. Joffre, “Spectral resolution and sampling issues in Fourier-transform spectral interferometry,” J. Opt. Soc. Am. B 17, 1795–1802 (2000). [CrossRef]

]. The magnitude of the decrease, R(z), can be shown to be:

R(z)=(sinζζ)2·exp[w22ln2ζ2],
(2)

3.2 Resolution

Figure 3(b) shows the FWHM axial resolution determined from a Gaussian fit to the PSF in amplitude (reflectance profile) at various depths. At small depths below 0.5 mm, the free-space axial resolution was measured to be approximately 12 µm, which is close to the 11.6 µm calculated from the source bandwidth. However, the resolution decreased with the depth; it was 13.5 µm at a depth of 1.7 mm, and beyond 1.7 mm, a significant degradation of resolution was measured. These results indicate that the interpolation process is quite sensitive to small errors at high fringe frequencies, leading to poorer than transform-limited resolution at the maximum depth. These errors may be minimized by utilizing a linear array with more pixels. Future developments in InGaAs CCD array technology are merited to improve the ranging depth of high-speed 1.3 µm SD-OCT. Alternatively the ranging depth may be increased by a factor of two by measuring the complex spectral density from quadrature signals [13

13. M. Wojtkowski, A. Kowalczyk, R. Leitgeb, and A. F. Fercher, “Full range complex spectral optical coherence tomography technique in eye imaging,” Opt. Lett. 27, 1415–1417 (2002). [CrossRef]

].

3.3 Images

Figure 4(a) depicts the ventral portion of a volunteer’s finger acquired by SD-OCT at 38 frames per second (fps). The image comprised 256 axial and 500 transverse pixels and was plotted using a logarithmic inverse grayscale lookup table. The focal point of the imaging lens was positioned in the middle of the depth range. For comparison, we also show an image of the same sample obtained with a state-of-the-art TD-OCT system from our laboratory (Fig. 4b) [2

2. G. J. Tearney, H. Yabushita, S. L. Houser, H. T. Aretz, I. K. Jang, K. Schlendorf, C. R. Kauffman, M. Shishkov, E. F. Halpern, and B. E. Bouma, “Quantification of macrophage content in atherosclerotic plaques by optical coherence tomography,” Circulation 106, 113–119 (2003). [CrossRef]

], which uses the same light source and interferometer configuration. The TD-OCT system obtains images at 4 fps, has a ranging depth of 2.5 mm, and a shot-noise-limited detection sensitivity of approximately 108 dB. Despite the nearly 10-fold higher imaging speed, the SD-OCT image exhibits similar resolution, contrast, and imaging penetration to the TD-OCT image.

Fig. 4. (a) Image of a human finger acquired in vivo with the SD-OCT system at 38 fps (256 axial×500 transverse pixels, 2.1×5.0 mm). (b) Image of the same human finger (250 axial×500 transverse pixels, 2.5×5.0 mm) acquired at 4 fps using a state-of-the-art TD-OCT system. The scale bars represent 0.5 mm.

4. Conclusion

Acknowledgements

This research was supported in part by research grants from the National Institutes of Health (R01-RR019768), Department of Defense (F4 9620-01-1-0014), Center for Integration of Medicine and Innovative Technology (CIMIT), and a gift from Dr. and Mrs. J.S. Chen to the optical diagnostics program of the Wellman Center of Photomedicine.

References and links

1.

E. A. Swanson, D. Huang, M. R. Hee, J. G. Fujimoto, C. P. Lin, and C. A. Puliafito, “High-speed optical coherence domain reflectometry,” Opt. Lett. 17, 151–153 (1992). [CrossRef] [PubMed]

2.

G. J. Tearney, H. Yabushita, S. L. Houser, H. T. Aretz, I. K. Jang, K. Schlendorf, C. R. Kauffman, M. Shishkov, E. F. Halpern, and B. E. Bouma, “Quantification of macrophage content in atherosclerotic plaques by optical coherence tomography,” Circulation 106, 113–119 (2003). [CrossRef]

3.

P. Andretzky, M. W. Lindner, J. M. Hermann, A. Schultz, M. Konzog, F. Kiesewetter, and G. Hausler, “Optical coherence tomography by spectral radar: dynamic range estimation and in vivo measurements of skin,” Proc. SPIE 3567, 78–87 (1998).

4.

M. Wojtkowski, R. Leitgeb, A. Kowalczyk, T. Bajraszewski, and A. F. Fercher, “In vivo human retinal imaging by Fourier domain optical coherence tomography,” J. Biomed. Opt. 7, 457–463 (2002). [CrossRef] [PubMed]

5.

R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of Fourier domain vs. time domain optical coherence tomography,” Opt. Express 11, 889–894 (2003), http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-8-889 [CrossRef] [PubMed]

6.

J. F. de Boer, B. Cense, B.H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28, 2067–2069 (2003). [CrossRef] [PubMed]

7.

M. A. Choma, M. V. Sarunic, C. Uang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11, 2183–2189 (2003), http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-18-2183 [CrossRef] [PubMed]

8.

S. H. Yun, G. J. Tearney, J. F. de Boer, N. Iftimia, and B. E. Bouma, “High-speed optical frequency-domain imaging,” Opt. Express 11, 2953–2963 (2003), http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-18-2183 [CrossRef] [PubMed]

9.

S. R. Chinn, E. Swanson, and J. G. Fujimoto, “Optical coherence tomography using a frequency-tunable optical source,” Opt. Lett. 22, 340–342 (1997). [CrossRef] [PubMed]

10.

B. Golubovic, B. E. Bouma, G. J. Tearney, and J. G. Fujimoto, “Optical frequency-domain reflectometry using rapid wavelength tuning of a Cr4+:forsterite laser,” Opt. Lett. 22, 1704–1706 (1997). [CrossRef]

11.

A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. El-Zaiat, “Measurements of intraocular distances by backscattering spectral interferometry,” Opt. Comm. 117, 43–48 (1995). [CrossRef]

12.

G. Hausler and M. W. Lindner, “Coherence radar and spectral radar - new tools for dermatological diagnosis,” J. Biomed. Opt. 3, 21–31 (1998). [CrossRef]

13.

M. Wojtkowski, A. Kowalczyk, R. Leitgeb, and A. F. Fercher, “Full range complex spectral optical coherence tomography technique in eye imaging,” Opt. Lett. 27, 1415–1417 (2002). [CrossRef]

14.

M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett. 28, 1745–1747 (2003). [CrossRef] [PubMed]

15.

N. Nassif, B. Cense, B. H. Park, S. H. Yun, T. C. Chen, B. E. Bouma, G. J. Tearney, and J. F. de Boer, “In-vivo human retinal imaging by ultra high-speed spectral domain optical coherence tomography,” Opt. Lett. (in press)

16.

B. E. Bouma and G. J. Tearney, “Clinical imaging with optical coherence tomography,” Acad. Radiol. 9, 942–953 (2002). [CrossRef] [PubMed]

17.

W. V. Sorin and D. M. Baney, “A simple intensity noise reduction technique for optical low-coherence reflectometry,” IEEE Photon. Technol. Lett. 4, 1404–1406 (1994). [CrossRef]

18.

C. Dorrer, N. Belabas, J-P Likforman, and M. Joffre, “Spectral resolution and sampling issues in Fourier-transform spectral interferometry,” J. Opt. Soc. Am. B 17, 1795–1802 (2000). [CrossRef]

OCIS Codes
(110.4500) Imaging systems : Optical coherence tomography
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(170.4500) Medical optics and biotechnology : Optical coherence tomography

ToC Category:
Research Papers

History
Original Manuscript: November 25, 2003
Revised Manuscript: December 10, 2003
Published: December 29, 2003

Citation
S. Yun, G. Tearney, B. Bouma, B. Park, and Johannes de Boer, "High-speed spectral-domain optical coherence tomography at 1.3 μm wavelength," Opt. Express 11, 3598-3604 (2003)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-26-3598


Sort:  Journal  |  Reset  

References

  1. E. A. Swanson, D. Huang, M. R. Hee, J. G. Fujimoto, C. P. Lin, and C. A. Puliafito, �??High-speed optical coherence domain reflectometry,�?? Opt. Lett. 17, 151-153 (1992). [CrossRef] [PubMed]
  2. G. J. Tearney, H. Yabushita, S. L. Houser, H. T. Aretz, I. K. Jang, K. Schlendorf, C. R. Kauffman, M. Shishkov, E. F. Halpern, and B. E. Bouma, �??Quantification of macrophage content in atherosclerotic plaques by optical coherence tomography,�?? Circulation 106, 113-119 (2003). [CrossRef]
  3. P. Andretzky, M. W. Lindner, J. M. Hermann, A. Schultz, M. Konzog, F. Kiesewetter, and G. Hausler, �??Optical coherence tomography by spectral radar: dynamic range estimation and in vivo measurements of skin,�?? Proc. SPIE 3567, 78-87 (1998).
  4. M. Wojtkowski, R. Leitgeb, A. Kowalczyk, T. Bajraszewski, and A. F. Fercher, �??In vivo human retinal imaging by Fourier domain optical coherence tomography,�?? J. Biomed. Opt. 7, 457-463 (2002). [CrossRef] [PubMed]
  5. R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, �??Performance of Fourier domain vs. time domain optical coherence tomography,�?? Opt. Express 11, 889-894 (2003), <a href="http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-8-889">http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-8-889</a>. [CrossRef] [PubMed]
  6. J. F. de Boer, B. Cense, B.H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, �??Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,�?? Opt. Lett. 28, 2067-2069 (2003). [CrossRef] [PubMed]
  7. M. A. Choma, M. V. Sarunic, C. Uang, and J. A. Izatt, �??Sensitivity advantage of swept source and Fourier domain optical coherence tomography,�?? Opt. Express 11, 2183-2189 (2003), <a href="http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-18-2183">http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-18-2183</a>. [CrossRef] [PubMed]
  8. S. H. Yun, G. J. Tearney, J. F. de Boer, N. Iftimia, and B. E. Bouma, �??High-speed optical frequency domain imaging,�?? Opt. Express 11, 2953-2963 (2003), <a href="http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-22-2953">http://www.opticsexpress.org/abstract.cfm?URI=OPEX-11-22-2953</a>. [CrossRef] [PubMed]
  9. S. R. Chinn, E. Swanson, and J. G. Fujimoto, �??Optical coherence tomography using a frequency-tunable optical source,�?? Opt. Lett. 22, 340-342 (1997). [CrossRef] [PubMed]
  10. B. Golubovic, B. E. Bouma, and G. J. Tearney, and J. G. Fujimoto, �??Optical frequency-domain reflectometry using rapid wavelength tuning of a Cr4+:forsterite laser,�?? Opt. Lett. 22, 1704-1706 (1997). [CrossRef]
  11. A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. El-Zaiat, �??Measurements of intraocular distances by backscattering spectral interferometry,�?? Opt. Comm. 117, 43-48 (1995). [CrossRef]
  12. G. Hausler and M. W. Lindner, �??Coherence radar and spectral radar - new tools for dermatological diagnosis,�?? J. Biomed. Opt. 3, 21-31 (1998). [CrossRef]
  13. M. Wojtkowski, A. Kowalczyk, R. Leitgeb, and A. F. Fercher, �??Full range complex spectral optical coherence tomography technique in eye imaging,�?? Opt. Lett. 27, 1415-1417 (2002). [CrossRef]
  14. M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, �??Real time in vivo imaging by high-speed spectral optical coherence tomography,�?? Opt. Lett. 28, 1745-1747 (2003). [CrossRef] [PubMed]
  15. N. Nassif, B. Cense, B. H. Park, S. H. Yun, T. C. Chen, B. E. Bouma, G. J. Tearney, J. F. de Boer, �??In-vivo human retinal imaging by ultra high-speed spectral domain optical coherence tomography,�?? Opt. Lett. (in press)
  16. B. E. Bouma and G. J. Tearney, �??Clinical imaging with optical coherence tomography,�?? Acad. Radiol. 9, 942-953 (2002). [CrossRef] [PubMed]
  17. W. V. Sorin and D. M. Baney, �??A simple intensity noise reduction technique for optical low-coherence reflectometry,�?? IEEE Photon. Technol. Lett. 4, 1404-1406 (1994). [CrossRef]
  18. C. Dorrer, N. Belabas, J-P Likforman, and M. Joffre, �??Spectral resolution and sampling issues in Fourier-transform spectral interferometry,�?? J. Opt. Soc. Am. B 17, 1795-1802 (2000). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1. Fig. 2. Fig. 3.
 
Fig. 4.
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited