Frequency measurement of acetylene-stabilized lasers using a femtosecond optical comb without carrier-envelope offset frequency control

Jie Jiang, Atsushi Onae, Hirokazu Matsumoto, and Feng-Lei Hong  »View Author Affiliations


Optics Express, Vol. 13, Issue 6, pp. 1958-1965 (2005)
http://dx.doi.org/10.1364/OPEX.13.001958


View Full Text Article

Acrobat PDF (518 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We have established a frequency measurement system for frequency-stabilized lasers operating in telecommunication wavelength bands, by using a femtosecond optical comb without the need for carrier envelope offset frequency control. This system has been used to measure the frequency of an acetylene-stabilized laser operating at 1542 nm for a period of over 10 hours. The frequency stability of the acetylene-stabilized laser is estimated to be 3×10-12 for a 10-s averaging time, improving toward 1×10-13 after 10000 s. We have measured three acetylene-stabilized lasers, including one commercially available laser, and confirmed that the frequency values are in good agreement (a frequency scatter of 2.1 kHz) with previously measured results reported by different institutes. In addition to the P(16) line of acetylene at 1542 nm, we measured the absolute frequencies of the P(24) line at 1547 nm, the P(1) line at 1534 nm, and the R(5) line at 1530 nm with a view to improving the accuracy of the acetylene frequency atlas. The acetylene-stabilized laser serves as an important optical frequency standard for telecommunication applications.

© 2005 Optical Society of America

1. Introduction

Recent advances in optical communication systems that include high-density wavelength-division multiplexing (DWDM) have motivated research on optical frequency standards in the 1.5-µm wavelength region. Doppler-free high-resolution spectroscopy of acetylene was demonstrated and used to stabilize the frequency of a laser diode in the 1.5-µm wavelength region [1

1. M. de Labachelerie, K. Nakagawa, Y. Awaji, and M. Ohtsu, “High-frequency-stability laser at 1.5 µm using Doppler-free molecular lines,” Opt. Lett. 20, 572–574 (1995). [CrossRef]

, 2

2. K. Nakagawa, M. de Labachelerie, Y. Awaji, and M. Kourogi, “Accurate optical frequency atlas of the 1.5-µm bands of acetylene,” J. Opt. Soc. Am. B 13, 2708–2714 (1996). [CrossRef]

]. Acetylene provides several overtone and combination bands, of which a relatively strong band corresponds to the combination band of a C-H stretching vibrational mode (assigned as ν13). This ν13 band of acetylene is coincident with the International Telecommunication Union’s C band (1530–1560 nm) and has attracted considerable attention as a frequency standard for telecommunication applications [2

2. K. Nakagawa, M. de Labachelerie, Y. Awaji, and M. Kourogi, “Accurate optical frequency atlas of the 1.5-µm bands of acetylene,” J. Opt. Soc. Am. B 13, 2708–2714 (1996). [CrossRef]

, 3

3. A. Onae, K. Minoshima, J. Yoda, K. Nakagawa, A. Yamaguchi, M. Kourogi, K. Imai, and B. Widiyatomoko, “Toward an accurate frequency standard at 1.5 µm based on the acetylene overtone band transition,” IEEE Trans. Instrum. Meas. 48, 563–566 (1999). [CrossRef]

].

Nowadays, optical frequency combs based on femtosecond lasers have revolutionized the field of optical frequency metrology, making possible the absolute measurement of optical frequencies with unprecedented accuracy [6

6. Th. Udem, J. Reichert, R. Holzwarth, and T. W. Haensch, “Absolute optical frequency measurement of the Cesium D1 line with a mode-locked laser,” Phys. Rev. Lett. 82, 3568–3571 (1999). [CrossRef]

, 7

7. D. J. Jones, S. A. Diddams, J. K. Ranka, A. Stentz, R. S. Windeler, J. L. Hall, and S. T. Cundiff, “Carrier-envelope phase control of femtosecond mode-locked lasers and direct optical frequency synthesis,” Science 288, 635–639 (2000). [CrossRef] [PubMed]

]. In 2003, for the first time, we measured the absolute frequency of a diode laser that was locked on the P(16) line of acetylene by using a femtosecond optical comb based on a Kerr-lens mode-locked (KLM) Ti:sapphire (Ti:s) laser [8

8. F. -L. Hong, A. Onae, J. Jiang, R. Guo, H. Inaba, K. Minoshima, T. R. Schibli, H. Matsumoto, and K. Nakagawa, “Absolute frequency measurement of an acetylene-stabilized laser at 1542 nm,” Opt. Lett. 28, 2324–2326 (2003). [CrossRef] [PubMed]

]. Although the P(16) line does not lie within the bandwidth of the comb based on the Ti:s laser (~500–1000 nm), we frequency-doubled the acetylene-stabilized laser to 771 nm. In this experiment, the measurement accuracy was limited not by the uncertainty of the frequency measurement but by the reproducibility of the acetylene stabilized laser. The frequency of the P(16) line based on the Ti:s comb has also been measured by other groups in different institutes [9

9. C. S. Edwards, H. S. Margolis, G. P. Barwood, S. N. Lea, P. Gill, G. Huang, and W. R. C. Rowley, “Absolute frequency measurement of a 1.5-µm acetylene standard by use of a combined frequency chain and femtosecond comb,” Opt. Lett. 29, 566–568 (2004). [CrossRef] [PubMed]

, 10

10. A. Czajkowski, J. E. Bernard, A. A. Madej, and R. S. Windeler, “Absolute frequency measurement of acetylene transitions in the region of 1540 nm,” Appl. Phys. B 79, 45–50 (2004). [CrossRef]

].

During frequency measurements based on the optical comb, it is usually necessary to stabilize two degrees of freedom of the comb, namely the repetition rate (f rep) and the carrier-envelope offset (CEO) frequency (f CEO). The f CEO of a femtosecond comb spanning more than one octave was directly measured with an f-to-2f interferometer [7

7. D. J. Jones, S. A. Diddams, J. K. Ranka, A. Stentz, R. S. Windeler, J. L. Hall, and S. T. Cundiff, “Carrier-envelope phase control of femtosecond mode-locked lasers and direct optical frequency synthesis,” Science 288, 635–639 (2000). [CrossRef] [PubMed]

]. An octave-spanning comb is often obtained by injecting the output of a KLM Ti:s laser into a photonic crystal fiber (PCF) [11

11. J. C. Knight, “Photonic crystal fibers,” Nature 424, 847–851 (2003). [CrossRef] [PubMed]

]. Due to the small core diameter of the PCF (~1 µm), the efficiency with which the laser beam couples into the PCF varies during the operation. This affects the stability of the signal-to-noise ratio (S/N) of f CEO, and hence limits the long-term operation of the frequency measurement system. There are several methods for improving the long-term operation of the frequency measurement system.

a) The core area of the PCF input side has been increased by splicing a short piece of tapered single-mode fiber onto the PCF. Meanwhile, active control of the coupling stage has been introduced to optimize the coupling efficiency [12

12. Ronald Holzwarth, Max Planck Inst Quantum Opt, D-85748 Garching, and Menlo Systems GmbH, D-82152 Munich, Germany (personal communication, 2004).

].

b) With a mode-locked fiber laser, a highly nonlinear fiber has been spliced onto the laser fiber to remove the coupling problem [13

13. T. R. Schibli, K. Minoshima, F.-L. Hong, H. Inaba, A. Onae, H. Matsumoto, I. Hartl, and M. E. Fermann, “Frequency metrology with a turnkey all-fiber system,” Opt. Lett. 29, 2467–2469 (2004). [CrossRef] [PubMed]

].

c) With a broadband mode-locked laser, f CEO has been observed without a PCF [14

14. T. M. Ramond, S. A. Diddams, L. Hollberg, and A. Bartels, “Phase-coherent link from optical to microwave frequencies by means of the broadband continuum from a 1-GHz Ti:sapphire femtosecond oscillator,” Opt. Lett. 27, 1842–1844 (2002). [CrossRef]

].

Alternatively, the sum-frequency generation (SFG) of an optical comb is considered an excellent method for measuring optical frequencies without the need for f CEO control. SFG between an infrared optical frequency standard and the low-frequency wing of an optical comb yields an SFG comb. Assuming the SFG comb can be tuned to realize a spectral overlap with the high-frequency wing of the original comb, the beat note between the SFG comb and the high-frequency wing (both combs have the same origin) allows us to make an f CEO-independent frequency measurement. This scheme was employed for measuring the absolute optical frequency of an OsO4-stabilized CO2 laser [15

15. A. Amy-Klein, A. Goncharov, C. Daussy, C. Grain, O. Lopez, G. Santarelli, and C. Chardonnet, “Absolute frequency measurement in the 28-THz spectral region with a femtosecond laser comb and a long-distance optical link to a primary standard,” Appl. Phys. B 78, 25–30 (2004). [CrossRef]

] and implementing optical clockwork [16

16. O. D. Mücke, O. Kuzucu, F. N. C. Wong, E. P. Ippen, F. X. Kaertner, S. M. Foreman, D. J. Jones, L.-S. Ma, J. L. Hall, and J. Ye, “Experimental implementation of optical clockwork without carrier-envelope phase control,” Opt. Lett. 29, 2806–2808 (2004). [CrossRef] [PubMed]

]. An SFG comb was evaluated and found to have excellent accuracy and stability [17

17. M. Zimmermann, Ch. Gohle, R. Holzwarth, Th. Udem, and T. Haensch, “Optical clockwork with an offset-free difference-frequency comb: accuracy of sum-and difference-frequency generation,” Opt. Lett. 29, 310–312 (2004). [CrossRef] [PubMed]

].

Fig. 1. Schematic diagram of the experimental setup. Ti:s, Ti:sapphire; CM, chirped mirror; OC, output coupler; PZT, piezoelectric transducer; BS, beam splitter; MO, micro-objective lens; PCF, photonic crystal fiber; SC, supercontinuum; DM, dichroic mirror; PBS, polarization beam splitter; BP, band-pass filter; APD, avalanche photo detector; EDFA, erbium-doped fiber amplifier; PLL, phase-lock loop.

2. Experimental setup

Figure 1 shows a schematic diagram of our experimental setup. The light source is a home-made KLM ring-cavity Ti:s laser with an f rep of 380 MHz. A 4-mm-long Ti:s crystal with an absorption coefficient of ~5 cm-1 at 532 nm was mounted at the Brewster angle. All the cavity mirrors except the output coupler were chirped mirrors. As a pump light, the output of a frequency-doubled Nd:YVO4 laser (Coherent, Verdi V10) was focused into the crystal by using a lens with a 40-mm focal length. With a pump power of 5 W, we obtained an average output power of 560 mW from the Ti:s laser with a pulse duration of 40 fs. The central wavelength and the spectrum bandwidth were 800 and 25 nm (FWHM), respectively.

Part of the output of the mode-locked laser was detected with an avalanche photo detector (APD1) to allow us to measure f rep. The repetition rate, f rep was beat down to approximately 20 MHz by using a synthesizer with a fixed frequency of 360 MHz. The beat-down frequency was fed to a phase-locked loop to control the length of the piezoelectric transducer attached to the mode-locked laser. The synthesizer was phase locked to an H-maser.

The output pulses were again chirp-compensated and split into two parts by a 40/60 beam splitter. One part, with an average power of 330 mW, was launched into a PCF for supercontinuum (SC) generation. A cw acetylene-stabilized laser at 1542 nm was amplified by an erbium-doped fiber amplifier (EDFA) and overlapped with the other part of the laser pulses from the Ti:s laser by a dichroic mirror. The overlapped beams with dual wavelengths were focused into a KTiOPO4 (KTP) crystal to generate an SFG comb.

The SFG comb was arranged so that it overlapped with the SC comb at a polarization beam splitter (PBS1). A delay line was introduced in the path of the SC comb to match the timing of the two combs (which were actually fs pulses in the time domain). A half-wave plate and another polarization beam splitter (PBS2) were introduced to adjust the power ratio of the two combs in the beat measurement. The beat frequency of the two combs was detected after a band-pass filter by a second avalanche photo detector (APD2) and recorded by a counter (Agilent 53132A Universal Counter). The counter used a time base obtained from the H-maser.

3. Experimental results

3.1 Sum-frequency comb generation and beat frequency measurement

In Fig. 2(a), the spectra of the original Ti:s comb are indicated as the dashed curve. This comb is used to generate the sum-frequency comb with a cw acetylene-stabilized laser at 1542 nm. The SFG was realized by using a 5-mm-long Type-II phase matching (θ=59.6°, Φ=0°) KTP crystal. In Fig. 2(a), the solid curve indicates the spectra of the SFG comb, with a central wavelength and bandwidth of 527 and 7 nm, respectively. The spectrum bandwidth of the SFG comb was limited by the phase matching condition of the KTP crystal. The average power of the SFG comb was about 100 nW. The power level of each comb is relatively adjusted in Fig. 2(a) for better presentation. The cw laser power after the EDFA was about 160 mW.

Fig. 2. (a) Spectra of the original Ti:s comb (dashed curve), the SC comb after the PCF (dotted curve), the SFG comb (solid curve), and the acetylene-stabilized laser (straight line). (b) Beat frequency observed between the SFG and SC combs at 527 nm. The resolution bandwidth was 300 kHz.

The spectrum of the SC after the PCF ranged from 480 to 1120 nm (-20 dB) and is shown by the dotted curve in Fig. 2. The PCF used in this experiment had a core diameter of 1.8 µm and was 30 cm long. The power of the generated SC after the PCF was 160 mW. The SFG comb and part of the SC comb in the same wavelength region (527 nm) were combined spatially and temporally to generate a heterodyne beat note (f beat) [shown in Fig. 2(b)]. The S/N of f beat exceeds 30 dB at a resolution bandwidth of 300 kHz. The average power of the SC comb used for the beat measurement around 527 nm was about 10 mW.

With the measured f beat, the frequency of the cw laser at 1.5 µm (f laser) can be expressed as f laser=n×f rep±f beat. We notice that f CEO does not appear here in the formula because both the SFG and SC combs have the same origin.

Table 1. Selected operating parameters and characteristics of the laser systems

table-icon
View This Table
| View All Tables

3.2 Long-term stability of acetylene-stabilized laser

The stability of A4 was 3×10-12 for a 10-s averaging time, improving to 1×10-13 after 10000 s, without showing a flicker floor. Stable long-term operation of the comb measurement system was achieved by removing one critical servo port (f CEO) using the SFG scheme. For comparison, the measured stability of A1 [8

8. F. -L. Hong, A. Onae, J. Jiang, R. Guo, H. Inaba, K. Minoshima, T. R. Schibli, H. Matsumoto, and K. Nakagawa, “Absolute frequency measurement of an acetylene-stabilized laser at 1542 nm,” Opt. Lett. 28, 2324–2326 (2003). [CrossRef] [PubMed]

] is also indicated in Fig. 3 as a straight line. The stability of A4 is slightly better than that of A1 in the 10–1000 s averaging time range.

Fig. 3. Measured root Allan variance of A4 (solid line with filled circles). The root Allan variance of A1 and H-maser are shown as a dashed line and a solid line, respectively, for comparison.

3.3 Frequency measurement of acetylene-stabilized lasers

As shown in Table 2, the present frequency value of A1 is about 2.2 kHz lower than our previously measured result [8

8. F. -L. Hong, A. Onae, J. Jiang, R. Guo, H. Inaba, K. Minoshima, T. R. Schibli, H. Matsumoto, and K. Nakagawa, “Absolute frequency measurement of an acetylene-stabilized laser at 1542 nm,” Opt. Lett. 28, 2324–2326 (2003). [CrossRef] [PubMed]

]. This frequency difference is within the reproducibility of A1. A similar discussion is also applied to the frequency difference of A2. The absolute frequency of A4 is measured for the first time. The accuracy of the H-maser was calibrated to UTC and confirmed to be better than 1×10-14 during the measurements.

Table 2. Absolute frequency of three different lasers locked on the P(16) line of 13C2H2.

table-icon
View This Table
| View All Tables

Fig. 4. Absolute frequency of acetylene-stabilized lasers obtained in different institutes. NPL-1 and NPL-2 are the lasers from NPL [9]. NRC is the laser from NRC [10].

Table 3. Absolute frequency values of acetylene lines

table-icon
View This Table
| View All Tables

3.4 Absolute frequency measurement of multiple acetylene lines in 1.5-µm band

The acetylene lines in the 1515 to 1550 nm wavelength range [2

2. K. Nakagawa, M. de Labachelerie, Y. Awaji, and M. Kourogi, “Accurate optical frequency atlas of the 1.5-µm bands of acetylene,” J. Opt. Soc. Am. B 13, 2708–2714 (1996). [CrossRef]

] provide an attractive frequency atlas for telecommunication applications. The absolute frequency of the P(24), P(1) and R(5) lines are measured, for the first time, using the SFG comb. The measured frequencies are listed in Table 3. The frequency measurement was performed with A2. The absolute frequency of each line was measured by taking an average of 10 to 50 beat frequency data (depending on the measurement run), where each beat frequency datum was measured with a 10-s gate time by the counter. The P(24) line is at the low frequency end of the operation range of our laser due to the S/N limitation of the absorption signal. In contrast, the lines with frequencies higher than the R(5) line are outside the amplification range of our EDFA. We notice that the frequency measurement range with the present SFG scheme is much wider than the phase-matching frequency range of the periodically poled LiNbO3 (PPLN) crystal, which was used to double the frequency of the acetylene-stabilized laser from 1542 to 771 nm for frequency measurements with a Ti:s based comb [8

8. F. -L. Hong, A. Onae, J. Jiang, R. Guo, H. Inaba, K. Minoshima, T. R. Schibli, H. Matsumoto, and K. Nakagawa, “Absolute frequency measurement of an acetylene-stabilized laser at 1542 nm,” Opt. Lett. 28, 2324–2326 (2003). [CrossRef] [PubMed]

].

The frequency of the line spacing was measured by measuring the heterodyne beat between two acetylene-stabilized lasers bridged with an optical frequency comb localized at 1.5 µm [19

19. A. Onae, K. Okumura, F.-L. Hong, and H. Matsumoto, “Accurate frequency atlas of 1.5 µm band of acetylene measured by a mode-locked fiber laser,” presented at the 2004 Conference on Precision Electromagnetic Measurements, London, United Kingdom, 27 June-2 July, 2004.

]. From the metrological point of view, it is important to confirm this line spacing measurement with a different method. By using the absolute frequency values listed in Table 3, we calculated the frequency spacing between the lines and listed them in Table 4. These results are compared with the measured line spacing in [19

19. A. Onae, K. Okumura, F.-L. Hong, and H. Matsumoto, “Accurate frequency atlas of 1.5 µm band of acetylene measured by a mode-locked fiber laser,” presented at the 2004 Conference on Precision Electromagnetic Measurements, London, United Kingdom, 27 June-2 July, 2004.

] (also listed in Table 4). Good agreement is obtained between the two measurements. These results should also be included in the CIPM recommendation.

Table 4. Line spacing of the acetylene frequency atlas

table-icon
View This Table
| View All Tables

4. Conclusion

We have established a frequency measurement system without f CEO control for frequency-stabilized lasers operating in the 1.5-µm telecommunication band by using an SFG comb. We demonstrated continuous frequency measurement without interruption for over 10 hours. The measurement scheme can be easily extended to any IR wavelength region (longer than 1.5 µm) by employing a suitable SFG crystal.

The results of frequency measurements of three different lasers show that the acetylene-stabilized laser is a reliable optical frequency standard for wavelength-division multiplexing (WDM) applications. We suggest adjusting the value and the uncertainty of the CIPM recommendation based on these new measurement results.

Acknowledgments

The authors are grateful to Y. Kobayashi, T. R. Schibli, K. Minoshima and R. Holzwarth for their help and fruitful discussions. The authors are also grateful to S. Hatano and O. Takano from Neoark Corp. for their cooperation in commercializing the acetylene-stabilized laser. This research is conducted as part of the program “Promotion of Leading Research” in Special Coordination Funds for Promoting Science and Technology by the Ministry of Education, Culture, Sports, Science and Technology, Japan.

References and Links

1.

M. de Labachelerie, K. Nakagawa, Y. Awaji, and M. Ohtsu, “High-frequency-stability laser at 1.5 µm using Doppler-free molecular lines,” Opt. Lett. 20, 572–574 (1995). [CrossRef]

2.

K. Nakagawa, M. de Labachelerie, Y. Awaji, and M. Kourogi, “Accurate optical frequency atlas of the 1.5-µm bands of acetylene,” J. Opt. Soc. Am. B 13, 2708–2714 (1996). [CrossRef]

3.

A. Onae, K. Minoshima, J. Yoda, K. Nakagawa, A. Yamaguchi, M. Kourogi, K. Imai, and B. Widiyatomoko, “Toward an accurate frequency standard at 1.5 µm based on the acetylene overtone band transition,” IEEE Trans. Instrum. Meas. 48, 563–566 (1999). [CrossRef]

4.

A. Onae, T. Ikegami, K. Sugiyama, F-L. Hong, K. Minoshima, H. Matsumoto, K. Nakagawa, M. Yoshida, and S. Harada, “Optical frequency link between an acetylene stabilized laser at 1542 nm and an Rb stabilized laser at 778 nm using a two-colour mode-locked fiber laser,” Opt. Commun. 183, 181–187 (2000). [CrossRef]

5.

T. J. Quinn, “Practical realization of the definition of the metre, including recommended radiations of the optical frequency standards (2001),” Metrologia 40, 103–133 (2003). [CrossRef]

6.

Th. Udem, J. Reichert, R. Holzwarth, and T. W. Haensch, “Absolute optical frequency measurement of the Cesium D1 line with a mode-locked laser,” Phys. Rev. Lett. 82, 3568–3571 (1999). [CrossRef]

7.

D. J. Jones, S. A. Diddams, J. K. Ranka, A. Stentz, R. S. Windeler, J. L. Hall, and S. T. Cundiff, “Carrier-envelope phase control of femtosecond mode-locked lasers and direct optical frequency synthesis,” Science 288, 635–639 (2000). [CrossRef] [PubMed]

8.

F. -L. Hong, A. Onae, J. Jiang, R. Guo, H. Inaba, K. Minoshima, T. R. Schibli, H. Matsumoto, and K. Nakagawa, “Absolute frequency measurement of an acetylene-stabilized laser at 1542 nm,” Opt. Lett. 28, 2324–2326 (2003). [CrossRef] [PubMed]

9.

C. S. Edwards, H. S. Margolis, G. P. Barwood, S. N. Lea, P. Gill, G. Huang, and W. R. C. Rowley, “Absolute frequency measurement of a 1.5-µm acetylene standard by use of a combined frequency chain and femtosecond comb,” Opt. Lett. 29, 566–568 (2004). [CrossRef] [PubMed]

10.

A. Czajkowski, J. E. Bernard, A. A. Madej, and R. S. Windeler, “Absolute frequency measurement of acetylene transitions in the region of 1540 nm,” Appl. Phys. B 79, 45–50 (2004). [CrossRef]

11.

J. C. Knight, “Photonic crystal fibers,” Nature 424, 847–851 (2003). [CrossRef] [PubMed]

12.

Ronald Holzwarth, Max Planck Inst Quantum Opt, D-85748 Garching, and Menlo Systems GmbH, D-82152 Munich, Germany (personal communication, 2004).

13.

T. R. Schibli, K. Minoshima, F.-L. Hong, H. Inaba, A. Onae, H. Matsumoto, I. Hartl, and M. E. Fermann, “Frequency metrology with a turnkey all-fiber system,” Opt. Lett. 29, 2467–2469 (2004). [CrossRef] [PubMed]

14.

T. M. Ramond, S. A. Diddams, L. Hollberg, and A. Bartels, “Phase-coherent link from optical to microwave frequencies by means of the broadband continuum from a 1-GHz Ti:sapphire femtosecond oscillator,” Opt. Lett. 27, 1842–1844 (2002). [CrossRef]

15.

A. Amy-Klein, A. Goncharov, C. Daussy, C. Grain, O. Lopez, G. Santarelli, and C. Chardonnet, “Absolute frequency measurement in the 28-THz spectral region with a femtosecond laser comb and a long-distance optical link to a primary standard,” Appl. Phys. B 78, 25–30 (2004). [CrossRef]

16.

O. D. Mücke, O. Kuzucu, F. N. C. Wong, E. P. Ippen, F. X. Kaertner, S. M. Foreman, D. J. Jones, L.-S. Ma, J. L. Hall, and J. Ye, “Experimental implementation of optical clockwork without carrier-envelope phase control,” Opt. Lett. 29, 2806–2808 (2004). [CrossRef] [PubMed]

17.

M. Zimmermann, Ch. Gohle, R. Holzwarth, Th. Udem, and T. Haensch, “Optical clockwork with an offset-free difference-frequency comb: accuracy of sum-and difference-frequency generation,” Opt. Lett. 29, 310–312 (2004). [CrossRef] [PubMed]

18.

F.-L. Hong, J. Ishikawa, Y. Zhang, R. Guo, A. Onae, and H. Matsumoto, “Frequency reproducibility of an iodine-stabilized Nd:YAG laser at 532 nm,” Opt. Commun. 235, 377–385 (2004). [CrossRef]

19.

A. Onae, K. Okumura, F.-L. Hong, and H. Matsumoto, “Accurate frequency atlas of 1.5 µm band of acetylene measured by a mode-locked fiber laser,” presented at the 2004 Conference on Precision Electromagnetic Measurements, London, United Kingdom, 27 June-2 July, 2004.

OCIS Codes
(120.3940) Instrumentation, measurement, and metrology : Metrology
(320.7110) Ultrafast optics : Ultrafast nonlinear optics

ToC Category:
Research Papers

History
Original Manuscript: January 21, 2005
Revised Manuscript: March 4, 2005
Published: March 21, 2005

Citation
Jie Jiang, Atsushi Onae, Hirokazu Matsumoto, and Feng-Lei Hong, "Frequency measurement of acetylene-stabilized lasers using a femtosecond optical comb without carrier-envelope offset frequency control," Opt. Express 13, 1958-1965 (2005)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-6-1958


Sort:  Journal  |  Reset  

References

  1. M. de Labachelerie, K. Nakagawa, Y. Awaji, M. Ohtsu, �??High-frequency-stability laser at 1.5 μm using Doppler-free molecular lines,�?? Opt. Lett. 20, 572-574 (1995) [CrossRef]
  2. K. Nakagawa, M. de Labachelerie, Y. Awaji, M. Kourogi, �??Accurate optical frequency atlas of the 1.5-μm bands of acetylene,�?? J. Opt. Soc. Am. B 13, 2708-2714 (1996) [CrossRef]
  3. A. Onae, K. Minoshima, J. Yoda, K. Nakagawa, A. Yamaguchi, M. Kourogi, K. Imai, and B. Widiyatomoko, �??Toward an accurate frequency standard at 1.5 μm based on the acetylene overtone band transition,�?? IEEE Trans. Instrum. Meas. 48, 563-566 (1999) [CrossRef]
  4. A. Onae, T. Ikegami, K. Sugiyama, F-L. Hong, K. Minoshima, H. Matsumoto, K. Nakagawa, M. Yoshida, S. Harada, �??Optical frequency link between an acetylene stabilized laser at 1542 nm and an Rb stabilized laser at 778 nm using a two-colour mode-locked fiber laser,�?? Opt. Commun. 183, 181-187 (2000) [CrossRef]
  5. T. J. Quinn, �??Practical realization of the definition of the metre, including recommended radiations of the optical frequency standards (2001),�?? Metrologia 40, 103-133 (2003) [CrossRef]
  6. Th. Udem, J. Reichert, R. Holzwarth, T. W. Haensch, �??Absolute optical frequency measurement of the Cesium D1 line with a mode-locked laser,�?? Phys. Rev. Lett. 82, 3568-3571 (1999) [CrossRef]
  7. D. J. Jones, S. A. Diddams, J. K. Ranka, A. Stentz, R. S. Windeler, J. L. Hall, S. T. Cundiff, �??Carrier-envelope phase control of femtosecond mode-locked lasers and direct optical frequency synthesis,�?? Science 288, 635-639 (2000) [CrossRef] [PubMed]
  8. F. -L. Hong, A. Onae, J. Jiang, R. Guo, H. Inaba, K. Minoshima, T. R. Schibli, H. Matsumoto, K. Nakagawa, �??Absolute frequency measurement of an acetylene-stabilized laser at 1542 nm,�?? Opt. Lett. 28, 2324-2326 (2003) [CrossRef] [PubMed]
  9. C. S. Edwards, H. S. Margolis, G. P. Barwood, S. N. Lea, P. Gill, G. Huang, W. R. C. Rowley, �??Absolute frequency measurement of a 1.5-μm acetylene standard by use of a combined frequency chain and femtosecond comb,�?? Opt. Lett. 29, 566-568 (2004) [CrossRef] [PubMed]
  10. A. Czajkowski, J. E. Bernard, A. A. Madej, R. S. Windeler, �??Absolute frequency measurement of acetylene transitions in the region of 1540 nm,�?? Appl. Phys. B 79, 45-50 (2004) [CrossRef]
  11. J. C. Knight, �??Photonic crystal fibers,�?? Nature 424, 847-851 (2003) [CrossRef] [PubMed]
  12. Ronald Holzwarth, Max Planck Inst Quantum Opt, D-85748 Garching, and Menlo Systems GmbH, D-82152 Munich, Germany (personal communication, 2004).
  13. T. R. Schibli, K. Minoshima, F.-L. Hong, H. Inaba, A. Onae, H. Matsumoto, I. Hartl, M. E. Fermann, �??Frequency metrology with a turnkey all-fiber system,�?? Opt. Lett. 29, 2467-2469 (2004) [CrossRef] [PubMed]
  14. T. M. Ramond, S. A. Diddams, L. Hollberg, A. Bartels, �??Phase-coherent link from optical to microwave frequencies by means of the broadband continuum from a 1-GHz Ti:sapphire femtosecond oscillator,�?? Opt. Lett. 27, 1842-1844 (2002) [CrossRef]
  15. A. Amy-Klein, A. Goncharov, C. Daussy, C. Grain, O. Lopez, G. Santarelli, C. Chardonnet, �??Absolute frequency measurement in the 28-THz spectral region with a femtosecond laser comb and a long-distance optical link to a primary standard,�?? Appl. Phys. B 78, 25-30 (2004) [CrossRef]
  16. O. D. Mücke, O. Kuzucu, F. N. C. Wong, E. P. Ippen, F. X. Kaertner, S. M. Foreman, D. J. Jones, L.-S. Ma, J. L. Hall, J. Ye, �??Experimental implementation of optical clockwork without carrier-envelope phase control,�?? Opt. Lett. 29, 2806-2808 (2004) [CrossRef] [PubMed]
  17. M. Zimmermann, Ch. Gohle, R. Holzwarth, Th. Udem, T. Haensch, �??Optical clockwork with an offset-free difference-frequency comb: accuracy of sum- and difference-frequency generation,�?? Opt. Lett. 29, 310-312 (2004) [CrossRef] [PubMed]
  18. F.-L. Hong, J. Ishikawa, Y. Zhang, R. Guo, A. Onae, H. Matsumoto, �??Frequency reproducibility of an iodine-stabilized Nd:YAG laser at 532 nm,�?? Opt. Commun. 235, 377-385 (2004) [CrossRef]
  19. A. Onae, K. Okumura, F.-L. Hong, H. Matsumoto, �??Accurate frequency atlas of 1.5 μm band of acetylene measured by a mode-locked fiber laser,�?? presented at the 2004 Conference on Precision Electromagnetic Measurements, London, United Kingdom, 27 June �?? 2 July, 2004

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1. Fig. 2. Fig. 3.
 
Fig. 4.
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited