Optical signal degradation study in fixed human skin using confocal microscopy and higher-harmonic optical microscopy

Tsung-Han Tsai, Shih-Peng Tai, Wen-Jeng Lee, Hsin-Yi Huang, Yi-Hua Liao, and Chi-Kuang Sun  »View Author Affiliations


Optics Express, Vol. 14, Issue 2, pp. 749-758 (2006)


View Full Text Article

Acrobat PDF (906 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations
  • Export Citation/Save Click for help

Abstract

Confocal and nonlinear optical microscopies have been applied for dermatological studies because of their capability to provide sub-surface three-dimensional images with sub-µm spatial resolutions. Optical signal degradation as the imaging plane being moved toward deeper regions in skin specimens is the key factor that limits the observation depth for the laser scanning based linear or nonlinear imaging modalities. In this article, we studied the signal degradation in fixed human skin specimens using reflection confocal microscopy and higher-harmonic optical microscopy based on a Cr:forsterite femtosecond laser centered at 1230-nm. By analyzing the optical properties through these linear and nonlinear imaging modalities, we found that the optical signal degradation in the studied human skin specimen is dominated by the distortion of the point spread function.

© 2006 Optical Society of America

1. Introduction

Optical imaging and microscopy such as optical coherence tomography (OCT) [1–3

1. G. J. Tearney, M. E. Brezinski, B. E. Bouma, S. A. Boppart, C. Pitris, J. F. Southern, and J. G. Fujimoto, “In vivo endoscopic optical biopsy with optical coherent coherent tomography,” Science 276, 2037–2039 (1997). [CrossRef] [PubMed]

], confocal microscopy [4

4. M. Minsky , Microscopy apparatus. U.S. Patent03013467 (1957).

,5

5. C. J. R. Sheppard and D. M. Shotton , Confocal Laser Scanning Microscopy (BIOS Scientific Publisher, Oxford, 1997).

], and nonlinear optical microscopy [6–10

6. P. T. C. So, H. Kim, and I. E. Kochevar, “Two photon deep tissue ex vivo imaging of mouse dermal and subcutaneous structures,” Opt. Express 3, 339–350 (1998). [CrossRef] [PubMed]

] have been widely applied for dermatological studies because of their high penetration and three-dimensional (3D) imaging capability with high spatial resolutions. OCT can provide high penetration using 1.3-μm wavelength light [1

1. G. J. Tearney, M. E. Brezinski, B. E. Bouma, S. A. Boppart, C. Pitris, J. F. Southern, and J. G. Fujimoto, “In vivo endoscopic optical biopsy with optical coherent coherent tomography,” Science 276, 2037–2039 (1997). [CrossRef] [PubMed]

] or a sub-μm axial resolution with a 725-nm broadband light source [3

3. B. Povazay, K. Bizheva, A. Unterhuber, B. Hermann, H. Sattmann, A. F. Fercher, W. Drexler, A. Apolonski, W. J. Wadsworth, J. C. Knight, P. St. J. Russell, M. Vetterlein, and E. Scherzer, ”Submicron axial resolution optical coherence tomography,” Opt. Lett. 27, 1800–1802 (2002). [CrossRef]

]. Compared with OCT, laser scanning confocal and nonlinear microscopies provide much higher lateral resolution while the sectioned two-dimensional images are in the plane parallel to the surface, in contrast to the cross-sectional (tangential) images provided by OCT. Reflection confocal microscopy has provided a window into living skin for basic and clinical research. Confocal reflectance images of normal skin correlate well with images from conventional histology, with contrast being provided by refractive index differences of subcellular structures such as cell layers and nuclear details in the epidermis, collagen, and microcirculation in the dermis [11–13

11. B. Selkin, M. Rajadhyaksha, S. González, and R. G. Langley RG, “In vivo confocal microscopy in dermatology,” Dermatol. Clin. 19369–77, ix–x (2001).

]. It was also found that pigmented skin and pigmented lesions are especially suitable for laser scanning confocal microscopy because melanin functions as a natural source of contrast [12

12. M. Rajadhyaksha, M. Grossman, D. Esterowitz, R. H. Webb, and R. R. Anderson, “In vivo confocal scanning laser microscopy of human skin: melanin provides strong contrast,” J. Invest. Dermatol. 104, 946–952 (1995). [CrossRef] [PubMed]

]. Two-photon fluorescence microscopy [14

14. W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248, 73–76 (1990). [CrossRef] [PubMed]

] of skin based on 780-nm femtosecond light also provides high resolution imaging from the skin surface through the epidermal-dermal junction [7

7. B. R. Masters and P. T. C. So, “Confocal microscopy and multi-photon excitation microscopy of human skin in vivo,” Opt. Express 8, 2–10 (2001). [CrossRef] [PubMed]

]. Recently we demonstrated that higher-harmonic optical microscopy (HOM) based on a 1230-nm femtosecond light source could provide sub-micrometer-resolution deep-tissue biopsy images of skin without the use of fluorescence or exogenous markers [9

9. C.-K. Sun, C.-C. Chen, S.-W. Chu, T.-H. Tsai, Y.-C. Chen, and B.-L. Lin, “Multi-harmonic generation biopsy of skin,” Opt. Lett. 28, 2488–2490 (2003). [CrossRef] [PubMed]

,10

10. S.-P. Tai, T.-H. Tsai, W.-J. Lee, D.-B. Shieh, Y.-H. Liao, H.-Y. Huang, K. Y.-J. Zhang, H.-L. Liu, and C.-K. Sun, “Optical biopsy of fixed human skin with backward-collected optical harmonics signals,” Opt. Express 13, 8231–8242 (2005). [CrossRef] [PubMed]

], which could also be considered as a powerful tool for dermatological studies.

Optical signal degradation as the imaging plane being moved toward deeper regions in skin specimens is the key factor that limits the observation depth for the laser scanning based linear or nonlinear imaging modality mentioned above. The signal degradation in biological tissues might probably be owing to the excitation light attenuation, the distortion and broadening of the 3D excitation point spread function (PSF), and the broadening of the femtosecond excitation light pulses [15

15. V. R. Daria, C. Saloma, and S. Kawata, “Excitation with a focused, pulsed optical beam in scattering media: diffraction effects,” Appl. Opt. 39, 5244–5255 (2000). [CrossRef]

,16

16. M. Gu, X. Gan, A. Kisteman, and M. G. Xu, “Comparison of penetration depth between two-photon excitation and single-photon excitation in imaging through turbid tissue media,” Appl. Phys. Lett. 77, 1551–1553 (2000). [CrossRef]

]. The refractive index differences in tissues, which are significant for lenses with high numerical apertures (NA), cause not only aberration effects that distort the PSF profiles [17

17. A. Egner and S. Hell, “Equivalence of the Huygens-Fresnel and Debye approach for the calculation of high aperture pointspread functions in the presence of refractive index mismatch,” J. Microsc. 193, 244–249 (1999). [CrossRef]

], but are also responsible to move part of the incident light away from the focal spot, which can be considered as part of the scattering loss. Due to the combination of scattering and absorption, the average power delivered at the focal plane also attenuates with increasing depth, as predicted by the Beer’s law. While the broadening of the femtosecond pulsewidth will severely reduce the generated nonlinear signals, the reduced average excitation power and the distorted PSF would reduce both the linear and nonlinear optical signal intensities, smooth the image details with poorer imaging resolutions, and degrade the relative contrast of the generated images.

In this article, we studied the signal degradation of linear and nonlinear imaging modalities in fixed human skin specimens using reflection confocal microscopy and HOM based on a Cr:forsterite femtosecond laser centered at 1230-nm. By minimizing the combined effects of scattering and absorption of common tissue constituents such as water, melanin, and hemoglobin, 1.3-μm light is known to provide the deepest penetration capability in live biological specimens [1

1. G. J. Tearney, M. E. Brezinski, B. E. Bouma, S. A. Boppart, C. Pitris, J. F. Southern, and J. G. Fujimoto, “In vivo endoscopic optical biopsy with optical coherent coherent tomography,” Science 276, 2037–2039 (1997). [CrossRef] [PubMed]

,18–20

18. R. R. Anderson and J. A. Parrish, “The optics of human skin,” J. Invest. Dermatol. 77, 13–19 (1981). [CrossRef] [PubMed]

]. By analyzing the optical properties through reflection confocal (RC), second harmonic generation (SHG), and third harmonic generation (THG) modalities including spatial resolutions and measured signal power at different imaging depths in fixed human skin specimens, we show that the optical signal degradation in the studied human skin specimens with a light source in the optical penetration window is dominated by the distortion induced broadening of the point spread function, rather than light attenuation or femtosecond pulse-broadening.

2. Materials and methods

The study of the signal degradation in fixed human skin specimen was performed with a femtosecond Cr:forsterite laser centered at 1230-nm with a 140-fs pulse width at a 110-MHz repetition rate, using a harmonic laser scanning microscope combining RC microscopy (that could be considered as the fundamental harmonic mode) with higher-harmonic modalities, as shown in Fig. 1. The 1230 nm infrared (IR) laser beam was first shaped by a telescope and then directed into a modified beam scanning system (Olympus Fluoview 300) and a microscope (Olympus BX-51) with an IR water-immersion objective (LUMplanFL/IR 60×/NA 0.9). The RC signals were collected in the reflection geometry by use of the same objective. The reflection 1230 nm signal passed through the same scanner and a confocal pinhole with the size of 0.25-AU (1-AU=1.22λ/NA) [21

21. P. N. Prasad, Introduction to Biophotonics (John Wiley & Sons, Hoboken, 2003).

] and was detected by an IR-sensitive photodetector (PDA400, Thorlabs). With a momentum conservation law, most of the SHG and THG photons are with a forward propagating characteristic. We collected the forward transmitted SHG (615 nm light) and THG (410 nm light) signals by a 1.4-NA achromatic visible condenser, divided them by a chromatic beam splitter, and detected them by two separate photomultiplier tubes with 410- and 615-nm narrowband interference filters in front. Average illumination power of 90-mW was applied to the sample surface during the study.

The fixed integument samples were removed from the back of a seventy-year-old female patient and preserved in formalin at 4□. The experimental protocols were approved the National Taiwan University Hospital Institutional Review Board (NTUH-IRB). The thickness of the samples under observation was on the order of 1 mm.

Fig. 1. Set up of the harmonic laser scanning microscope combining the reflection confocal with the higher-harmonic generation modalities. BF: barrier filter; BS: beamspliter; DB: dichroic beamsplitter; PMT: photomultiplier tube; GM: galvanometric mirrors; PD: photodetector; VIC: V-I converter circuit; PH: confocal pinhole.

3. Experimental results and discussion

Figure 2 and Fig. 3 show examples of the cross-sectional images taken at depths of 90- and 150-μm from the studied skin surface by the harmonic laser scanning microscope. With the signals oriented from refractive index differences, RC can reflect subcellular structures such as cell layers and nuclear details in the epidermis, as well as collagens and microcirculation in the dermis [12

12. M. Rajadhyaksha, M. Grossman, D. Esterowitz, R. H. Webb, and R. R. Anderson, “In vivo confocal scanning laser microscopy of human skin: melanin provides strong contrast,” J. Invest. Dermatol. 104, 946–952 (1995). [CrossRef] [PubMed]

,13

13. J. A. Veiro and P. G. Cummins, “Imaging of skin epidermis from various origins using confocal laser scanning microscopy,” Dermatology 189: 16–22 (1994). [CrossRef] [PubMed]

] while the general histological structures in both epidermis and dermal layers can also be identified through the THG modality due to its sensitivity to local optical inhomogeneities [22–24

22. S.-W. Chu, I-H. Chen, T.-M. Liu, P. C. Cheng, C.-K. Sun, and B.-L. Lin, “Multimodal nonlinear spectral microscopy based on a femtosecond Cr:forsterite laser,” Opt. Lett. 26, 1909–1921 (2001). [CrossRef]

], including refractive index (or susceptibility χ(1)) and third-order susceptibility χ(3) differences which usually occurs at the same bio-locations. Fig. 2(a) and Fig. 2(B) show cross-sectional RC and THG images in the stratum basal layer on top of the dermo-epidermal junction. The highly-correlated images of the densely packed basal cells resolved by the RC and THG signals clearly indicate that the RC and THG modalities have the similar sources of contrast in human skin, i.e., optical susceptibility discontinuities. In the dermal layer, as shown Fig. 3(a), RC signals can still pick up fine structures including wavy collagen fibril distribution in the connective tissues with relatively less brightness than the RC image shown in Fig. 2(a). THG signals, on the other hand, decreases more significantly deep into the dermal layer (see Fig. 3(b)), compared with the linear-optics based RC signals [16

16. M. Gu, X. Gan, A. Kisteman, and M. G. Xu, “Comparison of penetration depth between two-photon excitation and single-photon excitation in imaging through turbid tissue media,” Appl. Phys. Lett. 77, 1551–1553 (2000). [CrossRef]

]. With strong optical second harmonic generation in collagen fibers [8

8. Y. Guo, P. P. Ho, H. Savage, D. Harris, P. Sacks, S. Schantz, F. Liu, N. Zhadin, and R. R. Alfano, “Second-harmonic tomography of tissues,” Opt. Lett. 22, 1323–1325 (1997). [CrossRef]

,25

25. P. J. Campagnola, A. C Millard, M. Terasaki, P. E. Hoppe, C. J. Malone, and W. A. Mohler, “Three-dimensional high-resolution second-harmonic generation imaging of endogenous structural proteins in biological tissues,” Biophy. J. 81, 493–508 (2002). [CrossRef]

,26

26. S.-W. Chu, I-S. Chen, T.-M. Liu, C.-K. Sun, S.-P. Lee, B.-L. Lin, P.-C. Cheng, M.-X. Kuo, D.-J. Lin, and H.-L. Liu, “Nonlinear Bio-photonic Crystal Effects Revealed with Multi-modal Nonlinear Microscopy,” J. Microsc. 208, Part 3, 190–200 (2002). [CrossRef] [PubMed]

], the SHG modality provides the unique capability to reflect the distribution of connected tissues in the dermal layer [10

10. S.-P. Tai, T.-H. Tsai, W.-J. Lee, D.-B. Shieh, Y.-H. Liao, H.-Y. Huang, K. Y.-J. Zhang, H.-L. Liu, and C.-K. Sun, “Optical biopsy of fixed human skin with backward-collected optical harmonics signals,” Opt. Express 13, 8231–8242 (2005). [CrossRef] [PubMed]

] (Fig. 3(c)). The uneven distribution of SHG signals in Fig. 2(c) is due to the Rete ridge structures of the dermal papillae [27

27. H. Elias, J. E. Pauly, and E. R. Burns, Histology and Human Microanatomy, 4th Edition, Piccin Medical Books (Wiley, New York, 1978).

] inter-digitated with the downward projections of the epidermis.

Fig. 2. Cross-sectional (a) RC, (b) THG, (c) SHG, and (d) combined images of human skin taken at a depth of 90-|j,m from the skin surface around the dermo-epidermal junction. Images size: 120×120μm.
Fig. 3. Cross-sectional (a) RC, (b) THG, (c) SHG, and (d) combined images of human skin taken at a depth of 150-μm from the skin surface in the dermal layer. Images size: 120×120μm.

Fig. 4. Lateral resolutions of RC (black square), SHG (red circle), and THG (green triangle) images versus depth in the studied human skin sample.

Table 1. Formula for the focal beam diameter and the resolutions of different imaging modalities.

table-icon
View This Table
Fig. 5. Focal diameter degradation with depth in the studied human skin sample derived from the lateral resolution of the RC (black square), SHG (red circle), and THG (green triangle) images respectively.

The incident light peak intensity at the focal plane is the critical factor to the generation of signals, especially to nonlinear SHG and THG signals. Because of nonlinearity, the generated SHG intensity depends on the square power of the incident light intensity, while the generated THG intensity depends on the third power of the incident light intensity. When the incident light peak intensity decreases, the generated SHG and THG intensities will also severely decay. Due to distortion of the point spread functions as the observation depth move to deeper regions, the focal area also get larger that causes the reduction of the proportional RC signals that passing through the confocal pinhole. Therefore, the measured RC signal power would be inversely proportional to the focal area, that is, inversely quadratic to the focal diameter [32

32. H. Beyer , Handbuch der Mikroskopie, 2nd Edition (VEB Verlag Technik, Berlin, 1985).

]. Since the generated THG intensity depends on the third power of the incident light intensity and the measured signal intensity depends on inversely quadratic of the focal diameter of the incident light, the THG signal intensity depends on the inverse 6th power of the focal diameter of the incident light. The THG signal power thus depends on inversely quartic (inverse 4th power) of the focal diameter of the incident light. Figure 6 shows the collected RC signal power versus focal diameter taken at different sample depths. The values of the focal diameters at different sample depths are based those shown in Fig. 5. We did not analyze the power dependency of the SHG signal because the intensity of the forward SHG signal is strongly dependent on the collagen fibril thickness [10

10. S.-P. Tai, T.-H. Tsai, W.-J. Lee, D.-B. Shieh, Y.-H. Liao, H.-Y. Huang, K. Y.-J. Zhang, H.-L. Liu, and C.-K. Sun, “Optical biopsy of fixed human skin with backward-collected optical harmonics signals,” Opt. Express 13, 8231–8242 (2005). [CrossRef] [PubMed]

,34

34. J. Mertz and L. Moreaux, “Second-harmonic generation by focused excitation of inhomogeneously distributed scatterers,” Opt. Comm. 196, 325–330 (2001). [CrossRef]

] while the thickness of the collagen fibrils in the dermis layer is very uneven. The distribution of the SHG signal is also very different from those of RC and THG signals. Collagen-induced SHG is only present in the dermal layer [10

10. S.-P. Tai, T.-H. Tsai, W.-J. Lee, D.-B. Shieh, Y.-H. Liao, H.-Y. Huang, K. Y.-J. Zhang, H.-L. Liu, and C.-K. Sun, “Optical biopsy of fixed human skin with backward-collected optical harmonics signals,” Opt. Express 13, 8231–8242 (2005). [CrossRef] [PubMed]

] and its uneven signal distribution is further modified by the Rete ridge structures of the dermal papillae [27

27. H. Elias, J. E. Pauly, and E. R. Burns, Histology and Human Microanatomy, 4th Edition, Piccin Medical Books (Wiley, New York, 1978).

] inter-digitated with the downward projections of the epidermis. Based on a chi-square-based fitting method [35

35. W.H. Press, S.A. Teukolsky, W.T. Vetterling, and B.P. Flannery, Numerical Recipes in C (Cambridge Univ. Press, Cambridge, 1992).

] (solid black line in Fig. 6), the dependency of the RC signal power on the focal diameter broadening as the focal plane moving into the sample can be found to be roughly negatively quadratic with a slope of -2.24, confirming that the degradation of signal power is mainly affected by the distortion of the focused beam size in our study. The difference between slope = -2.24 and slope = -2 in Fig. 6 could be attributed to the attenuation of optical signals in the skin specimen, which is another important signal degradation mechanism. Pulse broadening will not affect the intensity of the RC signals. Based on this assumption, we could also obtain an effective attenuation coefficient of 6.3-cm-1 (for 1230-nm light) based on analyzing the difference of these two slopes. The effective attenuation coefficient a for 1230-nm light was calculated according to the Beer’s law, i.e., α= (lnP1-lnP2)/2Δd, where P1 is the power calculated from the solid line with slope = -2 at the depth of Δd, P2 is the power measured from the skin specimen at the depth of Δd, and the factor of 2 is due to the roundtrip measurement in the reflection geometry.

Fig. 6. RC signal power (square) versus focal diameter degradation with different depths in the studied human skin sample. The black solid line is the best fit with a slope = -2.24 and the gray solid line is the slope = -2 fitting. Note that both axes are log-scaled.

Figure 7 shows the dependency of the THG signal power on the focal diameter degradation as the focal plane moving into the sample, where the solid black line is the best fit obtained based on a chi-square-based fitting method [34

34. J. Mertz and L. Moreaux, “Second-harmonic generation by focused excitation of inhomogeneously distributed scatterers,” Opt. Comm. 196, 325–330 (2001). [CrossRef]

] with a slope of -4.5, confirming that the degradation of THG signal power is also mainly affected by the distortion of the focused beam size. Compared with the ideal case with a slope of -4 where only the distortion of the point spread function is considered, the slight difference between the experimental data and the theoretical upper limit (-4.5 versus -4) could be attributed to light attenuation and the broadening of the femtosecond pulsewidth. When the imaging depth was increased with Δd, the average intensity of the 1230nm excitation light would be decreased with a factor of eαΔd, while the generated THG average intensity at the focal spot would be decreased with a factor of e3αΔd, assuming that the excitation pulsewidth was not broadened. With a forward collection geometry and an increasing excitation depth, the total propagation length for THG photons will be decreased. With a THG photon attenuation constant α’, the measured THG intensity should thus experience a degradation factor of e(3α-α’)Δd, due to the light attenuation effect. Due to stronger light attenuation at 410 nm than that at 1230 nm [18

18. R. R. Anderson and J. A. Parrish, “The optics of human skin,” J. Invest. Dermatol. 77, 13–19 (1981). [CrossRef] [PubMed]

], the signal degradation due to the light attenuation effect in forward THG study should thus be weaker than the RC study (with a factor of e2αΔd). The relatively stronger discrepancy in the THG study (-4.5 versus -4), compared with the result in the RC study (-2.24 versus -2), indicates the contribution of the femtosecond pulsewidth broadening to the observed nonlinear signal degradation. However our analysis suggests that the degradation of optical signal power versus imaging depth, for linear confocal or nonlinear imaging modalities in the studied human skin sample with a 1230-nm light in the optical penetration window, is dominated by the distortion induced focal diameter broadening rather than the light attenuation [16

16. M. Gu, X. Gan, A. Kisteman, and M. G. Xu, “Comparison of penetration depth between two-photon excitation and single-photon excitation in imaging through turbid tissue media,” Appl. Phys. Lett. 77, 1551–1553 (2000). [CrossRef]

] or dispersion induced pulse broadening. This is understandable due to the fact that the applied laser wavelength is located in the optical penetration window of the studied specimens [18

18. R. R. Anderson and J. A. Parrish, “The optics of human skin,” J. Invest. Dermatol. 77, 13–19 (1981). [CrossRef] [PubMed]

] and is close to the dispersion zero wavelengths of normal materials. Due to higher dependency of the nonlinear signal degradation on the focal diameter broadening, the nonlinear optical signal (for example THG) power will decay more rapidly while the observation plane moves deeper into the skin specimens.

Fig. 7. THG signal power (square) versus focal diameter degradation with different depths in the studied human skin sample. The solid black line is the best fit with a slope = -4.5. The solid gray line is with a slope = -4. Note that both axes are log-scaled.

4. Conclusion

In summary, we studied the optical signal degradation mechanisms in fixed human skin specimens using RC microscopy and higher-harmonic optical microscopy based on a Cr:forsterite femtosecond laser centered at 1230-nm. By analyzing the optical properties through these linear and nonlinear imaging modalities, we found that the optical signal degradation in the studied fixed human skin specimen is primarily dominated by the distortion of the point spread function, while the excitation light attenuation or dispersion induced optical pulse broadening can be considered as secondary effects. The capability of THG modality to resolve the shapes, distribution, and sizes of nuclei in human skin basal cells with high spatial resolution could also provide an imaging solution for the diagnosis of skin cancer with minimized invasion.

Acknowledgments

The authors gratefully acknowledge financial support under the National Health Research Institute (NHRI-EX94-9201EI) of Taiwan and the National Taiwan University Center for Genomic Medicine.

References and Links

1.

G. J. Tearney, M. E. Brezinski, B. E. Bouma, S. A. Boppart, C. Pitris, J. F. Southern, and J. G. Fujimoto, “In vivo endoscopic optical biopsy with optical coherent coherent tomography,” Science 276, 2037–2039 (1997). [CrossRef] [PubMed]

2.

Y. Yasuno, S. Makita, Y. Sutoh, M. Itoh, and T. Yatagai, “Birefringence imaging of human skin by polarization-sensitive spectral interferometric optical coherent tomography,” Opt. Lett. 27, 1803–1805 (2002). [CrossRef]

3.

B. Povazay, K. Bizheva, A. Unterhuber, B. Hermann, H. Sattmann, A. F. Fercher, W. Drexler, A. Apolonski, W. J. Wadsworth, J. C. Knight, P. St. J. Russell, M. Vetterlein, and E. Scherzer, ”Submicron axial resolution optical coherence tomography,” Opt. Lett. 27, 1800–1802 (2002). [CrossRef]

4.

M. Minsky , Microscopy apparatus. U.S. Patent03013467 (1957).

5.

C. J. R. Sheppard and D. M. Shotton , Confocal Laser Scanning Microscopy (BIOS Scientific Publisher, Oxford, 1997).

6.

P. T. C. So, H. Kim, and I. E. Kochevar, “Two photon deep tissue ex vivo imaging of mouse dermal and subcutaneous structures,” Opt. Express 3, 339–350 (1998). [CrossRef] [PubMed]

7.

B. R. Masters and P. T. C. So, “Confocal microscopy and multi-photon excitation microscopy of human skin in vivo,” Opt. Express 8, 2–10 (2001). [CrossRef] [PubMed]

8.

Y. Guo, P. P. Ho, H. Savage, D. Harris, P. Sacks, S. Schantz, F. Liu, N. Zhadin, and R. R. Alfano, “Second-harmonic tomography of tissues,” Opt. Lett. 22, 1323–1325 (1997). [CrossRef]

9.

C.-K. Sun, C.-C. Chen, S.-W. Chu, T.-H. Tsai, Y.-C. Chen, and B.-L. Lin, “Multi-harmonic generation biopsy of skin,” Opt. Lett. 28, 2488–2490 (2003). [CrossRef] [PubMed]

10.

S.-P. Tai, T.-H. Tsai, W.-J. Lee, D.-B. Shieh, Y.-H. Liao, H.-Y. Huang, K. Y.-J. Zhang, H.-L. Liu, and C.-K. Sun, “Optical biopsy of fixed human skin with backward-collected optical harmonics signals,” Opt. Express 13, 8231–8242 (2005). [CrossRef] [PubMed]

11.

B. Selkin, M. Rajadhyaksha, S. González, and R. G. Langley RG, “In vivo confocal microscopy in dermatology,” Dermatol. Clin. 19369–77, ix–x (2001).

12.

M. Rajadhyaksha, M. Grossman, D. Esterowitz, R. H. Webb, and R. R. Anderson, “In vivo confocal scanning laser microscopy of human skin: melanin provides strong contrast,” J. Invest. Dermatol. 104, 946–952 (1995). [CrossRef] [PubMed]

13.

J. A. Veiro and P. G. Cummins, “Imaging of skin epidermis from various origins using confocal laser scanning microscopy,” Dermatology 189: 16–22 (1994). [CrossRef] [PubMed]

14.

W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248, 73–76 (1990). [CrossRef] [PubMed]

15.

V. R. Daria, C. Saloma, and S. Kawata, “Excitation with a focused, pulsed optical beam in scattering media: diffraction effects,” Appl. Opt. 39, 5244–5255 (2000). [CrossRef]

16.

M. Gu, X. Gan, A. Kisteman, and M. G. Xu, “Comparison of penetration depth between two-photon excitation and single-photon excitation in imaging through turbid tissue media,” Appl. Phys. Lett. 77, 1551–1553 (2000). [CrossRef]

17.

A. Egner and S. Hell, “Equivalence of the Huygens-Fresnel and Debye approach for the calculation of high aperture pointspread functions in the presence of refractive index mismatch,” J. Microsc. 193, 244–249 (1999). [CrossRef]

18.

R. R. Anderson and J. A. Parrish, “The optics of human skin,” J. Invest. Dermatol. 77, 13–19 (1981). [CrossRef] [PubMed]

19.

C.-K. Sun, S.-W. Chu, S.-Y. Chen, T.-H. Tsai, T.-M. Liu, C.-Y. Lin, and H.-J. Tsai, “Higher harmonic generation microscopy for developmental biology,” J. Struct. Bio. 147, 19–30 (2004). [CrossRef]

20.

S.-W. Chu, S.-Y. Chen, T.-H. Tsai, T.-M. Liu, C.-Y. Lin, H.-J. Tsai, and C.-K. Sun, ”In vivo developmental biology study using noninvasive multi-harmonic generation microscopy,” Opt. Express 11, 3093–3099 (2003). [CrossRef] [PubMed]

21.

P. N. Prasad, Introduction to Biophotonics (John Wiley & Sons, Hoboken, 2003).

22.

S.-W. Chu, I-H. Chen, T.-M. Liu, P. C. Cheng, C.-K. Sun, and B.-L. Lin, “Multimodal nonlinear spectral microscopy based on a femtosecond Cr:forsterite laser,” Opt. Lett. 26, 1909–1921 (2001). [CrossRef]

23.

D. Yelin and Y. Silberberg, “Laser scanning third-harmonic-generation microscopy in biology,” Opt. Express 5, 169–175 (1999). [CrossRef] [PubMed]

24.

J. A. Squier, M. Muller, G. J. Brakenhoff, and K. R. Wilson, “Third harmonic generation microscopy,” Opt. Express 3, 315–324 (1999). [CrossRef]

25.

P. J. Campagnola, A. C Millard, M. Terasaki, P. E. Hoppe, C. J. Malone, and W. A. Mohler, “Three-dimensional high-resolution second-harmonic generation imaging of endogenous structural proteins in biological tissues,” Biophy. J. 81, 493–508 (2002). [CrossRef]

26.

S.-W. Chu, I-S. Chen, T.-M. Liu, C.-K. Sun, S.-P. Lee, B.-L. Lin, P.-C. Cheng, M.-X. Kuo, D.-J. Lin, and H.-L. Liu, “Nonlinear Bio-photonic Crystal Effects Revealed with Multi-modal Nonlinear Microscopy,” J. Microsc. 208, Part 3, 190–200 (2002). [CrossRef] [PubMed]

27.

H. Elias, J. E. Pauly, and E. R. Burns, Histology and Human Microanatomy, 4th Edition, Piccin Medical Books (Wiley, New York, 1978).

28.

M. Rajadhyaksha, S. González, J. M. Zavislan, R. R. Anderson, and R. H. Webb, “In vivo confocal scanning laser microscopy of human skin II: Advances in instrumentation and comparison with histology,” J. Invest. Dermatol. 113, 293–303 (1999). [CrossRef] [PubMed]

29.

S. González and Z. Tannous, “Real-time, in vivo confocal reflectance microscopy of basal cell carcinoma,” J. Am. Acad. Dermatol. 47, 869–874 (2002). [CrossRef] [PubMed]

30.

T. Yamashita, T. Kuwahara, S. González, and M. Takahashi, “Non-invasive visualization of melanin and melanocytes by reflectance-mode confocal microscopy,” J. Invest. Dermotal. 124, 235–240 (2005). [CrossRef]

31.

A. Dunn and R. Richards-Kortum, “Three-dimensional computation of light scattering from cells,” IEEE J. Sel. Top. Quantum Electron. 2, 898–905 (1996). [CrossRef]

32.

H. Beyer , Handbuch der Mikroskopie, 2nd Edition (VEB Verlag Technik, Berlin, 1985).

33.

J. Squier and M. Müller, “High resolution nonlinear microscopy: A review of sources and methods for achieving optimal imaging,” Rev. Sci. Instrum. 72, 2855–2867 (2001). [CrossRef]

34.

J. Mertz and L. Moreaux, “Second-harmonic generation by focused excitation of inhomogeneously distributed scatterers,” Opt. Comm. 196, 325–330 (2001). [CrossRef]

35.

W.H. Press, S.A. Teukolsky, W.T. Vetterling, and B.P. Flannery, Numerical Recipes in C (Cambridge Univ. Press, Cambridge, 1992).

OCIS Codes
(170.1790) Medical optics and biotechnology : Confocal microscopy
(170.1870) Medical optics and biotechnology : Dermatology
(180.5810) Microscopy : Scanning microscopy
(180.6900) Microscopy : Three-dimensional microscopy
(190.4160) Nonlinear optics : Multiharmonic generation

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: December 9, 2005
Revised Manuscript: January 17, 2006
Manuscript Accepted: January 18, 2006
Published: January 23, 2006

Virtual Issues
Vol. 1, Iss. 2 Virtual Journal for Biomedical Optics

Citation
, "Optical signal degradation study in fixed human skin using confocal microscopy and higher-harmonic optical microscopy," Opt. Express 14, 749-758 (2006)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-14-2-749


Sort:  Journal  |  Reset  

References

References are not available for this paper.

Cited By

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited