Continuum generation in a novel photonic crystal fiber for ultrahigh resolution optical coherence tomography at 800 nm and 1300 nm
Optics Express, Vol. 14, Issue 3, pp. 1145-1160 (2006)
http://dx.doi.org/10.1364/OE.14.001145
Acrobat PDF (3916 KB)
Abstract
Ultrahigh resolution optical coherence tomography (OCT) is demonstrated at 800 nm and 1300 nm using continuum generation in a single photonic crystal fiber with a parabolic dispersion profile and two closely spaced zero dispersion wavelengths. Both wavelengths are generated simultaneously by pumping the fiber with ~78 mW average power at 1064 nm in a 52 MHz, 85 fs pulse train from a compact Nd:Glass oscillator. Continuum processes result in a double peak spectrum with > 110 nm and 30 mW average power at 800 nm and > 150 nm and 48 mW at 1300 nm. OCT imaging with < 5 μm resolution in tissue at 1300 nm and < 3 μm resolution at 800 nm is demonstrated. Numerical modeling of propagation was used to predict the spectrum and can be used for further optimization to generate smooth, broad spectra for OCT applications.
© 2006 Optical Society of America
1. Introduction
D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254, 1178–1181 (1991). [CrossRef] [PubMed]
J. G. Fujimoto, “Optical coherence tomography for ultrahigh resolution in vivo imaging,” Nat Biotechnol 21, 1361–1367 (2003). [CrossRef] [PubMed]
B. Bouma, G. J. Tearney, S. A. Boppart, M. R. Hee, M. E. Brezinski, and J. G. Fujimoto, “High-resolution optical coherence tomographic imaging using a mode-locked Ti:Al2O3 laser source,” Optics Letters 20, 1486–1488 (1995). [CrossRef] [PubMed]
W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, “In vivo ultrahigh-resolution optical coherence tomography,” Optics Letters 24, 1221–1223 (1999). [CrossRef]
W. Drexler, U. Morgner, R. K. Ghanta, F. X. Körtner, J. S. Schuman, and J. G. Fujimoto, “Ultrahigh-resolution ophthalmic optical coherence tomography,” Nat Med 7, 502–507 (2001). [CrossRef] [PubMed]
A. M. Kowalevicz, T. R. Schibli, F. X. Kartner, and J. G. Fujimoto, “Ultralow-threshold Kerr-lens mode-locked Ti:Al2O3 laser,” Optics Letters 27, 2037–2039 (2002). [CrossRef]
A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, V. Yakovlev, G. Tempea, C. Schubert, E. M. Anger, P. K. Ahnelt, M. Stur, J. E. Morgan, A. Cowey, G. Jung, T. Le, and A. Stingl, “Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography,” Optics letters 28, 905–907 (2003). [CrossRef] [PubMed]
D. L. Marks, A. L. Oldenburg, J. J. Reynolds, and S. A. Boppart, “Study of an ultrahigh-numerical-aperture fiber continuum generation source for optical coherence tomography,” Optics Letters 27, 2010–2012 (2002). [CrossRef]
P. C. Wagenblast, T. H. Ko, J. G. Fujimoto, F. X. Kaertner, and U. Morgner, “Ultrahigh-resolution optical coherence tomography with a diode-pumped broadband Cr3+: LiCAF laser,” Optics Express 12, 3257–3263 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80534. [CrossRef] [PubMed]
L. Vabre, A. Dubois, and A. C. Boccara, “Thermal-light full-field optical coherence tomography,” Optics Letters 27, 530–532 (2002). [CrossRef]
A. M. Kowalevicz, T. Ko, I. Hartl, J. G. Fujimoto, M. Pollnau, and R. P. Salathe, “Ultrahigh resolution optical coherence tomography using a superluminescent light source,” Optics Express 10, 349–353 (2002), http://www.opticsinfobase.org/abstract.cfm?id=68496. [PubMed]
J. M. Schmitt, A. Knuttel, M. Yadlowsky, and M. A. Eckhaus, “Optical-coherence tomography of a dense tissue: statistics of attenuation and backscattering,” Physics in Medicine and Biology 39, 1705–1720 (1994). [CrossRef] [PubMed]
B. E. Bouma, G. J. Tearney, I. P. Bilinsky, B. Golubovic, and J. G. Fujimoto, “Self-phase-modulated Kerr-lens mode-locked Cr:forsterite laser source for optical coherence tomography,” Optics Letters 21, 1839–1841 (1996). [CrossRef] [PubMed]
P. R. Herz, Y. Chen, A. D. Aguirre, J. C. Fujimoto, H. Mashimo, J. Schmitt, A. Koski, J. Goodnow, and C. Petersen, “Ultrahigh resolution optical biopsy with endoscopic optical coherence tomography,” Optics Express 12 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80642. [CrossRef] [PubMed]
J. K. Ranka, R. S. Windeler, and A. J. Stentz, “Visible continuum generation in air-silica microstructure optical fibers with anomalous dispersion at 800 nm,” Optics Letters 25, 25–27 (2000). [CrossRef]
I. Hartl, X. D. Li, C. Chudoba, R. K. Hganta, T. H. Ko, J. G. Fujimoto, J. K. Ranka, and R. S. Windeler, “Ultrahigh-resolution optical coherence tomography using continuum generation in an air-silica microstructure optical fiber,” Optics Letters 26, 608–610 (2001). [CrossRef]
I. Hartl, X. D. Li, C. Chudoba, R. K. Hganta, T. H. Ko, J. G. Fujimoto, J. K. Ranka, and R. S. Windeler, “Ultrahigh-resolution optical coherence tomography using continuum generation in an air-silica microstructure optical fiber,” Optics Letters 26, 608–610 (2001). [CrossRef]
B. Povazay, K. Bizheva, B. Hermann, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, C. Schubert, P. K. Ahnelt, M. Mei, R. Holzwarth, W. J. Wadsworth, J. C. Knight, and P. S. Russel, “Enhanced visualization of choroidal vessels using ultrahigh resolution ophthalmic OCT at 1050 nm,” Optics Express 11, 1980–1986 (2003), http://www.opticsinfobase.org/abstract.cfm?id=74008. [CrossRef] [PubMed]
Y. Wang, J. S. Nelson, Z. Chen, B. J. Reiser, R. S. Chuck, and R. S. Windeler, “Optimal wavelength for ultrahigh-resolution optical coherence tomography,” Optics Express 11, 1411–1417 (2003), http://www.opticsinfobase.org/abstract.cfm?id=72649. [CrossRef] [PubMed]
N. Nishizawa, Y. Chen, P. Hsiung, E. P. Ippen, and J. G. Fujimoto, “Real-time, ultrahigh-resolution, optical coherence tomography with an all-fiber, femtosecond fiber laser continuum at 1.5 mu m,” Optics Letters 29, 2846–2848 (2004). [CrossRef]
T. Hori, N. Nishizawa, T. Goto, and M. Yoshida, “Wideband and nonmechanical sonogram measurement by use of an electronically controlled, wavelength-tunable, femtosecond soliton pulse,” J Opt Soc Am B 20, 2410–2417 (2003). [CrossRef]
S. Bourquin, A. D. Aguirre, I. Hartl, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bunting, and D. Kopf, “Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd : Glass laser and nonlinear fiber,” Optics Express 11, 3290–3297 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78091. [CrossRef] [PubMed]
P. R. Herz, Y. Chen, A. D. Aguirre, J. C. Fujimoto, H. Mashimo, J. Schmitt, A. Koski, J. Goodnow, and C. Petersen, “Ultrahigh resolution optical biopsy with endoscopic optical coherence tomography,” Optics Express 12 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80642. [CrossRef] [PubMed]
K. L. Corwin, N. R. Newbury, J. M. Dudley, S. Coen, S. A. Diddams, K. Weber, and R. S. Windeler, “Fundamental noise limitations to supercontinuum generation in microstructure fiber,” Physical Review Letters 90, 113904/1–4 (2003). [CrossRef] [PubMed]
K. Bizheva, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, M. Mei, R. Holzwarth, T. Hoelzenbein, V. Wacheck, and H. Pehamherger, “Compact, broad-bandwidth fiber laser for sub-2- mu m axial resolution optical coherence tomography in. The 1300-nm wavelength region,” Optics Letters 28, 707–709 (2003). [CrossRef] [PubMed]
P. L. Hsiung, Y. Chen, T. H. Ko, J. G. Fujimoto, C. J. S. de Matos, S. V. Popov, J. R. Taylor, and V. P. Gapontsev, “Optical coherence tomography using a continuous-wave, high-power, Raman continuum light source,” Optics Express 12, 5287–5295 (2004), http://www.opticsinfobase.org/abstract.cfm?id=81626. [CrossRef] [PubMed]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
S. Bourquin, A. D. Aguirre, I. Hartl, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bunting, and D. Kopf, “Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd : Glass laser and nonlinear fiber,” Optics Express 11, 3290–3297 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78091. [CrossRef] [PubMed]
2. Numerical simulations
T. Hori, N. Nishizawa, T. Goto, and M. Yoshida, “Experimental and numerical analysis of widely broadened supercontinuum generation in highly nonlinear dispersion-shifted fiber with a femtosecond pulse,” J Opt Soc Am B 21, 1969–1980 (2004). [CrossRef]
2.1 Fiber length dependence of the continuum
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
S. Coen, A. H. L. Chau, R. Leonhardt, J. D. Harvey, J. C. Knight, W. J. Wadsworth, and P. S. J. Russell, “Supercontinuum generation by stimulated Raman scattering and parametric four-wave mixing in photonic crystal fibers,” J Opt Soc Am B 19, 753–764 (2002). [CrossRef]
A. V. Husakou and J. Herrmann, “Supercontinuum generation, four-wave mixing, and fission of higher-order solitons in photonic-crystal fibers,” J Opt Soc Am B 19, 2171–2182 (2002). [CrossRef]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
M. H. Frosz, P. Falk, and O. Bang, “The role of the second zero-dispersion wavelength in generation of supercontinua and bright-bright soliton-pairs across the zero-dispersion wavelength,” Optics Express 13, 6181–6192 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85282. [CrossRef] [PubMed]
M. H. Frosz, P. Falk, and O. Bang, “The role of the second zero-dispersion wavelength in generation of supercontinua and bright-bright soliton-pairs across the zero-dispersion wavelength,” Optics Express 13, 6181–6192 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85282. [CrossRef] [PubMed]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
2.2 Pump wavelength dependence of the continuum
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
S. R. Chinn and E. A. Swanson, “Blindness Limitations in Optical Coherence Domain Reflectometry,” Electronics Letters 29, 2025–2027 (1993). [CrossRef]
3. Experimental measurements of the continuum
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
K. L. Corwin, N. R. Newbury, J. M. Dudley, S. Coen, S. A. Diddams, K. Weber, and R. S. Windeler, “Fundamental noise limitations to supercontinuum generation in microstructure fiber,” Physical Review Letters 90, 113904/1–4 (2003). [CrossRef] [PubMed]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
S. Bourquin, A. D. Aguirre, I. Hartl, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bunting, and D. Kopf, “Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd : Glass laser and nonlinear fiber,” Optics Express 11, 3290–3297 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78091. [CrossRef] [PubMed]
4. OCT imaging results and discussion
4.1 OCT system setup
4.2 System performance evaluation
W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, “In vivo ultrahigh-resolution optical coherence tomography,” Optics Letters 24, 1221–1223 (1999). [CrossRef]
I. Hartl, X. D. Li, C. Chudoba, R. K. Hganta, T. H. Ko, J. G. Fujimoto, J. K. Ranka, and R. S. Windeler, “Ultrahigh-resolution optical coherence tomography using continuum generation in an air-silica microstructure optical fiber,” Optics Letters 26, 608–610 (2001). [CrossRef]
4.3 OCT imaging results
I. Hartl, X. D. Li, C. Chudoba, R. K. Hganta, T. H. Ko, J. G. Fujimoto, J. K. Ranka, and R. S. Windeler, “Ultrahigh-resolution optical coherence tomography using continuum generation in an air-silica microstructure optical fiber,” Optics Letters 26, 608–610 (2001). [CrossRef]
T. H. Ko, D. C. Adler, J. G. Fujimoto, D. Mamedov, V. Prokhorov, V. Shidlovski, and S. Yakubovich, “Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source,” Optics Express 12, 2112–2119 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79925. [CrossRef] [PubMed]
P. R. Herz, Y. Chen, A. D. Aguirre, J. C. Fujimoto, H. Mashimo, J. Schmitt, A. Koski, J. Goodnow, and C. Petersen, “Ultrahigh resolution optical biopsy with endoscopic optical coherence tomography,” Optics Express 12 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80642. [CrossRef] [PubMed]
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
M. H. Frosz, P. Falk, and O. Bang, “The role of the second zero-dispersion wavelength in generation of supercontinua and bright-bright soliton-pairs across the zero-dispersion wavelength,” Optics Express 13, 6181–6192 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85282. [CrossRef] [PubMed]
P. Falk, M. H. Frosz, and O. Bang, “Supercontinuum generation in a photonic crystal fiber with two zero-dispersion wavelengths tapered to normal dispersion at all wavelengths,” Optics Express 13, 7535–7540 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85486. [CrossRef] [PubMed]
T. Hori, J. Takayanagi, N. Nishizawa, and T. Goto, “Flatly broadened, wideband and low noise supercontinuum generation in highly nonlinear hybrid fiber,” Optics Express 12, 317–324 (2004), http://www.opticsinfobase.org/abstract.cfm?id=78593. [CrossRef] [PubMed]
P. Falk, M. H. Frosz, and O. Bang, “Supercontinuum generation in a photonic crystal fiber with two zero-dispersion wavelengths tapered to normal dispersion at all wavelengths,” Optics Express 13, 7535–7540 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85486. [CrossRef] [PubMed]
M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by highspeed spectral optical coherence tomography,” Optics Letters 28, 1745–1747 (2003). [CrossRef] [PubMed]
4. Conclusion
Acknowledgments
References and links
D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254, 1178–1181 (1991). [CrossRef] [PubMed] | |
J. G. Fujimoto, “Optical coherence tomography for ultrahigh resolution in vivo imaging,” Nat Biotechnol 21, 1361–1367 (2003). [CrossRef] [PubMed] | |
B. Bouma, G. J. Tearney, S. A. Boppart, M. R. Hee, M. E. Brezinski, and J. G. Fujimoto, “High-resolution optical coherence tomographic imaging using a mode-locked Ti:Al2O3 laser source,” Optics Letters 20, 1486–1488 (1995). [CrossRef] [PubMed] | |
W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, “In vivo ultrahigh-resolution optical coherence tomography,” Optics Letters 24, 1221–1223 (1999). [CrossRef] | |
W. Drexler, U. Morgner, R. K. Ghanta, F. X. Körtner, J. S. Schuman, and J. G. Fujimoto, “Ultrahigh-resolution ophthalmic optical coherence tomography,” Nat Med 7, 502–507 (2001). [CrossRef] [PubMed] | |
A. M. Kowalevicz, T. R. Schibli, F. X. Kartner, and J. G. Fujimoto, “Ultralow-threshold Kerr-lens mode-locked Ti:Al2O3 laser,” Optics Letters 27, 2037–2039 (2002). [CrossRef] | |
A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, V. Yakovlev, G. Tempea, C. Schubert, E. M. Anger, P. K. Ahnelt, M. Stur, J. E. Morgan, A. Cowey, G. Jung, T. Le, and A. Stingl, “Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography,” Optics letters 28, 905–907 (2003). [CrossRef] [PubMed] | |
D. L. Marks, A. L. Oldenburg, J. J. Reynolds, and S. A. Boppart, “Study of an ultrahigh-numerical-aperture fiber continuum generation source for optical coherence tomography,” Optics Letters 27, 2010–2012 (2002). [CrossRef] | |
P. C. Wagenblast, T. H. Ko, J. G. Fujimoto, F. X. Kaertner, and U. Morgner, “Ultrahigh-resolution optical coherence tomography with a diode-pumped broadband Cr3+: LiCAF laser,” Optics Express 12, 3257–3263 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80534. [CrossRef] [PubMed] | |
L. Vabre, A. Dubois, and A. C. Boccara, “Thermal-light full-field optical coherence tomography,” Optics Letters 27, 530–532 (2002). [CrossRef] | |
A. M. Kowalevicz, T. Ko, I. Hartl, J. G. Fujimoto, M. Pollnau, and R. P. Salathe, “Ultrahigh resolution optical coherence tomography using a superluminescent light source,” Optics Express 10, 349–353 (2002), http://www.opticsinfobase.org/abstract.cfm?id=68496. [PubMed] | |
C. Grivas, T. C. May-Smith, D. P. Shepherd, R. W. Eason, M. Pollnau, and M. Jelinek, “Broadband single-transverse-mode fluorescence sources based on ribs fabricated in pulsed laser deposited Ti : sapphire waveguides,” Appl Phys a-Mater 79, 1195–1198 (2004). | |
T. H. Ko, D. C. Adler, J. G. Fujimoto, D. Mamedov, V. Prokhorov, V. Shidlovski, and S. Yakubovich, “Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source,” Optics Express 12, 2112–2119 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79925. [CrossRef] [PubMed] | |
J. M. Schmitt, A. Knuttel, M. Yadlowsky, and M. A. Eckhaus, “Optical-coherence tomography of a dense tissue: statistics of attenuation and backscattering,” Physics in Medicine and Biology 39, 1705–1720 (1994). [CrossRef] [PubMed] | |
B. E. Bouma, G. J. Tearney, I. P. Bilinsky, B. Golubovic, and J. G. Fujimoto, “Self-phase-modulated Kerr-lens mode-locked Cr:forsterite laser source for optical coherence tomography,” Optics Letters 21, 1839–1841 (1996). [CrossRef] [PubMed] | |
P. R. Herz, Y. Chen, A. D. Aguirre, J. C. Fujimoto, H. Mashimo, J. Schmitt, A. Koski, J. Goodnow, and C. Petersen, “Ultrahigh resolution optical biopsy with endoscopic optical coherence tomography,” Optics Express 12 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80642. [CrossRef] [PubMed] | |
J. K. Ranka, R. S. Windeler, and A. J. Stentz, “Visible continuum generation in air-silica microstructure optical fibers with anomalous dispersion at 800 nm,” Optics Letters 25, 25–27 (2000). [CrossRef] | |
I. Hartl, X. D. Li, C. Chudoba, R. K. Hganta, T. H. Ko, J. G. Fujimoto, J. K. Ranka, and R. S. Windeler, “Ultrahigh-resolution optical coherence tomography using continuum generation in an air-silica microstructure optical fiber,” Optics Letters 26, 608–610 (2001). [CrossRef] | |
B. Povazay, K. Bizheva, A. Unterhuber, B. Hermann, H. Sattmann, A. F. Fercher, W. Drexler, A. Apolonski, W. J. Wadsworth, J. C. Knight, P. S. J. Russell, M. Vetterlein, and E. Scherzer, “Submicrometer axial resolution optical coherence tomography,” Optics Letters 27, 1800–1802 (2002). [CrossRef] | |
W. Drexler, “Ultrahigh-resolution optical coherence tomography,” Journal of Biomedical Optics 9, 47–74 (2004). [CrossRef] [PubMed] | |
B. Povazay, K. Bizheva, B. Hermann, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, C. Schubert, P. K. Ahnelt, M. Mei, R. Holzwarth, W. J. Wadsworth, J. C. Knight, and P. S. Russel, “Enhanced visualization of choroidal vessels using ultrahigh resolution ophthalmic OCT at 1050 nm,” Optics Express 11, 1980–1986 (2003), http://www.opticsinfobase.org/abstract.cfm?id=74008. [CrossRef] [PubMed] | |
Y. Wang, Y. Zhao, J. S. Nelson, Z. Chen, and R. S. Windeler, “Ultrahigh-resolution optical coherence tomography by broadband continuum generation from a photonic crystal fiber,” Optics Letters 28, 182–184 (2003). [CrossRef] [PubMed] | |
S. Bourquin, A. D. Aguirre, I. Hartl, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bunting, and D. Kopf, “Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd : Glass laser and nonlinear fiber,” Optics Express 11, 3290–3297 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78091. [CrossRef] [PubMed] | |
H. Lim, Y. Jiang, Y. Wang, Y. C. Huang, Z. Chen, and F. W. Wise, “Ultrahigh-resolution optical coherence tomography with a fiber laser source at 1 microm,” Opt Lett 30, 1171–1173 (2005). [CrossRef] [PubMed] | |
Y. Wang, J. S. Nelson, Z. Chen, B. J. Reiser, R. S. Chuck, and R. S. Windeler, “Optimal wavelength for ultrahigh-resolution optical coherence tomography,” Optics Express 11, 1411–1417 (2003), http://www.opticsinfobase.org/abstract.cfm?id=72649. [CrossRef] [PubMed] | |
N. Nishizawa, Y. Chen, P. Hsiung, E. P. Ippen, and J. G. Fujimoto, “Real-time, ultrahigh-resolution, optical coherence tomography with an all-fiber, femtosecond fiber laser continuum at 1.5 mu m,” Optics Letters 29, 2846–2848 (2004). [CrossRef] | |
T. Hori, N. Nishizawa, T. Goto, and M. Yoshida, “Wideband and nonmechanical sonogram measurement by use of an electronically controlled, wavelength-tunable, femtosecond soliton pulse,” J Opt Soc Am B 20, 2410–2417 (2003). [CrossRef] | |
K. L. Corwin, N. R. Newbury, J. M. Dudley, S. Coen, S. A. Diddams, K. Weber, and R. S. Windeler, “Fundamental noise limitations to supercontinuum generation in microstructure fiber,” Physical Review Letters 90, 113904/1–4 (2003). [CrossRef] [PubMed] | |
N. R. Newbury, B. R. Washburn, K. L. Corwin, and R. S. Windeler, “Noise amplification during supercontinuum generation in microstructure fiber,” Optics Letters 28, 944–946 (2003). [CrossRef] [PubMed] | |
A. Apolonski, B. Povazay, A. Unterhuber, W. Drexler, W. J. Wadsworth, J. C. Knight, and P. S. Russell, “Spectral shaping of supercontinuum in a cobweb photonic-crystal fiber with sub-20-fs pulses,” J Opt Soc Am B 19, 2165–2170 (2002). [CrossRef] | |
K. Bizheva, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, M. Mei, R. Holzwarth, T. Hoelzenbein, V. Wacheck, and H. Pehamherger, “Compact, broad-bandwidth fiber laser for sub-2- mu m axial resolution optical coherence tomography in. The 1300-nm wavelength region,” Optics Letters 28, 707–709 (2003). [CrossRef] [PubMed] | |
P. L. Hsiung, Y. Chen, T. H. Ko, J. G. Fujimoto, C. J. S. de Matos, S. V. Popov, J. R. Taylor, and V. P. Gapontsev, “Optical coherence tomography using a continuous-wave, high-power, Raman continuum light source,” Optics Express 12, 5287–5295 (2004), http://www.opticsinfobase.org/abstract.cfm?id=81626. [CrossRef] [PubMed] | |
K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, Y. Kristiansen, K. P. Hansen, and J. J. Larsen, “Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths,” Optics Express 12, 1045–1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed] | |
T. Hori, N. Nishizawa, T. Goto, and M. Yoshida, “Experimental and numerical analysis of widely broadened supercontinuum generation in highly nonlinear dispersion-shifted fiber with a femtosecond pulse,” J Opt Soc Am B 21, 1969–1980 (2004). [CrossRef] | |
G. P. Agrawal, Nonlinear Fiber Optics (Academic Press, London, 2001). | |
A. V. Husakou and J. Herrmann, “Supercontinuum generation of higher-order solitons by fission in photonic crystal fibers,” Physical Review Letters 8720, art. no.-203901 (2001). | |
A. L. Gaeta, “Nonlinear propagation and continuum generation in microstructured optical fibers,” Optics Letters 27, 924–926 (2002). [CrossRef] | |
B. R. Washburn, S. E. Ralph, and R. S. Windeler, “Ultrashort pulse propagation in air-silica microstructure fiber,” Optics Express 10, 575–580 (2002), http://www.opticsinfobase.org/abstract.cfm?id=69331. [PubMed] | |
G. Genty, M. Lehtonen, H. Ludvigsen, J. Broeng, and M. Kaivola, “Spectral broadening of femtosecond pulses into continuum radiation in microstructured fibers,” Optics Express 10, 1083–1098 (2002), http://www.opticsinfobase.org/abstract.cfm?id=70205. [PubMed] | |
S. Coen, A. H. L. Chau, R. Leonhardt, J. D. Harvey, J. C. Knight, W. J. Wadsworth, and P. S. J. Russell, “Supercontinuum generation by stimulated Raman scattering and parametric four-wave mixing in photonic crystal fibers,” J Opt Soc Am B 19, 753–764 (2002). [CrossRef] | |
A. V. Husakou and J. Herrmann, “Supercontinuum generation, four-wave mixing, and fission of higher-order solitons in photonic-crystal fibers,” J Opt Soc Am B 19, 2171–2182 (2002). [CrossRef] | |
M. H. Frosz, P. Falk, and O. Bang, “The role of the second zero-dispersion wavelength in generation of supercontinua and bright-bright soliton-pairs across the zero-dispersion wavelength,” Optics Express 13, 6181–6192 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85282. [CrossRef] [PubMed] | |
S. R. Chinn and E. A. Swanson, “Blindness Limitations in Optical Coherence Domain Reflectometry,” Electronics Letters 29, 2025–2027 (1993). [CrossRef] | |
P. Falk, M. H. Frosz, and O. Bang, “Supercontinuum generation in a photonic crystal fiber with two zero-dispersion wavelengths tapered to normal dispersion at all wavelengths,” Optics Express 13, 7535–7540 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85486. [CrossRef] [PubMed] | |
T. Hori, J. Takayanagi, N. Nishizawa, and T. Goto, “Flatly broadened, wideband and low noise supercontinuum generation in highly nonlinear hybrid fiber,” Optics Express 12, 317–324 (2004), http://www.opticsinfobase.org/abstract.cfm?id=78593. [CrossRef] [PubMed] | |
M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by highspeed spectral optical coherence tomography,” Optics Letters 28, 1745–1747 (2003). [CrossRef] [PubMed] | |
M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Optics Express 12, 2404–2422 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80147. [CrossRef] [PubMed] | |
N. Nassif, B. Cense, B. H. Park, S. H. Yun, T. C. Chen, B. E. Bouma, G. J. Tearney, and J. F. de Boer, “In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography,” Opt Lett 29, 480–482 (2004). [CrossRef] [PubMed] | |
S. H. Yun, G. J. Tearney, B. E. Bouma, B. H. Park, and J. F. de Boer, “High-speed spectral-domain optical coherence tomography at 1.3 mu m wavelength,” Optics Express 11, 3598–3604 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78225. [CrossRef] [PubMed] |
OCIS Codes
(060.7140) Fiber optics and optical communications : Ultrafast processes in fibers
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(170.4500) Medical optics and biotechnology : Optical coherence tomography
(320.7090) Ultrafast optics : Ultrafast lasers
ToC Category:
Medical Optics and Biotechnology
History
Original Manuscript: December 14, 2005
Revised Manuscript: January 30, 2006
Manuscript Accepted: February 1, 2006
Published: February 6, 2006
Virtual Issues
Vol. 1, Iss. 3 Virtual Journal for Biomedical Optics
Citation
Aaron Aguirre, Norihiko Nishizawa, James Fujimoto, Wolfgang Seitz, Max Lederer, and Daniel Kopf, "Continuum generation in a novel photonic crystal fiber for ultrahigh resolution optical coherence tomography at 800 nm and 1300 nm," Opt. Express 14, 1145-1160 (2006)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-14-3-1145
Sort: Year | Journal | Reset
References
- D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, "Optical coherence tomography," Science 254, 1178-1181 (1991). [CrossRef] [PubMed]
- J. G. Fujimoto, "Optical coherence tomography for ultrahigh resolution in vivo imaging," Nat Biotechnol 21, 1361-1367 (2003). [CrossRef] [PubMed]
- B. Bouma, G. J. Tearney, S. A. Boppart, M. R. Hee, M. E. Brezinski, and J. G. Fujimoto, "High-resolution optical coherence tomographic imaging using a mode-locked Ti:Al2O3 laser source," Opt. Lett. 20, 1486-1488 (1995). [CrossRef] [PubMed]
- W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, "In vivo ultrahigh-resolution optical coherence tomography," Opt. Lett. 24, 1221-1223 (1999). [CrossRef]
- W. Drexler, U. Morgner, R. K. Ghanta, F. X. Kärtner, J. S. Schuman, and J. G. Fujimoto, "Ultrahigh-resolution ophthalmic optical coherence tomography," Nat Med 7, 502-507 (2001). [CrossRef] [PubMed]
- A. M. Kowalevicz, T. R. Schibli, F. X. Kartner, and J. G. Fujimoto, "Ultralow-threshold Kerr-lens mode-locked Ti:Al2O3 laser," Opt. Lett. 27, 2037-2039 (2002). [CrossRef]
- A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, V. Yakovlev, G. Tempea, C. Schubert, E. M. Anger, P. K. Ahnelt, M. Stur, J. E. Morgan, A. Cowey, G. Jung, T. Le, and A. Stingl, "Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography," Opt. Lett. 28, 905-907 (2003). [CrossRef] [PubMed]
- D. L. Marks, A. L. Oldenburg, J. J. Reynolds, and S. A. Boppart, "Study of an ultrahigh-numerical-aperture fiber continuum generation source for optical coherence tomography," Opt. Lett. 27, 2010-2012 (2002). [CrossRef]
- P. C. Wagenblast, T. H. Ko, J. G. Fujimoto, F. X. Kaertner, and U. Morgner, "Ultrahigh-resolution optical coherence tomography with a diode-pumped broadband Cr3+: LiCAF laser," Opt. Express 12, 3257-3263 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80534. [CrossRef] [PubMed]
- L. Vabre, A. Dubois, and A. C. Boccara, "Thermal-light full-field optical coherence tomography," Opt. Lett. 27, 530-532 (2002). [CrossRef]
- A. M. Kowalevicz, T. Ko, I. Hartl, J. G. Fujimoto, M. Pollnau, and R. P. Salathe, "Ultrahigh resolution optical coherence tomography using a superluminescent light source," Opt. Express 10, 349-353 (2002), http://www.opticsinfobase.org/abstract.cfm?id=68496. [PubMed]
- C. Grivas, T. C. May-Smith, D. P. Shepherd, R. W. Eason, M. Pollnau, and M. Jelinek, "Broadband single-transverse-mode fluorescence sources based on ribs fabricated in pulsed laser deposited Ti : sapphire waveguides," Appl. Phys.A-Mater 79, 1195-1198 (2004).
- T. H. Ko, D. C. Adler, J. G. Fujimoto, D. Mamedov, V. Prokhorov, V. Shidlovski, and S. Yakubovich, "Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source," Opt. Express 12, 2112-2119 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79925. [CrossRef] [PubMed]
- J. M. Schmitt, A. Knuttel, M. Yadlowsky, and M. A. Eckhaus, "Optical-coherence tomography of a dense tissue: statistics of attenuation and backscattering," Physics in Medicine and Biology 39, 1705-1720 (1994). [CrossRef] [PubMed]
- B. E. Bouma, G. J. Tearney, I. P. Bilinsky, B. Golubovic, and J. G. Fujimoto, "Self-phase-modulated Kerr-lens mode-locked Cr:forsterite laser source for optical coherence tomography," Opt. Lett. 21, 1839-1841 (1996). [CrossRef] [PubMed]
- P. R. Herz, Y. Chen, A. D. Aguirre, J. C. Fujimoto, H. Mashimo, J. Schmitt, A. Koski, J. Goodnow, and C. Petersen, "Ultrahigh resolution optical biopsy with endoscopic optical coherence tomography," Opt. Express 12 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80642. [CrossRef] [PubMed]
- J. K. Ranka, R. S. Windeler, and A. J. Stentz, "Visible continuum generation in air-silica microstructure optical fibers with anomalous dispersion at 800 nm," Opt. Lett. 25, 25-27 (2000). [CrossRef]
- I. Hartl, X. D. Li, C. Chudoba, R. K. Hganta, T. H. Ko, J. G. Fujimoto, J. K. Ranka, and R. S. Windeler, "Ultrahigh-resolution optical coherence tomography using continuum generation in an air-silica microstructure optical fiber," Opt. Lett. 26, 608-610 (2001). [CrossRef]
- B. Povazay, K. Bizheva, A. Unterhuber, B. Hermann, H. Sattmann, A. F. Fercher, W. Drexler, A. Apolonski, W. J. Wadsworth, J. C. Knight, P. S. J. Russell, M. Vetterlein, and E. Scherzer, "Submicrometer axial resolution optical coherence tomography," Opt. Lett. 27, 1800-1802 (2002). [CrossRef]
- W. Drexler, "Ultrahigh-resolution optical coherence tomography," J. Biomedical Opt. 9, 47-74 (2004). [CrossRef] [PubMed]
- B. Povazay, K. Bizheva, B. Hermann, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, C. Schubert, P. K. Ahnelt, M. Mei, R. Holzwarth, W. J. Wadsworth, J. C. Knight, and P. S. Russel, "Enhanced visualization of choroidal vessels using ultrahigh resolution ophthalmic OCT at 1050 nm," Opt. Express 11, 1980-1986 (2003), http://www.opticsinfobase.org/abstract.cfm?id=74008. [CrossRef] [PubMed]
- Y. Wang, Y. Zhao, J. S. Nelson, Z. Chen, and R. S. Windeler, "Ultrahigh-resolution optical coherence tomography by broadband continuum generation from a photonic crystal fiber," Opt. Lett. 28, 182-184 (2003). [CrossRef] [PubMed]
- S. Bourquin, A. D. Aguirre, I. Hartl, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bunting, and D. Kopf, "Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd : Glass laser and nonlinear fiber," Opt. Express 11, 3290-3297 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78091. [CrossRef] [PubMed]
- H. Lim, Y. Jiang, Y. Wang, Y. C. Huang, Z. Chen, and F. W. Wise, "Ultrahigh-resolution optical coherence tomography with a fiber laser source at 1 microm," Opt Lett 30, 1171-1173 (2005). [CrossRef] [PubMed]
- Y. Wang, J. S. Nelson, Z. Chen, B. J. Reiser, R. S. Chuck, and R. S. Windeler, "Optimal wavelength for ultrahigh-resolution optical coherence tomography," Opt. Express 11, 1411-1417 (2003), http://www.opticsinfobase.org/abstract.cfm?id=72649. [CrossRef] [PubMed]
- N. Nishizawa, Y. Chen, P. Hsiung, E. P. Ippen, and J. G. Fujimoto, "Real-time, ultrahigh-resolution, optical coherence tomography with an all-fiber, femtosecond fiber laser continuum at 1.5 mu m," Opt. Lett. 29, 2846-2848 (2004). [CrossRef]
- T. Hori, N. Nishizawa, T. Goto, and M. Yoshida, "Wideband and nonmechanical sonogram measurement by use of an electronically controlled, wavelength-tunable, femtosecond soliton pulse," J. Opt. Soc. Am. B 20, 2410-2417 (2003). [CrossRef]
- K. L. Corwin, N. R. Newbury, J. M. Dudley, S. Coen, S. A. Diddams, K. Weber, and R. S. Windeler, "Fundamental noise limitations to supercontinuum generation in microstructure fiber," Phys. Rev. Lett. 90, 113904 (2003). [CrossRef] [PubMed]
- N. R. Newbury, B. R. Washburn, K. L. Corwin, and R. S. Windeler, "Noise amplification during supercontinuum generation in microstructure fiber," Opt. Lett. 28, 944-946 (2003). [CrossRef] [PubMed]
- A. Apolonski, B. Povazay, A. Unterhuber, W. Drexler, W. J. Wadsworth, J. C. Knight, and P. S. Russell, "Spectral shaping of supercontinuum in a cobweb photonic-crystal fiber with sub-20-fs pulses," J. Opt. Soc. Am. B 19, 2165-2170 (2002). [CrossRef]
- K. Bizheva, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, M. Mei, R. Holzwarth, T. Hoelzenbein, V. Wacheck, and H. Pehamherger, "Compact, broad-bandwidth fiber laser for sub-2- mu m axial resolution optical coherence tomography in. The 1300-nm wavelength region," Opt. Lett. 28, 707-709 (2003). [CrossRef] [PubMed]
- P. L. Hsiung, Y. Chen, T. H. Ko, J. G. Fujimoto, C. J. S. de Matos, S. V. Popov, J. R. Taylor, and V. P. Gapontsev, "Optical coherence tomography using a continuous-wave, high-power, Raman continuum light source," Opt. Express 12, 5287-5295 (2004), http://www.opticsinfobase.org/abstract.cfm?id=81626. [CrossRef] [PubMed]
- K. M. Hilligsoe, T. V. Andersen, H. N. Paulsen, C. K. Nielsen, K. Molmer, S. Keiding, R. Kristiansen, K. P. Hansen, and J. J. Larsen, "Supercontinuum generation in a photonic crystal fiber with two zero dispersion wavelengths," Opt. Express 12, 1045-1054 (2004), http://www.opticsinfobase.org/abstract.cfm?id=79252. [CrossRef] [PubMed]
- T. Hori, N. Nishizawa, T. Goto, and M. Yoshida, "Experimental and numerical analysis of widely broadened supercontinuum generation in highly nonlinear dispersion-shifted fiber with a femtosecond pulse," J. Opt. Soc. Am. B 21, 1969-1980 (2004). [CrossRef]
- G. P. Agrawal, Nonlinear Fiber Optics (Academic Press, London, 2001).
- A. V. Husakou, and J. Herrmann, "Supercontinuum generation of higher-order solitons by fission in photonic crystal fibers," Phys. Rev. Lett. 8720, (2001).
- A. L. Gaeta, "Nonlinear propagation and continuum generation in microstructured optical fibers," Opt. Lett. 27, 924-926 (2002). [CrossRef]
- B. R. Washburn, S. E. Ralph, and R. S. Windeler, "Ultrashort pulse propagation in air-silica microstructure fiber," Opt. Express 10, 575-580 (2002), http://www.opticsinfobase.org/abstract.cfm?id=69331. [PubMed]
- G. Genty, M. Lehtonen, H. Ludvigsen, J. Broeng, and M. Kaivola, "Spectral broadening of femtosecond pulses into continuum radiation in microstructured fibers," Opt. Express 10, 1083-1098 (2002), http://www.opticsinfobase.org/abstract.cfm?id=70205. [PubMed]
- S. Coen, A. H. L. Chau, R. Leonhardt, J. D. Harvey, J. C. Knight, W. J. Wadsworth, and P. S. J. Russell, "Supercontinuum generation by stimulated Raman scattering and parametric four-wave mixing in photonic crystal fibers," J. Opt. Soc. Am. B 19, 753-764 (2002). [CrossRef]
- A. V. Husakou, and J. Herrmann, "Supercontinuum generation, four-wave mixing, and fission of higher-order solitons in photonic-crystal fibers," J. Opt. Soc. Am. B 19, 2171-2182 (2002). [CrossRef]
- M. H. Frosz, P. Falk, and O. Bang, "The role of the second zero-dispersion wavelength in generation of supercontinua and bright-bright soliton-pairs across the zero-dispersion wavelength," Opt. Express 13, 6181-6192 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85282. [CrossRef] [PubMed]
- S. R. Chinn, and E. A. Swanson, "Blindness Limitations in Optical Coherence Domain Reflectometry," Electronics Letters 29, 2025-2027 (1993). [CrossRef]
- P. Falk, M. H. Frosz, and O. Bang, "Supercontinuum generation in a photonic crystal fiber with two zero-dispersion wavelengths tapered to normal dispersion at all wavelengths," Opt. Express 13, 7535-7540 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85486. [CrossRef] [PubMed]
- T. Hori, J. Takayanagi, N. Nishizawa, and T. Goto, "Flatly broadened, wideband and low noise supercontinuum generation in highly nonlinear hybrid fiber," Opt. Express 12, 317-324 (2004), http://www.opticsinfobase.org/abstract.cfm?id=78593. [CrossRef] [PubMed]
- M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, "Real-time in vivo imaging by high-speed spectral optical coherence tomography," Opt. Lett. 28, 1745-1747 (2003). [CrossRef] [PubMed]
- M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, "Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation," Opt. Express 12, 2404-2422 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80147. [CrossRef] [PubMed]
- N. Nassif, B. Cense, B. H. Park, S. H. Yun, T. C. Chen, B. E. Bouma, G. J. Tearney, and J. F. de Boer, "In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography," Opt Lett 29, 480-482 (2004). [CrossRef] [PubMed]
- S. H. Yun, G. J. Tearney, B. E. Bouma, B. H. Park, and J. F. de Boer, "High-speed spectral-domain optical coherence tomography at 1.3 mu m wavelength," Opt. Express 11, 3598-3604 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78225. [CrossRef] [PubMed]
Cited By |
OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.





OSA is a member of 