Abstract
Line-field spectral domain optical coherence tomography (LF-SDOCT) has been developed for very high-speed three-dimensional (3D) retinal imaging. By this technique, the A-line rate significantly improved to 823,200 A-lines/s for single frame imaging and 51,500 A-lines/s for continues frame imaging. The frame rate at continues frame imaging is 201 fps. This 3D acquisition speed is more than two fold higher acquisition speed than the standard flying spot SD-OCT. In this paper, the integration time of the camera was optimized for the in vivo retinal measurement and the degradation of the lateral resolution due to the ocular aberrations was suppressed by introducing the pupil stop. Owing to an optimal integration time, the motion artifact can be significantly suppressed. Also a pupil stop was employed in order to enhance the contrast of the OCT image for the effect of ocular aberrations. The in vivo 3D retinal imaging with 256 cross-sectional images (256 A-lines/image) was successfully performed in 1.3 seconds, corresponding to 0.8 volume/s. The maximum on-axis system sensitivity was measured to be 89.4 dB at a depth of 112 μm with an axial resolution of 7.4 μm in tissue. It is shown that LF-SDOCT might have a sensitivity advantage in comparison to the flying spot SD-OCT in the ultra high-speed acquisition mode.
© 2007 Optical Society of America
1. Introduction
D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography, Science 254, 1178–1181 (1991). [CrossRef] [PubMed]
G. Hausler and M. W. Lindner, ““Coherence radar”and“spectral radar”-new tools for dermatological diagnosis,” J. Biomed. Opt. 3, 21–31 (1998). [CrossRef]
R. Leitgeb, C. Hitzenberger, and A. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express 11, 889–894 (2003). [CrossRef] [PubMed]
J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28, 2067–2069 (2003). [CrossRef] [PubMed]
N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12, 367–376 (2004). [CrossRef] [PubMed]
N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12, 367–376 (2004). [CrossRef] [PubMed]
M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett. 28, 1745–1747 (2003). [CrossRef] [PubMed]
M. Wojtkowski, V. Srinivasan, J. G. Fujimoto, T. Ko, J. S. Schuman, A. Kowalczyk, and J. S. Duker, “Three-dimensional Retinal Imaging with High-Speed Ultrahigh-Resolution Optical Coherence Tomography,” Ophthalmology 112, 1734–1746 (2005). [CrossRef] [PubMed]
M. Mujat, R. Chan, B. Cense, B. Park, Chulmin Joo, T. Akkin, T. Chen, and J. de Boer, “Retinal nerve fiber layer thickness map determined from optical coherence tomography images,” Opt. Express 13, 9480–9491 (2005). [CrossRef] [PubMed]
S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14, 7821–7840 (2006). [CrossRef] [PubMed]
S. Jiao, C. Wu, R. W. Knighton, G. Gregori, and C. A. Puliafito, “Registration of high-density cross sectional images to the fundus image in spectral-domain ophthalmic optical coherence tomography,” Opt. Express 14, 3368–3376 (2006). [CrossRef] [PubMed]
S. Bourquin, P. Seitz, and R. P. Salathe, “Optical coherence topography based on a two-dimensional smart detector array,” Opt. Lett. 26, 512–514 (2001). [CrossRef]
M. Laubscher, M. Ducros, B. Karamata, T. Lasser, and R. Salathe, “Video-rate three-dimensional optical coherence tomography,” Opt. Express 10, 429–435 (2002). [PubMed]
M. Lebec, L. Blanchot, H. Saint-Jalmes, E. Beaurepaire, and A. C. Boccara, “Full-field optical coherence microscopy,” Opt. Lett. 23, 244–246 (1998). [CrossRef]
M. Akiba, K. P. Chan, and N. Tanno, “Full-field optical coherence tomography by two-dimensional heterodyne detection with a pair of CCD camera,” Opt. Lett. 28, 816–818 (2003). [CrossRef] [PubMed]
W. Y. Oh, B. E. Bouma, N. Iftimia, R. Yelin, and G. J. Tearney “Spectrally-modulated full-field optical coherence microscopy for ultrahigh-resolution endoscopic imaging,” Opt. Express 14, 8675–8684 (2006). [CrossRef] [PubMed]
S. Bourquin, P. Seitz, and R. P. Salathe, “Optical coherence topography based on a two-dimensional smart detector array,” Opt. Lett. 26, 512–514 (2001). [CrossRef]
M. Laubscher, M. Ducros, B. Karamata, T. Lasser, and R. Salathe, “Video-rate three-dimensional optical coherence tomography,” Opt. Express 10, 429–435 (2002). [PubMed]
M. Lebec, L. Blanchot, H. Saint-Jalmes, E. Beaurepaire, and A. C. Boccara, “Full-field optical coherence microscopy,” Opt. Lett. 23, 244–246 (1998). [CrossRef]
M. Akiba, K. P. Chan, and N. Tanno, “Full-field optical coherence tomography by two-dimensional heterodyne detection with a pair of CCD camera,” Opt. Lett. 28, 816–818 (2003). [CrossRef] [PubMed]
W. Y. Oh, B. E. Bouma, N. Iftimia, R. Yelin, and G. J. Tearney “Spectrally-modulated full-field optical coherence microscopy for ultrahigh-resolution endoscopic imaging,” Opt. Express 14, 8675–8684 (2006). [CrossRef] [PubMed]
I. Zeylikovich, A. Gilerson, and R. R. Alfano, “Nonmechanical grating-generated scanning coherence microscopy,” Opt. Lett. 23, 1797–1799 (1998). [CrossRef]
Y. Watanabe, K. Yamada, and M. Sato, “In vivo nonmechanical scanning grating-generated optical coherence tomography using an InGaAs digital camera,” Opt. Commun. 261, 376–380 (2006). [CrossRef]
Y. Watanabe, K. Yamada, and M. Sato, “Three-dimensional imaging by ultrahigh-speed axial-lateral parallel time domain optical coherence tomography,”Opt. Express 14, 5201–5209 (2006). [CrossRef] [PubMed]
Y. Watanabe, K. Yamada, and M. Sato, “Three-dimensional imaging by ultrahigh-speed axial-lateral parallel time domain optical coherence tomography,”Opt. Express 14, 5201–5209 (2006). [CrossRef] [PubMed]
A. Zuluaga and R. Richards-Kortum, “Spatially resolved spectral interferometry for determination of subsurface structure,” Opt. Lett. 24, 519–521 (1999). [CrossRef]
B. Grajciar, M. Pircher, A. Fercher, and R. Leitgeb, “Parallel Fourier domain optical coherence tomography for in vivo measurement of the human eye,” Opt. Express 13, 1131–1137 (2005). [CrossRef] [PubMed]
Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13, 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?id=84362. [CrossRef] [PubMed]
T. Endo, Y. Yasuno, S. Makita, M. Itoh, and T. Yatagai, “Profilometry with line-field Fourier-domain interferom-etry,” Opt. Express 13, 695–701 (2005). [CrossRef] [PubMed]
Y. Yasuno, T. Endo, S. Makita, G. Aoki, M. Itoh, and T. Yatagai, “Three-dimensional line-field Fourier domain optical coherence tomography for in vivo dermatological investigation,” J. Biomed. Opt. 11, 014014–014020 (2006). [CrossRef] [PubMed]
2. System
2.1. Optical set-up
2.2. Synchronization
2.3. Post processing
M. Takeda, H. Ina, and S. Kobayashi,“Fourier-transform method of fringe-pattern analysis for computer-based topography and interferometry,”J. Opt. Soc. Am. 72, 156–160 (1982). [CrossRef]
Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K. -P. Chan, M. Itoh, and T. Yatagai,“Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments,”Opt. Express 13, 10652–10664 (2005). [CrossRef] [PubMed]
R. Leitgeb, W. Drexler, A. Unterhuber, B. Hermann, T. Bajraszewski, T. Le, A. Stingl, and A. Fercher, “Ultrahigh resolution Fourier domain optical coherence tomography,” Opt. Express 12, 2156–2165 (2004). [CrossRef] [PubMed]
3. Optimizations for retinal measurement
3.1. Integration time
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12, 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80320. [CrossRef] [PubMed]
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Pulsed-source and swept-source spectral-domain optical coherence tomography with reduced motion artifacts,”Opt. Express 12, 5614–5624 (2004). [CrossRef] [PubMed]
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12, 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80320. [CrossRef] [PubMed]
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12, 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80320. [CrossRef] [PubMed]
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Pulsed-source and swept-source spectral-domain optical coherence tomography with reduced motion artifacts,”Opt. Express 12, 5614–5624 (2004). [CrossRef] [PubMed]
S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14, 7821–7840 (2006). [CrossRef] [PubMed]
3.2. Ocular aberrations
Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13, 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?id=84362. [CrossRef] [PubMed]
Zawadzki, S. Jones, S. Olivier, M. Zhao, B. Bower, J. Izatt, S. Choi, S. Laut, and J. Werner, “Adaptive-optics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging,” Opt. Express 13, 8532–8546 (2005). [CrossRef] [PubMed]
B. Hermann, E. J. Fernandez, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, P. M. Prieto, and P. Artal, “Adaptive-optics ultrahigh-resolution optical coherence tomography,” Opt. Lett. 29, 2142–2144 (2004). [CrossRef] [PubMed]
E. J. Fernandez, B. Povazay, B. Hermann, A. Unterhuber, H. Sattmann, P. M. Prieto, R. Leitgeb, P. Ahnelt, P. Artal, and W. Drexler, “Three dimensional adaptive optics ultrahigh-resolution optical coherence tomography using a liquid crystal spatial light modulator,” Vision Res. 45, 3432–3444 (2005). [CrossRef] [PubMed]
E. J. Fernandez and W. Drexler, “Influence of ocular chromatic aberration and pupil size on transverse resolution in ophthalmic adaptive optics optical coherence tomography,”Opt. Express 13, 8184–8197 (2005). [CrossRef] [PubMed]
E. J. Fernandez, A. Unterhuber, B. Povazay, B. Hermann, P. Artal, and W. Drexler, “Chromatic aberration correction of the human eye for retinal imaging in the near infrared,” Opt. Express 14, 6213–6225 (2006), http://www.opticsinfobase.org/abstract.cfm?URI=oe-14-13-6213 [CrossRef] [PubMed]
Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13, 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?id=84362. [CrossRef] [PubMed]
W. J. Donnelly and A. Roorda, “Optimal pupil size in the human eye for axial resolution,” J. Opt. Soc. Am. A 20, 2010–2015 (2003). [CrossRef]
4. Result and discussions
4.1. In vivo high-speed 3D retinal imaging
S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14, 7821–7840 (2006). [CrossRef] [PubMed]
4.2. Sensitivity
4.2.1. Maximum on-axis sensitivity
S. Yun, G. Tearney, B. Bouma, B. Park, and J. de Boer, “High-speed spectral-domain optical coherence tomography at 1.3 m wavelength,” Opt. Express 11, 35983604 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78225. [CrossRef]
4.2.2. Lateral vertical position dependence of sensitivity
Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13, 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?id=84362. [CrossRef] [PubMed]
4.3. Sensitivity advantage in ultra high-speed acquisition
D. X. Hammer, R. D. Ferguson, T. E. Bigelow, N. V. Iftimia, and R. H. Webb, “Line-scanning laser ophthalmoscope,” J. Biomed. Opt. 11, 041126 (2006). [CrossRef] [PubMed]
5. Conclusion
Acknowledgments
References and links
D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography, Science 254, 1178–1181 (1991). [CrossRef] [PubMed] | |
G. Hausler and M. W. Lindner, ““Coherence radar”and“spectral radar”-new tools for dermatological diagnosis,” J. Biomed. Opt. 3, 21–31 (1998). [CrossRef] | |
R. Leitgeb, C. Hitzenberger, and A. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express 11, 889–894 (2003). [CrossRef] [PubMed] | |
J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28, 2067–2069 (2003). [CrossRef] [PubMed] | |
N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12, 367–376 (2004). [CrossRef] [PubMed] | |
M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett. 28, 1745–1747 (2003). [CrossRef] [PubMed] | |
M. Wojtkowski, V. Srinivasan, J. G. Fujimoto, T. Ko, J. S. Schuman, A. Kowalczyk, and J. S. Duker, “Three-dimensional Retinal Imaging with High-Speed Ultrahigh-Resolution Optical Coherence Tomography,” Ophthalmology 112, 1734–1746 (2005). [CrossRef] [PubMed] | |
M. Mujat, R. Chan, B. Cense, B. Park, Chulmin Joo, T. Akkin, T. Chen, and J. de Boer, “Retinal nerve fiber layer thickness map determined from optical coherence tomography images,” Opt. Express 13, 9480–9491 (2005). [CrossRef] [PubMed] | |
S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14, 7821–7840 (2006). [CrossRef] [PubMed] | |
S. Jiao, C. Wu, R. W. Knighton, G. Gregori, and C. A. Puliafito, “Registration of high-density cross sectional images to the fundus image in spectral-domain ophthalmic optical coherence tomography,” Opt. Express 14, 3368–3376 (2006). [CrossRef] [PubMed] | |
S. Bourquin, P. Seitz, and R. P. Salathe, “Optical coherence topography based on a two-dimensional smart detector array,” Opt. Lett. 26, 512–514 (2001). [CrossRef] | |
M. Laubscher, M. Ducros, B. Karamata, T. Lasser, and R. Salathe, “Video-rate three-dimensional optical coherence tomography,” Opt. Express 10, 429–435 (2002). [PubMed] | |
M. Lebec, L. Blanchot, H. Saint-Jalmes, E. Beaurepaire, and A. C. Boccara, “Full-field optical coherence microscopy,” Opt. Lett. 23, 244–246 (1998). [CrossRef] | |
M. Akiba, K. P. Chan, and N. Tanno, “Full-field optical coherence tomography by two-dimensional heterodyne detection with a pair of CCD camera,” Opt. Lett. 28, 816–818 (2003). [CrossRef] [PubMed] | |
W. Y. Oh, B. E. Bouma, N. Iftimia, R. Yelin, and G. J. Tearney “Spectrally-modulated full-field optical coherence microscopy for ultrahigh-resolution endoscopic imaging,” Opt. Express 14, 8675–8684 (2006). [CrossRef] [PubMed] | |
I. Zeylikovich, A. Gilerson, and R. R. Alfano, “Nonmechanical grating-generated scanning coherence microscopy,” Opt. Lett. 23, 1797–1799 (1998). [CrossRef] | |
Y. Watanabe, K. Yamada, and M. Sato, “In vivo nonmechanical scanning grating-generated optical coherence tomography using an InGaAs digital camera,” Opt. Commun. 261, 376–380 (2006). [CrossRef] | |
Y. Watanabe, K. Yamada, and M. Sato, “Three-dimensional imaging by ultrahigh-speed axial-lateral parallel time domain optical coherence tomography,”Opt. Express 14, 5201–5209 (2006). [CrossRef] [PubMed] | |
A. Zuluaga and R. Richards-Kortum, “Spatially resolved spectral interferometry for determination of subsurface structure,” Opt. Lett. 24, 519–521 (1999). [CrossRef] | |
B. Grajciar, M. Pircher, A. Fercher, and R. Leitgeb, “Parallel Fourier domain optical coherence tomography for in vivo measurement of the human eye,” Opt. Express 13, 1131–1137 (2005). [CrossRef] [PubMed] | |
Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13, 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?id=84362. [CrossRef] [PubMed] | |
T. Endo, Y. Yasuno, S. Makita, M. Itoh, and T. Yatagai, “Profilometry with line-field Fourier-domain interferom-etry,” Opt. Express 13, 695–701 (2005). [CrossRef] [PubMed] | |
Y. Yasuno, T. Endo, S. Makita, G. Aoki, M. Itoh, and T. Yatagai, “Three-dimensional line-field Fourier domain optical coherence tomography for in vivo dermatological investigation,” J. Biomed. Opt. 11, 014014–014020 (2006). [CrossRef] [PubMed] | |
American National Standards Institute, American National Standard for Safe Use of Lasers: ANSI Z136.1 (Laser Institute of America, Orlando, Florida, 2000). | |
M. Takeda, H. Ina, and S. Kobayashi,“Fourier-transform method of fringe-pattern analysis for computer-based topography and interferometry,”J. Opt. Soc. Am. 72, 156–160 (1982). [CrossRef] | |
Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K. -P. Chan, M. Itoh, and T. Yatagai,“Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments,”Opt. Express 13, 10652–10664 (2005). [CrossRef] [PubMed] | |
R. Leitgeb, W. Drexler, A. Unterhuber, B. Hermann, T. Bajraszewski, T. Le, A. Stingl, and A. Fercher, “Ultrahigh resolution Fourier domain optical coherence tomography,” Opt. Express 12, 2156–2165 (2004). [CrossRef] [PubMed] | |
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12, 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?id=80320. [CrossRef] [PubMed] | |
S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Pulsed-source and swept-source spectral-domain optical coherence tomography with reduced motion artifacts,”Opt. Express 12, 5614–5624 (2004). [CrossRef] [PubMed] | |
Zawadzki, S. Jones, S. Olivier, M. Zhao, B. Bower, J. Izatt, S. Choi, S. Laut, and J. Werner, “Adaptive-optics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging,” Opt. Express 13, 8532–8546 (2005). [CrossRef] [PubMed] | |
B. Hermann, E. J. Fernandez, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, P. M. Prieto, and P. Artal, “Adaptive-optics ultrahigh-resolution optical coherence tomography,” Opt. Lett. 29, 2142–2144 (2004). [CrossRef] [PubMed] | |
E. J. Fernandez, B. Povazay, B. Hermann, A. Unterhuber, H. Sattmann, P. M. Prieto, R. Leitgeb, P. Ahnelt, P. Artal, and W. Drexler, “Three dimensional adaptive optics ultrahigh-resolution optical coherence tomography using a liquid crystal spatial light modulator,” Vision Res. 45, 3432–3444 (2005). [CrossRef] [PubMed] | |
E. J. Fernandez and W. Drexler, “Influence of ocular chromatic aberration and pupil size on transverse resolution in ophthalmic adaptive optics optical coherence tomography,”Opt. Express 13, 8184–8197 (2005). [CrossRef] [PubMed] | |
E. J. Fernandez, A. Unterhuber, B. Povazay, B. Hermann, P. Artal, and W. Drexler, “Chromatic aberration correction of the human eye for retinal imaging in the near infrared,” Opt. Express 14, 6213–6225 (2006), http://www.opticsinfobase.org/abstract.cfm?URI=oe-14-13-6213 [CrossRef] [PubMed] | |
W. J. Donnelly and A. Roorda, “Optimal pupil size in the human eye for axial resolution,” J. Opt. Soc. Am. A 20, 2010–2015 (2003). [CrossRef] | |
S. Yun, G. Tearney, B. Bouma, B. Park, and J. de Boer, “High-speed spectral-domain optical coherence tomography at 1.3 m wavelength,” Opt. Express 11, 35983604 (2003), http://www.opticsinfobase.org/abstract.cfm?id=78225. [CrossRef] | |
D. X. Hammer, R. D. Ferguson, T. E. Bigelow, N. V. Iftimia, and R. H. Webb, “Line-scanning laser ophthalmoscope,” J. Biomed. Opt. 11, 041126 (2006). [CrossRef] [PubMed] |
OCIS Codes
(110.4500) Imaging systems : Optical coherence tomography
(110.6880) Imaging systems : Three-dimensional image acquisition
(170.4470) Medical optics and biotechnology : Ophthalmology
(170.4500) Medical optics and biotechnology : Optical coherence tomography
ToC Category:
Medical Optics and Biotechnology
History
Original Manuscript: February 27, 2007
Revised Manuscript: April 23, 2007
Manuscript Accepted: April 26, 2007
Published: May 29, 2007
Virtual Issues
Vol. 2, Iss. 7 Virtual Journal for Biomedical Optics
Citation
Yoshifumi Nakamura, Shuichi Makita, Masahiro Yamanari, Masahide Itoh, Toyohiko Yatagai, and Yoshiaki Yasuno, "High-speed three-dimensional human retinal imaging by line-field spectral
domain optical coherence tomography," Opt. Express 15, 7103-7116 (2007)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-15-12-7103
Sort: Year | Journal | Reset
References
- D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, "Optical coherence tomography," Science 254, 1178-1181 (1991). [CrossRef] [PubMed]
- G. Hausler and M. W. Lindner, "Coherence radar" and "spectral radar"-new tools for dermatological diagnosis," J. Biomed. Opt. 3, 21-31 (1998). [CrossRef]
- R. Leitgeb, C. Hitzenberger, and A. Fercher, "Performance of fourier domain vs. time domain optical coherence tomography," Opt. Express 11, 889-894 (2003). [CrossRef] [PubMed]
- J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, "Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography," Opt. Lett. 28, 2067-2069 (2003). [CrossRef] [PubMed]
- N. Nassif, B. Cense, B. Park, M. Pierce, S. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, "In vivo highresolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve," Opt. Express 12, 367-376 (2004). [CrossRef] [PubMed]
- M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, "Real-time in vivo imaging by high-speed spectral optical coherence tomography," Opt. Lett. 28, 1745-1747 (2003). [CrossRef] [PubMed]
- M. Wojtkowski, V. Srinivasan, J. G. Fujimoto, T. Ko, J. S. Schuman, A. Kowalczyk and J. S. Duker, "Three dimensional retinal imaging with high-speed ultrahigh-resolution Optical Coherence Tomography," Ophthalmology 112, 1734-1746 (2005). [CrossRef] [PubMed]
- M. Mujat, R. Chan, B. Cense, B. Park, Chulmin Joo, T. Akkin, T. Chen, and J. de Boer, "Retinal nerve fiber layer thickness map determined from optical coherence tomography images," Opt. Express 13, 9480-9491 (2005). [CrossRef] [PubMed]
- S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, "Optical coherence angiography," Opt. Express 14, 7821-7840 (2006). [CrossRef] [PubMed]
- S. Jiao, C. Wu, R. W. Knighton, G. Gregori, and C. A. Puliafito, "Registration of high-density cross sectional images to the fundus image in spectral-domain ophthalmic optical coherence tomography," Opt. Express 14, 3368-3376 (2006). [CrossRef] [PubMed]
- S. Bourquin, P. Seitz, and R. P. Salathe, "Optical coherence topography based on a two-dimensional smart detector array," Opt. Lett. 26, 512-514 (2001). [CrossRef]
- M. Laubscher, M. Ducros, B. Karamata, T. Lasser, and R. Salathe, "Video-rate three-dimensional optical coherence tomography," Opt. Express 10, 429-435 (2002). [PubMed]
- M. Lebec, L. Blanchot, H. Saint-Jalmes, E. Beaurepaire, and A. C. Boccara, "Full-field optical coherence microscopy," Opt. Lett. 23, 244-246 (1998). [CrossRef]
- M. Akiba, K. P. Chan, and N. Tanno, "Full-field optical coherence tomography by two-dimensional heterodyne detection with a pair of CCD camera," Opt. Lett. 28, 816-818 (2003). [CrossRef] [PubMed]
- W. Y. Oh, B. E. Bouma, N. Iftimia, R. Yelin, and G. J. Tearney "Spectrally-modulated full-field optical coherence microscopy for ultrahigh-resolution endoscopic imaging," Opt. Express 14, 8675-8684 (2006). [CrossRef] [PubMed]
- I. Zeylikovich, A. Gilerson, and R. R. Alfano, "Nonmechanical grating-generated scanning coherence microscopy," Opt. Lett. 23, 1797-1799 (1998). [CrossRef]
- Y. Watanabe, K. Yamada, and M. Sato, "In vivo nonmechanical scanning grating-generated optical coherence tomography using an InGaAs digital camera," Opt. Commun. 261, 376-380 (2006). [CrossRef]
- Y. Watanabe, K. Yamada, and M. Sato, "Three-dimensional imaging by ultrahigh-speed axial-lateral parallel time domain optical coherence tomography," Opt. Express 14, 5201-5209 (2006). [CrossRef] [PubMed]
- A. Zuluaga and R. Richards-Kortum, "Spatially resolved spectral interferometry for determination of subsurface structure," Opt. Lett. 24, 519-521 (1999). [CrossRef]
- B. Grajciar, M. Pircher, A. Fercher, and R. Leitgeb, "Parallel Fourier domain optical coherence tomography for in vivo measurement of the human eye," Opt. Express 13, 1131-1137 (2005). [CrossRef] [PubMed]
- Y. Zhang, J. Rha, R. Jonnal, and D. Miller, "Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina," Opt. Express 13, 4792-4811 (2005). [CrossRef] [PubMed]
- T. Endo, Y. Yasuno, S. Makita, M. Itoh, and T. Yatagai, "Profilometry with line-field Fourier-domain interferometry," Opt. Express 13, 695-701 (2005). [CrossRef] [PubMed]
- Y. Yasuno, T. Endo, S. Makita, G. Aoki, M. Itoh, and T. Yatagai, "Three-dimensional line-field Fourier domain optical coherence tomography for in vivo dermatological investigation," J. Biomed. Opt. 11, 014014-014020 (2006). [CrossRef] [PubMed]
- American National Standards Institute, American National Standard for Safe Use of Lasers: ANSI Z136.1 (Laser Institute of America, Orlando, Florida, 2000).
- M. Takeda, H. Ina, and S. Kobayashi,"Fourier-transform method of fringe-pattern analysis for computer-based topography and interferometry,"J. Opt. Soc. Am. 72, 156-160 (1982). [CrossRef]
- Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K. -P. Chan, M. Itoh, and T. Yatagai,"Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments," Opt. Express 13, 10652-10664 (2005). [CrossRef] [PubMed]
- R. Leitgeb,W. Drexler, A. Unterhuber, B. Hermann, T. Bajraszewski, T. Le, A. Stingl, and A. Fercher, "Ultrahigh resolution Fourier domain optical coherence tomography," Opt. Express 12, 2156-2165 (2004). [CrossRef] [PubMed]
- S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, "Motion artifacts in optical coherence tomography with frequency-domain ranging," Opt. Express 12, 2977-2998 (2004). [CrossRef] [PubMed]
- S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, "Pulsed-source and swept-source spectral-domain optical coherence tomography with reduced motion artifacts," Opt. Express 12, 5614-5624 (2004). [CrossRef] [PubMed]
- R. Zawadzki, S. Jones, S. Olivier, M. Zhao, B. Bower, J. Izatt, S. Choi, S. Laut, and J. Werner, "Adaptive-optics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging," Opt. Express 13, 8532-8546 (2005). [CrossRef] [PubMed]
- B. Hermann, E. J. Fernandez, A. Unterhuber, H. Sattmann, A. F. Fercher, and W. Drexler, P. M. Prieto and P. Artal, "Adaptive-optics ultrahigh-resolution optical coherence tomography," Opt. Lett. 29, 2142-2144 (2004). [CrossRef] [PubMed]
Cited By |
OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.
Multimedia
| Multimedia Files | Recommended Software |
| » Media 1: AVI (8182 KB) | |
| » Media 2: AVI (2160 KB) | |
| » Media 3: AVI (2270 KB) | |
| » Media 4: AVI (12323 KB) |





OSA is a member of 