Microring-resonator-based sensor measuring both the concentration and temperature of a solution

Min-Suk Kwon and William H. Steier  »View Author Affiliations


Optics Express, Vol. 16, Issue 13, pp. 9372-9377 (2008)
http://dx.doi.org/10.1364/OE.16.009372


View Full Text Article

Acrobat PDF (3865 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We propose and investigate experimentally a micro-ring-resonator-based sensor with which we can measure both the concentration and temperature of glucose solution. It consists of two micro-ring resonators consecutively coupled to a bus waveguide by the overlap between them. The resonance wavelengths of the two resonators change similarly with the temperature but differently with the concentration. For that purpose, the core of just one micro-ring resonator is exposed directly to the solution. Using polymers, conventional processes, and a polymer lift-off process, we implement the sensor. Through the measurement of the fabricated sensor, we obtain its characteristics of measuring the temperature and concentration.

© 2008 Optical Society of America

1. Introduction

In this paper, we propose and investigate experimentally a MRR-based sensor measuring both the concentration and temperature of glucose solution. In the following sections, we describe its operation principle and fabrication. Then, we present experimental results and discussion. Finally, concluding remarks follow.

2. Description of the sensor

The proposed sensor consists of two MRRs with the same radius R coupled consecutively to a bus waveguide. Its schematic diagram is shown in Fig. 1(a). The cores of the MRRs and the bus waveguide are in the form of a square channel on a substrate as shown in Fig. 1(b). The two MRRs are sufficiently apart from each other such that there is no interaction between them. Obviously, they should be different for the measurement of two physical parameters. For that purpose, the core of MRR 1 is covered by a cladding, but the most part of the core of MRR 2 is uncovered. When the sensor is dipped into glucose solution, the uncovered core of MRR 2 contacts directly it.

Fig. 1. (a). Schematic diagram of the proposed sensor, (b) waveguide structures of MRR 1 (top) and 2 (bottom).

As shown in Fig. 1(a), the coupling between the MRRs and the bus waveguide is achieved by overlapping them appropriately [7

7. W.-Y. Chen, V. Van, T. N. Ding, M. Du, W. N. Herman, and P.-T. Ho, “Benzocyclobutene negative-gap micro-ring notch filters,” in Frontiers in Optics, OSA Technical Digest Series (Optical Society of America, 2005), paper SWA6.

]. Usually, it has been obtained from the directional coupling mechanism. However, as well known, the gap between the MRR and the bus waveguide of a MRR device employing the latter becomes smaller as the index difference between the core and the cladding of the device increases. Consequently, it becomes harder to fabricate such a device using a conventional photolithography process. In contrast, the overlap-based coupling mechanism removes the difficulty in fabrication. This is because there is no need to pattern lines with sub-micrometer spacing for a MRR device based on it. The efficiency of the coupling between the MRR and the bus waveguide of such a device is easily controlled by adjusting the degree of overlapping. Using the finite-difference semi-vectorial beam propagation method [8

8. BeamPROP™ is available from RSoft, Inc., http://www.rsoftinc.com.

], we simulated a structure comprised of straight and bent waveguides shown in Fig. 2(a). In the simulation, the refractive indexes of the core, the cladding, and the substrate were 1.567, 1.49, and 1.45, respectively. In addition, the width and height of the core were 2 and 2 µm, respectively, and the cladding was sufficiently thick. The radius of curvature of the bent waveguide was 1000 µm. Figure 2(b) shows the efficiency of coupling to the bent waveguide and that of transmission through the straight waveguide as functions of the gap defined in Fig. 2(a). The coupling efficiency takes almost all the values between 0 and 1 as the gap decreases from 2 to 1 µm. The simulation results show that we can obtain a desired coupling efficiency using the overlap-based coupling mechanism although it changes sensitively depending on the degree of overlapping.

Fig. 2. (a). Simulated structure. In the simulation, θ=11.5°. The gap between the straight and bent waveguides is defined as the distance between their centers. (b) Simulated efficiency of coupling to the bent waveguide and that of transmission through the straight waveguide.

The transmission spectrum through the bus waveguide of the proposed sensor has two distinct groups of resonance bands whose center wavelengths are given by λc,i=2πRNeff,i/mi, where i is 1 and 2 for MRR 1 and 2, respectively. In this equation, Neff,i is the effective index of a MRR mode, and mi is an integer. From this expression, a center wavelength change Δλc,i is obtained as follows:

Δλc,i=2πRmi(Neff,inaΔna+Neff,iTΔT+Neff,iλΔλc,i),
(1)

where T and na denote the temperature and refractive index of analyte, i.e. glucose solution, respectively. If we assume that (2πR/mi)·(∂Neff,i/∂λ)≪1 and that na is proportional to glucose concentration C, then Eq. (1) becomes

[Δλc,1Δλc,2]=[SC,1ST,1SC,2ST,2][ΔCΔT],
(2)

3. Fabrication and measurement

To implement the proposed sensor, we used a SiO2-deposited silicon wafer as a substrate. SU-8 (Microchem Corporation) was used as a core material. Using conventional processes, we made SU-8 patterns. First, we spin-coated SU-8 on the SiO2 layer at 2100 rpm for 30 seconds. Then, it was pre-baked on a hot plate at 65 °C for 2 minutes, and soft-baked at 95 °C for 3 minutes. After the soft baking, we made the patterns of the MRRs and the bus waveguide with a mask aligner (MJB3, SUSS MicroTec Inc.). At this step, exposure energy measured at the wavelength of 405 nm was 70 mJ/cm2. The exposed SU-8 layer was post-baked sequentially at 65 °C for 2 minutes and at 95 °C for 3 minutes. Finally, it was developed in SU-8 developer, rinsed with isopropyl alcohol, and dried with nitrogen gas. The height and width of a formed SU-8 pattern were almost 2 and 2 µm, respectively. According to a simulation, the channel waveguide of this size supports two higher-order modes, which are near cut-off, as well as a fundamental mode.

For the most part of the SU-8 pattern for MRR 2 to be uncovered by a cladding, we used a polymer lift-off process. This process was done as follows. Over the core region of MRR 2 required to be uncovered, an 18-µm-thick ∩-shaped AZ P4620 (Clariant Co.) pattern was made. To make the pattern, we spin-coated AZ P4620 at 1000 rpm for 60 seconds, and baked on a hot plate at 110 °C for 1 minute. Then, the pattern was formed by a photolithography process, and developed in a mixture of 1 part AZ 400K (Clariant Co.) and 3 parts DI water (by volume). After making the AZ P4620 pattern, we spin-coated UFC170A (Uray Co.) used as a cladding and UV-cured it. Dipping the sample into AZ 400K, we removed the AZ P4620 pattern, and we baked the sample at 160 °C for 2 hours. Finally, using a dicing process, we prepared the input and output facets of the sensor. The size of the accomplished sensor chip was 8×8 mm2.

A microscope image of the fabricated MRR 2 is shown in Fig. 3(a). In Fig. 3(a), thin lines correspond to the SU-8 patterns for the MRR and the bus waveguide. A Cross-section image of MRR 2 are shown in Fig. 3(b). As shown in Fig. 3(b), the cladding is thickest at the boundaries of the uncovered region of MRR 2, and it becomes gradually thinner as the distance from the region increases. The thickness of the cladding on the core of MRR 1 was 3 µm. In the fabricated sensor, both MRR 1 and 2 had a radius of 1000 µm. The gaps for MRR 1 and 2 were 1.4 and 1.5 µm, respectively, so d 1 and d 2 in Fig. 1(a) were 3.4 and 3.5 µm.

Fig. 3. Images of the fabricated sensor: (a) the top view of MRR 2, and (b) the cross-section of MRR 2.

For the characterization of the sensor, we pigtailed small-core fibers with a core diameter of 4 µm to the input and output facets, using glass ferrules and UV-curable epoxy [9

9. B. Bhola and W. H. Steier, “A novel optical microring resonator accelerometer,” IEEE Sens. J. 7, 1759–1765 (2007). [CrossRef]

]. As a light source, we used a tunable laser (Model 6200, New Focus Co.), and monitored the signal from the output fiber with a photo-detector. Dissolving w1-gram glucose (G7528, Sigma Aldrich Co.) into w 2-gram DI water, we prepared glucose solution with the concentration C=100·w 1/(w 1+w 2) (%). Into the prepared glucose solution with a volume of about 1 litter, we dipped the pigtailed sensor, and measured its transmission spectrum while controlling the temperature of the solution.

4. Results and discussion

Fig. 4. (a). The measured transmission spectrum of the sensor dipped into pure DI water at 22 °C. (b) Changes of the transmission spectrum due to a temperature increase of 1 °C (the green line with triangles) and a concentration increase of 1.5 % (the red line with circles). Each number means that the resonance band corresponds to the MRR with the number.

The center wavelength changes Δλ c,1 and Δλ c,2 are shown as functions of the temperature for several concentrations in Figs. 6(a) and 6(b), respectively. The measured S T,1 and S T,2 were -81.0±7.26 and -92.5±6.02 pm/°C, respectively. Light is more strongly confined to the core of MRR 2 than that of MRR 1 since the cladding over that has a larger refractive index than glucose solution. Thus, |S T,2|>|S T,1|. The absolute values of S T,1 and S T,2 are a little smaller than the previously reported value of about 150 pm/°C, which was measured around the wavelength of 1550 nm [10

10. P. Rabiei, W. H. Steier, C. Zhang, and L. R. Dalton, “Polymer microring filters and modulators,” IEEE J. Lightwave Technol. 20, 1968–1975 (2002). [CrossRef]

].

As mentioned above, we can calculate and find ΔC and ΔT of unknown glucose solution, using Eq. (2) with the obtained values of S C,1, S T,1, S C,2, and S T,2 after measuring Δλ c,1 and Δλ c,2. The (1, 1), (1, 2), (2, 1), and (2, 2) elements of the inverse of the 2×2 matrix in Eq. (2) are -0.0197±0.0107 %/pm, 0.0172±0.0098 %/pm, -0.0209±0.0118 °C/pm, and 0.00749±0.0051 °C/pm, respectively. Because of the errors in these values, the calculated values of ΔC and ΔT may have quite large errors. Therefore, as a further work, it is required to measure carefully S C,1, S T,1, S C,2, and S T,2 with a more precise thermometer.

Fig. 5. Center wavelength changes (a) Δλ c,1 and (b) Δλ c,2 related to MRR 1 and 2, respectively, as functions of the concentration for several temperatures. The solid lines are obtained from linear fitting of the data. The error bars denote the inaccuracy of the wavelength shift inferred from that of measured temperature.
Fig. 6. Center wavelength changes (a) Δλ c,1 and (b) Δλ c,2 related to MRR 1 and 2, respectively, as functions of the temperature for several concentrations. The solid lines are obtained from linear fitting of the data. The error bars denote the inaccuracy in measuring temperature.

5. Conclusions

Acknowledgment

The authors would like to thank Andrew Yick, Dr. Bipin Bhola, and Dr. Hidehisa Tazawa for helpful discussions about this work.

References and links

1.

P. V. Lambeck, “Integrated optical sensors for the chemical domain,” Meas. Sci. Technol. 17, R93–R116, (2006). [CrossRef]

2.

X. D. Hoa, A. G. Kirk, and M. Tabrizian, “Towards integrated and sensitive surface plasmon resonance biosensors: A review of recent progress,” Biosens. Bioelectron. 23, 151–160 (2007). [CrossRef] [PubMed]

3.

M.-S. Kwon and S.-Y. Shin, “Refractive index sensitivity measurement of a long-period waveguide grating,” IEEE Photon. Technol. Lett. 17, 1923–1925 (2005). [CrossRef]

4.

P. Debackere, S. Scheerlinck, P. Bienstman, and R. Baets, “Surface plasmon interferometer in silicon-on-insulator: novel concept for an integrated biosensor,” Opt. Express 14, 7063–7071 (2006) [CrossRef] [PubMed]

5.

C.-Y. Chao, W. Fung, and L. J. Guo, “Polymer microring resonators for biochemical sensing applications,” J. Sel. Top. Quantum Electron. 12, 134–142 (2006). [CrossRef]

6.

A. Ksendzov and Y. Lin, “Integrated optics ring-resonator sensors for protein detection,” Opt. Lett. 30, 3344–3346 (2005). [CrossRef]

7.

W.-Y. Chen, V. Van, T. N. Ding, M. Du, W. N. Herman, and P.-T. Ho, “Benzocyclobutene negative-gap micro-ring notch filters,” in Frontiers in Optics, OSA Technical Digest Series (Optical Society of America, 2005), paper SWA6.

8.

BeamPROP™ is available from RSoft, Inc., http://www.rsoftinc.com.

9.

B. Bhola and W. H. Steier, “A novel optical microring resonator accelerometer,” IEEE Sens. J. 7, 1759–1765 (2007). [CrossRef]

10.

P. Rabiei, W. H. Steier, C. Zhang, and L. R. Dalton, “Polymer microring filters and modulators,” IEEE J. Lightwave Technol. 20, 1968–1975 (2002). [CrossRef]

OCIS Codes
(130.3120) Integrated optics : Integrated optics devices
(130.6010) Integrated optics : Sensors
(230.5750) Optical devices : Resonators
(130.5460) Integrated optics : Polymer waveguides

ToC Category:
Integrated Optics

History
Original Manuscript: March 12, 2008
Revised Manuscript: June 3, 2008
Manuscript Accepted: June 8, 2008
Published: June 11, 2008

Virtual Issues
Vol. 3, Iss. 7 Virtual Journal for Biomedical Optics

Citation
Min-Suk Kwon and William H. Steier, "Microring-resonator-based sensor measuring both the concentration and temperature of a solution," Opt. Express 16, 9372-9377 (2008)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-16-13-9372


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. P. V. Lambeck, "Integrated optical sensors for the chemical domain," Meas. Sci. Technol. 17, R93-R116, (2006). [CrossRef]
  2. X. D. Hoa, A. G. Kirk, and M. Tabrizian, "Towards integrated and sensitive surface plasmon resonance biosensors: A review of recent progress," Biosens. Bioelectron. 23, 151-160 (2007). [CrossRef] [PubMed]
  3. M.-S. Kwon and S.-Y. Shin, "Refractive index sensitivity measurement of a long-period waveguide grating," IEEE Photon. Technol. Lett. 17, 1923-1925 (2005). [CrossRef]
  4. P. Debackere, S. Scheerlinck, P. Bienstman, and R. Baets, "Surface plasmon interferometer in silicon-on-insulator: novel concept for an integrated biosensor," Opt. Express 14, 7063-7071 (2006) [CrossRef] [PubMed]
  5. C.-Y. Chao, W. Fung, and L. J. Guo, "Polymer microring resonators for biochemical sensing applications," J. Sel. Top. Quantum Electron. 12, 134-142 (2006). [CrossRef]
  6. A. Ksendzov and Y. Lin, "Integrated optics ring-resonator sensors for protein detection," Opt. Lett. 30, 3344-3346 (2005). [CrossRef]
  7. W.-Y. Chen, V. Van, T. N. Ding, M. Du, W. N. Herman, and P.-T. Ho, "Benzocyclobutene negative-gap micro-ring notch filters," in Frontiers in Optics, OSA Technical Digest Series (Optical Society of America, 2005), paper SWA6.
  8. BeamPROPTM is available from RSoft, Inc., http://www.rsoftinc.com.
  9. B. Bhola and W. H. Steier, "A novel optical microring resonator accelerometer," IEEE Sens. J. 7, 1759-1765 (2007). [CrossRef]
  10. P. Rabiei, W. H. Steier, C. Zhang, and L. R. Dalton, "Polymer microring filters and modulators," IEEE J. Lightwave Technol. 20, 1968-1975 (2002). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited