Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo

Christoph J. Engelbrecht, Richard S. Johnston, Eric J. Seibel, and Fritjof Helmchen  »View Author Affiliations


Optics Express, Vol. 16, Issue 8, pp. 5556-5564 (2008)
http://dx.doi.org/10.1364/OE.16.005556


View Full Text Article

Acrobat PDF (479 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We present a small, lightweight two-photon fiberscope and demonstrate its suitability for functional imaging in the intact brain. Our device consists of a hollow-core photonic crystal fiber for efficient delivery of near-IR femtosecond laser pulses, a spiral fiber-scanner for resonant beam steering, and a gradient-index lens system for fluorescence excitation, dichroic beam splitting, and signal collection. Fluorescence light is remotely detected using a standard photomultiplier tube. All optical components have 1 mm dimensions and the microscope’s headpiece weighs only 0.6 grams. The instrument achieves micrometer resolution at frame rates of typically 25 Hz with a field-of-view of up to 200 microns. We demonstrate functional imaging of calcium signals in Purkinje cell dendrites in the cerebellum of anesthetized rats. The microscope will be easily portable by a rat or mouse and thus should enable functional imaging in freely behaving animals.

© 2008 Optical Society of America

1. Introduction

Advances in the development of miniaturized fiber-optic imaging devices are opening new avenues for biomedical research and clinical imaging. For example, flexible catheter-like probes for video endoscopy [1

1. E. Seibel, T. Soper, R. Johnston, and R. Glenny, “Ultrathin laser scanning bronchoscope and guidance system for the peripheral lung,” Lung Cancer 49, S162–S162 (2005). [CrossRef]

, 2

2. E. J. Seibel, R. S. Johnston, C. M. Brown, J. A. Dominitz, and M. B. Kimmey, “Novel ultrathin scanning fiber endoscope for cholangioscopy and pancreatoscopy,” Gastrointest. Endosc. 65, Ab125–Ab125 (2007). [CrossRef]

] or optical coherence tomography [3

3. A. F. Low, G. J. Tearney, B. E. Bouma, and I. K. Jang, “Technology insight: optical coherence tomography - current status and future development,” Nature Clinical Practice Cardiovascular Medicine 3, 154–162 (2006). [CrossRef] [PubMed]

, 4

4. Z. Yaqoob, J. G. Wu, E. J. McDowell, X. Heng, and C. H. Yang, “Methods and application areas of endoscopic optical coherence tomography,” J. Biomed. Opt. 11, 063001 (2006). [CrossRef]

] enable minimally invasive medical imaging, diagnosis, and potential treatment. While some applications rely on large field-of-views (FOV), others require a high spatial resolution in order to resolve cellular structures. Various approaches have been explored to build miniature high-resolution microscopes, including the design of microendoscopes, miniature confocal laser-scanning microscopes, coherent fiber-bundle imagers and nonlinear fluorescence microscopes (reviewed in [5–8

5. F. Helmchen, “Miniaturization of fluorescence microscopes using fibre optics,” Exp. Physiol. 87, 737–745 (2002). [CrossRef] [PubMed]

]). Besides their potential impact for clinical applications, fiber-optic devices have great promises as new tools in neuroscientific research. In combination with fluorescent dyes for optical probing of neural activity and miniature gradient-index (GRIN) lens designs, they may allow in vivo endoscopic imaging of cells in deep brain structures [9

9. J. C. Jung, A. D. Mehta, E. Aksay, R. Stepnoski, and M. J. Schnitzer, “In vivo mammalian brain Imaging using one- and two-photon fluorescence microendoscopy,” J. Neurophysiol. 92, 3121–3133 (2004). [CrossRef] [PubMed]

, 10

10. M. J. Levene, D. A. Dombeck, K. A. Kasischke, R. P. Molloy, and W. W. Webb, “In vivo multiphoton microscopy of deep brain tissue,” J. Neurophysiol. 91, 1908–1912 (2004). [CrossRef]

] as well as functional imaging of cellular activities in freely behaving animals [11–13

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

].

Two-photon microscopy [14

14. W. Denk, J. H. Strickler, and W. W. Webb, “Two-Photon Laser Scanning Fluorescence Microscopy,” Science 248, 73–76 (1990). [CrossRef] [PubMed]

] is especially well suited for imaging deep inside biological tissue [15

15. W. Denk and K. Svoboda, “Photon upmanship: why multiphoton imaging is more than a gimmick,” Neuron 18, 351–357 (1997). [CrossRef] [PubMed]

, 16

16. F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nature Methods 2, 932–940 (2005). [CrossRef] [PubMed]

]. Therefore miniaturization of two-photon microscopes has gained particular interest. Following the first demonstration of a fiber-optic two-photon microscope [11

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

] various instruments of decreasing size have been presented. Key technological achievements have been the application of micro-optical components such as GRIN lenses [17–19

17. J. Knittel, L. Schnieder, G. Buess, B. Messerschmidt, and T. Possner, “Endoscope-compatible confocal microscope using a gradient index-lens system,” Opt. Commun. 188, 267–273 (2001). [CrossRef]

], the use of special optical fibers for efficient laser pulse delivery [20–23

20. D. G. Ouzounov, K. D. Moll, M. A. Foster, W. R. Zipfel, W. W. Webb, and A. L. Gaeta, “Delivery of nanojoule femtosecond pulses through large-core microstructured fibers,” Opt. Lett. 27, 1513–1515 (2002). [CrossRef]

] and fluorescence collection [24

24. M. T. Myaing, D. J. MacDonald, and X. D. Li, “Fiber-optic scanning two-photon fluorescence endoscope,” Opt. Lett. 31, 1076–1078 (2006). [CrossRef] [PubMed]

, 25

25. L. Fu, X. Gan, and M. Gu, “Nonlinear optical microscopy based on double-clad photonic crystal fibers,” Opt. Express 13, 5528–5534 (2005). [CrossRef] [PubMed]

], respectively, and the development of compact laser scanning devices. Two scanning methods have prevailed: (1) mechanical deflection or vibration of fiber cantilevers [11

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

, 23

23. B. A. Flusberg, J. C. Lung, E. D. Cocker, E. P. Anderson, and M. J. Schnitzer, “In vivo brain imaging using a portable 3.9 gram two-photon fluorescence microendoscope,” Opt. Lett. 30, 2272–2274 (2005). [CrossRef] [PubMed]

, 26

26. J. Sawinski and W. Denk, “Miniature random-access fiber scanner for in vivo multiphoton imaging,” J. Appl. Phys. 102, (2007). [CrossRef]

], and (2) microelectromechanical systems (MEMS) with small mirrors [27

27. W. Piyawattanametha, R. P. J. Barretto, T. H. Ko, B. A. Flusberg, E. D. Cocker, H. J. Ra, D. S. Lee, O. Solgaard, and M. J. Schnitzer, “Fast-scanning two-photon fluorescence imaging based on a microelectromechanical systems two-dimensional scanning mirror,” Opt. Lett. 31, 2018–2020 (2006). [CrossRef] [PubMed]

, 28

28. L. Fu, A. Jain, H. Xie, C. Cranfield, and M. Gu, “Nonlinear optical endoscopy based on a double-clad photonic crystal fiber and a MEMS mirror,” Opt. Express 14, 1027–1032 (2006). [CrossRef] [PubMed]

]. Alternatively, a coherent fiber bundle can circumvent the need for scanning at the microscope headpiece albeit at the cost of reduced spatial resolution and sensitivity [19

19. W. Göbel, J. N. D. Kerr, A. Nimmerjahn, and F. Helmchen, “Miniaturized two-photon microscope based on a flexible coherent fiber bundle and a gradient-index lens objective,” Opt. Lett. 29, 2521–2523 (2004). [CrossRef] [PubMed]

]. Despite these technical advances, only few functional measurements have been demonstrated with miniature fiberscopes, possibly because of the challenges in terms of sensitivity and stability. Functional studies comprise blood flow measurements in rodent brain [11

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

] and cochlea [29

29. A. Monfared, N. H. Blevins, E. L. M. Cheung, J. C. Jung, G. Popelka, and M. J. Schnitzer, “In vivo Imaging of mammalian cochlear blood flow using fluorescence microendoscopy,” Otology & Neurotology 27, 144–152 (2006). [CrossRef] [PubMed]

] and measurements of neuronal activity using fluorescent calcium indicators in anesthetized animals [11

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

, 30

30. P. Vincent, U. Maskos, I. Charvet, L. Bourgeais, L. Stoppini, N. Leresche, J. P. Changeux, R. Lambert, P. Meda, and D. Paupardin-Tritsch, “Live imaging of neural structure and function by fibred fluorescence microscopy,” EMBO Rep 7, 1154–1161 (2006). [CrossRef] [PubMed]

]. So far, calcium imaging in awake, freely moving animals has only been achieved using fiber-optic approaches with low or no spatial resolution [12

12. H. Adelsberger, O. Garaschuk, and A. Konnerth, “Cortical calcium waves in resting newborn mice,” Nat. Neurosci. 8, 988–990 (2005). [CrossRef] [PubMed]

, 13

13. M. Murayama, E. Perez-Garci, H. R. Luscher, and M. E. Larkum, “Fiberoptic system for recording dendritic calcium signals in layer 5 neocortical pyramidal cells in freely moving rats,” J. Neurophysiol. 98, 1791–1805 (2007). [CrossRef] [PubMed]

]. A routine application of fiber-optic microscopes for spatially-resolved functional imaging in awake animals is still pending.

Here we present a novel ultra-compact design of a two-photon fiberscope combining a miniature fiber scanner originally developed for video endoscopy [31

31. E. J. Seibel and Q. Y. J. Smithwick, “Unique features of optical scanning, single fiber endoscopy,” Lasers Surg. Med. 30, 177–183 (2002). [CrossRef] [PubMed]

] with optical and fiber components suitable for two-photon excited fluorescence measurements. We provide an optical characterization of our instrument and we demonstrate in vivo calcium imaging of dendritic activity in Purkinje cells of rat cerebellum.

2. Methods and results

2.1 Optical setup and fiberscope design

Near-infrared laser pulses (λ=812 nm, τ≈100 fs) generated by an ultrafast Ti:Sapphire laser system (Tsunami and Millenia-X, Spectra-Physics, CA) were intensity modulated by a Pockels cell (Conoptics Inc., CT) and coupled into a hollow-core photonic crystal fiber (PCF, HC-800-01/-02, Crystal Fibre A/S, Birkerod, Denmark) for efficient pulse delivery with minimal distortions [22

22. W. Göbel, A. Nimmerjahn, and F. Helmchen, “Distortion-free delivery of nanojoule femtosecond pulses from a Ti : sapphire laser through a hollow-core photonic crystal fiber,” Opt. Lett. 29, 1285–1287 (2004). [CrossRef] [PubMed]

]. The laser beam was expanded and collimated with two telescopes (concave mirror pairs with focal lengths f 1=50 mm and f 2=100 mm, respectively). It was coupled to the PCF with a f=18.4 mm aspheric lens (Geltech) matching the numerical aperture (NA) of the PCF for maximum coupling efficiency. We achieved 30–40% transmission through the fiber with maximum average laser power levels of ~300 mW at the sample.

Laser pulses were guided through the PCF to the ultra-compact fiberscope headpiece (Fig. 1(a) and (b)). We used an ultra-thin spiral-pattern fiber scanner, which is based on a recently developed micro-endoscope system [2

2. E. J. Seibel, R. S. Johnston, C. M. Brown, J. A. Dominitz, and M. B. Kimmey, “Novel ultrathin scanning fiber endoscope for cholangioscopy and pancreatoscopy,” Gastrointest. Endosc. 65, Ab125–Ab125 (2007). [CrossRef]

, 31

31. E. J. Seibel and Q. Y. J. Smithwick, “Unique features of optical scanning, single fiber endoscopy,” Lasers Surg. Med. 30, 177–183 (2002). [CrossRef] [PubMed]

, 32

32. E. J. Seibel, R. S. Johnston, and C. D. Melville, “A full-color scanning fiber endoscope,” Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI. Proceedings of the SPIE. 6083, 9–16 (2006).

]. Resonant vibrations (resonance frequency ν≈5 kHz; quality factor Q≈50) were induced in a fiber cantilever of about 4.5-mm length by using a custom-designed tubular piezoelectric actuator equipped with quadrant electrodes. Application of drive signals to the piezo tube with a phase shift of ϕ=90° between x- and y-direction induced cyclic spiral motions of the fiber cantilever (Fig. 1(c) and (d)). Appropriate amplitude modulation was used to create a spiral scan pattern (using linear or sinusoidal segments as envelopes) [24

24. M. T. Myaing, D. J. MacDonald, and X. D. Li, “Fiber-optic scanning two-photon fluorescence endoscope,” Opt. Lett. 31, 1076–1078 (2006). [CrossRef] [PubMed]

, 32

32. E. J. Seibel, R. S. Johnston, and C. D. Melville, “A full-color scanning fiber endoscope,” Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI. Proceedings of the SPIE. 6083, 9–16 (2006).

]. Specifically, expanding outward spirals were used for the imaging portion. Each imaging period was followed by shorter periods of retrace and fiber settling. The durations of the three phases could be arbitrarily chosen. To create 128×128–512×512 pixel images we used 64–520 expanding spiral cycles, typically 25 cycles for retracing and 50–100 cycles as settling period. These parameters result in frame rates of 36–8 frames per second (fps). In the in vivo experiments, we typically used scanning cycles near video rate (25 Hz), producing 128×128 pixel images with 80–130 expanding spirals, 25 retrace spirals, and 50–100 settling spirals. Optional online averaging could be used to efficiently reduce noise at the cost of lowering the frame rates.

The spiral motion of the fiber cantilever was imaged to the sample plane by a custom-designed assembly of GRIN-lenses with 1-mm diameter and dichroic beam-splitting 1-mm prisms (GRINTECH GmbH, Jena, Germany). The dichroic coating was designed to reflect ~93% of the near-IR excitation light while transmitting ~95% of the fluorescence emission light in the visible regime (~530 nm). Both values are for mixed s- and p-polarized light. The back lens served as a tube lens (working distance 450 µm in air) and had a maximum NA of 0.2. The front objective lens (maximum NA=0.5) had a working distance of 300 µm in water, thus yielding a demagnification factor of 2.5x. The FOV in the sample plane was determined by the maximal fiber deflection and the lens demagnification factor. With our instrument, we achieved FOVs of up to 200 µm in diameter. Fluorescence emission photons in the visible regime were coupled to a large-core step-index fiber with an outer diameter of 1 mm (Edmund Optics, Karlsruhe, Germany) and were remotely detected by a photomultiplier tube (PMT, RU5600, Hamamatsu, Japan) after passing a 2-mm BG39 emission filter (AHF Analysentechnik, Tübingen, Germany). The analog PMT current-signal was amplified by a variable gain current amplifier (Femto Messtechnik GmbH, Berlin, Germany) and digitized at 12 bits at 8–10 MHz with a multifunction data acquisition board (PCI-6115, National Instruments, TX).

For image generation the sampled fluorescence intensity values were assigned to their corresponding pixels based on a static lookup-table (LUT). Figure 1(e) (left) shows a two-photon image of 10-micron fluorescent microspheres obtained with the fiberscope. Individual microspheres are clearly discernible but image distortions (e.g., angular lag and pin-cushion distortions) are noticeable near the center of the image. These distortions were caused by phase and amplitude deviations of the actual fiber cantilever motion from the “ideal” trajectory defined by the drive signals. This effect is illustrated in Figure 1(d) for a small number of spiral lines. The distortions of the scan trajectory could be compensated for by applying a remapping-LUT [32

32. E. J. Seibel, R. S. Johnston, and C. D. Melville, “A full-color scanning fiber endoscope,” Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI. Proceedings of the SPIE. 6083, 9–16 (2006).

] based on direct measurement of the actual cantilever motion using a 2D position-sensitive detector (S5991, Hamamatsu). Figure 1(e) (right) shows the image of 10-micron microspheres with a previously measured LUT applied. After remapping, microspheres had similar diameters near the center and at the periphery of the FOV and only minimal distortions remained in the center, which were only noticeable when slowly moving the FOV laterally. For functional imaging from specific cellular structures, correction of image distortions is optional.

Fig. 1. Schematic (a) and perspective (b) representations of the ultra-compact fiber-optic two-photon microscope. A hollow-core PCF guides near-IR ultrafast laser pulses (red) to the microscope headpiece. The fiber cantilever is scanned in a spiral pattern by applying appropriate piezo drive signals (PDS) to a resonant fiber scanner (RFS). The motion of the fiber cantilever is imaged onto the sample (S) by a custom-designed combination of GRINlenses and micro-beamsplitter prisms (BSP). Fluorescence emission light (green) passes the GRIN objective lens and the BSP, is guided through a large core fiber (LCF) and detected by a photomultiplier tube. All optical components in the headpiece have millimeter dimensions. (c) Piezo drive signals for x- (blue trace) and y-direction (orange trace; 90° phase shifted). Amplitude modulation is used to create the spiral pattern. (d) ‘Ideal’ (left) and measured (right) scan trajectories, exemplified for a very small number of spiral lines. (e) Two-photon fiberscope images of 10-micron diameter fluorescent microspheres using an ideal (left) and a measured (right) scan trajectory for image reconstruction, respectively. Average laser power levels of 1 mW and 10x averaging were used. Scale bars are 20 µm.

Scan signal generation, data acquisition, and image reconstruction were controlled by a software package custom-written in the LabVIEW-environment (National Instruments, TX). Due to the compactness of the ultra-thin fiber scanner and the 1-mm dimension of the optical components, the total weight of the fiberscope headpiece was only 0.6 grams.

2.2 Characterization of spatial resolution

Functional imaging of cellular and subcellular activity requires micrometer spatial resolution. We therefore characterized the spatial resolution of the two-photon fiberscope by imaging subresolution (500-nm diameter) fluorescent microspheres at different positions throughout the FOV (Fig. 2(a)). Focusing was achieved by attaching the fiberscope headpiece to a micromanipulator. We analyzed the widths of the fluorescence intensity profiles along the lateral and axial directions (the contribution of the finite bead-sizes was estimated to be negligible, in the range of 3–4%). Specifically, intensity profiles were fitted with Gaussian-shaped distributions (f(x)=y 0+C·exp[-2·(x-x 0)22]) to determine the lateral and axial standard deviations σ of the two-photon point-spread function (PSF) (Fig. 2(b) and (c)). The resulting σ-values were multiplied by a factor of [2 ln(2)]1/2 to obtain full-width-at-half-maximum (FWHM). The mean PSF-FWHM was 0.98±0.09 µm laterally and 7.68±1.30 µm axially (mean ±S.D., n=10).

To translate the measured FWHM values to an “effective NA” we used a theory derived from Heisenberg’s uncertainty principle [33

33. S. W. Grill and E. H. K. Stelzer, “Method to calculate lateral and axial gain factors of optical setups with a large solid angle,” J. Opt. Soc. Am. A 16, 2658–2665 (1999). [CrossRef]

]. For single-photon PSFs the lateral and axial FWHM values are given by [34

34. C. J. Engelbrecht and E. H. K. Stelzer, “Resolution enhancement in a light-sheet-based microscope (SPIM),” Opt. Lett. 31, 1477–1479 (2006). [CrossRef] [PubMed]

]

FWHM1p,lat=λ0n32cos(α)cos(2α)
(1)
FWHM1p,ax=λ0n[1cos(α)]
(2)

where λ 0 is the vacuum wavelength, n the refractive index of the immersion medium, and α the half opening angle of the lens (NA=n sin(α)). The widths of the theoretical two-photon PSF were calculated according to FWHM2p=FWHM1p 2-1/2 assuming Gaussian intensity distributions. Eqs. 1 and 2 were numerically solved for α following insertion of the experimentally determined FWHM values. The resultant effective NAs were 0.34±0.03 and 0.38±0.03 for the lateral and axial direction, respectively (Fig. 2(d)). Therefore the Gaussian-shaped beam emerging from the PCF nearly filled the back aperture of the front GRIN-lens. We conclude that the two-photon fiberscope provides micrometer resolution and thus should be well suited for imaging activity of cellular structures.

Fig. 2. Micrometer resolution of the two-photon fiberscope. (a) Two-photon image of 500-nm fluorescent microspheres. Scale bar 3 µm. Fluorescence intensity profiles of an example microsphere along the lateral (b) and the axial (c) dimension. Red traces are Gaussian-shaped curves fitted to the data points. (d) Calculation of the effective NA from the observed lateral (left, red) and axial (right, blue) FWHM. Horizontal lines represent mean ± S.D. The shaded areas correspond to the estimated error ranges of the calculated NAs in lateral and axial direction, respectively. Average laser power levels of 25 mW and 3x averaging were used.

2.3 Calcium measurements in rat cerebellum in vivo

To demonstrate the suitability of the ultra-compact two-photon fiberscope for functional measurements of neural activity, we applied the fiberscope to measure dendritic calcium signals in Purkinje cells of the rat cerebellum in vivo. All experiments were carried out according to the guidelines of the Institute for Laboratory Animal Science of the University Zurich, Switzerland and were approved by the Cantonal Veterinary Office. Long Evans rats (43–49 days old) were anesthetized by an intraperitoneal injection of urethane (1.5 g/kg body weight). A craniotomy was performed above cerebellum as previously described [35

35. M. R. Sullivan, A. Nimmerjahn, D. V. Sarkisov, F. Helmchen, and S. S. Wang, “In vivo calcium imaging of circuit activity in cerebellar cortex,” J. Neurophysiol. 94, 1636–1644 (2005). [CrossRef] [PubMed]

]. The dura was carefully removed and the exposed brain surface was superfused with normal rat Ringer (NRR) solution (in mM: 135 NaCl, 5.4 KCl, 5 HEPES, 1.8 CaCl2, pH 7.2 with NaOH). Population staining with the fluorescent calcium indicator Oregon Green-BAPTA 1 (OGB-1) was performed using the multi-cell bolus loading (MCBL) technique [35

35. M. R. Sullivan, A. Nimmerjahn, D. V. Sarkisov, F. Helmchen, and S. S. Wang, “In vivo calcium imaging of circuit activity in cerebellar cortex,” J. Neurophysiol. 94, 1636–1644 (2005). [CrossRef] [PubMed]

, 36

36. C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, “In vivo two-photon calcium imaging of neuronal networks,” Proc. Natl. Acad. Sci. U S A 100, 7319–7324 (2003). [CrossRef] [PubMed]

]. Briefly, 50 µg of OGB-1 (acetoxymethyl(AM)-ester form; Molecular Probes, CA) were dissolved in 20% Pluronic F-127 (BASF, Florham Park, NJ) in DMSO and diluted in NRR to a final concentration of 1 mM. This solution was pressure injected directly into the molecular layer of cerebellar cortex, about 100–150 µm below the surface. Within one hour, MCBL unspecifically stained various cellular components in the molecular layer including dendrites of Purkinje neurons [35

35. M. R. Sullivan, A. Nimmerjahn, D. V. Sarkisov, F. Helmchen, and S. S. Wang, “In vivo calcium imaging of circuit activity in cerebellar cortex,” J. Neurophysiol. 94, 1636–1644 (2005). [CrossRef] [PubMed]

, 37

37. W. Göbel and F. Helmchen, “New angles on neuronal dendrites in vivo,” J. Neurophysiol. 98, 3770–3779 (2007). [CrossRef] [PubMed]

].

The fiberscope headpiece was lowered onto the surface of the exposed cerebellum without penetrating the tissue. Time-lapse imaging was performed in the upper molecular layer at effective frame rates of 7–15 fps (2–3 frames averaged) and with ~150–200 mW laser power at the sample. We observed spontaneous fluorescence intensity fluctuations in parasagittal stripes, reminiscent of previously reported calcium signals in Purkinje cell dendrites [35

35. M. R. Sullivan, A. Nimmerjahn, D. V. Sarkisov, F. Helmchen, and S. S. Wang, “In vivo calcium imaging of circuit activity in cerebellar cortex,” J. Neurophysiol. 94, 1636–1644 (2005). [CrossRef] [PubMed]

, 37

37. W. Göbel and F. Helmchen, “New angles on neuronal dendrites in vivo,” J. Neurophysiol. 98, 3770–3779 (2007). [CrossRef] [PubMed]

] (Fig. 3(a)). These calcium signals were expressed as relative fluorescence intensity changes (ΔF/F) after background subtraction. Gaussian blurring was used to remove shot noise from the PMT. Pseudocolor-coded movies of fluorescence changes were subsequently rendered in ImageJ (NIH, MD) by merging the time series of ΔF/F images (as hue H and saturation S) with the average fluorescence image (as brightness B) into a single HSB-movie (Fig. 3(a)).

Fig. 3. (a) The cerebellar cortex has a well-defined anatomical organization with flat dendritic trees of Purkinje cells lying in parasagittal planes. In top view, dendritic excitation is evident as band-like calcium signals. Three snapshots from a ΔF/F movie of 33 s total duration are provided. (b and c) Examples of spontaneous ΔF/F-traces in ROIs comprising different dendrites. Note the different time scales. (d) Example of semi-automated ROI-definition by ICA. (e) Corresponding ΔF/F-traces color-coded according to the ROI selections in (d). All scale bars are 15 µm. [Media 1]

4. Discussion

Our fiberscope could be further improved in several ways. First, in the current design, pixel dwell times become shorter with increasing spiral radius because fiber velocities are higher at the periphery. As a consequence, image brightness levels are reduced in the outer part of the images. Modulation of the analog gain with radial position may improve image homogeneity, although it leads to increased noise levels in the periphery. A different compensation mechanism is laser power modulation. However, changing laser power typically affects the resonance properties of the fiber cantilever due to thermal heating. Therefore, a second improvement would be to incorporate a temperature stabilization unit into the spiral scanner. However, brightness normalization is not essential for recording relative fluorescence changes (ΔF/F). As a third improvement a focusing mechanism could be included in the fiberscope headpiece. At present, our microscope is designed for imaging in a fixed two-dimensional plane. Studies on freely moving animals will therefore require positioning and subsequent fixation of the microscope headpiece. If recordings from different optical sections are required during the experiment, remote focusing can be achieved with a miniature DC motor, which adds only little weight [11

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

, 23

23. B. A. Flusberg, J. C. Lung, E. D. Cocker, E. P. Anderson, and M. J. Schnitzer, “In vivo brain imaging using a portable 3.9 gram two-photon fluorescence microendoscope,” Opt. Lett. 30, 2272–2274 (2005). [CrossRef] [PubMed]

]. Most simply, adjustment of the distance between the GRIN tube lens and the fiber scanner can be used to axially shift the focal plane.

Most in vivo two-photon imaging studies have used anesthetized animals. However, correlation of neuronal network activity with behavior is not possible during anesthesia. Two principal approaches may extend functional two-photon microscopy to awake preparations: head-restraint animals [39

39. D. A. Dombeck, A. N. Khabbaz, F. Collman, T. L. Adelman, and D. W. Tank, “Imaging large-scale neural activity with cellular resolution in awake, mobile mice,” Neuron 56, 43–57 (2007). [CrossRef] [PubMed]

] and head-mounted microscopes [11

11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

]. The first approach benefits from the use of conventional two-photon microscopes providing good image quality but may be applicable to a limited set of behavioral paradigms only. The second approach, for which our two-photon fiberscope is designed, promises to enable a wider range of behavioral studies at the cost of slightly deteriorated image quality. Thus, complementary results can be expected from these different approaches. Fiberscope imaging in unrestrained animals promises comprehensive insights into neural network activity and its correlation to specific behaviors. Fields of applications include the functional studies of neocortex, cerebellum and other superficial brain areas as well as endoscopic targeting of subcortical regions [9

9. J. C. Jung, A. D. Mehta, E. Aksay, R. Stepnoski, and M. J. Schnitzer, “In vivo mammalian brain Imaging using one- and two-photon fluorescence microendoscopy,” J. Neurophysiol. 92, 3121–3133 (2004). [CrossRef] [PubMed]

, 10

10. M. J. Levene, D. A. Dombeck, K. A. Kasischke, R. P. Molloy, and W. W. Webb, “In vivo multiphoton microscopy of deep brain tissue,” J. Neurophysiol. 91, 1908–1912 (2004). [CrossRef]

, 18

18. J. C. Jung and M. J. Schnitzer, “Multiphoton endoscopy,” Opt. Lett. 28, 902–904 (2003). [CrossRef] [PubMed]

]. Finally, clinical applications comprising medical imaging and treatment can be envisioned.

5. Conclusion

We have presented a novel ultra-compact two-photon microscope for functional calcium imaging in vivo. Our device is based on miniaturized components for beam steering and epifluorescence detection. The headpiece weighs only 0.6 grams and should be easily portable by a rat or a mouse. Spatial resolution is ~1 µm. Fluorescence collection sensitivity and achievable frame rates are sufficient for reliable recordings of functional calcium transients. We conclude that our fiberscope will be well suited for optical recording of neural network dynamics in awake, freely behaving animals.

Acknowledgments

The authors thank Hansjörg Kasper and Stefan Giger for designing and assembling electronic circuits and mechanical parts, Werner Göbel, David Melville, Cameron Lee and Bernie Murray for helpful discussions, Werner Göbel for comments on the manuscript, Eva Hochreutener for help with the figures, David Melville for building the first PCF-scanner, and Ryland Bryant for introducing C.J.E. to the manufacturing processes of the scan engines at the University of Washington. This work was supported by the NCCR on Neural Plasticity and Repair and a Human Frontier Science Program grant to F.H.

References and links

1.

E. Seibel, T. Soper, R. Johnston, and R. Glenny, “Ultrathin laser scanning bronchoscope and guidance system for the peripheral lung,” Lung Cancer 49, S162–S162 (2005). [CrossRef]

2.

E. J. Seibel, R. S. Johnston, C. M. Brown, J. A. Dominitz, and M. B. Kimmey, “Novel ultrathin scanning fiber endoscope for cholangioscopy and pancreatoscopy,” Gastrointest. Endosc. 65, Ab125–Ab125 (2007). [CrossRef]

3.

A. F. Low, G. J. Tearney, B. E. Bouma, and I. K. Jang, “Technology insight: optical coherence tomography - current status and future development,” Nature Clinical Practice Cardiovascular Medicine 3, 154–162 (2006). [CrossRef] [PubMed]

4.

Z. Yaqoob, J. G. Wu, E. J. McDowell, X. Heng, and C. H. Yang, “Methods and application areas of endoscopic optical coherence tomography,” J. Biomed. Opt. 11, 063001 (2006). [CrossRef]

5.

F. Helmchen, “Miniaturization of fluorescence microscopes using fibre optics,” Exp. Physiol. 87, 737–745 (2002). [CrossRef] [PubMed]

6.

A. D. Mehta, J. C. Jung, B. A. Flusberg, and M. J. Schnitzer, “Fiber optic in vivo imaging in the mammalian nervous system,” Curr. Opin. Neurobiol. 14, 617–628 (2004). [CrossRef] [PubMed]

7.

B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, “Fiber-optic fluorescence imaging,” Nature Methods 2, 941–950 (2005). [CrossRef] [PubMed]

8.

L. Fu and M. Gu, “Fibre-optic nonlinear optical microscopy and endoscopy,” J. Microsc. 226, 195–206 (2007). [CrossRef] [PubMed]

9.

J. C. Jung, A. D. Mehta, E. Aksay, R. Stepnoski, and M. J. Schnitzer, “In vivo mammalian brain Imaging using one- and two-photon fluorescence microendoscopy,” J. Neurophysiol. 92, 3121–3133 (2004). [CrossRef] [PubMed]

10.

M. J. Levene, D. A. Dombeck, K. A. Kasischke, R. P. Molloy, and W. W. Webb, “In vivo multiphoton microscopy of deep brain tissue,” J. Neurophysiol. 91, 1908–1912 (2004). [CrossRef]

11.

F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, “A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals,” Neuron 31, 903–912 (2001). [CrossRef] [PubMed]

12.

H. Adelsberger, O. Garaschuk, and A. Konnerth, “Cortical calcium waves in resting newborn mice,” Nat. Neurosci. 8, 988–990 (2005). [CrossRef] [PubMed]

13.

M. Murayama, E. Perez-Garci, H. R. Luscher, and M. E. Larkum, “Fiberoptic system for recording dendritic calcium signals in layer 5 neocortical pyramidal cells in freely moving rats,” J. Neurophysiol. 98, 1791–1805 (2007). [CrossRef] [PubMed]

14.

W. Denk, J. H. Strickler, and W. W. Webb, “Two-Photon Laser Scanning Fluorescence Microscopy,” Science 248, 73–76 (1990). [CrossRef] [PubMed]

15.

W. Denk and K. Svoboda, “Photon upmanship: why multiphoton imaging is more than a gimmick,” Neuron 18, 351–357 (1997). [CrossRef] [PubMed]

16.

F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nature Methods 2, 932–940 (2005). [CrossRef] [PubMed]

17.

J. Knittel, L. Schnieder, G. Buess, B. Messerschmidt, and T. Possner, “Endoscope-compatible confocal microscope using a gradient index-lens system,” Opt. Commun. 188, 267–273 (2001). [CrossRef]

18.

J. C. Jung and M. J. Schnitzer, “Multiphoton endoscopy,” Opt. Lett. 28, 902–904 (2003). [CrossRef] [PubMed]

19.

W. Göbel, J. N. D. Kerr, A. Nimmerjahn, and F. Helmchen, “Miniaturized two-photon microscope based on a flexible coherent fiber bundle and a gradient-index lens objective,” Opt. Lett. 29, 2521–2523 (2004). [CrossRef] [PubMed]

20.

D. G. Ouzounov, K. D. Moll, M. A. Foster, W. R. Zipfel, W. W. Webb, and A. L. Gaeta, “Delivery of nanojoule femtosecond pulses through large-core microstructured fibers,” Opt. Lett. 27, 1513–1515 (2002). [CrossRef]

21.

F. Helmchen, D. W. Tank, and W. Denk, “Enhanced two-photon excitation through optical fiber by single-mode propagation in a large core,” Appl. Opt. 41, 2930–2934 (2002). [CrossRef] [PubMed]

22.

W. Göbel, A. Nimmerjahn, and F. Helmchen, “Distortion-free delivery of nanojoule femtosecond pulses from a Ti : sapphire laser through a hollow-core photonic crystal fiber,” Opt. Lett. 29, 1285–1287 (2004). [CrossRef] [PubMed]

23.

B. A. Flusberg, J. C. Lung, E. D. Cocker, E. P. Anderson, and M. J. Schnitzer, “In vivo brain imaging using a portable 3.9 gram two-photon fluorescence microendoscope,” Opt. Lett. 30, 2272–2274 (2005). [CrossRef] [PubMed]

24.

M. T. Myaing, D. J. MacDonald, and X. D. Li, “Fiber-optic scanning two-photon fluorescence endoscope,” Opt. Lett. 31, 1076–1078 (2006). [CrossRef] [PubMed]

25.

L. Fu, X. Gan, and M. Gu, “Nonlinear optical microscopy based on double-clad photonic crystal fibers,” Opt. Express 13, 5528–5534 (2005). [CrossRef] [PubMed]

26.

J. Sawinski and W. Denk, “Miniature random-access fiber scanner for in vivo multiphoton imaging,” J. Appl. Phys. 102, (2007). [CrossRef]

27.

W. Piyawattanametha, R. P. J. Barretto, T. H. Ko, B. A. Flusberg, E. D. Cocker, H. J. Ra, D. S. Lee, O. Solgaard, and M. J. Schnitzer, “Fast-scanning two-photon fluorescence imaging based on a microelectromechanical systems two-dimensional scanning mirror,” Opt. Lett. 31, 2018–2020 (2006). [CrossRef] [PubMed]

28.

L. Fu, A. Jain, H. Xie, C. Cranfield, and M. Gu, “Nonlinear optical endoscopy based on a double-clad photonic crystal fiber and a MEMS mirror,” Opt. Express 14, 1027–1032 (2006). [CrossRef] [PubMed]

29.

A. Monfared, N. H. Blevins, E. L. M. Cheung, J. C. Jung, G. Popelka, and M. J. Schnitzer, “In vivo Imaging of mammalian cochlear blood flow using fluorescence microendoscopy,” Otology & Neurotology 27, 144–152 (2006). [CrossRef] [PubMed]

30.

P. Vincent, U. Maskos, I. Charvet, L. Bourgeais, L. Stoppini, N. Leresche, J. P. Changeux, R. Lambert, P. Meda, and D. Paupardin-Tritsch, “Live imaging of neural structure and function by fibred fluorescence microscopy,” EMBO Rep 7, 1154–1161 (2006). [CrossRef] [PubMed]

31.

E. J. Seibel and Q. Y. J. Smithwick, “Unique features of optical scanning, single fiber endoscopy,” Lasers Surg. Med. 30, 177–183 (2002). [CrossRef] [PubMed]

32.

E. J. Seibel, R. S. Johnston, and C. D. Melville, “A full-color scanning fiber endoscope,” Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI. Proceedings of the SPIE. 6083, 9–16 (2006).

33.

S. W. Grill and E. H. K. Stelzer, “Method to calculate lateral and axial gain factors of optical setups with a large solid angle,” J. Opt. Soc. Am. A 16, 2658–2665 (1999). [CrossRef]

34.

C. J. Engelbrecht and E. H. K. Stelzer, “Resolution enhancement in a light-sheet-based microscope (SPIM),” Opt. Lett. 31, 1477–1479 (2006). [CrossRef] [PubMed]

35.

M. R. Sullivan, A. Nimmerjahn, D. V. Sarkisov, F. Helmchen, and S. S. Wang, “In vivo calcium imaging of circuit activity in cerebellar cortex,” J. Neurophysiol. 94, 1636–1644 (2005). [CrossRef] [PubMed]

36.

C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, “In vivo two-photon calcium imaging of neuronal networks,” Proc. Natl. Acad. Sci. U S A 100, 7319–7324 (2003). [CrossRef] [PubMed]

37.

W. Göbel and F. Helmchen, “New angles on neuronal dendrites in vivo,” J. Neurophysiol. 98, 3770–3779 (2007). [CrossRef] [PubMed]

38.

J.-F. Cardoso, “BLIND SOURCE SEPARATION and INDEPENDENT COMPONENT ANALYSIS,” http://www.tsi.enst.fr/~cardoso/guidesepsou.html.

39.

D. A. Dombeck, A. N. Khabbaz, F. Collman, T. L. Adelman, and D. W. Tank, “Imaging large-scale neural activity with cellular resolution in awake, mobile mice,” Neuron 56, 43–57 (2007). [CrossRef] [PubMed]

OCIS Codes
(170.0110) Medical optics and biotechnology : Imaging systems
(170.2150) Medical optics and biotechnology : Endoscopic imaging
(170.2520) Medical optics and biotechnology : Fluorescence microscopy
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(180.2520) Microscopy : Fluorescence microscopy
(180.5810) Microscopy : Scanning microscopy

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: February 12, 2008
Revised Manuscript: March 28, 2008
Manuscript Accepted: April 1, 2008
Published: April 4, 2008

Virtual Issues
Vol. 3, Iss. 5 Virtual Journal for Biomedical Optics

Citation
Christoph J. Engelbrecht, Richard S. Johnston, Eric J. Seibel, and Fritjof Helmchen, "Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo," Opt. Express 16, 5556-5564 (2008)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-16-8-5556


Sort:  Year  |  Journal  |  Reset  

References

  1. E. Seibel, T. Soper, R. Johnston, and R. Glenny, "Ultrathin laser scanning bronchoscope and guidance system for the peripheral lung," Lung Cancer 49, S162-S162 (2005). [CrossRef]
  2. E. J. Seibel, R. S. Johnston, C. M. Brown, J. A. Dominitz, and M. B. Kimmey, "Novel ultrathin scanning fiber endoscope for cholangioscopy and pancreatoscopy," Gastrointest. Endosc. 65, Ab125-Ab125 (2007). [CrossRef]
  3. A. F. Low, G. J. Tearney, B. E. Bouma, and I. K. Jang, "Technology insight: optical coherence tomography - current status and future development," Nature Clinical Practice Cardiovascular Medicine 3, 154-162 (2006). [CrossRef] [PubMed]
  4. Z. Yaqoob, J. G. Wu, E. J. McDowell, X. Heng, and C. H. Yang, "Methods and application areas of endoscopic optical coherence tomography," J. Biomed. Opt. 11, 063001 (2006). [CrossRef]
  5. F. Helmchen, "Miniaturization of fluorescence microscopes using fibre optics," Exp. Physiol. 87, 737-745 (2002). [CrossRef] [PubMed]
  6. A. D. Mehta, J. C. Jung, B. A. Flusberg, and M. J. Schnitzer, "Fiber optic in vivo imaging in the mammalian nervous system," Curr. Opin. Neurobiol. 14, 617-628 (2004). [CrossRef] [PubMed]
  7. B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, "Fiber-optic fluorescence imaging," Nature Methods 2, 941-950 (2005). [CrossRef] [PubMed]
  8. L. Fu and M. Gu, "Fibre-optic nonlinear optical microscopy and endoscopy," J. Microsc. 226, 195-206 (2007). [CrossRef] [PubMed]
  9. J. C. Jung, A. D. Mehta, E. Aksay, R. Stepnoski, and M. J. Schnitzer, "In vivo mammalian brain Imaging using one- and two-photon fluorescence microendoscopy," J. Neurophysiol. 92, 3121-3133 (2004). [CrossRef] [PubMed]
  10. M. J. Levene, D. A. Dombeck, K. A. Kasischke, R. P. Molloy, and W. W. Webb, "In vivo multiphoton microscopy of deep brain tissue," J. Neurophysiol. 91, 1908-1912 (2004). [CrossRef]
  11. F. Helmchen, M. S. Fee, D. W. Tank, and W. Denk, "A miniature head-mounted two-photon microscope: High-resolution brain imaging in freely moving animals," Neuron 31, 903-912 (2001). [CrossRef] [PubMed]
  12. H. Adelsberger, O. Garaschuk, and A. Konnerth, "Cortical calcium waves in resting newborn mice," Nat. Neurosci. 8, 988-990 (2005). [CrossRef] [PubMed]
  13. M. Murayama, E. Perez-Garci, H. R. Luscher, and M. E. Larkum, "Fiberoptic system for recording dendritic calcium signals in layer 5 neocortical pyramidal cells in freely moving rats," J. Neurophysiol. 98, 1791-1805 (2007). [CrossRef] [PubMed]
  14. W. Denk, J. H. Strickler, and W. W. Webb, "Two-Photon Laser Scanning Fluorescence Microscopy," Science 248, 73-76 (1990). [CrossRef] [PubMed]
  15. W. Denk and K. Svoboda, "Photon upmanship: why multiphoton imaging is more than a gimmick," Neuron 18, 351-357 (1997). [CrossRef] [PubMed]
  16. F. Helmchen and W. Denk, "Deep tissue two-photon microscopy," Nature Methods 2, 932-940 (2005). [CrossRef] [PubMed]
  17. J. Knittel, L. Schnieder, G. Buess, B. Messerschmidt, and T. Possner, "Endoscope-compatible confocal microscope using a gradient index-lens system," Opt. Commun. 188, 267-273 (2001). [CrossRef]
  18. J. C. Jung and M. J. Schnitzer, "Multiphoton endoscopy," Opt. Lett. 28, 902-904 (2003). [CrossRef] [PubMed]
  19. W. Göbel, J. N. D. Kerr, A. Nimmerjahn, and F. Helmchen, "Miniaturized two-photon microscope based on a flexible coherent fiber bundle and a gradient-index lens objective," Opt. Lett. 29, 2521-2523 (2004). [CrossRef] [PubMed]
  20. D. G. Ouzounov, K. D. Moll, M. A. Foster, W. R. Zipfel, W. W. Webb, and A. L. Gaeta, "Delivery of nanojoule femtosecond pulses through large-core microstructured fibers," Opt. Lett. 27, 1513-1515 (2002). [CrossRef]
  21. F. Helmchen, D. W. Tank, and W. Denk, "Enhanced two-photon excitation through optical fiber by single-mode propagation in a large core," Appl. Opt. 41, 2930-2934 (2002). [CrossRef] [PubMed]
  22. W. Göbel, A. Nimmerjahn, and F. Helmchen, "Distortion-free delivery of nanojoule femtosecond pulses from a Ti : sapphire laser through a hollow-core photonic crystal fiber," Opt. Lett. 29, 1285-1287 (2004). [CrossRef] [PubMed]
  23. B. A. Flusberg, J. C. Lung, E. D. Cocker, E. P. Anderson, and M. J. Schnitzer, "In vivo brain imaging using a portable 3.9 gram two-photon fluorescence microendoscope," Opt. Lett. 30, 2272-2274 (2005). [CrossRef] [PubMed]
  24. M. T. Myaing, D. J. MacDonald, and X. D. Li, "Fiber-optic scanning two-photon fluorescence endoscope," Opt. Lett. 31, 1076-1078 (2006). [CrossRef] [PubMed]
  25. L. Fu, X. Gan, and M. Gu, "Nonlinear optical microscopy based on double-clad photonic crystal fibers," Opt. Express 13, 5528-5534 (2005). [CrossRef] [PubMed]
  26. J. Sawinski and W. Denk, "Miniature random-access fiber scanner for in vivo multiphoton imaging," J. Appl. Phys. 102, (2007). [CrossRef]
  27. W. Piyawattanametha, R. P. J. Barretto, T. H. Ko, B. A. Flusberg, E. D. Cocker, H. J. Ra, D. S. Lee, O. Solgaard, and M. J. Schnitzer, "Fast-scanning two-photon fluorescence imaging based on a microelectromechanical systems two-dimensional scanning mirror," Opt. Lett. 31, 2018-2020 (2006). [CrossRef] [PubMed]
  28. L. Fu, A. Jain, H. Xie, C. Cranfield, and M. Gu, "Nonlinear optical endoscopy based on a double-clad photonic crystal fiber and a MEMS mirror," Opt. Express 14, 1027-1032 (2006). [CrossRef] [PubMed]
  29. A. Monfared, N. H. Blevins, E. L. M. Cheung, J. C. Jung, G. Popelka, and M. J. Schnitzer, "In vivo Imaging of mammalian cochlear blood flow using fluorescence microendoscopy," Otology & Neurotology 27, 144-152 (2006). [CrossRef] [PubMed]
  30. P. Vincent, U. Maskos, I. Charvet, L. Bourgeais, L. Stoppini, N. Leresche, J. P. Changeux, R. Lambert, P. Meda, and D. Paupardin-Tritsch, "Live imaging of neural structure and function by fibred fluorescence microscopy," EMBO Rep 7, 1154-1161 (2006). [CrossRef] [PubMed]
  31. E. J. Seibel and Q. Y. J. Smithwick, "Unique features of optical scanning, single fiber endoscopy," Lasers Surg. Med. 30, 177-183 (2002). [CrossRef] [PubMed]
  32. E. J. Seibel, R. S. Johnston, and C. D. Melville, "A full-color scanning fiber endoscope," Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI.Proceedings of the SPIE. 6083, 9-16 (2006).
  33. S. W. Grill and E. H. K. Stelzer, "Method to calculate lateral and axial gain factors of optical setups with a large solid angle," J. Opt. Soc. Am. A 16, 2658-2665 (1999). [CrossRef]
  34. C. J. Engelbrecht and E. H. K. Stelzer, "Resolution enhancement in a light-sheet-based microscope (SPIM)," Opt. Lett. 31, 1477-1479 (2006). [CrossRef] [PubMed]
  35. M. R. Sullivan, A. Nimmerjahn, D. V. Sarkisov, F. Helmchen, and S. S. Wang, "In vivo calcium imaging of circuit activity in cerebellar cortex," J. Neurophysiol. 94, 1636-1644 (2005). [CrossRef] [PubMed]
  36. C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, "In vivo two-photon calcium imaging of neuronal networks," Proc. Natl. Acad. Sci. U S A 100, 7319-7324 (2003). [CrossRef] [PubMed]
  37. W. Göbel and F. Helmchen, "New angles on neuronal dendrites in vivo," J. Neurophysiol. 98, 3770-3779 (2007). [CrossRef] [PubMed]
  38. J.-F. Cardoso, "BLIND SOURCE SEPARATION and INDEPENDENT COMPONENT ANALYSIS," http://www.tsi.enst.fr/~cardoso/guidesepsou.html.
  39. D. A. Dombeck, A. N. Khabbaz, F. Collman, T. L. Adelman, and D. W. Tank, "Imaging large-scale neural activity with cellular resolution in awake, mobile mice," Neuron 56, 43-57 (2007). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.