Temperature analysis of low-expansion Fabry-Perot cavities

Richard W. Fox  »View Author Affiliations


Optics Express, Vol. 17, Issue 17, pp. 15023-15031 (2009)
http://dx.doi.org/10.1364/OE.17.015023


View Full Text Article

Acrobat PDF (738 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Temperature dependent structural distortion at the contacted mirrors of low-expansion glass cavities can introduce changes to the cavity length independent of the length of the spacer. There are resulting temperature sensitivities of the path (m/K) at each end of a cavity that are proportional to the difference of the coefficient of thermal expansion (α) at the contact. The temperature sensitivity of the resonant frequency can be a minimum at a temperature TC if the product of length times α(TC) of the spacer approximately offsets the combined sensitivities of the ends, even if α(TC) of the spacer is significantly nonzero.

© 2009 OSA

1. Introduction

This paper presents an approach that should considerably improve predictions of TC for optical cavities when the material parameters are well known. In fact, there are two requirements for making such predictions: (1) a model of frequency versus temperature that includes all relevant dependences and (2) accurately calibrated CTE values for the glass components. This work focuses mainly on the first issue but has implications for the second. The approach is presented by comparing frequency versus temperature measurements of cavities built with the same spacer but different optically-contacted mirrors to highlight the significance of structural distortion at the ends. Finite element analysis (FEA) is used to make an a priori calculation of this effect. Using the frequency measurements from one cavity, this calculation is used to determine the CTE of the spacer. The results are consistent with the frequency measurements of a second cavity, as well as with those made on a third cavity that was built on this spacer without optical contacting in order to minimize structural distortion. This analysis method should be especially useful in the design of shorter cavities, as end effects have a proportionally larger influence as the cavity length decreases. Furthermore, it may lead to a proper choice of spacer glass to achieve a conveniently warm TC when fused silica mirror substrates are employed, which is of interest for thermal noise reasons [10

10. K. Numata, A. Kemery, and J. Camp, “Thermal-noise limit in the frequency stabilization of lasers with rigid cavities,” Phys. Rev. Lett. 93(25), 250602 (2004). [CrossRef]

]. Such an analysis may also be useful in measurements of the CTE of ultra-low expansion materials, which are sometimes accomplished by monitoring the frequency of optical cavities [11

11. J. W. Berthold III and S. F. Jacobs, “Ultraprecise thermal expansion measurements of seven low expansion materials,” Appl. Opt. 15(10), 2344–2347 (1976). [CrossRef]

,12

12. S. F. Jacobs, “Dimensional stability of materials useful in optical engineering,” Opt. Acta (Lond.) 33(11), 1377–1388 (1986). [CrossRef]

].

2. Optical frequency versus temperature

The nominal resonant optical frequencies of the longitudinal modes of a Fabry-perot cavity in vacuum are commonly written
ν=mc2l,
(1)
with m a unit-less integer denoting the longitudinal mode number. For Δl << l o, the dependence of the frequency on length is given by
ν=νo(1Δllo).
(2)
In this equation l o is the physical length of the cavity, defined as the distance between the first surfaces of the opposing dielectric stacks, when the frequency of the m th mode is νo. An incremental change of the physical length from l o could occur from expansion of the spacer, which in the case of a constant CTE is given by Δl = l oαSΔT. In the more pertinent case of a variable CTE material, Δl is found by integrating l oαS(T) over the interval from T o to T. However, Δl may be due to more than just the spacer’s thermal expansion or contraction. Although l oαS(T) describes the temperature sensitivity of the spacer in units of m/K, a significant portion of the total temperature sensitivity of l may be due to distortion at the ends, completely independent of the length of the spacer. The familiar expressions of Eqs. (1) and (2) are modified here by including other factors that can affect a mode’s resonant frequency, including terms to describe the temperature sensitivity of the cavity length at each end. Taking into account the phase shift upon reflection (ϕ) due to the mirrors that occurs twice per round-trip, one may write

ν=c2l(m+2ϕ2π).
(3)

The terms l1(T) and l2(T) account for structural deformation at each end of a Fabry-Perot cavity due to the CTE differences between the mirror substrate and spacer, and to a lesser extent, the coatings. The flat contacting surfaces of the cavity will distort along with the mirror substrate as the constrained components expand differentially. The exact change of the on-axis optical path at each end of a cavity with temperature is likely to be a function of the physical geometry such as the mirror shape, thickness and contacted area, thickness of the coatings, the shape of the spacer near the contact area, and also of the material parameters such as the elastic modulus and Poisson’s ratio in addition to the CTE. The temperature sensitivity of each end of the cavity is included individually because the ends may differ, for instance because one end may have a flat mirror that is contacted over a larger area than the opposing curved mirror. Or alternatively, the mirror substrates may be from different lots of glass or even from locations in the same melt that exhibit a different CTE.

This simple model assumes a uniform CTE of each part and a uniform temperature distribution. Small departures from these ideals will be a perturbation on the model results. However it is possible that the transverse spatial variation of the CTE in some spacers may be large enough that bending of the spacer induces a displacement of the mode spot on the mirror, affecting the frequency versus temperature in a manner not accounted for here. Likewise, effects of other variations of the glass such as air inclusions (bubbles) are not taken into account here, nor are dynamic effects that may become important with sufficient optical power circulating in the cavity. Temperature dependence of the Gouy phase [15

15. A. E. Siegman, Lasers, (University Research Books, 1986).

] and of the glass modulus and Poisson’s ratio [16

16. M. Fukuhara and A. Sampei, “Effects on high-temperature-elastic properties on α-/β-quartz phase transition of fused quartz,” J. Mater. Sci. Lett. 18(10), 751–753 (1999). [CrossRef]

] are not included since these effects are very small. Another influence of temperature on the resonant frequency that is not included in this model may come from the mounting. Strain may develop from the relative movement of fixed mounting points, for instance if a cavity is tightly held at several points by an external structure that has a different CTE than the glass spacer.

The key to this formulism of temperature sensitivities at the cavity ends is a method to approximate l1(T) and l2(T) with sufficient accuracy. The main problem is that the CTE of both the mirror substrate and the end of the spacer are temperature dependent, and generally different. This will often be the case even if ULE substrates are contacted to a ULE spacer, for the glass specifications allow for as much as a 60 ppb/K (ppb = x 10−9) difference in CTE between two samples. The CTE curve of fused silica and CTE bounds of the low-expansion glasses ULE and TSG [3

3. ULE and TSG are trademarks of Corning Inc., and mention here is for technical clarity only and does not imply endorsement.

] are given in Fig. 1
Fig. 1 (Top) Nominal CTE of fused silica per NIST Standard Reference Material 739 [18]. (Lower) CTE bounds of ULE (dashed) and TSG. The ULE(TSG) glass specifications state that within the range of 5 to 35 C, the mean value of the CTE is within the limits of ± 30 ( ± 100) ppb/K [19]. The nominal temperature dependence, which is given graphically in references [12] and [19], is approximately α(T) = K0 + 2.21T – 0.0122T 2 ppb/K [20] over this limited temperature range shown. No data concerning the statistical variability of the higher-order coefficients of the CTE was found during this work.
. (The average CTE of a given length of glassexhibits a curve parallel to and within the bounds shown, dependent upon the TiO2 concentration [17

17. R. R. VanBrocklin, M. J. Edwards, and B. Wells, “Review of Corning’s capabilities for ULE mirror blank manufacturing for an extremely large telescope,” Proc. SPIE 6273–01, 1–11 (2006).

]). A further difficulty is the mechanical strain due to the HR coating. The extremely low-loss ion-beam-sputtered quarter-wave stacks used in high-finesse cavities have a higher modulus than the glass and a much larger CTE.

3. Experiment

Optically contacting mirrors of different materials to the same spacer can readily reveal the effect of structural distortion. Here, a single TSG glass spacer was used with pairs of fused silica and ULE mirrors to form two cavities, and resonant frequency versus temperature was monitored using an Iodine-stabilized HeNe laser as a reference. The spacer was 98.04 mm in length, 25.4 mm in diameter with a 6.3 mm diameter borehole and an air relief hole. The ends were polished to λ/10 flatness for optical contacting. In addition, the outside of the cylinder had a slight flat of width 9 mm that was ground along the spacer length as shown in Fig. 2
Fig. 2 (Left) A cross section of the 98.04 mm long spacer. (Right) Red: optical frequency versus temperature of a cavity mode with fused silica mirrors contacted. Blue: frequency of a cavity mode with ULE mirrors contacted to the same spacer. The missing data are times when the laser is unlocked.
. No information concerning the CTE of this particular piece was known other than the general TSG glass specifications.

First, fused silica mirrors (Corning 7980 standard grade) were contacted to the ends of the spacer to form a cavity resonant at 633 nm, and frequency versus temperature data was collected as detailed in ref [21

21. R. W. Fox, “Fabry-Perot temperature dependence and surface-mounted optical cavities,” Proc. SPIE 7099 (2008). (available as arXiv:0807.0656v1 [physics.optics]).

] (this reference incorrectly states the spacer length as 99 mm). The high-finesse mirrors were 25.4 mm in diameter, 5.75 mm thick with a 50 cm radius of curvature and the SiO2/Ta2O5 layers were coated by ion-beam sputtering, as was the backside AR coating. An annular flat region for optical contacting extended from a central diameter of 19.1 mm to the chamfered edge, and the resulting sagitta at the center was 91 μm.

Next, the fused silica mirrors were carefully removed from the TSG spacer and a set of ULE mirrors contacted. These substrates had the same geometry and had the same coatings as the fused silica mirrors but were 5.96 mm thick. No information was known about the CTE of the ULE substrates. As with the prior test, the frequency was recorded with the cavity inside a thick-walled blackbody cell while the outside temperature of the vacuum chamber was very slowly changed over the course of approximately 70 hours. The temperature was monitored by the same thermistor still affixed directly to the TSG spacer with a small (< 1 mm3) quantity of epoxy. Traces of longitudinal mode frequency of the two cavities over a temperature range limited by the experimental set-up are shown in Fig. 2. A potential systematic error exists if the rate of temperature change was insufficiently small, due to the lag between thetemperature indication on the surface of the spacer and the actual temperature distribution inside the glass components. This would likely affect the fused silica data more than the ULE data since the dimensional change per degree is much larger.

Clearly the frequency data is not consistent with a simple picture of expansion of the spacer and mirror sagitta without distortion of the geometry. In each cavity the mirror sagitta (91 μm) account for less than 0.2% of the total cavity length. Yet the frequency traces corresponding to the ULE and fused silica mirrors form parabolic curves that peak at TC = + 16.6 and TC ≅ − 15 ± 5 C respectively. Instead, as will be concluded in the following section, the outward displacement at the center of the constrained fused silica mirror per degree is much larger (≅ 3 nm/K) than would be expected of a simple expansion of the mirror radius sagitta (≅ 40 pm/K).

4. Finite Element Analysis

Our FEA package was not designed to handle calculations over a temperature interval if the CTE values are temperature dependent [22

22. Ansys Workbench Simulation, V11.0. Mentioned only for technical clarity, no endorsement implied.

]. (Although the CTE may be entered into the software as a function of temperature, over a temperature interval from T1 to T2 only the value α(T2) is used). However, if the appropriate constant CTE values are assumed for each component, the on-axis temperature sensitivity of the end of a cavity at a discrete temperature can be readily calculated. Each FEA analysis modeled a spacer with one HR coated mirror attached with no slippage of the optical contact. The spacer was held by specifying the opposing flat surface of the spacer as a fixed support, and the general procedure was as follows here. First, CTE values were assigned to the spacer, mirror substrate and HR coating, αS, αM and αC respectively. Next, the displacement of the center of the coating surface along the optical axis due to a + 1 K temperature shift was calculated. Finally, a second FEA calculation with the substrate and coating suppressed determined the change in the length of the spacer itself in the absence of end distortion. Although the change in the FEA spacer’s length in one degree may be closely approximated by l SαS, the second calculation was performed instead as there is a small perturbation of the length caused by constraining the end with a fixed support. The difference between the two FEA calculations yields ln, the change of the cavity length per K due to the structural changes at one end corresponding to these static values of αS, αM, and αC. An example calculation output is shown in Fig. 3
Fig. 3 The FEA graphical results of deformation in the direction of the optical axis caused by a + 1 K temperature change for the case of a 5.75 mm thick fused silica mirror contacted to the one inch diameter spacer, with αS = 0 ppb/K and αM = 520 ppb/K. The range of the color bar is + 3.5 nm to −2.5 nm, with positive numbers indicating an elongation of the cavity. (Left) The coated mirror with the spacer hidden. The center of the coating moves + 3.2 nm/K in the z direction, while the contacted area (outside the white ring) distorts in the opposite direction. (Right) The end of the spacer with the coating and substrate hidden.
.

A caveat to this analysis is that the CTE differences must be such that a linear relationship exists between stress and strain, i.e., Hooke’s Law must apply. The stress at the optical contact region was investigated using FEA under the condition that the CTE difference between the spacer and a substrate was 620 ppb/K, the most expected from a fused silica mirror on a low-expansion substrate. With a temperature shift of 15 K the maximum stress was found to be less than 1% of the tensile strength. Consequently, the stress/strain relationship should be linear in the range of 20 ± 15 C for the mirror configurations analyzed in this work.

An elastic modulus and Poisson’s ratio of 67.6 GPa and 0.17 respectively are used for the ULE and TSG glass in these calculations, and 72.7 GPa and 0.16 respectively for fused silica. The coating was modeled as a single 5 μm thick layer. Following work on the effective parameters of ion-beam sputtered Ta2O5/SiO2 thin-film coatings, the HR coating parameters ascribed here are a modulus of 106 GPa, a Poisson’s ratio of 0.21, and a temperature-independent CTE equal to 1.8 × 10−6 K−1 [23

23. D. Coyne, “Beamsplitter coating strain induced radius of curvature,” LIGO document LIGO-T050057–00-D (2005).

]. Care was taken to ensure appropriate finite-element mesh sizes consistent with the sizes of the three parts and the area of the two interfaces. A mesh of 50,000 finite elements was employed, and increasing the number of elements changed the typical result by 0.1%.

With the CTE of each glass part specified by a temperature dependent polynomial, the on-axis distortion at the end of a cavity structure as a function of temperature may be found in the manner described above by assigning the appropriate CTE values and repeating the calculation procedure at a sufficient number of distinct temperatures. However, if a different spacer or mirror glass CTE is considered during the cavity design process, a new set of FEA calculations is necessary. Alternatively, the observation that the on-axis distortion (and subsequent change in TC [9

9. M. Notcutt, C. T. Taylor, A. G. Mann, and D. G. Blair, “Temperature compensation for cryogenic cavity stabilized lasers,” J. Phys. D Appl. Phys. 28(9), 1807–1810 (1995). [CrossRef]

]) is proportional to the CTE difference between a mirror substrate and the spacer offers a way to avoid additional FEA calculations. The on-axis distortion is linear with respect to αM − αS for a given physical structure and material parameters but does but not depend on the absolute value of either CTE. Figure 4
Fig. 4 The temperature sensitivity of one end of a cavity as a function of the CTE difference between a fused silica mirror and the TSG spacer shown in Fig. 2. Red dots, calculated ln values for the case of a 25.4 mm diameter, 5.75 mm thick mirror with the geometry and coating described in the text. A linear fit, and the fit residuals (crosses, right axis) are also shown.
shows the distortion calculated for arbitrary values of αM and αS that are expected within the temperature range of interest. To extract numbers from the graphical output a cursor was placed manually at the center of the coating. The possible error from a slight misplacement of the probe from the center was small relative to the ln magnitudes, but likely contributed to the residuals shown in Fig. 4.

The linear dependence of the end distortion on the CTE difference can be established from two or more points by determining the coefficients to Eq. (5).
ln(T)=A(αMαS)+B.
(5)
The coefficients are thus independent of the CTE of the spacer and the substrates, but unique to the physical geometry, the elastic modulus and Poisson’s ratio and the coating parameters. The temperature dependence of ln follows naturally since the CTE of each part may be expressed as, for instance, α = K0 + K1T + K2T2.

With coated fused silica mirrors and spacer of the dimensions given, the calculations shown in Fig. 4 are described by ln(T) = [0.00616(αM−αS) − 0.03] × 10−9 m/K, with the CTE in ppb/K. The identical calculation was performed for the case of the slightly thicker ULE mirror, and in that case the result is ln(T) = [0.00596(αM−αS) − 0.04] × 10−9 m/K.

Initial calculations included a fourth part, a thin (250 nm) AR coating on the substrate’s back side. By simulations of the frequency versus temperature, using Eq. (4), it was observed that introducing the AR coating to the model of these 6 mm thick mirrors caused TC of the 10 cm long cavity to shift by much less than 100 mK. Consequently the AR coating is neglected here.

5. Comparison with experiment

The hypothesis presented is that the cavity mode’s frequency versus temperature behavior can be predicted by Eq. (4) in conjunction with FEA calculations of the end’s temperature sensitivity. This hypothesis can be tested by comparing Eq. (4) to parabolic frequency measurements of a single cavity, but only if the absolute CTE of the spacer glass is known with certainty. Evaluating Eq. (4) against measurements gleaned from different mirrors on the same spacer that are conducted over a wide temperature range would allow independent conclusions of the spacer CTE to be compared, thus testing the hypothesis. That is the route followed here, except that the data traces shown in Fig. 2 are over a limited temperature span. Comparing any two polynomials (such as Eq. (4) and a measured frequency profile) over a limited range of the independent variable is difficult since a good fit can be obtained for a range of any one coefficient as long as the other coefficients are allowed sufficient freedom. This problem is avoided here by accepting the well established CTE curvature of the spacer glass (Fig. 1), and comparing only the absolute value (or y-intercept) of the spacer’s CTE that each frequency measurement indicates. In fitting Eq. (4) to each data trace of Fig. 2 only one free parameter was adjusted, the coefficient K0 of the polynomial defining the spacer’s absolute CTE at 0 C.

The CTE of fused silica is taken as αM(T) = 435.1 + 2.66T − 0.014T2 ppb/K, with T in Celsius, which was derived from the more complicated exponential expression of ref [18

18. Fused Silica CTE from 5 − 35 C adapted from the expression in SRM 739: http://ts.nist.gov/MeasurementServices/ReferenceMaterials/archived_certificates/739.pdf

] over the temperature range from 5 − 35 C. All of the remaining terms in Eq. (4) have been given previously except for ν0 and T0, which were selected at random from the frequency versus temperature data pairs. Of course, uncertainty of the “best fit” K0 of the spacer CTE results from uncertainty in the CTE of the mirror substrates.

The fused silica measurements and material uncertainty of −30 to + 30 ppb/K [18

18. Fused Silica CTE from 5 − 35 C adapted from the expression in SRM 739: http://ts.nist.gov/MeasurementServices/ReferenceMaterials/archived_certificates/739.pdf

] correspond to a range of zero crossings of the spacer’s CTE curve from 17 to 22 C respectively (Fig. 5
Fig. 5 The spacer CTE as deduced from a fit to the frequency data of Fig. 2 using Eq. (4) and the FEA calculations of end distortion. The expansivity, or curvature of the spacer’s CTE is assumed as given in Fig. 1 and only the absolute value (y-intercept) is solved for. The band between the red traces is the range of the spacer CTE resulting from a fit to the cavity data with fused silica mirrors contacted. Similarly, the band between the blue traces is the range of the spacer CTE resulting from a fit to the cavity data with ULE mirrors contacted. In both cases the width of the uncertainty band is due to the mirror CTE uncertainty. Consistent with both measurements are CTE curves that cross zero CTE between 17 and 19 C (cross hatched).
, between the red traces). Likewise, the ULE measurements and mirror substrate uncertainty (K0 of ULE is between −68.5 and −8.5 ppb/K) leads to a range of probable zero crossings of the same spacer from about 14.6 to 19 C respectively (between the blue traces), and partially overlapping the fused silica results in the region from 17 to 19 C.

The fused silica and ULE measurements taken together indicate that the spacer CTE crosses zero between 17 and 19 C. Notably, a second reference of the CTE of fused silica at 273 K [24

24. M. Okaji, N. Yamada, K. Nara, and H. Kato, “Laser interferometric dilatometer at low temperatures: application to fused silica SRM 739,” Cryogenics 35(12), 887–891 (1995). [CrossRef]

] is in agreement with reference [18

18. Fused Silica CTE from 5 − 35 C adapted from the expression in SRM 739: http://ts.nist.gov/MeasurementServices/ReferenceMaterials/archived_certificates/739.pdf

] only if αM(0) < 420 ppb/K, corresponding to the upper portion of the band between the red traces in Fig. 5 that partially overlaps with the ULE results. Finally, frequency versus temperature of a third cavity built on this spacer was measured previously [21

21. R. W. Fox, “Fabry-Perot temperature dependence and surface-mounted optical cavities,” Proc. SPIE 7099 (2008). (available as arXiv:0807.0656v1 [physics.optics]).

]. As described in the reference, the mirrors were not contacted on the ends of the spacer, but instead were lightly bonded to the flat of the spacer shown in Fig. 2 in a manner designed not to provide stability but to minimize differential thermal expansion between the glass parts. The frequency versus temperature peaked at TC ~17.5 C, in excellent agreement with the overlapping region of Fig. 5.

6. Summary

References and links

1.

Ch. Eisele, M. Okhapkin, A. Yu. Nevsky, and S. Schiller, “A crossed optical cavities apparatus for a precision test of the isotropy of light propagation,” Opt. Commun. 281(5), 1189–1196 (2008). [CrossRef]

2.

A. D. Ludlow, X. Huang, M. Notcutt, T. Zanon-Willette, S. M. Foreman, M. M. Boyd, S. Blatt, and J. Ye, “Compact, thermal-noise-limited optical cavity for diode laser stabilization at 1x10(-15).,” Opt. Lett. 32(6), 641–643 (2007). [CrossRef] [PubMed]

3.

ULE and TSG are trademarks of Corning Inc., and mention here is for technical clarity only and does not imply endorsement.

4.

S. A. Webster, M. Oxborrow, S. Pugla, J. Millo, and P. Gill, “Thermal-noise-limited optical cavity,” Phys. Rev. A 77(3), 033847 (2008). [CrossRef]

5.

N. M. Sampas, E. K. Gustafson, and R. L. Byer, “Long-term stability of two diode-laser-pumped nonplanar ring lasers independently stabilized to two Fabry-Perot interferometers,” Opt. Lett. 18(12), 947–949 (1993). [CrossRef] [PubMed]

6.

N. Poli, R. E. Drullinger, G. Ferrari, M. Prevedelli, F. Sorrentino, M. G. Tarallo, and G. M. Tino, “Prospect for a compact strontium optical lattice clock,” Proc. SPIE 6673 (2007).

7.

J. Alnis, A. Matveev, N. Kolachevsky, T. Wilken, Th. Udem, and T. W. Hänsch, “Subhertz linewidth diode lasers by stabilization to vibrationally and thermally compensated ultralow-expansion glass Fabry-Pérot cavities,” Phys. Rev. A 77(5), 053809 (2008). [CrossRef]

8.

S. F. Jacobs, J. W. Berthold, and J. Osmundsen, “Ultraprecise measurement of thermal expansion coefficients – recent progress,” AIP conf. Proceedings, New York (1970).

9.

M. Notcutt, C. T. Taylor, A. G. Mann, and D. G. Blair, “Temperature compensation for cryogenic cavity stabilized lasers,” J. Phys. D Appl. Phys. 28(9), 1807–1810 (1995). [CrossRef]

10.

K. Numata, A. Kemery, and J. Camp, “Thermal-noise limit in the frequency stabilization of lasers with rigid cavities,” Phys. Rev. Lett. 93(25), 250602 (2004). [CrossRef]

11.

J. W. Berthold III and S. F. Jacobs, “Ultraprecise thermal expansion measurements of seven low expansion materials,” Appl. Opt. 15(10), 2344–2347 (1976). [CrossRef]

12.

S. F. Jacobs, “Dimensional stability of materials useful in optical engineering,” Opt. Acta (Lond.) 33(11), 1377–1388 (1986). [CrossRef]

13.

H. Takashashi, “Temperature stability of thin-film narrow-bandpass filters produced by ion-assisted deposition,” Appl. Opt. 34(4), 667–675 (1995). [CrossRef] [PubMed]

14.

Based on the slope of the phase of the Ta2O5-SiO2 coatings used here, 10 mr/nm, and a spectral shift of such coatings on low expansion glass of Δν/ν ~−10−5 K−1 [13].

15.

A. E. Siegman, Lasers, (University Research Books, 1986).

16.

M. Fukuhara and A. Sampei, “Effects on high-temperature-elastic properties on α-/β-quartz phase transition of fused quartz,” J. Mater. Sci. Lett. 18(10), 751–753 (1999). [CrossRef]

17.

R. R. VanBrocklin, M. J. Edwards, and B. Wells, “Review of Corning’s capabilities for ULE mirror blank manufacturing for an extremely large telescope,” Proc. SPIE 6273–01, 1–11 (2006).

18.

Fused Silica CTE from 5 − 35 C adapted from the expression in SRM 739: http://ts.nist.gov/MeasurementServices/ReferenceMaterials/archived_certificates/739.pdf

19.

ULE and TSG Product Information Sheets, Corning Advanced Optics, 334 County Rt., 16, Canton, New York 13617 (2006). http://www.corning.com/docs/specialtymaterials/pisheets/TSGBro91106.pdfhttp://www.corning.com/docs/specialtymaterials/pisheets/UleBro91106.pdf

20.

M. Edwards, Corning Inc., private communication, α(T) = K0 + 2.21T - 0.0122T2 + 1.88e-5T3 ppb/K, T in Celsius (2006).

21.

R. W. Fox, “Fabry-Perot temperature dependence and surface-mounted optical cavities,” Proc. SPIE 7099 (2008). (available as arXiv:0807.0656v1 [physics.optics]).

22.

Ansys Workbench Simulation, V11.0. Mentioned only for technical clarity, no endorsement implied.

23.

D. Coyne, “Beamsplitter coating strain induced radius of curvature,” LIGO document LIGO-T050057–00-D (2005).

24.

M. Okaji, N. Yamada, K. Nara, and H. Kato, “Laser interferometric dilatometer at low temperatures: application to fused silica SRM 739,” Cryogenics 35(12), 887–891 (1995). [CrossRef]

OCIS Codes
(020.0020) Atomic and molecular physics : Atomic and molecular physics
(120.2230) Instrumentation, measurement, and metrology : Fabry-Perot
(120.3930) Instrumentation, measurement, and metrology : Metrological instrumentation
(160.2750) Materials : Glass and other amorphous materials

ToC Category:
Materials

History
Original Manuscript: June 1, 2009
Revised Manuscript: July 22, 2009
Manuscript Accepted: July 23, 2009
Published: August 10, 2009

Citation
Richard W. Fox, "Temperature analysis of low-expansion Fabry-Perot cavities," Opt. Express 17, 15023-15031 (2009)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-17-17-15023


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. Ch. Eisele, M. Okhapkin, A. Yu. Nevsky, and S. Schiller, “A crossed optical cavities apparatus for a precision test of the isotropy of light propagation,” Opt. Commun. 281(5), 1189–1196 (2008). [CrossRef]
  2. A. D. Ludlow, X. Huang, M. Notcutt, T. Zanon-Willette, S. M. Foreman, M. M. Boyd, S. Blatt, and J. Ye, “Compact, thermal-noise-limited optical cavity for diode laser stabilization at 1x10(-15).,” Opt. Lett. 32(6), 641–643 (2007). [CrossRef] [PubMed]
  3. ULE and TSG are trademarks of Corning Inc., and mention here is for technical clarity only and does not imply endorsement.
  4. S. A. Webster, M. Oxborrow, S. Pugla, J. Millo, and P. Gill, “Thermal-noise-limited optical cavity,” Phys. Rev. A 77(3), 033847 (2008). [CrossRef]
  5. N. M. Sampas, E. K. Gustafson, and R. L. Byer, “Long-term stability of two diode-laser-pumped nonplanar ring lasers independently stabilized to two Fabry-Perot interferometers,” Opt. Lett. 18(12), 947–949 (1993). [CrossRef] [PubMed]
  6. N. Poli, R. E. Drullinger, G. Ferrari, M. Prevedelli, F. Sorrentino, M. G. Tarallo, and G. M. Tino, “Prospect for a compact strontium optical lattice clock,” Proc. SPIE 6673 (2007).
  7. J. Alnis, A. Matveev, N. Kolachevsky, T. Wilken, Th. Udem, and T. W. Hänsch, “Subhertz linewidth diode lasers by stabilization to vibrationally and thermally compensated ultralow-expansion glass Fabry-Pérot cavities,” Phys. Rev. A 77(5), 053809 (2008). [CrossRef]
  8. S. F. Jacobs, J. W. Berthold, and J. Osmundsen, “Ultraprecise measurement of thermal expansion coefficients – recent progress,” AIP conf. Proceedings, New York (1970).
  9. M. Notcutt, C. T. Taylor, A. G. Mann, and D. G. Blair, “Temperature compensation for cryogenic cavity stabilized lasers,” J. Phys. D Appl. Phys. 28(9), 1807–1810 (1995). [CrossRef]
  10. K. Numata, A. Kemery, and J. Camp, “Thermal-noise limit in the frequency stabilization of lasers with rigid cavities,” Phys. Rev. Lett. 93(25), 250602 (2004). [CrossRef]
  11. J. W. Berthold and S. F. Jacobs, “Ultraprecise thermal expansion measurements of seven low expansion materials,” Appl. Opt. 15(10), 2344–2347 (1976). [CrossRef]
  12. S. F. Jacobs, “Dimensional stability of materials useful in optical engineering,” Opt. Acta (Lond.) 33(11), 1377–1388 (1986). [CrossRef]
  13. H. Takashashi, “Temperature stability of thin-film narrow-bandpass filters produced by ion-assisted deposition,” Appl. Opt. 34(4), 667–675 (1995). [CrossRef] [PubMed]
  14. Based on the slope of the phase of the Ta2O5-SiO2 coatings used here, 10 mr/nm, and a spectral shift of such coatings on low expansion glass of Δν/ν ~−10−5 K−1 [13].
  15. A. E. Siegman, Lasers, (University Research Books, 1986).
  16. M. Fukuhara and A. Sampei, “Effects on high-temperature-elastic properties on α-/β-quartz phase transition of fused quartz,” J. Mater. Sci. Lett. 18(10), 751–753 (1999). [CrossRef]
  17. R. R. VanBrocklin, M. J. Edwards, and B. Wells, “Review of Corning’s capabilities for ULE mirror blank manufacturing for an extremely large telescope,” Proc. SPIE 6273–01, 1–11 (2006).
  18. Fused Silica CTE from 5 − 35 C adapted from the expression in SRM 739: http://ts.nist.gov/MeasurementServices/ReferenceMaterials/archived_certificates/739.pdf
  19. ULE and TSG Product Information Sheets, Corning Advanced Optics, 334 County Rt., 16, Canton, New York 13617 (2006). http://www.corning.com/docs/specialtymaterials/pisheets/TSGBro91106.pdf http://www.corning.com/docs/specialtymaterials/pisheets/UleBro91106.pdf
  20. M. Edwards, Corning Inc., private communication, α(T) = K0 + 2.21T - 0.0122T2 + 1.88e-5T3 ppb/K, T in Celsius (2006).
  21. R. W. Fox, “Fabry-Perot temperature dependence and surface-mounted optical cavities,” Proc. SPIE 7099 (2008). (available as arXiv:0807.0656v1 [physics.optics]).
  22. Ansys Workbench Simulation, V11.0. Mentioned only for technical clarity, no endorsement implied.
  23. D. Coyne, “Beamsplitter coating strain induced radius of curvature,” LIGO document LIGO-T050057–00-D (2005).
  24. M. Okaji, N. Yamada, K. Nara, and H. Kato, “Laser interferometric dilatometer at low temperatures: application to fused silica SRM 739,” Cryogenics 35(12), 887–891 (1995). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 
Fig. 4 Fig. 5
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited