Enhanced fluorescence signal in nonlinear microscopy through supplementary fiber-optic light collection

Christoph J. Engelbrecht, Werner Göbel, and Fritjof Helmchen  »View Author Affiliations


Optics Express, Vol. 17, Issue 8, pp. 6421-6435 (2009)
http://dx.doi.org/10.1364/OE.17.006421


View Full Text Article

Acrobat PDF (1597 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Nonlinear microscopy techniques crucially rely on efficient signal detection. Here, we present a ring of large-core optical fibers for epi-collection of fluorescence photons that are not transmitted through the objective and thus normally wasted. Theoretical treatments indicated that such a supplementary fiber-optic light collection system (SUFICS) can provide an up to 4-fold signal gain. In typical in vivo imaging experiments, the fiber-ring channel was brighter than the objective channel down to 800 μm depth, thus providing a gain >2. Moreover, SUFICS reduced noise levels in calcium imaging experiments by about 23%. We recommend SUFICS as a generally applicable, effective add-on to nonlinear microscopes for enhancing fluorescence signals.

© 2009 Optical Society of America

1. Introduction

Nonlinear optical microscopy is the current gold standard for high-resolution in vivo imaging deep inside biological tissue [1

1. W. R. Zipfel, R. M. Williams, and W. W. Webb, “Nonlinear magic: multiphoton microscopy in the biosciences,” Nat. Biotechnol. 21, 1369–1377 (2003). [CrossRef] [PubMed]

, 2

2. J. Mertz, “Nonlinear microscopy: new techniques and applications,” Curr. Opin. Neurobiol. 14, 610–616 (2004). [CrossRef] [PubMed]

]. Most prominently, two-photon excited fluorescence laser-scanning microscopy (2PLSM) [3

3. W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248, 73–76 (1990). [CrossRef] [PubMed]

] is widely applied in the life sciences and particularly in neuroscience [4

4. F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nat. Methods 2, 932–940 (2005). [CrossRef] [PubMed]

, 5

5. K. Svoboda and R. Yasuda, “Principles of two-photon excitation microscopy and its applications to neuroscience,” Neuron 50, 823–839 (2006). [CrossRef] [PubMed]

]. A key advantage of nonlinear fluorescence excitation is its three-dimensional confinement to the focus of the microscope objective, resulting in a point-like, isotropically emitting fluorescence source. As a consequence, all fluorescence photons -whether scattered or not - carry useful signal information and as many of them as possible should be collected on the photodetector to maximize the signal-to-noise ratio (SNR) [6

6. W. Denk, D. W. Piston, and W. W. Webb, “Multi-photon molecular excitation in laser-scanning microscopy,” in Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum Press, New York, 2005).

].

Several approaches have been devised to optimize fluorescence collection. First, whole-area detection in non-descanned positions [6

6. W. Denk, D. W. Piston, and W. W. Webb, “Multi-photon molecular excitation in laser-scanning microscopy,” in Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum Press, New York, 2005).

] is used by default nowadays. Second, low-magnification, high numerical aperture (NA) objectives with large effective field-of-views are especially well suited for epi-fluorescence collection from scattering tissue [7

7. M. Oheim, E. Beaurepaire, E. Chaigneau, J. Mertz, and S. Charpak, “Two-photon microscopy in brain tissue: parameters influencing the imaging depth (vol 111, pg 29, 2001),” J. Neurosci. Meth. 112, 205–205 (2001). [CrossRef]

, 8

8. E. Beaurepaire and J. Mertz, “Epifluorescence collection in two-photon microscopy,” Appl. Opt. 41, 5376–5382 (2002). [CrossRef] [PubMed]

]. Third, an up to 3-fold gain in fluorescence can be achieved for tissue slices by additional fluorescence collection through a high-NA condensor in the trans-illumination pathway [9

9. H. J. Koester, D. Baur, R. Uhl, and S. W. Hell, “Ca2+ fluorescence imaging with pico- and femtosecond two-photon excitation: signal and photodamage,” Biophys. J. 77, 2226–2236 (1999). [CrossRef] [PubMed]

, 10

10. Z. F. Mainen, M. Maletic-Savatic, S. H. Shi, Y. Hayashi, R. Malinow, and K. Svoboda, “Two-photon imaging in living brain slices,” Methods 18, 231–239 (1999). [CrossRef] [PubMed]

]. The latter method cannot be utilized, however, for in vivo imaging experiments because photon collection from intact tissue is restricted to the ‘epi’-hemisphere that is oriented towards the objective.

Despite the general acknowledgment of the importance of maximizing fluorescence collection, many signal-carrying photons are still wasted in current microscope implementations. For example, in vivo imaging experiments typically require long working-distance objectives with NAobj in the range of 0.8-1.0. For such objectives and for transparent samples, the maximum collection efficiency η obj is only 10-17% (normalized to the full solid angle 4π). Thus, considering epifluorescence collection only, a three- to five-fold signal gain is in principle possible.

Additional photons can be collected outside of the objective. Such an ‘external detection’ scheme [6

6. W. Denk, D. W. Piston, and W. W. Webb, “Multi-photon molecular excitation in laser-scanning microscopy,” in Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum Press, New York, 2005).

] may be realized with the help of reflective surfaces [11

11. D. Vucinic, T. M. Bartol, and T. J. Sejnowski, “Hybrid reflecting objectives for functional multiphoton microscopy in turbid media,” Opt. Lett. 31, 2447–2449 (2006). [CrossRef] [PubMed]

, 12

12. C. A. Combs, A. V. Smirnov, J. D. Riley, A. H. Gandjbakhche, J. R. Knutson, and R. S. Balaban, “Optimization of multiphoton excitation microscopy by total emission detection using a parabolic light reflector,” J. Microsc. 228, 330–337 (2007). [CrossRef] [PubMed]

]. Here, we demonstrate a simple alternative approach for supplementary epifluorescence collection using large-core optical fibers similar to those previously used for fluorescence collection in miniaturized fiber-optic microscopes [13

13. B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, “Fiber-optic fluorescence imaging,” Nat. Methods 2, 941–950 (2005). [CrossRef] [PubMed]

, 14

14. C. J. Engelbrecht, R. S. Johnston, E. J. Seibel, and F. Helmchen, “Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo,” Opt. Express 16, 5556–5564 (2008). [CrossRef] [PubMed]

]. We demonstrate theoretically and experimentally that a ring-like arrangement of multiple optical fibers efficiently collects fluorescence photons, providing an extra image channel that is brighter than the image obtained through a 40x/0.8W objective. Using in vivo two-photon imaging of mouse neocortex we show that this more than 2-fold signal gain (total collected fluorescence divided by fluorescence in objective channel) is maintained throughout the cortical depths and that it improves the SNR of calcium measurements from neuronal populations.

2. Materials and methods

2.1 Optical setup

The principle of our method is to collect photons that emerge into the epi-hemisphere but outside of the collection cone of the objective. The idea is to use a set of large-core optical fibers and direct their tips towards the fluorescence source (Fig. 1(a)). We used a custom-built two-photon laser scanning microscope with a Ti:sapphire laser system (Tsunami and Millenia-X; Spectra-Physics). Laser pulses were prechirped with a pair of SF10 prisms to compensate for pulse broadening in the microscope beam path. We modulated laser intensity with a Pockel’s cell (Conoptics) and expanded beam size with a lens telescope (Thorlabs). For single-fiber characterization we used a 10x air objective (UPlanFLN, NA 0.3; Olympus). For all other experiments we used a 40x water immersion objective (LUMPlanFl/IR, NA 0.8; Olympus). Fluorescence images were acquired with custom software written in the LabView environment (National Instruments).

Fig. 1. Supplementary epifluorescence collection through a ring of optical fibers. (a) Top: CAD-drawing of a custom fiber-ring holder placed under an objective. Bottom: Closeup view showing the ring-like arrangement of the fiber tips. Only five of eight fibers are shown. Fluorophores are 2-photon excited in the focus of an infrared laser beam (red), causing isotropic fluorescence emission (green). (b) Left: Top view of the ring-like arrangement of eight 1-mm diameter fibers. Right: Dual-channel detection in a custom 2PLSM setup. Optical fibers were bundled and placed in front of a second PMT.

The normal epi-fluorescence detection pathway included a dichroic mirror (KS 93/45°; Linos), a large collector lens (01 LAG 014/066; Melles Griot), a secondary collector lens (G063012000; Linos), and a condensor lens (G063097000; Linos) in front of the photomultiplier tube (PMT).

For fiber-optic fluorescence collection we used optical grade fiber optics from Edmund Optics (Karlsruhe, Germany). These fibers exhibit low attenuation (max. 0.15 dB/m) and have numerical apertures of NAf = 0.51 (according to NAf = (n2 1-n2 2)1/2 with core and cladding refractive indices n1 = 1.492 and n2 = 1.402, respectively). For single-fiber measurements we used a 1.5-mm diameter fiber (NT02-535, 1.47 mm core diameter). To maximize fiber collection we arranged 8 large-core optical fibers (1-mm outer diameter; r0 = 0.49 mm; NAf = 0.51) in a ring such that their tips were touching each other (Fig. 1(a)). We bundled the remote ends of the fibers by inserting them into an array of holes in a custom Teflon piece and projected their output light onto an additional PMT (Fig. 1(b)). This detection scheme thus provided a second ‘fiber-ring’ signal channel in addition to the regular ‘objective’ channel. To make collection conditions as similar as possible we used identical PMTs (R6357, Hamamatsu) in the two channels with always the same supply voltage. A 2-mm thick BG39 color glass served as emission filter in both detection pathways.

For imaging in non-scattering tissue our custom fiber-ring consisted of eight 1-mm diameter fibers (NT02-534, 0.98 mm core diameter, fiber radius r0 = 0.49 mm) with tilt angles ϕ = 23° (Fig. 2). Collection efficiencies were characterized with a single fiber or the fiber-ring attached to a 3-axes manipulator (Luigs&Neumann, Germany) for horizontal and vertical positioning with respect to the sample. For imaging transparent samples the fiber-ring holder could in principle be attached to the microscope objective. For collecting photons emerging from scattering tissue, however, it is advantageous to place the fiber-ring at a fixed position near the tissue surface, in particular for functional imaging in 3 dimensions [15

15. W. Göbel, B. M. Kampa, and F. Helmchen, “Imaging cellular network dynamics in three dimensions using fast 3D laser scanning,” Nat. Methods 4, 73–79 (2007). [CrossRef]

]. As the fiber ring cannot be lowered below the tissue surface we adapted our fiber-ring holder for subsurface in vivo imaging by slightly increasing the tilt angle to ϕ = 30° so that the fiber axes pointed towards a location about 250 μm below the surface (Fig. 5).

2.2 Animal preparation and in vivo imaging

All animal experiments were carried out according to the guidelines of the Center for Laboratory Animals of the University of Zurich and were approved by the Cantonal Veterinary Office. We anesthetized C57BL/6 mice (wildtype or thy1-YFP/line H, respectively; P35-40) with urethane (~1.5 g/kg body weight) and secured a stainless steel plate to the exposed skull with dental acrylic cement. We thinned the skull over an area of about 3 mm diameter and prepared a small cranial window (< 1 mm diameter) above the somatosensory area of neocortex. The dura within the cranial window was carefully removed and the exposed brain surface was superfused with normal rat Ringer solution (in mM: 135 NaCl, 5.4 KCl, 5 HEPES, 1.8 CaCl2, pH 7.2 with NaOH). We stained cell populations in the supragranular layers of neocortex with the calcium indicator Oregon Green-BAPTA-1 (OGB-1; Invitrogen) using the multi-cell bolus loading technique [16

16. C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, “In vivo two-photon calcium imaging of neuronal networks,” Proc. Natl. Acad. Sci. U.S.A. 100, 7319–7324 (2003). [CrossRef] [PubMed]

]. Briefly, the acetoxymethyl (AM) ester of OGB-1 was dissolved in DMSO plus 20% Pluronic F-127 and diluted in calcium-free Ringer solution to a final concentration of about 1 mM. The dye was pressure-injected into the superficial layers of neocortex using a micropipette. Alexa-594 (20 μM) was included in the pipette solution for visualization.

For subsurface in vivo imaging the fiber-ring holder was mounted with two screws on the head-plate against four springs, which allowed us to carefully lower the fiber tips onto the cranial window. Slot holes enabled us to center the fiber-ring above the target brain area within the craniotomy. We estimate that the lower edges of the fibers were placed 0-100 μm above the pial surface. A custom O-ring made of rubber between fiber-ring holder and head plate prevented leakage of water-immersion. The depth dependence of the signal gain was assessed from image stacks acquired at 2-μm z-steps. Intensity values from nine consecutive planes were averaged to obtain the gain plot in Fig. 6(d). Laser intensity was increased with depth. Movies of spontaneous calcium signals were acquired at 10.42 frames/sec (128×64 pixel; 1.5 ms per line). Fluorescence signals in individual cells were analyzed from defined regions-of-interest (ROIs) in both the objective and fiber-ring channel and, in addition, in a combined SUFICS channel that was obtained by summating the two background-corrected channels in ImageJ. To minimize contamination of cellular signals with neuropil calcium signals that reflect ongoing cortical activity [17

17. J. N. D. Kerr, D. Greenberg, and F. Helmchen, “Imaging input and output of neocortical networks in vivo” Proc. Natl. Acad. Sci. U.S.A. 102, 14063–14068 (2005). [CrossRef] [PubMed]

, 18

18. W. Göbel and F. Helmchen, “In vivo calcium imaging of neural network function,” Physiology (Bethesda) 22, 358–365 (2007). [CrossRef]

], we used small ROIs for analysis (less than half of the apparent cell diameter). Calcium signals in all channels were expressed as relative fluorescence changes ΔF/F. Background signal was estimated from an unstained region, typically a blood vessel lumen. The noise levels were quantified as the widths σ of Gaussian fits to the histogram of baseline ΔF/F fluorescence. To correct for slow drifts in the fluorescence traces histograms were calculated from high-pass filtered traces (>0.5 Hz cut-off frequency).

2.3 Monte Carlo simulation of epifluorescence collection

3. Theory and experimental results

3.1 Theory of fiber-optic light collection

The physical dimensions of the objective are typically larger than the size of the accepted light cone at the front lens as given by the numerical aperture (ensuring high-NA imaging over a certain field-of-view). In our configuration the point-like fluorescence source is excited through the same objective; therefore it will always be at focal distance and within the field-of-view so that the solid angle ΩNA,obj, defined by the half-opening angle αobj = arcsin(NAobj/n), is the only limiting factor (Fig. 2(a); n being the refractive index of the immersion medium). The light collection efficiency of an objective (normalized to 4π) is thus given by

ηobj=ΩNA,obj4π=12(1cosαobj)
(1)
Fig. 2. Theory of fiber-optic fluorescence collection. (a) Large-core optical fibers with a tilt angle ϕ positioned outside of the acceptance cone of a microscope objective. Maximum acceptance angles of objective and fibers are αobj and αf, respectively. (b) Fiber collection from an arbitrarily positioned source. The fiber solid angle Ωf depends on the core radius r0, the source distance R and the off-axis angle γ. See main text for further details. (c) The maximum acceptance angle αf determines the transmitted fraction of photons that hit the fiber core, indicated by the circle-circle intersection area (green). Top: fiber-limited case with a distal source; the fiber core size (red area) limits the maximum number of transmitted photons. Bottom: NA-limited case with a proximal source; the maximum solid angle determined by the NA (green area) limits light collection. (d) Theoretical dependence of single-fiber collection efficiency ηf on axial source position (top) and on lateral position (bottom) for three axial distances (0.5·z0 blue, 1.0·z0 black, 1.5·z0 red; vertical lines in d). Efficiencies are normalized to ηmax, z′ to z0, and r′ to r0. (e) Experimental data points for the axial (top) and lateral dependence (bottom) as obtained from fluorescent bead measurements (n = 5 beads; S.D. error bars). Data points are normalized to the plateau level for z′ < z0.

Deriving a general equation for the collection efficiency of an optical fiber is more complex as the fluorescence source may be located anywhere relative to the fiber tip. We need to take into account both the solid angle covered by the fiber entrance (Ωf) and the fraction of photons that are accepted (Af). Both factors depend on the fiber core radius r0 and the position of the source relative to the fiber tip, i.e., its distance along the fiber axis z’ and its off-axis position r’ (Fig. 2(b); primes refer to the fiber coordinate system).

For a single fiber the solid angle Ωf is given by S/R2 where S is the area of a cap (in general of ellipsoid shape) cut out by the fiber entrance from a sphere of radius R (R being the distance of the source from the center of the fiber end; Fig. 2(b)). If the source is located on the fiber axis the solid angle is given by

Ωf(r=0,z')=2π(1cosβ)
(2)

with β = arctan(r0/z’). For off-fiber-axis positions only the projection of the fiber entrance area is relevant, which scales with the cosine of the off-axis angle γ. Formally, this can be approximated by substituting r0 with a base radius rγ = r0·(cos γ)1/2 (assuming a circular instead of elliptic projection with identical areas). In addition, the distance R scales with cos γ, thus yielding

Ωf=(r,z)=2π(1cos(arctan(rγR))=2π(1cos(arctan(r0z(cosγ)3/2)))
(3)

For small angles γ this equation simplifies to

Ωf(r,z)=2π(1cosβ)(cosγ)2
(4)

The solid angle that can contribute to light collection is, however, limited by the maximal collection efficiency of the fiber η max, which can be expressed as a maximum solid angle ΩNA,f, defined by the maximum acceptance angle of the fiber αf = arcsin(NAf/n) (NAf is the numerical aperture of the fiber):

ηmax=ΩNA,f4π=12(1cosαf)
(5)

It is useful to define z0 = r0/tanαf as characteristic length scale along the fiber axis. Depending on the source’s distance from the fiber two cases need to be distinguished (Fig. 2(c)): if the source is relatively close to the fiber (z′ < z0) then the maximal number of transmittable photons is limited by ΩNA,f (NA-limited case); if the source is far away (z′ > z0) the photon number is limited by Ωf, i.e., by the physical size of the fiber (fiber-limited case). In general, the maximum possible transmission is determined by the smaller one of the two solid angles ΩNA,f and Ωf.

The second factor is that photons that reach the fiber core will only be transmitted if their entrance angle is smaller than αf. This acceptance criterion is illustrated by the overlap of the circular fiber core with a circular area in the same plane determined by a cone with opening half-angle αf that is oriented parallel to the fiber axis (Fig. 2(b), (c)). Only photons reaching the fiber entry within the intersection area of these two circles fulfill the angle criterion (indicated by the dark green areas in Fig. 2(c)). Effectively, this problem boils down to solving a circle-circle intersection problem to reveal the percentage fraction of photons transmitted Af. In the Appendix we provide a general expression of Af as a function of normalized fiber coordinates u = r′/r0 and v = z′/z0 (Fig. 2(c)). In summary, the collection efficiency of a single fiber is given by

ηf(u,v)=min(Ωf(u,v),ΩNA,f)Af(u,v)4π
(6)

3.2 Measurement of fiber collection efficiency

To verify equation 6 experimentally we measured fluorescence collection through a single large-core fiber using two-photon excitation of fluorescent beads. We embedded 1-μm fluorescent beads in a block of agar that was placed on a microscope slide. We then excited fluorescent beads located at the edge of the agar block with the two-photon microscope (10x/0.3 air objective) and collected fluorescence light through the fiber in air. We systematically varied axial and off-axis position of the bead relative to the fiber end. For comparison with the analytical theory we normalized detected photocurrents to the maximum current, which was reached for short distances (z’ < z0) on the fiber axis. In addition, we normalized the radial and axial distances to r0 and z0, respectively. Collection efficiency η f decreased for large axial distances as well as for lateral displacements (Fig. 2(d), (e)). Experimental data agreed well with the theoretical predictions except for a “soft edge” around z’ = z0, the characteristic length of the fiber. These results highlight the importance of the exact position and orientation of the fiber relative to the fluorescence source in order to achieve maximum collection efficiency. In particular, it is crucial to position the fiber end at a distance of about the characteristic length or closer to the source.

3.3 A ring of optical fibers for supplementary epifluorescence collection

Because an individual fiber, even with relatively high NAf, has a rather limited collection efficiency (e.g., η max = 3.8% for NAf = 0.51), we combined multiple fibers in a ring-like arrangement. From equations 1 and 6 we can calculate the theoretically expected signal gain G of a fiber-ring containing N fibers:

G=ηobj+Nηf(u,v)ηobj=1+Nmin(Ωf(u,v)ΩNA,f)Af(u,v)ΩNA,obj
(7)

If the maximal collection power of the fibers is exploited, the maximum theoretical gain of a fiber ring is given by

Gmax=ηobj+N·ηmaxηobj=1+N1cosαf1cosαobj=1+N11(NAf/n)211(NAobj/n)2
(8)

For our fiber-ring of 8 fibers with NAobj = 0.8 and NAf = 0.51 in water immersion Gmax = 4.04. N = 8 was chosen because the polar half-angle available for each of the touching fiber tips (360°/N=22.50°) matches the NA-governed αf =22.54°. In addition, αobj is well complemented by this arrangement (αobj+2·22.5°=82°) so that the objective’s light cone is not obstructed by the fiber-ring (Fig. 3(c)). The distance of the fiber tips from the intersection of all fiber axes (1.31 mm) was, however, slightly longer than z0 for this fiber type (1.28 mm), thus reducing Gmax to 3.6 (Fig. 3(b)). In general, there is no ideal number of fibers and the choices of N, NAf, and r0 critically depend on the objective used and the additional geometrical constraints. For example, for a 20x/0.95W objective we calculated Gmax = 2.46 for an appropriate design consisting of 10 fibers with NAf = 0.39 and r0 = 0.4 mm.

To derive an explicit expression of G as a function of the fiber-ring offset Δz (distance of the fiber-axes intersection from the objective’s focal plane) we expressed r’ and z’ as a function of Δz (r’ = Δz cosϕ and z′ = Δz sinϕ + ϕ and inserted this coordinate transformation into equation 7. The resulting function G(Δz) describes the theoretically expected signal gain for our fiber-ring and is plotted in Fig. 3(b). We measured the signal gain as a function of Δz by two-photon imaging of a fluorescent pollen grain through a 40x water-immersion objective (NAobj = 0.8). The fiber-ring was attached to a 3-axes manipulator for vertical and horizontal positioning independent of the objective. Pollen grains were placed on the tip of a glass pipette, which was glued vertically onto a microscope slide. This arrangement provided sufficient space to vary the height of the fiber-ring both below and above the microscope’s focal plane. For analysis, fluorescence intensity was averaged over the pollen grain after background correction. The gain factor was determined by dividing the sum of the fluorescence intensities in the two channels by the fluorescence intensity in the objective channel at Δz = 0. Consistent with our theory, the fiber-ring channel displayed maximum intensity when the fiber axes were aligned with the objective’s focus (Δz = 0) and decreased for both downward and upward displacements (Fig. 3(a), (b)). The shape of the experimental curve was similar to the theoretical one but the maximum gain was lower than predicted (2.2 compared to Gmax = 3.6). This difference is presumably due to coupling and transmission losses not considered here. Nonetheless, it is remarkable that around Δz = 0 fluorescence intensity was brighter in the fiber-ring channel than in the objective channel.

Fig. 3. Large signal gain by epifluorescence collection through a ring of optical fibers. (a) Signal gain obtained with the fiber-ring. Pollen grain images in objective and fiber-ring channels at various axial offsets Δz. (b) Comparison of theoretical and experimental gains as a function of Δz (solid red curve: analytical prediction; solid red symbols: Monte Carlo simulation; black curve: experiment). (c) Objective channel intensity as a function of axial offset Δz of the fiber-ring. The signal only decreases when the elevated fiber-ring partially obstructs the objective’s light cone as schematically indicated.

The fiber-ring could potentially obstruct the illumination and detection pathway through the objective. However, with the fiber-ring centered on the microscope axis, no obstruction occurred except for very high elevations of the fiber-ring above the focal plane (Fig. 3(c); see also Fig. 3(a), Δz = +1.0 mm). We furthermore characterized the spatial distribution of the SUFICS gain-factor by systematically moving the fiber-ring laterally in x- and y-directions (Fig. 4). The signal gain was defined relative to the non-obstructed objective collection efficiency with the fiber-ring centered. A false reference to a locally reduced objective signal would otherwise lead to an artificially high gain factor. We found an almost homogenous gain within a central area of about 0.8 mm diameter (Fig. 4(b)) with reduced values toward the edges of the field-of-view. We conclude that SUFICS employing a simple ring of 8 fibers can substantially (more than 2-fold) enhance the fluorescence signal in 2PLSM over a broad imaging field.

Fig. 4. Spatial dependence of individual channel intensities and SUFICS signal gain. (a) Horizontal map of the fluorescence intensity for objective (left) and fiber-ring channel (right) as a function of lateral x- and y-offset of the fiber-ring (fiber-axes intersection aligned with objective’s focal plane). Fluorescence intensity decreases when the objective’s light cone is partially obstructed by fibers as indicated. (b) Left: Horizontal map of the gain factor determined relative to the central fluorescence intensity of the non-obstructed objective channel. Right: Surface map of the gain factor within the fiber-ring arrangement. A non-obstructed field with a diameter of ~0.8 mm was obtained.

3.4 Monte Carlo simulation of fiber-collection in scattering tissue

Epifluorescence collection from scattering tissue was substantially different from the transparent case (Fig. 5(b), (c)). For instance, epi-collection was enhanced near the tissue surface. The reason is that photons originally emitted in the downward direction can be back-scattered and eventually may reach a detector. Nevertheless, collection efficiencies decrease at large imaging depth because ‘memory’ about the initial angular distribution is progressively lost. Only few photons emerging from the diffusely fluorescing surface will enter the objective’s pupil and the fiber cores below their maximum acceptance angles. The surface enhancement of collection was particularly evident for smooth surfaces, reaching down to a depth of about the field-of-view radius of the objective (Fig. 5(b)). With rough surfaces, this additional randomization of angles led to a steeper decline of collection efficiency (data not shown). The resulting signal gain was around 3 near the tissue surface (Fig. 5(c)). The absolute gain remained relatively constant with increasing depth, while the exact shape of the depth-dependence varied considerably with simulation parameters, as shown for different anisotropy factors g in Fig. 5(c). These results of the simulations indicate that SUFICS can substantially enhance fluorescence signals not only for transparent samples but also under highly scattering conditions.

Fig. 5. Monte Carlo simulation of SUFICS in scattering tissue. (a) General setup used for Monte-Carlo simulations. Photons were launched from a fluorescence source at different imaging depths zS. After being scattered within the tissue (grainy area) photons either reached the objective (red trajectory) or one of the fiber cores (blue trajectory) or they missed any of them. (b) Collection efficiencies η from Monte-Carlo simulations for fiber-ring (blue) and objective (red) as a function of imaging depth zS for smooth tissue-surfaces. The fiber-ring position was fixed to the tissue surface in all simulations. Analytical curves for transparent samples are displayed for comparison. Collection efficiencies are normalized to 4π. (c) Gain factors as a function of imaging depth zS for different values of the anisotropy factor g (g = 0.95 black squares; g = 0.90 black triangles up; g = 0.85 black circles; g = 0.70 black triangles down). Values above the tissue surface (zS < 0) are academic (no fluorescence staining) and provided for consistency check only. The analytical curve for non-scattering tissue is provided for comparison (line).

3.5 Signal improvement under in vivo imaging conditions

To demonstrate the advantages of SUFICS for in vivo imaging we applied the fiber-ring to two-photon imaging of neurons and neuronal activity in mouse neocortex (Figs. 6, 7). The tips of the fibers were positioned on the neocortical surface using a special holder that allowed flexible positioning of the fiber tips within the cranial window (Fig. 6(a)).

We first determined the depth-dependence of the SUFICS signal gain by imaging YFP-expressing deep-layer pyramidal neurons in transgenic mice. In both, the objective and the fiber-ring channels, the apical dendrites were bright and could be followed down to the cell somata about 650-800 μm below the pial surface (Fig. 6(b), (c)). In all 3 transgenic mice tested, the fiber-ring channel was brighter than the objective channel at all depths resulting in a nearly constant, more than 2-fold signal gain over the entire depth range (Fig. 6(d)). Experimental gain factors were slightly lower than predicted values. The shape of the depth dependence matched best Monte Carlo simulations of a smooth surface with g = 0.85. Rough surfaces led to steeply increasing gain factors with depth, which were not consistent with experiments. We conclude that SUFICS signal enhancement persists at large imaging depths and therefore may help to resolve very deep structures, even though excitation effects eventually limit depth penetration [22

22. P. Theer and W. Denk, “On the fundamental imaging-depth limit in two-photon microscopy,” J. Opt. Soc. Am. A Opt. Image Sci. Vis. 23, 3139–3149 (2006). [CrossRef] [PubMed]

, 23

23. A. Leray, C. Odin, and Y. Le Grand, “Out-of-focus fluorescence collection in two-photon microscopy of scattering media,” Opt. Commun. 281, 6139–6144 (2008). [CrossRef]

].

Fig. 6. Signal gain and deep imaging with SUFICS in vivo. (a) Photographs of the fiber-ring holder (top) and the fiber tips (bottom) centered above an exposed surface area of mouse neocortex. (b) Two-photon image stack of YFP-expressing neocortical pyramidal cells. Sample images for both, objective and fiber-ring channel, are shown at identical grey scales for 3 focal depths as indicated. (c) Maximum-intensity side projections of the image stacks shown in b for both objective and fiber-ring channel. (d) Signal gain as a function of imaging depth (black circles: experimental data; red: theoretical prediction by Monte-Carlo simulation with smooth surface and g = 0.85).

Finally we examined whether SUFICS can improve SNR in typical in vivo calcium imaging experiments. The occurrence of neuronal action potentials can be indirectly inferred from the associated calcium signals [17

17. J. N. D. Kerr, D. Greenberg, and F. Helmchen, “Imaging input and output of neocortical networks in vivo” Proc. Natl. Acad. Sci. U.S.A. 102, 14063–14068 (2005). [CrossRef] [PubMed]

], however the amplitude of action potential-evoked calcium transients is small and any noise reduction is therefore highly beneficial. We loaded a cell population in layer 2/3 of mouse somatosensory cortex with the calcium indicator Oregon Green BAPTA-1 (OGB-1) [16

16. C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, “In vivo two-photon calcium imaging of neuronal networks,” Proc. Natl. Acad. Sci. U.S.A. 100, 7319–7324 (2003). [CrossRef] [PubMed]

]. Similar to the YFP-mice, stained cells were brighter in the fiber-ring channel compared to the objective channel in almost all z-planes (2.47 ± 0.16 near the surface and 1.90 ± 0.15 at 290 μm depth; n = 3 mice) (Fig. 7(a)). We acquired spontaneous fluorescence traces from OGB-1 loaded neurons in cortical layer 2/3 (Fig. 7(b)) and analyzed the high frequency noise distribution of the relative fluorescence changes ΔF/F (Fig. 7(c)). Comparable ΔF/F traces were observed in the two detection channels but high-frequency baseline noise was reduced by about 23% in the combined channel as judged from the width of the baseline histogram (Fig. 7(d); SD of baseline: 5.3 ± 2.3% objective, 5.5 ± 2.6% fiber-ring, 4.3 ± 1.9% combined; n = 60 cells from 3 mice). This value is comparable to the expected value of 29% (1-1/√2) noise reduction for a gain of 2 in a shot-noise limited system. Overall, these examples clearly demonstrate two key benefits of SUFICS for in vivo imaging.

Fig. 7. Noise reduction of neuronal calcium transients in vivo. (a) Images of OGB1-AM-stained layer 2/3 cells in mouse neocortex for the objective channel (top), the fiber-ring channel (middle), and the combined channel (bottom; all identical grey scales). (b) Examples of spontaneous calcium signals in the 4 cells marked in a. (c) Left: Close-up views of the calcium transient indicated by the box in b. Right: Histograms with Gaussian fits of high-frequency baseline noise (f > 0.5 Hz). (d) Mean widths of Gaussian fits to noise histograms (3 mice, n = 20 cells each) were reduced in the combined channel compared to the objective channel.

4. Discussion

In this study we introduced SUFICS as a simple, inexpensive, and generally applicable method to enhance signal collection in nonlinear microscopy. A more than 2-fold signal gain was achieved for both non-scattering and scattering samples. In particular, a signal gain of ≥2 was still present at about 800 μm depth in mouse neocortex with the fiber-ring positioned on the surface of the brain. Moreover, a noise reduction was associated with the signal gain that will be helpful for optical recordings of neural activity.

Although the possibility of external fluorescence detection (outside of the normal detection pathways through either objective or condensor) has been recognized early [6

6. W. Denk, D. W. Piston, and W. W. Webb, “Multi-photon molecular excitation in laser-scanning microscopy,” in Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum Press, New York, 2005).

], only few theoretical concepts or experimental designs have been proposed. One report presented a novel concept of hybrid objective lenses, in which large portions of the epifluorescence light are captured by a reflector built around an inner refractive or reflective lens system for excitation [11

11. D. Vucinic, T. M. Bartol, and T. J. Sejnowski, “Hybrid reflecting objectives for functional multiphoton microscopy in turbid media,” Opt. Lett. 31, 2447–2449 (2006). [CrossRef] [PubMed]

]. This theoretical concept promises large signal gains if most of the fluorescence emerging in the epi-hemisphere can be redirected to a detector. Its realization will, however, require custom design of such an objective. An alternative design is based on a parabolic reflector that redirects extra fluorescence photons to a photomultiplier in the transillumination pathway [12

12. C. A. Combs, A. V. Smirnov, J. D. Riley, A. H. Gandjbakhche, J. R. Knutson, and R. S. Balaban, “Optimization of multiphoton excitation microscopy by total emission detection using a parabolic light reflector,” J. Microsc. 228, 330–337 (2007). [CrossRef] [PubMed]

]. While this design has been built and shown to provide signal enhancement, its current form is solely applicable to extracted tissue, e.g., brain slices. Unlike SUFICS, it cannot be used for in vivo imaging.

SUFICS is an alternative fiber-optic approach that is easy to implement as an add-on to existing microscope setups, at least doubling the microscope’s fluorescence collection efficiency. Besides the fiber-holder that should be designed to allow easy positioning, only one extra PMT is necessary. In principle, it is also possible to project the output of the fiber-bundle onto the same PMT that receives the fluorescence collected by the objective, which might be beneficial in terms of further minimizing detector noise. SUFICS is furthermore compatible with in vivo electrophysiology because omitting a single fiber would provide sufficient space for the insertion of a pipette electrode. With our 8-fiber ring we achieved a 23-fold signal enhancement. Similar to a recent study [12

12. C. A. Combs, A. V. Smirnov, J. D. Riley, A. H. Gandjbakhche, J. R. Knutson, and R. S. Balaban, “Optimization of multiphoton excitation microscopy by total emission detection using a parabolic light reflector,” J. Microsc. 228, 330–337 (2007). [CrossRef] [PubMed]

], the gain factor was lower than the theoretically calculated value, most probably due to unconsidered coupling and transmission losses. Although we attempted to make conditions in the two detection pathways as equal as possible, we cannot fully exclude slight differences in transmission losses due to aperture effects and surface reflections. In practice, it is advisable to maximize total detection efficiency by optimizing both channels independently. The benefits of SUFICS or other fiber collection schemes [24

24. E. J. Seibel, R. S. Johnston, and C. D. Melville, “A full-color scanning fiber endoscope,” Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI. Proceedings of the SPIE. 6083, 9–16 (2006).

, 25

25. J. D. McMullen and W. Zipfel, “A Scheme for Increasing the Collection Efficiency of Multiphoton Microscopy,” Biophys. J. 96, 639a–639a (2009). [CrossRef]

] may still be further increased by optimization of the fiber-ring holders, by using special light guides, or by combinations of large-core fibers with different diameters covering even larger solid angles. Moreover, custom-shaped photodiode arrays arranged in a similar ring-like fashion may enable comparable gain factors.

Enhancing signal detection in 2PLSM is significant for multiple reasons: first, sensitivity for detecting dim objects is increased; second, lower excitation power can be used to obtain sufficiently bright signal intensities which reduces photodamage and phototoxicity [9

9. H. J. Koester, D. Baur, R. Uhl, and S. W. Hell, “Ca2+ fluorescence imaging with pico- and femtosecond two-photon excitation: signal and photodamage,” Biophys. J. 77, 2226–2236 (1999). [CrossRef] [PubMed]

, 26

26. A. Hopt and E. Neher, “Highly nonlinear photodamage in two-photon fluorescence microscopy,” Biophys. J. 80, 2029–2036 (2001). [CrossRef] [PubMed]

]; and third, SNRs are improved, e.g., for dynamic measurements. Here, we found that SUFICS provides signal enhancement even at large depths, which may help to enable imaging experiments in deeper cortical layers [4

4. F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nat. Methods 2, 932–940 (2005). [CrossRef] [PubMed]

]. SUFICS may also be combined with proposed schemes for optimization of fluorescence excitation, either in terms of fluorescence yield [27

27. G. Donnert, C. Eggeling, and S. W. Hell, “Major signal increase in fluorescence microscopy through dark-state relaxation,” Nat. Methods 4, 81–86 (2007). [CrossRef]

] or in terms of reducing photobleaching and photodamage [28

28. N. Ji, J. C. Magee, and E. Betzig, “High-speed, low-photodamage nonlinear imaging using passive pulse splitters,” Nat. Methods 5, 197–202 (2008). [CrossRef] [PubMed]

]. SUFICS may help to either compensate for signal losses associated with special excitation modes or to further reduce the excitation intensity necessary to achieve a certain signal level.

Finally, SUFICS is applicable to and may prove highly beneficial for other nonlinear imaging modalities such as second- or third-harmonic generation microscopy [2

2. J. Mertz, “Nonlinear microscopy: new techniques and applications,” Curr. Opin. Neurobiol. 14, 610–616 (2004). [CrossRef] [PubMed]

, 29

29. D. Debarre, N. Olivier, and E. Beaurepaire, “Signal epidetection in third-harmonic generation microscopy of turbid media,” Opt. Express 15, 8913–8924 (2007). [CrossRef] [PubMed]

] and coherent anti-Stokes Raman scattering microscopy [30

30. J. X. Cheng, Y. K. Jia, G. Zheng, and X. S. Xie, “Laser-scanning coherent anti-Stokes Raman scattering microscopy and applications to cell biology,” Biophys. J. 83, 502–509 (2002). [CrossRef] [PubMed]

]. Enhanced fluorescence collection by SUFICS will also help to gain signal in miniaturized nonlinear microscopes [13

13. B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, “Fiber-optic fluorescence imaging,” Nat. Methods 2, 941–950 (2005). [CrossRef] [PubMed]

, 14

14. C. J. Engelbrecht, R. S. Johnston, E. J. Seibel, and F. Helmchen, “Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo,” Opt. Express 16, 5556–5564 (2008). [CrossRef] [PubMed]

]. Overall, we envision numerous applications of SUFICS in many research fields.

5. Appendix

The problem of calculating the fraction of the photons that hit the fiber core entrance and that are also accepted and transmitted, reduces to a circle-circle intersection problem (Fig. 2(c)). For two circles with radii R1 and R2 that are laterally offset by a distance d the intersection area A is given by [31

31. E. W. Weisstein, “Circle-Circle Intersection,” http://mathworld.wolfram.com/Circle-CircleIntersection.html.

]

A(R1,R2,d)=Re{R12arccos(d2+R12R222dR1)+R22arccos(d2+R22R122dR2)12(R12(R2d)2)((R2+d)2R12)}
(9)

Here, we apply this equation to obtain a general expression for the percentage fraction Af of transmitted photons as a function the position of the fluorescence source relative to the fiber tip. The two relevant circular areas are (1) the fiber core with radius r0 and (2) a circle in the same plane as the fiber entrance defined by the light cone emerging from the source that contains photon trajectories with angles below the maximum acceptance angle αf of the fiber (Fig. 2(c)). The center point of the second circle is defined by the off-axis shift r’ of the light source and its radius given by rα = z’ tan(αf). As the problem is fully defined by the two fiber parameters r0 and αf, we use normalized fiber coordinates u = r′/r0 and v = z′/z0, accounting for the scalability of the problem with regard to the fiber core size and numerical aperture. We defined Af as percentage transmission with respect to the maximally possible transmission. Thus, we normalized the intersection area to the area of the smaller one of the two circles (the fiber core area in the fiber-limited case and the acceptance area in the NA-limited case). Formally, this leads to the following set of equations:

Af(u,v)=1πmin(1,u)2Re{arccos(ν2u2+12ν)+u2arccos(ν2+u212uv)12(1(uν)2)((u+ν)21)}
forν>max(1,u)min(1,u)(partialoverlapofcircles)
Af(u,ν)=1νmax(1,u)min(1,u)(fulloverlapofcircles)
Af(u,v)=0forν>u+1(nooverlapofcircles)
(10)

Of special interest are the limiting cases v = 0 and u = 1. For v = 0 the axial dependence is given by Af(0,u) = 1. The radial dependence at z’ = z0 simplifies to

Af(ν,1)=1π{2arccos(ν2)ν24ν2}ν<=2
Af(ν,1)=0ν>2
(11)

To calculate fiber collection efficiencies this expression for Af(u,v) was then inserted into equations 6 and 7.

Acknowledgments

We thank S. Giger and H.J. Kasper for expert technical assistance and B. Grewe for fruitful discussions. This work was supported by Swiss National Science Foundation Grant 3100A0-114624, the National Center of Competence in Research “Neural Plasticity and Repair”, and a grant from the Swiss SystemsX.ch initiative, evaluated by the Swiss National Science Foundation. F.H. had the original idea for SUFICS. C.J.E. and F.H. developed the analytical theory and Monte-Carlo simulations. C.J.E. and W.G. performed the fiber collection efficiency measurements and designed and optimized the fiber rings. W.G. and F.H. performed the in vivo experiments. F.H. and C.J.E. wrote the manuscript.

References and links

1.

W. R. Zipfel, R. M. Williams, and W. W. Webb, “Nonlinear magic: multiphoton microscopy in the biosciences,” Nat. Biotechnol. 21, 1369–1377 (2003). [CrossRef] [PubMed]

2.

J. Mertz, “Nonlinear microscopy: new techniques and applications,” Curr. Opin. Neurobiol. 14, 610–616 (2004). [CrossRef] [PubMed]

3.

W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248, 73–76 (1990). [CrossRef] [PubMed]

4.

F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nat. Methods 2, 932–940 (2005). [CrossRef] [PubMed]

5.

K. Svoboda and R. Yasuda, “Principles of two-photon excitation microscopy and its applications to neuroscience,” Neuron 50, 823–839 (2006). [CrossRef] [PubMed]

6.

W. Denk, D. W. Piston, and W. W. Webb, “Multi-photon molecular excitation in laser-scanning microscopy,” in Handbook of Biological Confocal Microscopy, J. B. Pawley, ed. (Plenum Press, New York, 2005).

7.

M. Oheim, E. Beaurepaire, E. Chaigneau, J. Mertz, and S. Charpak, “Two-photon microscopy in brain tissue: parameters influencing the imaging depth (vol 111, pg 29, 2001),” J. Neurosci. Meth. 112, 205–205 (2001). [CrossRef]

8.

E. Beaurepaire and J. Mertz, “Epifluorescence collection in two-photon microscopy,” Appl. Opt. 41, 5376–5382 (2002). [CrossRef] [PubMed]

9.

H. J. Koester, D. Baur, R. Uhl, and S. W. Hell, “Ca2+ fluorescence imaging with pico- and femtosecond two-photon excitation: signal and photodamage,” Biophys. J. 77, 2226–2236 (1999). [CrossRef] [PubMed]

10.

Z. F. Mainen, M. Maletic-Savatic, S. H. Shi, Y. Hayashi, R. Malinow, and K. Svoboda, “Two-photon imaging in living brain slices,” Methods 18, 231–239 (1999). [CrossRef] [PubMed]

11.

D. Vucinic, T. M. Bartol, and T. J. Sejnowski, “Hybrid reflecting objectives for functional multiphoton microscopy in turbid media,” Opt. Lett. 31, 2447–2449 (2006). [CrossRef] [PubMed]

12.

C. A. Combs, A. V. Smirnov, J. D. Riley, A. H. Gandjbakhche, J. R. Knutson, and R. S. Balaban, “Optimization of multiphoton excitation microscopy by total emission detection using a parabolic light reflector,” J. Microsc. 228, 330–337 (2007). [CrossRef] [PubMed]

13.

B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, “Fiber-optic fluorescence imaging,” Nat. Methods 2, 941–950 (2005). [CrossRef] [PubMed]

14.

C. J. Engelbrecht, R. S. Johnston, E. J. Seibel, and F. Helmchen, “Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo,” Opt. Express 16, 5556–5564 (2008). [CrossRef] [PubMed]

15.

W. Göbel, B. M. Kampa, and F. Helmchen, “Imaging cellular network dynamics in three dimensions using fast 3D laser scanning,” Nat. Methods 4, 73–79 (2007). [CrossRef]

16.

C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, “In vivo two-photon calcium imaging of neuronal networks,” Proc. Natl. Acad. Sci. U.S.A. 100, 7319–7324 (2003). [CrossRef] [PubMed]

17.

J. N. D. Kerr, D. Greenberg, and F. Helmchen, “Imaging input and output of neocortical networks in vivo” Proc. Natl. Acad. Sci. U.S.A. 102, 14063–14068 (2005). [CrossRef] [PubMed]

18.

W. Göbel and F. Helmchen, “In vivo calcium imaging of neural network function,” Physiology (Bethesda) 22, 358–365 (2007). [CrossRef]

19.

L. Wang, S. L. Jacques, and L. Zheng, “MCML--Monte Carlo modeling of light transport in multi-layered tissues,” Comput. Methods Programs Biomed. 47, 131–146 (1995). [CrossRef] [PubMed]

20.

L. G. Henyey and J. L. Greenstein, “Diffuse radiation in the galaxy,” Astrophys. J. 93, 70–83 (1941). [CrossRef]

21.

S. L. Jacques and S. A. Prahl, “Steady-state Monte Carlo: A minimal program, "mc321.c”." (1998), http://omlc.ogi.edu/classroom/ece532/class4/ssmc/roulette.html.

22.

P. Theer and W. Denk, “On the fundamental imaging-depth limit in two-photon microscopy,” J. Opt. Soc. Am. A Opt. Image Sci. Vis. 23, 3139–3149 (2006). [CrossRef] [PubMed]

23.

A. Leray, C. Odin, and Y. Le Grand, “Out-of-focus fluorescence collection in two-photon microscopy of scattering media,” Opt. Commun. 281, 6139–6144 (2008). [CrossRef]

24.

E. J. Seibel, R. S. Johnston, and C. D. Melville, “A full-color scanning fiber endoscope,” Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI. Proceedings of the SPIE. 6083, 9–16 (2006).

25.

J. D. McMullen and W. Zipfel, “A Scheme for Increasing the Collection Efficiency of Multiphoton Microscopy,” Biophys. J. 96, 639a–639a (2009). [CrossRef]

26.

A. Hopt and E. Neher, “Highly nonlinear photodamage in two-photon fluorescence microscopy,” Biophys. J. 80, 2029–2036 (2001). [CrossRef] [PubMed]

27.

G. Donnert, C. Eggeling, and S. W. Hell, “Major signal increase in fluorescence microscopy through dark-state relaxation,” Nat. Methods 4, 81–86 (2007). [CrossRef]

28.

N. Ji, J. C. Magee, and E. Betzig, “High-speed, low-photodamage nonlinear imaging using passive pulse splitters,” Nat. Methods 5, 197–202 (2008). [CrossRef] [PubMed]

29.

D. Debarre, N. Olivier, and E. Beaurepaire, “Signal epidetection in third-harmonic generation microscopy of turbid media,” Opt. Express 15, 8913–8924 (2007). [CrossRef] [PubMed]

30.

J. X. Cheng, Y. K. Jia, G. Zheng, and X. S. Xie, “Laser-scanning coherent anti-Stokes Raman scattering microscopy and applications to cell biology,” Biophys. J. 83, 502–509 (2002). [CrossRef] [PubMed]

31.

E. W. Weisstein, “Circle-Circle Intersection,” http://mathworld.wolfram.com/Circle-CircleIntersection.html.

OCIS Codes
(110.4280) Imaging systems : Noise in imaging systems
(170.3660) Medical optics and biotechnology : Light propagation in tissues
(180.2520) Microscopy : Fluorescence microscopy
(180.5810) Microscopy : Scanning microscopy
(190.4180) Nonlinear optics : Multiphoton processes
(180.4315) Microscopy : Nonlinear microscopy

ToC Category:
Microscopy

History
Original Manuscript: February 17, 2009
Revised Manuscript: March 29, 2009
Manuscript Accepted: March 30, 2009
Published: April 2, 2009

Virtual Issues
Vol. 4, Iss. 6 Virtual Journal for Biomedical Optics

Citation
Christoph J. Engelbrecht, Werner Göbel, and Fritjof Helmchen, "Enhanced fluorescence signal in nonlinear microscopy through supplementary fiber-optic light collection," Opt. Express 17, 6421-6435 (2009)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-17-8-6421


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. W. R. Zipfel, R. M. Williams, and W. W. Webb, "Nonlinear magic: multiphoton microscopy in the biosciences," Nat. Biotechnol. 21, 1369-1377 (2003). [CrossRef] [PubMed]
  2. J. Mertz, "Nonlinear microscopy: new techniques and applications," Curr. Opin. Neurobiol. 14, 610-616 (2004). [CrossRef] [PubMed]
  3. W. Denk, J. H. Strickler, and W. W. Webb, "Two-photon laser scanning fluorescence microscopy," Science 248, 73-76 (1990). [CrossRef] [PubMed]
  4. F. Helmchen and W. Denk, "Deep tissue two-photon microscopy," Nat. Methods 2, 932-940 (2005). [CrossRef] [PubMed]
  5. K. Svoboda and R. Yasuda, "Principles of two-photon excitation microscopy and its applications to neuroscience," Neuron 50, 823-839 (2006). [CrossRef] [PubMed]
  6. W. Denk, D. W. Piston, and W. W. Webb, "Multi-photon molecular excitation in laser-scanning microscopy," in Handbook of Biological Confocal Microscopy, J. B. Pawley, ed., (Plenum Press, New York, 2005).
  7. M. Oheim, E. Beaurepaire, E. Chaigneau, J. Mertz, and S. Charpak, "Two-photon microscopy in brain tissue: parameters influencing the imaging depth (Vol. 111, pg 29, 2001)," J. Neurosci. Meth. 112, 205 (2001). [CrossRef]
  8. E. Beaurepaire and J. Mertz, "Epifluorescence collection in two-photon microscopy," Appl. Opt. 41, 5376-5382 (2002). [CrossRef] [PubMed]
  9. H. J. Koester, D. Baur, R. Uhl, and S. W. Hell, "Ca2+ fluorescence imaging with pico- and femtosecond two-photon excitation: signal and photodamage," Biophys. J. 77, 2226-2236 (1999). [CrossRef] [PubMed]
  10. Z. F. Mainen, M. Maletic-Savatic, S. H. Shi, Y. Hayashi, R. Malinow, and K. Svoboda, "Two-photon imaging in living brain slices," Methods 18, 231-239 (1999). [CrossRef] [PubMed]
  11. D. Vucinic, T. M. Bartol, and T. J. Sejnowski, "Hybrid reflecting objectives for functional multiphoton microscopy in turbid media," Opt. Lett. 31, 2447-2449 (2006). [CrossRef] [PubMed]
  12. C. A. Combs, A. V. Smirnov, J. D. Riley, A. H. Gandjbakhche, J. R. Knutson, and R. S. Balaban, "Optimization of multiphoton excitation microscopy by total emission detection using a parabolic light reflector," J. Microsc. 228, 330-337 (2007). [CrossRef] [PubMed]
  13. B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, "Fiber-optic fluorescence imaging," Nat. Methods 2, 941-950 (2005). [CrossRef] [PubMed]
  14. C. J. Engelbrecht, R. S. Johnston, E. J. Seibel, and F. Helmchen, "Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo," Opt. Express 16, 5556-5564 (2008). [CrossRef] [PubMed]
  15. W. Göbel, B. M. Kampa, and F. Helmchen, "Imaging cellular network dynamics in three dimensions using fast 3D laser scanning," Nat. Methods 4, 73-79 (2007). [CrossRef]
  16. C. Stosiek, O. Garaschuk, K. Holthoff, and A. Konnerth, "In vivo two-photon calcium imaging of neuronal networks," Proc. Natl. Acad. Sci. U.S.A. 100, 7319-7324 (2003). [CrossRef] [PubMed]
  17. J. N. D. Kerr, D. Greenberg, and F. Helmchen, "Imaging input and output of neocortical networks in vivo " Proc. Natl. Acad. Sci. U.S.A. 102, 14063-14068 (2005). [CrossRef] [PubMed]
  18. W. Göbel and F. Helmchen, "In vivo calcium imaging of neural network function," Physiology (Bethesda) 22, 358-365 (2007). [CrossRef]
  19. L. Wang, S. L. Jacques, and L. Zheng, "MCML--Monte Carlo modeling of light transport in multi-layered tissues," Comput. Methods Programs Biomed. 47, 131-146 (1995). [CrossRef] [PubMed]
  20. L. G. Henyey and J. L. Greenstein, "Diffuse radiation in the galaxy," Astrophys. J. 93, 70-83 (1941). [CrossRef]
  21. S. L. Jacques and S. A. Prahl, "Steady-state Monte Carlo: A minimal program, "mc321.c"." (1998), http://omlc.ogi.edu/classroom/ece532/class4/ssmc/roulette.html.
  22. P. Theer and W. Denk, "On the fundamental imaging-depth limit in two-photon microscopy," J. Opt. Soc. Am. A Opt. Image Sci. Vis. 23, 3139-3149 (2006). [CrossRef] [PubMed]
  23. A. Leray, C. Odin, and Y. Le Grand, "Out-of-focus fluorescence collection in two-photon microscopy of scattering media," Opt. Commun. 281, 6139-6144 (2008). [CrossRef]
  24. E. J. Seibel, R. S. Johnston, and C. D. Melville, "A full-color scanning fiber endoscope," Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI.Proc. SPIE. 6083, 9-16 (2006).
  25. J. D. McMullen and W. Zipfel, "A Scheme for Increasing the Collection Efficiency of Multiphoton Microscopy," Biophys. J. 96, 639a (2009). [CrossRef]
  26. A. Hopt and E. Neher, "Highly nonlinear photodamage in two-photon fluorescence microscopy," Biophys. J. 80, 2029-2036 (2001). [CrossRef] [PubMed]
  27. G. Donnert, C. Eggeling, and S. W. Hell, "Major signal increase in fluorescence microscopy through dark-state relaxation," Nat. Methods 4, 81-86 (2007). [CrossRef]
  28. N. Ji, J. C. Magee, and E. Betzig, "High-speed, low-photodamage nonlinear imaging using passive pulse splitters," Nat. Methods 5, 197-202 (2008). [CrossRef] [PubMed]
  29. D. Debarre, N. Olivier, and E. Beaurepaire, "Signal epidetection in third-harmonic generation microscopy of turbid media," Opt. Express 15, 8913-8924 (2007). [CrossRef] [PubMed]
  30. J. X. Cheng, Y. K. Jia, G. Zheng, and X. S. Xie, "Laser-scanning coherent anti-Stokes Raman scattering microscopy and applications to cell biology," Biophys. J. 83, 502-509 (2002). [CrossRef] [PubMed]
  31. E. W. Weisstein, "Circle-Circle Intersection," http://mathworld.wolfram.com/Circle-CircleIntersection.html.

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited