Theoretical aspects of Fourier Transform Spectrometry and common path triangular interferometers

Alessandro Barducci, Donatella Guzzi, Cinzia Lastri, Paolo Marcoionni, Vanni Nardino, and Ivan Pippi  »View Author Affiliations


Optics Express, Vol. 18, Issue 11, pp. 11622-11649 (2010)
http://dx.doi.org/10.1364/OE.18.011622


View Full Text Article

Acrobat PDF (1700 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Recent investigations have induced relevant advancements of imaging interferometry, which is becoming a viable option for Earth remote sensing. Various research programs have chosen the Sagnac configuration for new imaging interferometers. Due to the growing diffusion of this technique, we have developed a self-contained theory for describing the signal produced by triangular FTSs and its optimal processing. We investigate the relevant disadvantages of multiplexing, and compare dispersive with FTS instruments. The paper addresses some methods for correcting the phase error, and the non-unitary transformation performed by a Sagnac interferometer. The effect of noise on spectral estimations is discussed.

© 2010 OSA

1. Introduction

Recent investigations have induced relevant advancements of imaging interferometry, which is becoming a viable option for Earth remote sensing [1

1. M. J. Persky, “A review of spaceborne infrared Fourier transform spectrometers for remote sensing,” Rev. Sci. Instrum. 66(10), 4763–4797 (1995). [CrossRef]

,2

2. B. Harnisch, W. Posselt, K. Holota, H. O. Tittel, and M. Rost, “Compact Fourier-transform imaging spectrometer for small satellite missions,” Acta Astronaut. 52(9-12), 803–811 (2003). [CrossRef]

]. Various research programs have focused the opportunity to adopt the common path configuration for new imaging interferometers under analysis. More specifically, the Sagnac triangular configuration has been adopted for the development of some sensor prototypes, as reported by [3

3. M.-L. Junttila, “Stationary Fourier transform spectrometer,” Appl. Opt. 31(21), 4106–4112 (1992). [CrossRef] [PubMed]

6

6. Y. Ferrec, J. Taboury, H. Sauer, and P. Chavel, “Optimal geometry for Sagnac and Michelson interferometers used as spectral imagers,” Opt. Eng. 45(11), 115601-115606 (2006). [CrossRef]

].

The Fourier Transform Hyperspectral Imager (FTHSI) was the first spaceborne imaging interferometer that operated on board of the AirForce Research Laboratory (AFRL) – U. S. Department of Defence (DoD) technological satellite MightySat II.1 (Sindri P99-1) after its launch in 2000 [4

4. L. J. Otten, A. D. Meigs, B. A. Jones, P. Prinzing, and D. S. Fronterhouse, “Payload Qualification and Optical Performance Test Results for the MightySat II.1 Hyperspectral Imager,” Proc. SPIE 3498, 231–238 (1998). [CrossRef]

,7

7. L. J. Otten, R. G. Sellar, and J. B. Rafert, “MightySatII.1 Fourier transform hyperspectral imager payload performance,” Proc. SPIE 2583, 566–575 (1995). [CrossRef]

]. Precursors of the FTHSI were the HyperCam and the IrCam developed by the Kestrel Corporation (USA) for airborne applications, the Spatially Modulated Fourier Transform Spectrometer (SMIFTS) developed by Hawaii University [5

5. P. G. Lucey, T. Williams, K. Horton, C. Budney, J. B. Ratfer, and E. T. Risk, “SMIFTS: A cryogenically cooled spatially modulate, imaging, Fourier transform spectrometer for remote sensing applications,” Proceeding of the International Conference on Spectral Sensing Research, Vol. 1, 251 – 262, (1992).

], the High Ètendue Imaging Fourier Transform Spectrometer (HEIFTS) by Science Application International Corporation (USA) [8

8. P. G. Lucey, K. A. Horton, and T. Williams, “Performance of a long-wave infrared hyperspectral imager using a Sagnac interferometer and an uncooled microbolometer array,” Appl. Opt. 47(28), F107–F113 (2008). [CrossRef] [PubMed]

], and the imaging interferometer developed by Applied Spectral Imaging (Israel) for laboratory applications [6

6. Y. Ferrec, J. Taboury, H. Sauer, and P. Chavel, “Optimal geometry for Sagnac and Michelson interferometers used as spectral imagers,” Opt. Eng. 45(11), 115601-115606 (2006). [CrossRef]

]. An additional scientific initiative that will exploit the use of a hyperspectral imaging interferometer (the ALISEO sensor [9

9. A. Barducci, P. Marcoionni, I. Pippi, and M. Poggesi, “Simulation of the Performance of a Stationary Imaging Interferometer for High Resolution Monitoring of the Earth,” Proc. SPIE 4540, 112–121 (2001). [CrossRef]

] and [10

10. A. Barducci, A. Casini, F. Castagnoli, P. Marcoionni, M. Morandi, and I. Pippi, “Performance assessment of a Stationary Interferometer for High-Resolution Remote Sensing,” Proc. SPIE 4725, 547–555 (2002). [CrossRef]

]) is the MIOSAT mission of the Italian Space Agency (ASI). MIOSAT is a technological compact satellite (mass around 120 kg) that will host three payloads: a panchromatic camera, a MEMS Mach-Zehnder interferometer for atmospheric sounding, and the ALISEO sensor. Other remote sensing instruments adopting an interferometric approach have been reported in [11

11. M. Bliss, “Demonstration of a static Fourier transform spectrometer,” Proc. SPIE 3541, 103–109 (1999). [CrossRef]

14

14. S. Subramaniam, B. Y. Ravindra, B. Rabindranath, B. G. Basheerullah, P. V. Viswanath, and O. P. Bajpai, “Stationary spatially modulated fourier transform spectro-radiometer,” J. Indian Soc. Remote Sens. 31(3), 187–196 (2003). [CrossRef]

].

Some advantages are implicit in using imaging interferometers, due to their high signal (Jacquinot’s effect), and the option to adjust the sampled spectral range and resolution by changing the sensor sampling step and the instrument Field-Of-View (FOV) [2

2. B. Harnisch, W. Posselt, K. Holota, H. O. Tittel, and M. Rost, “Compact Fourier-transform imaging spectrometer for small satellite missions,” Acta Astronaut. 52(9-12), 803–811 (2003). [CrossRef]

6

6. Y. Ferrec, J. Taboury, H. Sauer, and P. Chavel, “Optimal geometry for Sagnac and Michelson interferometers used as spectral imagers,” Opt. Eng. 45(11), 115601-115606 (2006). [CrossRef]

,15

15. R. F. Horton, “Optical Design for High Ètendue Imaging Fourier Transform Spectrometer,” Proc. SPIE 2819, 300–315 (1996). [CrossRef]

,16

16. D. Cabib, R. A. Buckwald, Y. Garin, and D. G. Soenksen, “Spatially resolved Fourier transform spectroscopy (spectral imaging): a powerful tool for quantitative analytical microscopy”, in Optical diagnostics of living cells on biofluids,” Proc. SPIE 2678, 278–291 (1996). [CrossRef]

]. Critical points are instead connected with the high data-rate requested, the need to pre-filter the incoming radiation in order to avoid aliasing in the retrieved spectra, and the heavy data pre-processing for compensating instrument response and possible acquisition artefacts [17

17. J. Genest, P. Tremblay, and A. Villemaire, “Throughput of tilted interferometers,” Appl. Opt. 37(21), 4819–4822 (1998). [CrossRef]

20

20. R. G. Sellar and G. D. Boreman, “Comparison of relative signal-to-noise ratios of different classes of imaging spectrometer,” Appl. Opt. 44(9), 1614–1624 (2005). [CrossRef] [PubMed]

]. In this connection achieving new insights on the performance of imaging interferometers is an important topic that should originate relevant contributions to remote sensing applications and spectroscopic studies.

In this work we propose a theoretical analysis of Fourier Transform Spectrometry (FTS) with specific relations to the common-path Sagnac configuration. We gives a brief description of the optical configuration (Sagnac, without input slit) of the new ALISEO sensor, and discusses the theory describing its signal and the spectral estimations derived by it. The remainder of this paper is organized as follow. In Sect. 2 the main optical characteristics of the developed instrument are outlined. Sect. 3 discusses the theory of Sagnac common-path interferometers and describes a self-contained model of interferogram and spectral estimation. In this section we demonstrate that multiplexing usually produces a severe radiometric disadvantage in FTS (as compared with dispersive spectrometry), while the absence of exit slit (here termed as Fellgett’s advantage) is confirmed. We indicate some theoretical properties of the interferometric signal that can be useful for correcting the phase error originated by uncontrolled scanning offset. We also point out how this phase correction scheme interferes with experimental noise affecting the interferogram. We also find an analytical model describing the error introduced by the not unitary cosine-like transformation operated by the instrument, obtaining implicitly an algorithm for correcting its effects. Let us note that most of these findings are proved under very general assumptions, such that they represents as well the behaviour of almost each interferometer. Sect. 4 discusses the main consequences of our theoretical findings, while Sect. 5 summarizes the outcomes of our investigation, and sketches future work and open problems.

2. The Sagnac interferometer

Sagnac is an optical configuration for a class of interferometers that have a triangular ray-path, common for the two rays of the interferometer. Due to this reason this type of instruments are also called common-path interferometers, as opposed to tilted interferometers like the Michelson-Morley and Mach Zehnder instruments. Figure 1
Fig. 1 Layout of a stationary interferometer in the Sagnac (triangular common path) configuration. The light from the objective is first collimated by the input lens L1, and the semi-reflecting surface of the beam-splitter BS originates two rays (reflected and transmitted beam), which travel the instrument on the same triangular path limited by the two folding mirrors M1 and M2. On the exit port, light is focused onto a CCD plane by the lens L2.
shows the instrument’s optical layout, so describing the main characteristics of the Sagnac configuration. The light is first collimated by the lens L1, and travels the interferometer by means of a beam-splitter BS and two folding mirrors M1 and M2. Light emerging on the output port is then focused onto the output focal plane that holds an image detector by the lens L2. The instrument produces in its focal plane a stationary pattern of interference fringes of equal thickness that are localized at infinity. Let us note that interference fringes of a Sagnac instrument are sometimes termed as “equal inclination” fringes, as in Katzberg and Statham (1996) [12

12. S. J. Katzberg, and R. B. Statham, “Performance Assessment of the Digital Array Scanned Interferometers (DASI) Concept,” NASA Technical Paper 3570, August 1996.

]. However, equal inclination fringes are usually produced in Michelson-Morley interferometers, and are characterized by a non-uniform fringe distribution in the instrument focal plane, as discussed in [21

21. F. A. Jankins and H. E. White, Fundamentals of Optics, (Mcgraw-Hill College; 4th edition, 1976).

]. This feature is quite different from the behaviour of a narrow Field-Of-View (FOV) Sagnac instrument, in the focal plane of which fringe thickness is truly constant.

It can be shown that the BS provides part of phase-delay between the two interfering rays, the remaining part being originated outside the BS, and that the overall OPD linearly changes with varying the angle of the entering ray with respect to the instrumental optical axis [9

9. A. Barducci, P. Marcoionni, I. Pippi, and M. Poggesi, “Simulation of the Performance of a Stationary Imaging Interferometer for High Resolution Monitoring of the Earth,” Proc. SPIE 4540, 112–121 (2001). [CrossRef]

,10

10. A. Barducci, A. Casini, F. Castagnoli, P. Marcoionni, M. Morandi, and I. Pippi, “Performance assessment of a Stationary Interferometer for High-Resolution Remote Sensing,” Proc. SPIE 4725, 547–555 (2002). [CrossRef]

]. Due to the absence of entrance slit, the device acquires the image of an object superimposed to a fixed pattern of across-track interference fringes. Then, introducing a relative motion between the sensor and the object, each scene pixel exploits the entire interference pattern, hence its interferogram and spectrum can be assessed. We call this kind of optical layout “Leap frog” configuration. In remote sensing applications, each ground point is observed under several viewing angles while the sensor moves with respect to the target, so a 3-dim array of data (image stack) of varying phase offset is collected. This data-cube is first processed in order to extract the complete interferogram of every image pixel, then it is inverse cosine transformed to yield a wavenumber hyperspectral data-cube.

One of the main drawbacks of the Sagnac configuration is that the image at the output port is affected by relevant vignetting frequently. This phenomenon evidently influences the fringe visibility, and demands for complex data pre-processing necessary for restoring the correct radiometric level of the signal in the far wings of the interferogram.

3. Theory of a Sagnac imaging interferometer

In this Section an attempt is made to develop a comprehensive theory of an imaging interferometer operating in the common-path, triangular configuration. Nonetheless, the properties discussed in Sections from 3.1 until 3.6 have been deduced on such a quite general ground that they can be applied to almost any kind of Fourier Transform Spectrometer (FTS).

In a Fourier Transform Imaging Spectrometer (FTIS) the acquired physical information is the interferogram I(x) , that is the power of the interference pattern generated by the two rays at the position x in the focal plane of the instrument, as stated in the following relationship.
I(x)=A0+S(κ)tr(κ)E(κ)exp(jφ)+tt(κ)E(κ)exp(jφ+2πjκOPD(x,κ))2dκ.
(1)
In this equation j is the imaginary unit, φ the initial phase of the impinging radiation field, tr(κ) and tt(κ) are the overall amplitude attenuation factors affecting the electric field amplitude E(κ) of the reflected and transmitted rays when passing from the input to the output port of the interferometer. The coefficient A relates the square modulus of the field amplitude E(κ) to the corresponding radiance i(κ)=AE(κ)2 (specific intensity of the radiation field), and OPD is the optical path difference introduced by the interferometer in the considered propagation direction. In Eq. (1), the wavenumber κ corresponds to the wavelength λ=1/κ , and the integration over κ takes into account for a non-monochromatic light spectrum. Let us note that the OPD depends on the inclination ϑ of the entering ray, and that the corresponding focal plane position x is the projection of ϑ by the equivalent focal length of the focusing element of the instrument.

Actually, the complete raw interferogram of the energy coming from a certain pixel of the observed scene is convolved by the pixel dimension, and sampled with a (square) grid whose extension is limited by the detector size D. This is true for a system equipped with a (2 dim) array detector, like a CCD or a CMOS sensor. Therefore, the sampled interferogram IS(x) would be expressed as::
IS(x)=[I(OPD(x,κ))1drect(xd)]comb(xp)1Drect(xx0D)D=Mp,
(5)
where the symbol ∗ indicates the standard convolution product, p is the distance between two adjacent pixels, d is the pixel size, M is the number of pixels of the detector, comb(x/p)=pm=m=+δ(xmp) is the sampling function, and x0 is the centre of the sampling grid. Imposing a not null value for x00 gives rise to sampling asymmetry, a condition in which the number of samples collected on the two sides of the interferogram is not balanced. Since the interferogram is a function having even symmetry, the instrument may be designed in order to collect interferogram samples only on one of its sides, i.e. on the left of the interferogram center. We take x0 to be an integer multiple of the pixel pitch p, so assuming that a pixel exactly falls on the interferogram centre. In Sect. 3.2 we will discuss the important effects introduced by a supplementary sampling offset in the form of a fraction of p. The symbol comb(x/p) is coherent with the notation used by Goodman [22

22. W. Goodman, Introduction to Fourier Optics, (McGraw-Hill, New York, 1968).

], and is proportional to the shah function adopted by Bracewell [23

23. R. Bracewell, The Fourier transform and its applications, (McGraw-Hill, New York, 1965).

], and Walmsley et al [24

24. D. A. Walmsley, T. A. Clark, and R. E. Jennings, “Correction of off-center sampled interferograms by a change of origin in the fourier transform; the important effect of overlapping aliases,” Appl. Opt. 11(5), 1148–1151 (1972). [CrossRef] [PubMed]

]. The ratio (d/p)2 is the areal fill factor, a characteristic specific of the adopted detector. We define dOPD=ϖ(κ)d as the OPD(x,κ) variation along the pixel area, and δOPD=ϖ(κ)p as the smallest OPD variation between adjacent pixels. With this naming convention, the maximum OPD in the sampled interferogram ( OPDmax ) is given by:
OPDmax=ϖ(κ)(x0+D/2)=MbδOPDb=(x0+D/2)D.
(6)
The parameter b takes into account the interferogram sampling asymmetry. For a null x0 , b equals ½ and the interferogram sampling is symmetric having half samples on the left of the origin and half samples on its right. When b=1 only one side of the interferogram is sampled. Let us note that in Eqs. (5) and (6) we suppose that the interferogram central sample always is measured, regardless of the actual b value.

3.1 Spectral dependence of the optical path difference

3.2 Phase error due to sampling offset

Perfect interferogram sampling is an infrequent circumstance, and in real applications no sample is exactly collected on the interferogram center. We term this phenomenon as “sampling offset”, and introduce the new symbol Δx that indicates an unknown x translation of the sampling grid. Obviously, the sampling offset has to be less than the sampling step Δxp , otherwise its integral part has to be included in the asymmetry parameter x0 for a coherent analysis of sensor performance. In the past, a great effort has been devoted to mitigate or remove the phase error introduced by uncontrolled sampling offset [24

24. D. A. Walmsley, T. A. Clark, and R. E. Jennings, “Correction of off-center sampled interferograms by a change of origin in the fourier transform; the important effect of overlapping aliases,” Appl. Opt. 11(5), 1148–1151 (1972). [CrossRef] [PubMed]

,28

28. L. Mertz, “Auxiliary computation for Fourier spectrometry,” Infrared Phys. 7(1), 17–23 (1967). [CrossRef]

30

30. M. L. Forman, W. H. Steel, and G. A. Vanasse, “Correction of Asymmetric Interferograms Obtained in Fourier Spectroscopy,” J. Opt. Soc. Am. 56(1), 59–63 (1966). [CrossRef]

]. As a consequence of offset, no sample is exactly matching the interferogram center, introducing a phase error exp[2πjκϖ(κ)x+jπ] in the source spectrum estimation computed as inverse complex interferometer transform IFT1{ISS(x)} . The phase term disturbance can be removed taking the norm of the inverse complex transform F{IFT1{ISS(x)}} , an operation that restores the result in the right hand-side of Eq. (20). This type of correction was devised by Connes [31

31. J. Connes, “Recherches sur la spectroscopie par transformation de Fourier,” Revue d’Optique 40, 45–265 (1961).

], and it was investigated in depth by Walmsley et al [24

24. D. A. Walmsley, T. A. Clark, and R. E. Jennings, “Correction of off-center sampled interferograms by a change of origin in the fourier transform; the important effect of overlapping aliases,” Appl. Opt. 11(5), 1148–1151 (1972). [CrossRef] [PubMed]

]. In the past, the main shortcoming of this correction scheme was its requirement for a significant computation burden that prevented its application.

Considering that the bias affecting spectral estimations can partially removed with a simple calibration measurement, the corrected estimator F{IFT1{ISS(x)}} is always preferable because it prevents possible phase errors. It can be affirmed that the analyzed phase correction scheme provides reliable spectral estimations independent of possible sampling offset errors at the price of enhancing the weight of random noise and a possible residual estimation bias originated by photonic noise contribution (source dependent) that can’t be calibrated and removed.

3.3 Effects of the direct term

As shown in Eq. (3) the interferogram contains a constant contribution independent of x and proportional to the half power of the observed source. This contribution does not bring information about the source spectrum, and should be considered as a disturbance that may degrade the interferogram and the associated spectral estimations. The above reasoning is summarized in Fig.s (3) and (4). Figure (3)
Fig. 3 Schematic of signals composing the measured interferogram (black curve). The interferogram is made up of two components, a dc term (green line) that holds the source half-energy and the Fourier (cosine) Transform of the source spectrum (orange curve) that contains the desired spectral information. While the measured signal (interferogram) always is high and does not decreases with increasing the spectral resolution, the useful signal (orange curve) is tiny for high OPDs. The interferogram model here described arises from Eq. (3) and constitutes a basic property obeyed by any interferometers. This behavior is the main drawback of Fourier Transform Spectrometers.
indicates the informative part of the interferometric signal (the small amplitude ripple of the orange curve) and the useless part of the interferogram: the dc term plotted as a green straight line. In Fig. 4
Fig. 4 Schematic of signals composing the spectral estimate obtained stemming from a generic interferogram measurement. The interferogram is made up of two components, a constant dc term that inverse transforms in a pulse at zero wavenumber (green line), and the true source signal that inverse transforms in the desired spectrum (orange curve). The meaning of these two interferogram component in the conjugate OPD domain is discussed in the text and illustrated in Fig. 3. This picture undoubtedly shows that the dc term, which is at the core of the multuplexing advantage, does not bring source information apart its energy.
we show the inverse Fourier Transform (spectra) of the interferogram components already shown in Fig. 3.

Often, this direct term is removed before the inverse transformation that leads to the source spectrum estimation, although the presence in the sampled interferogram of a residual direct term does not affect the spectrum estimation significantly in many spectroscopic applications. In order to shows this property let us consider the presence of an additional constant term I0 in the interferogram. Due to this constant Eq. (23) changes as stated by the following equation:
E{F{IFT1{ISS(x)+z(x)}2}σz2}12=                 ={[i(κ)+I0δ(κ)]ϖ(κ)ϖ(κ)+ϖ(κ)κsinc(κdOPD)}sinc(2κOPDmax);
(26)
updating of Eq. (24) follows straightforwardly. The perturbation I0δ(κ) is negligible for wavenumbers κ greater than 0.5/OPDmax , where the sinc(2κOPDmax) function on the right hand-side of Eq. (26) has its first null. The condition for which the perturbation is negligible λ2OPDmax defines an interval (λmin;λmax) that evidently limits the operating spectral range of the interferometer, as stated by the following relationships:
λmax=2OPDmax=2δOPDMbλmin=2δOPD.
(27)
Let us note that frequently λmax is hundreds (even thousands) times λmin for many sampled imaging interferometers where M is between 500 and 5000. The limits of the operating spectral interval due to the constant term I0 are so wide to not affect at all the spectrum of source observed by a Sagnac interferometer and estimated by Eqs. (23) or (24). The sole exception is constituted by those cases in which the overall sensor response S(κ)H(κ)V(κ) is not properly compensated, apodization is applied before inverse transformation, or uncontrolled vignetting introduces a broader bell-shaped profile whose convolution with the pulse in Eq. (26) gives rise to a disturbance that might interfere with spectral estimations (significantly lowering the λmax limit).

3.4 Fellgett’s advantage of Fourier Transform Spectrometers

Evidently, the multiplexing effect has been misinterpreted in the past frequently, also assuming that a higher spectral resolving power can be achieved by a Fourier Transform Spectrometer avoiding the effect typical of dispersive spectrometers where a higher spectral resolution implies the measurement of a fainter signal. It is easy to demonstrate that the above apparently accurate reasoning is very often untrue for a generic FTS that obeys Eq. (3). Hence, it is interesting stating on more firm basis (quantitatively) possible radiometric advantages of FTS, stemming from a physical and mathematical investigation of the signal produced by FTS instruments.

In the following we assume that Fellgett’s advantage is associated to the absence of exit slit, while possible radiometric advantages connected with the typical FT multiplexing are termed multiplexing advantage. This definition of possible radiometric advantages of FTS help us to include in the following analysis new features of modern detectors that weren’t available sixty years ago (e.g. array detectors). Let us note that in many works on interferometry the phrases “Fellgett advantage” and “multiplex advantage” are used synonymously. However, our purpose is not getting a better representation the original thinking of Fellgett or other authors. We arbitrarily assume the above definitions for sake of simplicity, aiming at obtaining a clear discussion of their implications.

3.4.1 Noise and multiplexing advantage

Here as follows we will make use of continuous representations of signals involved in the interferogram and spectral estimations, avoiding the mathematical complexity of handling with sampled signals. Any results are easily extended to the case of a sampled signals. Let us write the single-sided interferogram signal model as:
I(x)={FT{i(κ)}|x=0+FT{i(κ)}2+z(x)       0xOPDmax0                                                         elsewhere.
(28)
In the above equation I(x) is as usual the interferogram, i(κ) the source spectrum, z(x) the additive zero-mean noise (e.g. AWN), and the Fourier Transform has been employed in lieu of the basic cosine transform. In this case we neglect the possible dependence of the OPD on the wavenumber κ, this characteristic being unnecessary for the following analysis. An unreal folk-lore piece frequently encountered when discussing the multiplexing advantage, is that white noise should combine incoherently in the inverse Fourier Transform which leads to spectral estimations. Apparently, some authors retain that spectral estimations should benefit from a reduced noise amplitude thanks to the FT1{I(x)} operator. As an example, Katzberg and Statham [12

12. S. J. Katzberg, and R. B. Statham, “Performance Assessment of the Digital Array Scanned Interferometers (DASI) Concept,” NASA Technical Paper 3570, August 1996.

] write: “Fourier Transform Spectrometers, on the other hand, observe a linear sum of elements of all portions of the input spectrum. When reconstructed, the presumably uncorrelated noise samples combine incoherently, while the signal adds coherently”. This claim is erroneous because it applies a property of the arithmetic addition of uncorrelated random samples to the FT of the same random sequence. As shown in Eq. (22), and in many textbooks on random fields and stochastic processes (e.g [33

33. A. Papoulis, Probability Random Variables, and Stochastic Processes, (McGraw-Hill International Editions, Third Edition, 1991).

].) the FT of a random sequence obeys a different law, and the signal-to-noise ratio (SNR) is an invariant quantity with respect to Fourier transformation. For the interferogram of Eq. (23), the average variance σz2 of an auto-uncorrelated noise z(x) can be estimated as:
σz2=1OPDmax+z(x)2dx,
(29)
while the average square interferogram is:
I(x)2=1OPDmax+I(x)2dx.
(30)
We note that Eq. (29) represents a common form of noise amplitude estimator, adopted in many works. Interestingly, this equation holds true in strict sense if the involved noise term is at least a variance-ergodic field (process). The variance ergodicity simply requires an auto-uncorrelated normal field. Evidently, any form of white noise or of auto-uncorrelated normal noise is variance ergodic. The SNR in the interferogram domain is given by:
SNRI(x)2=+I(x)2dx+z(x)2dx.
(31)
With a similar reasoning, we obtain in the conjugated domain (spectrum) the following expression:
SNRFT1{I(x)}2=+FT1{I(x)}2dκ+ζ(κ)2dκ,
(32)
where ζ(κ) is the inverse Fourier Transform of z(x) , hence ζ(κ) always is white noise. Relying on the Plancherel theorem it is easy proving that the average SNR is the same in both domains SNRFT1{I(x)}2=SNRI(x)2 , therefore the application of the inverse Fourier transform operator doesn’t add radiometric benefit to the result. This outcome is a general characteristic of the FT operator, and is based on the FT property of being a complete representation of the original function. In a different wording, the FT operator exactly preserves the information content held in the processed data, and the original separation in its random and source (ideal) contributions. Let us note that Eqs. (31) and (32) show the average SNR of the interferometer physical signal in both domains, but that they do not represent the effective SNR of interferometric measurements. As shown in Section 3.3 the signal I(x) (and its FT1{I(x)} ) has a large amplitude due to the presence of a non-informative component, which beautifies the total SNR.

3.4.2 The constant term in the interferogram

The measured interferogram takes its high amplitude from a useless contribution represented by the first term in the right hand-side of Eq. (28). This term is not useful because it does not bring information concerning the source, with exception of its panchromatic energy. The truly source-informative signal in the interferogram is a tiny undulation FT{i(κ)}/2 around the constant term FT{i(κ)}|x=0/2 . This behavior is evident when we consider the inverse FT of the constant term: it originates a pulse located κ=0 that doesn’t add information about the source spectrum, a circumstance that has been shown in Eq. (26).

It is worth noting that Plancherel theorem implies that the integrated power carried by the informative component of the interferogram equates a quarter of the integrated power carried by the source spectrum, ideally measured by a dispersive instrument.
+Iinformative(x)2dx=14+i(κ)2dκ.
(33)
In comparison with traditional dispersive techniques, FTS can maintain some étendue advantage as long as no input slit is adopted (Jaquinot’s advantage), or when the interferometer exhibits less optical losses of a grating spectrometer, but multiplexing doesn’t involve any radiometric advantage for the effective signal.

3.4.3 The informative part of the interferogram

In this Section we will show that multiplexing by trigonometric orthonormal functions originates a serious radiometric disadvantage connected to the circumstance that the power carried by the effective signal in the interferogram is concentrated at lower OPD s. In other words, spectral information concerning the source at high resolution (the finest spectrum details) is held in the subtle tails of the interferogram [39

39. F. D. Kahn, “The signal: noise ratio of a suggested spectral analyzer,” Astrophys. J. 129, 518–520 (1959). [CrossRef]

]. In order to elucidate this point with an example let us consider a radiation source i(κ) having a rectangular spectrum with bandwidth B as indicated in the following equation:

i(κ)=I0rect(κκ0B)IInformative(OPD(x))=BI02sinc(OPD(x))exp(2πjκ0OPD(x)).
(34)

3.5 Interferogram quantization and noise effects

In this Section we analyses the possible benefits of using interferometric and dispersive techniques. Since we have shown that the average SNR is the same in the interferogram domain and in its spectral estimation, performance of FTS and dispersive instruments can be computed in the domain in which the measurement is executed (interferogram for the FTS and spectral domain for dispersive devices).

The FT{i(κ)}|OPD=0/2 to δiFTS ratio is an important estimate of the radiometric accuracy required to the interferometer, in order to reach the spectral resolution δκ . From Eq. (38) the following relationship is easily obtained:
SNRminFT{i(κ)}|OPD=02δiFTS1AFTS(Kmax)l+1Kmax=Bδκ,
(40)
where we have introduced the minimal signal-to-noise ratio SNRmin necessary for measuring the tiny amplitude FT{i(κ)}|OPDOPDmax/2 on the top of the constant value FT{i(κ)}|OPD=0/2 , and the maximum number Kmax of in-band independent spectral channels that can be interpolated from such an interferometric measurement when observing a source having bandwidth B. Equation (40) makes even more evident the possible disadvantage of using an interferometer for high spectral resolution measurements when l>1 .

Evidently, Eqs. (40) and (41) implicitly assume that the standard deviation σz of the overall noise affecting the interferogram measurement is less than the small amplitude ripple FT{i(κ)}/2 to be observed. In a different wording, the maximum digitized OPD (the OPDmax ) has to be less than the value OPDn where the envelope of the informative signal FT{i(κ)}/2 equates the noise level σz :

OPDmax<OPDn.
(42)

It is well known that noise affecting radiometric measurements comes from three main sources: quantization noise (the round-off error), photonic noise, and instrumental noise. Noise caused by quantization and electronic detector and circuitry is not strictly connected to the optical configuration of the spectrometer under examination, and we can suppose that they affect in the same way the SNR of dispersive and FTS measurements. Therefore, we will focus our attention mainly on the photonic noise when investigating the advantages or disadvantages of FTSs. Photonic noise instead originates effects that are specific of the interferometer optical configuration and further degrade the performance of FTS. The point here examined regards the informative component of the interferometer signal, the only component that contains source spectral information. While the constant term doesn’t add contribution to the estimation of the source spectrum, the photonic noise originated by it strongly affects any spectral estimations. And this photonic noise contribution can’t be separated from the informative component of the signal. Moreover, since the non-informative term holds most of the power of the measured signal, this photonic noise contribution might be overwhelming at high spectral resolution (large OPD s) with respect to the tiny informative component of the signal.

Due to the above behavior, measurements performed in the visible spectral range where the photonic noise has a large impact can’t be executed at high spectral resolution with a FTS. Nevertheless, in the infrared spectral range the photonic noise is mitigated by the lower photon energy, and high spectral resolution interferometric measurements are a viable alternative to dispersive instruments.

3.6 Lost samples during data collection or transmission

With the phrase lost samples we indicate measurements that, due to a transmission or acquisition error, assume a trivial (not informative) value, e.g. null. When such a circumstance occurs in the measurement performed with a dispersive spectrometer the concerned spectral channel has definitively gone. In general this lacking channel can’t be in any way recovered by interpolation or other predictive modeling, unless a large error is tolerated. The important point is that the resulting error (information lack) is concentrated in a single spectral channel that becomes useless.

The corresponding situation with FTS devices is quite different and less disruptive. After the inverse transform procedure that leads to the spectral estimation, the effect of a missing interferogram sample (error) is spread as a small amplitude perturbation over all the interpolated spectral channels. Moreover, the corresponding spectral error is quite low in any spectral channels, and no channel is missing. This is an advantage typical of FTSs and all those spectrometric devices which impose mutually orthogonal modulations to the spectral radiance impinging on separate detector elements (multiplexing).

4. Discussion

The following list recaps the aspects relevant to the power efficiency of an generic FTS as compared to the traditional dispersive techniques.

  • 1. The multiplexing measurement approach doesn’t produce advantage in reducing the experimental noise, nor the final SNR is optimized by the inverse FT operation.
  • 2. The interferometric physical signal has a power high above that of a grating instrument observing the same source in the same experimental conditions. Most of this power is carried by a non-informative component of the interferometric signal (a constant term);
  • 3. Using the Plancherel theorem it can be shown that the power held in the informative part of the interferometer signal (the raw interferometer signal minus the constant factor) is on average the same as the power of the spectrally dispersed signal available for a dispersive spectrometer;
  • 4. Unfortunately, the cosine-like transformation operated by the FTS concentrates this equal power level (in the informative component of the signal) at lower OPD s, strongly reducing the power efficiency of the FTS with respect to dispersive spectrometers when high spectral resolution measurements are performed (large OPD s).
  • 5. The interferometer has to measure a much more tiny signal as the dispersive spectrometer does, but this signal is found to be superimposed on a high-amplitude continuous radiation plateau. From a practical point of view, this feature represents a shortcoming since most of the digitalization accuracy of the employed detector is used to sense this useless (non informative) high signal.
  • 6. This high-amplitude constant signal level originates a photonic noise of large standard deviation, so that the effective average SNR available in interferometric measurements is usually significantly lower than that obtained in dispersive observations with high spectral resolution. Decreasing is caused by both: the reduced level of effective signal amplitude at large OPD s (see lemma 1) and the noise increase due to the augmented photonic noise (large amplitude of the non-informative interferogram component).

The above outcomes demonstrate that multiplexing is a radiometric disadvantage real for Fourier Transform Spectrometers. This characteristic was not fully recognized in many works about interferometry due to a misinterpretation of the noise role and the effect of the constant term. Fellgett’s advantage is maintained in FTS as long as it is defined as the signal amplitude increase connected to the absence of exit slit. Today, this advantage no longer is a prerogative of FT spectrometers due to the availability of low-cost array detector. Similar conclusions has been drawn by Schumann and Lomheim [40

40. W. Schumann, and T. S. Lomheim, “Infrared hyperspectral imaging Fourier transform and dispersive spectrometers: comparison of signal-to-noise-based performance,” Imaging Spectrometry VII, San Diego, CA, USA, SPIE vol. 4480, (2001).

]. The following Table 1

Table 1. Main differences between dispersive and interferometric technique about throughput, multiplexing, radiometric accuracy and SNR issues.

table-icon
View This Table
resumes the main differences among the traditional dispersive spectrometers and the most recent triangular interferometers that have been examined in this paper.

Furthermore, we point out that:

  • Sources having a spectrum with almost rectangular profile can be observed by a FTS that would closely match the optimal radiometric- resolution / spectral resolution compromise achieved by dispersive instruments;
  • Sources observed in the Infrared spectral range may show a spectrum closely approximated by a negative exponential function of the wavenumber. It can be shown that the Fourier Transform of such kind of spectra approach a Lorentz function of the OPD . In such a case the radiometric accuracy required to interpolate the source spectrum with spectral resolution δκ follows a rather favorable law δiFTSAFTS2FT{i(κ)}|OPD=0(δκB)2    , that can be reliably measured with a FTS;
  • In the TIR spectral range the available signal contains radiative contributions from the optical elements along the ray path requiring cooling of the whole instrument. Usually, an FT spectrometer has lesser optical losses than a dispersive instrument, and sometime a minor number of optical surfaces along the raypath. Hence, the source-informative signal may be higher in FT spectrometers and be degraded at a lesser extent with respect to a dispersive instrument. This difference may mitigate the needing for cooling in spaceborne devices, originating a possible advantage of FTS in the TIR.
  • In the Visible spectral range, where due to the photonic noise the FTS performance is weakened, Fourier Transform Spectrometers can be advantageously employed for conducting low and medium spectral resolution observations of a broad-band source. In fact our theoretical modeling of FTSs performance of Eq. (37) provides us with a prediction of the required radiometric accuracy only as an asymptotic approximation valid for large OPD s. Hence, in spectral applications where the maximum observed OPD is not huge, the FTS technique can be a viable tool.
  • When observing a narrow-band source (e.g. a laser source with B<<1 ), it is possible to perform high spectral resolution measurements while maintaining a mild radiometric resolution of the FTS even for applications covering the visible spectral range.

Possible applications of multiplexing advantage potentially remain confined to instruments adopting a non trigonometric set of orthonormal functions for representing the observed spectrum. Our theoretical investigations only exclude that this advantage holds true for Fourier Transform Spectrometers.

5. Conclusions

In this paper a deep theoretical analysis has been performed about the principles of Fourier Spectrometry pertaining the Sagnac triangular configuration. A special effort has been devoted to study the cosine-like instrument transformation in order to take into account the circumstance that the Optical Path Difference function depends also on the wavenumber. This investigation led us to introduce the concept of group OPD , which expressly contains the spectral dispersion law of the material composing the beam-splitter (i.e., the essential element which originates by amplitude splitting the two interfering beams). It has been shown that the phase to group OPD ratio is the amplitude of the non-unitary transformation operated by the instrument. This theoretical result may be useful for correcting the spectral estimations obtained with a FTS or for other interferometric applications.

An analytical expression of the sampled interferogram has been found, which includes the effects of on-pixel integration, spatial sampling offset, truncation of the interferometric measurement, and additional instrumental effects such as fringe visibility. Attention has been devoted to the effects of a null-mean additive white noise affecting the interferometric measurement, and to signal degradation produced by the photonic noise. If the complex inverse (Fourier) transform estimator is adopted in lieu of the standard cosine-like operator (for an instance, in order to automatically remove the phase distortion in spectral estimations due to the sampling offset error), the retrieved spectrum is affected by a bias proportional to the noise standard deviation, and subject to an enhanced noise amplitude (40% relative increment). We have shown that this behavior is introduced by the fact that the basic cosine-like inverse transform selects only the even part of the noise record.

Finally, a new interpretation of the so called Fellgett’s advantage has been discussed based on an analytical comparison of system performance (estimation of amplitude of the effective signal and Signal-to-Noise ratio) between the interferometric technique and the traditional dispersive spectrometers. This discussion has led to the conclusion that Fellgett’s advantage has been misunderstood frequently in the past. We have proved that the informative tail of the interferogram to be resolved requires a radiometric resolution much finer (depending on the illumination source’s bandwidth) than that is needed for a dispersive spectrometer operating at the same high spectral resolution. This circumstance also reveals the advantage of dispersive techniques of an exponential factor of l+1/2 pertaining the maximum Signal-to-Noise ratio with respect to the interferometric technique. In the Visible spectral range, where due to the photonic noise the FTS performance is weakened, Fourier Transform Spectrometers can be advantageously employed for conducting low and medium spectral resolution observations of a broad-band source. On the other hand, when observing a Visible narrow-band or an Infrared broad-band source, it is possible to perform high spectral resolution measurements while maintaining a mild radiometric resolution of the interferometer.

Acknowledgements

This work was carried out with the support of the Italian Space Agency (ASI).

References and links

1.

M. J. Persky, “A review of spaceborne infrared Fourier transform spectrometers for remote sensing,” Rev. Sci. Instrum. 66(10), 4763–4797 (1995). [CrossRef]

2.

B. Harnisch, W. Posselt, K. Holota, H. O. Tittel, and M. Rost, “Compact Fourier-transform imaging spectrometer for small satellite missions,” Acta Astronaut. 52(9-12), 803–811 (2003). [CrossRef]

3.

M.-L. Junttila, “Stationary Fourier transform spectrometer,” Appl. Opt. 31(21), 4106–4112 (1992). [CrossRef] [PubMed]

4.

L. J. Otten, A. D. Meigs, B. A. Jones, P. Prinzing, and D. S. Fronterhouse, “Payload Qualification and Optical Performance Test Results for the MightySat II.1 Hyperspectral Imager,” Proc. SPIE 3498, 231–238 (1998). [CrossRef]

5.

P. G. Lucey, T. Williams, K. Horton, C. Budney, J. B. Ratfer, and E. T. Risk, “SMIFTS: A cryogenically cooled spatially modulate, imaging, Fourier transform spectrometer for remote sensing applications,” Proceeding of the International Conference on Spectral Sensing Research, Vol. 1, 251 – 262, (1992).

6.

Y. Ferrec, J. Taboury, H. Sauer, and P. Chavel, “Optimal geometry for Sagnac and Michelson interferometers used as spectral imagers,” Opt. Eng. 45(11), 115601-115606 (2006). [CrossRef]

7.

L. J. Otten, R. G. Sellar, and J. B. Rafert, “MightySatII.1 Fourier transform hyperspectral imager payload performance,” Proc. SPIE 2583, 566–575 (1995). [CrossRef]

8.

P. G. Lucey, K. A. Horton, and T. Williams, “Performance of a long-wave infrared hyperspectral imager using a Sagnac interferometer and an uncooled microbolometer array,” Appl. Opt. 47(28), F107–F113 (2008). [CrossRef] [PubMed]

9.

A. Barducci, P. Marcoionni, I. Pippi, and M. Poggesi, “Simulation of the Performance of a Stationary Imaging Interferometer for High Resolution Monitoring of the Earth,” Proc. SPIE 4540, 112–121 (2001). [CrossRef]

10.

A. Barducci, A. Casini, F. Castagnoli, P. Marcoionni, M. Morandi, and I. Pippi, “Performance assessment of a Stationary Interferometer for High-Resolution Remote Sensing,” Proc. SPIE 4725, 547–555 (2002). [CrossRef]

11.

M. Bliss, “Demonstration of a static Fourier transform spectrometer,” Proc. SPIE 3541, 103–109 (1999). [CrossRef]

12.

S. J. Katzberg, and R. B. Statham, “Performance Assessment of the Digital Array Scanned Interferometers (DASI) Concept,” NASA Technical Paper 3570, August 1996.

13.

P. D. Hammer, F. P. J. Valero, and D. L. Peterson, “An imaging interferometer for terrestrial remote sensing,” Proc. SPIE 1937, 244–255 (1993). [CrossRef]

14.

S. Subramaniam, B. Y. Ravindra, B. Rabindranath, B. G. Basheerullah, P. V. Viswanath, and O. P. Bajpai, “Stationary spatially modulated fourier transform spectro-radiometer,” J. Indian Soc. Remote Sens. 31(3), 187–196 (2003). [CrossRef]

15.

R. F. Horton, “Optical Design for High Ètendue Imaging Fourier Transform Spectrometer,” Proc. SPIE 2819, 300–315 (1996). [CrossRef]

16.

D. Cabib, R. A. Buckwald, Y. Garin, and D. G. Soenksen, “Spatially resolved Fourier transform spectroscopy (spectral imaging): a powerful tool for quantitative analytical microscopy”, in Optical diagnostics of living cells on biofluids,” Proc. SPIE 2678, 278–291 (1996). [CrossRef]

17.

J. Genest, P. Tremblay, and A. Villemaire, “Throughput of tilted interferometers,” Appl. Opt. 37(21), 4819–4822 (1998). [CrossRef]

18.

P. Jacquinot, “The luminosity of spectrometers with Prisms, Grating, or Fabry-Perot Etalons,” J. Opt. Soc. Am. 44(10), 761–765 (1954). [CrossRef]

19.

M. R. Descour, “The Throughput Advantage In Imaging Fourier-Transform Spectrometers,” Proc. SPIE 2819, 285–290 (1997). [CrossRef]

20.

R. G. Sellar and G. D. Boreman, “Comparison of relative signal-to-noise ratios of different classes of imaging spectrometer,” Appl. Opt. 44(9), 1614–1624 (2005). [CrossRef] [PubMed]

21.

F. A. Jankins and H. E. White, Fundamentals of Optics, (Mcgraw-Hill College; 4th edition, 1976).

22.

W. Goodman, Introduction to Fourier Optics, (McGraw-Hill, New York, 1968).

23.

R. Bracewell, The Fourier transform and its applications, (McGraw-Hill, New York, 1965).

24.

D. A. Walmsley, T. A. Clark, and R. E. Jennings, “Correction of off-center sampled interferograms by a change of origin in the fourier transform; the important effect of overlapping aliases,” Appl. Opt. 11(5), 1148–1151 (1972). [CrossRef] [PubMed]

25.

M.-L. Junttila, J. Kauppinen, and E. Ikonen, “Performance limits of stationary Fourier spectrometers,” J. Opt. Soc. Am. A 8(9), 1457–1462 (1991). [CrossRef]

26.

R. L. Hilliard and G. G. Shepherd, “Wide-Angle Michelson Interferometer for Measuring Doppler Line Widths,” J. Opt. Soc. Am. 56(3), 362–369 (1966). [CrossRef]

27.

P. Hlubina, D. Ciprian, J. Lunacek, and R. Chlebus, “Phase retrieval from the spectral interference signal used to measure thickness of SiO2 thin film on silicon wafer,” Appl. Phys. B 88(3), 397–403 (2007), doi:. [CrossRef]

28.

L. Mertz, “Auxiliary computation for Fourier spectrometry,” Infrared Phys. 7(1), 17–23 (1967). [CrossRef]

29.

H. Sakai, G. A. Vanasse, and M. L. Forman, “Spectral Recovery in Fourier Spectroscopy,” J. Opt. Soc. Am. 58(1), 84–90 (1968). [CrossRef]

30.

M. L. Forman, W. H. Steel, and G. A. Vanasse, “Correction of Asymmetric Interferograms Obtained in Fourier Spectroscopy,” J. Opt. Soc. Am. 56(1), 59–63 (1966). [CrossRef]

31.

J. Connes, “Recherches sur la spectroscopie par transformation de Fourier,” Revue d’Optique 40, 45–265 (1961).

32.

T. Okamoto, S. Kawata, and S. Minami, “Fourier transform spectrometer with a self-scanning photodiode array,” Appl. Opt. 23(2), 269–273 (1984). [CrossRef] [PubMed]

33.

A. Papoulis, Probability Random Variables, and Stochastic Processes, (McGraw-Hill International Editions, Third Edition, 1991).

34.

P. R. Griffiths, H. J. Sloane, and R. W. Hannah, “Interferometers vs monochromators: separating the optical and digital advantages,” Appl. Spectrosc. 31(6), 485–495 (1977). [CrossRef]

35.

R. J. Bell, Introductory Fourier Transform Spectroscopy, (Academic Press, New York and London, 1972).

36.

P. B. Fellgett, “Conclusions on multiplex methods,” Journal de Physique, Colloque C2, Supplément au n. 3–4 Tome 28, mars-avril 1967, pp.: 165–171, (1967).

37.

P. B. Fellgett, “I. — les principes généraux des méthodes nouvelles en spectroscopie interférentielle A propos de la théorie du spectromètre interférentiel multiplex,” J. Phys. Radium 19(3), 187–191 (1958). [CrossRef]

38.

P. B. Fellgett, “The nature and origin of multiplex Fourier spectrometry,” Notes Rec. R. Soc. 60(1), 91–93 (2006). [CrossRef]

39.

F. D. Kahn, “The signal: noise ratio of a suggested spectral analyzer,” Astrophys. J. 129, 518–520 (1959). [CrossRef]

40.

W. Schumann, and T. S. Lomheim, “Infrared hyperspectral imaging Fourier transform and dispersive spectrometers: comparison of signal-to-noise-based performance,” Imaging Spectrometry VII, San Diego, CA, USA, SPIE vol. 4480, (2001).

OCIS Codes
(120.0280) Instrumentation, measurement, and metrology : Remote sensing and sensors
(120.3180) Instrumentation, measurement, and metrology : Interferometry
(300.6190) Spectroscopy : Spectrometers
(300.6300) Spectroscopy : Spectroscopy, Fourier transforms
(280.4991) Remote sensing and sensors : Passive remote sensing

ToC Category:
Spectroscopy

History
Original Manuscript: February 24, 2010
Revised Manuscript: March 17, 2010
Manuscript Accepted: March 19, 2010
Published: May 18, 2010

Citation
Alessandro Barducci, Donatella Guzzi, Cinzia Lastri, Paolo Marcoionni, Vanni Nardino, and Ivan Pippi, "Theoretical aspects of Fourier Transform Spectrometry and common path triangular interferometers," Opt. Express 18, 11622-11649 (2010)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-18-11-11622


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. M. J. Persky, “A review of spaceborne infrared Fourier transform spectrometers for remote sensing,” Rev. Sci. Instrum. 66(10), 4763–4797 (1995). [CrossRef]
  2. B. Harnisch, W. Posselt, K. Holota, H. O. Tittel, and M. Rost, “Compact Fourier-transform imaging spectrometer for small satellite missions,” Acta Astronaut. 52(9-12), 803–811 (2003). [CrossRef]
  3. M.-L. Junttila, “Stationary Fourier transform spectrometer,” Appl. Opt. 31(21), 4106–4112 (1992). [CrossRef] [PubMed]
  4. L. J. Otten, A. D. Meigs, B. A. Jones, P. Prinzing, and D. S. Fronterhouse, “Payload Qualification and Optical Performance Test Results for the MightySat II.1 Hyperspectral Imager,” Proc. SPIE 3498, 231–238 (1998). [CrossRef]
  5. P. G. Lucey, T. Williams, K. Horton, C. Budney, J. B. Ratfer, and E. T. Risk, “SMIFTS: A cryogenically cooled spatially modulate, imaging, Fourier transform spectrometer for remote sensing applications,” Proceeding of the International Conference on Spectral Sensing Research, Vol. 1, 251 – 262, (1992).
  6. Y. Ferrec, J. Taboury, H. Sauer, and P. Chavel, “Optimal geometry for Sagnac and Michelson interferometers used as spectral imagers,” Opt. Eng. 45(11), 115601-115606 (2006). [CrossRef]
  7. L. J. Otten, R. G. Sellar, and J. B. Rafert, “MightySatII.1 Fourier transform hyperspectral imager payload performance,” Proc. SPIE 2583, 566–575 (1995). [CrossRef]
  8. P. G. Lucey, K. A. Horton, and T. Williams, “Performance of a long-wave infrared hyperspectral imager using a Sagnac interferometer and an uncooled microbolometer array,” Appl. Opt. 47(28), F107–F113 (2008). [CrossRef] [PubMed]
  9. A. Barducci, P. Marcoionni, I. Pippi, and M. Poggesi, “Simulation of the Performance of a Stationary Imaging Interferometer for High Resolution Monitoring of the Earth,” Proc. SPIE 4540, 112–121 (2001). [CrossRef]
  10. A. Barducci, A. Casini, F. Castagnoli, P. Marcoionni, M. Morandi, and I. Pippi, “Performance assessment of a Stationary Interferometer for High-Resolution Remote Sensing,” Proc. SPIE 4725, 547–555 (2002). [CrossRef]
  11. M. Bliss, “Demonstration of a static Fourier transform spectrometer,” Proc. SPIE 3541, 103–109 (1999). [CrossRef]
  12. S. J. Katzberg, and R. B. Statham, “Performance Assessment of the Digital Array Scanned Interferometers (DASI) Concept,” NASA Technical Paper 3570, August 1996.
  13. P. D. Hammer, F. P. J. Valero, and D. L. Peterson, “An imaging interferometer for terrestrial remote sensing,” Proc. SPIE 1937, 244–255 (1993). [CrossRef]
  14. S. Subramaniam, B. Y. Ravindra, B. Rabindranath, B. G. Basheerullah, P. V. Viswanath, and O. P. Bajpai, “Stationary spatially modulated fourier transform spectro-radiometer,” J. Indian Soc. Remote Sens. 31(3), 187–196 (2003). [CrossRef]
  15. R. F. Horton, “Optical Design for High Ètendue Imaging Fourier Transform Spectrometer,” Proc. SPIE 2819, 300–315 (1996). [CrossRef]
  16. D. Cabib, R. A. Buckwald, Y. Garin, and D. G. Soenksen, “Spatially resolved Fourier transform spectroscopy (spectral imaging): a powerful tool for quantitative analytical microscopy”, in Optical diagnostics of living cells on biofluids,” Proc. SPIE 2678, 278–291 (1996). [CrossRef]
  17. J. Genest, P. Tremblay, and A. Villemaire, “Throughput of tilted interferometers,” Appl. Opt. 37(21), 4819–4822 (1998). [CrossRef]
  18. P. Jacquinot, “The luminosity of spectrometers with Prisms, Grating, or Fabry-Perot Etalons,” J. Opt. Soc. Am. 44(10), 761–765 (1954). [CrossRef]
  19. M. R. Descour, “The Throughput Advantage In Imaging Fourier-Transform Spectrometers,” Proc. SPIE 2819, 285–290 (1997). [CrossRef]
  20. R. G. Sellar and G. D. Boreman, “Comparison of relative signal-to-noise ratios of different classes of imaging spectrometer,” Appl. Opt. 44(9), 1614–1624 (2005). [CrossRef] [PubMed]
  21. F. A. Jankins and H. E. White, Fundamentals of Optics, (Mcgraw-Hill College; 4th edition, 1976).
  22. W. Goodman, Introduction to Fourier Optics, (McGraw-Hill, New York, 1968).
  23. R. Bracewell, The Fourier transform and its applications, (McGraw-Hill, New York, 1965).
  24. D. A. Walmsley, T. A. Clark, and R. E. Jennings, “Correction of off-center sampled interferograms by a change of origin in the fourier transform; the important effect of overlapping aliases,” Appl. Opt. 11(5), 1148–1151 (1972). [CrossRef] [PubMed]
  25. M.-L. Junttila, J. Kauppinen, and E. Ikonen, “Performance limits of stationary Fourier spectrometers,” J. Opt. Soc. Am. A 8(9), 1457–1462 (1991). [CrossRef]
  26. R. L. Hilliard and G. G. Shepherd, “Wide-Angle Michelson Interferometer for Measuring Doppler Line Widths,” J. Opt. Soc. Am. 56(3), 362–369 (1966). [CrossRef]
  27. P. Hlubina, D. Ciprian, J. Lunacek, and R. Chlebus, “Phase retrieval from the spectral interference signal used to measure thickness of SiO2 thin film on silicon wafer,” Appl. Phys. B 88(3), 397–403 (2007), doi:. [CrossRef]
  28. L. Mertz, “Auxiliary computation for Fourier spectrometry,” Infrared Phys. 7(1), 17–23 (1967). [CrossRef]
  29. H. Sakai, G. A. Vanasse, and M. L. Forman, “Spectral Recovery in Fourier Spectroscopy,” J. Opt. Soc. Am. 58(1), 84–90 (1968). [CrossRef]
  30. M. L. Forman, W. H. Steel, and G. A. Vanasse, “Correction of Asymmetric Interferograms Obtained in Fourier Spectroscopy,” J. Opt. Soc. Am. 56(1), 59–63 (1966). [CrossRef]
  31. J. Connes, “Recherches sur la spectroscopie par transformation de Fourier,” Revue d’Optique 40, 45–265 (1961).
  32. T. Okamoto, S. Kawata, and S. Minami, “Fourier transform spectrometer with a self-scanning photodiode array,” Appl. Opt. 23(2), 269–273 (1984). [CrossRef] [PubMed]
  33. A. Papoulis, Probability Random Variables, and Stochastic Processes, (McGraw-Hill International Editions, Third Edition, 1991).
  34. P. R. Griffiths, H. J. Sloane, and R. W. Hannah, “Interferometers vs monochromators: separating the optical and digital advantages,” Appl. Spectrosc. 31(6), 485–495 (1977). [CrossRef]
  35. R. J. Bell, Introductory Fourier Transform Spectroscopy, (Academic Press, New York and London, 1972).
  36. P. B. Fellgett, “Conclusions on multiplex methods,” Journal de Physique, Colloque C2, Supplément au n. 3–4 Tome 28, mars-avril 1967, pp.: 165–171, (1967).
  37. P. B. Fellgett, “I. — les principes généraux des méthodes nouvelles en spectroscopie interférentielle A propos de la théorie du spectromètre interférentiel multiplex,” J. Phys. Radium 19(3), 187–191 (1958). [CrossRef]
  38. P. B. Fellgett, “The nature and origin of multiplex Fourier spectrometry,” Notes Rec. R. Soc. 60(1), 91–93 (2006). [CrossRef]
  39. F. D. Kahn, “The signal: noise ratio of a suggested spectral analyzer,” Astrophys. J. 129, 518–520 (1959). [CrossRef]
  40. W. Schumann and T. S. Lomheim, “Infrared hyperspectral imaging Fourier transform and dispersive spectrometers: comparison of signal-to-noise-based performance,” Imaging Spectrometry VII, San Diego, CA, USA, SPIE 4480,1-14 (2001).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.