Collecting back-reflected photons in photoacoustic microscopy

Hao F. Zhang, Jing Wang, Qing Wei, Tan Liu, Shuliang Jiao, and Carmen A. Puliafito  »View Author Affiliations


Optics Express, Vol. 18, Issue 2, pp. 1278-1282 (2010)
http://dx.doi.org/10.1364/OE.18.001278


View Full Text Article

Acrobat PDF (204 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Since the photoacoustic effect relies only on the absorbed optical energy, the back-reflected photons from samples in optical-resolution photoacoustic microscopy are usually discarded. By employing a 2 × 2 single-mode fiber optical coupler in a laser-scanning optical-resolution photoacoustic microscope for delivering the illuminating laser light and collecting the back reflected photons, a fiber-optic confocal microscope is integrated with the photoacoustic microscope. Thus, simultaneous multimodal imaging can be achieved with a single light source and images from the two modalities are intrinsically registered. Such capabilities are demonstrated in imaging both phantoms and small animals in vivo.

© 2010 OSA

1. Introduction

Photoacoustic microscopy (PAM) is one of the fastest-growing imaging technologies due to its unique capability to achieve high-spatial resolution imaging of optical absorption contrast in biological tissue. Two forms of PAM, ultrasonic-resolution PAM (UR-PAM) [1

1. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

3

3. E. Z. Zhang, J. G. Laufer, R. B. Pedley, and P. C. Beard, “In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy,” Phys. Med. Biol. 54(4), 1035–1046 (2009). [CrossRef] [PubMed]

] and optical-resolution PAM (OR-PAM) [4

4. K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

,5

5. Z. X. Xie, S. L. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]

], have been developed for different targeted applications.

UR-PAM depends on ultrasonic localization and usually has a spatial resolution from 15 µm to 100 µm, depending on the bandwidth and center frequency of the ultrasonic detector. Since ultrasonic scattering is two orders of magnitude weaker than optical scattering in biological tissue, UR-PAM exceeds the existing depth limit (~1 mm) of high-resolution optical imaging and can probe deep (up to 30 mm) into tissue [1

1. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

,3

3. E. Z. Zhang, J. G. Laufer, R. B. Pedley, and P. C. Beard, “In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy,” Phys. Med. Biol. 54(4), 1035–1046 (2009). [CrossRef] [PubMed]

]. Although the spatial resolution becomes worse, the depth-to-resolution ratio is usually kept around 100 [6

6. L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nat. Photonics 3(9), 503–509 (2009). [CrossRef]

]. Hence, UR-PAM aims at deep-tissue applications such as imaging of the sentinel lymph node for diagnosing breast cancer [7

7. K. H. Song, C. H. Kim, C. M. Cobley, Y. N. Xia, and L. V. Wang, “Near-infrared gold nanocages as a new class of tracers for photoacoustic sentinel lymph node mapping on a rat model,” Nano Lett. 9(1), 183–188 (2009). [CrossRef]

] and small animal brain imaging [8

8. J. Laufer, E. Zhang, G. Raivich, and P. Beard, “Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner,” Appl. Opt. 48(10), D299–D306 (2009). [CrossRef] [PubMed]

].

OR-PAM relies on optical focusing to provide lateral resolution (can be as high as a few micrometers) and the ultrasonic detection can be either strongly focused [4

4. K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

] or unfocused [5

5. Z. X. Xie, S. L. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]

]. OR-PAM has the potential to be integrated with existing optical microscopic imaging modalities, such as optical coherence tomography (OCT) [9

9. S. Jiao, R. Knighton, X. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]

], confocal microscopy, and two-photon microcopy, to provide multimodal imaging with complementary contrasts.

The development of laser-scanning optical-resolution PAM (LSOR-PAM) [5

5. Z. X. Xie, S. L. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]

] solved the incompatibility between OR-PAM and existing optical microscopy by eliminating the mechanical scanning of the optical and ultrasonic components. LSOR-PAM has been successfully fused with spectral-domain OCT [10

10. S. L. Jiao, Z. X. Xie, H. F. Zhang, and C. A. Puliafito, “Simultaneous multimodal imaging with integrated photoacoustic microscopy and optical coherence tomography,” Opt. Lett. 34(19), 2961–2963 (2009). [CrossRef] [PubMed]

] to achieve simultaneous imaging of the microvasculature and micro-anatomy in small animals in vivo, which laid the foundation for potential applications in ophthalmic imaging.

Here, we report on the integration of LSOR-PAM with a fiber-optic confocal microscope (FOCON) [11

11. T. Dabbs and M. Glass, “Fiber-optic confocal microscope: FOCON,” Appl. Opt. 31(16), 3030–3035 (1992). [CrossRef] [PubMed]

] to achieve multimodal imaging of both optical absorption and optical scattering contrasts. The axial resolution of the confocal microscopy was quantified both theoretically and experimentally. The complementary contrast was demonstrated in both phantoms and small animals in vivo.

2. Methods and materials

Figure 1
Fig. 1 Schematic of the fused laser-scanning confocal microscopy and photoacoustic microscopy. PD: photodiode; 2×2: fiber coupler.
shows a schematic of the experimental system. A tunable dye laser (Cobra, Sirah Laser and Plasmatchnik GmbH) pumped by a Nd:YLF laser system (IS8II-E, EdgeWave GmbH, pulse duration 6 ns, repetition rate: 1 kHz) was used as the irradiation source. The output wavelength was 578 nm. The laser light from the dye laser was spatially filtered by an iris and attenuated by a neutral density filter before coupled into a 2×2 single-mode optical fiber coupler (FC-632, Thorlabs). The output light from arm O1 was collimated and expanded to 10 mm in diameter and then scanned by a 2D galvanometer scanner (6230H, Cambridge Technology). An achromatic lens with a focal length of 40 mm (NT49-664, Edmund Optics) was used as the objective lens (please refer to Fig. 2(a)
Fig. 2 Axial resolution of the confocal microscopy. (a) schematic of the optical illumination and detection for theoretical estimation; (b) experimental measurement of axial resolution.
for more details). The output light from the second output arm O2 was detected by a photodiode (PD2, DET10A, Thorlabs) to record the energy of each laser pulse to compensate for the energy instability. The laser pulse energy out of the output arm O1 was 0.04 μJ.

The reflected light from the sample was detected by a 10-MHz Si photodiode (PD3, 2107-FC, New Focus) to form an FOCON image. A custom-built, high-frequency, unfocused ultrasonic transducer (center frequency: 40 MHz, active element size: 1 mm) detected the induced PA signals to form a PAM image. The data acquisition was triggered by laser pulses detected by PD1 (DET10A, Thorlabs) to avoid laser jittering.

This detection scheme permitted the natural integration of LSOR-PAM with FOCON, which allows simultaneous imaging the optical absorption and scattering contrasts without employing a second light source.

We applied the multimodal system to image the ears of a Swiss Webster mouse (body weight 20 g, Charles River Labs) in vivo. After the hairs on one of the ears were gently removed with commercial non-irritating hair removing lotion (Surgi Cream, Ardell), the animal was placed on a homemade animal holder for imaging. During experiments, a mixture of 1% isoflurane with medical grade oxygen was ventilated to the animal through a commercial non-rebreathing anesthesia system (M3000, LEI Medical) at a flow rate of 1 l/min to keep the animal motionless. All experimental animal procedures conformed to the laboratory animal protocol approved by the Animal Care and Use Committee of the University of Wisconsin–Milwaukee (UWM).

3. Results and discussions

3.1 Spatial resolution of LSOR-PAM

3.2 Spatial resolution of FOCON

Figure 2(a) gives a detailed schematic of the optical illumination and detection in the FOCON. The optical fiber (SM600, Thorlabs) has a numerical aperture (NA) of 0.12 and a mode field diameter of 4.6 µm. The fiber collimator (PAF-X-7-A, Thorlabs) has a NA of 0.29 and an output beam waist diameter of 1.2 mm. The beam was further expanded to 10 mm in diameter before it was focused by the objective lens, which yields an objective NA of 0.125.

Because the optical fiber acts as the pinhole, the lateral resolution of FOCON improves from the lateral resolution of the LSOR-PAM and can be calculated as kλ/NA, where λ is the optical wavelength, NA is the numerical aperture of objective lens, k is a coefficient from 0.37 to 0.51 depending on the mode field radius of the optical fiber [11

11. T. Dabbs and M. Glass, “Fiber-optic confocal microscope: FOCON,” Appl. Opt. 31(16), 3030–3035 (1992). [CrossRef] [PubMed]

]. Taking λ = 0.578 µm, NA = 0.125, and the mode field radius of 2.3 µm, the lateral resolution of the confocal microscope is around 2 µm in air.

The axial resolution of FOCON was first estimated theoretically. The axial resolution r can be expressed as r = λA/(πnNA 1 2) [12

12. M. Gu, Principles of three-dimensional imaging in confocal microscopes (Word Scientific, 1996).

], where λ is the optical wavelength, n is the refractive index of the medium, NA 1 is the NA of the objective lens, A is the normalized fiber spot size and is expressed as (2πr 0 NA 2)2, where r 0 is the mode field radius of the optical fiber and NA 2 is the minimum of the NA of the optical fibers and the collimating lens. Taking λ = 0.578 µm, NA 1 = 0.125, r 0 = 2.3 µm, and NA 2 = 0.12, we have r = 106 µm in air.

The axial resolution of FOCON in air was further quantified experimentally. A silver mirror was used as the sample and was translated along the z axis. A total distance of 300 µm was translated at a step size of 20 µm. The normalized reading from PD3 at each step is shown in Fig. 2(b). The experimental data were then fit by a Gaussian function aexp(-((x-b)/c)2)+d, where a, b, c, and d are unknown constants and d counts for the light reflection directly from the fiber tip of arm O1. The full-width-half-max (FWHM) value of the fitted Gaussian function gives the experimental measurement of axial resolution at 108 µm, which agrees with the theoretical estimation very well.

3.3 Phantom imaging

To demonstrate the complementary contrast of multimodal imaging, we imaged a printed mesh grid (grid size: 865 µm) as used in [1

1. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

]. The grid has carbon lines on a printing transparency, where the dark lines have strong optical absorption. Figure 3
Fig. 3 Images of a printed mesh grid using the fused PAM and FOCON. (a) LSOR-PAM image; (b) FOCON image.
shows the imaging results. As can be observed, FOCON and PAM imaged opposite contrasts: the black printed lines induced strong PA signal [Fig. 3(a)] but very weak back-scattered photons for confocal microscopy [Fig. 3(b)] due to stronger optical absorption compared with the transparency. Since the LSOR-PAM and confocal microscopy share the same optical focus, the two images are intrinsically registered.

3.4 In vivo animal imaging

The in vivo images of a mouse ear are shown in Fig. 4
Fig. 4 In vivo multimodal imaging of a mouse ear. (a) LSOR-PAM image showing microvasculature; (b) FOCON image showing skin structure. SB: sebaceous gland.
. Because the source of optical absorption is mainly hemoglobin and the Swiss Webster mouse has low skin melanin concentration, LSOR-PAM revealed the anatomy of microvessels [Fig. 4(a)] well. Sebaceous glands are embedded in the dermis and have higher optical absorption compared with the dermis; hence, they create shadows in the LSOR-PAM image, as highlighted by arrows.

Completely different anatomical features are observed in the FOCON image [Fig. 4(b)]. The detected photons are primarily reflected from epidermis and dermis. Since the sebaceous gland is more optically absorptive compared with the dermal tissue, fewer photons were back-reflected and the glands appeared to be dark regions in the image, which agrees with other reported results [13

13. E. Salomatina, B. Jiang, J. Novak, and A. N. Yaroslavsky, “Optical properties of normal and cancerous human skin in the visible and near-infrared spectral range,” J. Biomed. Opt. 11(6), 064026 (2006). [CrossRef]

]. As highlighted in Fig. 4, the spatial locations of the glands correspond well in the two images. The contrast-to-noise ratios of the LSOR-PAM and FOCON images are 36 dB and 12 dB, respectively, which clearly demonstrates the advantage of LSOR-PAM when optical absorption contrast dominates. We noticed that blood vessels do not manifest themselves as dark regions in the FOCON image. It is, presumably, because that LSOR-PAM relies on only single-trip photon propagation while FOCON requires round-trip photons propagation and thus has less sensitivity in order to image blood vessels without contrast agents at a low NA.

4. Conclusion

The novelty of this work comes from the integration, for the first time, of confocal microscopy with photoacoustic microscopy based on shared optical delivery, optical scanning, and light source for multimodal imaging. The significance of this work is that it exhibits the feasibility of integrating LSOR-PAM with the well-established confocal scanning laser ophthalmoscope (cSLO) [14

14. R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]

] for potential multimodal ophthalmic imaging based on a single light source. The 0.125 NA of the objective lens used in this study was specifically chosen to be close to the optical parameters of human eyes [15

15. D. A. Atchison, and G. Smith, Optics of human eye (Butterworth-Heinemann, 2000).

]. Although not demonstrated experimentally, this system is readily expandable to conduct confocal fluorescence microscopy [16

16. S. Schmitz-Valckenberg, M. Fleckenstein, A. P. Göbel, K. Sehmi, F. W. Fitzke, F. G. Holz, and A. Tufail, “Evaluation of autofluorescence imaging with the scanning laser ophthalmoscope and the fundus camera in age-related geographic atrophy,” Am. J. Ophthalmol. 146(2), 183–192 (2008). [CrossRef] [PubMed]

] by adding appropriate band-pass filters and a high-sensitivity optical detector (such as a photon multiplication tube), yet still sharing the same light source with LSOR-PAM.

Acknowledgments

This work is supported in part by the UWM start-up fund, the UWM Research Growth Initiative grant, the Shaw Scientist Award, and the Juvenile Diabetes Research Foundation Innovative Grant 5-2009-498 to H. F. Zhang. Support is also provided in part by the National Institutes of Health grant 7R21EB008800-02 to S. Jiao. Both H. F. Zhang and S. Jiao are corresponding authors to this paper.

References and Links

1.

H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

2.

A. De La Zerda, C. Zavaleta, S. Keren, S. Vaithilingam, S. Bodapati, Z. Liu, J. Levi, B. R. Smith, T. J. Ma, O. Oralkan, Z. Cheng, X. Y. Chen, H. J. Dai, B. T. Khuri-Yakub, and S. S. Gambhir, “Carbon nanotubes as photoacoustic molecular imaging agents in living mice,” Nat. Nanotechnol. 3(9), 557–562 (2008). [CrossRef] [PubMed]

3.

E. Z. Zhang, J. G. Laufer, R. B. Pedley, and P. C. Beard, “In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy,” Phys. Med. Biol. 54(4), 1035–1046 (2009). [CrossRef] [PubMed]

4.

K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

5.

Z. X. Xie, S. L. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]

6.

L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nat. Photonics 3(9), 503–509 (2009). [CrossRef]

7.

K. H. Song, C. H. Kim, C. M. Cobley, Y. N. Xia, and L. V. Wang, “Near-infrared gold nanocages as a new class of tracers for photoacoustic sentinel lymph node mapping on a rat model,” Nano Lett. 9(1), 183–188 (2009). [CrossRef]

8.

J. Laufer, E. Zhang, G. Raivich, and P. Beard, “Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner,” Appl. Opt. 48(10), D299–D306 (2009). [CrossRef] [PubMed]

9.

S. Jiao, R. Knighton, X. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]

10.

S. L. Jiao, Z. X. Xie, H. F. Zhang, and C. A. Puliafito, “Simultaneous multimodal imaging with integrated photoacoustic microscopy and optical coherence tomography,” Opt. Lett. 34(19), 2961–2963 (2009). [CrossRef] [PubMed]

11.

T. Dabbs and M. Glass, “Fiber-optic confocal microscope: FOCON,” Appl. Opt. 31(16), 3030–3035 (1992). [CrossRef] [PubMed]

12.

M. Gu, Principles of three-dimensional imaging in confocal microscopes (Word Scientific, 1996).

13.

E. Salomatina, B. Jiang, J. Novak, and A. N. Yaroslavsky, “Optical properties of normal and cancerous human skin in the visible and near-infrared spectral range,” J. Biomed. Opt. 11(6), 064026 (2006). [CrossRef]

14.

R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]

15.

D. A. Atchison, and G. Smith, Optics of human eye (Butterworth-Heinemann, 2000).

16.

S. Schmitz-Valckenberg, M. Fleckenstein, A. P. Göbel, K. Sehmi, F. W. Fitzke, F. G. Holz, and A. Tufail, “Evaluation of autofluorescence imaging with the scanning laser ophthalmoscope and the fundus camera in age-related geographic atrophy,” Am. J. Ophthalmol. 146(2), 183–192 (2008). [CrossRef] [PubMed]

OCIS Codes
(110.0180) Imaging systems : Microscopy
(110.5120) Imaging systems : Photoacoustic imaging
(170.1790) Medical optics and biotechnology : Confocal microscopy

ToC Category:
Microscopy

History
Original Manuscript: November 30, 2009
Revised Manuscript: December 31, 2009
Manuscript Accepted: January 4, 2010
Published: January 11, 2010

Virtual Issues
Vol. 5, Iss. 3 Virtual Journal for Biomedical Optics

Citation
Hao F. Zhang, Jing Wang, Qing Wei, Tan Liu, Shuliang Jiao, and Carmen A. Puliafito, "Collecting back-reflected photons in photoacoustic microscopy," Opt. Express 18, 1278-1282 (2010)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-18-2-1278


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]
  2. A. De La Zerda, C. Zavaleta, S. Keren, S. Vaithilingam, S. Bodapati, Z. Liu, J. Levi, B. R. Smith, T. J. Ma, O. Oralkan, Z. Cheng, X. Y. Chen, H. J. Dai, B. T. Khuri-Yakub, and S. S. Gambhir, “Carbon nanotubes as photoacoustic molecular imaging agents in living mice,” Nat. Nanotechnol. 3(9), 557–562 (2008). [CrossRef] [PubMed]
  3. E. Z. Zhang, J. G. Laufer, R. B. Pedley, and P. C. Beard, “In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy,” Phys. Med. Biol. 54(4), 1035–1046 (2009). [CrossRef] [PubMed]
  4. K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]
  5. Z. X. Xie, S. L. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]
  6. L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nat. Photonics 3(9), 503–509 (2009). [CrossRef]
  7. K. H. Song, C. H. Kim, C. M. Cobley, Y. N. Xia, and L. V. Wang, “Near-infrared gold nanocages as a new class of tracers for photoacoustic sentinel lymph node mapping on a rat model,” Nano Lett. 9(1), 183–188 (2009). [CrossRef]
  8. J. Laufer, E. Zhang, G. Raivich, and P. Beard, “Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner,” Appl. Opt. 48(10), D299–D306 (2009). [CrossRef] [PubMed]
  9. S. Jiao, R. Knighton, X. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]
  10. S. L. Jiao, Z. X. Xie, H. F. Zhang, and C. A. Puliafito, “Simultaneous multimodal imaging with integrated photoacoustic microscopy and optical coherence tomography,” Opt. Lett. 34(19), 2961–2963 (2009). [CrossRef] [PubMed]
  11. T. Dabbs and M. Glass, “Fiber-optic confocal microscope: FOCON,” Appl. Opt. 31(16), 3030–3035 (1992). [CrossRef] [PubMed]
  12. M. Gu, Principles of three-dimensional imaging in confocal microscopes (Word Scientific, 1996).
  13. E. Salomatina, B. Jiang, J. Novak, and A. N. Yaroslavsky, “Optical properties of normal and cancerous human skin in the visible and near-infrared spectral range,” J. Biomed. Opt. 11(6), 064026 (2006). [CrossRef]
  14. R. H. Webb, G. W. Hughes, and F. C. Delori, “Confocal scanning laser ophthalmoscope,” Appl. Opt. 26(8), 1492–1499 (1987). [CrossRef] [PubMed]
  15. D. A. Atchison, and G. Smith, Optics of human eye (Butterworth-Heinemann, 2000).
  16. S. Schmitz-Valckenberg, M. Fleckenstein, A. P. Göbel, K. Sehmi, F. W. Fitzke, F. G. Holz, and A. Tufail, “Evaluation of autofluorescence imaging with the scanning laser ophthalmoscope and the fundus camera in age-related geographic atrophy,” Am. J. Ophthalmol. 146(2), 183–192 (2008). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 
Fig. 4
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited