Imaging vibrating vocal folds with a high speed 1050 nm swept source OCT and ODT

Gangjun Liu, Marc Rubinstein, Arya Saidi, Wenjuan Qi, Allen Foulad, Brian Wong, and Zhongping Chen  »View Author Affiliations


Optics Express, Vol. 19, Issue 12, pp. 11880-11889 (2011)
http://dx.doi.org/10.1364/OE.19.011880


View Full Text Article

Acrobat PDF (1349 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Vocal fold vibration is vital in voice production and the correct pitch of speech. We have developed a high speed functional optical coherence tomography (OCT) system with a center wavelength of 1050 nm and an imaging speed of 100,000 A-lines per second. We imaged the vibration of an ex-vivo swine vocal fold. At an imaging speed of 100 frames per second, we demonstrated high quality vocal fold images during vibration. Functional information, such as vibration frequency and vibration amplitude, was obtained by analyzing the tissue surface during vibration. The axial direction velocity distribution in the cross-sectional images of the vibrating vocal folds was obtained with the Doppler OCT. The quantitative transverse direction velocity distribution in the cross-sectional images was obtained with the Doppler variance images.

© 2011 OSA

1. Introduction

Optical coherence tomography (OCT) is a powerful interferometric technology used to obtain cross-sectional tissue images noninvasively with micrometer resolution, millimeter penetration depth, and a video-rate imaging speed [1

1. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]

]. Due to its non-contact and high resolution nature, OCT has become a valuable tool in a number of medical fields. Recently, OCT technique has been increasingly used to perform functional imaging as well. Doppler optical coherence tomography (DOCT) or optical Doppler tomography (ODT) is one kind of functional extension of OCT, which combines the Doppler principle with OCT. ODT has been widely used for in-vivo imaging of blood flow in live animals and human beings [2

2. Z. Chen, T. E. Milner, D. Dave, and J. S. Nelson, “Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media,” Opt. Lett. 22(1), 64–66 (1997), http://www.opticsinfobase.org/abstract.cfm?URI=ol-22-1-64. [CrossRef] [PubMed]

7

7. B. White, M. Pierce, N. Nassif, B. Cense, B. Park, G. Tearney, B. Bouma, T. Chen, and J. de Boer, “In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical coherence tomography,” Opt. Express 11(25), 3490–3497 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-25-3490. [CrossRef] [PubMed]

].

Clinically, vocal fold vibration has been widely imaged using laryngeal videostroboscopy and high speed video, as these methods provide clinically relevant important information on vocal fold behavior in health and pathology. Lohscheller et al. obtained functional information regarding the vibration vocal folds, such as vibrating frequency, velocity, and acceleration [8

8. R. Patel, S. Dailey, and D. Bless, “Comparison of high-speed digital imaging with stroboscopy for laryngeal imaging of glottal disorders,” Ann. Otol. Rhinol. Laryngol. 117(6), 413–424 (2008). [PubMed]

10

10. J. Lohscheller and U. Eysholdt, “Phonovibrogram visualization of entire vocal fold dynamics,” Laryngoscope 118(4), 753–758 (2008). [CrossRef] [PubMed]

]. Although videostroboscopy provides an excellent method to dynamically assess the vocal folds, it only provides information on the surface of the vocal folds; therefore, the condition of the vocal folds underneath the surface remains unknown using these systems. There is a wide spectrum of diseases that can occur in the vocal folds, including benign polyps, premalignant and malignant lesions. Differentiating these afflictions using only direct visualization can be difficult and a biopsy is often required. The key element to differentiate these lesions has to do with visualizing the integrity of the basement membrane. A loss of the basement membrane integrity is a hallmark of cancers of the vocal fold. Currently, there is no reliable noninvasive method to diagnose laryngeal cancer without introducing a biopsy. However, doing a biopsy in the vocal folds can come with its own risk of creating permanent damage to the vocal folds; therefore, the importance of using a noninvasive imaging method that can visualize below the surface of the vocal cords, such as Ultrasound, and OCT, is highly practical. Ultrasound has also been used to image the vibrating vocal folds [11

11. T. Y. Hsiao, C. L. Wang, C. N. Chen, F. J. Hsieh, and Y. W. Shau, “Elasticity of human vocal folds measured in vivo using color Doppler imaging,” Ultrasound Med. Biol. 28(9), 1145–1152 (2002). [CrossRef] [PubMed]

13

13. M. Dedecjus, Z. Adamczewski, J. Brzeziński, and A. Lewiński, “Real-time, high-resolution ultrasonography of the vocal folds--a prospective pilot study in patients before and after thyroidectomy,” Langenbecks Arch. Surg. 395(7), 859–864 (2010). [CrossRef] [PubMed]

]. Although functional information can be obtained from an ultrasound, color Doppler ultrasound images suffer from low resolution and low frame rate. Hence, there is immense value in being able to image dynamically and in real-time image the structure and characteristics of the vibrating vocal folds, as much pathology is below the thin subsurface of this organ. Recently, imaging vibration vocal folds using OCT has been demonstrated by several groups [14

14. K. Lüerßen, H. Lubatschowski, N. Radicke, and M. Ptok, “Optical characterization of vocal folds using optical coherence tomography,” Med. Laser Appl. 21(3), 185–190 (2006). [CrossRef]

16

16. J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]

]. Lüerßen et al. demonstrated the vibration vocal fold OCT image at an imaging speed of 10 frames per second [14

14. K. Lüerßen, H. Lubatschowski, N. Radicke, and M. Ptok, “Optical characterization of vocal folds using optical coherence tomography,” Med. Laser Appl. 21(3), 185–190 (2006). [CrossRef]

]. Our group has demonstrated in-vivo imaging of human vibrating vocal folds with a 1.3 µm, 20 kHz swept source OCT system and a hand-held probe [15

15. L. Yu, G. Liu, M. Rubinstein, A. Saidi, B. J. Wong, and Z. Chen, “Office-based dynamic imaging of vocal cords in awake patients with swept-source optical coherence tomography,” J. Biomed. Opt. 14(6), 064020 (2009). [CrossRef]

]. Functional information, such as vibrating frequency, was obtained by analysis of the OCT structure images. Kober et al. used a triggered 10 kHz swept source OCT system to image the excised half calf larynx [16

16. J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]

]. With the help of the particle image analysis method, the authors obtained the velocity vector in the cross section images from the OCT structure images.

For humans, the actual vocal fundamental frequencies vary by sex. In females it is approximately 200 Hz and in males it is approximately 120 Hz. For imaging such high frequency movement, a high speed imaging system is essential to provide high frame rate images for the analysis. In addition, Doppler OCT requires much more dense scanning between A-lines. In order to cover a large enough field of view and obtain high quality Doppler images at the same time, a fast system is essential to provide high frame rate.

In this paper, we demonstrate functional imaging of vibrating vocal folds ex-vivo with a high speed swept source OCT and ODT system. The system has a maximum imaging speed of 100 kHz A-line per second, a central wavelength of 1.05 µm, and a depth resolution of 7 µm. The functional information regarding the vibrating vocal folds, such as vibrating frequency, vibrating amplitude, and speed was obtained by fitting the surface curve of the vibrating vocal fold. To the best of our knowledge, this is the first time high quality and cross-sectional velocity distribution images of the vibrating vocal fold were obtained with an OCT and ODT system at a frame rate of 100 frames per second.

2. Experiment setup and sample preparation

2.1 OCT system setup

The schematic of the OCT and ODT system is shown in Fig. 1
Fig. 1 Schematic of the high speed OCT and ODT system. GM: galvanometer mirror; CG: cover glass; BD: balanced photo detector.
. The laser source is a high speed swept laser source with a central wavelength of 1050 nm and a sweeping speed of 100 kHz (Axsun Technology, Billerica, MA). The MEMS-based compact laser source has a high coherence length (>10 mm), and these MEMS-based swept source OCT systems have demonstrated a high sensitivity roll-off in ophthalmology applications [17

17. B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-19-20029. [CrossRef] [PubMed]

]. The laser source output is split into the reference and sample arms by a 20:80 coupler with 80% in the reference arm and 20% in the sample arm. In the reference arm, the light is further split by a 95:5 coupler. About 95% of the light in the reference is sent to cause interference with the collected back-reflection or backscattering signal from the sample. This interference signal is detected by a balanced photon detector and digitized by a high speed digitizer (ATS 9350, Alazar Technologies Inc., Pointe-Claire, QC, Canada). The remaining 5% of the light in the reference arm is sent to a non-balanced photon-detector after passing through cover slides and is then finally digitized by another channel in the digitizer. The system can work in two modes, i.e. OCT or ODT modes. In OCT mode, only the signal from the channel that connects to the balanced photon-detector is acquired and digitized. In ODT mode, signals from both channels are acquired. The measured system sensitivity is 101.2 dB around the zero path difference.

The phase stability of the system is very important for a phase resolved Doppler OCT system. Due to the mechanical scanning components used in the tunable filters of the swept source laser, the swept source laser based FD ODT systems encounter worse phase stability issues than spectrometer-based FD ODT systems. Usually, a static surface may be added as a reference to correct phase error and improve the phase stability. Our group has used the top surface of chick chorioallantoic membrane (CAM) as a reference to obtain the blood flow in a CAM [18

18. J. Zhang and Z. Chen, “In vivo blood flow imaging by a swept laser source based Fourier domain optical Doppler tomography,” Opt. Express 13(19), 7449–7457 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-19-7449. [CrossRef] [PubMed]

]. Vakoc et al. have proposed a method to tap 1% of the sample arm and direct it at a calibration mirror, which is positioned near the maximum imaging range of the system [19

19. B. Vakoc, S. Yun, J. de Boer, G. Tearney, and B. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13(14), 5483–5493 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-14-5483. [CrossRef] [PubMed]

]. In this manuscript, a common-path method was used to correct phase errors similar to that presented by Adler et al. [20

20. D. C. Adler, R. Huber, and J. G. Fujimoto, “Phase-sensitive optical coherence tomography at up to 370,000 lines per second using buffered Fourier domain mode-locked lasers,” Opt. Lett. 32(6), 626–628 (2007), http://www.opticsinfobase.org/abstract.cfm?URI=ol-32-6-626. [CrossRef] [PubMed]

]. We tapped 5% of the light from the reference arm and pass it through a cover glass with a thickness of 1mm. The two surfaces of the cover glass generated an inference fringe. This produced a reference surface at a depth corresponding to the thickness of the cover glass. By subtracting a portion of the phase difference of the reference surface location from the phase difference of the sample signal, the phase error was corrected [19

19. B. Vakoc, S. Yun, J. de Boer, G. Tearney, and B. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13(14), 5483–5493 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-14-5483. [CrossRef] [PubMed]

]. Figures 2 (a)
Fig. 2 (a) OCT images of a mirror; (b) Adjacent A-line phase difference before correction; (c) Adjacent A-line phase difference after correction.
and 2(b) show the OCT structure image and the phase difference between successive A-lines before correction for a mirror respectively. Figure 2(c) shows the phase difference between successive A-lines after the correction. The improvement can be clearly seen by comparing the images in Fig. 2(b) and Fig. 2(c). In the method proposed here, the reference surface calibration was obtained by taking a portion of the reference arm light in order to be detected by another channel of the digitizer. The sample signals were not affected by the calibration procedure, so that the system parameters, such as sensitivity and imaging range, were kept just the same as in a normal non-calibration procedure.

2.2 Sample preparation

Fresh porcine larynges, with an intact trachea, were obtained from a local biological tissues supply company. Then, larynges were dissected with removal of supraglottic tissue exposing the vocal fold, but leaving key structures (arytenoid cartilages, anterior commissure, thyroid cartilage) intact. A nylon suture was placed to approximate position the arytenoid cartilages and thus create adduction of the vocal folds. Once the vocal folds were exposed, the larynx was mounted on a custom made mount and air supply device. A cuffed endotracheal tube was placed from below into the trachea, and to avoid air leakage the cuff was inflated. Then, warm air at different flow velocities was delivered through the endotracheal tube through the trachea and past the glottis to vibrate the vocal folds. Figures 3 (a) and (b)
Fig. 3 (a) Photograph of side view of the porcine larynx. (b) Photograph of top view of the porcine larynx.
show the side view and top view photographs of the larynges mounted on the holder. A total of 6 larynges were prepared and investigated in this study.

3. Experimental results and analysis

3.1 M-mode imaging

The M-mode was first used to image the vibrating vocal fold. In M-mode imaging, the laser beam is not scanning and OCT images provide the depth profile of a single location at different times. The location of the incident beam is shown as the red dot in Fig. 3(b). Figure 4
Fig. 4 M-mode OCT image of the vibrating vocal fold.
shows the M-mode OCT structure image. The oscillation pattern of the vocal fold can be clearly seen from the images. The parameters such as oscillation period, amplitude, and speed are important to analysis of the imaged sample. In order to get these parameters, we first found the surface curve of the vocal fold by an intensity threshold method. The surface is overlapped with the OCT image as depicted by the solid red line in Fig. 4. Hsaio et al. used a bilaterally fixed-ends vibrating string model to simulate the vocal vibration [11

11. T. Y. Hsiao, C. L. Wang, C. N. Chen, F. J. Hsieh, and Y. W. Shau, “Elasticity of human vocal folds measured in vivo using color Doppler imaging,” Ultrasound Med. Biol. 28(9), 1145–1152 (2002). [CrossRef] [PubMed]

]. At a fixed location, the surface curve can be described by a sine function:
y(t)=y0+Asin[π(ttc)w],
(1)
where y(t) is the surface location at time t, A is the amplitude of the vibration, wis the period of the vibration, tc and y0 are two parameters decided by the acquisition start time and initial surface location of the sample at rest respectively. By fitting the surface curve in Fig. 4 with Eq. (1), these parameters can be obtained. Figure 5
Fig. 5 Fitting of the surface of the vibrating vocal fold with Eq. (1).
shows the curve fitting results together with the data. Clearly, the sine function fits well with the curve of tissue surface movement, especially in the down slope of the vibration. The amplitude Awas found to be 709 μmand w was 5.3 ms. The period was 10.6 ms and the vibration frequency was 94.3 Hz. From equation, the velocity of the tissue surface movement can also be obtained. This can be obtained with the following equation:
v(t)=d[y(t)]dt=Aπwcos[π(ttc)w],
(2)
where v(t)the velocity at the tissue surface and the other parameters are the same as in Eq. (1). The maximum velocity is shown in Eq. (3), which is 0.415 m/s in this case.

vmax=Aπw.
(3)

We obtained the velocity of the tissue surface with the above mentioned fitting method. In addition to the velocity of the tissue surface, the velocity distribution beneath the tissue surface is also valuable. Kobler et al. have used a particle image velocimetry to obtain the velocity distribution [16

16. J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]

]. In our case, we used the phase resolved ODT to obtain the velocity distribution. Phase resolved ODT has been used to image blood vessels in tissue; in addition, it utilizes the phase difference between adjacent A-lines to estimate the velocity value along the incident light beam direction. The Doppler frequency caused by the sample movement in the axial direction can be obtained by the following equation:
fd=2vzλc=Δθ2πT,
(4)
where fd is the Doppler frequency, vz is the sample velocity along the light beam direction, λcis the central wavelength of the incident beam, Δθis the phase difference between adjacent A-lines, and Tis the time difference between adjacent A-lines. Figure 6
Fig. 6 Color Doppler image of the vibrating vocal fold.
shows the velocity distribution of a cross-sectional image obtained with the phase resolved Doppler method. In Fig. 6 a quasi-periodic pattern was caused by phase wrapping, and the phase difference is wrapped between π and π. However, wrapped phase images also give the qualitative information regarding the acceleration. The absolute value of the velocity can be obtained using the following simple method.

According to Eq. (4), we can find that the velocity may be expressed as:

vz=λcΔθ4πT.
(5)

3.2 B-Mode imaging

Due to the limited penetration depth of the OCT, the velocity distribution at greater depths inside the tissue cannot be resolved with the current technique. Slower vibrations may ease the requirement for deeper penetration, and some mechanical properties may be found from the superficial layers of the vocal fold. By controlling the volume of the air flow rate, we are able to control the frequency and amplitude of the vocal fold vibration. Figure 8
Fig. 8 OCT images of a vibrating vocal fold with frequency of around 1.1Hz. (a) B-mode OCT structure image. (b) B-mode color Doppler OCT image. (c) B-mode Doppler variance OCT image. Scale bar: 500 μm .
shows a B-mode OCT structure, color Doppler, and Doppler variance images of the vocal fold vibrating at slow frequency and small amplitude. The images are acquired at 100 frames per second. Figure 9
Fig. 9 OCT Movies of a vibrating vocal fold with frequency of around 1.1Hz. (a) B-mode OCT structure image (Media 1). (b) B-mode color Doppler OCT image (Media 2). (c) B-mode Doppler variance OCT image (Media 3).
is the movie of the B-mode images replayed at a slower 10 frames per second. The velocity distribution at different time points can be seen from the movie. Figure 10
Fig. 10 B-mode color Doppler images of a slow vibrating vocal fold. Scale bar: 500 um.
shows four ODT images extracted from the movie with the time difference between adjacent images being 12 milliseconds. The velocity distribution in the cross-section of the vocal fold can be obtained by the color Doppler images. An interesting phenomenon is that the change in velocity is in the radial direction, as pointed out by the blue arrows in Fig. 10. Although this phenomenon has been found in the waist region of the fast vibration case, as shown in Fig. 6 and Fig. 7, this phenomenon is more evident in the case of slower vibration like that in Fig. 10. Furthermore, the radial directed velocity in this area is changing when the wave travels from right to the left, as shown in the progression of Fig. 10(a)-(d). For these images, we can also quantify the wave traveling speed in the horizontal direction. The distance between the two yellow vertical lines in Fig. 10(a) and Fig. 10(c) is 224 μm. Since the time difference between Fig. 10(a) and Fig. 10(c) is 24 ms (12 ms x 2 = 24 ms), the velocity for the transverse wave is 9.3 mm/s (assume a constant velocity for the wave).

4. Discussion

For phase resolved Doppler OCT, the minimum detectable velocity is decided by the phase stability of the system. For the current setup, the phase stability of the system is measured to be less than 5 milliradians. According to Eq. (5) in this manuscript, this phase stability corresponds to a speed of 42 μm/s. The maximum speed with Doppler OCT is usually decided by the phase wrapping, which corresponds to a phase difference of2π. Furthermore, this phase difference corresponds to a speed of 0.0525 m/s. With the method introduced in section 3.1 of this paper, or phase unwrapping methods, we are able to obtain a velocity larger than 0.0525 m/s. The accurate results that the system can provide are dependent upon the velocity, frequency, and amplitude of the vibration sample. The velocity is decided by the amplitude and frequency of the vibration sample. When the sample is vibrating too fast, the Doppler effect will induce erroneous depth information, and the axial resolution will be degraded [16

16. J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]

, 22

22. S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12(13), 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?URI=oe-12-13-2977. [CrossRef] [PubMed]

]. The duty cycle of the laser used in this study is around 50%, and the actual integration time period for one single A-line is around 5 μs. If we set the intolerable axial sample movement during one integration period as the axial resolution (7 μm) of the current system [16

16. J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]

, 22

22. S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12(13), 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?URI=oe-12-13-2977. [CrossRef] [PubMed]

], then the axial direction velocity that will not introduce an intolerable depth is 1.4 m/s in our setup. The largest velocity demonstrated in this experiment is 0.415 m/s. This corresponds to an axial movement of around 2 μm during the integration time of one A-line.

As mentioned in section 3.2, the B-mode images are not “snap-shot” images of the vibration sample. When the frame rate of the system is slower than the vibration frequency of the sample, the information for the whole frame is not obtained at the same time and taking the B-mode image as a “snap-shot” image may cause misleading results. However, the velocity and acceleration information provided by color Doppler and Doppler variance images are corrected because the phase resolved methods use 2 adjacent A-lines to obtain this information. A sliding window covering 50-100 A-lines was used to analyze these B-mode images in this study. The information provided in this window can be considered as an “instant” or “snap-shot” image. In order to obtain a “snap-shot” like B-scan image, the frame rate of the system must be much higher than the vibration frequency. This may be realized by decreasing the number of A-lines per frame or by increasing the A-line speed of the swept source laser.

Although bulk motion is not applicable in our current experiment, the human bulk motion will affect the results of the Doppler OCT for awake patient imaging. There is extensive experience regarding bulk motion from OCT ophthalmology applications [23

23. S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14(17), 7821–7840 (2006), http://www.opticsinfobase.org/abstract.cfm?URI=oe-14-17-7821. [CrossRef] [PubMed]

,24

24. L. An, H. M. Subhush, D. J. Wilson, and R. K. Wang, “High-resolution wide-field imaging of retinal and choroidal blood perfusion with optical microangiography,” J. Biomed. Opt. 15(2), 026011 (2010). [CrossRef] [PubMed]

]. The bulk motion will introduce bulk phase and usually the structural images will not be affected. The vibration frequency, amplitude and period are extracted from the structure images and we estimate that the bulk motion will not affect the extraction of these parameters. However, the velocity information is obtained with the phase-revolved method and will be affected by bulk motion. The bulk motion artifacts may not be able to be eliminated with the histogram-based statistic method usually adopted in ophthalmology application. A way to minimize this effect is to increase the speed of the system so that the imaging time is reduced. In the current setup for B-mode imaging, the imaging speed is 100 frames per second with 1000 A lines per frame. We can increase the frame rate further by decreasing the number of A lines per frame or by increasing the sweeping rate the laser by a buffering technique [17

17. B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-19-20029. [CrossRef] [PubMed]

].

5. Summary

Functional imaging of vibrating ex-vivo porcine vocal folds was demonstrated with a high speed swept source OCT and ODT system. The functional information regarding the vibrating vocal folds was obtained with this high speed system. The tissue surface of the vibrating vocal folds was extracted to obtain functional information such as vibration amplitude, vibration frequency, velocity, and acceleration. Color Doppler and Doppler variance methods were used to obtain the velocity distribution characteristics in the cross sections. Essentially, the use of this system or one similar to it by physicians in otolaryngology could show how laryngeal carcinomas and other afflictions differ from baseline characteristics, because factors such as velocity are monitored to show how the inertial movement of the tissue is affected.

Acknowledgment

This work was supported by the National Institutes of Health (EB-00293, EB-10090, RR-01192, HL-103764, HL-105215), Air Force Office of Scientific Research (FA9550-04-0101), and the Beckman Laser Institute Endowment.

Reference and links:

1.

D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]

2.

Z. Chen, T. E. Milner, D. Dave, and J. S. Nelson, “Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media,” Opt. Lett. 22(1), 64–66 (1997), http://www.opticsinfobase.org/abstract.cfm?URI=ol-22-1-64. [CrossRef] [PubMed]

3.

Z. Chen, T. E. Milner, S. Srinivas, X. J. Wang, A. Malekafzali, M. J. C. van Gemert, and J. S. Nelson, “Noninvasive imaging of in vivo blood flow velocity using optical Doppler tomography,” Opt. Lett. 22(14), 1119–1121 (1997), http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-22-14-1119. [CrossRef] [PubMed]

4.

J. A. Izatt, M. D. Kulkarni, S. Yazdanfar, J. K. Barton, and A. J. Welch, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett. 22(18), 1439–1441 (1997), http://www.opticsinfobase.org/abstract.cfm?URI=ol-22-18-1439. [CrossRef]

5.

Y. Zhao, Z. Chen, C. Saxer, Q. Shen, S. Xiang, J. F. de Boer, and J. S. Nelson, “Doppler standard deviation imaging for clinical monitoring of in vivo human skin blood flow,” Opt. Lett. 25(18), 1358–1360 (2000), http://www.opticsinfobase.org/abstract.cfm?URI=ol-25-18-1358. [CrossRef]

6.

Y. Zhao, Z. Chen, C. Saxer, S. Xiang, J. F. de Boer, and J. S. Nelson, “Phase-resolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity,” Opt. Lett. 25(2), 114–116, (2000), http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-25-2-114. [CrossRef]

7.

B. White, M. Pierce, N. Nassif, B. Cense, B. Park, G. Tearney, B. Bouma, T. Chen, and J. de Boer, “In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical coherence tomography,” Opt. Express 11(25), 3490–3497 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-25-3490. [CrossRef] [PubMed]

8.

R. Patel, S. Dailey, and D. Bless, “Comparison of high-speed digital imaging with stroboscopy for laryngeal imaging of glottal disorders,” Ann. Otol. Rhinol. Laryngol. 117(6), 413–424 (2008). [PubMed]

9.

J. Lohscheller, U. Eysholdt, H. Toy, and M. Dollinger, “Phonovibrography: mapping high-speed movies of vocal fold vibrations into 2-D diagrams for visualizing and analyzing the underlying laryngeal dynamics,” IEEE Trans. Med. Imaging 27(3), 300–309 (2008). [CrossRef] [PubMed]

10.

J. Lohscheller and U. Eysholdt, “Phonovibrogram visualization of entire vocal fold dynamics,” Laryngoscope 118(4), 753–758 (2008). [CrossRef] [PubMed]

11.

T. Y. Hsiao, C. L. Wang, C. N. Chen, F. J. Hsieh, and Y. W. Shau, “Elasticity of human vocal folds measured in vivo using color Doppler imaging,” Ultrasound Med. Biol. 28(9), 1145–1152 (2002). [CrossRef] [PubMed]

12.

C. G. Tsai, J. H. Chen, Y. W. Shau, and T. Y. Hsiao, “Dynamic B-mode ultrasound imaging of vocal fold vibration during phonation,” Ultrasound Med. Biol. 35(11), 1812–1818 (2009). [CrossRef] [PubMed]

13.

M. Dedecjus, Z. Adamczewski, J. Brzeziński, and A. Lewiński, “Real-time, high-resolution ultrasonography of the vocal folds--a prospective pilot study in patients before and after thyroidectomy,” Langenbecks Arch. Surg. 395(7), 859–864 (2010). [CrossRef] [PubMed]

14.

K. Lüerßen, H. Lubatschowski, N. Radicke, and M. Ptok, “Optical characterization of vocal folds using optical coherence tomography,” Med. Laser Appl. 21(3), 185–190 (2006). [CrossRef]

15.

L. Yu, G. Liu, M. Rubinstein, A. Saidi, B. J. Wong, and Z. Chen, “Office-based dynamic imaging of vocal cords in awake patients with swept-source optical coherence tomography,” J. Biomed. Opt. 14(6), 064020 (2009). [CrossRef]

16.

J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]

17.

B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-19-20029. [CrossRef] [PubMed]

18.

J. Zhang and Z. Chen, “In vivo blood flow imaging by a swept laser source based Fourier domain optical Doppler tomography,” Opt. Express 13(19), 7449–7457 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-19-7449. [CrossRef] [PubMed]

19.

B. Vakoc, S. Yun, J. de Boer, G. Tearney, and B. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13(14), 5483–5493 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-14-5483. [CrossRef] [PubMed]

20.

D. C. Adler, R. Huber, and J. G. Fujimoto, “Phase-sensitive optical coherence tomography at up to 370,000 lines per second using buffered Fourier domain mode-locked lasers,” Opt. Lett. 32(6), 626–628 (2007), http://www.opticsinfobase.org/abstract.cfm?URI=ol-32-6-626. [CrossRef] [PubMed]

21.

H. Ren, K. M. Brecke, Z. Ding, Y. Zhao, J. S. Nelson, and Z. Chen, “Imaging and quantifying transverse flow velocity with the Doppler bandwidth in a phase-resolved functional optical coherence tomography,” Opt. Lett. 27(6), 409–411 (2002), http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-27-6-409. [CrossRef]

22.

S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12(13), 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?URI=oe-12-13-2977. [CrossRef] [PubMed]

23.

S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14(17), 7821–7840 (2006), http://www.opticsinfobase.org/abstract.cfm?URI=oe-14-17-7821. [CrossRef] [PubMed]

24.

L. An, H. M. Subhush, D. J. Wilson, and R. K. Wang, “High-resolution wide-field imaging of retinal and choroidal blood perfusion with optical microangiography,” J. Biomed. Opt. 15(2), 026011 (2010). [CrossRef] [PubMed]

OCIS Codes
(170.3340) Medical optics and biotechnology : Laser Doppler velocimetry
(170.3890) Medical optics and biotechnology : Medical optics instrumentation
(170.4500) Medical optics and biotechnology : Optical coherence tomography

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: February 10, 2011
Revised Manuscript: April 29, 2011
Manuscript Accepted: May 12, 2011
Published: June 3, 2011

Virtual Issues
Vol. 6, Iss. 7 Virtual Journal for Biomedical Optics

Citation
Gangjun Liu, Marc Rubinstein, Arya Saidi, Wenjuan Qi, Allen Foulad, Brian Wong, and Zhongping Chen, "Imaging vibrating vocal folds with a high speed 1050 nm swept source OCT and ODT," Opt. Express 19, 11880-11889 (2011)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-19-12-11880


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). [CrossRef] [PubMed]
  2. Z. Chen, T. E. Milner, D. Dave, and J. S. Nelson, “Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media,” Opt. Lett. 22(1), 64–66 (1997), http://www.opticsinfobase.org/abstract.cfm?URI=ol-22-1-64 . [CrossRef] [PubMed]
  3. Z. Chen, T. E. Milner, S. Srinivas, X. J. Wang, A. Malekafzali, M. J. C. van Gemert, and J. S. Nelson, “Noninvasive imaging of in vivo blood flow velocity using optical Doppler tomography,” Opt. Lett. 22(14), 1119–1121 (1997), http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-22-14-1119 . [CrossRef] [PubMed]
  4. J. A. Izatt, M. D. Kulkarni, S. Yazdanfar, J. K. Barton, and A. J. Welch, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett. 22(18), 1439–1441 (1997), http://www.opticsinfobase.org/abstract.cfm?URI=ol-22-18-1439 . [CrossRef]
  5. Y. Zhao, Z. Chen, C. Saxer, Q. Shen, S. Xiang, J. F. de Boer, and J. S. Nelson, “Doppler standard deviation imaging for clinical monitoring of in vivo human skin blood flow,” Opt. Lett. 25(18), 1358–1360 (2000), http://www.opticsinfobase.org/abstract.cfm?URI=ol-25-18-1358 . [CrossRef]
  6. Y. Zhao, Z. Chen, C. Saxer, S. Xiang, J. F. de Boer, and J. S. Nelson, “Phase-resolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity,” Opt. Lett. 25(2), 114–116, (2000), http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-25-2-114 . [CrossRef]
  7. B. White, M. Pierce, N. Nassif, B. Cense, B. Park, G. Tearney, B. Bouma, T. Chen, and J. de Boer, “In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical coherence tomography,” Opt. Express 11(25), 3490–3497 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-25-3490 . [CrossRef] [PubMed]
  8. R. Patel, S. Dailey, and D. Bless, “Comparison of high-speed digital imaging with stroboscopy for laryngeal imaging of glottal disorders,” Ann. Otol. Rhinol. Laryngol. 117(6), 413–424 (2008). [PubMed]
  9. J. Lohscheller, U. Eysholdt, H. Toy, and M. Dollinger, “Phonovibrography: mapping high-speed movies of vocal fold vibrations into 2-D diagrams for visualizing and analyzing the underlying laryngeal dynamics,” IEEE Trans. Med. Imaging 27(3), 300–309 (2008). [CrossRef] [PubMed]
  10. J. Lohscheller and U. Eysholdt, “Phonovibrogram visualization of entire vocal fold dynamics,” Laryngoscope 118(4), 753–758 (2008). [CrossRef] [PubMed]
  11. T. Y. Hsiao, C. L. Wang, C. N. Chen, F. J. Hsieh, and Y. W. Shau, “Elasticity of human vocal folds measured in vivo using color Doppler imaging,” Ultrasound Med. Biol. 28(9), 1145–1152 (2002). [CrossRef] [PubMed]
  12. C. G. Tsai, J. H. Chen, Y. W. Shau, and T. Y. Hsiao, “Dynamic B-mode ultrasound imaging of vocal fold vibration during phonation,” Ultrasound Med. Biol. 35(11), 1812–1818 (2009). [CrossRef] [PubMed]
  13. M. Dedecjus, Z. Adamczewski, J. Brzeziński, and A. Lewiński, “Real-time, high-resolution ultrasonography of the vocal folds--a prospective pilot study in patients before and after thyroidectomy,” Langenbecks Arch. Surg. 395(7), 859–864 (2010). [CrossRef] [PubMed]
  14. K. Lüerßen, H. Lubatschowski, N. Radicke, and M. Ptok, “Optical characterization of vocal folds using optical coherence tomography,” Med. Laser Appl. 21(3), 185–190 (2006). [CrossRef]
  15. L. Yu, G. Liu, M. Rubinstein, A. Saidi, B. J. Wong, and Z. Chen, “Office-based dynamic imaging of vocal cords in awake patients with swept-source optical coherence tomography,” J. Biomed. Opt. 14(6), 064020 (2009). [CrossRef]
  16. J. B. Kobler, E. W. Chang, S. M. Zeitels, and S. H. Yun, “Dynamic imaging of vocal fold oscillation with four-dimensional optical coherence tomography,” Laryngoscope 120(7), 1354–1362 (2010). [CrossRef] [PubMed]
  17. B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-19-20029 . [CrossRef] [PubMed]
  18. J. Zhang and Z. Chen, “In vivo blood flow imaging by a swept laser source based Fourier domain optical Doppler tomography,” Opt. Express 13(19), 7449–7457 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-19-7449 . [CrossRef] [PubMed]
  19. B. Vakoc, S. Yun, J. de Boer, G. Tearney, and B. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13(14), 5483–5493 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-14-5483 . [CrossRef] [PubMed]
  20. D. C. Adler, R. Huber, and J. G. Fujimoto, “Phase-sensitive optical coherence tomography at up to 370,000 lines per second using buffered Fourier domain mode-locked lasers,” Opt. Lett. 32(6), 626–628 (2007), http://www.opticsinfobase.org/abstract.cfm?URI=ol-32-6-626 . [CrossRef] [PubMed]
  21. H. Ren, K. M. Brecke, Z. Ding, Y. Zhao, J. S. Nelson, and Z. Chen, “Imaging and quantifying transverse flow velocity with the Doppler bandwidth in a phase-resolved functional optical coherence tomography,” Opt. Lett. 27(6), 409–411 (2002), http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-27-6-409 . [CrossRef]
  22. S. H. Yun, G. Tearney, J. de Boer, and B. Bouma, “Motion artifacts in optical coherence tomography with frequency-domain ranging,” Opt. Express 12(13), 2977–2998 (2004), http://www.opticsinfobase.org/abstract.cfm?URI=oe-12-13-2977 . [CrossRef] [PubMed]
  23. S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, “Optical coherence angiography,” Opt. Express 14(17), 7821–7840 (2006), http://www.opticsinfobase.org/abstract.cfm?URI=oe-14-17-7821 . [CrossRef] [PubMed]
  24. L. An, H. M. Subhush, D. J. Wilson, and R. K. Wang, “High-resolution wide-field imaging of retinal and choroidal blood perfusion with optical microangiography,” J. Biomed. Opt. 15(2), 026011 (2010). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Multimedia

Multimedia FilesRecommended Software
» Media 1: AVI (5893 KB)      QuickTime
» Media 2: AVI (3119 KB)      QuickTime
» Media 3: AVI (4225 KB)      QuickTime

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited