Wavelength encoding technique for particle analyses in hematology analyzer

Nelly Rongeat, Patrick Brunel, Jean-Philippe Gineys, Didier Cremien, Vincent Couderc, and Philippe Nérin  »View Author Affiliations


Optics Express, Vol. 19, Issue 15, pp. 14076-14082 (2011)
http://dx.doi.org/10.1364/OE.19.014076


View Full Text Article

Acrobat PDF (1051 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

The aim of this study is to combine multiple excitation wavelengths in order to improve accuracy of fluorescence characterization of labeled cells. The experimental demonstration is realized with a hematology analyzer based on flow cytometry and a CW laser source emitting two visible wavelengths. A given optical encoding associated to each wavelength allows fluorescence identification coming from specific fluorochromes and avoiding the use of noisy compensation method.

© 2011 OSA

1. Introduction

Flow cytometers are currently used to detect blood diseases [1

1. H. M. Shapiro, Practical Flow Cytometry, 2nd ed. (A. R. Liss, Inc., 1988).

3

3. J. L. Faucher, C. Lacronique-Gazaille, E. Frébet, F. Trimoreau, M. Donnard, D. Bordessoule, F. Lacombe, and J. Feuillard, ““6 markers/5 colors” extended white blood cell differential by flow cytometry,” Cytometry A 71A(11), 934–944 (2007). [CrossRef] [PubMed]

] and to monitor pathology evolution such as residual disease [4

4. C. Arnoulet, M. C. Béné, F. Durrieu, J. Feuillard, C. Fossat, B. Husson, H. Jouault, M. Maynadié, and F. Lacombe, “Four- and five-color flow cytometry analysis of leukocyte differentiation pathways in normal bone marrow: a reference document based on a systematic approach by the GTLLF and GEIL,” Cytometry B Clin. Cytom. 78(1), 4–10 (2010). [PubMed]

]. Technically, a flow cytometer gates stained cells previously labeled with antibodies conjugated to dyes or directly stained with dyes or fluorochromes, and determines their relative percentages. The simultaneous use of several dyes introduces difficulties, as spectral overlaps of their beam excitation and/or fluorescence emission spectra. To overcome these problems, specialists generally use a correction method called “compensation” [5

5. C. C. Stewart and S. J. Stewart, “Four color compensation,” Cytometry B 38(4), 161–175 (1999). [CrossRef]

10

10. K. Luong, P. Lundquist, R. Dalal, J. Lyle, and S. Turner, “Methods and systems for analysis of fluorescent reactions with modulated excitation,” CA patent 2711560 A1 (9 January 2009).

]. It consists in the removing a part of collected energy by an amount proportional to the spectral overlap between two fluorescence signals. Unfortunately, the main problem induced by the use of this method is the calculation of compensation factor which is based on the average values of chemical properties of fluorochromes. For a same fluorochrome family, these factors can be drastically different and are directly supplier dependent [6

6. M. Roederer, “Spectral compensation for flow cytometry: visualization artefacts, limitations and caveats,” Cytometry A 45(3), 194–205 (2001). [CrossRef]

]. A same product coming from different manufacturers can give different sets of compensations, causing a critical problem for the calibration of instruments and giving rise to erroneous measurements which could induce dangerous interpretation of results. In order to identify a given fluorescence among multiple signals or to enhance fluorescence detection, encoding techniques have already been used in several systems [10

10. K. Luong, P. Lundquist, R. Dalal, J. Lyle, and S. Turner, “Methods and systems for analysis of fluorescent reactions with modulated excitation,” CA patent 2711560 A1 (9 January 2009).

16

16. M. Dinkelmann and C. Rogers, “Expanding fluorescence detection options with the accuri C6 flow cytometer system,” Nat. Meth. 7, ••• (2010).

]. Indeed, fluorescence detection can be drastically improved if only one fluorochrome is excited at a given time by a single wavelength [17

17. K. Mossberg and M. Ericsson, “Detection of doubly stained fluorescent specimens using confocal microscopy,” J. Microsc. 158(Pt 2), 215–224 (1990). [CrossRef] [PubMed]

,18

18. X. Wang and I. Kurtz, “H+/base transport in principal cells characterized by confocal fluorescence imaging,” Am. J. Physiol. 259(2 Pt 1), C365–C373 (1990). [PubMed]

]. Unfortunately this method induces a long recording time when several fluorochromes are used. A new technique for simultaneous recording of multiple fluorochromes has been published in 1993 by Aslund et al. [13

13. N. Åslund and K. Carlsson, “Confocal scanning microfluorometry of dual-labelled specimens using two excitation wavelengths and lock-in detection technique,” Micron 24(6), 603–609 (1993). [CrossRef]

]. It consists in intensity-modulating the illumination wavelengths at different frequencies and sending them simultaneously on samples. A lock-in detection technique permits to separate the fluorescence contributions arising from each wavelength.

In this paper, we present results on simultaneous measurements of wavelength resolved scattering and fluorescences in a hematology analyzer. An encoding optical technique of incident light beams allows to decrease or to eliminate problems of multiparametric flow cytometry such as fluorescence “compensation”. In flow cytometry, this technique brings new insights in multiple fluorescence issue for multiparametric monitoring of a single particle. Our measurements are performed by using fluorescent polymer microspheres.

2. Experimental set-up and method

Experiments are realized with a hematology analyzer integrating principle of flow cytometry based on hydrodynamic focusing. In addition to flow cytometry technique, this apparatus allows blood cell population counting and identification by using scattering and fluorescence measurements. Particle flow circulating at the speed of 5 m/s is coupled with a specific optical layout presented on Fig. 1
Fig. 1 Optical system of the hematology analyzer. The particle flow circulates out of the plane, in our direction.
. The optical laser source (COBOLT Dual Calypso) produces two CW lines at 491 nm and 532 nm. It is composed of two different gain medium (Nd:YAG, NdYVO4) pumped by a single CW diode at 808 nm. Sum frequency generation between wavelengths at 914 nm and 1064 nm and second harmonic generation of the 1064 nm radiation allow the generation of the two visible wavelengths by means of a PPKTP crystal with two grating periods. The linearly polarized output laser beams are TEM00 with a narrow spectral bandwidth (<0.01pm) and a low noise (<0.3% rms). An acousto-optic modulator (AOTF nC.TN, company AA Opto Electronic) independently selects and encodes the two radiations. This special acousto-optic tunable filter can control simultaneously up to 8 distinct lines with narrow resolution (1-2 nm) and high extinction ratio. The laser beams are spatially shaped through an optical system composed of two spherical and cylindrical lenses to adjust the elliptical beam to flow chamber dimensions. A standard microfluidic device (HORIBA ABX Pentra 60 instrument, Montpellier, FR) is used to sample and to mix green and red fluorescent microspheres. Particle flow is analyzed during 15 seconds.

Particle counting is realized by using impedance measurement placed close to the electric flow chamber. The particles propagate through a micro-hole immersed in a liquid with known conductivity. The impedance modification between electrodes placed on each side of the micro-aperture reveals the presence of a particle. The impedance value also allows obtaining information on particle volumes [19

19. F. De Bisshop, “Electronic gating for particle/cell counting and sizing, DSP-operated,” IEEE Trans. Instrum. Meas. 58(9), 3159–3166 (2009). [CrossRef]

].

Optical measurements are performed after the electrical gating. In our set-up, the optical detection is composed of three channels. The first one is used to measure the forward light scattering (FSC) obtained in the vicinity of the beam axis. This scattered light is detected by a photodiode (S1223 Hamamatsu) placed behind a bar stop. The FSC is more sensitive to cell size than to other cellular characteristics. The axial light forms two half-moon of 1-3° apparent diameter coming from the scattering region. Fluorescence emissions at the red and green wavelengths coming from lit microspheres (respectively Duke Scientific Corporation R0300 and G0300) are selected through a Semrock dichroïc filter (FF01-513/17 and FF01-582/75). Excitation and emission spectra of these microspheres are shown on Fig. 2
Fig. 2 Absorption and emission spectra of green fluorescent microspheres. Absorption and emission spectra of red fluorescent microspheres. Excitation wavelengths of laser (λ = 491 nm and λ = 532 nm). Bandpass filters for two fluorescent detections (rectangles).
. Signal channel 1 (S1) is composed of a Semrock filter FF01-513/17, a Semrock dichroïc filter FF562-Di02, a focusing lens and a photomultiplier tube (Hamamatsu) to selectively record green fluorescence. Signal channel 2 (S2) is composed of a Semrock filter FF01-582/75, a focal lens and a photomultiplier (Hamamatsu) to measure red signal close to 610 nm. It is important to note that spectral emission of the two microspheres are overlapping in the 560-650 nm spectral window, thus increasing signal to noise ratio and decreasing accuracy of characterizations obtained from channel 2.

In our system, a programmable optical synthesizer based on an acousto-optic modulator allows the selection and the encoding of each incident beam. A given sinusoidal intensity modulation at 200 kHz and 300 kHz is associated to λ1 = 491 nm and λ2 = 532 nm respectively. In these conditions the incident multicolor beam expression is given by Eq. (1) and Eq. (2), sm1 and sm2 are the first and the second terms of Eq. (2) respectively.
I(t)=sm1(w1)+sm2(w2).
(1)
I(t)=I10(t)2[1+M1cos(w1t)]+I20(t)2[1+M2cos(w2t)].
(2)
I10 and I20 are the maximum intensities of optical signals; M1 and M2 represent the modulation amplitudes and w 1 and w 2 are the modulation frequencies.

The optical tunable filter is able to induce different modulations on optical signals with variable efficiency. This parameter has been measured for the two input wavelengths at 491 nm and 532 nm and is illustrated on Fig. 3
Fig. 3 Optical modulation efficiency of the tunable acousto-optic filter versus frequency for laser beams at 491 nm and 532 nm.
. A modulation efficiency higher than 40% can be experimentally achieved for both wavelengths but only for modulation frequency between 0 and 300 kHz.

3. Results

The incident and encoded radiations (Fig. 4(a)
Fig. 4 (a) Incident light beams at 491 nm and 532 nm modulated with frequencies w 1(200 kHz) et w 2(300 kHz) respectively. (b) Fluorescence signal at 510 nm from channel 1. (c) Fluorescence signal at 610 nm from channel 2 (experimental data).
) interact with green and red fluorescent microspheres propagating in microfluidic exposition chamber at a speed of 5m/s. Because of the microsphere size and speed, the interaction time with the CW multi-wavelength laser beam gives rise to output pulsed signal with a quasi-Gaussian shape and a duration of ~50µs (Fig. 4(b)). The modulated output radiations (sm, Eq. (3) and Eq. (4)) are shown on Fig. 4(b) and Fig. 4(c).
Sm(t)=exp[tt0]²[(kG+k'R)sm1(w1)+kRsm2(w2)]
(3)
Sm(t)=exp[tt0]²[(kG+k'R)I10(t)2[1+M1cos(w1t)]+kRI20(t)2[1+M2cos(w2t)]]
(4)
I10 and I20 are the maximum intensities of optical signals; M1, M2 represent the modulation amplitudes, w 1 and w 2 are the modulation frequencies; kG, kR and k’R (Eq. (5)) are factors depending on quantum yield of fluorescent beads (ρG and ρR of R0300 and G0300 beads respectively (Duke Scientific Corporation)).
kG=0,55.ρGkR=0,44.ρRk'R=1.ρR
(5)
The emitted fluorescence signal is detected close to 510 nm and 610 nm by two photomultipliers placed in channels 1 and 2 respectively. The photomultiplier of channel 1 (signal S1) only collects part of the green microsphere fluorescence signal (a.kG.sm1(w 1)) (Fig. 4(b)). Detector 2 (signal S2) collects a great part of the red microsphere fluorescent signal excited by both 491 nm (b.kR.sm2(w 2)) and 561 nm lights (c.k’R.sm1(w 1), and a small part of the green microsphere fluorescence (d.kG.sm1(w 1)) (Fig. 4(c) and Fig. 2) (a, b, c and d factors are depending of bandpass filters). The digital signal processing is realized with a home-made algorithm based on the Labview software allowing computation of each frequency component and identification of fluorescent signals coming from green or red microspheres.

4. Conclusion

We demonstrated microsphere characterization by using a dual laser source emitting two CW radiations at 491 nm and 532 nm. Measurements are realized with a specific hematology analyzer based on microfluidic technique. Encoding of each incident optical signal allows to discriminating fluorescence signals coming from green and red fluorescent microspheres despite their spectral overlap in the 550 - 650 nm spectral window. This new technique of spectral encoding opens new ways to improve diagnostic accuracy in the hematology field when fluorescence multiplexing is required. Digital signal processing of scattering and fluorescence channels enables to deliver accurate optical information for quantitative and qualitative experiments. This technique can be extended with a broadband source such supercontinuum light source operating in CW regime [20

20. A. Kudlinski, G. Bouwmans, O. Vanvincq, Y. Quiquempois, A. Le Rouge, L. Bigot, G. Mélin, and A. Mussot, “White-light cw-pumped supercontinuum generation in highly GeO(2)-doped-core photonic crystal fibers,” Opt. Lett. 34(23), 3631–3633 (2009). [CrossRef] [PubMed]

22

22. J. C. Travers, R. E. Kennedy, S. V. Popov, J. R. Taylor, H. Sabert, and B. Mangan, “Extended continuous-wave supercontinuum generation in a low-water-loss holey fiber,” Opt. Lett. 30(15), 1938–1940 (2005). [CrossRef] [PubMed]

].

Acknowledgments

This work was supported by ANRT.

References and links

1.

H. M. Shapiro, Practical Flow Cytometry, 2nd ed. (A. R. Liss, Inc., 1988).

2.

F. E. Craig and K. A. Foon, “Flow cytometric immunophenotyping for hematologic neoplasms,” Blood 111(8), 3941–3967 (2008). [CrossRef] [PubMed]

3.

J. L. Faucher, C. Lacronique-Gazaille, E. Frébet, F. Trimoreau, M. Donnard, D. Bordessoule, F. Lacombe, and J. Feuillard, ““6 markers/5 colors” extended white blood cell differential by flow cytometry,” Cytometry A 71A(11), 934–944 (2007). [CrossRef] [PubMed]

4.

C. Arnoulet, M. C. Béné, F. Durrieu, J. Feuillard, C. Fossat, B. Husson, H. Jouault, M. Maynadié, and F. Lacombe, “Four- and five-color flow cytometry analysis of leukocyte differentiation pathways in normal bone marrow: a reference document based on a systematic approach by the GTLLF and GEIL,” Cytometry B Clin. Cytom. 78(1), 4–10 (2010). [PubMed]

5.

C. C. Stewart and S. J. Stewart, “Four color compensation,” Cytometry B 38(4), 161–175 (1999). [CrossRef]

6.

M. Roederer, “Spectral compensation for flow cytometry: visualization artefacts, limitations and caveats,” Cytometry A 45(3), 194–205 (2001). [CrossRef]

7.

P. Nérin and D. Lefevre, “Device and method for multiparametric analysis of microscopic elements,” U.S. patent 7,777,869 (14 April 2006).

8.

R. Kapoor and M. Casstevens, “Evaluation of multicomponent mixtures using modulated light beam,” U.S. 2007/0096039 A1 (14 April 2006).

9.

G. Durack and J. Hatcher, “System and method for the measurement of multiple fluorescence emissions in a flow cytometry system,” U.S. patent 2008/0213915 A1 (23 April 2007).

10.

K. Luong, P. Lundquist, R. Dalal, J. Lyle, and S. Turner, “Methods and systems for analysis of fluorescent reactions with modulated excitation,” CA patent 2711560 A1 (9 January 2009).

11.

D. Jameson, E. Gratton, and R. Hall, “The measurement and analysis of heterogenous emissions by multifrequency phase and modulation fluorometry,” Appl. Spectrosc. Rev. 20(1), 55–106 (1984). [CrossRef]

12.

G. Durack and J. P. Robinson, Emerging Tools for Single-Cell Analysis: Advances in Optical Measurement Technologies, ed. (Wiley-Liss., 2000).

13.

N. Åslund and K. Carlsson, “Confocal scanning microfluorometry of dual-labelled specimens using two excitation wavelengths and lock-in detection technique,” Micron 24(6), 603–609 (1993). [CrossRef]

14.

S. W. Lin, C. H. Chang, D. Y. Wu, and C. H. Lin, “Digitally synchronized LCD projector for multi-color fluorescence excitation in parallel capillary electrophoresis detection,” Biosens. Bioelectron. 26(2), 717–722 (2010). [CrossRef] [PubMed]

15.

A. D. Donnenberg and V. S. Donnenberg, “Configuration of initial control parameters in photodetectors for multi-color flow cytometry,” U.S. patent 2010/0256943 A1 (1 April 2010).

16.

M. Dinkelmann and C. Rogers, “Expanding fluorescence detection options with the accuri C6 flow cytometer system,” Nat. Meth. 7, ••• (2010).

17.

K. Mossberg and M. Ericsson, “Detection of doubly stained fluorescent specimens using confocal microscopy,” J. Microsc. 158(Pt 2), 215–224 (1990). [CrossRef] [PubMed]

18.

X. Wang and I. Kurtz, “H+/base transport in principal cells characterized by confocal fluorescence imaging,” Am. J. Physiol. 259(2 Pt 1), C365–C373 (1990). [PubMed]

19.

F. De Bisshop, “Electronic gating for particle/cell counting and sizing, DSP-operated,” IEEE Trans. Instrum. Meas. 58(9), 3159–3166 (2009). [CrossRef]

20.

A. Kudlinski, G. Bouwmans, O. Vanvincq, Y. Quiquempois, A. Le Rouge, L. Bigot, G. Mélin, and A. Mussot, “White-light cw-pumped supercontinuum generation in highly GeO(2)-doped-core photonic crystal fibers,” Opt. Lett. 34(23), 3631–3633 (2009). [CrossRef] [PubMed]

21.

B. A. Cumberland, J. C. Travers, S. V. Popov, and J. R. Taylor, “Toward visible cw-pumped supercontinua,” Opt. Lett. 33(18), 2122–2124 (2008). [CrossRef] [PubMed]

22.

J. C. Travers, R. E. Kennedy, S. V. Popov, J. R. Taylor, H. Sabert, and B. Mangan, “Extended continuous-wave supercontinuum generation in a low-water-loss holey fiber,” Opt. Lett. 30(15), 1938–1940 (2005). [CrossRef] [PubMed]

OCIS Codes
(140.3580) Lasers and laser optics : Lasers, solid-state
(160.2540) Materials : Fluorescent and luminescent materials
(170.1530) Medical optics and biotechnology : Cell analysis
(170.4090) Medical optics and biotechnology : Modulation techniques
(170.4580) Medical optics and biotechnology : Optical diagnostics for medicine
(260.2510) Physical optics : Fluorescence

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: March 18, 2011
Revised Manuscript: April 29, 2011
Manuscript Accepted: May 10, 2011
Published: July 8, 2011

Virtual Issues
Vol. 6, Iss. 8 Virtual Journal for Biomedical Optics

Citation
Nelly Rongeat, Patrick Brunel, Jean-Philippe Gineys, Didier Cremien, Vincent Couderc, and Philippe Nérin, "Wavelength encoding technique for particle analyses in hematology analyzer," Opt. Express 19, 14076-14082 (2011)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-19-15-14076


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. H. M. Shapiro, Practical Flow Cytometry, 2nd ed. (A. R. Liss, Inc., 1988).
  2. F. E. Craig and K. A. Foon, “Flow cytometric immunophenotyping for hematologic neoplasms,” Blood 111(8), 3941–3967 (2008). [CrossRef] [PubMed]
  3. J. L. Faucher, C. Lacronique-Gazaille, E. Frébet, F. Trimoreau, M. Donnard, D. Bordessoule, F. Lacombe, and J. Feuillard, ““6 markers/5 colors” extended white blood cell differential by flow cytometry,” Cytometry A 71A(11), 934–944 (2007). [CrossRef] [PubMed]
  4. C. Arnoulet, M. C. Béné, F. Durrieu, J. Feuillard, C. Fossat, B. Husson, H. Jouault, M. Maynadié, and F. Lacombe, “Four- and five-color flow cytometry analysis of leukocyte differentiation pathways in normal bone marrow: a reference document based on a systematic approach by the GTLLF and GEIL,” Cytometry B Clin. Cytom. 78(1), 4–10 (2010). [PubMed]
  5. C. C. Stewart and S. J. Stewart, “Four color compensation,” Cytometry B 38(4), 161–175 (1999). [CrossRef]
  6. M. Roederer, “Spectral compensation for flow cytometry: visualization artefacts, limitations and caveats,” Cytometry A 45(3), 194–205 (2001). [CrossRef]
  7. P. Nérin and D. Lefevre, “Device and method for multiparametric analysis of microscopic elements,” U.S. patent 7,777,869 (14 April 2006).
  8. R. Kapoor and M. Casstevens, “Evaluation of multicomponent mixtures using modulated light beam,” U.S. 2007/0096039 A1 (14 April 2006).
  9. G. Durack and J. Hatcher, “System and method for the measurement of multiple fluorescence emissions in a flow cytometry system,” U.S. patent 2008/0213915 A1 (23 April 2007).
  10. K. Luong, P. Lundquist, R. Dalal, J. Lyle, and S. Turner, “Methods and systems for analysis of fluorescent reactions with modulated excitation,” CA patent 2711560 A1 (9 January 2009).
  11. D. Jameson, E. Gratton, and R. Hall, “The measurement and analysis of heterogenous emissions by multifrequency phase and modulation fluorometry,” Appl. Spectrosc. Rev. 20(1), 55–106 (1984). [CrossRef]
  12. G. Durack and J. P. Robinson, Emerging Tools for Single-Cell Analysis: Advances in Optical Measurement Technologies, ed. (Wiley-Liss., 2000).
  13. N. Åslund and K. Carlsson, “Confocal scanning microfluorometry of dual-labelled specimens using two excitation wavelengths and lock-in detection technique,” Micron 24(6), 603–609 (1993). [CrossRef]
  14. S. W. Lin, C. H. Chang, D. Y. Wu, and C. H. Lin, “Digitally synchronized LCD projector for multi-color fluorescence excitation in parallel capillary electrophoresis detection,” Biosens. Bioelectron. 26(2), 717–722 (2010). [CrossRef] [PubMed]
  15. A. D. Donnenberg and V. S. Donnenberg, “Configuration of initial control parameters in photodetectors for multi-color flow cytometry,” U.S. patent 2010/0256943 A1 (1 April 2010).
  16. M. Dinkelmann and C. Rogers, “Expanding fluorescence detection options with the accuri C6 flow cytometer system,” Nat. Meth. 7, ••• (2010).
  17. K. Mossberg and M. Ericsson, “Detection of doubly stained fluorescent specimens using confocal microscopy,” J. Microsc. 158(Pt 2), 215–224 (1990). [CrossRef] [PubMed]
  18. X. Wang and I. Kurtz, “H+/base transport in principal cells characterized by confocal fluorescence imaging,” Am. J. Physiol. 259(2 Pt 1), C365–C373 (1990). [PubMed]
  19. F. De Bisshop, “Electronic gating for particle/cell counting and sizing, DSP-operated,” IEEE Trans. Instrum. Meas. 58(9), 3159–3166 (2009). [CrossRef]
  20. A. Kudlinski, G. Bouwmans, O. Vanvincq, Y. Quiquempois, A. Le Rouge, L. Bigot, G. Mélin, and A. Mussot, “White-light cw-pumped supercontinuum generation in highly GeO(2)-doped-core photonic crystal fibers,” Opt. Lett. 34(23), 3631–3633 (2009). [CrossRef] [PubMed]
  21. B. A. Cumberland, J. C. Travers, S. V. Popov, and J. R. Taylor, “Toward visible cw-pumped supercontinua,” Opt. Lett. 33(18), 2122–2124 (2008). [CrossRef] [PubMed]
  22. J. C. Travers, R. E. Kennedy, S. V. Popov, J. R. Taylor, H. Sabert, and B. Mangan, “Extended continuous-wave supercontinuum generation in a low-water-loss holey fiber,” Opt. Lett. 30(15), 1938–1940 (2005). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.