Abstract

This paper proposes an automatic algorithm for the montage of OCT data sets, which produces a composite 3D OCT image over a large field of view out of several separate, partially overlapping OCT data sets. First the OCT fundus images (OFIs) are registered, using blood vessel ridges as the feature of interest and a two step iterative procedure to minimize the distance between all matching point pairs over the set of OFIs. Then the OCT data sets are merged to form a full 3D montage using cross-correlation. The algorithm was tested using an imaging protocol consisting of 8 OCT images for each eye, overlapping to cover a total retinal region of approximately 50x35 degrees. The results for 3 normal eyes and 3 eyes with retinal degeneration are analyzed, showing registration errors of 1.5±0.3 and 2.0±0.8 pixels respectively.

© 2011 OSA

1. Introduction

A number of recent papers have analyzed stitching algorithms for 3D microscopic images [5

5. M. Emmenlauer, O. Ronneberger, A. Ponti, P. Schwarb, A. Griffa, A. Filippi, R. Nitschke, W. Driever, and H. Burkhardt, “XuvTools: free, fast and reliable stitching of large 3D datasets,” J. Microsc. 233(1), 42–60 (2009). [CrossRef] [PubMed]

7

7. Y. Yu and H. Peng, “Automated high speed stitching of large 3D microscopic images,” Proc. of IEEE 2011 International Symposium on Biomedical Imaging: From Nano to Macro 238–241 (2011).

]. In general these algorithms are restricted to translations and are not well suited to work with clinical OCT images, where the position of the eye and the instrument alignment can vary substantially and unpredictably across different scans, even during a single imaging session. Furthermore OCT data sets suffer from axial eye motions which cause shifts (and sometimes rotations) between different B-scans. Therefore OCT data sets cannot really be registered through rigid 3D transformations. A paper [8

8. J. Yoo, I. V. Larina, K. V. Larin, M. E. Dickinson, and M. Liebling, “Increasing the field-of-view of dynamic cardiac OCT via post-acquisition mosaicing without affecting frame-rate or spatial resolution,” Biomed. Opt. Express 2(9), 2614–2622 (2011). [CrossRef] [PubMed]

] about increasing OCT FOV involves a very different type of OCT data sets and it is very much in the spirit of stitching of microscopic images.

Each 3D OCT data set has an associated 2D OCT fundus image (OFI) obtained by summing the OCT intensities along the A-scans [9

9. S. L. Jiao, R. Knighton, X. R. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]

12

12. Z. Hu, M. Niemeijer, M. D. Abràmoft, K. Lee, and M. K. Garvin, “Automated segmentation of 3-D spectral OCT retinal blood vessels by neural canal opening false positive suppression,” Med. Image Comput. Comput. Assist. Interv. 13(Pt 3), 33–40 (2010). [PubMed]

]. OFIs are qualitatively similar to traditional en face retinal images and many of the standard retinal features visible on fundus photography can be recognized, in particular the retinal vasculature. This paper proposes an automated algorithm for the montage of OCT data sets based on a montage of the corresponding OFIs. Once a 2D OFI montage is obtained, a process involving resampling, interpolation, and cross-correlation of appropriate portions of the original OCT data sets can be used to construct a full 3D OCT montage data set, whose OFI is the given OFI montage.

When working with OFIs we need to consider issues like changes in illumination and/or reflectivity unrelated to retinal characteristics, as well as speckle noise [28

28. J. M. Schmitt, S. H. Xiang, and K. M. Yung, “Speckle in optical coherence tomography,” J. Biomed. Opt. 4(1), 95–105 (1999). [CrossRef]

]. Therefore methods based on intensity correlation or SURF are not likely to produce optimal results. Blood vessels still appear to be a good candidate for registration features. Since we have previously [29

29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

] shown that blood vessel ridges can be used successfully to register OFIs, we decided to use them here as the relevant features for a montage algorithm.

It is well known that constructing a montage by simply registering series of individual image pairs together does not provide satisfying results [15

15. H. Y. Shum and R. Szeliski, “Construction of panoramic image mosaics with global and local alignment,” Int. J. Comput. Vis. 36(2), 101–130 (2000). [CrossRef]

]. Global alignment by minimizing the distance between matched features is widely used [15

15. H. Y. Shum and R. Szeliski, “Construction of panoramic image mosaics with global and local alignment,” Int. J. Comput. Vis. 36(2), 101–130 (2000). [CrossRef]

, 20

20. D. E. Becker, A. Can, J. N. Turner, H. L. Tanenbaum, and B. Roysam, “Image processing algorithms for retinal montage synthesis, mapping, and real-time location determination,” IEEE Trans. Biomed. Eng. 45(1), 105–118 (1998). [CrossRef] [PubMed]

, 21

21. A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based technique for joint, linear estimation of high-order image-to-mosaic transformations: Mosaicing the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell. 24(3), 412–419 (2002). [CrossRef]

]. The method by Can et al [21

21. A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based technique for joint, linear estimation of high-order image-to-mosaic transformations: Mosaicing the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell. 24(3), 412–419 (2002). [CrossRef]

] uses features of blood vessel branching points and is likely to have no registration constraints for two images with small overlap. Yang et al. [23

23. G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).

] improved the method by generating new constraints driven by the covariance matrix for image pairs after initial alignment. Sofka et al. [33

33. M. Sofka, Y. Gehua, and C. V. Stewart, “Simultaneous Covariance Driven Correspondence (CDC) and Transformation Estimation in the Expectation Maximization Framework,” Computer Vision and Pattern Recognition, 2007. CVPR '07. IEEE Conference on 1–8 (2007).

] compared the performance of generating point correspondences between the method of Covariance Driven Correspondences (CDC) and Iterative Closest Point (ICP) algorithms. Since the CDC method makes use of gradient information of the registration error map, it has a broader domain of convergence than ICP algorithms; however the number of unknowns increases quadratically with the number of transformation parameters.

We used a montage technique similar to Yang et al. [23

23. G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).

] by adding new constraints for image pairs with small overlap. However, several modifications are made to adapt the montage technique to our particular application. We use a type of ICP algorithm in the global alignment step to avoid extra computation time incurred by the CDC method. Furthermore we update all registration constraints (including the ones for the image pairs first aligned by pairwise registration) at each iteration.

A standard OFI is obtained summing the OCT intensities along the A-scans in the OCT data set. However, it has been shown that filtering, for instance using only the intensities corresponding to pixels close to the retinal pigment epithelium (RPE), can produce images with improved blood vessel contrast [9

9. S. L. Jiao, R. Knighton, X. R. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]

, 11

11. M. Niemeijer, M. K. Garvin, B. V. Ginneken, M. Sonka, and M. D. Abramoff, “Vessel segmentation in 3D spectral OCT scans of the retina,” Proc. SPIE 6914, 69141R (2008).

, 12

12. Z. Hu, M. Niemeijer, M. D. Abràmoft, K. Lee, and M. K. Garvin, “Automated segmentation of 3-D spectral OCT retinal blood vessels by neural canal opening false positive suppression,” Med. Image Comput. Comput. Assist. Interv. 13(Pt 3), 33–40 (2010). [PubMed]

]. We use a similar preprocessing step, based on a quick and rough RPE segmentation, before applying the montage algorithm.

2. Methods

For the purposes of this paper we will assume that we have a number n of partially overlapping OCT data sets I1,…,In of a given eye. The corresponding OFIs give us a set of partially overlapping 2D images O1,…,On. As an example we utilized an imaging protocol consisting of 8 images as shown in Fig. 1
Fig. 1 Protocol for sample image acquisition. This montage involves 8 partially overlapping OCT data sets for a given eye.
.

Our goal is to merge the various OCT data sets into a single large, 3D OCT data set covering an extended retinal region. In order to achieve this we first register every pair of OFIs with sufficient overlap, using blood vessel ridges as described in [29

29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

,34

34. Y. Li, N. Hutchings, R. W. Knighton, G. Gregori, R. J. Lujan, and J. G. Flanagan, “Ridge-branch-based blood vessel detection algorithm for multimodal retinal images,” Proc. SPIE. 7259, 72594K (2009).

]. A global alignment step then creates an initial OFI montage and the process is iterated, updating the registration constraints at every step. Once the OFIs are globally registered, a fairly straightforward process of resampling, interpolation, and cross-correlation can be used to merge the full OCT data sets.

2.1. Preprocessing

It is well known that it is possible to improve the contrast of various features in the OFI by the use of A-scan filters adapted to the retinal anatomy [9

9. S. L. Jiao, R. Knighton, X. R. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]

]. For our purposes we decided to use a quick and simple approach, based on the location of the intensity peaks in the A-scans, to obtain a rough, but satisfactory RPE segmentation. After some smoothing we can select essentially the second (from the top) large peak in the A-scan (with some error correction using a continuity constraint, as well as a spacing constraint between peaks; the details are not essential) as a surrogate RPE surface and use a window of 20 pixels above and 100 pixels below the surface to build the enhanced OFI (Fig. 2
Fig. 2 Comparison between the OFI (sum of all intensities along the A-scans) and the enhanced OFI for two sample data sets. Panel A shows a typical B-scan. The same B-scan appears in B with a superimposed RPE segmentation line. C.1 and D.1 are the OFIs. C.2 and D.2 are the corresponding enhanced OFIs. Notice that our very simple algorithm can results in some artifacts (arrows), but its performance is quite satisfactory for the present purposes.
). In this context it is not crucial for the RPE segmentation to be extremely accurate and speed is a more important factor in the choice of algorithm.

2.2. Registration between individual image pairs

We consider all pairs of OFIs (Oi,Oj) in our sample and select those with sufficient overlap. We will assume that our set of images is such that for each two images Oi and Oj in it there is a connecting chain of images between them, that is a series of images OK1,…,OKm, such that Oi = OK1, Oj = OKm, and each image in the chain has sufficient overlap with the image following it. Typically, an overlap of more than a quarter of the image area was defined to be sufficient overlap [33

33. M. Sofka, Y. Gehua, and C. V. Stewart, “Simultaneous Covariance Driven Correspondence (CDC) and Transformation Estimation in the Expectation Maximization Framework,” Computer Vision and Pattern Recognition, 2007. CVPR '07. IEEE Conference on 1–8 (2007).

]. This requirement can be relaxed somewhat by using for instance some a priori information on the relative location of the images. For instance, when using a precise acquisition protocol like the one described in Fig. 1, we can be more efficient in covering the retina and use an approximately 20% overlap as sufficient.

For all pairs (Oi,Oj) with sufficient overlap we can compute a linear transformation Tj,i that registers Tj,i(Oj) with Oi directly, using the approach described in [29

29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

]. Blood vessel ridges (roughly the center lines of blood vessels) are detected. A ridge pixel in Oi and the ridge pixel in Tj,i(Oj) closest to it form a pair of matching pixels and a similarity function is introduced, based on a specially defined distance between pairs of matching pixels. The registration transformation Tj,i is obtained by minimizing this similarity function. For all other pairs, a (potentially not unique) transformation Tj,i can be defined indirectly using a connecting chain of images between them.

2.3. Global alignment

For simplicity we select O1 as the anchor image: we keep this image fixed and register all the other images to it. Also the transformations Tj,1 will be simply denoted by Tj from now on. Based on the registration parameters Tj (j = 2,…,n) between the other OFIs and the anchor OFI, the transformed OFIs can be pieced together to construct a montage (see Fig. 3
Fig. 3 Global alignment for a simplified montage of three OFIs. (A) A direct registration relationship exists for two image pairs: (Oa, Oc), and (Ob, Oc). (B) Let Oa be the anchor, and construct a montage. (C) Consider every possible combination of overlapping image pairs, and identify matching point pairs for each. New registration constraints are introduced for the image pair (Oa,Ob).
).

We can now examine the overlap between images without sufficient overlap and define new registration constraints for these image pairs. These registration constraints are analogous to the ones used for the direct registration, i.e. a set of matching point pairs is determined on the region of overlap [27

27. J. Zheng, J. Tian, K. Deng, X. Dai, X. Zhang, and M. Xu, “Salient feature region: a new method for retinal image registration,” IEEE Trans. Inf. Technol. Biomed. 15(2), 221–232 (2011). [CrossRef] [PubMed]

]. For a ridge pixel on the reference image, we find the closest pixel on the registered target image; if their distance (determined by the Euclidean distance and the difference between the normal directions of the two pixels along each of their vessel ridge curves [29

29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

]) is less than a given threshold (experimentally determined, about 0.4 degrees of FOV), the pair of pixels is a matching point pair and is included in the updated registration constraints.

We can now compute new transformations Tj using a global minimization of the distance between all the matching point pairs (the ones used in the direct registration and the additional new ones). Let (x,y) be a point in Oj and (X,Y) its image after the transformation Tj. We can write the linear transformation relationship Tj as:

[X,Y,1]=[x,y,1]Tj=[x,y,1][a1,ja2,j0a2,ja1,j0a3,ja4,j1]

If (x,y) and (x',y') are the coordinates of a pair of matching points in Oi and Oj respectively, the global alignment problem involves minimizing the sum of the distances (over all pairs of matching points) between [x,y,1]Ti and [x',y',1]Tj. This is easily set up as a linear least square problem. LetΝdenote the column vector of montage parameters with 4(n-1) parameters, i.e.
Ν =(A2,,An)T.
where Aj = (a1,j,a2,j,a3,j,a4,j), for j = 2,…,n, contains the 4 parameters defining the transformation Tj.

Let M be a 2k by 4(n-1) matrix, b a 2k column vector, where k is the total number of matching point pairs, and (x,y), (x',y'), be the coordinates, respectively in Oi and Oj, of the m-th pair of matching points. The (m-1)-th and m-th rows of Mare given by the following matrix:
(0,x,y,1,0,0,x',y',1,0,00,y,x,0,1,0,y',x',0,1,0)
where the only non zero blocks start at the (4i-3)-th and (4j-3)-th position. The corresponding entries in bare 0s. (The entries in M and b are slightly different when one of the indexes i, j is equal to 1. We leave the easy adjustments to the reader). The parameters for the global montage Νcan be obtained as the least square solution to the linear system

MN=b

This global alignment is iterated several times (3 in our experimental setting) updating the registration constraints after each iteration. This iterative procedure is a kind of iterative closest point (ICP) algorithm.

2.4. 3D registration

Once we have obtained a montage of the OFIs, we can use it to piece together the full OCT data sets using a procedure based on resampling, interpolation, and cross-correlation. Let x and y be coordinates in the en face plane, defined by the axes of the anchor OFI O1 (with x along the instrument scanning direction), while z denotes the coordinate in the axial direction. Let R be a rectangle in the x-y plane covering the entire OFI montage and with sides parallel to the x and y axes. We can then define a set of n+1 (where n is the number of OCT data sets) connected, non overlapping regions Rj such that Rj is contained in TjOj for j = 1,...,n and R1Rn+1 = R. A new global sampling grid sij (where the second index runs along the x direction) over R is generated by extending homogeneously the orthogonal sampling grid of the anchor image. The montage OCT is built by defining an A-scan for each point in the grid sij.

Let m1, m2 be the dimensions of the grid over R. For each given value of i, 1≤im1, let Bi be the B-scan over sij, for j = 1,…,m2. For the sake of simplicity we can assume that Bi only intersects R1 and R2, i.e. sijR1 for j = 1,…,m and sijR2 for j = m+1,…,m2. Then essentially the first m A-scans in Bi can be obtained using a weighted closest point interpolation of the A-scans in I1 and the rest by a weighted closest point interpolation of the A-scans in I2. In order to have the A-scans match smoothly across the transition some consistency constraints are needed. We define a transition region (a few grid points around sim which are contained both in T1O1 and T2O2) and at each point in this region we use cross-correlation to find the optimal shift that matches the A-scan in I2 with the corresponding A-scan in I1. A linear fit of the shifts in the transitional region is used to compute shifts over sij for j=m+1,…,m2. This approach can be extended in a straightforward manner to the case when a B-scan intersects several or the sets in the partition of R and, correspondingly, multiple transitional regions.

3. Experimental Results

We tested the algorithm in this paper using an imaging protocol that consisted of 8 partially overlapping OCT data sets as shown in Fig. 1. For each eye, the 8 OCT data sets were acquired in a single imaging session, using a Cirrus HD-OCT (Carl Zeiss Meditec Inc, Dublin, California). Each data set consisted of a raster scan covering a 6 X 6 mm area with 200 X 200 A-scans. The scans O1 and O2 are centered at the fovea and the optic disc respectively (or vice versa, depending whether it was a right or left eye). The total retinal area covered is approximately 50x35 degrees. Approval for the collection and analysis of SDOCT images was obtained from the Institutional Review Board at the University of Miami Miller School of Medicine and all participants signed research informed consent.

We analyzed the sets of OCT images of 3 normal eyes and 3 eyes with retinal degeneration. Figure 4
Fig. 4 Example montage results of enhanced OFIs for one normal eye (left) and one eye with retinitis pigmentosa (right).
shows the results of the montage algorithm on the OFIs for one normal eye and eye with retinitis pigmentosa, while Fig. 5
Fig. 5 OFI montage (left) and a B-scan (right) from the corresponding OCT montage. The position of the B-scan is shown by the red horizontal line. The width of the transition regions is a single pixel.
shows a B-scan in the OCT montage of a normal eye.

Registration errors are measured by calculating the average root mean square error (RMSE) of manually labeled matching point pairs. Matching point pairs are normally blood vessel branching points, and we manually label all point pairs that can be identified, typically several point pairs for each image pair (see [29

29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

]). Using our imaging protocol we have 13 (non trivially) overlapping image pairs. Table 1

Table 1. Averages and standard deviations (within parentheses) of RMSE values (in pixels) for the different image pairs as well as the whole montage. For our data sets a pixel corresponds to 30 µm.

table-icon
View This Table
shows the means and standard deviations of RMSE values for the individual image pairs, as well as for the whole montage.

4. Discussion and Conclusion

The core of our montage algorithm is a 2D montage algorithm for OFIs. The 2D montage algorithm is similar to the one in Yang et al. [23

23. G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).

], which addresses the montage of traditional retinal images such as color fundus photographs. For the specific application to OFIs, the choice of registration feature can be somewhat problematic. Blood vessel ridges are selected here as registration feature due to their relative stability [29

29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

]. However, standard OFIs can sometime show (areas of) poor blood vessel contrast which, when used in the montage algorithm, can lead to incorrect results. In order to overcome this problem, we use a simple and efficient approximate RPE boundary detection, and enhance the OFIs by filtering out the OCT intensities away from the segmentation. Of course a more sophisticated RPE segmentation algorithm [35

35. T. Fabritius, S. Makita, M. Miura, R. Myllylä, and Y. Yasuno, “Automated segmentation of the macula by optical coherence tomography,” Opt. Express 17(18), 15659–15669 (2009). [CrossRef] [PubMed]

37

37. Q. Yang, C. A. Reisman, Z. G. Wang, Y. Fukuma, M. Hangai, N. Yoshimura, A. Tomidokoro, M. Araie, A. S. Raza, D. C. Hood, and K. P. Chan, “Automated layer segmentation of macular OCT images using dual-scale gradient information,” Opt. Express 18(20), 21293–21307 (2010). [CrossRef] [PubMed]

], and/or filtering approach, could be used here, but we did not want to introduce unnecessary complexities and, although our segmentation algorithm is very basic, the resulting OFIs show a sufficient improvement in contrast for our purposes.

Yang et al. [23

23. G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).

] did not update the constraints for the image pairs that were registered directly (step 2.2), because they claimed that the pairwise constraints are sufficient for accurate results in the joint alignment step. This does not appear to be the case for the application to OFIs. Hence, we update registration constraints for every image pair with overlap. In addition, the ICP is used instead of the CDC in [23

23. G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).

], to save computation time. In practice information about the approximate location of an OCT data set with respect to fixed retinal landmarks is stored during image acquisition and could be available to the montage algorithm, therefore the improvement in convergence domain associated with CDC was not considered a crucial feature.

Eye movements during acquisition are an important factor producing artifacts/distortions in OCT data sets. These types of artifacts are often easily recognizable in the OFIs and the affected data sets can be discarded by a skilled operator during the imaging session. In principle it would be possible to split a data sets marred by eye-motion artifacts in several artifact free pieces and (as long as the single pieces contained enough recognizable features) use these pieces as input for the montage algorithm.

When building the full OCT montage (step 2.4) there are typically several different possible choices for the partition R1Rn + 1 = R. Different choices of partition sets will result in slightly different OCT montages. Normally these differences are non essential (equivalent to taking two different OCT images of the same eye), but there are situations when the choice of one OCT image has definite advantages over its competitors on a region of overlap, i.e. better image quality, absence of artifacts, etc. In these cases it can make sense to pursue an optimal choice of partition. One possible way to accomplish it is by defining weights over (regions of) each individual OFI. These weight could incorporate both objective information (image quality scores, local OCT intensity values) and subjective judgment from the operator.

During image acquisition, factors like the beam entrance point in the pupil and the location of the fixation target can influence considerably the retinal tilt in the imaging cube. It is then reasonable to be concerned about possible higher order discontinuities (kinks in the retinal geometry) in the OCT montage across the boundaries of the partition sets. Our definition of transition regions is meant to perform a first order correction for this potential issue by introducing an additional linear adjustment of the imaging plane in each individual OCT data set. In practice though it turns out that it is almost always sufficient (at least in our experience, operator’s skill and other factors may play a role) to consider transition regions of width equal to a single pixel, corresponding to a zero order correction of the imaging plane (i.e. a simple uniform shift in the z-direction across each OCT data set).

The ability to montage OCT data sets implies the possibility to piece together any kind of information about the underlying OCT data sets. For instance it is certainly possible to extend segmentations maps of each OCT data sets to the montage. Figure 6
Fig. 6 Retinal thickness maps montage for a normal eye.
shows the retinal thickness maps for the OCT montage of a normal eye obtained from the individual retinal thickness maps.

The algorithm was realized in MATLAB. No effort was made to optimize for speed. Given a set of 8 OCT data sets, as described in Fig. 1, the computation time for the montage was approximately 8 minutes.

Acknowledgments

Supported by NIH Center Core Grant P30EY014801, Research to Prevent Blindness Unrestricted Grant, Department of Defense (DOD-Grant#W81XWH-09-1-0675), Carl Zeiss Meditec.

References and links

1.

Z. Yehoshua, P. J. Rosenfeld, G. Gregori, and F. Penha, “Spectral domain optical coherence tomography imaging of dry age-related macular degeneration,” Ophthalmic Surg. Lasers Imaging 41(6Suppl), S6–S14 (2010). [CrossRef] [PubMed]

2.

B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010). [CrossRef] [PubMed]

3.

B. Povazay, B. Hermann, B. Hofer, V. Kajić, E. Simpson, T. Bridgford, and W. Drexler, “Wide-field optical coherence tomography of the choroid in vivo,” Invest. Ophthalmol. Vis. Sci. 50(4), 1856–1863 (2008). [CrossRef] [PubMed]

4.

T. Klein, W. Wieser, C. M. Eigenwillig, B. R. Biedermann, and R. Huber, “Megahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laser,” Opt. Express 19(4), 3044–3062 (2011). [CrossRef] [PubMed]

5.

M. Emmenlauer, O. Ronneberger, A. Ponti, P. Schwarb, A. Griffa, A. Filippi, R. Nitschke, W. Driever, and H. Burkhardt, “XuvTools: free, fast and reliable stitching of large 3D datasets,” J. Microsc. 233(1), 42–60 (2009). [CrossRef] [PubMed]

6.

S. Preibisch, S. Saalfeld, and P. Tomancak, “Globally optimal stitching of tiled 3D microscopic image acquisitions,” Bioinformatics 25(11), 1463–1465 (2009). [CrossRef] [PubMed]

7.

Y. Yu and H. Peng, “Automated high speed stitching of large 3D microscopic images,” Proc. of IEEE 2011 International Symposium on Biomedical Imaging: From Nano to Macro 238–241 (2011).

8.

J. Yoo, I. V. Larina, K. V. Larin, M. E. Dickinson, and M. Liebling, “Increasing the field-of-view of dynamic cardiac OCT via post-acquisition mosaicing without affecting frame-rate or spatial resolution,” Biomed. Opt. Express 2(9), 2614–2622 (2011). [CrossRef] [PubMed]

9.

S. L. Jiao, R. Knighton, X. R. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express 13(2), 444–452 (2005). [CrossRef] [PubMed]

10.

M. Wojtkowski, T. Bajraszewski, I. Gorczyńska, P. Targowski, A. Kowalczyk, W. Wasilewski, and C. Radzewicz, “Ophthalmic imaging by spectral optical coherence tomography,” Am. J. Ophthalmol. 138(3), 412–419 (2004). [CrossRef] [PubMed]

11.

M. Niemeijer, M. K. Garvin, B. V. Ginneken, M. Sonka, and M. D. Abramoff, “Vessel segmentation in 3D spectral OCT scans of the retina,” Proc. SPIE 6914, 69141R (2008).

12.

Z. Hu, M. Niemeijer, M. D. Abràmoft, K. Lee, and M. K. Garvin, “Automated segmentation of 3-D spectral OCT retinal blood vessels by neural canal opening false positive suppression,” Med. Image Comput. Comput. Assist. Interv. 13(Pt 3), 33–40 (2010). [PubMed]

13.

R. Szeliski, “Image alignment and stitching: A tutorial,” MSR-TR-2004–92, Microsoft Research (2004).

14.

X. Fang, B. Luo, H. Zhao, J. Tang, and S. Zhai, “New multi-resolution image stitching with local and global alignment,” IET Comput. Vision 4(4), 231–246 (2010). [CrossRef]

15.

H. Y. Shum and R. Szeliski, “Construction of panoramic image mosaics with global and local alignment,” Int. J. Comput. Vis. 36(2), 101–130 (2000). [CrossRef]

16.

D. L. Milgram, “Computer Methods for Creating Photomosaics,” IEEE Trans. Comput. C-24(11), 1113–1119 (1975). [CrossRef]

17.

G. H. Yang, C. V. Stewart, M. Sofka, and C. L. Tsai, “Registration of challenging image pairs: initialization, estimation, and decision,” IEEE Trans. Pattern Anal. Mach. Intell. 29(11), 1973–1989 (2007). [CrossRef] [PubMed]

18.

A. A. Mahurkar, M. A. Vivino, B. L. Trus, E. M. Kuehl, M. B. Datiles 3rd, and M. I. Kaiser-Kupfer, “Constructing retinal fundus photomontages. A new computer-based method,” Invest. Ophthalmol. Vis. Sci. 37(8), 1675–1683 (1996). [PubMed]

19.

P. C. Cattin, H. Bay, L. Van Gool, and G. Szekely, “Retina mosaicing using local features,” Med. Image Comput. Comput. Assist. Interv. 4191, 185–192 (2006).

20.

D. E. Becker, A. Can, J. N. Turner, H. L. Tanenbaum, and B. Roysam, “Image processing algorithms for retinal montage synthesis, mapping, and real-time location determination,” IEEE Trans. Biomed. Eng. 45(1), 105–118 (1998). [CrossRef] [PubMed]

21.

A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based technique for joint, linear estimation of high-order image-to-mosaic transformations: Mosaicing the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell. 24(3), 412–419 (2002). [CrossRef]

22.

S. Lee, M. D. Abramoff, and J. M. Reinhardt, “Retinal image mosaicing using the radial distortion correction model,” Proc. SPIE Medical Imaging 6914, 91435 (2008).

23.

G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).

24.

T. E. Choe, I. Cohen, M. Lee, and G. Medioni, “Optimal global mosaic generation from retinal images,” Proceedings of 18th International Conference on Pattern Recognition 3, 681–684 (2006).

25.

W. Aguilar, M. E. Martinez-Perez, Y. Frauel, F. Escolano, M. A. Lozano, and A. Espinosa-Romero, “Graph-based methods for retinal mosaicing and vascular characterization,” Lect. Notes Comput. Sci. 4538, 25–36 (2007). [CrossRef]

26.

D. G. Lowe, “Distinctive image features from scale-invariant keypoints,” Int. J. Comput. Vis. 60(2), 91–110 (2004). [CrossRef]

27.

J. Zheng, J. Tian, K. Deng, X. Dai, X. Zhang, and M. Xu, “Salient feature region: a new method for retinal image registration,” IEEE Trans. Inf. Technol. Biomed. 15(2), 221–232 (2011). [CrossRef] [PubMed]

28.

J. M. Schmitt, S. H. Xiang, and K. M. Yung, “Speckle in optical coherence tomography,” J. Biomed. Opt. 4(1), 95–105 (1999). [CrossRef]

29.

Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express 19(1), 7–16 (2011). [CrossRef] [PubMed]

30.

A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based, robust, hierarchical algorithm for registering pairs of images of the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell. 24(3), 347–364 (2002). [CrossRef]

31.

T. Chanwimaluang, G. L. Fan, and S. R. Fransen, “Hybrid retinal image registration,” IEEE Trans. Inf. Technol. Biomed. 10(1), 129–142 (2006). [CrossRef] [PubMed]

32.

S. Lee, M. D. Abramoff, and J. M. Reinhardt, “Retinal image mosaicing using the radial distortion correction model - art. no. 691435,” Proc. SPIE 6914, 91435 (2008).

33.

M. Sofka, Y. Gehua, and C. V. Stewart, “Simultaneous Covariance Driven Correspondence (CDC) and Transformation Estimation in the Expectation Maximization Framework,” Computer Vision and Pattern Recognition, 2007. CVPR '07. IEEE Conference on 1–8 (2007).

34.

Y. Li, N. Hutchings, R. W. Knighton, G. Gregori, R. J. Lujan, and J. G. Flanagan, “Ridge-branch-based blood vessel detection algorithm for multimodal retinal images,” Proc. SPIE. 7259, 72594K (2009).

35.

T. Fabritius, S. Makita, M. Miura, R. Myllylä, and Y. Yasuno, “Automated segmentation of the macula by optical coherence tomography,” Opt. Express 17(18), 15659–15669 (2009). [CrossRef] [PubMed]

36.

G. Gregori, F. H. Wang, P. J. Rosenfeld, Z. Yehoshua, N. Z. Gregori, B. J. Lujan, C. A. Puliafito, and W. J. Feuer, “Spectral domain optical coherence tomography imaging of drusen in nonexudative age-related macular degeneration,” Ophthalmology 118(7), 1373–1379 (2011). [PubMed]

37.

Q. Yang, C. A. Reisman, Z. G. Wang, Y. Fukuma, M. Hangai, N. Yoshimura, A. Tomidokoro, M. Araie, A. S. Raza, D. C. Hood, and K. P. Chan, “Automated layer segmentation of macular OCT images using dual-scale gradient information,” Opt. Express 18(20), 21293–21307 (2010). [CrossRef] [PubMed]

OCIS Codes
(100.0100) Image processing : Image processing
(110.4500) Imaging systems : Optical coherence tomography
(170.4460) Medical optics and biotechnology : Ophthalmic optics and devices
(170.5755) Medical optics and biotechnology : Retina scanning

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: September 19, 2011
Revised Manuscript: November 17, 2011
Manuscript Accepted: November 20, 2011
Published: December 8, 2011

Virtual Issues
Vol. 7, Iss. 2 Virtual Journal for Biomedical Optics

Citation
Ying Li, Giovanni Gregori, Byron L. Lam, and Philip J. Rosenfeld, "Automatic montage of SD-OCT data sets," Opt. Express 19, 26239-26248 (2011)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-19-27-26239


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. Z. Yehoshua, P. J. Rosenfeld, G. Gregori, and F. Penha, “Spectral domain optical coherence tomography imaging of dry age-related macular degeneration,” Ophthalmic Surg. Lasers Imaging41(6Suppl), S6–S14 (2010). [CrossRef] [PubMed]
  2. B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker, and J. G. Fujimoto, “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express18(19), 20029–20048 (2010). [CrossRef] [PubMed]
  3. B. Povazay, B. Hermann, B. Hofer, V. Kajić, E. Simpson, T. Bridgford, and W. Drexler, “Wide-field optical coherence tomography of the choroid in vivo,” Invest. Ophthalmol. Vis. Sci.50(4), 1856–1863 (2008). [CrossRef] [PubMed]
  4. T. Klein, W. Wieser, C. M. Eigenwillig, B. R. Biedermann, and R. Huber, “Megahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laser,” Opt. Express19(4), 3044–3062 (2011). [CrossRef] [PubMed]
  5. M. Emmenlauer, O. Ronneberger, A. Ponti, P. Schwarb, A. Griffa, A. Filippi, R. Nitschke, W. Driever, and H. Burkhardt, “XuvTools: free, fast and reliable stitching of large 3D datasets,” J. Microsc.233(1), 42–60 (2009). [CrossRef] [PubMed]
  6. S. Preibisch, S. Saalfeld, and P. Tomancak, “Globally optimal stitching of tiled 3D microscopic image acquisitions,” Bioinformatics25(11), 1463–1465 (2009). [CrossRef] [PubMed]
  7. Y. Yu and H. Peng, “Automated high speed stitching of large 3D microscopic images,” Proc. of IEEE 2011 International Symposium on Biomedical Imaging: From Nano to Macro 238–241 (2011).
  8. J. Yoo, I. V. Larina, K. V. Larin, M. E. Dickinson, and M. Liebling, “Increasing the field-of-view of dynamic cardiac OCT via post-acquisition mosaicing without affecting frame-rate or spatial resolution,” Biomed. Opt. Express2(9), 2614–2622 (2011). [CrossRef] [PubMed]
  9. S. L. Jiao, R. Knighton, X. R. Huang, G. Gregori, and C. A. Puliafito, “Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography,” Opt. Express13(2), 444–452 (2005). [CrossRef] [PubMed]
  10. M. Wojtkowski, T. Bajraszewski, I. Gorczyńska, P. Targowski, A. Kowalczyk, W. Wasilewski, and C. Radzewicz, “Ophthalmic imaging by spectral optical coherence tomography,” Am. J. Ophthalmol.138(3), 412–419 (2004). [CrossRef] [PubMed]
  11. M. Niemeijer, M. K. Garvin, B. V. Ginneken, M. Sonka, and M. D. Abramoff, “Vessel segmentation in 3D spectral OCT scans of the retina,” Proc. SPIE6914, 69141R (2008).
  12. Z. Hu, M. Niemeijer, M. D. Abràmoft, K. Lee, and M. K. Garvin, “Automated segmentation of 3-D spectral OCT retinal blood vessels by neural canal opening false positive suppression,” Med. Image Comput. Comput. Assist. Interv.13(Pt 3), 33–40 (2010). [PubMed]
  13. R. Szeliski, “Image alignment and stitching: A tutorial,” MSR-TR-2004–92, Microsoft Research (2004).
  14. X. Fang, B. Luo, H. Zhao, J. Tang, and S. Zhai, “New multi-resolution image stitching with local and global alignment,” IET Comput. Vision4(4), 231–246 (2010). [CrossRef]
  15. H. Y. Shum and R. Szeliski, “Construction of panoramic image mosaics with global and local alignment,” Int. J. Comput. Vis.36(2), 101–130 (2000). [CrossRef]
  16. D. L. Milgram, “Computer Methods for Creating Photomosaics,” IEEE Trans. Comput.C-24(11), 1113–1119 (1975). [CrossRef]
  17. G. H. Yang, C. V. Stewart, M. Sofka, and C. L. Tsai, “Registration of challenging image pairs: initialization, estimation, and decision,” IEEE Trans. Pattern Anal. Mach. Intell.29(11), 1973–1989 (2007). [CrossRef] [PubMed]
  18. A. A. Mahurkar, M. A. Vivino, B. L. Trus, E. M. Kuehl, M. B. Datiles, and M. I. Kaiser-Kupfer, “Constructing retinal fundus photomontages. A new computer-based method,” Invest. Ophthalmol. Vis. Sci.37(8), 1675–1683 (1996). [PubMed]
  19. P. C. Cattin, H. Bay, L. Van Gool, and G. Szekely, “Retina mosaicing using local features,” Med. Image Comput. Comput. Assist. Interv.4191, 185–192 (2006).
  20. D. E. Becker, A. Can, J. N. Turner, H. L. Tanenbaum, and B. Roysam, “Image processing algorithms for retinal montage synthesis, mapping, and real-time location determination,” IEEE Trans. Biomed. Eng.45(1), 105–118 (1998). [CrossRef] [PubMed]
  21. A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based technique for joint, linear estimation of high-order image-to-mosaic transformations: Mosaicing the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell.24(3), 412–419 (2002). [CrossRef]
  22. S. Lee, M. D. Abramoff, and J. M. Reinhardt, “Retinal image mosaicing using the radial distortion correction model,” Proc. SPIE Medical Imaging 6914, 91435 (2008).
  23. G. H. Yang and C. V. Stewart, “Covariance-driven mosaic formation from sparsely-overlapping image sets with application to retinal image mosaicing,” Proceedings of the 2004 IEEE Computer Society Conference on Computer Vision and Pattern Recognition 1, 804–810 (2004).
  24. T. E. Choe, I. Cohen, M. Lee, and G. Medioni, “Optimal global mosaic generation from retinal images,” Proceedings of 18th International Conference on Pattern Recognition 3, 681–684 (2006).
  25. W. Aguilar, M. E. Martinez-Perez, Y. Frauel, F. Escolano, M. A. Lozano, and A. Espinosa-Romero, “Graph-based methods for retinal mosaicing and vascular characterization,” Lect. Notes Comput. Sci.4538, 25–36 (2007). [CrossRef]
  26. D. G. Lowe, “Distinctive image features from scale-invariant keypoints,” Int. J. Comput. Vis.60(2), 91–110 (2004). [CrossRef]
  27. J. Zheng, J. Tian, K. Deng, X. Dai, X. Zhang, and M. Xu, “Salient feature region: a new method for retinal image registration,” IEEE Trans. Inf. Technol. Biomed.15(2), 221–232 (2011). [CrossRef] [PubMed]
  28. J. M. Schmitt, S. H. Xiang, and K. M. Yung, “Speckle in optical coherence tomography,” J. Biomed. Opt.4(1), 95–105 (1999). [CrossRef]
  29. Y. Li, G. Gregori, R. W. Knighton, B. J. Lujan, and P. J. Rosenfeld, “Registration of OCT fundus images with color fundus photographs based on blood vessel ridges,” Opt. Express19(1), 7–16 (2011). [CrossRef] [PubMed]
  30. A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based, robust, hierarchical algorithm for registering pairs of images of the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell.24(3), 347–364 (2002). [CrossRef]
  31. T. Chanwimaluang, G. L. Fan, and S. R. Fransen, “Hybrid retinal image registration,” IEEE Trans. Inf. Technol. Biomed.10(1), 129–142 (2006). [CrossRef] [PubMed]
  32. S. Lee, M. D. Abramoff, and J. M. Reinhardt, “Retinal image mosaicing using the radial distortion correction model - art. no. 691435,” Proc. SPIE 6914, 91435 (2008).
  33. M. Sofka, Y. Gehua, and C. V. Stewart, “Simultaneous Covariance Driven Correspondence (CDC) and Transformation Estimation in the Expectation Maximization Framework,” Computer Vision and Pattern Recognition, 2007. CVPR '07. IEEE Conference on 1–8 (2007).
  34. Y. Li, N. Hutchings, R. W. Knighton, G. Gregori, R. J. Lujan, and J. G. Flanagan, “Ridge-branch-based blood vessel detection algorithm for multimodal retinal images,” Proc. SPIE.7259, 72594K (2009).
  35. T. Fabritius, S. Makita, M. Miura, R. Myllylä, and Y. Yasuno, “Automated segmentation of the macula by optical coherence tomography,” Opt. Express17(18), 15659–15669 (2009). [CrossRef] [PubMed]
  36. G. Gregori, F. H. Wang, P. J. Rosenfeld, Z. Yehoshua, N. Z. Gregori, B. J. Lujan, C. A. Puliafito, and W. J. Feuer, “Spectral domain optical coherence tomography imaging of drusen in nonexudative age-related macular degeneration,” Ophthalmology118(7), 1373–1379 (2011). [PubMed]
  37. Q. Yang, C. A. Reisman, Z. G. Wang, Y. Fukuma, M. Hangai, N. Yoshimura, A. Tomidokoro, M. Araie, A. S. Raza, D. C. Hood, and K. P. Chan, “Automated layer segmentation of macular OCT images using dual-scale gradient information,” Opt. Express18(20), 21293–21307 (2010). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited