Electron beam lithography designed silver nano-disks used as label free nano-biosensors based on localized surface plasmon resonance

Neval A. Cinel, Serkan Bütün, and Ekmel Özbay  »View Author Affiliations


Optics Express, Vol. 20, Issue 3, pp. 2587-2597 (2012)
http://dx.doi.org/10.1364/OE.20.002587


View Full Text Article

Acrobat PDF (1323 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We present a label-free, optical nano-biosensor based on the Localized Surface Plasmon Resonance (LSPR) that is observed at the metal-dielectric interface of silver nano-disk arrays located periodically on a sapphire substrate by Electron-Beam Lithography (EBL). The nano-disk array was designed by finite-difference and time-domain (FDTD) algorithm-based simulations. Refractive index sensitivity was calculated experimentally as 221-354 nm/RIU for different sized arrays. The sensing mechanism was first tested with a biotin-avidin pair, and then a preliminary trial for sensing heat-killed Escherichia coli (E. coli) O157:H7 bacteria was done. Although the study is at an early stage, the results indicate that such a plasmonic structure can be applied to bio-sensing applications and then extended to the detection of specific bacteria species as a fast and low cost alternative.

© 2012 OSA

1. Introduction

The need for rapid, specific, sensitive, inexpensive, in-field and real-time detection of target analytes in solutions with an unknown content has made “biosensors” gain importance especially in fields such as environmental monitoring, pollutant detection, medical diagnostics, and biological warfare defense. A subcategory that biosensors are used is the detection of food-borne pathogens. Food borne pathogens cause infectious or toxic diseases upon the consumption of contaminated water and/or food. Escherichia coli (E.Coli O157:H7) is among the most serious food borne infections that also causes severe complications and sometimes fatal health problems, particularly among infants, children and the elderly.

There are several classical techniques to detect the presence of bacteria. Traditional methods like culture collection are effective but tests may take days to produce results and have limited ability in the discrimination of various bacterial strains [1

1. Anonymous, “ISO 9308-1,” (ISO, Geneva, Switzerland, 2000).

]. There are newer, more rapid methods such as polymerase chain reaction (PCR) and Enzyme linked immunosorbent assay (ELISA) [2

2. P. Daly, T. Collier, and S. Doyle, “PCR-ELISA detection of Escherichia coli in milk,” Lett. Appl. Microbiol. 34(3), 222–226 (2002). [CrossRef] [PubMed]

4

4. R. P. Johnson, R. J. Durham, S. T. Johnson, L. A. MacDonald, S. R. Jeffrey, and B. T. Butman, “Detection of Escherichia coli O157:H7 in meat by an enzyme-linked immunosorbent assay, EHEC-Tek,” Appl. Environ. Microbiol. 61(1), 386–388 (1995). [PubMed]

]. However, these methods require extensive pre-treatment prior to analysis, such as an enrichment step that involves growing bacteria on a nutrient agar to detectable levels; equipped laboratory and highly trained personnel who can operate the instruments and interpret the results.

These difficulties faced in classical detection methods have led to the development of different types of biosensors each relying on a different transduction mechanism. There are numerous optical [5

5. M. A. Cooper, “Optical biosensors in drug discovery,” Nat. Rev. Drug Discov. 1(7), 515–528 (2002). [CrossRef] [PubMed]

, 6

6. J. Homola, S. S. Yee, and G. Gauglitz, “Surface plasmon resonance sensors: review,” Sens. Actuators B Chem. 54(1-2), 3–15 (1999). [CrossRef]

], amperometric [7

7. X. Y. Liu, X. D. Zeng, N. N. Mai, Y. Liu, B. Kong, Y. H. Li, W. Z. Wei, and S. L. Luo, “Amperometric glucose biosensor with remarkable acid stability based on glucose oxidase entrapped in colloidal gold-modified carbon ionic liquid electrode,” Biosens. Bioelectron. 25(12), 2675–2679 (2010). [CrossRef] [PubMed]

, 8

8. I. Vostiar, J. Tkac, E. Sturdik, and P. Gemeiner, “Amperometric urea biosensor based on urease and electropolymerized toluidine blue dye as a pH-sensitive redox probe,” Bioelectrochemistry 56(1-2), 113–115 (2002). [CrossRef] [PubMed]

], impedometric [9

9. A. Malave, M. Tewes, T. Gronewold, and M. Lohndorf, “Development of impedance biosensors with nanometer gaps for marker-free analytical measurements,” Microelectron. Eng. 78–79, 587–592 (2005). [CrossRef]

], electro chemical [10

10. W. Zhao, J. J. Xu, and H. Y. Chen, “Electrochemical biosensors based on layer-by-layer assemblies,” Electroanalysis 18(18), 1737–1748 (2006). [CrossRef]

], and magnetic [11

11. M. H. Meyer, M. Stehr, S. Bhuju, H. J. Krause, M. Hartmann, P. Miethe, M. Singh, and M. Keusgen, “Magnetic biosensor for the detection of Yersinia pestis,” J. Microbiol. Methods 68(2), 218–224 (2007). [CrossRef] [PubMed]

, 12

12. J. Llandro, J. J. Palfreyman, A. Ionescu, and C. H. Barnes, “Magnetic biosensor technologies for medical applications: a review,” Med. Biol. Eng. Comput. 48(10), 977–998 (2010).

] biosensors described in the literature. These sensors are designed for tracking interactions at the classical physics limit, such as measuring mass, conductivity, pH, ionic mobility, and resonant RF frequency. However, optical sensors offer advantages such as immunity to electromagnetic interference, high sensitivity, and a fast response. They are capable of performing remote sensing and can provide multiplexed detection within a single device [13

13. X. Fan, I. M. White, S. I. Shopova, H. Zhu, J. D. Suter, and Y. Sun, “Sensitive optical biosensors for unlabeled targets: a review,” Anal. Chim. Acta 620(1-2), 8–26 (2008). [CrossRef] [PubMed]

] and, therefore, are preferable.

When compared to other optical Propagating Surface Plasmon Resonance (PSPR) [6

6. J. Homola, S. S. Yee, and G. Gauglitz, “Surface plasmon resonance sensors: review,” Sens. Actuators B Chem. 54(1-2), 3–15 (1999). [CrossRef]

] or Surface Enhanced Raman Spectroscopy (SERS) [14

14. K. E. Shafer-Peltier, C. L. Haynes, M. R. Glucksberg, and R. P. Van Duyne, “Toward a glucose biosensor based on surface-enhanced Raman scattering,” J. Am. Chem. Soc. 125(2), 588–593 (2003). [CrossRef] [PubMed]

16

16. F. S. Ligler and C. A. R. Taitt, Optical Biosensors: Present and Future, 1st ed. (Elsevier, 2002), pp. viii, 607 p.

] sensors, Localized Surface Plasmon Resonance (LSPR) systems are easy to manufacture and less expensive, in addition to being portable and practical. Working with small samples of analytes is possible and no special geometry (no specific angular conditions of excitation, no needs of prism coupler-based, grating coupler-based or optical waveguide) for detection is needed when compared with PSPR sensors.

PSPR and LSPR sensors both rely on detecting small changes in refractive index in the vicinity of a noble metal’s surface. Their sensitivities are caused by different mechanisms, but their overall sensitivities are approximately equivalent. PSPR sensors exhibit large refractive index sensitivities (2 × 106 nm RIU–1) [17

17. L. S. Jung, C. T. Campbell, T. M. Chinowsky, M. N. Mar, and S. S. Yee, “Quantitative interpretation of the response of surface plasmon resonance sensors to adsorbed films,” Langmuir 14(19), 5636–5648 (1998). [CrossRef]

, 18

18. A. J. Haes and R. P. Van Duyne, “A unified view of propagating and localized surface plasmon resonance biosensors,” Anal. Bioanal. Chem. 379(7-8), 920–930 (2004). [CrossRef] [PubMed]

]. The LSPR nano sensor, on the other hand, has modest refractive index sensitivity (2 × 102 nm RIU–1) [18

18. A. J. Haes and R. P. Van Duyne, “A unified view of propagating and localized surface plasmon resonance biosensors,” Anal. Bioanal. Chem. 379(7-8), 920–930 (2004). [CrossRef] [PubMed]

, 19

19. M. D. Malinsky, K. L. Kelly, G. C. Schatz, and R. P. Van Duyne, “Chain length dependence and sensing capabilities of the localized surface plasmon resonance of silver nanoparticles chemically modified with alkanethiol self-assembled monolayers,” J. Am. Chem. Soc. 123(7), 1471–1482 (2001). [CrossRef]

]. However, the LSPR sensors’ enhanced sensitivity is due to the short and tunable characteristic electromagnetic field decay length that is on the order of 5-15 nm, whereas it is on the order of 200-300 nm for PSPR sensors [17

17. L. S. Jung, C. T. Campbell, T. M. Chinowsky, M. N. Mar, and S. S. Yee, “Quantitative interpretation of the response of surface plasmon resonance sensors to adsorbed films,” Langmuir 14(19), 5636–5648 (1998). [CrossRef]

, 18

18. A. J. Haes and R. P. Van Duyne, “A unified view of propagating and localized surface plasmon resonance biosensors,” Anal. Bioanal. Chem. 379(7-8), 920–930 (2004). [CrossRef] [PubMed]

]. In addition, in PSPR systems, there is strong environmental temperature dependency due to the large refractive index sensitivity, whereas LSPR sensors do not require temperature control.

Moreover, PSPR sensors require at least a 10 × 10 μm2 area for sensing. However, in LSPR sensing, confocal or near-field measurement techniques help minimize the area to a large number of individual sensing elements and even up to a single nanoparticle (NP). Finally, a UV-visible spectrometer is sufficient to obtain the extinction spectra, so that no complex equipment is necessary [20

20. J. C. Riboh, A. J. Haes, A. D. McFarland, C. Ranjit Yonzon, and R. P. Van Duyne, “A nanoscale optical biosensor: Real-time immunoassay in physiological buffer enabled by improved nanoparticle adhesion,” J. Phys. Chem. B 107(8), 1772–1780 (2003). [CrossRef]

].

LSPR is highly dependent on the size, shape, and period of the nanoparticles as well as the uniformity of the resultant structure. Chemical patterning methods such as Nano sphere lithography (NSL) are usually inexpensive and parallel-natured processes with high throughput. However, only limited shapes and arrays with hexagonal symmetry are feasible. Post deposition steps like thermal annealing or multi angle deposition are necessary for modification in shapes of nanostructures. Moreover, uniform areas are difficult to obtain [27

27. F. Y. Lee, K. H. Fung, T. L. Tang, W. Y. Tam, and C. T. Chan, “Fabrication of gold nano-particle arrays using two-dimensional templates from holographic lithography,” Curr. Appl. Phys. 9(4), 820–825 (2009). [CrossRef]

].

EBL, on the other hand, is a modified SEM system that has superior properties compared to other optical or chemical patterning techniques, such as the precise placement and design of arbitrarily shaped NPs with a large selection of geometries of various sizes with fine features. It offers a high resolution of around 10 nm since it is not limited by diffraction limits unlike other optical lithography methods [28

28. A. Hohenau, H. Ditlbacher, B. Lamprecht, J. R. Krenn, A. Leitner, and F. R. Aussenegg, “Electron beam lithography, a helpful tool for nanooptics,” Microelectron. Eng. 83(4-9), 1464–1467 (2006). [CrossRef]

, 29

29. C. Vieu, F. Carcenac, A. Pepin, Y. Chen, M. Mejias, A. Lebib, L. Manin-Ferlazzo, L. Couraud, and H. Launois, “Electron beam lithography: resolution limits and applications,” Appl. Surf. Sci. 164(1-4), 111–117 (2000). [CrossRef]

]. Therefore, although advanced precision results in a higher cost and greater time, EBL serves academic research and development purposes the best [30

30. A. A. Tseng, K. Chen, C. D. Chen, and K. J. Ma, “Electron beam lithography in nanoscale fabrication: recent development,” IEEE Trans. Electron. Packag. Manuf. 26(2), 141–149 (2003). [CrossRef]

, 31

31. P. Rai-Choudhukry, Handbook of Microlithography, Micromachining, and Microfabrication, IEE materials and devices series (SPIE Optical Engineering Press, 1997).

].

In this study, the size and period dependency of the LSPR wavelength and the way it can be tuned was shown through simulations and justified by reflection measurements. Surface functionalization was done for a biotin-avidin pair and verified by transmission measurements. The concentration dependency of the LSPR shifts was observed by changing the analyte concentrations. Real time binding measurements and refractive index sensitivity calculations were made. Finally, the sensor structure was applied and verified to detect E. coli bacteria.

2. Materials and methods

2.1 Chemicals

2.2 Fabrications

2.3. Surface functionalization

High affinity chemical pairs are used to test the validity of a biosensor before trials with real antigens and antibodies start. The avidin and biotin (Vitamin H) pair is the most extensively studied in bio-sensing applications, which is known for one of the strongest non-covalent interactions and their extraordinary affinity towards each other (Ka = 1015M−1). Biotin is a 244 Da vitamin found in small amounts in all living cells and avidin is a tetrameric protein that is usually found in egg whites [32

32. Pierce, Avidin-Biotin Technical Handbook (Thermo Scientific, 2010).

]. In the current context, the surface functionalization refers to the steps taken for the immobilization of biotin on the Ag nano-disks and, therefore, to be ready to bind with (detect) the avidin in the target analyte. The same procedure is then applied to immobilize biotin labeled antibody and to sense the heat killed E. coli bacteria. Antibody and antigen pairs also have a high affinity towards each other. The use of affinity purified antibody ensures the high specificity for the target antigen and prevents cross-reactivity since they have lower backgrounds and lower nonspecific binding.

For the immobilization of biotin labeled affinity purified antibody to E. coli O157:H7), an SAM layer was formed following the same steps as described herein and then the samples were incubated in 0.5 ug/ml of antibody solution for at least 1 hour. E. coli O157:H7 Positive Control was prepared as described in the materials section and applied to the sensor surface.

2.4. Simulations

The simulations are done to computationally aid the design of nanostructures that give sharp and intense resonances that are necessary for sensitive detection, in turn avoiding the labor of fabrication and optical measurements.

The commercial software package “Lumerical”, which relies on “Finite-Difference Time-Domain Method” has been used in the simulations. The material data of sapphire and silver were taken from the literature [36

36. E. D. Palik, Handbook of Optical Constants of Solids (Academic Press, 1985).

]. The mesh sizes were set to values less than or equal to λ/14, where λ is the source wavelength divided by the refractive index of material of interest. The boundary conditions were set as perfectly matched layer (PML) in the direction of illumination for eliminating the undesired reflections from boundaries and periodic in the perpendicular plane for simulating a single unit cell and thereby saving simulation time. Numerical stability is ensured by setting the time step less than 0.02 fs based on the chosen mesh sizes.

Two sets of simulations were carried out to see the effect of changes in the period and diameter of nano-disks. In the first set of simulations, radii and the heights are held constant at 30 nm, and the period was varied 150-300 nm. The increase of the period without a change in radius, leads to a decrease in the density of the nanostructures and thereby a decrease in resonance intensity. See Fig. 2(a)
Fig. 2 (a) Reflection simulations: Radii and the height are held constant at 30 nm, and periods vary 150-300 nm. (b) Reflection simulations: Periods held constant at 150 nm, height is 30 nm, and the diameter varies 30-73 nm. (c) Reflection measurements: Periods held constant at 150 nm, height is 30 nm, and the diameter varies 30-73 nm.
. In the second set of simulations, the period was held constant at 150 nm, height at 30 nm, and the diameter was varied between 30 and 73 nm. The increase in radius resulted in a red shift in the resonance wavelength and an increase in the intensity of the resonance wavelength. See Fig. 2(b).

Simulation results and reflection measurements are in good accordance, which can be seen from a direct comparison of Fig. 2(b) and 2(c). Any discrepancy may be attributed to the differences in the material data and physical dimensions used at the simulations and the fabricated design.

2.5. Measurements

In the present study, a reflection and transmission set-up has been prepared separately for the measurements using the same “Ocean-Optics” spectrometer, personal computer (PC), and a xenon light source (400-700 nm).

In reflection measurements, light from the Xenon source was transmitted to an Olympus microscope with a moving stage and camera, through an optical fiber ended with a collimator, and focused onto the sample using the collimating lenses of the microscope. The light was reflected from the sample and collected back with an optical fiber.

In transmission measurements light is transmitted by a multimode optical fiber to a 10x collimating lens that collects the light at the input and then to a 40x lens that illuminates the biosensor. The light is transmitted through the biosensor sample and collected by a 20x focus lens and then a 10 x collimating lens attached to the multimode fiber.

The collected light in both measurements are fed into the monochromator that diffracts it into its different frequency components. The photomultiplier of the monochromator that measures the intensity of each component is connected to the PC and the reflectance/transmittance data is saved using special software.

Three measurements are taken to obtain the respective spectrum: one to measure the total reflected/transmitted light, second to measure the background error due to losses in the optical path and the third to obtain the reflected/transmitted light from the nano-disk arrays. The reflectance/transmittance is the ratio of the sample light intensity to reference light intensity, where background noise is subtracted from both measurements separately. Similar measurement schemes are available from the literature [25

25. K. A. Willets and R. P. Van Duyne, “Localized surface plasmon resonance spectroscopy and sensing,” Annu. Rev. Phys. Chem. 58(1), 267–297 (2007). [CrossRef] [PubMed]

, 26

26. J. N. Anker, W. P. Hall, O. Lyandres, N. C. Shah, J. Zhao, and R. P. Van Duyne, “Biosensing with plasmonic nanosensors,” Nat. Mater. 7(6), 442–453 (2008). [CrossRef] [PubMed]

, 37

37. S. L. Zhu, F. Li, C. L. Du, and Y. Q. Fu, “A localized surface plasmon resonance nanosensor based on rhombic Ag nanoparticle array,” Sens. Actuators B Chem. 134(1), 193–198 (2008). [CrossRef]

].

Graphs showing the reflection with respect to the wavelength can be seen from Fig. 2(c). And the transmission measurements taken from the bio samples at every step of the surface functionalization are shown in Fig. 3
Fig. 3 Transmission measurement for silver NPs before application of chemicals (1), after the application of chemicals (2), after the application of biotin (3), and after the application of avidin (4). Inset: the schematic illustration of surface functionalization steps.
.

3. Results and discussion

In this study, silver nano-disks on sapphire substrate are used as a sensor platform. Silver is selected since it has sharper and more intense LSPR than gold [38

38. B. Sepúlveda, P. C. Angelome, L. M. Lechuga, and L. M. Liz-Marzan, “LSPR-based nanobiosensors,” Nano Today 4(3), 244–251 (2009). [CrossRef]

]. The chemical instability and oxidation issue of silver is overcome by designing a vacuum box to keep the samples in between measurements and fabrication steps. Sapphire is transparent at optical frequencies which makes it suitable for reflection and transmission measurements. And the dimensions are selected to keep the operation at around 400-500 nm. By this way it is guaranteed to stay at optical frequencies after surface modification steps that cause red shift. This will gain more importance in future studies that involve the detection of real pathogenic bacteria where more complicated surface functionalization may be necessary since working at this portion of the spectrum provides a means for gentle detection that does not destroy the structure of matter [39

39. N. Menon, “Optical biosensors:applying photonics products to the biomedical diagnostics market,” in Optical Fiber Communication Conference (OFC), (Los Angeles, California, 2004).

].

The first set of simulations and reflection measurements are in good correlation as can be seen in Fig. 2. In both, the increase in radius resulted in a red shift in the resonance wavelength and an increase in intensity of the resonance. Mie theory may help explaining these results. Although Mie theory is originally developed for describing the scattering properties of spherical nanoparticles, it is frequently used in getting an intuition about LSPR shifts in other geometries, too [40

40. G. P. Wiederrecht, Handbook of Nanoscale Optics and Electronics (Elsevier, 2010).

]. In Mie Theory, extinction (sum of absorption and scattering) is directly proportional with the density and cube of radius of the nanostructures. (See Eq. (1) below).

E(λ)=24πNAr3εm3/2λ.ln(10)[ε1(λ)(εr(λ)+2εm)2+ε12(λ)]
(1)

Here, λ is the radiation wavelength and E(λ) is the extinction. NA is the areal density and “r” is the radius of the nanoparticles. εm is the dielectric constant of the medium surrounding the metallic nanoparticles. εi is the imaginary portion and εr is the real portion of the metallic nanoparticles’ dielectric function.

When Mie theory is correlated with our first case (Fig. 2(a)) the increase of the period without a change in radius leads to a decrease in the density of the nanostructures. According to Mie Theory, this should lead to a decrease in reflection intensity, which is the case in the simulations where increasing the period led to a decrease in the amplitude of intensity.

In the second case (Fig. 2(b)), the increase in radius keeping the period constant is expected to increase the areal density of nanoparticles as well as the amplitude of resonance intensity that is fulfilled in both the simulations and measurements (Fig. 2(b), 2(c)). The wavelength dependency of the extinction on the dielectric constants of the medium and nanoparticles is also obvious from the equation. Since the equation is derived for spherical geometry, the dependence of LSPR shifts on the shape and the height is not very explicit.

Transmission measurements were performed after every step of surface functionalization to verify their success. Figure 3 depicts the LSPR wavelength shifts for the Ag nano-disks of the period 150 nm, height 30 nm, and diameter 60 nm. After the formation of SAM, the application of biotin and avidin; 15.9nm, 7.6 nm, 19.55 nm red-shifts were observed, respectively. These amounts of shifts are reasonable and sufficient to show that the binding events take place and the sensor has detected the applied material without any uncertainty.

An important feature of biosensors is their ability to determine the concentration of the analyte to be detected. Experiments have been conducted to verify that the designed sensor has different responses at different analyte concentrations and the measurements taken showed that the sensor has an adequate response at the 1nM-100nM range, which can be fitted to a sigmoidal curve as shown in Fig. 4
Fig. 4 LSPR wavelength shifts with respect to avidin concentration. A sigmoidal dependence was observed and fit by hill function (red curve; inflection point slope ~0.99). Inset: Real time binding, LSPR wavelength versus time after the application of avidin is shown. Real time measurements after 1 minute, 2, 4, 12, 32 minutes and 3 hours are taken.
. Sigmoidal curves are used for describing the specific binding of bio-molecules due to the nature of binding [20

20. J. C. Riboh, A. J. Haes, A. D. McFarland, C. Ranjit Yonzon, and R. P. Van Duyne, “A nanoscale optical biosensor: Real-time immunoassay in physiological buffer enabled by improved nanoparticle adhesion,” J. Phys. Chem. B 107(8), 1772–1780 (2003). [CrossRef]

, 34

34. A. J. Haes, W. P. Hall, L. Chang, W. L. Klein, and R. P. Van Duyne, “A localized surface plasmon resonance biosensor: First steps toward an assay for Alzheimer's disease,” Nano Lett. 4(6), 1029–1034 (2004). [CrossRef]

, 35

35. T. Arai, P. K. R. Kumar, C. Rockstuhl, K. Awazu, and J. Tominaga, “An optical biosensor based on localized surface plasmon resonance of silver nanostructured films,” J. Opt. A, Pure Appl. Opt. 9(7), 699–703 (2007). [CrossRef]

, 41

41. A. J. Haes and R. P. Van Duyne, “A nanoscale optical biosensor: sensitivity and selectivity of an approach based on the localized surface plasmon resonance spectroscopy of triangular silver nanoparticles,” J. Am. Chem. Soc. 124(35), 10596–10604 (2002). [CrossRef] [PubMed]

]. As the concentration of the analyte increases, the surface-receptors, namely biotin bound to the SAM formation on Ag nano-disks, start to saturate. The sigmoidal fit justifies the saturation as well as the steric hindrance effects that are expected as the concentration of the analyte increases.

Another experiment was conducted to see the real time binding, after the application of avidin. The measurements showed that the greatest response was observed shortly after the application of avidin within the first few minutes and the response was well saturated within half an hour. This experiment showed that the designed LSPR sensor has a fast response enabling real time detection, which is a very important feature for biosensors where the need of detection is urgent. The real time binding measurement results are provided in the inset of Fig. 4.

After the verification of the sensor structure with a biotin-avidin pair, another trial with heat-killed E. coli was conducted. Since active E. coli bacteria can only be studied in BL2 (biosafety level 2) laboratories [42

42. S. M. Radke and E. C. Alocilja, “A high density microelectrode array biosensor for detection of E. coli O157:H7,” Biosens. Bioelectron. 20(8), 1662–1667 (2005). [CrossRef] [PubMed]

], heat-killed, potentially harmless positive controls were used in order to test the sensor structure [43

43. C. Ruan, K. Zeng, O. K. Varghese, and C. A. Grimes, “Magnetoelastic immunosensors: amplified mass immunosorbent assay for detection of Escherichia coli O157:H7,” Anal. Chem. 75(23), 6494–6498 (2003). [CrossRef] [PubMed]

]. The samples with immobilized antibody were subjected to a heat killed E. coli bacteria solution, and the transmission measurements were performed after rinsing and drying. To check for the repeatability of the results, a second measurement was taken after a second incubation of the sensor with the same bacteria solution. The results show that there is a red-shift of 4.5 nm after the application of bacteria and this measurement is stable and repeatable (Fig. 5
Fig. 5 Transmission measurements after incubation in antibody and E. coli positive control solutions inset: Dark-field Electron-microscope image of the sensor surface after the bacteria is applied. The applied bacteria look like cylindrical rods. The square shaped sensor area is 50um x 50um.
). For the time being, this trial only shows the possibility of such a sensing method. Specificity and concentration dependency should be studied before calling the sensor an E. coli sensor. A limit of detection study is not done yet. However, our current E. coli detection concentration of ~107 cfu/ml is comparable to the direct assay detection limit of 106 cfu/ml for a surface plasmon immunosensor for the detection of E. coli although it is much smaller than the sandwich assay detection limit of 103 cfu/ml for the same sensor [44

44. A. Subramanian, J. Irudayaraj, and T. Ryan, “A mixed self-assembled monolayer-based surface plasmon immunosensor for detection of E. coli O157:H7,” Biosens. Bioelectron. 21(7), 998–1006 (2006). [CrossRef] [PubMed]

]. It should be noted that no optimization studies have been performed yet, so there is still room for improvement.

An important feature of biosensors is refractive index sensitivity, which can be defined as the response of the resonance peak wavelength to changes in the bulk refractive index of the surrounding environment. In this study, we obtained the refractive index sensitivity through calculations based on transmission measurements. The sensors are exposed to iso-propanol (n = 1.3776) and water (n = 1.3325) and transmission measurements were taken while the samples were incubated in the solutions. The refractive index sensitivity was then calculated using the peak resonance wavelengths obtained through the transmission measurements using the formula Δλ/Δn; where Δn is the change in the refractive index of the dielectric environment and Δλ is the experimentally measured shift in the peak resonance wavelength. A refractive index sensitivity value of 221 nm/RIU was obtained for 40 nm diameter silver nano-disks and 354 nm/RIU for 80 nm nano-disks both with period 150 nm. These values are comparable to those obtained in the literature for different types of LSPR sensors that range between 191 and 366 nm/RIU [19

19. M. D. Malinsky, K. L. Kelly, G. C. Schatz, and R. P. Van Duyne, “Chain length dependence and sensing capabilities of the localized surface plasmon resonance of silver nanoparticles chemically modified with alkanethiol self-assembled monolayers,” J. Am. Chem. Soc. 123(7), 1471–1482 (2001). [CrossRef]

, 37

37. S. L. Zhu, F. Li, C. L. Du, and Y. Q. Fu, “A localized surface plasmon resonance nanosensor based on rhombic Ag nanoparticle array,” Sens. Actuators B Chem. 134(1), 193–198 (2008). [CrossRef]

, 45

45. S. L. Zhu, C. L. Du, and Y. Q. Fu, “Localized surface plasmon resonance-based hybrid Au-Ag nanoparticles for detection of Staphylococcus aureus enterotoxin B,” Opt. Mater. 31(11), 1608–1613 (2009). [CrossRef]

48

48. M. M. Miller and A. A. Lazarides, “Sensitivity of metal nanoparticle surface plasmon resonance to the dielectric environment,” J. Phys. Chem. B 109(46), 21556–21565 (2005). [CrossRef] [PubMed]

].

Another important comparative parameter of sensing devices is the figure of merit FOM=(Δλ/Δn).(1/Δw), where Δw is the FWHM (full-width at half-maximum) of the resonant dip. This parameter figures out the ability of the sensor to accurately measure small changes in the resonance wavelength considering the sharpness of the resonance behavior. Although there is an improvement in the sensitivity factor with the increase in diameter, FOMs calculated for the two cases described above are close to each other: 2.33 and 2.81 since the FWHM increases from 95 nm to 126 nm as the diameter increases. In the experimental trials we have chosen 60 nm diameter nanoparticles for their good sensitivity and moderate FWHM values considering the above calculations.

4. Conclusions

In this study, EBL designed silver nano-disks are shown to be used as label free nano-biosensors based on LSPR and verification is done through simulations and optical measurements.

Simulations are important since they can avoid unnecessary fabrications especially when the structure has to be tuned to a desired wavelength. In many other methodologies, LSPR substrates are hardly repeatable and the resultant sizes are not predictable so that the structure has to be fabricated to gain an intuition about the resonance frequency [33

33. J. X. Fu, A. Collins, and Y. P. Zhao, “Optical properties and biosensor application of ultrathin silver films prepared by oblique angle deposition,” J. Phys. Chem. C 112(43), 16784–16791 (2008). [CrossRef]

, 35

35. T. Arai, P. K. R. Kumar, C. Rockstuhl, K. Awazu, and J. Tominaga, “An optical biosensor based on localized surface plasmon resonance of silver nanostructured films,” J. Opt. A, Pure Appl. Opt. 9(7), 699–703 (2007). [CrossRef]

].

It is possible to extend the detection of heat killed E. coli bacteria for the detection of other pathogenic bacteria since biotin is capable of being conjugated to many proteins without altering their biological activity. If biotin conjugated antibody of specific bacteria is immobilized onto the sensor then similar steps can be taken to detect the presence of its antigen in target solution. By this method, one can even achieve multi-output sensor chips that can detect several types of bacteria simultaneously.

Acknowledgments

This work is supported by the European Union under the projects PHOME, ECONAM, and N4E, and TUBITAK under the Project Nos. 109E301, 107A004, and 107A012, and DPT under the project DPT-HAMIT. One of the authors (E.O.) also acknowledges partial support from the Turkish Academy of Sciences.

References and links

1.

Anonymous, “ISO 9308-1,” (ISO, Geneva, Switzerland, 2000).

2.

P. Daly, T. Collier, and S. Doyle, “PCR-ELISA detection of Escherichia coli in milk,” Lett. Appl. Microbiol. 34(3), 222–226 (2002). [CrossRef] [PubMed]

3.

B. L. Dylla, E. A. Vetter, J. G. Hughes, and F. R. Cockerill 3rd, “Evaluation of an immunoassay for direct detection of Escherichia coli O157 in stool specimens,” J. Clin. Microbiol. 33(1), 222–224 (1995). [PubMed]

4.

R. P. Johnson, R. J. Durham, S. T. Johnson, L. A. MacDonald, S. R. Jeffrey, and B. T. Butman, “Detection of Escherichia coli O157:H7 in meat by an enzyme-linked immunosorbent assay, EHEC-Tek,” Appl. Environ. Microbiol. 61(1), 386–388 (1995). [PubMed]

5.

M. A. Cooper, “Optical biosensors in drug discovery,” Nat. Rev. Drug Discov. 1(7), 515–528 (2002). [CrossRef] [PubMed]

6.

J. Homola, S. S. Yee, and G. Gauglitz, “Surface plasmon resonance sensors: review,” Sens. Actuators B Chem. 54(1-2), 3–15 (1999). [CrossRef]

7.

X. Y. Liu, X. D. Zeng, N. N. Mai, Y. Liu, B. Kong, Y. H. Li, W. Z. Wei, and S. L. Luo, “Amperometric glucose biosensor with remarkable acid stability based on glucose oxidase entrapped in colloidal gold-modified carbon ionic liquid electrode,” Biosens. Bioelectron. 25(12), 2675–2679 (2010). [CrossRef] [PubMed]

8.

I. Vostiar, J. Tkac, E. Sturdik, and P. Gemeiner, “Amperometric urea biosensor based on urease and electropolymerized toluidine blue dye as a pH-sensitive redox probe,” Bioelectrochemistry 56(1-2), 113–115 (2002). [CrossRef] [PubMed]

9.

A. Malave, M. Tewes, T. Gronewold, and M. Lohndorf, “Development of impedance biosensors with nanometer gaps for marker-free analytical measurements,” Microelectron. Eng. 78–79, 587–592 (2005). [CrossRef]

10.

W. Zhao, J. J. Xu, and H. Y. Chen, “Electrochemical biosensors based on layer-by-layer assemblies,” Electroanalysis 18(18), 1737–1748 (2006). [CrossRef]

11.

M. H. Meyer, M. Stehr, S. Bhuju, H. J. Krause, M. Hartmann, P. Miethe, M. Singh, and M. Keusgen, “Magnetic biosensor for the detection of Yersinia pestis,” J. Microbiol. Methods 68(2), 218–224 (2007). [CrossRef] [PubMed]

12.

J. Llandro, J. J. Palfreyman, A. Ionescu, and C. H. Barnes, “Magnetic biosensor technologies for medical applications: a review,” Med. Biol. Eng. Comput. 48(10), 977–998 (2010).

13.

X. Fan, I. M. White, S. I. Shopova, H. Zhu, J. D. Suter, and Y. Sun, “Sensitive optical biosensors for unlabeled targets: a review,” Anal. Chim. Acta 620(1-2), 8–26 (2008). [CrossRef] [PubMed]

14.

K. E. Shafer-Peltier, C. L. Haynes, M. R. Glucksberg, and R. P. Van Duyne, “Toward a glucose biosensor based on surface-enhanced Raman scattering,” J. Am. Chem. Soc. 125(2), 588–593 (2003). [CrossRef] [PubMed]

15.

M. Kahraman, M. M. Yazici, F. Sahin, O. F. Bayrak, and M. Culha, “Reproducible surface-enhanced Raman scattering spectra of bacteria on aggregated silver nanoparticles,” Appl. Spectrosc. 61(5), 479–485 (2007). [CrossRef] [PubMed]

16.

F. S. Ligler and C. A. R. Taitt, Optical Biosensors: Present and Future, 1st ed. (Elsevier, 2002), pp. viii, 607 p.

17.

L. S. Jung, C. T. Campbell, T. M. Chinowsky, M. N. Mar, and S. S. Yee, “Quantitative interpretation of the response of surface plasmon resonance sensors to adsorbed films,” Langmuir 14(19), 5636–5648 (1998). [CrossRef]

18.

A. J. Haes and R. P. Van Duyne, “A unified view of propagating and localized surface plasmon resonance biosensors,” Anal. Bioanal. Chem. 379(7-8), 920–930 (2004). [CrossRef] [PubMed]

19.

M. D. Malinsky, K. L. Kelly, G. C. Schatz, and R. P. Van Duyne, “Chain length dependence and sensing capabilities of the localized surface plasmon resonance of silver nanoparticles chemically modified with alkanethiol self-assembled monolayers,” J. Am. Chem. Soc. 123(7), 1471–1482 (2001). [CrossRef]

20.

J. C. Riboh, A. J. Haes, A. D. McFarland, C. Ranjit Yonzon, and R. P. Van Duyne, “A nanoscale optical biosensor: Real-time immunoassay in physiological buffer enabled by improved nanoparticle adhesion,” J. Phys. Chem. B 107(8), 1772–1780 (2003). [CrossRef]

21.

Z. T. Liu, M. D. Thoreson, A. V. Kildishev, and V. M. Shalaev, “Translation of nanoantenna hot spots by a metal-dielectric composite superlens,” Appl. Phys. Lett. 95, 033114 (2009).

22.

Z. Liu, A. Boltasseva, R. H. Pedersen, R. M. Bakker, A. V. Kildishev, V. P. Drachev, and V. M. Shalaev, “Plasmonic nanoantenna arrays for the visible,” Metamaterials (Amst.) 2(1), 45–51 (2008). [CrossRef]

23.

E. Hutter and J. H. Fendler, “Exploitation of localized surface plasmon resonance,” Adv. Mater. (Deerfield Beach Fla.) 16(19), 1685–1706 (2004). [CrossRef]

24.

J. Zhao, X. Zhang, C. R. Yonzon, A. J. Haes, and R. P. Van Duyne, “Localized surface plasmon resonance biosensors,” Nanomedicine (Lond) 1(2), 219–228 (2006). [CrossRef] [PubMed]

25.

K. A. Willets and R. P. Van Duyne, “Localized surface plasmon resonance spectroscopy and sensing,” Annu. Rev. Phys. Chem. 58(1), 267–297 (2007). [CrossRef] [PubMed]

26.

J. N. Anker, W. P. Hall, O. Lyandres, N. C. Shah, J. Zhao, and R. P. Van Duyne, “Biosensing with plasmonic nanosensors,” Nat. Mater. 7(6), 442–453 (2008). [CrossRef] [PubMed]

27.

F. Y. Lee, K. H. Fung, T. L. Tang, W. Y. Tam, and C. T. Chan, “Fabrication of gold nano-particle arrays using two-dimensional templates from holographic lithography,” Curr. Appl. Phys. 9(4), 820–825 (2009). [CrossRef]

28.

A. Hohenau, H. Ditlbacher, B. Lamprecht, J. R. Krenn, A. Leitner, and F. R. Aussenegg, “Electron beam lithography, a helpful tool for nanooptics,” Microelectron. Eng. 83(4-9), 1464–1467 (2006). [CrossRef]

29.

C. Vieu, F. Carcenac, A. Pepin, Y. Chen, M. Mejias, A. Lebib, L. Manin-Ferlazzo, L. Couraud, and H. Launois, “Electron beam lithography: resolution limits and applications,” Appl. Surf. Sci. 164(1-4), 111–117 (2000). [CrossRef]

30.

A. A. Tseng, K. Chen, C. D. Chen, and K. J. Ma, “Electron beam lithography in nanoscale fabrication: recent development,” IEEE Trans. Electron. Packag. Manuf. 26(2), 141–149 (2003). [CrossRef]

31.

P. Rai-Choudhukry, Handbook of Microlithography, Micromachining, and Microfabrication, IEE materials and devices series (SPIE Optical Engineering Press, 1997).

32.

Pierce, Avidin-Biotin Technical Handbook (Thermo Scientific, 2010).

33.

J. X. Fu, A. Collins, and Y. P. Zhao, “Optical properties and biosensor application of ultrathin silver films prepared by oblique angle deposition,” J. Phys. Chem. C 112(43), 16784–16791 (2008). [CrossRef]

34.

A. J. Haes, W. P. Hall, L. Chang, W. L. Klein, and R. P. Van Duyne, “A localized surface plasmon resonance biosensor: First steps toward an assay for Alzheimer's disease,” Nano Lett. 4(6), 1029–1034 (2004). [CrossRef]

35.

T. Arai, P. K. R. Kumar, C. Rockstuhl, K. Awazu, and J. Tominaga, “An optical biosensor based on localized surface plasmon resonance of silver nanostructured films,” J. Opt. A, Pure Appl. Opt. 9(7), 699–703 (2007). [CrossRef]

36.

E. D. Palik, Handbook of Optical Constants of Solids (Academic Press, 1985).

37.

S. L. Zhu, F. Li, C. L. Du, and Y. Q. Fu, “A localized surface plasmon resonance nanosensor based on rhombic Ag nanoparticle array,” Sens. Actuators B Chem. 134(1), 193–198 (2008). [CrossRef]

38.

B. Sepúlveda, P. C. Angelome, L. M. Lechuga, and L. M. Liz-Marzan, “LSPR-based nanobiosensors,” Nano Today 4(3), 244–251 (2009). [CrossRef]

39.

N. Menon, “Optical biosensors:applying photonics products to the biomedical diagnostics market,” in Optical Fiber Communication Conference (OFC), (Los Angeles, California, 2004).

40.

G. P. Wiederrecht, Handbook of Nanoscale Optics and Electronics (Elsevier, 2010).

41.

A. J. Haes and R. P. Van Duyne, “A nanoscale optical biosensor: sensitivity and selectivity of an approach based on the localized surface plasmon resonance spectroscopy of triangular silver nanoparticles,” J. Am. Chem. Soc. 124(35), 10596–10604 (2002). [CrossRef] [PubMed]

42.

S. M. Radke and E. C. Alocilja, “A high density microelectrode array biosensor for detection of E. coli O157:H7,” Biosens. Bioelectron. 20(8), 1662–1667 (2005). [CrossRef] [PubMed]

43.

C. Ruan, K. Zeng, O. K. Varghese, and C. A. Grimes, “Magnetoelastic immunosensors: amplified mass immunosorbent assay for detection of Escherichia coli O157:H7,” Anal. Chem. 75(23), 6494–6498 (2003). [CrossRef] [PubMed]

44.

A. Subramanian, J. Irudayaraj, and T. Ryan, “A mixed self-assembled monolayer-based surface plasmon immunosensor for detection of E. coli O157:H7,” Biosens. Bioelectron. 21(7), 998–1006 (2006). [CrossRef] [PubMed]

45.

S. L. Zhu, C. L. Du, and Y. Q. Fu, “Localized surface plasmon resonance-based hybrid Au-Ag nanoparticles for detection of Staphylococcus aureus enterotoxin B,” Opt. Mater. 31(11), 1608–1613 (2009). [CrossRef]

46.

S. L. Zhu, C. L. Du, and Y. Q. Fu, “Fabrication and characterization of rhombic silver nanoparticles for biosensing,” Opt. Mater. 31(6), 769–774 (2009). [CrossRef]

47.

C. D. Chen, S. F. Cheng, L. K. Chau, and C. R. C. Wang, “Sensing capability of the localized surface plasmon resonance of gold nanorods,” Biosens. Bioelectron. 22(6), 926–932 (2007). [CrossRef] [PubMed]

48.

M. M. Miller and A. A. Lazarides, “Sensitivity of metal nanoparticle surface plasmon resonance to the dielectric environment,” J. Phys. Chem. B 109(46), 21556–21565 (2005). [CrossRef] [PubMed]

OCIS Codes
(170.0170) Medical optics and biotechnology : Medical optics and biotechnology
(240.6680) Optics at surfaces : Surface plasmons
(280.1415) Remote sensing and sensors : Biological sensing and sensors
(220.4241) Optical design and fabrication : Nanostructure fabrication

ToC Category:
Sensors

History
Original Manuscript: December 1, 2011
Revised Manuscript: December 31, 2011
Manuscript Accepted: January 3, 2012
Published: January 20, 2012

Virtual Issues
Vol. 7, Iss. 3 Virtual Journal for Biomedical Optics

Citation
Neval A. Cinel, Serkan Bütün, and Ekmel Özbay, "Electron beam lithography designed silver nano-disks used as label free nano-biosensors based on localized surface plasmon resonance," Opt. Express 20, 2587-2597 (2012)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-20-3-2587


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. Anonymous, “ISO 9308-1,” (ISO, Geneva, Switzerland, 2000).
  2. P. Daly, T. Collier, and S. Doyle, “PCR-ELISA detection of Escherichia coli in milk,” Lett. Appl. Microbiol.34(3), 222–226 (2002). [CrossRef] [PubMed]
  3. B. L. Dylla, E. A. Vetter, J. G. Hughes, and F. R. Cockerill, “Evaluation of an immunoassay for direct detection of Escherichia coli O157 in stool specimens,” J. Clin. Microbiol.33(1), 222–224 (1995). [PubMed]
  4. R. P. Johnson, R. J. Durham, S. T. Johnson, L. A. MacDonald, S. R. Jeffrey, and B. T. Butman, “Detection of Escherichia coli O157:H7 in meat by an enzyme-linked immunosorbent assay, EHEC-Tek,” Appl. Environ. Microbiol.61(1), 386–388 (1995). [PubMed]
  5. M. A. Cooper, “Optical biosensors in drug discovery,” Nat. Rev. Drug Discov.1(7), 515–528 (2002). [CrossRef] [PubMed]
  6. J. Homola, S. S. Yee, and G. Gauglitz, “Surface plasmon resonance sensors: review,” Sens. Actuators B Chem.54(1-2), 3–15 (1999). [CrossRef]
  7. X. Y. Liu, X. D. Zeng, N. N. Mai, Y. Liu, B. Kong, Y. H. Li, W. Z. Wei, and S. L. Luo, “Amperometric glucose biosensor with remarkable acid stability based on glucose oxidase entrapped in colloidal gold-modified carbon ionic liquid electrode,” Biosens. Bioelectron.25(12), 2675–2679 (2010). [CrossRef] [PubMed]
  8. I. Vostiar, J. Tkac, E. Sturdik, and P. Gemeiner, “Amperometric urea biosensor based on urease and electropolymerized toluidine blue dye as a pH-sensitive redox probe,” Bioelectrochemistry56(1-2), 113–115 (2002). [CrossRef] [PubMed]
  9. A. Malave, M. Tewes, T. Gronewold, and M. Lohndorf, “Development of impedance biosensors with nanometer gaps for marker-free analytical measurements,” Microelectron. Eng.78–79, 587–592 (2005). [CrossRef]
  10. W. Zhao, J. J. Xu, and H. Y. Chen, “Electrochemical biosensors based on layer-by-layer assemblies,” Electroanalysis18(18), 1737–1748 (2006). [CrossRef]
  11. M. H. Meyer, M. Stehr, S. Bhuju, H. J. Krause, M. Hartmann, P. Miethe, M. Singh, and M. Keusgen, “Magnetic biosensor for the detection of Yersinia pestis,” J. Microbiol. Methods68(2), 218–224 (2007). [CrossRef] [PubMed]
  12. J. Llandro, J. J. Palfreyman, A. Ionescu, and C. H. Barnes, “Magnetic biosensor technologies for medical applications: a review,” Med. Biol. Eng. Comput.48(10), 977–998 (2010).
  13. X. Fan, I. M. White, S. I. Shopova, H. Zhu, J. D. Suter, and Y. Sun, “Sensitive optical biosensors for unlabeled targets: a review,” Anal. Chim. Acta620(1-2), 8–26 (2008). [CrossRef] [PubMed]
  14. K. E. Shafer-Peltier, C. L. Haynes, M. R. Glucksberg, and R. P. Van Duyne, “Toward a glucose biosensor based on surface-enhanced Raman scattering,” J. Am. Chem. Soc.125(2), 588–593 (2003). [CrossRef] [PubMed]
  15. M. Kahraman, M. M. Yazici, F. Sahin, O. F. Bayrak, and M. Culha, “Reproducible surface-enhanced Raman scattering spectra of bacteria on aggregated silver nanoparticles,” Appl. Spectrosc.61(5), 479–485 (2007). [CrossRef] [PubMed]
  16. F. S. Ligler and C. A. R. Taitt, Optical Biosensors: Present and Future, 1st ed. (Elsevier, 2002), pp. viii, 607 p.
  17. L. S. Jung, C. T. Campbell, T. M. Chinowsky, M. N. Mar, and S. S. Yee, “Quantitative interpretation of the response of surface plasmon resonance sensors to adsorbed films,” Langmuir14(19), 5636–5648 (1998). [CrossRef]
  18. A. J. Haes and R. P. Van Duyne, “A unified view of propagating and localized surface plasmon resonance biosensors,” Anal. Bioanal. Chem.379(7-8), 920–930 (2004). [CrossRef] [PubMed]
  19. M. D. Malinsky, K. L. Kelly, G. C. Schatz, and R. P. Van Duyne, “Chain length dependence and sensing capabilities of the localized surface plasmon resonance of silver nanoparticles chemically modified with alkanethiol self-assembled monolayers,” J. Am. Chem. Soc.123(7), 1471–1482 (2001). [CrossRef]
  20. J. C. Riboh, A. J. Haes, A. D. McFarland, C. Ranjit Yonzon, and R. P. Van Duyne, “A nanoscale optical biosensor: Real-time immunoassay in physiological buffer enabled by improved nanoparticle adhesion,” J. Phys. Chem. B107(8), 1772–1780 (2003). [CrossRef]
  21. Z. T. Liu, M. D. Thoreson, A. V. Kildishev, and V. M. Shalaev, “Translation of nanoantenna hot spots by a metal-dielectric composite superlens,” Appl. Phys. Lett.95, 033114 (2009).
  22. Z. Liu, A. Boltasseva, R. H. Pedersen, R. M. Bakker, A. V. Kildishev, V. P. Drachev, and V. M. Shalaev, “Plasmonic nanoantenna arrays for the visible,” Metamaterials (Amst.)2(1), 45–51 (2008). [CrossRef]
  23. E. Hutter and J. H. Fendler, “Exploitation of localized surface plasmon resonance,” Adv. Mater. (Deerfield Beach Fla.)16(19), 1685–1706 (2004). [CrossRef]
  24. J. Zhao, X. Zhang, C. R. Yonzon, A. J. Haes, and R. P. Van Duyne, “Localized surface plasmon resonance biosensors,” Nanomedicine (Lond)1(2), 219–228 (2006). [CrossRef] [PubMed]
  25. K. A. Willets and R. P. Van Duyne, “Localized surface plasmon resonance spectroscopy and sensing,” Annu. Rev. Phys. Chem.58(1), 267–297 (2007). [CrossRef] [PubMed]
  26. J. N. Anker, W. P. Hall, O. Lyandres, N. C. Shah, J. Zhao, and R. P. Van Duyne, “Biosensing with plasmonic nanosensors,” Nat. Mater.7(6), 442–453 (2008). [CrossRef] [PubMed]
  27. F. Y. Lee, K. H. Fung, T. L. Tang, W. Y. Tam, and C. T. Chan, “Fabrication of gold nano-particle arrays using two-dimensional templates from holographic lithography,” Curr. Appl. Phys.9(4), 820–825 (2009). [CrossRef]
  28. A. Hohenau, H. Ditlbacher, B. Lamprecht, J. R. Krenn, A. Leitner, and F. R. Aussenegg, “Electron beam lithography, a helpful tool for nanooptics,” Microelectron. Eng.83(4-9), 1464–1467 (2006). [CrossRef]
  29. C. Vieu, F. Carcenac, A. Pepin, Y. Chen, M. Mejias, A. Lebib, L. Manin-Ferlazzo, L. Couraud, and H. Launois, “Electron beam lithography: resolution limits and applications,” Appl. Surf. Sci.164(1-4), 111–117 (2000). [CrossRef]
  30. A. A. Tseng, K. Chen, C. D. Chen, and K. J. Ma, “Electron beam lithography in nanoscale fabrication: recent development,” IEEE Trans. Electron. Packag. Manuf.26(2), 141–149 (2003). [CrossRef]
  31. P. Rai-Choudhukry, Handbook of Microlithography, Micromachining, and Microfabrication, IEE materials and devices series (SPIE Optical Engineering Press, 1997).
  32. Pierce, Avidin-Biotin Technical Handbook (Thermo Scientific, 2010).
  33. J. X. Fu, A. Collins, and Y. P. Zhao, “Optical properties and biosensor application of ultrathin silver films prepared by oblique angle deposition,” J. Phys. Chem. C112(43), 16784–16791 (2008). [CrossRef]
  34. A. J. Haes, W. P. Hall, L. Chang, W. L. Klein, and R. P. Van Duyne, “A localized surface plasmon resonance biosensor: First steps toward an assay for Alzheimer's disease,” Nano Lett.4(6), 1029–1034 (2004). [CrossRef]
  35. T. Arai, P. K. R. Kumar, C. Rockstuhl, K. Awazu, and J. Tominaga, “An optical biosensor based on localized surface plasmon resonance of silver nanostructured films,” J. Opt. A, Pure Appl. Opt.9(7), 699–703 (2007). [CrossRef]
  36. E. D. Palik, Handbook of Optical Constants of Solids (Academic Press, 1985).
  37. S. L. Zhu, F. Li, C. L. Du, and Y. Q. Fu, “A localized surface plasmon resonance nanosensor based on rhombic Ag nanoparticle array,” Sens. Actuators B Chem.134(1), 193–198 (2008). [CrossRef]
  38. B. Sepúlveda, P. C. Angelome, L. M. Lechuga, and L. M. Liz-Marzan, “LSPR-based nanobiosensors,” Nano Today4(3), 244–251 (2009). [CrossRef]
  39. N. Menon, “Optical biosensors:applying photonics products to the biomedical diagnostics market,” in Optical Fiber Communication Conference (OFC), (Los Angeles, California, 2004).
  40. G. P. Wiederrecht, Handbook of Nanoscale Optics and Electronics (Elsevier, 2010).
  41. A. J. Haes and R. P. Van Duyne, “A nanoscale optical biosensor: sensitivity and selectivity of an approach based on the localized surface plasmon resonance spectroscopy of triangular silver nanoparticles,” J. Am. Chem. Soc.124(35), 10596–10604 (2002). [CrossRef] [PubMed]
  42. S. M. Radke and E. C. Alocilja, “A high density microelectrode array biosensor for detection of E. coli O157:H7,” Biosens. Bioelectron.20(8), 1662–1667 (2005). [CrossRef] [PubMed]
  43. C. Ruan, K. Zeng, O. K. Varghese, and C. A. Grimes, “Magnetoelastic immunosensors: amplified mass immunosorbent assay for detection of Escherichia coli O157:H7,” Anal. Chem.75(23), 6494–6498 (2003). [CrossRef] [PubMed]
  44. A. Subramanian, J. Irudayaraj, and T. Ryan, “A mixed self-assembled monolayer-based surface plasmon immunosensor for detection of E. coli O157:H7,” Biosens. Bioelectron.21(7), 998–1006 (2006). [CrossRef] [PubMed]
  45. S. L. Zhu, C. L. Du, and Y. Q. Fu, “Localized surface plasmon resonance-based hybrid Au-Ag nanoparticles for detection of Staphylococcus aureus enterotoxin B,” Opt. Mater.31(11), 1608–1613 (2009). [CrossRef]
  46. S. L. Zhu, C. L. Du, and Y. Q. Fu, “Fabrication and characterization of rhombic silver nanoparticles for biosensing,” Opt. Mater.31(6), 769–774 (2009). [CrossRef]
  47. C. D. Chen, S. F. Cheng, L. K. Chau, and C. R. C. Wang, “Sensing capability of the localized surface plasmon resonance of gold nanorods,” Biosens. Bioelectron.22(6), 926–932 (2007). [CrossRef] [PubMed]
  48. M. M. Miller and A. A. Lazarides, “Sensitivity of metal nanoparticle surface plasmon resonance to the dielectric environment,” J. Phys. Chem. B109(46), 21556–21565 (2005). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 
Fig. 4 Fig. 5
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited