Sub-pg mass sensing and measurement with an optomechanical oscillator

Fenfei Liu, Seyedhamidreza Alaie, Zayd C. Leseman, and Mani Hossein-Zadeh  »View Author Affiliations


Optics Express, Vol. 21, Issue 17, pp. 19555-19567 (2013)
http://dx.doi.org/10.1364/OE.21.019555


View Full Text Article

Acrobat PDF (2969 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Mass sensing based on mechanical oscillation frequency shift in micro/nano scale mechanical oscillators is a well-known and widely used technique. Piezo-electric, electronic excitation/detection and free-space optical detection are the most common techniques used for monitoring the minute frequency shifts induced by added mass. The advent of optomechanical oscillator (OMO), enabled by strong interaction between circulating optical power and mechanical deformation in high quality factor optical microresonators, has created new possibilities for excitation and interrogation of micro/nanomechanical resonators. In particular, radiation pressure driven optomechanical oscillators (OMOs) are excellent candidates for mass detection/measurement due to their simplicity, sensitivity and all-optical operation. In an OMO, a high quality factor optical mode simultaneously serves as an efficient actuator and a sensitive probe for precise monitoring of the mechanical eigen-frequencies of the cavity structure. Here, we show the narrow linewidth of optomechanical oscillation combined with harmonic optical modulation generated by nonlinear optical transfer function, can result in sub-pg mass sensitivity in large silica microtoroid OMOs. Moreover by carefully studying the impact of mechanical mode selection, device dimensions, mass position and noise mechanisms we explore the performance limits of OMO both as a mass detector and a high resolution mass measurement system. Our analysis shows that femtogram level resolution is within reach even with relatively large OMOs.

© 2013 OSA

1. Introduction

During the past ten years various micro and nano-electromechanical (MEMS and NEMS) mass sensors have been proposed and demonstrated based on monitoring the shift of the mechanical resonant frequency (dynamic mode), or the bending induced by the added mass (static mode) [1

1. A. Boisen, S. Dohn, S. S. Keller, S. Schmid, and M. Tenje, “Cantilever-like micromechanical sensors,” Rep. Prog. Phys. 74(036101), 1–30 (2011).

12

12. J. W. Yi, W. Y. Shih, and W.-H. Shih, “Effect of length, width, and mode on the mass detection sensitivity of piezoelectric unimorph cantilevers,” J. Appl. Phys. 91(3), 1680–1686 (2002). [CrossRef]

]. Typically in the dynamic operational mode, the mechanical modes of the resonator/oscillator are excited thermally [5

5. N. V. Lavrik and P. G. Datskos, “Femtogram mass detection using photothermally actuated nanomechanical resonators,” Appl. Phys. Lett. 82(16), 2697–2699 (2003). [CrossRef]

,10

10. Y. H. Tao, X. X. Li, T. G. Xu, H. T. Yu, P. C. Xu, B. Xiong, and C. Z. Wei, “Resonant cantilever sensors operated in a high-Q in-plane mode for real-time bio/chemical detection in liquids,” Sens. Actuators B Chem. 157(2), 606–614 (2011). [CrossRef]

], piezoelectrically [11

11. Y. Lee, G. Lim, and W. Moon, “A self-excited micro cantilever biosensor actuated by PZT using the mass micro balancing technique,” Sens. Actuators A Phys. 130-131, 105–110 (2006). [CrossRef]

,12

12. J. W. Yi, W. Y. Shih, and W.-H. Shih, “Effect of length, width, and mode on the mass detection sensitivity of piezoelectric unimorph cantilevers,” J. Appl. Phys. 91(3), 1680–1686 (2002). [CrossRef]

] or electro-statically [6

6. A. Cagliani and Z. J. Davis, “Ultrasensitive bulk disk microresonator-based sensor for distributed mass sensing,” J. Micromech. Microeng. 21(045016), 1–6 (2011).

10

10. Y. H. Tao, X. X. Li, T. G. Xu, H. T. Yu, P. C. Xu, B. Xiong, and C. Z. Wei, “Resonant cantilever sensors operated in a high-Q in-plane mode for real-time bio/chemical detection in liquids,” Sens. Actuators B Chem. 157(2), 606–614 (2011). [CrossRef]

]. The vibrational frequencies are then monitored using piezo-resistive [8

8. M. Li, H. X. Tang, and M. L. Roukes, “Ultra-sensitive NEMS-based cantilevers for sensing, scanned probe and very high-frequency applications,” Nat. Nanotechnol. 2(2), 114–120 (2007). [CrossRef] [PubMed]

,11

11. Y. Lee, G. Lim, and W. Moon, “A self-excited micro cantilever biosensor actuated by PZT using the mass micro balancing technique,” Sens. Actuators A Phys. 130-131, 105–110 (2006). [CrossRef]

], capacitive [6

6. A. Cagliani and Z. J. Davis, “Ultrasensitive bulk disk microresonator-based sensor for distributed mass sensing,” J. Micromech. Microeng. 21(045016), 1–6 (2011).

] or optical mechanisms [7

7. M. W. Pruessner, T. H. Stievater, M. S. Ferraro, W. S. Rabinovich, J. L. Stepnowski, and R. A. McGill, “Waveguide micro-opto-electro-mechanical resonant chemical sensors,” Lab Chip 10(6), 762–768 (2010). [CrossRef] [PubMed]

,9

9. M. A. Taylor, A. Szorkovszky, J. Knittel, K. H. Lee, T. G. McRae, and W. P. Bowen, “Cavity optoelectromechanical regenerative amplification,” Opt. Express 20(12), 12742–12751 (2012). [CrossRef] [PubMed]

,10

10. Y. H. Tao, X. X. Li, T. G. Xu, H. T. Yu, P. C. Xu, B. Xiong, and C. Z. Wei, “Resonant cantilever sensors operated in a high-Q in-plane mode for real-time bio/chemical detection in liquids,” Sens. Actuators B Chem. 157(2), 606–614 (2011). [CrossRef]

].

Recently we have shown that optomechanical oscillator (OMO) can function as a compact, all-optical and low-power mass sensor [13

13. F. Liu and M. Hossein-Zadeh, “Mass Sensing with Optomechanical Oscillation,” IEEE Sensors 13(1), 146–147 (2013). [CrossRef]

]. Because of the high quality factor (Q-factor) of the circulating optical modes inside an OMO, a sub-mW optical beam can simultaneously function as an actuator and a sensitive probe. Beyond simplicity and low-power consumption the narrow linewidth (low-noise) of the OMO enables high-resolution frequency tracking and therefore high sensitivity mass measurement.

An optomechanical resonator is simply a high quality factor (high-Q) optical microcavity that also supports high-Q mechanical modes that can efficiently modulate the optical path length [14

14. T. J. Kippenberg and K. J. Vahala, “Cavity Opto-Mechanics,” Opt. Express 15(25), 17172–17205 (2007). [CrossRef] [PubMed]

]. In a high-Q optomechanical resonator, the large circulating optical power results in a strong coupling between the optical and mechanical modes through radiation pressure. The interplay between resonant optical wavelength and the mechanical deformation may result in optomechanically induced gain or loss for the motion of the corresponding mechanical mode (depending on pump laser wavelength detuning) [14

14. T. J. Kippenberg and K. J. Vahala, “Cavity Opto-Mechanics,” Opt. Express 15(25), 17172–17205 (2007). [CrossRef] [PubMed]

]. In the blue detuned regime (Δλ = λreslaser > 0, λres: resonant wavelength of the optical mode, λlaser: laser wavelength) above certain threshold power (Pth) the optomechanical gain cancels the mechanical loss and self-sustained optomechanical oscillation ensues [15

15. H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Radiation-pressure-driven micro-mechanical oscillator,” Opt. Express 13(14), 5293–5301 (2005). [CrossRef] [PubMed]

18

18. M. Hossein-Zadeh and K. J. Vahala, “Optomechanical Oscillator on a Silicon Chip”, Invited Paper, J Sel. Top. Quantum Electron. Special Issue on Silicon Photonics 16(1), 276–287 (2010).

]. Consequently the transmitted optical power (coupled out of microresonator) will be modulated at the mechanical oscillation frequency with an amplitude proportional to optical input power (pump power). So in an optomechanical oscillator (OMO) the optical power serves both as an efficient actuator and sensitive probe for monitoring the resonant frequencies of the oscillating mechanical modes. The optomechanical oscillation frequency (fOMO) is almost equal to the frequency of the corresponding mechanical mode (f0) and therefore depends on the size, geometry and material properties of the cavity structure [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

]. The sensitivity slope for mass sensing can be defined as η = |Δf0m|, where Δf0 is the frequency shift of the corresponding mechanical mode due to the added mass Δm. Accordingly, the sensing resolution is limited by η and the resolution of frequency shift (Δf0fOMO) measurement. While η is determined by the dimensions of the sensor [12

12. J. W. Yi, W. Y. Shih, and W.-H. Shih, “Effect of length, width, and mode on the mass detection sensitivity of piezoelectric unimorph cantilevers,” J. Appl. Phys. 91(3), 1680–1686 (2002). [CrossRef]

], position of the mass and the selected mechanical mode, the minimum resolution for frequency measurement (ΔfOMO-min) is fundamentally limited by the linewidth of the mechanical oscillation (δfL) and frequency jitter. The extremely narrow linewidth (δfL) of OMO [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

,19

19. H. Rokhsari, M. Hossein-Zadeh, A. Hajimiri, and K. J. Vahala, “Brownian noise in radiation-pressure-driven micromechanical oscillators,” Appl. Phys. Lett. 89(6), 261109 (2006).

] makes it a high-resolution mass sensor [13

13. F. Liu and M. Hossein-Zadeh, “Mass Sensing with Optomechanical Oscillation,” IEEE Sensors 13(1), 146–147 (2013). [CrossRef]

].

It is worth pointing out that in the absence of the optomechanical interaction, high-Q modes of optical microresonators have been extensively used for molecule detection and quantification based on effective refractive index change induced by molecular polarization [20

20. F. Vollmer and S. Arnold, “Whispering-gallery-mode biosensing: label-free detection down to single molecules,” Nat. Methods 5(7), 591–596 (2008). [CrossRef] [PubMed]

]. Basically direct interaction between molecules and the evanescent field of the optical mode results in a detectable resonant optical wavelength shift proportional to the total number of interacting molecules. In contrast, in an optomechanical oscillator the particles/molecules interact with the mechanical mode that is activated and interrogated by a high-Q optical mode. In an optomechanical mass sensor the evanescent optical field does not play any role and the optical mode is completely isolated from the particle/molecules (this is an advantage, because direct optical interaction degrades the optical-Q and therefore the performance of the device).

Here, we demonstrate sub-pg mass sensitivity of relatively large microtoroid OMOs enabled by frequency multiplication in optical domain. We also present a detailed analysis of optomechanical mass sensing mechanism and identify three possible operational modes. Our calculations show that when driven by a low noise and stable optical source, even large OMOs (diameter ~100 μm) can reach femtogram (fg) level mass sensitivity.

We have chosen the silica microtoroid OMO for our study because of its low phase noise, large capture area and well-studied optomechanical oscillatory behavior [15

15. H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Radiation-pressure-driven micro-mechanical oscillator,” Opt. Express 13(14), 5293–5301 (2005). [CrossRef] [PubMed]

18

18. M. Hossein-Zadeh and K. J. Vahala, “Optomechanical Oscillator on a Silicon Chip”, Invited Paper, J Sel. Top. Quantum Electron. Special Issue on Silicon Photonics 16(1), 276–287 (2010).

]. Recently monolithic micro/nano scale optomechanical oscillators have been proposed and demonstrated [21

21. W. C. Jiang, X. Y. Lu, J. D. Zhang, and Q. Lin, “High-frequency silicon optomechanical oscillator with an ultralow threshold,” Opt. Express 20(14), 15991–15996 (2012). [CrossRef] [PubMed]

25

25. M. W. Pruessner, T. H. Stievater, J. B. Khurgin, and W. S. Rabinovich, “Integrated waveguide-DBR microcavity opto-mechanical system,” Opt. Express 19(22), 21904–21918 (2011). [CrossRef] [PubMed]

], however all of them have larger phase noise (due to the smaller effective mass) and relatively small capture areas. Also most nano-scale OMOs have very low-mechanical quality factor in air and can only oscillate in vacuum environment. The large optical-Q of silica microtoroidal resonator not only reduces the optomechanical threshold power but also enables high sensitivity monitoring of the mechanical modes [26

26. A. Schliesser, G. Anetsberger, R. Riviere, O. Arcizet, and T. J. Kippenberg, “High-sensitivity monitoring of micromechanical vibration using optical whispering gallery mode resonators,” New J. Phys. 10(9), 095015 (2008). [CrossRef]

]. Even in atmospheric pressure and room temperature silica microtoroid OMO have relatively low threshold power and very narrow oscillation line-width (sub-Hz linewidth has been reported with only 300 μW of optical input power [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

]). Compared to other OMO cavities, the silica microtoroid has a large sensitive area that is crucial for mass detection and measurement. Finally the fabrication process of silica microtoroids is relatively simple [27

27. D. K. Armani, T. J. Kippenberg, S. M. Spillane, and K. J. Vahala, “Ultra-high-Q toroid microcavity on a chip,” Nature 421(6926), 925–928 (2003). [CrossRef] [PubMed]

] compared to other OMOs.

2. Experimental results

Figure 1(a)
Fig. 1 (a) Experimental arrangement used for mass measurement using microtoroid optomechanical oscillator. (b) SEM image of one of the silica microtoroids used in this study. D = 131 µm, Dp = 5.2 µm and d = 7.8 µm. Inset: a polyethylene microbead landed on the toroidal region of the resonator. (c) Schematic diagram showing the cross–section of a microtoroid, mechanical deformation of the silica membrane (associated with the 3rd mechanical mode), and the trajectory of the circulating WG optical mode (red line).
shows the experimental arrangement used for mass measurement using microtoroid optomechanical oscillator. Optical input power is provided by a near-IR tunable laser (λ~1550 nm) that is coupled into and out of the microtoroid cavity using silica fiber-taper. When the laser wavelength is blue detuned (Δλ > 0) and the input power exceeds the optomechanical oscillation threshold power (i.e. Pin > Pth), the amplitude of the transmitted optical power will be modulated at fOMO with a modulation depth proportional to Pin/Pth [17

17. H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Theoretical and Experimental Study of Radiation Pressure-Induced Mechanical Oscillations (Parametric Instability) in Optical Microcavities,” IEEE J. Sel. Top. Quantum Electron. 12(1), 96–107 (2006).

]. Upon detection in a photodetector the optical signal generates an electric signal that is measured and analyzed using an RF spectrum analyzer. Figure 1(b) shows the SEM image of one of the silica microtoroids used in this study. This microtoroid has a major diameter of D = 131 µm, minor diameter of d = 7.4 µm. The silica membrane is 2 µm thick and is attached on top of a silicon pillar with a diameter of Dp ~5.2 µm. The inset in Fig. 1(b) shows a micro-bead with a diameter of 1.3 μm landed on the toroidal section of the microresonator.

We used two distinct mechanical modes excited on two different microtoroids to evaluate the mass sensitivity of OMO and the enhancement obtained by monitoring optically generated harmonics. The wavelength detuning (Δλ) was kept constant during all measurements. Figure 2(a)
Fig. 2 (a) Measured RF spectrum (RBW = 30 Hz) of the OMO optical output power in the absence of the microbeads. For this OMO D = 131 µm, Dp = 5.2 µm and d = 7.8 µm and fOMO = 8.505 MHz. Here: Q0 = 6.8 × 107, Qtot = 6.72 × 106,(Qtot is the loaded optical-Q), Qmech = 1300, Pth = 553.7µW, Pin = 2.4Pth and ΔλN = 0.56 (ΔλN = Δλ/δλ, where δλ = λres/Qtot). (b) Measured RF spectrum of the OMO in the vicinity of the fundamental frequency as microbeads are added sequentially (2 through 5). (c) Images of the distribution of microbeads on the microtoroid OMO. (d) SEM image of the microtoroid OMO and the microbeads distributed on it (the distribution corresponds to c-5). (e) FEM modeling of the mechanical deformation of the corresponding mechanical mode; the contours show the total displacements. (f) Measured RF spectrum (RBW = 30 Hz) of the OMO in the vicinity of the fifth harmonic frequency as microbeads are added sequentially (1 through 5). (g) Measured frequency shift of the fundamental oscillation ΔfOMO and its 5th harmonic Δ(5fOMO) plotted against the loaded mass. The dashed red line depicts the calculated results using numerical modeling. The inset is a close-up view for Δm< 5 pg.
shows the measured RF spectrum of the first OMO in the absence of external mass. The higher harmonics are generated by the nonlinear oscillatory optical transfer function of the optical microresonator as described in Ref. 28

28. F. Liu and M. Hossein-Zadeh, “On the spectrum of the radiation-pressure-driven optomechanical oscillator and its application in sensing,” Opt. Commun. 294, 338–343 (2013). [CrossRef]

. Basically the oscillation of the Lorentzian transfer function at fOMO (stimulated by optomechanical interaction) modulates the transmitted optical power at fOMO and its harmonics (n × ΔfOMO, n = 1,2,3…). Figure 2(b) shows the behavior of the RF spectrum of the detected optical power near the fundamental optomechanical oscillation frequency (fOMO = 8.505 MHz) as micro-beads are loaded onto the OMO. For all spectral measurements the resolution bandwidth (RBW) of the RF spectrum analyzer is set to 30 Hz. Figure 2(c) shows the top-view micrograph of the microtoroid as the microbeads (red polyethylene microspheres) are randomly distributed on the microtoroid in four steps (1 is bare microtorid and 2-5 correspod to the measured spectrums and frequency shifts in Fig. 2(b), 2(f) and 2(g)). Figure 2(d) shows the corresponding Scanning Electron Microscope (SEM) images of the microtoroid and the added microbeads. The deformation of the corresponding mechanical mode (estimated by finite element modeling) is shown in Fig. 2(e). Figure 2(f) shows the measured RF spectrum of the OMO in the vicinity of the fifth harmonic frequency as microbeads are added sequentially (see Fig. 2(c)). A mass induced mechanical frequency shift of ΔfOMO is translated to optical amplitude modulation frequency shift of n × ΔfOMO (where n ≥ 2 is the harmonic component whose shift is measured). Since the linewidth of the oscillation is almost equal for all harmonics [28

28. F. Liu and M. Hossein-Zadeh, “On the spectrum of the radiation-pressure-driven optomechanical oscillator and its application in sensing,” Opt. Commun. 294, 338–343 (2013). [CrossRef]

], the intrinsic optical harmonic generation effectively functions as an amplifying mechanism that enhances the frequency shift. Using larger optical input power and optimized detuning higher harmonics can be excited and monitored. Figure 2(g) shows the measured optomechanical oscillation frequency shift of the fOMO and its fifth harmonic plotted versus the added mass. The values of the added mass are estimated using the mass of the microbeads and their distribution obtained from the SEM and optical images after each step (as show in Fig. 2(d)). The mass of each microbead is calculated from the density of the corresponding material (polyethylene) and the volume of each microbead is estimated from the SEM image. The nonlinear behavior is due to the fact that mass is not added on a specific location of the disk (see section 3). The average sensitivity of the fundamental oscillation frequency to mass variation is about 266 Hz/pg and is in good agreement with the simulated results using finite element (FEM) modeling (red circles connected with a dashed line). As expected the sensitivity of the 5th harmonic is about five times larger (1380 Hz/pg).

Figure 3(a)
Fig. 3 (a) Top view images of the distribution of microbeads on the second microtoroid OMO. For this OMO D = 133 µm, DP = 11.25 µm, d = 7.24 µm and fOMO = 24.88 MHz, (b) Frequency shift of the fundamental oscillation ΔfOMO and its 5th harmonic Δ(5fOMO) plotted against the loaded mass. The dashed line depicts the calculated results using numerical modeling. The inset shows FEM modeling of the mechanical deformation of the corresponding mechanical mode. (c) and (d) are the measured RF spectrum ((RBW = 30 Hz)) of the OMO in the vicinity of the fundamental frequency (c) and its fifth harmonic (d) as microbeads are added sequentially (1 through 4). Here: Q0 = 3.6 × 106, Qtot = 2.3 × 106(Qtot is the loaded optical-Q), Qmech = 1466, ΔλN = 0.58 (ΔλN = Δλ/δλ, where δλ = λres/Qtot), Pth = 212 µW, and Pin = 3.67 Pth.
shows the top-view photograph of another microtoroid as microbeads are deposited on its surface in 3 steps. This microtoroid has a major diameter of D = 133 µm, pillar diameter of DP = 11.25 µm and minor diameter of d = 7.24 µm. On this microtoroid we have excited a mechanical mode with a fundamental optomechanical oscillation frequency of fOMO = 24.88 MHz. Figure 3(b) shows the measured optomechanical oscillation frequency shift of the fundamntal and its 5th harmonic plotted versus the added mass. The sensitivity of the fundamental oscillation frequency to mass variation is about 72 Hz/pg and is in good agreement with the simulated results using FEM simulation (red circles connected with a dashed line). The sensitivity of the 5th harmonic is about five times larger (~334 Hz/pg). The inset shows FEM modeling of the deformation of the corresponding mechanical mode. Figure 3(c) and 3(d) show the RF spectrum of the detected optical power near the fundamental optomechanical oscillation frequency (24.88 MHz) and its 5th fifth harmonic (~124.4 MHz) as microbeads are loaded onto the OMO.

3. Analysis of mass sensing with optomechanical oscillation

Figure 4(b) shows the frequency shift for 1 pg mass (that is equal to sensitivity slope η with the unit of Hz/pg) plotted against radial position of mass (along θ = 0 direction) for the modes shown Fig. 4(a) using both the Energy and FEM Methods. The sensitivity of the first mode is negligible compared to the other modes both because the displacement (mechanical energy) is distributed all over the disk and the oscillation frequency is very small. As expected because 1 pg is small compared to the resonator mass (mr = 6 × 104 pg), the outcomes of the two methods are very close. The agreement between the two methods justifies the use of Energy method that significantly simplifies the calculation processes for multiple masses at different positions. In addition, validity of the Energy method verifies the linearity of the response when Δm << mr and the mass is added at a specific location (because based on Energy method η only depends on the location of the added mass and not its magnitude).

For mass measurement (where the particle under test can be placed at a desired position), the mode order (i) and optimal position (rm, θm) should be chosen such that the slope sensitivity is maximized (i.e. ηmax = ηi,max = ηi(rmm)). According to Fig. 4(b) for the microtoroid in Fig. 2, the best choice is the 3rd mode and the best position is r = 17 µm (with an arbitrary value of θ due to axial symmetry of this mode). Near this specific position η3 ~500 Hz/pg for the fundamental oscillation and n × 500 Hz/pg when the frequency of the nth harmonic is measured. Figure 5(a)
Fig. 5 (a) Calculated maximum sensitivity ηmax plotted against D/DP for two different values of D and three eigen modes. (b) Sensitive area (As) for detecting 1 pg mass plotted against D for three eigen modes (assuming ΔfOMO-min = 40 Hz). Inset: Optimized microtoroid diameter Dopt plotted against the values of added mass.
shows ηmax plotted against D/DP for two different values of D and the three selected eigen modes (i = 1,2,3,4). Generally ηmax increases as D is decreased and for a given D, shows an asymptotic behavior as a function D/DP reaching a saturated value for D/DP > 10. As D/DP increases, a larger portion of energy is stored in the disk (U is larger) but the resonant frequency (fOMOf0) is reduced. According to Eq. (1), ΔfOMOfOMO3 × U2 so these two effects balance each other for large values of D/DP.

Note that as expected this study suggests that the oscillation frequency of smaller OMOs and those with smaller effective mass, e.g. the zipper photonic crystal [23

23. M. Eichenfield, R. Camacho, J. Chan, K. J. Vahala, and O. Painter, “A picogram- and nanometre-scale photonic-crystal optomechanical cavity,” Nature 459(7246), 550–555 (2009). [CrossRef] [PubMed]

] and spoke-supported silica microtoroid or silicon nitride microrings [22

22. S. Tallur, S. Sridaran, and S. A. Bhave, “A monolithic radiation-pressure driven, low phase noise silicon nitride opto-mechanical oscillator,” Opt. Express 19(24), 24522–24529 (2011). [CrossRef] [PubMed]

] should be more sensitive to the added mass compared to the silica microtoroid. However, loading mass to the most sensitive location of these OMOs is a difficult task and the presence of the particle interferes with the resonant optical field degrading the optical-Q to a level that quenches the optomechanical oscillation. Moreover the smaller effective mass of these devices results in larger phase noise (larger oscillation linewidth) that limits the smallest measurable frequency shift.

For lumped mass detection, the goal is to detect the existence of added mass on the OMO. Therefore the capture probability should be maximized while maintaining the induced frequency shift above the minimum detectable frequency change (i.e. ΔfOMO-min). ΔfOMO-min is limited by different noise mechanisms and the resolution of RF spectrum analyzer (see Section 4). For a given mass change Δm, we define the sensitive area (As) using the condition ΔfOMO > ΔfOMO-min. Figure 5(b) shows As plotted against D for the three eigen-modes (2-4) when pillar diameter is DP = 11.5 µm, particle mass is Δm = 1 pg and ΔfOMO-min = 40 Hz (chosen based on the limitation of our measurement system). For each eigen mode an optimized diameter (Dopt) can be identified that maximizes As. The inset shows Dopt plotted versus mass value (Δm) suggesting for detection of heavier particles, larger microtoroids provide larger sensitive capture area.

Finally, the microtoroid OMO can be used for detection of uniformly distributed mass on its surface. This capability is effectively translated to measuring the surface density (ρs) of gas molecules that are adsorbed on the OMO surface. Since in thermodynamic equilibrium ρs is proportional to the concentration of the molecules in the surrounding medium, one can extract the volume density from oscillation frequency shift. Note that in the case of biomolecules the adsorption probability can be significantly enhanced by functionalizing the surface (e.g. an organic monolayer deposited on top of the microtoroid). However at large densities, when the molecules are close enough to interact with each other, the intermolecular forces affect the resonant frequency through surface stress effects. As shown in the context of cantilever sensor, in this case both surface stress and mass effects should be taken into account [2

2. J. Tamayo, P. M. Kosaka, J. J. Ruz, A. San Paulo, and M. Calleja, “Biosensors based on nanomechanical systems,” Chem. Soc. Rev. 42(3), 1287–1311 (2013). [CrossRef] [PubMed]

]. For the sake of simplicity here we only consider a low-density but uniform distribution of non-interacting particles (ignoring surface stress effect). For a given microtoroid and mechanical mode Eq. (1) can be modified as:
ΔfOMOπ2fOMO3ρ(r,θ)ΔaU2(r,θ)Etotal
(2)
where ρs(r, θ) is surface mass density at position (r, θ) on the microdisk and Δa is the area covered by the external mass. To evaluate the contribution of the whole capture area, the figure of merit Λ is defined as:
Λ=π2fOMO3EtotalAρ(r,θ)U2(r,θ)da
(3)
Λ has units of Hz and represents the total frequency shift due to the surface density ρs. Λ, also contains the effects of microtoroid dimensions (which directly defines the capture area), frequency of oscillation and mechanical deformation (for a specific mode) for a particular mass distribution.

Figure 6
Fig. 6 Calculated Λ plotted against D/DP for two different values of D and the four eigen modes.
shows Λ is plotted against D/DP for two different values of D and the three selected eigen modes (2,3,4) as shown in Fig. 4(a), assuming a uniform mass distribution with a constant surface mass density of ρs = 0.1 ng/mm2. As evident from the graph increasing D/DP beyond 10 does not affect Λ.

In conclusion depending on the desired mode of operation, the optimized dimensions of the microtoroid sensor correspond to maximum ηmaxfOMO-min (for lumped mass measurement), Λ/ΔfOMO-min (for detection/measurement of distributed mass) or AsfOMO-min (for detection of lumped mass).

4. Ultimate performance of optomechanical mass sensor

The minimum mass resolution for an OMO sensor (for lumped mass measurement mode) is limited by the maximum mass sensitivity slope (ηmax) and the minimum detectable frequency shift (ΔfOMO-min) through:
Δmmin=ΔfOMOminηmax
(4)
ΔfOMO-min is limited by OMO frequency noise and the resolution of frequency shift measurement instrument (δfI). In classical regime, optomechanical oscillation frequency noise (δfOMO) originates from fluctuations of the circulating optical power δ(Pcirc) and the thermo-mechanical (Brownian) noise. As shown before [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

,19

19. H. Rokhsari, M. Hossein-Zadeh, A. Hajimiri, and K. J. Vahala, “Brownian noise in radiation-pressure-driven micromechanical oscillators,” Appl. Phys. Lett. 89(6), 261109 (2006).

] at room temperature Brownian noise (∝ kBT) and mechanical quality factor of the corresponding mode (Qm) set the lower limit for optomechanical oscillation linewidth (δfL), while δ(Pcirc) generates oscillation frequency fluctuations through optical spring and opto-thermal effects (Δfp = ΔfS + ΔfT) [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

]. So the ΔfOMO-min can be expanded as:
ΔfOMOmin=δfL+δ(ΔfP)+δfI
(5)
where Δfp ( = fOMO - f0) is the shift due to the “optical spring effect” and “thermal effect” [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

]. ΔfP translates the fluctuation of input parameters, ΔνN ( = Δν/δν = νres− νlaser/δν, where δν is the loaded linewidth of the corresponding optical mode), and Pin to fluctuations of fOMO by affecting the circulating optical power inside the cavity. Although in our experiment δ(Qtot/Q0) also contributes to δ(ΔfP) (due to uncertainties in the coupling gap controlled by piezo stage), in new integrated OMOs [21

21. W. C. Jiang, X. Y. Lu, J. D. Zhang, and Q. Lin, “High-frequency silicon optomechanical oscillator with an ultralow threshold,” Opt. Express 20(14), 15991–15996 (2012). [CrossRef] [PubMed]

25

25. M. W. Pruessner, T. H. Stievater, J. B. Khurgin, and W. S. Rabinovich, “Integrated waveguide-DBR microcavity opto-mechanical system,” Opt. Express 19(22), 21904–21918 (2011). [CrossRef] [PubMed]

] this contribution is negligible. For given values of δ(ΔνN) and δPin, δ(ΔfP) depends on the microtoroid dimensions [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

]. Note that the photodetection noise (i.e. shot noise, dark noise, and thermal noise) have negligible impact on the precision and resolution of the frequency monitoring.

The harmonic components (fOMO(n) = n × fOMO) are generated in the optical domain by the oscillatory transfer function of the cavity [28

28. F. Liu and M. Hossein-Zadeh, “On the spectrum of the radiation-pressure-driven optomechanical oscillator and its application in sensing,” Opt. Commun. 294, 338–343 (2013). [CrossRef]

], so their oscillation linewidth that is limited by the thermo-mechanical noise [16

16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

,19

19. H. Rokhsari, M. Hossein-Zadeh, A. Hajimiri, and K. J. Vahala, “Brownian noise in radiation-pressure-driven micromechanical oscillators,” Appl. Phys. Lett. 89(6), 261109 (2006).

] are almost equal to that of the fundamental oscillation (δfL). Note that here we are ignoring the noise associated with the optical multiplication process because it is negligible compared to the thermo-mechanical noise. So When the nth optically generated harmonic of the OMO is used for mass sensing, only Δfp changes and Eq. (5) can be modified as:
ΔfOMOmin(n)=δfL+n×δ(ΔfP)+δfI
(6)
The corresponding minimum detectable mass is:
ΔmminδfL+δfIn×ηmax+δfPηmax
(7)
Using lasers with small relative intensity noise (RIN <100 dB/Hz) and assuming δ(Qtot/Q0) = 0, δ(Δfp) will be only limited by δ(ΔνΝ). Modern frequency stabilization techniques that have enabled laser frequency fluctuation as small as 1 Hz [31

31. S. Seel, R. Storz, G. Ruoso, J. Mlynek, and S. Schiller, “Cryogenic Optical Resonators: A New Tool for Laser Frequency Stabilization at the 1 Hz Level,” Phys. Rev. Lett. 78(25), 4741–4744 (1997). [CrossRef]

] combined with wavelength locking techniques, such as Hänsch–Couillaud [32

32. T. W. Hansch and B. Couillaud, “Laser Frequency Stabilization by Polarization Spectroscopy of a Reflecting Reference Cavity,” Opt. Commun. 35(3), 441–444 (1980). [CrossRef]

] or Pound-Drever-Hall [33

33. R. W. P. Drever, J. L. Hall, F. V. Kowalski, J. Hough, G. M. Ford, A. J. Munley, and H. Ward, “Laser Phase and Frequency Stabilization Using an Optical-Resonator,” Appl. Phys. B 31(2), 97–105 (1983). [CrossRef]

] methods, can easily keep δ(ΔνΝ) below 500 Hz. Such level of stability results in (Δfp) ~0.1 Hz < δfL enabling mass measurement/detection sub-fg resolutions (limited only by optomechanical oscillation linewidth).

Figure 7(a)
Fig. 7 (a) Δmmin plotted against normalized detuning ΔνN for the fifth harmonic (n = 5) of the mechanical modes of the microtoroid in Fig. 2 (D = 131 µm, DP = 5.2 µm). The mass is located at the position of maximum sensitivity (ηmax = ηi(rmm)) for each mode. Here Q0 = 6.8 × 107, Qmech = 1300, Qtot/Q0 = 0.1 and Pin = 2 Pth. We have assumed RIN<90dB, δ(Δν0) is 500 Hz and ignored δ(Qtot/Q0) and δfI. (b) Δmmin plotted against the order of harmonic frequency used for the measurement (n) for two cases: δfI = 0 dashed lines) and δfI = 30 Hz (solid lines). The insets show the membrane deformation for the 2nd and 4th modes. The mass is located at the position of maximum sensitivity (ηmax = ηi(rmm)) for each mode shown as red zones in the insets. ΔλN is fixed at 0.28 that is the optimal detuning according to part-a. Other parameters are the same as the ones in part-a. (c) Δmmin plotted against the D/Dp based on the fifth harmonic shift, for D = 103 µm (dashed lines) and 133 µm (solid lines). Here all parameters are chosen according to the actual experiment. ΔλN = 0.56, Q0 = 6.8 × 107, Qmech~1400, and Pin~2 Pth. Qtot/Q0 = 0.1. Uncertainty of Qtot/Q0 and detuning δ (Δλ) are 1% and 400 KHz, respectively. Pin = 2Pth, δfI = 30 Hz.
shows Δmmin plotted against ΔνN for the microtoroid in Fig. 2(a) when δ(Δν0) = 500 Hz, δ(Qtot/Q0) = 0 and δfI = 0. Here the mass is located at the position of the maximum sensitivity (ηmax = ηi(rmm)) for each mode. By fixing the normalized detuning (ΔνN) at its optimal value (~0.28), in Fig. 7(b) we have plotted Δmmin against the order of harmonic frequency used for the measurement (n) for two cases: δfI = 0 and δfI = 30 Hz. As evident from this plot measuring the shift of the 7th harmonic frequency (instead of the fundamental frequency) can improve the sensitivity by almost one order of magnitude.

Figure 7(c) shows Δmmin plotted against the D/Dp based on fifth harmonic shift, for D = 103 µm (dashed lines) and 133 µm (solid lines). Here all parameters are chosen according to the actual experiment. ΔλN = 0.56, Q0 = 6.8 × 107, Qmech~1400, and Pin~2 Pth. Qtot/Q0 = 0.1. Uncertainty of Qtot/Q0 and detuning δ(Δλ) are 1% and 400 KHz, respectively. Pin = 2Pth, δfI = 30 Hz. In principle using our measurement system and the OMO shown in Fig. 2 a mass detection limit of 150 fg (150 × 10−15) is within reach.

Note that in electro-optomechanical oscillator driven by electro-mechanical force through external feedback [9

9. M. A. Taylor, A. Szorkovszky, J. Knittel, K. H. Lee, T. G. McRae, and W. P. Bowen, “Cavity optoelectromechanical regenerative amplification,” Opt. Express 20(12), 12742–12751 (2012). [CrossRef] [PubMed]

], δ(Δfp) may become negligible by reducing the optical power resulting to a fundamental sensitivity limited by thermal noise. However the impact of the feedback noise and frequency pulling mechanisms should be taken into account in these devices.

5. Conclusion

This work lays the foundation for a new class of mass sensors that can be used for characterization and detection of particles and molecules. Beyond all-optical operation the main advantage of OMO mass sensor is the simple and low-noise actuation and read-out mechanisms based on optical force and resonant optical path length modulation. Optical read-out based on high-Q optical modes enables monitoring higher order mechanical modes with minimal out of plane displacement that are usually difficult to monitor using other techniques. The fact that due to reduced viscous damping high order mechanical modes have larger mechanical-Qs in air, enables narrow linewidth oscillation and therefore high sensitivity in atmospheric pressure. In principle all newly introduced integrated and nano-scale OMOs can be used as ultra-sensitive mass sensors, however the large capture area, narrow linewidth oscillation in air (due to larger effective mass) and simple fabrication method make the microtoroidal OMO a better candidate for practical applications. High-Q optical modes can be also used to excite and interrogate resonant surface acoustic waves (as opposed to bulk mechanical modes) [34

34. G. Bahl, K.-H. Kim, W. Lee, J. Liu, X. Fan, and T. Carmon, “Brillouin cavity optomechanics with microfluidic devices,” Nat Commun 4, 1994 (2013). [CrossRef] [PubMed]

]. Although these modes may be used for mass sensing, their confinement near the small optical mode volume results in very small sensitive area.

We have shown the intrinsic harmonic optical modulation inside OMO can effectively serve as a low-noise amplification mechanism. Basically, independent of the specific mechanical mode used for sensing the induced shift for the corresponding nth harmonic frequency will be n times larger than Δf0 for a given mass and position. In principle this approach can be applied to all kinds of OMO based mass sensors to increase the slope response and signal-to-noise ratio (SNR) for frequency shift measurement and therefore high-resolution mass sensing. We have identified three distinct operational modes of the microtoroidal OMO sensor for lumped and distributed mass detection/quantification. Our analysis shows using low noise lasers and detectors combined with advanced laser locking mechanisms, microtoroidal OMO can serve as a mass sensor with fg level resolution for metrological applications.

Although the ultimate sensitivity of the mass sensor presented here is lower than the record sensitivities obtained for state-of-the-art nano-scale cantilever mass sensors, the microtoroid OMO’s simplicity, all-optical excitation/read-out and operation in atmospheric pressure make it a useful tool for applications where ultra-low sensitivity is not required. MEMS/NEMS mass sensors are fundamentally limited by the complexity and speed of the excitation mechanism (piezo, electro-static or thermal). The transducers of such sensors require relatively complex co-fabrication of the hybrid material systems integrated with a high-Q mechanical resonator resulting in high cost [35

35. K. A. Snook, J. Z. Zhao, C. H. F. Alves, J. M. Cannata, W. H. Chen, R. J. Meyer Jr, T. A. Ritter, and K. K. Shung, “Design, fabrication, and evaluation of high frequency, single-element transducers incorporating different materials,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 49(2), 169–176 (2002). [CrossRef] [PubMed]

]. Furthermore, piezoelectric and electrostatic sensors may have a limitation when working in the presence of strong electromagnetic fields. In contrast OMO based sensors have a relatively simple structure and material system and their all-optical operation make them immune to electromagnetic interference (EMI). An additional concern is that nano/microscale cantilever mass sensors commonly operate using out-of-plane vibrations that facilitate the read-out (specifically piezo-resistive techniques that rely on bending induced strain). These modes are affected by viscous damping when operating at atmospheric pressure and therefore have considerably lowered quality factors (approximately by factor of 10 or more; below 1000) [36

36. B. Ilic, Y. Yang, and H. G. Craighead, “Virus detection using nanoelectromechanical devices,” Appl. Phys. Lett. 85(13), 2604–2606 (2004). [CrossRef]

38

38. N. Kacem, J. Arcamone, F. Perez-Murano, and S. Hentz, “Dynamic range enhancement of nonlinear nanomechanical resonant cantilevers for highly sensitive NEMS gas/mass sensor applications,” J. Micromech. Microeng. 20(4), 045023 (2010). [CrossRef]

]. However the displacements of microtoroid OMO based sensors are mostly in-plane and as such display higher mechanical quality factors, typically above 2000 [18

18. M. Hossein-Zadeh and K. J. Vahala, “Optomechanical Oscillator on a Silicon Chip”, Invited Paper, J Sel. Top. Quantum Electron. Special Issue on Silicon Photonics 16(1), 276–287 (2010).

], leading to a higher sensitivity at atmospheric pressure.

Acknowledgments

F. Liu thanks Yang Deng for help with microtoroid fabrication and SEM imaging. FL and MHZ’s effort on this work were in part supported by National Science Foundation (Grant#: ECCS-1055959) and Air Force Office of Research (Grant#: FA9550-12-1-0049). SA and ZCL’s effort on this work were supported by National Science Foundation (CMMI-1056077).

References and links

1.

A. Boisen, S. Dohn, S. S. Keller, S. Schmid, and M. Tenje, “Cantilever-like micromechanical sensors,” Rep. Prog. Phys. 74(036101), 1–30 (2011).

2.

J. Tamayo, P. M. Kosaka, J. J. Ruz, A. San Paulo, and M. Calleja, “Biosensors based on nanomechanical systems,” Chem. Soc. Rev. 42(3), 1287–1311 (2013). [CrossRef] [PubMed]

3.

P. S. Waggoner and H. G. Craighead, “Micro- and nanomechanical sensors for environmental, chemical, and biological detection,” Lab Chip 7(10), 1238–1255 (2007). [CrossRef] [PubMed]

4.

N. V. Lavrik, M. J. Sepaniak, and P. G. Datskos, “Cantilever transducers as a platform for chemical and biological sensors,” Rev. Sci. Instrum. 75(7), 2229–2253 (2004). [CrossRef]

5.

N. V. Lavrik and P. G. Datskos, “Femtogram mass detection using photothermally actuated nanomechanical resonators,” Appl. Phys. Lett. 82(16), 2697–2699 (2003). [CrossRef]

6.

A. Cagliani and Z. J. Davis, “Ultrasensitive bulk disk microresonator-based sensor for distributed mass sensing,” J. Micromech. Microeng. 21(045016), 1–6 (2011).

7.

M. W. Pruessner, T. H. Stievater, M. S. Ferraro, W. S. Rabinovich, J. L. Stepnowski, and R. A. McGill, “Waveguide micro-opto-electro-mechanical resonant chemical sensors,” Lab Chip 10(6), 762–768 (2010). [CrossRef] [PubMed]

8.

M. Li, H. X. Tang, and M. L. Roukes, “Ultra-sensitive NEMS-based cantilevers for sensing, scanned probe and very high-frequency applications,” Nat. Nanotechnol. 2(2), 114–120 (2007). [CrossRef] [PubMed]

9.

M. A. Taylor, A. Szorkovszky, J. Knittel, K. H. Lee, T. G. McRae, and W. P. Bowen, “Cavity optoelectromechanical regenerative amplification,” Opt. Express 20(12), 12742–12751 (2012). [CrossRef] [PubMed]

10.

Y. H. Tao, X. X. Li, T. G. Xu, H. T. Yu, P. C. Xu, B. Xiong, and C. Z. Wei, “Resonant cantilever sensors operated in a high-Q in-plane mode for real-time bio/chemical detection in liquids,” Sens. Actuators B Chem. 157(2), 606–614 (2011). [CrossRef]

11.

Y. Lee, G. Lim, and W. Moon, “A self-excited micro cantilever biosensor actuated by PZT using the mass micro balancing technique,” Sens. Actuators A Phys. 130-131, 105–110 (2006). [CrossRef]

12.

J. W. Yi, W. Y. Shih, and W.-H. Shih, “Effect of length, width, and mode on the mass detection sensitivity of piezoelectric unimorph cantilevers,” J. Appl. Phys. 91(3), 1680–1686 (2002). [CrossRef]

13.

F. Liu and M. Hossein-Zadeh, “Mass Sensing with Optomechanical Oscillation,” IEEE Sensors 13(1), 146–147 (2013). [CrossRef]

14.

T. J. Kippenberg and K. J. Vahala, “Cavity Opto-Mechanics,” Opt. Express 15(25), 17172–17205 (2007). [CrossRef] [PubMed]

15.

H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Radiation-pressure-driven micro-mechanical oscillator,” Opt. Express 13(14), 5293–5301 (2005). [CrossRef] [PubMed]

16.

M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A 74(2), 023813 (2006). [CrossRef]

17.

H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Theoretical and Experimental Study of Radiation Pressure-Induced Mechanical Oscillations (Parametric Instability) in Optical Microcavities,” IEEE J. Sel. Top. Quantum Electron. 12(1), 96–107 (2006).

18.

M. Hossein-Zadeh and K. J. Vahala, “Optomechanical Oscillator on a Silicon Chip”, Invited Paper, J Sel. Top. Quantum Electron. Special Issue on Silicon Photonics 16(1), 276–287 (2010).

19.

H. Rokhsari, M. Hossein-Zadeh, A. Hajimiri, and K. J. Vahala, “Brownian noise in radiation-pressure-driven micromechanical oscillators,” Appl. Phys. Lett. 89(6), 261109 (2006).

20.

F. Vollmer and S. Arnold, “Whispering-gallery-mode biosensing: label-free detection down to single molecules,” Nat. Methods 5(7), 591–596 (2008). [CrossRef] [PubMed]

21.

W. C. Jiang, X. Y. Lu, J. D. Zhang, and Q. Lin, “High-frequency silicon optomechanical oscillator with an ultralow threshold,” Opt. Express 20(14), 15991–15996 (2012). [CrossRef] [PubMed]

22.

S. Tallur, S. Sridaran, and S. A. Bhave, “A monolithic radiation-pressure driven, low phase noise silicon nitride opto-mechanical oscillator,” Opt. Express 19(24), 24522–24529 (2011). [CrossRef] [PubMed]

23.

M. Eichenfield, R. Camacho, J. Chan, K. J. Vahala, and O. Painter, “A picogram- and nanometre-scale photonic-crystal optomechanical cavity,” Nature 459(7246), 550–555 (2009). [CrossRef] [PubMed]

24.

M. Li, W. H. P. Pernice, and H. X. Tang, “Ultra-high-frequency nano-optomechanical resonators in slot waveguide ring cavities,” Appl. Phys. Lett. 97(18), 183110 (2010).

25.

M. W. Pruessner, T. H. Stievater, J. B. Khurgin, and W. S. Rabinovich, “Integrated waveguide-DBR microcavity opto-mechanical system,” Opt. Express 19(22), 21904–21918 (2011). [CrossRef] [PubMed]

26.

A. Schliesser, G. Anetsberger, R. Riviere, O. Arcizet, and T. J. Kippenberg, “High-sensitivity monitoring of micromechanical vibration using optical whispering gallery mode resonators,” New J. Phys. 10(9), 095015 (2008). [CrossRef]

27.

D. K. Armani, T. J. Kippenberg, S. M. Spillane, and K. J. Vahala, “Ultra-high-Q toroid microcavity on a chip,” Nature 421(6926), 925–928 (2003). [CrossRef] [PubMed]

28.

F. Liu and M. Hossein-Zadeh, “On the spectrum of the radiation-pressure-driven optomechanical oscillator and its application in sensing,” Opt. Commun. 294, 338–343 (2013). [CrossRef]

29.

S. Dohn, W. Svendsen, A. Boisen, and O. Hansen, “Mass and position determination of attached particles on cantilever based mass sensors,” Rev. Sci. Instrum. 78(10), 103303 (2007). [CrossRef] [PubMed]

30.

S. Dohn, R. Sandberg, W. Svendsen, and A. Boisen, “Enhanced functionality of cantilever based mass sensors using higher modes,” Appl. Phys. Lett. 86(23), 233501 (2005). [CrossRef]

31.

S. Seel, R. Storz, G. Ruoso, J. Mlynek, and S. Schiller, “Cryogenic Optical Resonators: A New Tool for Laser Frequency Stabilization at the 1 Hz Level,” Phys. Rev. Lett. 78(25), 4741–4744 (1997). [CrossRef]

32.

T. W. Hansch and B. Couillaud, “Laser Frequency Stabilization by Polarization Spectroscopy of a Reflecting Reference Cavity,” Opt. Commun. 35(3), 441–444 (1980). [CrossRef]

33.

R. W. P. Drever, J. L. Hall, F. V. Kowalski, J. Hough, G. M. Ford, A. J. Munley, and H. Ward, “Laser Phase and Frequency Stabilization Using an Optical-Resonator,” Appl. Phys. B 31(2), 97–105 (1983). [CrossRef]

34.

G. Bahl, K.-H. Kim, W. Lee, J. Liu, X. Fan, and T. Carmon, “Brillouin cavity optomechanics with microfluidic devices,” Nat Commun 4, 1994 (2013). [CrossRef] [PubMed]

35.

K. A. Snook, J. Z. Zhao, C. H. F. Alves, J. M. Cannata, W. H. Chen, R. J. Meyer Jr, T. A. Ritter, and K. K. Shung, “Design, fabrication, and evaluation of high frequency, single-element transducers incorporating different materials,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 49(2), 169–176 (2002). [CrossRef] [PubMed]

36.

B. Ilic, Y. Yang, and H. G. Craighead, “Virus detection using nanoelectromechanical devices,” Appl. Phys. Lett. 85(13), 2604–2606 (2004). [CrossRef]

37.

R. Sandberg, K. Molhave, A. Boisen, and W. Svendsen, “Effect of gold coating on the Q-factor of a resonant cantilever,” J. Micromech. Microeng. 15(12), 2249–2253 (2005). [CrossRef]

38.

N. Kacem, J. Arcamone, F. Perez-Murano, and S. Hentz, “Dynamic range enhancement of nonlinear nanomechanical resonant cantilevers for highly sensitive NEMS gas/mass sensor applications,” J. Micromech. Microeng. 20(4), 045023 (2010). [CrossRef]

OCIS Codes
(130.6010) Integrated optics : Sensors
(140.4780) Lasers and laser optics : Optical resonators
(230.3990) Optical devices : Micro-optical devices
(230.4910) Optical devices : Oscillators
(280.1415) Remote sensing and sensors : Biological sensing and sensors

ToC Category:
Sensors

History
Original Manuscript: June 28, 2013
Revised Manuscript: July 30, 2013
Manuscript Accepted: August 1, 2013
Published: August 13, 2013

Virtual Issues
Vol. 8, Iss. 9 Virtual Journal for Biomedical Optics

Citation
Fenfei Liu, Seyedhamidreza Alaie, Zayd C. Leseman, and Mani Hossein-Zadeh, "Sub-pg mass sensing and measurement with an optomechanical oscillator," Opt. Express 21, 19555-19567 (2013)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-21-17-19555


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. A. Boisen, S. Dohn, S. S. Keller, S. Schmid, and M. Tenje, “Cantilever-like micromechanical sensors,” Rep. Prog. Phys.74(036101), 1–30 (2011).
  2. J. Tamayo, P. M. Kosaka, J. J. Ruz, A. San Paulo, and M. Calleja, “Biosensors based on nanomechanical systems,” Chem. Soc. Rev.42(3), 1287–1311 (2013). [CrossRef] [PubMed]
  3. P. S. Waggoner and H. G. Craighead, “Micro- and nanomechanical sensors for environmental, chemical, and biological detection,” Lab Chip7(10), 1238–1255 (2007). [CrossRef] [PubMed]
  4. N. V. Lavrik, M. J. Sepaniak, and P. G. Datskos, “Cantilever transducers as a platform for chemical and biological sensors,” Rev. Sci. Instrum.75(7), 2229–2253 (2004). [CrossRef]
  5. N. V. Lavrik and P. G. Datskos, “Femtogram mass detection using photothermally actuated nanomechanical resonators,” Appl. Phys. Lett.82(16), 2697–2699 (2003). [CrossRef]
  6. A. Cagliani and Z. J. Davis, “Ultrasensitive bulk disk microresonator-based sensor for distributed mass sensing,” J. Micromech. Microeng.21(045016), 1–6 (2011).
  7. M. W. Pruessner, T. H. Stievater, M. S. Ferraro, W. S. Rabinovich, J. L. Stepnowski, and R. A. McGill, “Waveguide micro-opto-electro-mechanical resonant chemical sensors,” Lab Chip10(6), 762–768 (2010). [CrossRef] [PubMed]
  8. M. Li, H. X. Tang, and M. L. Roukes, “Ultra-sensitive NEMS-based cantilevers for sensing, scanned probe and very high-frequency applications,” Nat. Nanotechnol.2(2), 114–120 (2007). [CrossRef] [PubMed]
  9. M. A. Taylor, A. Szorkovszky, J. Knittel, K. H. Lee, T. G. McRae, and W. P. Bowen, “Cavity optoelectromechanical regenerative amplification,” Opt. Express20(12), 12742–12751 (2012). [CrossRef] [PubMed]
  10. Y. H. Tao, X. X. Li, T. G. Xu, H. T. Yu, P. C. Xu, B. Xiong, and C. Z. Wei, “Resonant cantilever sensors operated in a high-Q in-plane mode for real-time bio/chemical detection in liquids,” Sens. Actuators B Chem.157(2), 606–614 (2011). [CrossRef]
  11. Y. Lee, G. Lim, and W. Moon, “A self-excited micro cantilever biosensor actuated by PZT using the mass micro balancing technique,” Sens. Actuators A Phys.130-131, 105–110 (2006). [CrossRef]
  12. J. W. Yi, W. Y. Shih, and W.-H. Shih, “Effect of length, width, and mode on the mass detection sensitivity of piezoelectric unimorph cantilevers,” J. Appl. Phys.91(3), 1680–1686 (2002). [CrossRef]
  13. F. Liu and M. Hossein-Zadeh, “Mass Sensing with Optomechanical Oscillation,” IEEE Sensors13(1), 146–147 (2013). [CrossRef]
  14. T. J. Kippenberg and K. J. Vahala, “Cavity Opto-Mechanics,” Opt. Express15(25), 17172–17205 (2007). [CrossRef] [PubMed]
  15. H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Radiation-pressure-driven micro-mechanical oscillator,” Opt. Express13(14), 5293–5301 (2005). [CrossRef] [PubMed]
  16. M. Hossein-Zadeh, H. Rokhsari, A. Hajimiri, and K. J. Vahala, “Characterization of a radiation-pressure-driven micromechanical oscillator,” Phys. Rev. A74(2), 023813 (2006). [CrossRef]
  17. H. Rokhsari, T. J. Kippenberg, T. Carmon, and K. J. Vahala, “Theoretical and Experimental Study of Radiation Pressure-Induced Mechanical Oscillations (Parametric Instability) in Optical Microcavities,” IEEE J. Sel. Top. Quantum Electron.12(1), 96–107 (2006).
  18. M. Hossein-Zadeh and K. J. Vahala, “Optomechanical Oscillator on a Silicon Chip”, Invited Paper, J Sel. Top. Quantum Electron. Special Issue on Silicon Photonics16(1), 276–287 (2010).
  19. H. Rokhsari, M. Hossein-Zadeh, A. Hajimiri, and K. J. Vahala, “Brownian noise in radiation-pressure-driven micromechanical oscillators,” Appl. Phys. Lett.89(6), 261109 (2006).
  20. F. Vollmer and S. Arnold, “Whispering-gallery-mode biosensing: label-free detection down to single molecules,” Nat. Methods5(7), 591–596 (2008). [CrossRef] [PubMed]
  21. W. C. Jiang, X. Y. Lu, J. D. Zhang, and Q. Lin, “High-frequency silicon optomechanical oscillator with an ultralow threshold,” Opt. Express20(14), 15991–15996 (2012). [CrossRef] [PubMed]
  22. S. Tallur, S. Sridaran, and S. A. Bhave, “A monolithic radiation-pressure driven, low phase noise silicon nitride opto-mechanical oscillator,” Opt. Express19(24), 24522–24529 (2011). [CrossRef] [PubMed]
  23. M. Eichenfield, R. Camacho, J. Chan, K. J. Vahala, and O. Painter, “A picogram- and nanometre-scale photonic-crystal optomechanical cavity,” Nature459(7246), 550–555 (2009). [CrossRef] [PubMed]
  24. M. Li, W. H. P. Pernice, and H. X. Tang, “Ultra-high-frequency nano-optomechanical resonators in slot waveguide ring cavities,” Appl. Phys. Lett.97(18), 183110 (2010).
  25. M. W. Pruessner, T. H. Stievater, J. B. Khurgin, and W. S. Rabinovich, “Integrated waveguide-DBR microcavity opto-mechanical system,” Opt. Express19(22), 21904–21918 (2011). [CrossRef] [PubMed]
  26. A. Schliesser, G. Anetsberger, R. Riviere, O. Arcizet, and T. J. Kippenberg, “High-sensitivity monitoring of micromechanical vibration using optical whispering gallery mode resonators,” New J. Phys.10(9), 095015 (2008). [CrossRef]
  27. D. K. Armani, T. J. Kippenberg, S. M. Spillane, and K. J. Vahala, “Ultra-high-Q toroid microcavity on a chip,” Nature421(6926), 925–928 (2003). [CrossRef] [PubMed]
  28. F. Liu and M. Hossein-Zadeh, “On the spectrum of the radiation-pressure-driven optomechanical oscillator and its application in sensing,” Opt. Commun.294, 338–343 (2013). [CrossRef]
  29. S. Dohn, W. Svendsen, A. Boisen, and O. Hansen, “Mass and position determination of attached particles on cantilever based mass sensors,” Rev. Sci. Instrum.78(10), 103303 (2007). [CrossRef] [PubMed]
  30. S. Dohn, R. Sandberg, W. Svendsen, and A. Boisen, “Enhanced functionality of cantilever based mass sensors using higher modes,” Appl. Phys. Lett.86(23), 233501 (2005). [CrossRef]
  31. S. Seel, R. Storz, G. Ruoso, J. Mlynek, and S. Schiller, “Cryogenic Optical Resonators: A New Tool for Laser Frequency Stabilization at the 1 Hz Level,” Phys. Rev. Lett.78(25), 4741–4744 (1997). [CrossRef]
  32. T. W. Hansch and B. Couillaud, “Laser Frequency Stabilization by Polarization Spectroscopy of a Reflecting Reference Cavity,” Opt. Commun.35(3), 441–444 (1980). [CrossRef]
  33. R. W. P. Drever, J. L. Hall, F. V. Kowalski, J. Hough, G. M. Ford, A. J. Munley, and H. Ward, “Laser Phase and Frequency Stabilization Using an Optical-Resonator,” Appl. Phys. B31(2), 97–105 (1983). [CrossRef]
  34. G. Bahl, K.-H. Kim, W. Lee, J. Liu, X. Fan, and T. Carmon, “Brillouin cavity optomechanics with microfluidic devices,” Nat Commun4, 1994 (2013). [CrossRef] [PubMed]
  35. K. A. Snook, J. Z. Zhao, C. H. F. Alves, J. M. Cannata, W. H. Chen, R. J. Meyer, T. A. Ritter, and K. K. Shung, “Design, fabrication, and evaluation of high frequency, single-element transducers incorporating different materials,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control49(2), 169–176 (2002). [CrossRef] [PubMed]
  36. B. Ilic, Y. Yang, and H. G. Craighead, “Virus detection using nanoelectromechanical devices,” Appl. Phys. Lett.85(13), 2604–2606 (2004). [CrossRef]
  37. R. Sandberg, K. Molhave, A. Boisen, and W. Svendsen, “Effect of gold coating on the Q-factor of a resonant cantilever,” J. Micromech. Microeng.15(12), 2249–2253 (2005). [CrossRef]
  38. N. Kacem, J. Arcamone, F. Perez-Murano, and S. Hentz, “Dynamic range enhancement of nonlinear nanomechanical resonant cantilevers for highly sensitive NEMS gas/mass sensor applications,” J. Micromech. Microeng.20(4), 045023 (2010). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.