High-sensitive and broad-dynamic-range quantitative phase imaging with spectral domain phase microscopy

Yangzhi Yan, Zhihua Ding, Yi Shen, Zhiyan Chen, Chen Zhao, and Yang Ni  »View Author Affiliations


Optics Express, Vol. 21, Issue 22, pp. 25734-25743 (2013)
http://dx.doi.org/10.1364/OE.21.025734


View Full Text Article

Acrobat PDF (2091 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Spectral domain phase microscopy for high-sensitive and broad-dynamic-range quantitative phase imaging is presented. The phase retrieval is realized in the depth domain to maintain a high sensitivity, while the phase information obtained in the spectral domain is exploited to extend the dynamic range of optical path difference. Sensitivity advantage of phase retrieved in the depth domain over that in the spectral domain is thoroughly investigated. The performance of the proposed depth domain phase based approach is illustrated by phase imaging of a resolution target and an onion skin.

© 2013 Optical Society of America

1. Introduction

Spectral domain optical coherence tomography (SD-OCT) [1

1. R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express 11(8), 889–894 (2003). [CrossRef] [PubMed]

4

4. M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett. 28(19), 1745–1747 (2003). [CrossRef] [PubMed]

] provides high-sensitive cross-sectional images of scattering tissues with axial resolution on the scale of several to tens of microns. It detects spectral interferograms as a function of optical frequency and produces A-scan data by applying Fourier transformation to the acquired interference spectrum. Spectral domain phase microscopy (SDPM) [5

5. M. A. Choma, A. K. Ellerbee, C. H. Yang, T. L. Creazzo, and J. A. Izatt, “Spectral-domain phase microscopy,” Opt. Lett. 30(10), 1162–1164 (2005). [CrossRef] [PubMed]

9

9. Z. Yaqoob, W. Choi, S. Oh, N. Lue, Y. Park, C. Fang-Yen, R. R. Dasari, K. Badizadegan, and M. S. Feld, “Improved phase sensitivity in spectral domain phase microscopy using line-field illumination and self phase-referencing,” Opt. Express 17(13), 10681–10687 (2009). [CrossRef] [PubMed]

] is a functional extension of SD-OCT. By employing common-path configuration and phase-sensitive measurement, the sensitivity of the axial displacement measurement by SDPM greatly exceeds the axial resolution of the SD-OCT system, allowing for various applications such as material inspection [10

10. Y. Verma, P. Nandi, K. D. Rao, M. Sharma, and P. K. Gupta, “Use of common path phase sensitive spectral domain optical coherence tomography for refractive index measurements,” Appl. Opt. 50(25), E7–E12 (2011). [CrossRef]

], cellular imaging [11

11. M. V. Sarunic, S. Weinberg, and J. A. Izatt, “Full-field swept-source phase microscopy,” Opt. Lett. 31(10), 1462–1464 (2006). [CrossRef] [PubMed]

, 12

12. P. Wang, R. Bista, R. Bhargava, R. E. Brand, and Y. Liu, “Spatial-domain low-coherence quantitative phase microscopy for cancer diagnosis,” Opt. Lett. 35(17), 2840–2842 (2010). [CrossRef] [PubMed]

], and Doppler flow measurements [13

13. B. J. Vakoc, S. H. Yun, J. F. de Boer, G. J. Tearney, and B. E. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13(14), 5483–5493 (2005). [CrossRef] [PubMed]

, 14

14. M. A. Choma, A. K. Ellerbee, S. Yazdanfar, and J. A. Izatt, “Doppler flow imaging of cytoplasmic streaming using spectral domain phase microscopy,” J. Biomed. Opt. 11(2), 024014 (2006). [CrossRef] [PubMed]

]. In conventional SDPM, the phase due to optical path difference (OPD) between the reference arm and sample arm is extracted in the depth domain. The range of OPD to be measured is restricted to less than half of the source center wavelength due to the well-known 2π ambiguity. To extend the OPD range of SDPM, different phase unwrapping algorithms have been proposed. Conventional phase unwrapping algorithm [15

15. T. J. Flynn, “Two-dimensional phase unwrapping with minimum weighted discontinuity,” J. Opt. Soc. Am. A 14(10), 2692–2701 (1997). [CrossRef]

] requires the corresponding OPD along neighboring sampling points to vary gradually, and fails if multiple wrapping of phase occurs. Synthetic wavelength phase unwrapping techniques [16

16. D. Parshall and M. K. Kim, “Digital holographic microscopy with dual-wavelength phase unwrapping,” Appl. Opt. 45(3), 451–459 (2006). [CrossRef] [PubMed]

18

18. M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, and A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett. 35(15), 2612–2614 (2010). [CrossRef] [PubMed]

] use two or more wavelengths to increase the OPD range without phase unwrapping at the price of increased phase noises. Instead of phase retrieval in the depth domain, phase retrieval in the spectral domain is proposed to extend the OPD range [19

19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

22

22. C. Wang, Z. H. Ding, S. T. Mei, H. Yu, W. Hong, Y. Z. Yan, and W. D. Shen, “Ultralong-range phase imaging with orthogonal dispersive spectral-domain optical coherence tomography,” Opt. Lett. 37(21), 4555–4557 (2012). [CrossRef] [PubMed]

]. However, the phase sensitivity is degraded since the relevant signals are distributed over a broadband of wavenumbers in the spectral domain in contrast to localization in depth domain, which leads to a reduced signal-to-noise ratio (SNR). Although by averaging spectral phases over multiple wavenumbers can significantly enhance the phase sensitivity [23

23. Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun. 303, 21–24 (2013). [CrossRef]

], the complicated spectral phase retrieval algorithm is time-consuming and error prone.

In this paper, we introduce a high-sensitive and broad-dynamic-range quantitative phase imaging method for SDPM. The phase retrieval is still realized in the depth domain to maintain a high sensitivity, while the phase information obtained in the spectral domain is further used for removing the phase ambiguity to extend the dynamic range of SDPM. The phase sensitivity in the spectral domain and in the depth domain will be compared theoretically and experimentally. Phase imaging of a resolution target and an onion skin are presented to evaluate the performance of the proposed phase imaging method.

2. Theory

2.1 Conventional depth domain phase based SDPM

Excluding autocorrelation and DC terms, the detected interference spectra of the return lights from a sample reflector and a reference reflector can be described as [2

2. M. A. Choma, M. V. Sarunic, C. H. Yang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]

, 19

19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

, 23

23. Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun. 303, 21–24 (2013). [CrossRef]

]
I(ki)=2S(kik0)RRRScos(2z0ki)+α(ki).
(1)
Here ki is the discrete wavenumber (i=1~N) in the spectral domain; k0 is the central wavenumber; S(kik0) represents power spectral density function of the light source; RR and RS are reflectivities of the sample reflector and the reference reflector, respectively; z0 denotes the OPD between the reference reflector and the sample reflector; α(ki) is the additive Gaussian white noise with a standard deviation of σk and a mean value of zero due to excluding of the DC term [1

1. R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express 11(8), 889–894 (2003). [CrossRef] [PubMed]

, 2

2. M. A. Choma, M. V. Sarunic, C. H. Yang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]

]. Conventional SDPM applies the fast Fourier transform (FFT) on Eq. (1) to get the depth signal as
I(zi)=RRRSΓ(ziz0)exp[j2k0(ziz0)]+α(zi),
(2)
here the coherent ghost image is omitted for simplicity; j is the basic imaginary unit sqrt(−1); zi is the discretely sampled depth position; Γ(ziz0)=F[S(ki)]δ(ziz0) with F[] denotes Fourier transformation; α(zi) is the Fourier Transform Spectrum of α(ki), it is the complex noise in the depth domain corresponding to α(ki) in the spectral domain. The standard deviation of α(zi) is σz.

The peak value in the depth domain will appear at zi=z0 if assuming a symmetry distribution of S(kik0) around its central wavenumber k0. However, the spectral density of the light source is usually not of an ideal Gaussian form but with asymmetry. The discretely sampled peak value in the depth domain might be at a position inconsistent with z0. Assuming a deviation of δzfrom z0, the nearest peaked value at sampling point zi=z0δzin the depth domain is then given by
I(z0δz)=RRRSΓ(δz)exp(j2k0δz)+α(z0δz).
(3)
The phase term φ(z0δz) of I(z0δz) is 2k0δz+Δφzi, here Δφzi denotes the equivalent phase uncertainty caused by the noise. Obviously, φ(z0δz) is a linear function of δz. Sub-resolution changes in OPD between the reference reflector and the sample reflector can be sensitively detected by measurement of this phase term. Unfortunately, the achieved phase from Eq. (3) is limited to the range of π to +π and can be written by
φw(z0δz)=φ(z0δz)2πfloor(φ(z0δz)+π2π),
(4)
here floor means round toward negative infinity. A problem known as 2π ambiguity occurs when phase change between consecutive measurement falls outside the range [π,+π], meaning a limitation of the dynamic OPD range less than half a central wavelength.

To extend the dynamic range of conventional depth domain phase based SDPM, synthetic wavelength based method is introduced. By windowing the signal spectrum into multiple spectra before applying the Fourier transformation, phase information at multiple center wavelengths are obtained and used for phase retrieval corresponding to a longer synthetic central wavelength. The dynamic range is thus extended but the sensitivity is degraded [16

16. D. Parshall and M. K. Kim, “Digital holographic microscopy with dual-wavelength phase unwrapping,” Appl. Opt. 45(3), 451–459 (2006). [CrossRef] [PubMed]

18

18. M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, and A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett. 35(15), 2612–2614 (2010). [CrossRef] [PubMed]

].

2.2 Spectral domain phase based SDPM

To measure OPD longer than half a central wavelength, spectral domain phase based SDPM [19

19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

, 20

20. E. D. Moore and R. R. McLeod, “Phase-sensitive swept-source interferometry for absolute ranging with application to measurements of group refractive index and thickness,” Opt. Express 19(9), 8117–8126 (2011). [CrossRef] [PubMed]

] is proposed. By Hilbert transformation of Eq. (1), the complex spectral signal is given by
I˜(ki)=S(kik0)RRRSexp(j2kiz0)+α˜(ki),
(5)
where α˜(ki) is the complex spectral noise. The standard deviation of α˜(ki) is σk. The phase term φ(ki) of I˜(ki) is 2kiz0+Δφki, here Δφki is the equivalent phase uncertainty due to the spectral noise. The wrapped phase φw(ki) can be obtained from the complex spectral signal described by Eq. (5) as [23

23. Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun. 303, 21–24 (2013). [CrossRef]

]
φw(ki)=φ(ki)2πfloor(φ(ki)+π2π).
(6)
By phase unwrapping and least-square fitting in the spectral domain [19

19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

, 20

20. E. D. Moore and R. R. McLeod, “Phase-sensitive swept-source interferometry for absolute ranging with application to measurements of group refractive index and thickness,” Opt. Express 19(9), 8117–8126 (2011). [CrossRef] [PubMed]

], an estimated phase of φ(ki) is then obtained by
φ(ki)=2kiz0=2kiz0+Δφki.
(7)
Here z0 denotes the fitted slope of the unwrapped phase versus wavenumber, Δφki denotes the phase errors in φ(ki). Since φ(ki) suffers from an amplified phase error [19

19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

], i.e. Δφki>Δφki, the estimated phase is only used for removal of 2π ambiguity of the phase in the spectral domain as [23

23. Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun. 303, 21–24 (2013). [CrossRef]

]
φ(ki)=φw(ki)+2πfloor(φ(ki)+π2π).
(8)
Unlike the algorithm presented in reference [19

19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

], addition of π is made here for synchronizing of phase jumps with opposite direction in both terms of the right side of Eq. (8) [23

23. Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun. 303, 21–24 (2013). [CrossRef]

]. This spectral domain phase based approach is capable of broad-dynamic-range phase imaging of phase specimen. However, its phase sensitivity is degraded in compared with that achieved in the depth domain and this will be thoroughly investigated in the following section.

2.3 A comparison of phase sensitivity

Phase sensitivity in SDPM is determined by SNR. A higher SNR means a higher phase sensitivity and hence a lower phase uncertainty. The phase uncertainty in SDPM can be expressed by [5

5. M. A. Choma, A. K. Ellerbee, C. H. Yang, T. L. Creazzo, and J. A. Izatt, “Spectral-domain phase microscopy,” Opt. Lett. 30(10), 1162–1164 (2005). [CrossRef] [PubMed]

]
Δφ=arctan(1SNR)1SNR=|σnoise||Isample|,
(9)
where σnoise denotes the noise component; Isample denotes the signal. To compare the phase sensitivity in the depth domain and that in the spectral domain, the SNR in both domains should be investigated.

To achieve the highest phase sensitivity, the phase information should be retrieved at their peak values in both domains. Assume a Gaussian distribution S(ki)=exp(4ln2(kik0)2Δk2) and FWHM bandwidth of Δk for the light source, the peak values of signal in the depth domain and spectral domain can be obtained from Eqs. (2) and (5) to be Imax(zi)=ΔkπRRRS2δkln2and I˜max(ki)=2RRRS, respectively. Here δk is the spectral resolution of the spectrometer in the SDPM. The peak signal ratio between depth domain and spectral domain is given by

|Imax(zi)||I˜max(ki)|=πΔk4ln2δk.
(10)

According to Parseval's theorem of discrete Fourier transformation, the noise variance ratio between depth domain and spectral domain is given by
σz2σk2=N.
(11)
N is the number of discrete wavenumbers in the spectral domain. The ratio of SNR between the depth domain and the spectral domain is thus deduced to be
SNRzSNRk=π16ln21N(Δkδk)2.
(12)
Using Eq. (9), the ratio of phase uncertaintyΔφz/Δφk between the depth domain and the spectral domain is determined by
ΔφzΔφk=4ln2πNδkΔk=4ln2πNNδkΔk.
(13)
Obviously, Δφz/Δφk is inversely proportional to square root of N since Nδk is comparable to Δk, and the phase sensitivity in the depth domain should be much better than that in the spectral domain as N>>1.

2.4 The proposed depth domain phase based SDPM

As discussed previously, the conventional depth domain phase based SDPM has the advantage in phase sensitivity and the disadvantage in limited dynamic range. To achieve high dynamic range while keeping high sensitivity, we propose a new depth domain phase retrieval method, in which phases in both domains are fully exploited.

As discussed in section 2.3, to achieve the highest phase sensitivity, the phase information should be retrieved at the peak values in depth domain. Due to asymmetry of spectral density distribution of the light source and discrete sampling, the peak value of the response profile in depth domain usually cannot be sampled. We take the sampling point zi=z0δz with a maximum value as the estimated peak point. At the estimated peak point, the corresponding wrapped phase is φw(z0δz). Now we introduce how to retrieve φ(z0δz) based on the spectral phase φ(ki) and the wrapped phase φw(z0δz).

The spectral phase at the central wavenumber φ(k0) can be determined by Eq. (8), which is then used to quantify an estimated phase of φ(z0δz) in the depth domain as
φ(z0δz)=φ(k0)2k0(z0δz)=2k0δz±Δφk0.
(14)
Since φ(z0δz) suffers from an amplified phase error, it is only used for removal of 2π ambiguity of the phase in the depth domain as
φ(z0δz)=φw(z0δz)+2πfloor(φ(z0δz)+π2π).
(15)
With the location information of the sample point in the depth domain where the phase is retrieved, the absolute OPD between the reference reflector and the sample reflector is thus determined by
OPD=z0δz+φ(z0δz)2k0.
(16)
Compared with the conventional depth domain phase based SDPM, this proposed depth domain phase retrieval method is immune of phase ambiguity.

3. Experiment

3.1 Coverslip

To illustrate the enhanced phase sensitivity of the proposed depth domain phase based SDPM over that achieved by the spectral domain phase based SDPM, 1024 A-scans of a stationary coverslip with nominal optical thickness of 210 ± 35 µm are performed. The OPDs obtained by the spectral domain phase based SDPM and the proposed depth domain phase based SDPM are shown in Fig. 3(a)
Fig. 3 (a) OPDs corresponding to 1024 repeated A-scans: by the spectral domain phase based SDPM (blue circles); by multiple spectral phases averaged SDPM (green diamonds); by the proposed depth domain phase based SDPM (red crosses). (b) The probability distribution of the measured OPD by spectral domain phase based SDPM. (c) The probability distribution of the measured OPD by multiple spectral phases averaged SDPM. (d) The probability distribution of the measured OPD by the proposed depth domain phase based SDPM.
. Evidently, the OPD measured by the proposed depth domain phase based SDPM is much more stable than that measured by the spectral domain phase based SDPM. The probability distribution of the measured OPD by spectral domain phase based SDPM and by the proposed depth domain phase based SDPM are shown in Figs. 3(b) and 3(d), respectively. The standard variation (STD) of the former one is 161.81 pm while the STD of the latter one is 21.41 pm. The phase sensitivity of the proposed depth domain phase based SDPM is 7.56 fold higher than that of the spectral domain phase based SDPM, which is in good agreement with the theoretical expectation according to Eq. (13).

3.2 Resolution target

To validate the ability of the proposed depth domain phase based SDPM for broad-dynamic-range OPD measurement, a resolution target (1951 USAF, Newport) is adopted as the sample. The sample configuration is shown in Fig. 4(a)
Fig. 4 Measurement of the OPD between coverslip and resolution target. (a) Sample configuration. (b) Quantitative phase image by conventional depth domain phase based SDPM. (c) Quantitative phase image by the proposed depth domain phase based SDPM. (d) Quantitative phase image by the spectral domain phase based SDPM.
where an optical fiber with diameter of 245 ± 5 µm is placed at one end between the coverslip and the resolution target to form a wedge realizing a large dynamic range of OPDs. Interference spectra of the return lights from the top surface of the coverslip and the topological surface of the resolution target are collected and used to measure the OPDs between these two surfaces. Figure 4(b) demonstrates the quantitative phase obtained by conventional depth domain phase based SDPM, where the measurement range is limited by half of the central wavenumber and the patterns on the resolution target are corrupted by phase ambiguity. Figure 4(c) shows the quantitative phase image obtained by the proposed depth domain phase based SDPM, where the wedge-shaped OPD-variations are well reconstructed. The patterns on the resolution target can be clearly resolved from the phase image. The height of the pattern is measured to be 120 nm, which is consistent with the reported value in the literatures [11

11. M. V. Sarunic, S. Weinberg, and J. A. Izatt, “Full-field swept-source phase microscopy,” Opt. Lett. 31(10), 1462–1464 (2006). [CrossRef] [PubMed]

, 12

12. P. Wang, R. Bista, R. Bhargava, R. E. Brand, and Y. Liu, “Spatial-domain low-coherence quantitative phase microscopy for cancer diagnosis,” Opt. Lett. 35(17), 2840–2842 (2010). [CrossRef] [PubMed]

]. The quantitative phase image obtained by the spectral domain phase based SDPM is also shown in Fig. 4(d) for comparison. However, due to the large dynamic range of OPD, it is hard to notice the difference between the phase images demonstrated in Figs. 4(c) and Fig. 4(d).

3.3 Single layer onion skin

To demonstrate the feasibility of the proposed depth domain phase based SDPM for biological sample, phase imaging of a single layer of onion skin is done. Figure 5(a)
Fig. 5 (a) Image of onion skin cells by optical microscope. The scale bar indicates 30 μm. (b) Reconstructed phase image of onion skin cells by the proposed depth domain phase based SDPM.
is the image recorded by an optical microscope showing the onion skin cells. The reconstructed phase image is presented in Fig. 5(b), where ellipse shape of onion skin cell (about 30 μm to 125 μm in lateral directions and 30 μm in depth direction) is clearly resolved.

4. Conclusion

By fully exploitation of phase information in the depth domain and the spectral domain, SDPM capable of high-sensitive and broad-dynamic-range quantitative phase imaging is presented. Sensitivity advantage of phase retrieved in the depth domain over that in the spectral domain is confirmed both theoretically and experimentally. Phase ambiguity in the depth domain is overcome by the proposed depth domain phase retrieval method, in which the unwrapped phase in the spectral domain and the location of the sample point in the depth domain is used to extend the dynamic range of SDPM. The performance of the proposed depth domain phase based SDPM with extended dynamic range is illustrated by phase imaging of a resolution target and an onion skin. Further application of this technique could be the study of dynamical phenomenon in biological sample, such as optical measurement of nerve activations.

Acknowledgments

The authors would like to acknowledge the financial supports from the Chinese Natural Science Foundation (61275196, 61335003, 61327007) and Zhejiang Province Science and Technology Grant (2012C33031).

References and links

1.

R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express 11(8), 889–894 (2003). [CrossRef] [PubMed]

2.

M. A. Choma, M. V. Sarunic, C. H. Yang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003). [CrossRef] [PubMed]

3.

J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28(21), 2067–2069 (2003). [CrossRef] [PubMed]

4.

M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett. 28(19), 1745–1747 (2003). [CrossRef] [PubMed]

5.

M. A. Choma, A. K. Ellerbee, C. H. Yang, T. L. Creazzo, and J. A. Izatt, “Spectral-domain phase microscopy,” Opt. Lett. 30(10), 1162–1164 (2005). [CrossRef] [PubMed]

6.

C. Joo, T. Akkin, B. Cense, B. H. Park, and J. F. de Boer, “Spectral-domain optical coherence phase microscopy for quantitative phase-contrast imaging,” Opt. Lett. 30(16), 2131–2133 (2005). [CrossRef] [PubMed]

7.

D. C. Adler, R. Huber, and J. G. Fujimoto, “Phase-sensitive optical coherence tomography at up to 370,000 lines per second using buffered Fourier domain mode-locked lasers,” Opt. Lett. 32(6), 626–628 (2007). [CrossRef] [PubMed]

8.

K. Singh, C. Dion, M. R. Lesk, T. Ozaki, and S. Costantino, “Spectral-domain phase microscopy with improved sensitivity using two-dimensional detector arrays,” Rev. Sci. Instrum. 82(2), 023706 (2011). [CrossRef] [PubMed]

9.

Z. Yaqoob, W. Choi, S. Oh, N. Lue, Y. Park, C. Fang-Yen, R. R. Dasari, K. Badizadegan, and M. S. Feld, “Improved phase sensitivity in spectral domain phase microscopy using line-field illumination and self phase-referencing,” Opt. Express 17(13), 10681–10687 (2009). [CrossRef] [PubMed]

10.

Y. Verma, P. Nandi, K. D. Rao, M. Sharma, and P. K. Gupta, “Use of common path phase sensitive spectral domain optical coherence tomography for refractive index measurements,” Appl. Opt. 50(25), E7–E12 (2011). [CrossRef]

11.

M. V. Sarunic, S. Weinberg, and J. A. Izatt, “Full-field swept-source phase microscopy,” Opt. Lett. 31(10), 1462–1464 (2006). [CrossRef] [PubMed]

12.

P. Wang, R. Bista, R. Bhargava, R. E. Brand, and Y. Liu, “Spatial-domain low-coherence quantitative phase microscopy for cancer diagnosis,” Opt. Lett. 35(17), 2840–2842 (2010). [CrossRef] [PubMed]

13.

B. J. Vakoc, S. H. Yun, J. F. de Boer, G. J. Tearney, and B. E. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13(14), 5483–5493 (2005). [CrossRef] [PubMed]

14.

M. A. Choma, A. K. Ellerbee, S. Yazdanfar, and J. A. Izatt, “Doppler flow imaging of cytoplasmic streaming using spectral domain phase microscopy,” J. Biomed. Opt. 11(2), 024014 (2006). [CrossRef] [PubMed]

15.

T. J. Flynn, “Two-dimensional phase unwrapping with minimum weighted discontinuity,” J. Opt. Soc. Am. A 14(10), 2692–2701 (1997). [CrossRef]

16.

D. Parshall and M. K. Kim, “Digital holographic microscopy with dual-wavelength phase unwrapping,” Appl. Opt. 45(3), 451–459 (2006). [CrossRef] [PubMed]

17.

H. C. Hendargo, M. Zhao, N. Shepherd, and J. A. Izatt, “Synthetic wavelength based phase unwrapping in spectral domain optical coherence tomography,” Opt. Express 17(7), 5039–5051 (2009). [CrossRef] [PubMed]

18.

M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, and A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett. 35(15), 2612–2614 (2010). [CrossRef] [PubMed]

19.

J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett. 34(21), 3442–3444 (2009). [CrossRef] [PubMed]

20.

E. D. Moore and R. R. McLeod, “Phase-sensitive swept-source interferometry for absolute ranging with application to measurements of group refractive index and thickness,” Opt. Express 19(9), 8117–8126 (2011). [CrossRef] [PubMed]

21.

Y. Zhu, N. T. Shaked, L. L. Satterwhite, and A. Wax, “Spectral-domain differential interference contrast microscopy,” Opt. Lett. 36(4), 430–432 (2011). [CrossRef] [PubMed]

22.

C. Wang, Z. H. Ding, S. T. Mei, H. Yu, W. Hong, Y. Z. Yan, and W. D. Shen, “Ultralong-range phase imaging with orthogonal dispersive spectral-domain optical coherence tomography,” Opt. Lett. 37(21), 4555–4557 (2012). [CrossRef] [PubMed]

23.

Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun. 303, 21–24 (2013). [CrossRef]

24.

K. Wang, Z. H. Ding, T. Wu, C. Wang, J. Meng, M. H. Chen, and L. Xu, “Development of a non-uniform discrete Fourier transform based high speed spectral domain optical coherence tomography system,” Opt. Express 17(14), 12121–12131 (2009). [CrossRef] [PubMed]

25.

J. Meng, Z. H. Ding, J. Li, K. Wang, and T. Wu, “Transit-time analysis based on delay-encoded beam shape for velocity vector quantification by spectral-domain Doppler optical coherence tomography,” Opt. Express 18(2), 1261–1270 (2010). [CrossRef] [PubMed]

OCIS Codes
(100.5070) Image processing : Phase retrieval
(170.4500) Medical optics and biotechnology : Optical coherence tomography
(180.3170) Microscopy : Interference microscopy

ToC Category:
Microscopy

History
Original Manuscript: September 3, 2013
Revised Manuscript: October 11, 2013
Manuscript Accepted: October 11, 2013
Published: October 21, 2013

Virtual Issues
Vol. 9, Iss. 1 Virtual Journal for Biomedical Optics

Citation
Yangzhi Yan, Zhihua Ding, Yi Shen, Zhiyan Chen, Chen Zhao, and Yang Ni, "High-sensitive and broad-dynamic-range quantitative phase imaging with spectral domain phase microscopy," Opt. Express 21, 25734-25743 (2013)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-21-22-25734


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express11(8), 889–894 (2003). [CrossRef] [PubMed]
  2. M. A. Choma, M. V. Sarunic, C. H. Yang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express11(18), 2183–2189 (2003). [CrossRef] [PubMed]
  3. J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett.28(21), 2067–2069 (2003). [CrossRef] [PubMed]
  4. M. Wojtkowski, T. Bajraszewski, P. Targowski, and A. Kowalczyk, “Real-time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett.28(19), 1745–1747 (2003). [CrossRef] [PubMed]
  5. M. A. Choma, A. K. Ellerbee, C. H. Yang, T. L. Creazzo, and J. A. Izatt, “Spectral-domain phase microscopy,” Opt. Lett.30(10), 1162–1164 (2005). [CrossRef] [PubMed]
  6. C. Joo, T. Akkin, B. Cense, B. H. Park, and J. F. de Boer, “Spectral-domain optical coherence phase microscopy for quantitative phase-contrast imaging,” Opt. Lett.30(16), 2131–2133 (2005). [CrossRef] [PubMed]
  7. D. C. Adler, R. Huber, and J. G. Fujimoto, “Phase-sensitive optical coherence tomography at up to 370,000 lines per second using buffered Fourier domain mode-locked lasers,” Opt. Lett.32(6), 626–628 (2007). [CrossRef] [PubMed]
  8. K. Singh, C. Dion, M. R. Lesk, T. Ozaki, and S. Costantino, “Spectral-domain phase microscopy with improved sensitivity using two-dimensional detector arrays,” Rev. Sci. Instrum.82(2), 023706 (2011). [CrossRef] [PubMed]
  9. Z. Yaqoob, W. Choi, S. Oh, N. Lue, Y. Park, C. Fang-Yen, R. R. Dasari, K. Badizadegan, and M. S. Feld, “Improved phase sensitivity in spectral domain phase microscopy using line-field illumination and self phase-referencing,” Opt. Express17(13), 10681–10687 (2009). [CrossRef] [PubMed]
  10. Y. Verma, P. Nandi, K. D. Rao, M. Sharma, and P. K. Gupta, “Use of common path phase sensitive spectral domain optical coherence tomography for refractive index measurements,” Appl. Opt.50(25), E7–E12 (2011). [CrossRef]
  11. M. V. Sarunic, S. Weinberg, and J. A. Izatt, “Full-field swept-source phase microscopy,” Opt. Lett.31(10), 1462–1464 (2006). [CrossRef] [PubMed]
  12. P. Wang, R. Bista, R. Bhargava, R. E. Brand, and Y. Liu, “Spatial-domain low-coherence quantitative phase microscopy for cancer diagnosis,” Opt. Lett.35(17), 2840–2842 (2010). [CrossRef] [PubMed]
  13. B. J. Vakoc, S. H. Yun, J. F. de Boer, G. J. Tearney, and B. E. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express13(14), 5483–5493 (2005). [CrossRef] [PubMed]
  14. M. A. Choma, A. K. Ellerbee, S. Yazdanfar, and J. A. Izatt, “Doppler flow imaging of cytoplasmic streaming using spectral domain phase microscopy,” J. Biomed. Opt.11(2), 024014 (2006). [CrossRef] [PubMed]
  15. T. J. Flynn, “Two-dimensional phase unwrapping with minimum weighted discontinuity,” J. Opt. Soc. Am. A14(10), 2692–2701 (1997). [CrossRef]
  16. D. Parshall and M. K. Kim, “Digital holographic microscopy with dual-wavelength phase unwrapping,” Appl. Opt.45(3), 451–459 (2006). [CrossRef] [PubMed]
  17. H. C. Hendargo, M. Zhao, N. Shepherd, and J. A. Izatt, “Synthetic wavelength based phase unwrapping in spectral domain optical coherence tomography,” Opt. Express17(7), 5039–5051 (2009). [CrossRef] [PubMed]
  18. M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, and A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett.35(15), 2612–2614 (2010). [CrossRef] [PubMed]
  19. J. Zhang, B. Rao, L. Yu, and Z. Chen, “High-dynamic-range quantitative phase imaging with spectral domain phase microscopy,” Opt. Lett.34(21), 3442–3444 (2009). [CrossRef] [PubMed]
  20. E. D. Moore and R. R. McLeod, “Phase-sensitive swept-source interferometry for absolute ranging with application to measurements of group refractive index and thickness,” Opt. Express19(9), 8117–8126 (2011). [CrossRef] [PubMed]
  21. Y. Zhu, N. T. Shaked, L. L. Satterwhite, and A. Wax, “Spectral-domain differential interference contrast microscopy,” Opt. Lett.36(4), 430–432 (2011). [CrossRef] [PubMed]
  22. C. Wang, Z. H. Ding, S. T. Mei, H. Yu, W. Hong, Y. Z. Yan, and W. D. Shen, “Ultralong-range phase imaging with orthogonal dispersive spectral-domain optical coherence tomography,” Opt. Lett.37(21), 4555–4557 (2012). [CrossRef] [PubMed]
  23. Y. Z. Yan, Z. H. Ding, L. Wang, C. Wang, and Y. Shen, “High-sensitive quantitative phase imaging with averaged spectral domain phase microscopy,” Opt. Commun.303, 21–24 (2013). [CrossRef]
  24. K. Wang, Z. H. Ding, T. Wu, C. Wang, J. Meng, M. H. Chen, and L. Xu, “Development of a non-uniform discrete Fourier transform based high speed spectral domain optical coherence tomography system,” Opt. Express17(14), 12121–12131 (2009). [CrossRef] [PubMed]
  25. J. Meng, Z. H. Ding, J. Li, K. Wang, and T. Wu, “Transit-time analysis based on delay-encoded beam shape for velocity vector quantification by spectral-domain Doppler optical coherence tomography,” Opt. Express18(2), 1261–1270 (2010). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.