Wideband laser locking to an atomic reference with modulation transfer spectroscopy

V. Negnevitsky and L. D. Turner  »View Author Affiliations


Optics Express, Vol. 21, Issue 3, pp. 3103-3113 (2013)
http://dx.doi.org/10.1364/OE.21.003103


View Full Text Article

Acrobat PDF (1908 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We demonstrate that conventional modulated spectroscopy apparatus, used for laser frequency stabilization in many atomic physics laboratories, can be enhanced to provide a wideband lock delivering deep suppression of frequency noise across the acoustic range. Using an acousto-optic modulator driven with an agile oscillator, we show that wideband frequency modulation of the pump laser in modulation transfer spectroscopy produces the unique single lock-point spectrum previously demonstrated with electro-optic phase modulation. We achieve a laser lock with 100 kHz feedback bandwidth, limited by our laser control electronics. This bandwidth is sufficient to reduce frequency noise by 30 dB across the acoustic range and narrows the imputed linewidth by a factor of five.

© 2013 OSA

1. Introduction

Closed-loop laser frequency stabilization — ‘laser locking’ — is required for laser cooling, precision spectroscopy and atomic clocks. These experiments demand laser light frequency-locked to an atomic transition to an absolute accuracy of better than 1 MHz. Such accuracy may be achieved by stabilizing the laser to a sub-Doppler spectroscopic feature of an atomic reference, with several locking schemes in common use. Schemes where the laser is frequency-modulated [1

1. J. L. Hall, L. Hollberg, T. Baer, and H. G. Robinson, “Optical heterodyne saturation spectroscopy,” Appl. Phys. Lett. 39, 680–682 (1981). [CrossRef]

3

3. S. C. Bell, D. M. Heywood, J. D. White, J. D. Close, and R. E. Scholten, “Laser frequency offset locking using electromagnetically induced transparency,” Appl. Phys. Lett. 90, 171120 (2007). [CrossRef]

] are often preferred to baseband schemes [4

4. C. P. Pearman, C. S. Adams, S. G. Cox, P. F. Griffin, D. A. Smith, and I. G. Hughes, “Polarization spectroscopy of a closed atomic transition: applications to laser frequency locking,” J. Phys. B 35, 5141–5151 (2002). [CrossRef]

, 5

5. N. P. Robins, B. J. J. Slagmolen, D. A. Shaddock, J. D. Close, and M. B. Gray, “Interferometric, modulation-free laser stabilization,” Opt. Lett. 27, 1905–1907 (2002). [CrossRef]

] as they operate at frequencies above most laser technical noise and are less sensitive to alignment drift, temperature variations and stray light. The most common modulation schemes are frequency modulation spectroscopy (FMS) [1

1. J. L. Hall, L. Hollberg, T. Baer, and H. G. Robinson, “Optical heterodyne saturation spectroscopy,” Appl. Phys. Lett. 39, 680–682 (1981). [CrossRef]

] and modulation transfer spectroscopy (MTS) [6

6. G. Camy, C. Bordé, and M. Ducloy, “Heterodyne saturation spectroscopy through frequency modulation of the saturating beam,” Opt. Commun. 41, 325–330 (1982). [CrossRef]

]. We show that modulation transfer spectroscopy is particularly suited to wideband locking, providing strong acoustic noise rejection and appreciable linewidth narrowing, advantages typically associated with systems using an optical cavity as a reference. The system retains the environmental immunity, robustness and absolute accuracy of a direct atomic lock. Using only the optics of a commonly used acousto-optic modulator (AOM) lock setup, we show that enhanced electronics delivers feedback bandwidths over 100 kHz, suppressing frequency noise across the acoustic range by 30 dB. Furthermore, we demonstrate that as the modulation frequency is increased into the megahertz range, the error signal evolves from the familiar cluster of overlapping dispersion curves into a characteristic modulation transfer spectrum with an unambiguous lock point located only at the closed atomic transition.

There are several reasons to prefer MTS over FMS for locking purposes. In both schemes the probe is coherently demodulated to obtain the error signal. The FMS lineshape arises due to the direct probing of the vapor absorption and dispersion by the probe beam [2

2. G. C. Bjorklund, M. D. Levenson, W. Lenth, and C. Ortiz, “Frequency modulation (FM) spectroscopy,” Appl. Phys. B 32, 145–152 (1983). [CrossRef]

], while MTS relies on the frequency-dependent transfer of modulation from the pump to the probe beam [7

7. J. H. Shirley, “Modulation transfer processes in optical heterodyne saturation spectroscopy,” Opt. Lett. 7, 537–539 (1982). [CrossRef] [PubMed]

]. The modulation frequency fm of an MTS setup is limited roughly to the natural linewidth, above which the error signal lock point develops a ‘kink’ [8

8. D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol. 19, 105601 (2008). [CrossRef]

]. FMS avoids this limitation, offering higher modulation frequencies and potentially higher signal-noise ratio (SNR). However, as it is still a linear process, FMS is encumbered with an undesirable Doppler background, and the amplitudes of its sub-Doppler features match those of the saturated absorption spectrum without emphasizing the closed transitions; which can lead to the lock point ‘hopping’ from one transition to another, among other pathologies. FMS is also sensitive to any dispersive element, including parasitic etalons, which may change the lock point by adding offsets to the demodulated error signal. In contrast, MTS produces symmetric sub-Doppler features free of Doppler background that are strong at closed transitions, such as 87Rb F = 2 → F′ = 3. Undesirable crossovers and other transitions are suppressed (given the correct selection of MTS parameters) and parasitic etalons do not contribute offsets.

Many laboratories employ a low-frequency variant of MTS, frequency modulating the pump beam with an acousto-optical modulator at relatively low modulation frequencies of order 100 kHz. This spectroscopy mode yields spectra without the Doppler background and free of etalon offsets, but the spectra otherwise resemble FM spectra showing all transitions and crossovers, and the usable servo bandwidth is less than 10 kHz. Recent work has revived interest in MTS-based laser locking, in particular due to the clean spectra and wide control bandwidths made possible by MTS at high modulation frequencies [8

8. D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol. 19, 105601 (2008). [CrossRef]

, 9

9. J. Eble and F. Schmidt-Kaler, “Optimization of frequency modulation transfer spectroscopy on the calcium 41S0 to 41P1 transition,” Appl. Phys. B 88, 563–568 (2007). [CrossRef]

]. These workers used resonant electro-optical modulators (EOMs) to phase modulate the pump beam at high modulation frequencies above 1 MHz. While AOMs have been used in both FMS and MTS systems [6

6. G. Camy, C. Bordé, and M. Ducloy, “Heterodyne saturation spectroscopy through frequency modulation of the saturating beam,” Opt. Commun. 41, 325–330 (1982). [CrossRef]

, 10

10. F. du Burck, O. Lopez, and A. El Basri, “Narrow-band correction of the residual amplitude modulation in frequency-modulation spectroscopy,” IEEE Trans. Instrum. Meas. 52, 288–291 (2003). [CrossRef]

] to provide improved optical isolation and intensity noise suppression [11

11. Q. Xiang-Hui, C. Wen-Lan, Y. Lin, Z. Da-Wei, Z. Tong, X. Qin, D. Jun, Z. Xiao-Ji, and C. Xu-Zong, “Ultra-stable rubidium-stabilized external-cavity diode laser based on the modulation transfer spectroscopy technique,” Chinese Phys. Lett. 26, 044205 (2009). [CrossRef]

, 12

12. F. du Burck, G. Tetchewo, A. N. Goncharov, and O. Lopez, “Narrow band noise rejection technique for laser frequency and length standards: application to frequency stabilization to I2 lines near dissociation limit at 501.7 nm,” Metrologia 46, 599–606 (2009). [CrossRef]

], to our knowledge there has been no detailed presentation of wideband AOM-based MTS locking. The key result of this paper is that a conventional modulated spectrometer using an acousto-optic modulator can deliver wideband modulation and coherent modulation transfer spectra using little, if any, additional optics. This AOM-based MTS spectrometer inherits the unambiguous spectrum, Doppler-free background, immunity to etalon interference effects and wide control bandwidth of EOM-based spectrometers, and requires only changes to the electronics operating a conventional AOM-based pump-probe spectrometer. We discuss the design and performance of such a system, and the various considerations in using it for robust laser locking.

2. Experimental setup

In our spectrometer (Fig. 1) an AOM [13

13. AOM: Crystal Tech. 3080-122; Laser controller: MOGlabs DLC-202.

] shifts the frequency of the pump beam by 160 MHz and imposes a fast frequency modulation. This results in a lock point 80 MHz below the rubidium D2F = 2 → F′ = 3 cycling transition, which is optimal for subsequent acousto-optical generation of several near-resonant beams. The AOM shift has the addition benefit of providing optical isolation between the diode laser and the spectrometer. We use an external-cavity diode laser [14

14. C. J. Hawthorn, K. P. Weber, and R. E. Scholten, “Littrow configuration tunable external cavity diode laser with fixed direction output beam,” Rev. Sci. Instrum. 72, 4477–4479 (2001). [CrossRef]

], with a commercial laser controller [13

13. AOM: Crystal Tech. 3080-122; Laser controller: MOGlabs DLC-202.

] for temperature stabilization and control of the diode injection current and laser cavity length. The laser produces a 40 mW elliptical beam with a large horizontal eccentricity of 0.8 and a D4σ (second moment) diameter of 1.4 mm [15

15. Four times the geometric mean of the horizontal and vertical standard deviations of the beam intensity distribution. D4σ diameter is equivalent to 1/e2 diameter for Gaussian beams, and is less affected by noise for non-Gaussian beams such as those used in the spectrometer.

]. We direct 2.7 mW to the spectrometer, of which 250 μW is used for the probe beam. The pump beam passes through a telescope (F1, F2) to reduce its diameter to 500 μm with a vertical eccentricity of 0.8, and is double-passed through the AOM using a ‘cat-eye’ retroreflector made up of F3 and M4 [16

16. E. A. Donley, T. P. Heavner, F. Levi, M. O. Tataw, and S. R. Jefferts, “Double-pass acousto-optic modulator system,” Rev. Sci. Instrum. 76, 063112 (2005). [CrossRef]

]. The double pass reduces unwanted wavevector modulation and amplitude modulation compared to single-pass arrangements such as those described in Refs. [17

17. E. Jaatinen, “An iodine stabilized laser source at two wavelengths for accurate dimensional measurements,” Rev. Sci. Instrum. 74, 1359–1361 (2003). [CrossRef]

] and [18

18. J. Zhang, D. Wei, C. Xie, and K. Peng, “Characteristics of absorption and dispersion for rubidium D2 lines with the modulation transfer spectrum,” Opt. Express 11, 1338–1344 (2003). [CrossRef] [PubMed]

]. Narrowing the beam improves AOM modulation bandwidth, and collimation minimizes the range of Bragg angles, reducing spatial inhomogeneity in the modulation of the pump beam. The frequency-modulated pump beam incident on the room-temperature rubidium vapor cell is 900 μW with the same vertical eccentricity, and a diameter of 2.4 mm.

Fig. 1 Electro-optical schematic of our acousto-optic MTS system. Frequency modulation is imparted to the pump beam via a double pass of the acousto-optic modulator (AOM). Modulation is transferred from the pump to the probe within the rubidium vapor cell. The beat of the probe frequency components at photodetector PD1 (12 MHz bandwidth) is demodulated to produce an error signal. F1, F2, F3 = +100, −35, +100 mm.

The 80 MHz rf drive for the AOM is generated by an agile voltage-controlled oscillator (VCO), locked to a 10 MHz reference using a phase-locked loop (PLL) evaluation board [19

19. VCO: Mini-Circuits (MCL) ZX95-78-S+; PLL board: Analog Devices EVAL-ADF411X-EB1; DDS board: Novatech DDS9m; Bias tee: MCL ZFBT-4R2GW+; Phase detector: MCL ZRPD-1.

]. A direct digital synthesizer (DDS) board generates the modulating signal. An fm = 3 MHz modulation was used for the laser lock results presented in Section 5, added to the PLL control signal on a bias tee [19

19. VCO: Mini-Circuits (MCL) ZX95-78-S+; PLL board: Analog Devices EVAL-ADF411X-EB1; DDS board: Novatech DDS9m; Bias tee: MCL ZFBT-4R2GW+; Phase detector: MCL ZRPD-1.

]. The PLL loop bandwidth is 5 kHz, being deliberately low to avoid the PLL feedback interfering with the modulation. The system is easily adaptable to other AOM center frequencies, provided a VCO is available with sufficient modulation bandwidth.

3. Modulation transfer lineshapes

Our use of acousto-optic modulation makes it much easier to vary the modulation frequency fm than with a resonant EOM, helping us find the modulating frequency giving the optimal lock and a clean MTS spectrum. We studied the MTS lineshape as fm was varied from 100 kHz to 8 MHz, limited by the mixer below 100 kHz and the AOM modulation bandwidth above 8 MHz. This provided novel information on MTS in rubidium below 1 MHz, a region where it is difficult to construct high-Q resonant electro-optic modulators [21

21. L. Mudarikwa, K. Pahwa, and J. Goldwin, “Sub-Doppler modulation spectroscopy of potassium for laser stabilization,” J. Phys. B 45, 065002 (2012). [CrossRef]

], bridging the gap between AOM-based schemes with fm < 200 kHz [18

18. J. Zhang, D. Wei, C. Xie, and K. Peng, “Characteristics of absorption and dispersion for rubidium D2 lines with the modulation transfer spectrum,” Opt. Express 11, 1338–1344 (2003). [CrossRef] [PubMed]

, 22

22. Z. Zhou, R. Wei, C. Shi, and Y. Wang, “Observation of modulation transfer spectroscopy in the deep modulation regime,” Chinese Phys. Lett. 27, 124211 (2010). [CrossRef]

] and EOM-based schemes with fm above several megahertz [8

8. D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol. 19, 105601 (2008). [CrossRef]

]. We maintained a constant frequency deviation of 2.4 MHz (i.e. maximum deviation of the rf from 80 MHz, limited by our VCO); larger deviations produce monotonically larger locking features, but the spectra are not significantly altered otherwise. Recent studies of MTS have used a constant modulation index[8

8. D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol. 19, 105601 (2008). [CrossRef]

, 21

21. L. Mudarikwa, K. Pahwa, and J. Goldwin, “Sub-Doppler modulation spectroscopy of potassium for laser stabilization,” J. Phys. B 45, 065002 (2012). [CrossRef]

], which precludes direct comparison of the results due to the nonlinear dependence of feature size upon frequency deviation. Demodulated spectra presented were filtered through a 20 kHz Bessel low-pass filter, rather than the wideband loop filter used for locking; Fig. 2(a) displays several example spectra.

Fig. 2 Variation of MTS spectra with modulation frequency: a) solid (dashed) spectra represent 0° (90°) components, with b) amplitudes and c) trough-to-peak widths of each spectral feature at 0°. Solid lines in b) and c) represent the predictions of Eq. 1 in [27] using a linewidth value of 9.5 MHz; the line in b) is scaled to the experimental amplitudes using a least-squares fit. Widths are omitted where signal amplitude is too small for their reliable determination. [blue circle] (F = 2 → F′ = {1, 2}) and [green triangle] (F = 2 → F′ = {1, 3}) are crossover resonances. [red square] (F = 2 → F′ = 3) is the closed ‘cooling’ transition.

Contributions to MTS spectra may be broadly classified as incoherent (modulated hole burning) or coherent (four-wave mixing). The most familiar process is hole burning, whereby the pump beam creates a Bennett hole in the ground state population, and the probe beam measures reduced absorption when both probe and pump beam are resonant with atoms of the same velocity class. This process occurs with unmodulated probe beams, and is commonly known as ‘saturated absorption’; it is well-known that in the alkalis the reduction of absorption on open transitions and crossovers is in fact largely due to the incoherent process of ‘velocity selective optical pumping’ into the other hyperfine ground state [24

24. D. A. Smith and I. G. Hughes, “The role of hyperfine pumping in multilevel systems exhibiting saturated absorption,” Am. J. Phys. 72, 631–637 (2004). [CrossRef]

].

At low modulation frequencies, the pump sweeps across the ground state velocity distribution, modulating the location of the Bennett hole on the distribution. The absorption of the velocity class resonant with the probe is thus modulated, causing in-phase amplitude modulation on the probe beam. This process is clearest in the fm = 100 kHz trace uppermost in Fig. 2. As a simple sub-Doppler modulation spectroscopy, the spectrum resembles the derivative of the unmodulated absorption spectrum and shows strong signals at crossover resonances (leftmost two features) and closed transitions (rightmost feature). It thus resembles an FMS spectrum, though without the Doppler background. At much higher modulation frequencies, the period of the frequency modulation becomes short compared to the mean transit time of an atom across the probe beam. In this limit there will be atoms still transiting the beam which have been optically pumped by the pump beam at any given optical frequency within its deviation range, and so there will be vanishing oscillation of the probe absorption at the modulation frequency. For our beam diameters and a room-temperature thermal distribution, the mean transit time is equal to the modulation period at approximately fm = 150 kHz. Figure 2(b) shows the heights of the crossover resonances falling sharply [25

25. The peak-to-peak height and width of each spectral feature were obtained by numerically locating the outermost pair of stationary points within the expected frequency range of the spectral feature and above a threshold amplitude, then calculating the frequency and amplitude difference between them. This method is immune to RAM, which causes a distortion with even symmetry around the transition frequency. The MTS features are odd-symmetric around this frequency, thus RAM distortion shifts both stationary points by a common distance in frequency and amplitude. Other stationary points, such as the central trough in the 7 MHz closed transition feature (due to high levels of RAM), are ignored by the algorithm.

] and becoming unmeasurable by fm = 400 kHz, consistent with this simple model of modulated optical pumping.

The coherent component of a modulation transfer spectrum has been attributed to four-wave mixing, whereby the pump carrier, probe carrier and a pump sideband interact via the χ3 susceptibility of the near-resonant vapor to create a new probe sideband [7

7. J. H. Shirley, “Modulation transfer processes in optical heterodyne saturation spectroscopy,” Opt. Lett. 7, 537–539 (1982). [CrossRef] [PubMed]

]. This process is inoperative at DC, is largely (if not entirely [26

26. H. Noh, S. E. Park, L. Z. Li, J. Park, and C. Cho, “Modulation transfer spectroscopy for 87Rb atoms: theory and experiment,” Opt. Express 19, 23444–23452 (2011). [CrossRef] [PubMed]

]) suppressed at crossover resonances, and is only efficient at closed transitions. The incoherent process weakens as the modulation frequency is increased, revealing the coherent single-feature spectra at frequencies above 1 MHz. Below 1 MHz modulated optical pumping is much more significant than four-wave mixing for the crossover lineshapes, however both processes contribute significantly to the closed transition. For example, at 100 kHz the crossovers are relatively undistorted, while the closed transition shows strong asymmetry. This distortion appears to be independent of RAM effects, which should affect all spectral features equally. We cannot conclusively pinpoint its cause, although it appears to be due to the interplay between the two MTS processes.

The spectra (Fig. 2(a)) and peak-to-peak heights (Fig. 2(b)) show that while the absolute height of the closed transition feature falls relatively slowly (approximately as the negative logarithm) with modulation frequency, the closed transition completely dominates the spectrum for fm > 500 kHz, providing an unambiguous lock point. The roll-off of the closed transition height in Fig. 2(b) is steeper than predicted by Eq. (1) in Ref. [27

27. E. Jaatinen, “Theoretical determination of maximum signal levels obtainable with modulation transfer spectroscopy,” Opt. Commun. 120, 91–97 (1995). [CrossRef]

] using a simple model of the atomic four-wave mixing process. We attribute this discrepancy to the limited modulation bandwidth of the AOM, discussed in the following section.

The peak-trough widths of the spectral features are shown in Fig. 2. We fit these to the same simple model, using the effective linewidth as the free parameter. Agreement is best with an effective linewidth of 9.5 MHz (1.6 times the natural linewidth). As noted elsewhere, the four-wave mixing process cannot be modeled as a simple two-level system [8

8. D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol. 19, 105601 (2008). [CrossRef]

]. This is clearly illustrated by the fact that a two-level model predicts a power-broadened linewidth of around 20 MHz.

We found that fm = 3 MHz was optimal for locking purposes, providing spectra with weak crossovers, a single clear lock point at the closed transition and a high SNR. Frequencies well below 3 MHz were insufficient to escape 1/f environmental and laser noise, and the limited modulation bandwidth of the AOM reduced the signal amplitude above 3 MHz. Modulating at 3 MHz should give a usable servo bandwidth of order 1 MHz for a laser controller able to respond sufficiently fast.

4. Optimizing the spectrometer

Our MTS-based lock employs a high modulation frequency of 3 MHz, delivering a usable feedback bandwidth of several hundred kHz. Working at such modulation frequencies places more critical demands on AOM alignment than those of commonly-used narrowband modulation at 100 kHz or below. We briefly summarize techniques we found useful in achieving a symmetrical error signal at these high frequencies.

AOMs diffract efficiently when the Bragg criterion is satisfied, connecting angles of incident, diffracted and radiofrequency waves. Conventionally this match is best made with the widest possible laser beam, limited by the ‘acoustic aperture’ of the AOM. Such wide beams present no problems at conventional low modulation frequencies: at fm = 100 kHz the acoustic wave travels several cm over a modulation cycle, and the laser beam illuminates a moving grating of spatial homogeneous period. At our high modulation frequency of fm = 3 MHz in our AOM crystal (TeO2, vsound = 4.2 mm/μs), the wave propagates only 1.4 mm during a cycle, so that the a 1 mm wide diffracted beam contains almost the full cycle of frequency deviations across its transverse profile. This ‘modulation bandwidth’ limit reduces contrast of the MTS error signal, and we reduce the incident beam diameter to 500 μm to ameliorate this effect. Focusing the pump beam through the AOM tends to produce too narrow a waist for reasonable focal lengths, and instead we found that forming a collimated beam of diameter 500 μm with a telescope (as shown in Fig. 1) yields the best compromise of sufficient diffraction efficiency with only slight compression of the MTS error signal due to the acoustic transit-time effect.

The final optical step is the careful superposition of the modulated pump beam with the probe beam in the vapor cell. This alignment was best made using the MTS signal as a diagnostic, adjusting beam overlap until a symmetric error signal was obtained. The signal was observed at several settings of the demodulation phase. We found that imperfect suppression of amplitude modulation in the AOM could be largely compensated with beam overlap [23

23. E. Jaatinen and D. J. Hopper, “Compensating for frequency shifts in modulation transfer spectroscopy caused by residual amplitude modulation,” Opt. Laser. Eng. 46, 69–74 (2008). [CrossRef]

].

Electronic adjustments complete the optimization. The demodulation phase was chosen to maximize the error signal amplitude, and the lineshape was balanced using the above techniques for this particular phase. Finally, we note that small frequency shifts in the lock point of up to several megahertz may be realized with great precision and stability by altering the AOM carrier frequency. We anticipate that such ‘frequency shimming’ will be useful to calibrate the MTS lock point to a more precise atomic reference, such as absorption measurements of cold atoms in regions of minimal stray magnetic field.

5. Lock performance

The major aims of this work were to produce a stable, reliable lock, with strong frequency noise suppression and significant linewidth reduction. Reliability was verified empirically over a period of months: typically the laser remains locked for many hours, and often days, at a time, with loss of lock caused most often by laser mode hops or temperature fluctuations exceeding the controller range. The lock is difficult to dislodge without mechanically disturbing the laser itself, thus no special care needs to be taken when working near the system.

To quantify noise suppression and linewidth, we measured laser frequency noise at Fourier frequencies up to 30 kHz using an optical spectrum analyzer (Sirah EagleEye) based on a high-finesse cavity. The EagleEye signal processor servos the cavity length to maintain the laser frequency half-way up an Airy fringe of the cavity transmission spectrum, and records time series of the transmission. Welch’s average periodogram method [28

28. W. H. Press, B. P. Flannery, S. A. Teukolsky, and W. T. Vetterling, “Numerical Recipes in C: The Art of Scientific Computing” (Cambridge University Press, 1992).

] was used to obtain the power spectral density (PSD) Sδν (f) of the laser frequency noise, shown in Fig. 3(a). From 30 kHz to 1.5 MHz the pre-filter error signal spectrum was measured electronically, and calibrated to the EagleEye data by measuring the response of both to modulation of the controller input.

Fig. 3 Laser stabilization properties. a) Composite Sδν (f), showing both EagleEye and MTS error signal frequency noise, for (top to bottom) piezo feedback with minimal gain, piezo and current feedback with minimal gains, piezo and current with optimal gains. Dark (light) traces represent EagleEye (MTS error signal) data. Below 30 kHz both are displayed, while above 30 kHz only MTS error signal data are shown. The β-separation line (red) and off-resonant MTS error signal noise floor (black) are shown (EagleEye photodetector noise floor is below this by 30 dB). b) (outer to inner) corresponding inferred lineshapes, calculated from spectra in (a) using Sδν (f) from EagleEye data below 30 kHz, and MTS error signal data from 30 kHz to 1.5 MHz.

The frequency noise spectrum is more fundamental than the Allan variance and generally a more helpful diagnostic [29

29. L. D. Turner, K. Weber, C. Hawthorn, and R. E. Scholten, “Frequency noise characterisation of narrow linewidth diode lasers,” Opt. Commun. 201, 391–397 (2002). [CrossRef]

], and was used to optimize the feedback gains. Figure 3 shows the noise suppression and imputed linewidth narrowing of the MTS system. A low-gain piezoelectric-only laser lock was used to estimate the free-running laser frequency noise, holding the laser frequency steady during measurements (top trace in Fig. 3(a)). To leverage the wideband design of the spectrometer, we added injection current feedback (second trace), which strongly attenuated noise below 2 kHz but failed to suppress mechanical resonances at 2 and 4 kHz. Piezo and current gains were then optimized for maximum noise suppression, raising the bandwidth up to 100 kHz with over 30 dB suppression across acoustic frequencies (third trace). The peaks in the noise spectrum below 5 kHz are due to mechanical resonances in the laser grating mount.

A common figure of merit for laser frequency stability is full-width at half-maximum (FWHM) linewidth, typically measured for sub-megahertz lines using a beatnote between two similar laser systems. Lacking two lasers, we instead inferred the optical spectrum from the PSD by determining the autocorrelation function, then applying the Wiener-Khinchin theorem to reconstruct the lineshape [30

30. D. S. Elliott, R. Roy, and S. J. Smith, “Extracavity laser band-shape and bandwidth modification,” Phys. Rev. A 26, 12–18 (1982). [CrossRef]

]. The effect on the FWHM linewidth of noise at Fourier frequency f hinges primarily on the strength of the noise Sδν (f) relative to the line 8ln(2) f/π2, which separates the PSD into regions of high and low modulation index [31

31. G. Di Domenico, S. Schilt, and P. Thomann, “Simple approach to the relation between laser frequency noise and laser line shape,” Appl. Opt. 49, 4801–4807 (2010). [CrossRef] [PubMed]

]. Noise above this ‘β-separation’ line (i.e. at frequencies where Sδν (f) > 8ln(2) f/π2) causes a relatively high frequency deviation compared to its Fourier frequency, and contributes significantly to broadening the line at its shoulders. The further Sδν (f) is above the line, the more the FWHM linewidth is increased. The converse applies to noise ‘below the line’, which affects only the wings of the lineshape, and its suppression is unimportant for reducing the FWHM linewidth.

The control bandwidth of our system was sufficient to suppress noise across the entire frequency range where it exceeded the β-separation line. Figure 3(a) shows the frequency noise suppressed to almost the system noise floor from near DC up to 60 kHz. As with all stable controllers to which the Bode sensitivity integral applies, a distinct ‘servo bump’ appeared above the 0 dB point (100 kHz) as a necessary consequence of noise suppression at lower frequencies. This bump is visible as a broadening of the base in Fig. 3(b), however because it falls largely below the β-separation line the FWHM linewidth is not significantly increased by its presence.

For controller gains higher than those used for the ‘optimal’ trace in Fig. 3, the noise in the error signal fell below the EagleEye frequency noise floor, as photodetector noise and laser intensity noise began to dominate frequency noise in the control loop. In this limit the error signal was an unreliable measure of linewidth. To further reduce the linewidth our system would require significantly higher SNR of the error signal; this could be achieved by increasing the power diverted to the spectrometer, increasing beam diameters, reducing noise by using intensity noise cancellation [12

12. F. du Burck, G. Tetchewo, A. N. Goncharov, and O. Lopez, “Narrow band noise rejection technique for laser frequency and length standards: application to frequency stabilization to I2 lines near dissociation limit at 501.7 nm,” Metrologia 46, 599–606 (2009). [CrossRef]

] and/or heating the vapor cell.

Laser systems are often compared using the root mean square (RMS) of their frequency noise to obtain a ‘linewidth’, especially where a frequency discriminator is used [32

32. S. Hädrich, P. Jauernik, L. McCrumb, and P. Feru, “Narrow linewidth ring laser with frequency doubling for titanium:sapphire and dye operation,” Proc. SPIE 6871, 68711S (2008). [CrossRef]

, 33

33. D. J. Thompson and R. E. Scholten, “Narrow linewidth tunable external cavity diode laser using wide bandwidth filter,” Rev. Sci. Instrum. 83, 023107 (2012). [CrossRef] [PubMed]

]. As shown in Table 1, RMS linewidths poorly reflect the valuable FWHM linewidth reduction visible in Fig. 3(b). RMS figures underestimate the linewidth when 1/f noise dominates and overestimate it when it is suppressed, due to the dominance of the servo bump.

Table 1. FWHM and RMS linewidths under the locking circumstances of Fig. 3. All quantities in kilohertz.

table-icon
View This Table

Our system has thus reduced the imputed laser linewidth by a factor of 5, limited by the controller bandwidth of 100 kHz and the noise floor of the system. We hope to increase this by extending the control bandwidth to at least several hundred kilohertz; the wideband MTS electronics should support this without modification. The PSD and laser lineshape are both useful for designing and characterizing such an extension.

6. Conclusions

In summary, this paper has outlined our approach for designing, characterizing and optimizing an AOM-based laser lock using modulation transfer spectroscopy. We have combined conventional, inexpensive AOM-based optics with a wideband electronic modulation and demodulation scheme, and shown that the error signal is of sufficient quality to reduce the inferred linewidth of a typical external-cavity diode laser by a factor of five. We have also demonstrated that to obtain the full range of benefits an MTS spectrometer provides, modulation frequencies above 500 kHz must be used. Heuristics for designing and optimizing such a system have been presented, which should help improve performance for adopters of this system. The laser frequency PSD has been used to quantify lock performance, and numerical calculation and analysis of the laser lineshape will inform future improvements to the controller.

Acknowledgments

This work was supported by a Monash University Faculty of Science ECR Grant, the J. L. William Fund and the Australian Research Council ( DP1094399). R. P. Anderson, M. Jasperse, A. A. Wood and L. M. Bennie assisted with the construction and optimization of the laser system. We are obliged to R. E. Scholten for loan of the EagleEye optical spectrum analyzer. The manuscript benefited greatly from a thorough review by R. P. Anderson and R. E. Scholten, for which we are very grateful. We thank A. Slavec for implementing the Chebyshev filter and A. Benci for electronics support.

References and links

1.

J. L. Hall, L. Hollberg, T. Baer, and H. G. Robinson, “Optical heterodyne saturation spectroscopy,” Appl. Phys. Lett. 39, 680–682 (1981). [CrossRef]

2.

G. C. Bjorklund, M. D. Levenson, W. Lenth, and C. Ortiz, “Frequency modulation (FM) spectroscopy,” Appl. Phys. B 32, 145–152 (1983). [CrossRef]

3.

S. C. Bell, D. M. Heywood, J. D. White, J. D. Close, and R. E. Scholten, “Laser frequency offset locking using electromagnetically induced transparency,” Appl. Phys. Lett. 90, 171120 (2007). [CrossRef]

4.

C. P. Pearman, C. S. Adams, S. G. Cox, P. F. Griffin, D. A. Smith, and I. G. Hughes, “Polarization spectroscopy of a closed atomic transition: applications to laser frequency locking,” J. Phys. B 35, 5141–5151 (2002). [CrossRef]

5.

N. P. Robins, B. J. J. Slagmolen, D. A. Shaddock, J. D. Close, and M. B. Gray, “Interferometric, modulation-free laser stabilization,” Opt. Lett. 27, 1905–1907 (2002). [CrossRef]

6.

G. Camy, C. Bordé, and M. Ducloy, “Heterodyne saturation spectroscopy through frequency modulation of the saturating beam,” Opt. Commun. 41, 325–330 (1982). [CrossRef]

7.

J. H. Shirley, “Modulation transfer processes in optical heterodyne saturation spectroscopy,” Opt. Lett. 7, 537–539 (1982). [CrossRef] [PubMed]

8.

D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol. 19, 105601 (2008). [CrossRef]

9.

J. Eble and F. Schmidt-Kaler, “Optimization of frequency modulation transfer spectroscopy on the calcium 41S0 to 41P1 transition,” Appl. Phys. B 88, 563–568 (2007). [CrossRef]

10.

F. du Burck, O. Lopez, and A. El Basri, “Narrow-band correction of the residual amplitude modulation in frequency-modulation spectroscopy,” IEEE Trans. Instrum. Meas. 52, 288–291 (2003). [CrossRef]

11.

Q. Xiang-Hui, C. Wen-Lan, Y. Lin, Z. Da-Wei, Z. Tong, X. Qin, D. Jun, Z. Xiao-Ji, and C. Xu-Zong, “Ultra-stable rubidium-stabilized external-cavity diode laser based on the modulation transfer spectroscopy technique,” Chinese Phys. Lett. 26, 044205 (2009). [CrossRef]

12.

F. du Burck, G. Tetchewo, A. N. Goncharov, and O. Lopez, “Narrow band noise rejection technique for laser frequency and length standards: application to frequency stabilization to I2 lines near dissociation limit at 501.7 nm,” Metrologia 46, 599–606 (2009). [CrossRef]

13.

AOM: Crystal Tech. 3080-122; Laser controller: MOGlabs DLC-202.

14.

C. J. Hawthorn, K. P. Weber, and R. E. Scholten, “Littrow configuration tunable external cavity diode laser with fixed direction output beam,” Rev. Sci. Instrum. 72, 4477–4479 (2001). [CrossRef]

15.

Four times the geometric mean of the horizontal and vertical standard deviations of the beam intensity distribution. D4σ diameter is equivalent to 1/e2 diameter for Gaussian beams, and is less affected by noise for non-Gaussian beams such as those used in the spectrometer.

16.

E. A. Donley, T. P. Heavner, F. Levi, M. O. Tataw, and S. R. Jefferts, “Double-pass acousto-optic modulator system,” Rev. Sci. Instrum. 76, 063112 (2005). [CrossRef]

17.

E. Jaatinen, “An iodine stabilized laser source at two wavelengths for accurate dimensional measurements,” Rev. Sci. Instrum. 74, 1359–1361 (2003). [CrossRef]

18.

J. Zhang, D. Wei, C. Xie, and K. Peng, “Characteristics of absorption and dispersion for rubidium D2 lines with the modulation transfer spectrum,” Opt. Express 11, 1338–1344 (2003). [CrossRef] [PubMed]

19.

VCO: Mini-Circuits (MCL) ZX95-78-S+; PLL board: Analog Devices EVAL-ADF411X-EB1; DDS board: Novatech DDS9m; Bias tee: MCL ZFBT-4R2GW+; Phase detector: MCL ZRPD-1.

20.

Photodetector: Thorlabs PDA36A, +10 dB gain setting, 12 MHz bandwidth.

21.

L. Mudarikwa, K. Pahwa, and J. Goldwin, “Sub-Doppler modulation spectroscopy of potassium for laser stabilization,” J. Phys. B 45, 065002 (2012). [CrossRef]

22.

Z. Zhou, R. Wei, C. Shi, and Y. Wang, “Observation of modulation transfer spectroscopy in the deep modulation regime,” Chinese Phys. Lett. 27, 124211 (2010). [CrossRef]

23.

E. Jaatinen and D. J. Hopper, “Compensating for frequency shifts in modulation transfer spectroscopy caused by residual amplitude modulation,” Opt. Laser. Eng. 46, 69–74 (2008). [CrossRef]

24.

D. A. Smith and I. G. Hughes, “The role of hyperfine pumping in multilevel systems exhibiting saturated absorption,” Am. J. Phys. 72, 631–637 (2004). [CrossRef]

25.

The peak-to-peak height and width of each spectral feature were obtained by numerically locating the outermost pair of stationary points within the expected frequency range of the spectral feature and above a threshold amplitude, then calculating the frequency and amplitude difference between them. This method is immune to RAM, which causes a distortion with even symmetry around the transition frequency. The MTS features are odd-symmetric around this frequency, thus RAM distortion shifts both stationary points by a common distance in frequency and amplitude. Other stationary points, such as the central trough in the 7 MHz closed transition feature (due to high levels of RAM), are ignored by the algorithm.

26.

H. Noh, S. E. Park, L. Z. Li, J. Park, and C. Cho, “Modulation transfer spectroscopy for 87Rb atoms: theory and experiment,” Opt. Express 19, 23444–23452 (2011). [CrossRef] [PubMed]

27.

E. Jaatinen, “Theoretical determination of maximum signal levels obtainable with modulation transfer spectroscopy,” Opt. Commun. 120, 91–97 (1995). [CrossRef]

28.

W. H. Press, B. P. Flannery, S. A. Teukolsky, and W. T. Vetterling, “Numerical Recipes in C: The Art of Scientific Computing” (Cambridge University Press, 1992).

29.

L. D. Turner, K. Weber, C. Hawthorn, and R. E. Scholten, “Frequency noise characterisation of narrow linewidth diode lasers,” Opt. Commun. 201, 391–397 (2002). [CrossRef]

30.

D. S. Elliott, R. Roy, and S. J. Smith, “Extracavity laser band-shape and bandwidth modification,” Phys. Rev. A 26, 12–18 (1982). [CrossRef]

31.

G. Di Domenico, S. Schilt, and P. Thomann, “Simple approach to the relation between laser frequency noise and laser line shape,” Appl. Opt. 49, 4801–4807 (2010). [CrossRef] [PubMed]

32.

S. Hädrich, P. Jauernik, L. McCrumb, and P. Feru, “Narrow linewidth ring laser with frequency doubling for titanium:sapphire and dye operation,” Proc. SPIE 6871, 68711S (2008). [CrossRef]

33.

D. J. Thompson and R. E. Scholten, “Narrow linewidth tunable external cavity diode laser using wide bandwidth filter,” Rev. Sci. Instrum. 83, 023107 (2012). [CrossRef] [PubMed]

OCIS Codes
(300.6290) Spectroscopy : Spectroscopy, four-wave mixing
(140.3425) Lasers and laser optics : Laser stabilization

ToC Category:
Lasers and Laser Optics

History
Original Manuscript: January 9, 2013
Manuscript Accepted: January 15, 2013
Published: January 31, 2013

Citation
V. Negnevitsky and L. D. Turner, "Wideband laser locking to an atomic reference with modulation transfer spectroscopy," Opt. Express 21, 3103-3113 (2013)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-21-3-3103


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. J. L. Hall, L. Hollberg, T. Baer, and H. G. Robinson, “Optical heterodyne saturation spectroscopy,” Appl. Phys. Lett.39, 680–682 (1981). [CrossRef]
  2. G. C. Bjorklund, M. D. Levenson, W. Lenth, and C. Ortiz, “Frequency modulation (FM) spectroscopy,” Appl. Phys. B32, 145–152 (1983). [CrossRef]
  3. S. C. Bell, D. M. Heywood, J. D. White, J. D. Close, and R. E. Scholten, “Laser frequency offset locking using electromagnetically induced transparency,” Appl. Phys. Lett.90, 171120 (2007). [CrossRef]
  4. C. P. Pearman, C. S. Adams, S. G. Cox, P. F. Griffin, D. A. Smith, and I. G. Hughes, “Polarization spectroscopy of a closed atomic transition: applications to laser frequency locking,” J. Phys. B35, 5141–5151 (2002). [CrossRef]
  5. N. P. Robins, B. J. J. Slagmolen, D. A. Shaddock, J. D. Close, and M. B. Gray, “Interferometric, modulation-free laser stabilization,” Opt. Lett.27, 1905–1907 (2002). [CrossRef]
  6. G. Camy, C. Bordé, and M. Ducloy, “Heterodyne saturation spectroscopy through frequency modulation of the saturating beam,” Opt. Commun.41, 325–330 (1982). [CrossRef]
  7. J. H. Shirley, “Modulation transfer processes in optical heterodyne saturation spectroscopy,” Opt. Lett.7, 537–539 (1982). [CrossRef] [PubMed]
  8. D. J. McCarron, S. A. King, and S. L. Cornish, “Modulation transfer spectroscopy in atomic rubidium,” Meas. Sci. Technol.19, 105601 (2008). [CrossRef]
  9. J. Eble and F. Schmidt-Kaler, “Optimization of frequency modulation transfer spectroscopy on the calcium 41S0 to 41P1 transition,” Appl. Phys. B88, 563–568 (2007). [CrossRef]
  10. F. du Burck, O. Lopez, and A. El Basri, “Narrow-band correction of the residual amplitude modulation in frequency-modulation spectroscopy,” IEEE Trans. Instrum. Meas.52, 288–291 (2003). [CrossRef]
  11. Q. Xiang-Hui, C. Wen-Lan, Y. Lin, Z. Da-Wei, Z. Tong, X. Qin, D. Jun, Z. Xiao-Ji, and C. Xu-Zong, “Ultra-stable rubidium-stabilized external-cavity diode laser based on the modulation transfer spectroscopy technique,” Chinese Phys. Lett.26, 044205 (2009). [CrossRef]
  12. F. du Burck, G. Tetchewo, A. N. Goncharov, and O. Lopez, “Narrow band noise rejection technique for laser frequency and length standards: application to frequency stabilization to I2 lines near dissociation limit at 501.7 nm,” Metrologia46, 599–606 (2009). [CrossRef]
  13. AOM: Crystal Tech. 3080-122; Laser controller: MOGlabs DLC-202.
  14. C. J. Hawthorn, K. P. Weber, and R. E. Scholten, “Littrow configuration tunable external cavity diode laser with fixed direction output beam,” Rev. Sci. Instrum.72, 4477–4479 (2001). [CrossRef]
  15. Four times the geometric mean of the horizontal and vertical standard deviations of the beam intensity distribution. D4σ diameter is equivalent to 1/e2 diameter for Gaussian beams, and is less affected by noise for non-Gaussian beams such as those used in the spectrometer.
  16. E. A. Donley, T. P. Heavner, F. Levi, M. O. Tataw, and S. R. Jefferts, “Double-pass acousto-optic modulator system,” Rev. Sci. Instrum.76, 063112 (2005). [CrossRef]
  17. E. Jaatinen, “An iodine stabilized laser source at two wavelengths for accurate dimensional measurements,” Rev. Sci. Instrum.74, 1359–1361 (2003). [CrossRef]
  18. J. Zhang, D. Wei, C. Xie, and K. Peng, “Characteristics of absorption and dispersion for rubidium D2 lines with the modulation transfer spectrum,” Opt. Express11, 1338–1344 (2003). [CrossRef] [PubMed]
  19. VCO: Mini-Circuits (MCL) ZX95-78-S+; PLL board: Analog Devices EVAL-ADF411X-EB1; DDS board: Novatech DDS9m; Bias tee: MCL ZFBT-4R2GW+; Phase detector: MCL ZRPD-1.
  20. Photodetector: Thorlabs PDA36A, +10 dB gain setting, 12 MHz bandwidth.
  21. L. Mudarikwa, K. Pahwa, and J. Goldwin, “Sub-Doppler modulation spectroscopy of potassium for laser stabilization,” J. Phys. B45, 065002 (2012). [CrossRef]
  22. Z. Zhou, R. Wei, C. Shi, and Y. Wang, “Observation of modulation transfer spectroscopy in the deep modulation regime,” Chinese Phys. Lett.27, 124211 (2010). [CrossRef]
  23. E. Jaatinen and D. J. Hopper, “Compensating for frequency shifts in modulation transfer spectroscopy caused by residual amplitude modulation,” Opt. Laser. Eng.46, 69–74 (2008). [CrossRef]
  24. D. A. Smith and I. G. Hughes, “The role of hyperfine pumping in multilevel systems exhibiting saturated absorption,” Am. J. Phys.72, 631–637 (2004). [CrossRef]
  25. The peak-to-peak height and width of each spectral feature were obtained by numerically locating the outermost pair of stationary points within the expected frequency range of the spectral feature and above a threshold amplitude, then calculating the frequency and amplitude difference between them. This method is immune to RAM, which causes a distortion with even symmetry around the transition frequency. The MTS features are odd-symmetric around this frequency, thus RAM distortion shifts both stationary points by a common distance in frequency and amplitude. Other stationary points, such as the central trough in the 7 MHz closed transition feature (due to high levels of RAM), are ignored by the algorithm.
  26. H. Noh, S. E. Park, L. Z. Li, J. Park, and C. Cho, “Modulation transfer spectroscopy for 87Rb atoms: theory and experiment,” Opt. Express19, 23444–23452 (2011). [CrossRef] [PubMed]
  27. E. Jaatinen, “Theoretical determination of maximum signal levels obtainable with modulation transfer spectroscopy,” Opt. Commun.120, 91–97 (1995). [CrossRef]
  28. W. H. Press, B. P. Flannery, S. A. Teukolsky, and W. T. Vetterling, “Numerical Recipes in C: The Art of Scientific Computing” (Cambridge University Press, 1992).
  29. L. D. Turner, K. Weber, C. Hawthorn, and R. E. Scholten, “Frequency noise characterisation of narrow linewidth diode lasers,” Opt. Commun.201, 391–397 (2002). [CrossRef]
  30. D. S. Elliott, R. Roy, and S. J. Smith, “Extracavity laser band-shape and bandwidth modification,” Phys. Rev. A26, 12–18 (1982). [CrossRef]
  31. G. Di Domenico, S. Schilt, and P. Thomann, “Simple approach to the relation between laser frequency noise and laser line shape,” Appl. Opt.49, 4801–4807 (2010). [CrossRef] [PubMed]
  32. S. Hädrich, P. Jauernik, L. McCrumb, and P. Feru, “Narrow linewidth ring laser with frequency doubling for titanium:sapphire and dye operation,” Proc. SPIE6871, 68711S (2008). [CrossRef]
  33. D. J. Thompson and R. E. Scholten, “Narrow linewidth tunable external cavity diode laser using wide bandwidth filter,” Rev. Sci. Instrum.83, 023107 (2012). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.