A simplified approach to quantitative coded aperture X-ray phase imaging

Peter R.T. Munro, Charlotte K. Hagen, Magdalena B. Szafraniec, and Alessandro Olivo  »View Author Affiliations


Optics Express, Vol. 21, Issue 9, pp. 11187-11201 (2013)
http://dx.doi.org/10.1364/OE.21.011187


View Full Text Article

Acrobat PDF (1617 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We recently demonstrated how quantitative X-ray phase contrast imaging may be performed with laboratory sources using the coded aperture technique. This technique required the knowledge of system parameters such as, for example, the source focal spot size and distances between elements of the imaging system. The method also assumes that the absorbing regions of the apertures are perfectly absorbing. In this paper we demonstrate how quantitative imaging can be performed without knowledge of individual system parameters and with partially absorbing apertures. We also show that this method is analogous to that employed in analyser based imaging which uses the rocking curve of an analyser crystal.

© 2013 OSA

1. Introduction

Coded aperture X-ray phase contrast imaging (CAXPCI) [1

1. A. Olivo and R. Speller, “A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources,” Appl. Phys. Lett. 91, 074106 (2007) [CrossRef] .

] was initially inspired by the synchrotron method known as analyser based imaging (ABI). ABI develops contrast using the rocking curve of an analyser crystal [2

2. T. Davis, D. Gao, T. Gureyev, A. Stevenson, and S. Wilkins, “Phase-contrast imaging of weakly absorbing materials using hard x-rays,” Nature 373, 595–598 (1995) [CrossRef] .

]. Algorithms were developed to extract sample phase and absorption from images taken on two opposing sides of the analyser crystal rocking curve [3

3. D. Chapman, W. Thomlinson, F. Arfelli, N. Gmür, Z. Zhong, R. Menk, R. E. Johnson, D. Washburn, E. Pisano, and D. Sayers, “Mammography imaging studies using a laue crystal analyzer,” The 9th National Conference on Synchrotron Radiation Instrumentation 67, 3360–3360 (1996).

, 4

4. D. Chapman, W. Thomlinson, R. Johnston, D. Washburn, E. Pisano, N. Gmür, Z. Zhong, R. Menk, F. Arfelli, and D. Sayers, “Diffraction enhanced x-ray imaging,” Phys. Med. Biol. 42, 2015–2025 (1997) [CrossRef] [PubMed] .

]. Algorithms employing a variety of assumptions about the sample and imaging system, and later, for more general systems were subsequently developed [5

5. D. Paganin, T. E. Gureyev, K. M. Pavlov, R. A. Lewis, and M. Kitchen, “Phase retrieval using coherent imaging systems with linear transfer functions,” Opt. Commun. 234, 87–105 (2004) [CrossRef] .

8

8. P. C. Diemoz, P. Coan, C. Glaser, and A. Bravin, “Absorption, refraction and scattering in analyzer-based imaging: comparison of different algorithms,” Opt. Express 18, 3494–3509 (2010) [CrossRef] [PubMed] .

]. Although this technique has been demonstrated using a laboratory source [9

9. D. J. Vine, D. M. Paganin, K. M. Pavlov, J. Kraeusslich, O. Wehrhan, I. Uschmann, and E. Foerster, “Analyzer-based phase contrast imaging and phase retrieval using a rotating anode x-ray source,” Appl. Phys. Lett. 91, 254110 (2007) [CrossRef] .

], it is generally restricted to synchrotron sources due to the narrow band of energies selected by the two crystals employed in the imaging system.

The restriction to synchrotron sources for applications not compatible with long exposure times motivated several people to develop alternative methods of performing X-ray phase contrast imaging (XPCI) using laboratory sources [1

1. A. Olivo and R. Speller, “A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources,” Appl. Phys. Lett. 91, 074106 (2007) [CrossRef] .

, 10

10. C. Kottler, F. Pfeiffer, O. Bunk, C. Gruenzweig, J. Bruder, R. Kaufmann, L. Tlustos, H. Walt, I. Briod, T. Weitkamp, and C. David, “Phase contrast x-ray imaging of large samples using an incoherent laboratory source,” Phys. Status Solidi. A 204, 2728–2733 (2007) [CrossRef] .

12

12. A. Momose, W. Yashiro, H. Kuwabara, and K. Kawabata, “Grating-based x-ray phase imaging using multiline x-ray source,” Jpn. J. Appl. Phys. 48, 076512 (2009) [CrossRef] .

]. It is beyond the scope of this paper to review the substantial number of contributions to this field, however we direct readers to a review on the subject [13

13. A. Momose, “Recent advances in x-ray phase imaging,” Jpn. J. Appl. Phys. 44, 6355–6367 (2005) [CrossRef] .

] for such an account. One such contribution resulted from initial experiments performed using the SYRMEP beamline of the Elettra synchrotron radiation facility in operation in Trieste [14

14. A. Olivo, F. Arfelli, G. Cantatore, R. Longo, R. Menk, S. Pani, M. Prest, P. Poropat, L. Rigon, G. Tromba, E. Vallazza, and E. Castelli, “An innovative digital imaging set-up allowing a low-dose approach to phase contrast applications in the medical field,” Med. Phys. 28, 1610–1619 (2001) [CrossRef] [PubMed] .

], an experiment directly inspired by analyser based imaging. Although initially demonstrated using synchrotron radiation, the technique was soon demonstrated using laboratory sources [1

1. A. Olivo and R. Speller, “A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources,” Appl. Phys. Lett. 91, 074106 (2007) [CrossRef] .

]. Recently this technique has been extended to a quantitative method for both laboratory [15

15. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Phase and absorption retrieval using incoherent x-ray sources,” Proc. Natl. Acad. Sci. USA 109, 13922–13927 (2012) [CrossRef] [PubMed] .

] and synchrotron sources [16

16. P. R. Munro, L. Rigon, K. Ignatyev, F. C. Lopez, D. Dreossi, R. D. Speller, and A. Olivo, “A quantitative, non-interferometric x-ray phase contrast imaging technique,” Opt. Express 21, 647–661 (2013) [CrossRef] [PubMed] .

].

The CAXPCI system is schematised in Fig. 1(a) where all parameters are defined in the figure caption. Phase contrast images are obtained by adjusting the position of A1 with respect to A2 such that the beams formed by A1 illuminate the edges of the A2. A2 is initially aligned with the detector pixels and remains stationary thereafter. Higher resolution images may be obtained by scanning the sample in the ξ direction as indicated in Fig. 1(b). CAXPCI imaging can be performed with synchrotron radiation [16

16. P. R. Munro, L. Rigon, K. Ignatyev, F. C. Lopez, D. Dreossi, R. D. Speller, and A. Olivo, “A quantitative, non-interferometric x-ray phase contrast imaging technique,” Opt. Express 21, 647–661 (2013) [CrossRef] [PubMed] .

] or laboratory sources [15

15. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Phase and absorption retrieval using incoherent x-ray sources,” Proc. Natl. Acad. Sci. USA 109, 13922–13927 (2012) [CrossRef] [PubMed] .

]. The is no limit on the spectral width of the source so long as the apertures are thick enough to generate contrast in the detected signal, in the absence of a sample, which exceeds the noise. The focal spot size should not exceed the pixel size of the detector [17

17. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Source size and temporal coherence requirements of coded aperture type x-ray phase contrast imaging systems,” Opt. Express 18, 19681–19692 (2010) [CrossRef] [PubMed] .

]. The system has recently been shown to have sensitivity which is at least comparable to that of grating interferometry [18

18. M. Marenzana, C. K. Hagen, P. D. N. Borges, M. Endrizzi, M. B. Szafraniec, K. Ignatyev, and A. Olivo, “Visualization of small lesions in rat cartilage by means of laboratory-based x-ray phase contrast imaging,” Phys. Med. Biol. 57, 8173–8184 (2012) [CrossRef] [PubMed] .

], the only other system to offer high phase sensitivity when employing a laboratory source. The difference between grating interferometry and CAXPCI has been discussed in detail previously [17

17. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Source size and temporal coherence requirements of coded aperture type x-ray phase contrast imaging systems,” Opt. Express 18, 19681–19692 (2010) [CrossRef] [PubMed] .

]. However, we note that the main difference is that CAXPCI employs the edge illumination principle where the edges of the sensitive region of a pixel are illuminated by X-ray beams. This means that CAXPCI doesn’t make use of Talbot’s self-imaging phenomenon and doesn’t employ phase stepping of the detector aperture.

Fig. 1 Schematic diagram of a CAXPCI system which is not to scale (a) and specification of the coordinate systems used within this paper (b). The system is formed by two apertures A1 and A2. The sample is placed immediately after A1. A1 is translated by an amount Δξ in order to select the operating regime of the system. A1 is placed a distance zso from the source and A2 is placed a distance zod from A1. The detector has pixels with a width of P and A1 has a period of p1. The distance along the optical axis is denoted z and is considered to begin at the centre of the source focal spot. The lateral dimensions within the source focal spot and within the planes of A1 and A2 are denoted xs, ξ and x respectively. Each lateral dimension has its origin at the optical axis.

The existing quantitative method [15

15. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Phase and absorption retrieval using incoherent x-ray sources,” Proc. Natl. Acad. Sci. USA 109, 13922–13927 (2012) [CrossRef] [PubMed] .

] was developed from a wave optical model of the CAXPCI system which took into account the principal system parameters such as source focal spot size and aperture dimensions and spacing. This technique thus requires such system parameters to be known to a reasonable degree of accuracy. However, some system parameters such as, for example, the source focal spot and spectrum are difficult to measure on a regular basis. It would be much more convenient if the effect of all system parameters could be aggregated into a single series of measurements, thus forming a simple method of calibrating the imaging system. We present such a method in this paper whereby a measurement, analogous to the rocking curve in ABI, is performed which forms the basis of the algorithm used to obtain quantitative phase measurements. We shall call this measurement the translation curve (TC) in order to distinguish it from the rocking curve used in ABI. We shall show in this paper how the measured TC may be used to perform quantitative phase imaging in a more general manner than our previously developed method.

2. Theory

2.1. Derivation of expression for the translation curve

In this section we develop a forward model which allows the detector pixel signal to be calculated taking into account the salient system parameters. This formula is then used to obtain an expression for the TC of the system. We begin by mathematically defining the transmission function of the apertures as:
𝒯i(x+nP)={1|x|<ηiP/2exp(μcaTi)otherwise,
(1)
where the subscript i takes the value 1 or 2 in correspondence with the apertures A1 and A2 respectively. μca is the energy dependent linear absorption coefficient of the absorbing material used to form the apertures (generally gold), Ti is the thickness of the absorbing material and ηi is the fraction of each aperture period which is non-absorbing. The sample is modelled by a multiplicative complex transparency function:
T(x)=exp(ikϕ(x)12𝒪μ(x,z)dz)
(2)
where μ is the linear absorption coefficient of the sample, k is the wave number, ϕ(x) = ∫𝒪δ(x, z)dz, δ is the refractive index decrement of the sample and 𝒪 is the extent of the sample. ϕ and μ both depend upon the photon energy, however, we develop the mathematical model by considering first a single energy before integrating over the source spectrum. Note also that the sample is assumed to satisfy the projection approximation, meaning that the sample is sufficiently thin and that inhomogeneities are large compared with the wavelength [19

19. S. Mayo, P. Miller, S. Wilkins, T. Davis, D. Gao, T. Gureyev, D. Paganin, D. Parry, A. Pogany, and A. Stevenson, “Quantitative x-ray projection microscopy: phase-contrast and multi-spectral imaging,” Journal of Microscopy-Oxford 207, 79–96 (2002) [CrossRef] .

]. In this context, one criterion for being sufficiently thin is that the thickness is much less than zso[20

20. T. Gureyev and S. Wilkins, “On x-ray phase imaging with a point source,” J. Opt. Soc. Am. A 15, 579–585 (1998) [CrossRef] .

] although Paganin [21

21. D. Paganin, Coherent X-ray optics, Oxford series on synchrotron radiation (Oxford University Press, Great Clarendon Street, Oxford OX2 6DP, 2006) [CrossRef] .

] presents a thorough discussion of this subject.

Without loss of generality, we consider the signal of the pixel bound by |x| < P/2. Then the signal detected by this pixel is given by
I=P/2P/2𝒯2(x)|U(x)|2dx
(5)
In order to simplify this expression into a more usable form, we approximate T(x) as
T(x)T˜(x)=exp(ik(ϕ(0)+ϕx(0)x)12𝒪μ(0,z)dz)
(6)
meaning that we assume that the object has a constant absorption and linear phase variation within the transmitting region of the pre-sample aperture. ϕx is thus the derivative of ϕ with respect to x and corresponds to the angle by which a ray is refracted by the sample. It is important to note that the coded aperture technique, like grating interferometry, measures refraction angle which, when multiplied by the wavenumber, yields the phase gradient. To avoid confusion arising due to the choice of wavenumber when employing a polychromatic source, we shall principally analyse refraction angle rather for the remainder of this paper.

2.2. Design of system parameters to enable quantitative imaging

As a final step before deriving the inversion formulae we show that |ITC′(Δξ)| is maximised at Δξ = ±Δξ0. To proceed we note that ITC′(Δξ) is given by
ITC(Δξ)=M|U0|2(zso+zod)2P/2P/2𝒯¯2(x)[S((xMΔξ+η1P/2)zsozod)S((xMΔξη1P/2)zsozod)]dx
(15)
Since we have not assumed a particular form of S, only that it is centred and maximised on xs = 0, we conclude that |ITC′(Δξ)| is indeed maximised at Δξ = ±Δξ0 since this maximises the overlap between one of the S terms with the 𝒯̄2(x) term and minimises the overlap with the other.

2.3. Derivation of inversion formulae

It is, however, possible to obtain a linear inversion formula valid for smaller refraction angles. By expanding IR and IL in a Taylor series and again making the approximation ITC (−Δξ0 − zodϕ̄x/M) + ITCξ0 − zodϕ̄x/M) ≈ 2ITCξ0), gives
IΔIΣ(zodϕ¯x/M)ITC(Δξ0)(zodϕ¯x/M)ITC(Δξ0)+O((zodϕ¯x)3)2ITC(Δξ0)=zodϕ¯xMITC(Δξ0)ITC(Δξ0)
(22)
where the remainder term in the numerator is of third order because the second derivative of ITC is approximately zero in the vicinity of Δξ0, since it was shown that the first derivative of ITC is minimised at Δξ0. Eq. (22) can then be manipulated to reveal the sample refraction angle image as:
ϕ¯x=MzodI^ΔI^ΣITC(Δξ0)ITC(Δξ0)
(23)
where ITCξ0)/I′TCξ0) is known from the experimentally measured TC.

3. Experimental example

The TC of the system is measured routinely whenever the pre-sample aperture is aligned with the detector aperture. An experimentally acquired TC is plotted in Fig. 2 using our laboratory CAXPCI system which has system parameters zso ≈1.6m and zod ≈0.4m. These quantities are approximate since the apertures and detector are initially placed in these positions and an alignment procedure is used to accurately position them. The pre-sample aperture, A1, has period p1=134μm and pixel width P=85μm. The detector aperture A2 has a period of 83.5μm to allow it to be placed in front of the detector. The width of the transmitting regions of A1 and A2 are 16μm and 20μm respectively. The quantity Δξ refers to the displacement of A1 from the position of perfect alignment with A2. Note that in practice, every second pixel was skipped to minimise detector pixel cross talk. We employed a Rigaku 007HF X-ray tube generator operated at 35 kVp/25 mA with a rotating Mo target, an Anrad SMAM flat panel detector and the gratings were fabricated by Creatv Microtech (Potomac, Maryland, USA). The TC was acquired by taking the average over all pixels for each displacement of the pre-sample aperture. Note that the measured data points are unequally spaced as two separate measurements have been combined and one of the measurements was acquired only for negative values of Δξ. The fitted curve was found by fitting a fourth order Fourier cosine series thus assuming symmetry about Δξ = 0. We showed in Sec. (2.2) that Δξ0 should be chosen to be equal to half of the width of the transmitting region of the pre-sample aperture, which is 8μm in our experimental system. Figure 3 shows plots of IL, IR and IΣ for the sampled TC which shows that IΣ is indeed flat for a range of values of Δξs = zodϕ̄x. We can, however, show using the experimentally acquired TC, that the selected value of Δξ0 results in IΣ being maximally flat over a range of values of Δξs. In order to show this we define a measure of flatness as:
ε(Δξ0)=maxΔξsΩΔξs(IΣ(Δξ0,Δξs))maxΔξsΩΔξs(IΣ(Δξ0,Δξs))meanΔξsΩΔξs(IΣ(Δξ0,Δξs))
(24)
where ΩΔξs = [−5 × 10−6, 5 × 10−6] m was chosen to accommodate extreme values measured experimentally [16

16. P. R. Munro, L. Rigon, K. Ignatyev, F. C. Lopez, D. Dreossi, R. D. Speller, and A. Olivo, “A quantitative, non-interferometric x-ray phase contrast imaging technique,” Opt. Express 21, 647–661 (2013) [CrossRef] [PubMed] .

]. The maximum, minimum and mean values were found using the fitted TC. εξ0) is plotted in Fig. 4 which shows that it exhibits a minimum value in the vicinity of Δξ0 as expected.

Fig. 2 Plot of experimentally acquired ITC. The fitted curve was found by minimising the least square error with Fourier series containing four harmonic terms.
Fig. 3 Plot of IΣ = ITC (−Δξ0 − Δξs) + ITCξ0 − Δξs), ITC (−Δξ0 − Δξs) and ITCξ0 − Δξs) using the fitted TC plotted in Fig. 2.
Fig. 4 Plot of εξ0), a measure of the flatness of IΣ for particular values of Δξ0 when Δξs takes values in the range [−5 × 10−6, 5 × 10−6] m.

Another important factor which influences the choice of Δξ0 is the sensitivity to sample refraction angles. Figure 5 shows a plot of the derivative of the curve fitted to the experimentally observed TC data. This plot verifies the result obtained in Sec. (2.2), that the optimal values of the derivative of the TC appear in the vicinity of Δξ = ±Δξ0. A similar approach was employed by Diemoz et al.[24

24. P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an x-ray grating interferometer,” Opt. Express 19, 1691–1698 (2011) [CrossRef] [PubMed] .

] for the case of grating interferometry. In particular, rather than phase stepping the detector grating, the pre-sample and detector gratings were offset by an amount which maximised the change in pixel signal for a given change to offset between the gratings. In their method, they make use of a principal analogous to the TC but, instead of separating phase and absorption, obtain a single image containing a well described combination of phase and absorption information. This technique carries the advantage that a single acquisition is acquired per angle of rotation when performing CT imaging.

Fig. 5 Plot of the derivative of the fitted ITCξ), the vertical lines denote Δξ = ±8μm.

The accuracy of Eq. (23) depends significantly on how constant the derivative of the TC is in the vicinity of Δξ0. Figure 6 shows how this derivative varies over a region which corresponds to a sample refraction angle satisfying |ϕx| < 10−5rad/m. The derivative of the rocking curve varies by at most approximately 15% over this range. Whilst this maximum error would be experienced only for extreme refraction angles, as explained in Sec. (2.3), it is possible to derive a refraction angle inversion formula without making such an assumption. Figure 7 shows a plot of IΔ/IΣ versus Δξs for the experimental system being considered, using the fitted TC plotted in Fig. 2. The plot shows that, in order to uniquely determine Δξs from an observed value of ÎΔΣ, it is necessary to restrict the domain of Δξs to |Δξs| ≲ 0.011 mm which implicitly restricts the observable refraction angles to the range |ϕ̄x| ≲ 0.034 rad/mm. The refraction angle may be extracted from an observed value of ÎΔΣ by numerically determining, from Fig. 7, the value of Δξs to which it corresponds. This numerical inversion may be performed very accurately and efficiently.

Fig. 6 Plot of ITCξ0 − Δξs) with Δξ0 = 8μm for a range of Δξs corresponding to |∂ϕ̄/∂x| < 10−5rad/m.
Fig. 7 Plot of IΔ/IΣ versus Δξs. The domain of the plot is restricted to the non-shaded region so that there is a one-to-one mapping from Δξs to IΔ/IΣ. The plot was calculated using the fitted TC in Fig. 2.

4. Examples and analysis

In this section we analyse the TC refraction angle retrieval method and show some examples of how it may be applied. In particular, we show how the TC method differs from that previously published for the experimental system considered in Sec. (3) using some experimentally acquired images. We also perform some numerical simulations to demonstrate the improvement offered by the TC method.

We begin by comparing the TC refraction angle extraction algorithm with that published previously, derived analytically and based upon the assumption of perfectly absorbing apertures. In particular, we have previously published [15

15. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Phase and absorption retrieval using incoherent x-ray sources,” Proc. Natl. Acad. Sci. USA 109, 13922–13927 (2012) [CrossRef] [PubMed] .

] a phase extraction algorithm which extracts the refraction angle according to the formula:
ϕ¯x=erf1(I^ΔI^Σ)σdzod
(25)
π2I^ΔI^Σσdzod
(26)
where erf is the standard error function erf(z)=2/π0zexp(t2)dt and σd=FWHMd/(2log(2)) and FWHMd is the full width at half maximum of the beams projected onto the detector aperture. Figure 8 shows a comparison between the analytic inversion formulae and those obtained using the TC plotted in Fig. 7. All of the plots relate to the system used to obtain the TC plotted in Fig. 2. σd was found to be approximately 12.5μm by determining the focal spot size which gave the best match between calculated and measured TCs. The TC approach should, in general, predict a refraction angle greater or equal in magnitude compared with the analytic method, since the TC method implicitly takes account of the partial absorption by the apertures.

Fig. 8 Comparison of the inversion formulae. TC-approx is described by Eq. (23), TC-full by inverting Eq. (21) using the experimental TC in Fig. 2, Analytic-approx by Eq. (26) and Analytic-full by Eq. (25), TC (ideal)-full is found using an ideal TC, calculated assuming fully absorbing apertures.

The refraction angle retrieval curves in Fig. 8 reveal that the full and approximate TC methods are equivalent for small refraction angles yet diverge strongly for large refraction angles. This will lead to the approximate TC method underestimating refraction angles. This phenomenon results from the partial transmission of X-rays through the apertures as is demonstrated by the remaining plots in Fig. 8. In particular, the full analytic method (Eq. (25)) and the ideal TC methods coincide. The ideal TC inversion curve was obtained from a simulated TC for the case of perfectly absorbing apertures. The approximate analytic curve (Eq. (26)) departs only slightly from the full analytic curve.

We now present an example showing how three of the refraction angle inversion relationships in Fig. 8 result in different values of refraction angle in an experimental case. We used the experimental system for which the TC in Fig. 2 was obtained, to image a titanium wire of radius 125μm. Images were taken at a total of 16 sub-pixel object positions in order to increase spatial resolution. Each such image had an 8.75 second exposure time. We selected this sample due to its very high value of δ throughout the source spectrum, even though it is also highly absorbing. High values of δ causes high values of |ÎΔΣ| to be measured which is where the greatest difference between the three inversion formulae occurs. We have also shown, for reference, the theoretical refraction angle at the estimated mean energy of the spectrum (18keV). We do not expect the experimentally acquired results to match this profile due to beam hardening which results from absorption by the wire [25

25. M. Chabior, T. Donath, C. David, O. Bunk, M. Schuster, C. Schroer, and F. Pfeiffer, “Beam hardening effects in grating-based x-ray phase-contrast imaging,” Med. Phys. 38, 1189–1195 (2011) [CrossRef] [PubMed] .

]. The main point of these plots is to show an experimental case where the full TC inversion formula is required for accuracy in order to account for large refraction angles. This difference arises because the linearity of the TC breaks down for large separations between the pre-sample and detector apertures. We also imaged aluminium, sapphire, polyetheretherketone and boron wires but these samples did not result in a difference between the full and approximate TC refraction angles. The expected refraction angle in Fig. 9 can only be calculated with accurate knowledge of the source and detector spectral responses and the sample absorption [25

25. M. Chabior, T. Donath, C. David, O. Bunk, M. Schuster, C. Schroer, and F. Pfeiffer, “Beam hardening effects in grating-based x-ray phase-contrast imaging,” Med. Phys. 38, 1189–1195 (2011) [CrossRef] [PubMed] .

]. In light of this and in order to further demonstrate the advantages of the TC inversion method we now present some numerical simulations.

Fig. 9 (left) an image of IΔ/IΣ for a 125μm titanium wire, (top right) line profiles of ϕ̄x through the titanium wire and (bottom right) a zoomed in version of the top right plots. Line profiles were calculated using the three inversion formula indicated by the legend. Ē is the estimated mean energy of the spectrum and ϕx at Ē is the theoretical value of ϕx at the mean energy.

Fig. 10 Plots of the source spectrum (left) and refractive index of PEEK (right) as employed in the simulation.

We have performed simulations for both the polychromatic and monochromatic cases, however, assuming a 60μm FWHM focal spot in both cases. The monochromatic simulation serves to demonstrate that the TC technique performs very accurate quantitative phase retrieval in a scenario when phase is well defined. When a polychromatic source is employed, the concept of phase is ambiguous since it implies an averaging of phase. Furthermore, any such measured phase will depend on the source, aperture, detector and sample spectral properties [25

25. M. Chabior, T. Donath, C. David, O. Bunk, M. Schuster, C. Schroer, and F. Pfeiffer, “Beam hardening effects in grating-based x-ray phase-contrast imaging,” Med. Phys. 38, 1189–1195 (2011) [CrossRef] [PubMed] .

].

Figure 11(a) shows the calculated TC for the polychromatic case for aperture gold thicknesses of 10, 20, 40 and 80μm respectively. We note that no significant variation in calculated TC or retrieved refraction angle was observed for aperture gold thicknesses exceeding 80μm. Figure 11(b) shows the calculated TC for the monochromatic case at the spectrum mean energy of 18keV. The two simulations assume the same source power density. Figure 11(c) shows the refraction angle retrieval relationship, Eq. (21), corresponding to the polychromatic TCs of Fig. 11a) and Fig. 11(d) shows the retrieval relationship corresponding to the monochromatic TCs of Fig. 11b). The two cases are similar, but not identical. In particular, we note that in the poly-chromatic case, thicker gold is required to achieve the ideal refraction angle relationship where there is insignificant transmission through the apertures.

Fig. 11 Plots of the translation and inversion curves for the four aperture thicknesses employed which are indicated in the legend. Plots a) and c) correspond to the polychromatic case whilst plots b) and d) correspond to the monochromatic case at the spectrum mean energy of 18keV.

We first explain how the numerical results are analysed. As previously discussed, the system results in a measurement of refraction angle, ϕ̄x. In this example we consider the object to have a circular cross section and so within the support of the object, ϕ̄ is given by ϕ¯=2δ¯R2x2 where, as usual, overlines are used to denote quantities averaged over the spectrum. Spatial variation is thus contained within the function which describes the shape of the object. In what follows, we plot the integrated refraction angle, ϕ̄ = ∫ ϕ̄xdx which enables δ̄ to be determined since the shape function is known. The integrated refraction angle can be used to obtain an measurement of integrated phase through multiplication by an effective wavenumber.

Figure 12 shows the integrated refraction angle for the polychromatic ((a) and (c)) and monochromatic ((b) and (d)) cases and using the TC refraction angle retrieval ((a) and (b)) and the approximate linear method Eq. (26) ((c) and (d)). The integrated refraction angle, calculated analytically, at the spectrum mean energy has also been plotted. We note that only in case b) do all integrated refraction angles match the analytic profile. Agreement should be expected in this case since it has not been necessary to assume the existence of an effective energy for the spectrum in order to obtain Eq. (10). In both c) and d), the approximate linear method results in a large variation between measured integrated refraction angle for the different aperture thicknesses, although we note that in the monochromatic case, the solution is very close to the analytic case for the 80μm thick apertures. Both the approximate linear cases exhibit wide variation simply because the method doesn’t take into account X-ray transmission through the apertures. The polychromatic case returns a generally greater phase change than the monochromatic simply because the average value of δ across the spectrum is greater than δ at the mean energy of the spectrum. The plots in Fig. 12(a), however, demonstrate the main result of this paper. Here we see a much smaller variation in the integrated refraction angle as aperture thickness varies. This shows that the TC method reduces the impact of the system’s spectral properties on the measured refraction angle. In order to more thoroughly analyse the results plotted in a) we present the values of δ̄ measured for each aperture thickness in Table 1. We expect the predicted value of δ̄ to converge towards the value of δ averaged over the spectrum, ie, δave = ∫δ(E)σ(E)dE = 1.16×10−6. Whilst this value is within the range of predicted values of δ̄, the values do not converge to this value. This error arises due to the approximation that the system has an effective energy made in order to simplify Eq. (9) to Eq. (10).

Fig. 12 Plots of integrated refraction angle across the simulated PEEK fiber of 100μm radius. Plot a) and c) correspond to the polychromatic case whilst plots b) and d) correspond to the monochromatic case at the spectrum mean energy of 18keV. Plots a) and b) have been obtained using the TC inversion technique whilst plots c) and d) have been obtained using the linear approximation. The analytic sample phase at the spectrum mean energy has been plotted on all axes for reference.

Table 1. Numerically determined values of δ̄ for four different values of aperture gold thickness. Ēδ̄ is the photon energy at which PEEK has δ = δ̄.

table-icon
View This Table

5. Conclusions

We have demonstrated a new quantitative technique for measuring X-ray refraction angle and absorption using the coded aperture method with a laboratory source. The quantitative technique requires knowledge of only the so-called translation curve which is obtained by recording the detector signal as the pre-sample aperture is translated relative to the detector aperture. The translation curve is measured routinely when the system is aligned.

We showed that the new technique is more accurate for measuring larger refraction angles and also is more accurate in the case where the apertures are only partially absorbing. We compared the new technique with that previously used which assumed that the apertures were perfectly absorbing. We showed that the new method is significantly more accurate in the case of partially absorbing apertures. Finally, we showed that the proposed technique measures an average value of refraction angle based upon the sample’s value of δ averaged over the spectrum, rather than, for example, measuring δ at the average energy of the spectrum.

Acknowledgments

This work was funded by the UK Engineering and Physical Sciences Research Council ( EP/G004250/1 and EP/I021884/1). P.M. is currently supported by a Discovery Early Career Research Award from the Australian Research Council ( DE120101331).

References and links

1.

A. Olivo and R. Speller, “A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources,” Appl. Phys. Lett. 91, 074106 (2007) [CrossRef] .

2.

T. Davis, D. Gao, T. Gureyev, A. Stevenson, and S. Wilkins, “Phase-contrast imaging of weakly absorbing materials using hard x-rays,” Nature 373, 595–598 (1995) [CrossRef] .

3.

D. Chapman, W. Thomlinson, F. Arfelli, N. Gmür, Z. Zhong, R. Menk, R. E. Johnson, D. Washburn, E. Pisano, and D. Sayers, “Mammography imaging studies using a laue crystal analyzer,” The 9th National Conference on Synchrotron Radiation Instrumentation 67, 3360–3360 (1996).

4.

D. Chapman, W. Thomlinson, R. Johnston, D. Washburn, E. Pisano, N. Gmür, Z. Zhong, R. Menk, F. Arfelli, and D. Sayers, “Diffraction enhanced x-ray imaging,” Phys. Med. Biol. 42, 2015–2025 (1997) [CrossRef] [PubMed] .

5.

D. Paganin, T. E. Gureyev, K. M. Pavlov, R. A. Lewis, and M. Kitchen, “Phase retrieval using coherent imaging systems with linear transfer functions,” Opt. Commun. 234, 87–105 (2004) [CrossRef] .

6.

Y. I. Nesterets, T. Gureyev, D. Paganin, K. Pavlov, and S. W. Wilkins, “Quantitative diffraction-enhanced x-ray imaging of weak objects,” J. Phys. D Appl. Phys. 37, 1262–1274 (2004) [CrossRef] .

7.

M. J. Kitchen, D. M. Paganin, K. Uesugi, B. J. Allison, R. A. Lewis, S. B. Hooper, and K. M. Pavlov, “X-ray phase, absorption and scatter retrieval using two or more phase contrast images,” Opt. Express 18, 19994–20012 (2010) [CrossRef] [PubMed] .

8.

P. C. Diemoz, P. Coan, C. Glaser, and A. Bravin, “Absorption, refraction and scattering in analyzer-based imaging: comparison of different algorithms,” Opt. Express 18, 3494–3509 (2010) [CrossRef] [PubMed] .

9.

D. J. Vine, D. M. Paganin, K. M. Pavlov, J. Kraeusslich, O. Wehrhan, I. Uschmann, and E. Foerster, “Analyzer-based phase contrast imaging and phase retrieval using a rotating anode x-ray source,” Appl. Phys. Lett. 91, 254110 (2007) [CrossRef] .

10.

C. Kottler, F. Pfeiffer, O. Bunk, C. Gruenzweig, J. Bruder, R. Kaufmann, L. Tlustos, H. Walt, I. Briod, T. Weitkamp, and C. David, “Phase contrast x-ray imaging of large samples using an incoherent laboratory source,” Phys. Status Solidi. A 204, 2728–2733 (2007) [CrossRef] .

11.

F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance x-ray sources,” Nat. Phys. 2, 258–261 (2006) [CrossRef] .

12.

A. Momose, W. Yashiro, H. Kuwabara, and K. Kawabata, “Grating-based x-ray phase imaging using multiline x-ray source,” Jpn. J. Appl. Phys. 48, 076512 (2009) [CrossRef] .

13.

A. Momose, “Recent advances in x-ray phase imaging,” Jpn. J. Appl. Phys. 44, 6355–6367 (2005) [CrossRef] .

14.

A. Olivo, F. Arfelli, G. Cantatore, R. Longo, R. Menk, S. Pani, M. Prest, P. Poropat, L. Rigon, G. Tromba, E. Vallazza, and E. Castelli, “An innovative digital imaging set-up allowing a low-dose approach to phase contrast applications in the medical field,” Med. Phys. 28, 1610–1619 (2001) [CrossRef] [PubMed] .

15.

P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Phase and absorption retrieval using incoherent x-ray sources,” Proc. Natl. Acad. Sci. USA 109, 13922–13927 (2012) [CrossRef] [PubMed] .

16.

P. R. Munro, L. Rigon, K. Ignatyev, F. C. Lopez, D. Dreossi, R. D. Speller, and A. Olivo, “A quantitative, non-interferometric x-ray phase contrast imaging technique,” Opt. Express 21, 647–661 (2013) [CrossRef] [PubMed] .

17.

P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Source size and temporal coherence requirements of coded aperture type x-ray phase contrast imaging systems,” Opt. Express 18, 19681–19692 (2010) [CrossRef] [PubMed] .

18.

M. Marenzana, C. K. Hagen, P. D. N. Borges, M. Endrizzi, M. B. Szafraniec, K. Ignatyev, and A. Olivo, “Visualization of small lesions in rat cartilage by means of laboratory-based x-ray phase contrast imaging,” Phys. Med. Biol. 57, 8173–8184 (2012) [CrossRef] [PubMed] .

19.

S. Mayo, P. Miller, S. Wilkins, T. Davis, D. Gao, T. Gureyev, D. Paganin, D. Parry, A. Pogany, and A. Stevenson, “Quantitative x-ray projection microscopy: phase-contrast and multi-spectral imaging,” Journal of Microscopy-Oxford 207, 79–96 (2002) [CrossRef] .

20.

T. Gureyev and S. Wilkins, “On x-ray phase imaging with a point source,” J. Opt. Soc. Am. A 15, 579–585 (1998) [CrossRef] .

21.

D. Paganin, Coherent X-ray optics, Oxford series on synchrotron radiation (Oxford University Press, Great Clarendon Street, Oxford OX2 6DP, 2006) [CrossRef] .

22.

P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “The relationship between wave and geometrical optics models of coded aperture type x-ray phase contrast imaging systems,” Opt. Express 18, 4103–4117 (2010) [CrossRef] [PubMed] .

23.

A. Olivo and R. Speller, “Experimental validation of a simple model capable of predicting the phase contrast imaging capabilities of any x-ray imaging system,” Phys. Med. Biol. 51, 3015–3030 (2006) [CrossRef] [PubMed] .

24.

P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an x-ray grating interferometer,” Opt. Express 19, 1691–1698 (2011) [CrossRef] [PubMed] .

25.

M. Chabior, T. Donath, C. David, O. Bunk, M. Schuster, C. Schroer, and F. Pfeiffer, “Beam hardening effects in grating-based x-ray phase-contrast imaging,” Med. Phys. 38, 1189–1195 (2011) [CrossRef] [PubMed] .

26.

B. Henke, E. Gullikson, and J. Davis, “X-ray interactions: Photoabsorption, scattering, transmission, and reflection at e = 50–30,000 ev, z = 1–92,” Atom. Data Nucl. Data 54, 181–342 (1993) [CrossRef] .

OCIS Codes
(110.7440) Imaging systems : X-ray imaging
(120.5050) Instrumentation, measurement, and metrology : Phase measurement

ToC Category:
X-ray Optics

History
Original Manuscript: December 7, 2012
Revised Manuscript: March 11, 2013
Manuscript Accepted: March 16, 2013
Published: April 30, 2013

Citation
Peter R.T. Munro, Charlotte K. Hagen, Magdalena B. Szafraniec, and Alessandro Olivo, "A simplified approach to quantitative coded aperture X-ray phase imaging," Opt. Express 21, 11187-11201 (2013)
http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-21-9-11187


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. A. Olivo and R. Speller, “A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources,” Appl. Phys. Lett.91, 074106 (2007). [CrossRef]
  2. T. Davis, D. Gao, T. Gureyev, A. Stevenson, and S. Wilkins, “Phase-contrast imaging of weakly absorbing materials using hard x-rays,” Nature373, 595–598 (1995). [CrossRef]
  3. D. Chapman, W. Thomlinson, F. Arfelli, N. Gmür, Z. Zhong, R. Menk, R. E. Johnson, D. Washburn, E. Pisano, and D. Sayers, “Mammography imaging studies using a laue crystal analyzer,” The 9th National Conference on Synchrotron Radiation Instrumentation67, 3360–3360 (1996).
  4. D. Chapman, W. Thomlinson, R. Johnston, D. Washburn, E. Pisano, N. Gmür, Z. Zhong, R. Menk, F. Arfelli, and D. Sayers, “Diffraction enhanced x-ray imaging,” Phys. Med. Biol.42, 2015–2025 (1997). [CrossRef] [PubMed]
  5. D. Paganin, T. E. Gureyev, K. M. Pavlov, R. A. Lewis, and M. Kitchen, “Phase retrieval using coherent imaging systems with linear transfer functions,” Opt. Commun.234, 87–105 (2004). [CrossRef]
  6. Y. I. Nesterets, T. Gureyev, D. Paganin, K. Pavlov, and S. W. Wilkins, “Quantitative diffraction-enhanced x-ray imaging of weak objects,” J. Phys. D Appl. Phys.37, 1262–1274 (2004). [CrossRef]
  7. M. J. Kitchen, D. M. Paganin, K. Uesugi, B. J. Allison, R. A. Lewis, S. B. Hooper, and K. M. Pavlov, “X-ray phase, absorption and scatter retrieval using two or more phase contrast images,” Opt. Express18, 19994–20012 (2010). [CrossRef] [PubMed]
  8. P. C. Diemoz, P. Coan, C. Glaser, and A. Bravin, “Absorption, refraction and scattering in analyzer-based imaging: comparison of different algorithms,” Opt. Express18, 3494–3509 (2010). [CrossRef] [PubMed]
  9. D. J. Vine, D. M. Paganin, K. M. Pavlov, J. Kraeusslich, O. Wehrhan, I. Uschmann, and E. Foerster, “Analyzer-based phase contrast imaging and phase retrieval using a rotating anode x-ray source,” Appl. Phys. Lett.91, 254110 (2007). [CrossRef]
  10. C. Kottler, F. Pfeiffer, O. Bunk, C. Gruenzweig, J. Bruder, R. Kaufmann, L. Tlustos, H. Walt, I. Briod, T. Weitkamp, and C. David, “Phase contrast x-ray imaging of large samples using an incoherent laboratory source,” Phys. Status Solidi. A204, 2728–2733 (2007). [CrossRef]
  11. F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance x-ray sources,” Nat. Phys.2, 258–261 (2006). [CrossRef]
  12. A. Momose, W. Yashiro, H. Kuwabara, and K. Kawabata, “Grating-based x-ray phase imaging using multiline x-ray source,” Jpn. J. Appl. Phys.48, 076512 (2009). [CrossRef]
  13. A. Momose, “Recent advances in x-ray phase imaging,” Jpn. J. Appl. Phys.44, 6355–6367 (2005). [CrossRef]
  14. A. Olivo, F. Arfelli, G. Cantatore, R. Longo, R. Menk, S. Pani, M. Prest, P. Poropat, L. Rigon, G. Tromba, E. Vallazza, and E. Castelli, “An innovative digital imaging set-up allowing a low-dose approach to phase contrast applications in the medical field,” Med. Phys.28, 1610–1619 (2001). [CrossRef] [PubMed]
  15. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Phase and absorption retrieval using incoherent x-ray sources,” Proc. Natl. Acad. Sci. USA109, 13922–13927 (2012). [CrossRef] [PubMed]
  16. P. R. Munro, L. Rigon, K. Ignatyev, F. C. Lopez, D. Dreossi, R. D. Speller, and A. Olivo, “A quantitative, non-interferometric x-ray phase contrast imaging technique,” Opt. Express21, 647–661 (2013). [CrossRef] [PubMed]
  17. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “Source size and temporal coherence requirements of coded aperture type x-ray phase contrast imaging systems,” Opt. Express18, 19681–19692 (2010). [CrossRef] [PubMed]
  18. M. Marenzana, C. K. Hagen, P. D. N. Borges, M. Endrizzi, M. B. Szafraniec, K. Ignatyev, and A. Olivo, “Visualization of small lesions in rat cartilage by means of laboratory-based x-ray phase contrast imaging,” Phys. Med. Biol.57, 8173–8184 (2012). [CrossRef] [PubMed]
  19. S. Mayo, P. Miller, S. Wilkins, T. Davis, D. Gao, T. Gureyev, D. Paganin, D. Parry, A. Pogany, and A. Stevenson, “Quantitative x-ray projection microscopy: phase-contrast and multi-spectral imaging,” Journal of Microscopy-Oxford207, 79–96 (2002). [CrossRef]
  20. T. Gureyev and S. Wilkins, “On x-ray phase imaging with a point source,” J. Opt. Soc. Am. A15, 579–585 (1998). [CrossRef]
  21. D. Paganin, Coherent X-ray optics, Oxford series on synchrotron radiation (Oxford University Press, Great Clarendon Street, Oxford OX2 6DP, 2006). [CrossRef]
  22. P. Munro, K. Ignatyev, R. Speller, and A. Olivo, “The relationship between wave and geometrical optics models of coded aperture type x-ray phase contrast imaging systems,” Opt. Express18, 4103–4117 (2010). [CrossRef] [PubMed]
  23. A. Olivo and R. Speller, “Experimental validation of a simple model capable of predicting the phase contrast imaging capabilities of any x-ray imaging system,” Phys. Med. Biol.51, 3015–3030 (2006). [CrossRef] [PubMed]
  24. P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an x-ray grating interferometer,” Opt. Express19, 1691–1698 (2011). [CrossRef] [PubMed]
  25. M. Chabior, T. Donath, C. David, O. Bunk, M. Schuster, C. Schroer, and F. Pfeiffer, “Beam hardening effects in grating-based x-ray phase-contrast imaging,” Med. Phys.38, 1189–1195 (2011). [CrossRef] [PubMed]
  26. B. Henke, E. Gullikson, and J. Davis, “X-ray interactions: Photoabsorption, scattering, transmission, and reflection at e = 50–30,000 ev, z = 1–92,” Atom. Data Nucl. Data54, 181–342 (1993). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited