Integrated endoscopy system for simultaneous imaging and spectroscopy for early lung cancer detection

Haishan Zeng, Mirjan Petek, Marjeta Tercelj Zorman, Annette McWilliams, Branko Palcic, and Stephen Lam  »View Author Affiliations


Optics Letters, Vol. 29, Issue 6, pp. 587-589 (2004)
http://dx.doi.org/10.1364/OL.29.000587


View Full Text Article

Acrobat PDF (1472 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

An integrated endoscopy system for simultaneous imaging and spectroscopy was developed to facilitate more accurate and convenient detection of early lung cancers. A specially designed three-CCD camera in combination with a dedicated light source permits capture of both white-light color images and tissue autofluorescence images without the need to switch between two different cameras. A mirror with an optical fiber at its center, placed at an interim imaging plane inside the camera unit, facilitates simultaneous imaging and spectroscopy measurements in either white-light reflectance mode or fluorescence mode. The system has been successfully tested in a clinic, demonstrating a practical approach to improve both diagnostic sensitivity and specificity at the same time.

© 2004 Optical Society of America

OCIS Codes
(110.2990) Imaging systems : Image formation theory
(170.0110) Medical optics and biotechnology : Imaging systems
(170.1610) Medical optics and biotechnology : Clinical applications
(170.2150) Medical optics and biotechnology : Endoscopic imaging
(170.6280) Medical optics and biotechnology : Spectroscopy, fluorescence and luminescence
(170.6510) Medical optics and biotechnology : Spectroscopy, tissue diagnostics

Citation
Haishan Zeng, Mirjan Petek, Marjeta Tercelj Zorman, Annette McWilliams, Branko Palcic, and Stephen Lam, "Integrated endoscopy system for simultaneous imaging and spectroscopy for early lung cancer detection," Opt. Lett. 29, 587-589 (2004)
http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-29-6-587


Sort:  Author  |  Year  |  Journal  |  Reset

References

  1. S. Lam, T. Kennedy, M. Unger, Y. E. Miller, D. Gelmont, V. Rusch, B. Gipe, D. Howard, J. C. LeRiche, A. Coldman, and A. F. Gazder, Chest 113, 696 (1998).
  2. B. Palcic, S. Lam, J. Hung, and C. MacAulay, Chest 99, 742 (1991).
  3. J. Hung, S. Lam, J. C. LeRiche, and B. Palcic, Lasers Surg. Med. 11, 99 (1991).
  4. H. Zeng, A. Weiss, R. Cline, and C. MacAulay, Bioimaging 6, 151 (1998).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.